Stamos S-LS-119 - Electronic test device

S-LS-119 - Electronic test device Stamos - Free user manual and instructions

Find the device manual for free S-LS-119 Stamos in PDF.

📄 432 pages English EN Download 💬 AI Question
Notice Stamos S-LS-119 - page 24
Pick your language and provide your email: we'll send you a specifically translated version.

User questions about S-LS-119 Stamos

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Electronic test device in PDF format for free! Find your manual S-LS-119 - Stamos and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. S-LS-119 by Stamos.

USER MANUAL S-LS-119 Stamos

text_image KORAC DC PAGUARABLE SLEVENS LOB S+ S- 1

1 - Leistung

Stamos S-LS-119 - 1

text_image LINE 190Vac - 110Vac 220Vac - 340Vac FUSE 1.25A 6.5A LAN USB RS232 1 2

1 - Externe

Schaltersteuerung

2 - Externe

Triggersteuerung

This User Manual has been translated for your convenience using machine translation. Reasonable efforts have been made to provide an accurate translation; however, no automated translation is perfect nor is it intended to replace human translators. The official User Manual is the English version. Any discrepancies or differences created in the translation are not binding and have no legal effect for compliance or enforcement purposes. If any questions arise related to the accuracy of the information contained in the User Manual, please refer to the English version of those contents which is the official version.

Specifications

Note: The specifications below are tested under the conditions of temperature 25^ C +5^ C and the warm-up for 20 minutes.

Product name: DC Electronic LoadModel name: S-LS-119
Input RatingPower1500W
Voltage0-150V
Current0-40A
CC ModeRange0-3A0-40A
Resolution0.1mA1mA
Accuracy± (0.05% of set + 0.045% off.s)
CV ModeRange0-50V0-150V
Resolution0.1mV1mV
Accuracy± (0.05% of set+0.025%off.s)
CR ModeRange0.05Ω - 7.5KΩ
Resolution16 bit
Accuracy± (0.05% of set+0.025%off.s)
CW ModeRange1500 W
Resolution0.01W
Accuracy± (0.1% of set+0.1%off.s)
SlopeRange0-3A0-40A
Rising0.0001-0.3A/μs0.001-2A/μs
Falling0.0001-0.3A/μs0.001-2A/μs
Voltage measurementRange0-18V0-150V
Resolution0.1mV1mV
Accuracy± (0.03% of set+0.025% off.s)
Current measurementRange0-3A0-40A
Resolution0.1mA1mA
Accuracy± (0.05% of set+0.045% off.s)
Power measurementRange1500W
Resolution0.01W
Accuracy± (0.1% of set+0.1% off.s)
Over power protection1550W
Over current protection42A
Over voltage protection155V
Over temperature protection85°C
Input impedance150KΩ
Weight14.85 kg
Dimensions (W*D*H)47 x 38 x 15 cm

1. Main interface

Stamos S-LS-119 - Main interface - 1

line | Parameter | Value | | --------------- | --------- | | OUTPUT | 15.979V | | OCP/OPP | 00.000A | | MODE | 00.000W | | CHOOSE MODE? | - |

1.1. Menu keys

STATIC(F1)Enter static CC, CV, CW, CR, or SHORT
OCP/OPP(F2)call tables OCP and OPP
BATTERY(F3)Battery measurement mode: VI curve
USB(F4)The panel will display the USB logo after inserting the U disk, and it will display blank when it is not inserted. Save data to U disk, import and export LIST to U disk.
DYNAMIC(F5)Dynamic CC, CV, CW, CR, RL, or PL
CONFIG(F6)Parameter settings

1.2. Keyboard function

ESCkey for escape and delete
F1 to F6Function keys
TABChange focus
STOP1. Pause during waveform display
2. Long press to take a screenshot
LOCKLong press to lock the keyboard, and long press again to unlock it
CC1. Short press to enter CC mode
2. Long press SHORT
CVEnter CV mode
CWEnter CW mode
CREnter CR mode
0 to 91. Short press the number keys
2. Long press numeric 1 to 6 to call static storage units 1 to 6
3. Long press 7 to set RTC, and long press again to save and exit
4. Long press 8 set LCD backlight, and long press again to save and exit
5. Long press 9 to enable the USB storage, and long press again to turn off the storage
6. In the main interface, long press numeric 0 for more than 3s to calibrate 0
WAVEPress TAB to select voltage or current, and press WAVE to display 4 measurement lines, among which 2 vertical lines measure time, and 2 horizontal lines measure voltage and current amplitude respectively.
1. Short press6 (left key) to select the vertical line on the left
2. Short press4 (right key) to select the vertical line on the right, and long press after STOP to move the waveform to the right
3. Short press (up key) to select the measured voltage amplitude, and long press to modify the voltage scale value
4. Short press (down key) to select the measured current amplitude, and long press to modify the current scale value
5. Enter to adjust the sampling time value
ONTurn on and turn off
Enter1. Short press to enter
2. Long press to save
Note:The long press time is about 1.5s

2. Static mode

2.1. Static CC

In constant current mode, the load consumes a constant current regardless of whether the input voltage is changed.

Operations:

1) Press CC or the soft key F1 on the keyboard.
2) Enter a value from 0 to 9 on the numeric keypad.
3) Press 6 and 4 to move the cursor, and press T and knob to adjust the corresponding value.
4) Press ENTER to turn ON/OFF.

2.2. Static CV

In constant voltage mode, the load keeps the device under test at the set voltage regardless of whether the input current is changed.

Operations: same as above.

2.3. Static CW

In constant voltage mode, the load keeps the device under test at the set power regardless of whether the input voltage and current are changed. Operations: same as above.

2.4. Static CR

In constant resistance mode, the load keeps the device under test at the set resistance regardless of whether the input voltage and current are changed. Operations: same as above

2.5. SHORT function

In SHORT mode, the load is output at the maximum current. Operations:

Long press CC to displays MODE: SHORT on the interface and press CC, CV, CW or CW to exit.

2.6. Call static storage

The load can store and call 100 groups of static set values.

1) Storage operation

(1) Press CAL/SAVE to switch the state to SAVE.
(2) Enter a number key to index a line in the list, press TAB to select it, and press ENTER to enter the edit mode. Edit to display a red background, and press6 and4 to select.
(3) Edit MODE. Press CC, CV, CW, or CR on the keyboard to modify.
(4) Edit data. Press 0 to 9 and ESC on the keyboard to modify.
(5) After modification, press ENTER, and long press ENTER again to save the data.
2) Call operation
(1) Press CALL/SAVE to switch to CALL.
(2) Enter a number key to index a line or press TAB to switch the list, select it with the knob, and press ENTER.
3) M1 to M6 quick call

Long press numeric 1 to 6 to call M1 to M6.

Stamos S-LS-119 - Call static storage - 1

other CALL OUTPUT: 15.978V 00.000A 00.000W CALL 12.200Ω NUM MODE Nepal 3 CW 10.400 4 CR 10.600 5(CC 10.811 6 CV 11.000 7 CW 11.200 8 CR 11.400 9(CC 11.600 10 CV 11.800 11 CW 12.000 12 CR 12.200 CC CV CW CR CALL SAVE HOME(1)

3. Dynamic Test Mode

In dynamic mode, the load is constantly switching between 2 different values A and B.

3.1. Dynamic CC

For output using two different currents with different duty cycles at a certain frequency.

Take the current slope A of 0.001A/ S, the current change slope B of 0.002A/ S, the current value A of 1A, the current value B of 2A, the cycle frequency of 1HZ, and the duty cycle of 40% as an example.

Operations:

Select (F5) Dynamic to enter the interface as shown in Fig. 3.1.

1) Press (F1) DYN/CALL, press TAB to select CALLDYN's drop-down box, and turn the knob to CC.
2) Press (F2) DYN/SAVE to display SAVE on the screen.
3) Press TAB to switch the focus. The knob selects the first row.
4) Press ENTER to select the red background.
5) Enter numeric 0 to 9, and ESC.
6) After editing as 0.001, press ENTER. Then 0.0010 will be displayed on the screen, and the background will be switched to blue.
7) Repeat the above steps to set 0.002A/ S, 1A, 2A, 1HZ, and 40% respectively.
8) Long press ENTER to save the edited data.
9) Press the soft key (F1) DYN/CALL to display CALL.
10) Press ENTER to call.
11) Turn ON/OFF.

Stamos S-LS-119 - Dynamic CC - 1

text_image SAVE MODE: Dynamic NO. DATA UNITS EXPLAIN 1 0.0010 A/uS A SLOPE 2 0.0010 A/uS B SLOPE 3 1.0000 A LEVEL A 4 1.0000 A LEVEL B 5 1.0000 Hz FREQUENCY 6 40 % DUTY DYN CALL DYN LIST LIST SAVE DYN: CC HOME

Figure 3.1 DYN CC

Note:

Set the slope and current value to be range dependent. The maximum slope of the small range is 0.24A/ S , the maximum current can be set to 3A, the maximum slope of the large range is 3.2A/ S , and the maximum current can be set to 40A.

The maximum frequency can be set to 40,000Hz. When the frequency is set to 40,000Hz, the maximum duty cycle is 50%.

Refer to Section "Load maximum setting" for the current maximum setting.

3.2. Dynamic CV

Used for output with different duty cycles at two different voltages at a certain frequency.

Take the A voltage of 1V, the B voltage of 2V, the cycle frequency of 1HZ, and the duty cycle of 40% as an example.

Operations:

1) Press (F1) DYN/CALL, TAB to select DYN/CALL, and turn the knob to CV.
2) Press (F2) DYN/SAVE to display SAVE on the screen.
3) Press TAB to switch the focus and turn the knob to select the first row.
4) Press ENTER to select the red background.
5) Enter numeric 0 to 9, and ESC.
6) After editing as 1.0001, press ENTER. Then 1.0000 will be displayed on the screen, and the background will be switched to blue.
7) Repeat the above steps to set 2V, 1HZ, and 40% respectively.
8) Long press ENTER to save the edited data.
9) Press (F1) DYN/CALL to display CALL.
10) Press ENTER to call DYN CV.
11) Turn ON/OFF.

3.3. Dynamic CW

The operation is the same as above.

3.4. Dynamic CR

The operation is the same as above.

3.5. Dynamic PL

It is set to the value A at the beginning. Each time a trigger signal is received, the load is switched to the value B, and switched to the value A again after the set time is maintained.

The following takes the current slope A of 0.001A/ S, the current slope B of 0.002A/ S, the current value A of 1A, the current value B of 2A, and the duration B of 1s as an example.

EN

Operations:

1) Press (F1) DYN/CALL, press TAB to select CALLYN, and turn the knob to PL.
2) Press (F2) DYN/SAVE to display SAVE on the screen.
3) Press TAB to switch the focus and turn the knob to select the first row.
4) Press ENTER to select the red background.
5) Enter numeric 0 to 9, and ESC.
6) After editing as 0.001, press ENTER. Then 0.0010 will be displayed on the screen, and the background will be switched to blue.
7) Repeat the above steps to set 0.002A/ S, 1A, 2A, and 1S respectively.
8) Long press ENTER to save the edited data.
9) Press the soft key (F1) DYN/CALL to display CALL.
10) Press ENTER to call DYN RL.
11) Turn ON/OFF.
12) Trigger the value B every time you press ENTER.

3.6. Dynamic RL

Each time a trigger signal is received, the load is switched back and forth between the value A and the value B.

Take the current slope A of 0.001A/ S, the current slope B of 0.002A/ S, the current value A of 1A, and the current value B of 2A as an example.

Operations:

1) Press (F1) DYN/CALL, press TAB to select DYN/CALL, and turn the knob to RL.
2) Press (F2) DYN/SAVE to display SAVE on the screen.
3) Press TAB to switch the focus and turn the knob to select the first row.
4) Press ENTER to select the red background.
5) Enter numeric 0 to 9, and ESC.
6) After editing as 0.001, press ENTER. Then 0.0010 will be displayed on the screen, and the background will be switched to blue.
7) Repeat the above steps to set 0.002A/ S, 1A, and 2A respectively.
8) Long press ENTER to save the edited data.
9) Press the soft key (F1) DYN/CALL to display CALL.
10) Press ENTER to call DYN RL.
11) Turn ON/OFF.

12) Trigger the value B every time you press ENTER.

4. Sequential Operation Mode

A maximum of 7 groups can be stored, each group can set up to 84 dynamically changing currents, and then the set currents can be switched in sequence. Take the settings stored in Group 1, the maximum current of 3A, and the number of the dynamic change current of 3; the first dynamic current of 1A, the change rate of 0.001A/ S, and the duration of 1s; the second dynamic current of 2A, the change rate of 0.002A/ S, and the duration of 2S; the third dynamic current of 3A, the change rate of 0.003A/ S, and the duration of 3s; the number of repeated operations of 5 as an example. As shown in Fig. 4.1 List. Operations:

1) Press (F3) LIST/CALL press TAB to select GROUP and turn the knob to Group 1.
2) Press (F4) LIST/SAVE to display SAVE on LIST.
3) Press TAB to switch the focus, select RANGE and edit the maximum value of 3A via numeric 0 to 9 and ESC.
4) Press TAB to select CYCLE. The number of editing cycles is 5.
5) Press TAB to the first row of LIST and press ENTER. At this time, the background turns red.

Edit the first item DATA value of 1A. After editing, press ENTER, and the background will be switched to blue.

Press left and right keys (E and 4) to the second item SLOPE (A/ μS), edit the value as 0.001A/ μS, and press ENTER, and the background will be switched to blue. Press left and right keys (6 and 4) to the third item TIME (S), edit the value as 1s, and press ENTER.

6) Repeat the above steps to start setting the second and third rows of the table.
7) Long press ENTER to save the edited data.
8) Press the soft key (F3) LIST/CALL to display CALL.
9) Press ENTER to call LIST1.

10) Turn ON/OFF.

Note:

To delete data after Row 3, select Row 4 in LIST SAVE mode.

Press ENTER to enter the editing background, and it turns red, and then press ESC to delete all data after Row 4. Long press ENTER to save the data.

Stamos S-LS-119 - Sequential Operation Mode - 1

text_image Cycle: 5 SAVE MODE: LIST Range(A): 3 NO. DATA(A) SLOPE(A/uS) TIME(S) 1 1.0000 0.0010 1.0000 2 2.0000 0.0020 2.0000 3 3.0000 0.0030 3.0000 4 0.1000 0.2400 2.3000 5 0.1000 0.2400 2.4000 6 0.1000 0.2400 2.5000 7 0.1000 0.2400 2.6000 8 0.1000 0.2400 2.7000 9 10 OUTPUT: 15.978V 00.000A 00.000W SAVE Group: 1 DYN CALL DYN SAVE LIST CALL LIST SAVE HOME

Figure 4.1 List

5. Battery Test Mode

A maximum of 7 groups of battery test parameters can be set, the battery is tested according to the set current, voltage, capacity, time, and the test is automatically turned off if one of the conditions is met.

5.1. Battery test setup

Operation instructions: it takes the settings stored in Group 1, the current range of 10A, the discharge current of 1A, the discharge cut-off voltage of 2V, the discharge cut-off capacity of 0.5AH, and the discharge duration of 200m (battery time unit: m) as an example.

1) Press (F3) Battery on the main interface to enter battery measurement.
2) Press (F1) BATT/CALL, and press TAB to select CALL GROUP 1.
3) Press (F2) BATT/SAVE to display SAVE on the table.
4) Press TAB to switch the focus and turn the knob to select the row position that needs to be modified.
5) Press ENTER to select the red background.
6) Enter numeric 0 to 9, and ESC.
7) After editing the range of 10A, press ENTER. Then 10.000 will be displayed on the screen, and the background will be switched to blue.
8) Repeat the above steps to set 1A, 2V, 0.5AH, and 200M respectively.
9) Long press ENTER to save the edited data.
10) Press (F1) BATT/CALL.
11) Press ENTER to call.
12) Turn ON/OFF.

6. VI Curve Mode

A maximum of 7 groups of VI test parameters can be set according to the set maximum current, minimum current, and step value.

VI Curve

Stamos S-LS-119 - VI Curve - 1

text_image SAVE MODE: VI-CURVE NO. DATA UNITS EXPLAIN 1 0.2000 A START CURR 2 5.0000 A END CURR 3 0.2000 A STEP CURR 4 1.0000 S STEP TIME Battery Call Battery Save Curve Call Curve Save OUTPUT: 15.978V 00.000A 00.000W SAVE Group: 1 HOME

Figure 6.1 VI Current Parameters

6.1. VI test setup

Operation instructions: it takes the settings stored in Group 1, the start current of 0.2A, the end current of 5A, the step current of 0.2A, and the step duration of 1s as an example.

1) Press (F3) Battery on the main interface to enter VI measurement.
2) Press (F3) Curve/CALL, and press TAB to select CALL GROUP 1
3) Press (F4) Curve/SAVE to display SAVE on the table.
4) Press TAB to switch to the table focus and turn the knob to select the row position to be modified.
5) Press ENTER to select the red background.
6) Press ENTER to select the red background.
7) After editing as 0.2, press ENTER. Then 0.2000 will be displayed on the screen, and the background will be switched to blue.
8) Repeat the above steps to set 5A, 2A, and 1.000s respectively.
9) Long press ENTER to save the edited data.
10) Press (F3) Curve/CALL.
11) Press ENTER to call.
12) Turn ON/OFF.

The operation effect is shown in Fig. 6.2.

Stamos S-LS-119 - VI test setup - 1

line | Current(A) | Voltage(V) | Power(W) | | ---------- | ---------- | -------- | | 0.200 | 0.000 | 0.000 | | 1.833 | 14.38 | 23.97 | | 1.867 | 19.17 | 47.93 | | 2.700 | 23.97 | 71.90 | | 3.533 | 27.97 | 95.87 | | 4.367 | 31.97 | 119.8 | | 5.200 | 35.97 | 143.8 |

Figure 6.2 VI Running Interface

7. OCP Mode

When the voltage reaches the VON value, the current output will be delayed for a period of time, and the step value will be decreased once every other period of time until the cut-off current or the voltage is higher than the OCP set voltage. If the stopped voltage is higher than the OCP voltage and the current is between the set maximum and minimum value, it is PASS, otherwise it is FAULT.

7.1. Setting function of OCP test

Note: 7 groups of OCP test parameters can be set at most.

Operation instructions: it takes the setting stored in Group 1, VON voltage of 10V, VON voltage delay of 5S, and current range of 3A; start current of 2A, each decrease of 0.1A, and each decreasing duration of 1s; end current of 1A, OCP voltage of 8V, maximum current of 1.9A, and minimum current of 1.1A as an example.

Stamos S-LS-119 - Setting function of OCP test - 1

text_image SAVE MODE: OCP NO. DATA UNITS EXPLAIN 1 10.000 V VON LEVEL 2 5.0000 S VON DELAY 3 3.0000 A RANGE 4 2.0000 A START 5 0.1000 A STEP 6 1.0000 S STEP DELAY 7 1.0000 A END 8 8.0000 V OCP VOLT 9 1.9000 A MAX TRIP 10 1.0000 A MIN TRIP OUTPUT: 15.978V 00.000A 00.000W SAVE Group: 1 OCP CALL OCP SAVE OPP CALL OPP SAVE HOME

Figure 7.1 OCP

1) Press (F2) OCP/OPP on home page.

2) Press (F1) OCP/CALL and TAB to select CALL GROUP 1.

3) Press (F2) OCP/SAVE to display SAVE on the table.

4) Press TAB to switch the focus and turn the knob to select the first row.

5) Press ENTER to select red background. Enter numeric 0 to 9, and ESC.

6) After editing VON as 10V, press ENTER. Then 10.000 will be displayed on the screen, and the background will be switched to blue.

7) Repeat the above steps to set 5S, 3A, 2A, 0.1A, 1S, 1A, 8V, 1.9A, and 1.1A respectively.
8) Long press ENTER to save the edited data.
9) Press (F1) OCP/CALL.
10) Press ENTER to call.
11) Turn ON/OFF.

8. OPP Mode

When voltage reaches the VON value, the power output should be delayed for a period of time, and the step value should be decreased every once in a while until cut-off power or voltage is higher than the OPP set voltage. After the delay and decrease stop, if the voltage is higher than OPP voltage, the power between the set maximum and minimum values means PASS, or it means FAULT.

8.1. Setting function of OPP test

7 groups of OPP test parameters can be set at most.

Operation instructions: it takes the setting stored in Group 1, VON voltage of 10V, VON voltage delay of 5S, and current range of 3A; start power of 20W, each decrease of 1W, and each decreasing duration of 1s; end power of 10W, OPP voltage of 8V, maximum power of 15W, and minimum power of 10W as an example.

Stamos S-LS-119 - Setting function of OPP test - 1

text_image SAVE MODE: OPP NO. DATA UNITS EXPLAIN 1 10.000 V VON LEVEL 2 5.0000 S VON DELAY 3 3.0000 A RANGE 4 20.000 W START 5 1.0000 W STEP 6 1.0000 S STEP DELAY 7 10.000 W END 8 8.0000 V OPP VOLT 9 15.000 W MAX TRIP 10 10.000 W MIN TRIP OUTPUT: 15.978V 00.000A 00.000W SAVE Group: 1 OCP CALL OCP SAVE OPP CALL OPP SAVE HOME

Figure 8.1 OPP

1) Press (F2) OCP/OPP on home page.
2) Press (F3) OPP/CALL and TAB to select CALL GROUP 1.
3) Press (F4) OPP/SAVE to display SAVE on the table.
4) Press TAB to switch the focus and turn the knob to select the first row.
5) Press ENTER to select DATA in the first row and display the red background.
6) Enter numeric 0 to 9, and ESC.
7) After editing VON as 10V, press ENTER. Then 10.000 will be displayed on the screen, and the background will be switched to blue.
8) Repeat the above steps to set 5S, 3A, 20W, 1W, 1S, 10W, 8V, 15W, and 10W respectively.

9) Long press ENTER to save the edited data.

10) Press (F3) OCP/CALL.

11) Press ENTER to call.

12) Turn ON/OFF.

9. WAVE Mode

9.1. WAVE measurement

1) Press TAB to display voltage, current and waveform.
2) Press WAVE to display the measuring column.
3) Measure the time scale. Press or4 to select the left or right measuring column and turn the knob to move left or right for the display of the difference between the two measuring lines.
4) Measure the negative value of voltage or current. Press T or to select the upper or lower measuring column and turn the knob to move up or down for the display of amplitude of the current measuring column.
5) Adjust the scale value of current. Long press and turn the knob to adjust the size.
6) Adjust the scale value of voltage. Long press and turn the knob to adjust the size.
7) Adjust the time value of sampling. Press ENTER and turn the knob to adjust the size.
8) When the wave pauses, press STOP.

Take the measurement of DYN CC mode as an example.

After editing the data of DYN CC, A SLOPE is 0.012 A/ S, B SLOPE is 0.08 A/ S, A is 0.2A, B is 1A, frequency is 20 HZ, duty cycle is 40%, as shown in Fig. 9.1. Please see Fig. 9.2 for the waveform measurement.

Stamos S-LS-119 - WAVE measurement - 1

text_image CALL MODE: Dynamic NO. DATA UNITS EXPLAIN 1 0.0120 A/uS A SLOPE 2 0.0800 A/uS B SLOPE 3 0.2000 A LEVEL A 4 1.0000 A LEVEL B 5 20.000 Hz FREQUENCY 6 40.000 % DUTY OUTPUT: 06.382V 00.000A 00.000W CALL DYN: CC DYN CALL DYN SAVE LIST CALL LIST SAVE HOME

Figure 9.1

Stamos S-LS-119 - WAVE measurement - 2

line | Parameter | Value | | --------- | --------- | | V:19.520V | 0.400 | | C: 0.320A | 2.400 | | M:50.000mS| 0.400 | | Label | Value | | --------- | --------- | | OUTPUT | 15.954V | | DYN | CC | | OUTPUT | 15.954V | | OUTPUT | 00.533A | | OUTPUT | 08.511W | | T:01398 | 01398 |

Figure 9.2 Waveform measurement

10. Setting of Parameter Function

Stamos S-LS-119 - Setting of Parameter Function - 1

text_image System Settings Serial Baud Rate: 115200 IP Address: 192.168.2.198 Gate Way: 192.168.2.1 Version: V1.10, 300W Communication Ver: V1.10 Subnet Mask: 255.255.255.0 IP Port: 18190 Function Settings USB Storage time(S): 1.20 Remote Compensation: OFF Voltage MAX: 61.000 V Power MAX: 300.10 W CC Slope: Low BEEP: ON EXT TRIG: OFF Current MAX: 15.000 A Resistance MAX: 7500.0 L2 Select Select Select Select Select Save Baudrate UsbStore Voltage UP DOWN Exit
Select Baudrate (F1)Select the parameter of baud rate
Select UsbStore (F2)Switch focus to select USB store
Select Voltage (F3)Switch the Max setting of focus voltage
Select UP (F4)Switch the focus up
Select DOWN (Fs)Switch the focus down
Save Exit (F6)Save and exit

10.1. Setting of communication interface

1) Setting of serial baud rate

Select TAB or (F1) SelBaudrate to turn the knob

(1) Setting of IP address

ESC on the keyboard is the delete key, and numeric 0 to 9 is for entering numbers.

(2) Setting of port number

ESC on the keyboard is the delete key, and numeric 0 to 9 is for entering numbers. The maximum port number is 65535, and the minimum port number is 1000. The value 18191 cannot be set.

10.2. Setting of CC slope rate

Press TAB to switch to CC SLOPE and turn the knob to select LOW/HIGH.

10.3. Storage time setting of USB flash disk

Press TAB to switch to Sel UsbStore. ESC is the delete key, and numeric 0 to 9 is for entering numbers. The minimum storage time can only be 0.05S, and the maximum storage time is 9999S;

Press TAB to switch to BEEP and turn the knob to select ON/OFF.

10.5. Setting of remote compensation

Press TAB to switch to Remote Comp, select ON, and connect the output end of the object to be tested to the terminal sense (+) or sense (-) on the front panel (this function is not saved during the blackout, otherwise it is closed).

10.6. Setting of external triggering

Press TAB to switch to EXIP TRIG and turn the knob to select Trig On/Switch On/OFF.

Trig On: triggering switch (open load after triggering)

Switch on: remote switch (ON/OFF function on the front panel will not work when this function works).

10.7. Setting of maximum values

1) Maximum voltage

Press TAB to switch to Voltage. ESC is the delete key, and numeric 0 to 9 is for entering numbers. The maximum voltage is 150V.

2) Maximum current

Press TAB to switch to Current. ESC is the delete key, and numeric 0 to 9 is for entering numbers. The maximum current is 40A.

3) Maximum power

Press TAB to switch to Power. ESC is the delete key, and numeric 0 to 9 is for entering numbers. The maximum power depends on the model.

4) Maximum resistance

Press TAB to switch to Resistance. ESC is the delete key, and numeric 0 to 9 is for entering numbers. The maximum resistance is 7500R.

Note: 18V and below is set to be the low voltage range (0 to 18V), and above

  1. 4V is set to be the high voltage range (0 to 150V).

3A and below is set to be the low current range (0 to 3A), and above 3.1A is set to be the high current range (0 to 40A).

OVP alarm value: it is (19.4V) at the low range and (155V) at the high range. OCP alarm value: it is 105% of the set current value. For example, if the current is set to be 5A, the alarm value is 5.25A.

OPP alarm value: it is 105% of the set power value. For example, if the power value is set to be 100W, the alarm value is 105W.

OTP alarm value: if the temperature is higher than 850, an alarm will be given, and the load will be closed.

10.8. RTC setting

Long press numeric 7, and then date will be displayed on the screen as shown in Fig. 10.2.

Stamos S-LS-119 - RTC setting - 1

text_image USB 2020-08-12 08:44 06 ent MODE:CC Scan OUTPUT: 0.0000V 0.0000A 0.0000W

Figure 10.2 RTC Date

Press or to move the focus and turn the knob to make modification. After modification, long press ENTER or numeric 7 to save, and press ESC to exit.

10.9. Backlight setting

After long pressing numeric 8, the progress bar will be displayed on the screen as shown in Fig. 10.3.

Stamos S-LS-119 - Backlight setting - 1

text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V

Figure 10.3 Backlight

Turn the knob to adjust the brightness, long press ENTER or numeric 8 to save, and press ESC to exit.

10.10. 0 calibration function

If the device has not been calibrated for a long time or the temperature is high or low, the data will not return to zero, and the 0-calibration function can be performed.

After entering home page, close the load, unplug the external test line, and long press numeric 0 for over 3 seconds for the 0 calibration of voltage and current. And the 0-calibration function will disappear when the load restarts, and 0-calibrate again when needed.

11. U Flash Disk Import and Export Function

Save real-time voltage and current data and export data list through U flash disk. As shown in Fig. 11.1, the USB [0] tree on the left shows that the file list is the acceptable format data loaded in the U flash disk. And the upper part on the right is the option of importing the load from the U flash disk, and the lower part is the data option of exporting the load to the U flash disk.

Press TAB to select the control. Press ◀▶ ▲▼ to move the focus left or right. Press F4 (Load CSV) to import the file from the USB flash disk and save CSV (F5) to export the CSV file to the USB flash disk. And the function keys are shown in Fig. 11.2 below:

Stamos S-LS-119 - U Flash Disk Import and Export Function - 1

text_image USB[0:] STATIC_set.csv OPP_1_set.csv OCP_1_set.csv LIST 1 set.csv DYN_RL_set.csv DYN_PL_set.csv DYN_CW_set.csv DYN_CV_set.csv DYN_CR_set.csv DYN_CC_s.csv BAT_1_s.csv Select the file import list from the left window LOAD CSV: LIST Group 1 Export files to USB flash drive SAVE CSV: LIST_1.csv LIST Group 1 Start Stop USB LOAD SAVE SaveData SaveData Discon. CSV CSV CSV HOME

Figure 11.1 Storage interface of USB Flash Disk

Start Save Data (F1)Create a CSV file with date as its name
Stop Save Data (F2)Stop writing files
USB Disconnect (F3)Disconnect USB
Load CSV (F4)Import files from USB flash disk
Save CSV (F5)Export CSV file
Home Page (F6)Home page

Figure 11.2

11.1. Import and save LIST

Take the table that data of BAT Group 1 is exported as an example.

1) After the USB flash disk is inserted, USB key is displayed on home page. Press (F4) USB.
2) Press TAB to switch to SAVE, turn the knob to select BAT mode, press6 or4 to select Group, and turn the knob to Group 1.
3) After pressing (F5) SAVE CSV, a message will prompt that it has been exported. Import DYN_PL_Set.csv from USB flash disk to DYN PL.

K12fx
ABCDE
1 NO.DATA(A)UNITSEXPLAIN
2 11.132A/uSA SLOPE
3 21.248A/uSB SLOPE
4 31.51ALEVEL A
5 43.201ALEVEL B
6 56.813STIME WIDTH
7

Figure 11.3 DYN_PL_Set.csv

4) Press TAB to switch to LOAD, adjust to DYN mode and turn the knob to PL.
5) List a tree of files on the left and turn the knob to select the file DYN_PL Set.CSV to be imported.

6) After pressing (F4) LOAD CSV, successful import will be prompted.

11.2. U Flash Disk Storage of Real-time Data of Load Test

If the real-time test data is saved with a U flash disk, the data volume is to save the voltage and current data 5 times per second.

The operation procedures are as follows:

1) Press F6 (Config), use Tab to switch the focus on USB Store (Fig. 10.1 Parameter Setting).

Use ESC to delete, enter the number key 0.2, which means to save 5 times per second.

2) There are 2 ways to open the save data file.

<1> Turn on or off data saving in the menu interface: enter the U flash disk page (Fig. 11.1 Storage Interface of USB Flash Disk). Press (F1) "Start Save Data" to start, then the top status bar will have a down arrow flashing, indicating that the data is being saved right now.

To stop writing, re-enter the USB flash disk page, then press (F2) "Stop Save Data". And the top arrow will disappear.

<2> Shortcut key to open or close the operation: in the state of inserting the U flash disk, long press the number key 9 to start the U flash disk saving, and long press the number key 9 again to stop saving.

Refer to Fig. 11.4 Data Write Icon

Stamos S-LS-119 - U Flash Disk Storage of Real-time Data of Load Test - 1

text_image USB current MODE:CC Sca

Fig. 11.4 Data Write Icon

Diagram of External Interfaces

Output compensation.

Stamos S-LS-119 - U Flash Disk Storage of Real-time Data of Load Test - 2

text_image RORO DC PAMWARIAS BLDVIRE LOB S+ S- 1

1 - Power

Stamos S-LS-119 - U Flash Disk Storage of Real-time Data of Load Test - 3

text_image LINE 180Vac - 110Vac 220Vac - 240Vac FUSE 1.25A 6.5A LAN USB RS232 1 2

1 – External switch control

2 – External Trigger control

12. Cleaning and maintenance

• Always unplug the device before cleaning or putting it away.
• Use only non-corrosive cleaners to clean the surface.
• After cleaning the device, all parts should be dried completely before using it again.
• Store the unit in a dry, cool place, free from moisture and direct exposure to sunlight.
- Do not spray the device with a water jet or submerge it in water.
- Do not allow water to get inside the device through vents in the housing of the device.
- Clean the vents with a brush and compressed air.

  • The device must be regularly inspected to check its technical efficiency and spot any damage.
  • Use a soft cloth for cleaning.
  • Do not use sharp and/or metal objects for cleaning (e.g. a wire brush or a metal spatula) because they may damage the surface material of the appliance.
  • Do not clean the device with an acidic substance, agents of medical purposes, thinners, fuel, oils or other chemical substances because it may damage the device.

DISPOSING OF USED DEVICES:

Do not dispose of this device in municipal waste systems. Hand it over to an electric and electrical device recycling and collection point. Check the symbol on the product, instruction manual and packaging. The plastics used to construct the device can be recycled in accordance with their markings. By choosing to recycle you are making a significant contribution to the protection of our environment. Contact local authorities for information on your local recycling facility.

Stamos S-LS-119 - DISPOSING OF USED DEVICES: - 1

text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca
text_image KORAC DC PAGNARIALS BLEDVRSI25 S+ S- 1

1 - Moc

Stamos S-LS-119 - DISPOSING OF USED DEVICES: - 2

text_image LINE 190Vac - 110Vac 220Vac - 240Vac FUSE 1.35A 6.5A LAN USB RS232 1 2
text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca

Obr. 11.4 Ikona zápisu dat

Figure 10.2 Date RTC

text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V

Figure 11.3 DYN_PL_Set.csv

text_image USB current MODE:CC Sca
text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca
text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca
text_image KORAC DC PAGNARALS BLDVERS LOS S+ S- 1

1 - Poder

Stamos S-LS-119 - DISPOSING OF USED DEVICES: - 3

text_image LINE 190Vac - 110Vac 220Vac - 240Vac FUSE 1.35A 6.5A LAN USB RS232 1 2

1 – Control de interruptor externo

2 – Control de disparo externo

text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca
text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca

Fig. 11.4 Ikon for dataskrivning

text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca
text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca
text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca

Fig. 11.4 Dataskriveikon

Diagram over eksterne grensesnitt

text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca

Fig. 11.4 Dataskrivikon

1. Interface principal

Stamos S-LS-119 - Interface principal - 1

line | Parameter | Value | | --------------- | --------- | | OUTPUT | 15.979V | | OCP/OPP | 00.000A | | Battery | 00.000W | | USB | - | | Dynamic | - | | Config | - |

1.1. Teclas de menu

text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V

Figura 10.3 Luz de fundo

text_image USB current MODE:CC Sca
text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca
text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca

Фиг. 11.4 Икона за запис на данни

text_image KORAC DC PAGUARABLE ELEVENTE LOSB S+ S- 1

1 - Мощност

Stamos S-LS-119 - Фиг. 11.4 Икона за запис на данни - 1

text_image LINE 190Vac - 110Vac 1.25A 220Vac - 240Vac 6.5A LAN USB RS232 1 2
text_image USB current MODE:CC Sca

10) Pritisnite (F3) OCP/CALL.

11) Pritisnite ENTER za poziv.

text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca

Slika 11.4 Ikona za pisanje podataka

line | Current (A) | Voltage (U) | Power (W) | |---|---|---| | 0.200 | 0.000 | 119.8 | | 1.833 | 4.793 | 95.87 | | 1.867 | 9.587 | 71.90 | | 2.780 | 14.38 | 47.93 | | 3.533 | 19.17 | 23.97 | | 4.367 | 23.97 | 0.000 | | 5.200 | 23.97 | 0.000 | The chart displays a single line plot of voltage and power over time, with annotations indicating maximum voltage (Max(V)) and maximum power (Max(W)). The output is labeled as "15.978V", "00.000A", and "00.000W". The bottom panel shows a button labeled 'VI-MODE' with group '1'.
text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V

10.3 pav. Foninis apšviemas

text_image USB current MODE:CC Sca
text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V

Figura 10.3 Iluminare de fundal

text_image USB current MODE:CC Sca

Fig. 11.4 Pictograma de scriere a datelor

Diagrama interfețelor externe

Compensarea ieşirii.

Stamos S-LS-119 - Фиг. 11.4 Икона за запис на данни - 2

text_image KORNO DC PAMINUARUS BLEDVRSI LOS $+ $- 1

1 – Putere

Stamos S-LS-119 - Фиг. 11.4 Икона за запис на данни - 3

text_image LINE 190Vac - 110Vac 220Vac - 240Vac FUSE 1.25A 6.5A LAN USB RS232 1 2

10) Pritisnite (F3) OCP/CALL.

11) Za klic pritisnite ENTER.

12) VKLOP/IZKLOP.

9. Način WAVE

9.1. Merjenje WAVE

1) Pritisnite TAB za prikaz napetosti, toka in valovne oblike.
2) Pritisnite WAVE za prikaz merilnega stolpca.
3) Izmerite časovno lestvico. Pritisnite ali4 za izbiro levega ali desnega merilnega stolpca in zavrtite gumb za pomik levo ali desno za prikaz razlike med obema merilnima linijama.
4) Izmerite negativno vrednost napetosti ali toka. Pritisnite T ali za izbiro zgornjega ali spodnjega merilnega stolpca in vrtenje gumba za pomik gor ali dol za prikaz amplitude trenutnega merilnega stolpca.
5) Prilagodite vrednost lestvice toka. Dolgo pritisnite in zavrtite gumb, da prilagodite velikost.
6) Prilagodite vrednost lestvice napetosti. Dolgo pritisnite in zavrtite gumb, da prilagodite velikost.
7) Prilagodite časovno vrednost vzorčenja. Pritisnite ENTER in zavrtite gumb, da prilagodite velikost.
8) Ko se valovanje ustavi, pritisnite STOP.

text_image 31.4 °C 16 58 Scan 42% OUTPUT: 00 000V
text_image USB current MODE:CC Sca
text_image KORNO DC PAGNUSABLE BLDVIRE LOB S+ S- 1

1 - moč

Stamos S-LS-119 - Merjenje WAVE - 1

text_image LINE 190Vac - 110Vac 220Vac - 340Vac FUSE 1.25A 6.5A LAN USB RS232 1 2

1 – Zunanje krmiljenje stikala

For the disposal of the device please consider and act according to the national and local rules and regulations.

CONTACT

expondo Polska sp. z o.o. sp. k.

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Stamos

Model : S-LS-119

Category : Electronic test device