USER MANUAL 877 SECA

Instructions for use

17-10-07-575-100c_2025-02S

SECA 877 - Instructions for use - 1

Deutsch - Gebrauchsanweisung 3
English-Instructions for use 26
Francais - Mode d'emploi 49
Italiano - Istruzioni per l'uso 73
Espanol - Instrucciones de uso 96
Portugues - Manual de instruções de'utilisation 121
Dansk - Brugsanvising 145
Svenska - Bruksanvisning 169
Norsk - Bruksanvising 192
Suomi-Kayttöohje 215
Nederlands - Gebruiksaanwijzing 238
Elambdavika-Odeltaicxpnoan 261
Česky - Návod k použití 286

INHALTSVERZEICHNIS

1 Description of device 27

1.1 Intended use 27
1.2 User qualification 27
1.3 Patient target group 27
1.4 Contraindications 27
1.5 Clinical benefit 27
1.6 Functional description.... 27

2 Safety information 28

2.1 Safety information in these instructions for use. 28
2.2 Basic safety information... 28

Handling device. 28
Preventing electric shock 30
Preventing injuries and infections 30
Preventing damage to device 31
Handling measuring results 32
Handling packaging material. 32
Handling batteries and rechargeable batteries.. 33

3 Overview 34

3.1 Controls 34
3.2 Symbols in the display .... 34
3.3 Labels 35

4 Starting up device 38

4.1 Scope of delivery 38
4.2 Inserting batteries 38
4.3 Setting up and aligning the device 39

5 Operation 40

5.1 Switching on the device ... 40
5.2 Weighing 40

Weighing a patient. 40

Weighing babies/ toddlers (2 in 1) 41

6 Hygiene treatment 42

6.1 Cleaning 42
6.2 Disinfecting 42
6.3 Sterilizing 43

7 Function check 43

8 Troubleshooting 44

9 Servicing 45

10 Technical data. 45

10.1 General technical data.... 45
10.2 Weight measurement 46

11 Optional accessories. 46

12 Disposal 47

12.1 Disposing of the device .... 47
12.2 Disposing of batteries and rechargeable batteries.... 47

13 Warranty 47

14 Declaration of conformity 48

1 DESCRIPTION OF DEVICE

1.1 Intended use

The electronic flat scale supports physicians in decision-making regarding diagnosis or therapy, based on the patient's weight or weight changes.

1.2 User qualification

The device may only be used by persons with sufficient expertise.

1.3 Patient target group

The device is intended for persons of any age who can stand independently and who do not exceed the maximum weight capacity of the device. Babies and toddlers who cannot stand independently can be weighed using the 2 in 1 function whilst being held by an adult who can stand independently.

1.4 Contraindications

No contraindications are known.

1.5 Clinical benefit

The electronic flat scale supports physicians in decision-making regarding diagnosis or therapy, based on the measured weight.

By that, the electronic flat scale only provides an indirect clinical benefit.

The physician must perform additional examinations for an accurate diagnosis.

1.6 Functional description

Weight on this electronic personal scale is detected by four load cells.

The 2 in 1 function can be used to determine the weight of babies and toddlers. The child is held in an adult's arms during weighing to do this.

2 SAFETY INFORMATION

2.1 Safety information in these instructions for use

SECA 877 - Safety information in these instructions for use - 1

DANGER!

Used to identify an extremely hazardous situation. If you fail to take note of this information, serious irreversible or fatal injuries will occur.

SECA 877 - DANGER! - 1

WARNING!

Used to identify an extremely hazardous situation. If you fail to take note of this information, serious irreversible or fatal injuries may result.

SECA 877 - WARNING! - 1

CAUTION!

Used to identify a hazardous situation. If you fail to take note of this information, minor to moderate injuries may result.

NOTICE!

Used to identify possible incorrect usage of the device. If you fail to take note of this information, you may damage the device, or the measuring results may be incorrect.

NOTE

Includes additional information about use of the device.

2.2 Basic safety information

Handling device

  • Please take note of the information in these instructions for use.
  • Keep the instructions for use in a safe place. The instructions for use are a component of the device and must be available at all times.

English

In the interest of patient safety, you and your patients are obliged to report serious events that occur in connection with this product to the manufacturer and to the authority responsible in your country.

SECA 877 - English - 1

DANGER!

Risk of explosion

Do not use the device in an environment in which one of the following gases has accumulated:

-oxygen
- flammable anesthetics
- other flammable substances/air mixtures

SECA 877 - Risk of explosion - 1

CAUTION!

Patient hazard, damage to device

Have servicing carried out regularly as described in the relevant section of this document.
Technical modifications may not be made to the device. The device does not contain any user-serviceable parts. Only have servicing and repairs performed by an authorized seca service partner. You can find your local service partner at www.seca.com.
Use only seca original accessories and spare parts, otherwise seca will not grant any warranty.

SECA 877 - Patient hazard, damage to device - 1

CAUTION!

Patient hazard, malfunction

  • Keep medical electrical devices such as high-frequency surgical equipment at a minimum distance of approx. 1 meter to prevent faulty measurements.
  • Keep high-frequency equipment such as cellphones at a minimum distance of approx. 1 meter to prevent faulty measurements.
    The actual transmission output of HF devices may require minimum distances of more than 1 meter. For details, go to www.seca.com.

Preventing electric shock

SECA 877 - Preventing electric shock - 1

WARNING! Electric shock

  • Set up devices which can be operated with a power supply unit so that the power supply socket is within easy reach and the power supply can be disconnected quickly.
  • Ensure that your local electricity supply matches the details on the power supply unit.
  • Never touch the power supply unit with wet hands.
    Do not use extension cables and multiple outlets.
  • Ensure that cables are not crushed or damaged by sharp edges.
  • Ensure that cables do not come into contact with hot objects.
    Do not operate the device at an altitude of more than 3000m above sea level.

Preventing injuries and infections

SECA 877 - Preventing injuries and infections - 1

WARNING! Injury from falling

Ensure that the device is steady and level.
- Route connector cables (if present) so that neither users nor the patient can trip over them.
Assist people with limited motor skills when they are getting up, e.g. from a wheelchair.
Ensure that the patient does not step directly onto or off the edges of the weighing platform.
- Ensure that the patient steps onto and off the weighing platform slowly and safely.

SECA 877 - WARNING! Injury from falling - 1

WARNING! Danger of slipping

  • Ensure that the weighing platform is dry before the patient steps onto it.
  • Ensure that the patient has dry feet before stepping onto the weighing platform.
  • Ensure that the patient steps onto and off the weighing platform slowly and safely.

SECA 877 - WARNING! Danger of slipping - 1

WARNING! Risk of infection

Before and after every measurement, wash your hands to reduce the risk of cross-contamination and nosocomial infections.
Should the patient have contagious diseases, subject the device to a hygiene treatment as described in the relevant section of these instructions for use immediately afterwards.
- Ensure that the patient does not have any open wounds or infectious skin alterations which may come into contact with the device.
Subject the device to a hygiene treatment at regular intervals.

Preventing damage to device

NOTICE! Damage to device

  • Ensure that fluids never get inside the device. These can destroy the electronics.
    For devices with power supply operation: Switch off the device before you disconnect the power supply unit from the power supply socket.
    For devices with power supply operation: If the device is not to be used for an extended period, disconnect the power supply unit from the power supply socket. Only then is the device de-energized.
    For devices with battery or rechargeable battery operation: If you are not using the device for an extended period of time, remove batteries or rechargeable batteries. Only then is the device de-energized.
    Do not drop the device.
    Do not subject the device to shocks or vibrations.
    Perform a function check before each use as described in the corresponding section in this document. Do not operate the device if it is not working properly or is damaged.

Do not place the device in direct sunlight and ensure that it is not placed in the direct proximity of a heat source. The excessive temperatures could damage the electronics.
- Avoid rapid temperature fluctuations. If the device is transported so that a temperature difference of over 20^ occurs, the device must be left to stand for at least 2 hours before it is switched on, otherwise condensation may form; this may damage the electronics.
Use the device only in the intended ambient conditions outlined in the section entitled "Technical data".
- Store the device only in the intended storage conditions outlined in the section entitled "Technical data".
Use only cleaning agents and disinfectants which match the details in the section entitled "Hygiene treatment".
For scales: Ensure that maximum capacity is not exceeded.

Handling measuring results

NOTICE!

Inconsistent measuring results

Before you save and continue using measured values determined using this device (e.g. in seca software or in an EMR system), ensure that the measured values are plausible.
If measured values have been sent to seca software or to an EMR system, ensure before continuing to use them that the measured values are plausible and assigned to the correct patient.

Handling packaging material

SECA 877 - Handling packaging material - 1

WARNING!

Risk of suffocation

Packaging materials made of plastic film (bags) present a risk of suffocation.

  • Store packaging material out of the reach of children.
    If the original packaging material is no longer available, only use plastic bags with safety holes to reduce the risk of suffocation. Use recyclable materials if possible.

NOTE

Store the original packaging material for future use (e.g. returning for servicing).

Handling batteries and rechargeable batteries

SECA 877 - Handling batteries and rechargeable batteries - 1

WARNING! Injury

Batteries and rechargeable batteries contain harmful substances which may explode and so escape if not handled properly.

Do not try to recharge batteries.
Do not expose batteries/rechargeable batteries to heat.
Do not burn batteries/rechargeable batteries.
If harmful substances are leaking out, avoid contact with the skin, eyes, and mucous membranes. Rinse affected areas of the body with plenty of clean water and seek medical help at once.

NOTICE!

Damage to device and malfunction due to incorrect handling

Use only the type of battery/rechargeable battery quoted in this document.
Always replace all batteries/rechargeable batteries at the same time.
Do not short-circuit batteries/rechargeable batteries.
If you do not use the device for an extended period of time, remove the batteries/rechargeable batteries. This prevents acid from leaking into the device.
If harmful substances have penetrated the device, do not continue using it. Have the device checked by an authorized seca service partner and repaired if necessary.

3

OVERVIEW

3.1 Controls

SECA 877 - Controls - 1

ItemControl Function
12 in 1 keyStarts the 2 in 1 function for weighing babies and tod- dlers
2Foot screw 4 pcs, for precise alignment
3Battery compartmentTakes 6 AA-type batteries (AA/1.5 V)
4Spirit level Indicates whether the device is horizontal

3.2 Symbols in the display

SECA 877 - Symbols in the display - 1

Item Symbol Meaning
1kgWeight value in kilograms
2NET Tare functionactive

English

Item Symbol Meaning
3 HOLD Hold function active
4Non-verifiable function active
5Batteries are low

3.3 Labels

Markings on the device and on the type plate
Text/symbolMeaning
Name and address of manufacturer, date of manufacture
UDIUnique Device Identification /product identification number as per Directive (EU) 2017/745)
REFArticle number
SNSerial number
GALValue in m/s² (model-dependent) • Gravitational acceleration on earth • Depends on the intended location
ProDIseca product identification number
Approval TypeType designation of design approval
Mat.No.Variant number
Follow instructions for use
Medical electrical device, Type B
IP20Type of protection to IEC 60529: • Protection against ingress of solid foreign bodies with a diameter of over 12.5 mm • Protection against access with fingers • No protection from ingress of water
eValue in units of mass Used to classify and verify a scale
IIIClass III scale according to Directive 2014/31/EU
CE M24 0102 0123Device complies with EU directives •M: Conformity label in compliance with directive 2014/31/EU for non-automatic scales (verified models) •24: (Example: 2024) Year in which declaration of conformity was implemented and the CE label was applied (verified models) •0102: Notified Body for Metrology (verified models) •0123: Notified Body for Medical Devices
MDMedical device in accordance with Regulation (EU) 2017/745
UK CA M24 xxx xxxxDevice complies with United Kingdom directives •M: Conformity label in compliance with UK directive SI 2016 no. 1152 for non-automatic scales (NAWIR) (verified models) •24: (Example: 2024) Year in which declaration of conformity was implemented and the UKCA label was applied (verified models) •xxxx: Notified Body for Medical Devices of the United Kingdom •yyyy: Notified Body for Metrology of the United Kingdom (verified models)
UK REPImporter/representative in the United Kingdom: seca Ltd 40 Barn Street B5 5QB Birmingham United Kingdom
CH REPImporter/representative in Switzerland: seca ag (schweiz) Medizinische Waagen und Messysteme Schönmitt Str. 2 CH-4153 REINACH
Do not dispose of device in household waste

English

Markings on the packaging
Symbol Meaning
Protect from moisture
Arrows indicate top of product Transport and store in an upright position
Fragile, do not throw or drop
Permitted min. and max. temperature for transport and storage
Permitted min. and max. humidity for transport and storage
Permitted min. and max. air pressure for transport and storage
Open packaging here
Packaging material can be disposed of through recycling programs

4 STARTING UP DEVICE

4.1 Scope of delivery

SECA 877 - Scope of delivery - 1

ItemComponent Pcs.
aPersonal scale 1
bBattery 6
- Instructions for use 1

4.2 Inserting batteries

SECA 877 - Inserting batteries - 1

You need 6 batteries (type AA/1.5 V, included in scope of delivery).

  1. Press on the latch of the battery compartment.
  2. Remove the cover of the battery compartment.

NOTICE!

Damage to device

Device may overheat and become damaged.

  • Ensure the correct polarity of the batteries (markings in the battery holder).

  • Insert the batteries.

NOTE

If bAtt appears in the display, you have inserted one of the batteries the wrong way round or the batteries have discharged.

  1. Close the battery compartment.

4.3 Setting up and aligning the device

SECA 877 - Setting up and aligning the device - 1

CAUTION!

Faulty measurement as a result of devices being set up incorrectly

Soft floors (wooden boards, for example) give under the patient's weight and falsify the measuring result. Incorrect ambient conditions or lack of alignment lead to measurement errors.

  • Select a setup location with a level, stable floor in order to achieve precise measuring results.
    Use the device only in the intended ambient conditions outlined in the section entitled "Technical data".
    Align the device before using it and every time it changes location.

SECA 877 - Faulty measurement as a result of devices being set up incorrectly - 1

  1. Place the device on a firm, level surface.
  2. Align the device by turning the foot screws so that the air bubble of the spirit level is in the exact center of the circle.
  3. Ensure that only the feet of the device are in contact with the ground.

5 OPERATION

CAUTION! Personal injury/faulty measurement

Before using the device each time, perform a function check as described in the corresponding section of these instructions for use.

5.1 Switching on the device

SECA 877 - Switching on the device - 1

The scale has no load.

Tap the scale with your foot.

The scale is operational when 0.0 appears in the display.

The scale switches off automatically 20 seconds after it was last used.

5.2 Weighing

Weighing a patient

The scale has no load.

  1. If the scale has switched off automatically after the previous measurement, tap the scale with your foot.

NOTICE!

Damage to device

The display is sensitive and may be damaged under load

Make sure that the patient does not step on the display.

SECA 877 - Damage to device - 1

SECA 877 - Damage to device - 2

  1. Ask the patient to step onto the scale.
  2. Read off the measuring result.

English

Weighing babies/ toddlers (2 in 1)

The 2 in 1 function can be used to determine the weight of babies and toddlers. The child is held in an adult's arms during weighing to do this. Proceed as follows:

  1. Ensure that there is no load on the scale.

  2. Ask the adult to stand on the scale.

  3. Read off the measuring result.

  4. Keep the 2 in 1 key depressed until the message NET appears in the display.

The weight is saved.
0.0 appears in the display.

SECA 877 - Weighing babies/ toddlers (2 in 1) - 1

#

WARNING!

Faulty measurement as a result of initial weight changing

If the child is weighed together with a different adult, it will not be possible to determine the child's weight correctly.

  • Ensure that the child is always weighed with the adult with whom the initial weight was determined.
  • Ensure that the adult's weight does not change - as a result of their taking off a piece of clothing, for example.

SECA 877 - Faulty measurement as a result of initial weight changing - 1

  1. Ask the adult to get onto the scale holding the child.

The child's weight is displayed.
The symbol and the messages HOLD and NET are displayed.

  1. Read off the measuring result.
  2. Ask the adult to get off the scale with the child.
  3. You have the following options for deactivating the 2 in 1 function:

Press the 2 in 1 key until the symbol and the messages HOLD and NET are no longer displayed.
Switch off the device.

6 HYGIENE TREATMENT

WARNING! Electric shock

The device is not de-energized when the on/off button is pressed and the display goes out. Use of fluids on the device may cause an electric shock.

Before each hygiene treatment, ensure that the device is switched off.
Take the batteries/battery packs out of the device before every hygiene treatment.
Ensure that no fluids penetrate the device.

WARNING! Risk of infection

Subject the device to a hygiene treatment at regular intervals as described in this section.

NOTICE! Damage to device

Unsuitable cleaning agents and disinfectants may damage the sensitive surfaces of the device and lead to transparent components becoming clouded.

Do not use aggressive or abrasive cleaning agents.
Do not use organic solvents (e.g. white spirit or petroleum spirit).

6.1 Cleaning

If required, moisten a soft cloth with a mild soap solution and wipe the device over with it.

6.2 Disinfecting

  1. Disinfect the device at regular intervals with a disinfectant suitable for sensitive surfaces and acrylic glass (e.g. 70% ethanol).
  2. Follow the instructions for use of the disinfectant.
  3. Disinfect the device as follows:

English

Moisten a soft cloth with disinfectant and wipe down the device with it.
Comply with the intervals, see table:

Interval Component
Before and after every measure-ment with direct skin contactWeighing plat-form
As requiredDisplay Carry case

6.3 Sterilizing

The device must not be sterilized.

7 FUNCTION CHECK

Perform a function check before each use.

A complete function check includes:

  • A visual inspection for mechanical damage
  • A test of the alignment of the device
  • A visual and function check of the display elements
  • A function check of all the controls shown in the section entitled "Overview"
  • A function check of the optional accessories

If you find faults or deviations during the function check, first try to remedy the fault with the aid of the "Troubleshooting" section in this document.

SECA 877 - FUNCTION CHECK - 1

CAUTION!

Personal injury

If you find faults or deviations during the function check which you are unable to remedy with the aid of the "Troubleshooting" section in this document, you must not use the device.

Have the device repaired by seca Service or an authorized service partner.
Follow the section entitled "Servicing" in this document.

8 TROUBLESHOOTING

If faults occur when operating the device, first attempt to remedy them yourself using the following table. If the fault persists, contact seca Service or an authorized service partner.

Fault Cause Remedy
The scale does not display a weight although a load was placed on itScale is switched off or without batteriesCheck that the scale is switched onCheck that batteries are inserted
The scale displays a weight following transport or a battery replacement even though no load has been placed on itScale has determined an incorrect zero pointSwitch off the scale, then switch it back on again
0.0 does not appear before weighingScale had load on it before being switched onRemove the load from the scaleSwitch off the scale, then switch it back on again
---- displayed before weighingScale had load on it before being switched onRemove the load from the scaleSwitch off the scale, then switch it back on again
Single segment is illuminated continuously or not at allDisplay faulty Inform seca Service
Display:Battery voltage is dropping Replace the batteries
Display: bAttBatteries discharged or not insertedReplace the batteriesInsert the batteries the right way round
Display: StOPMaximum capacity ex- ceededRemove the load from the scale
Display: Er + figureToo high a load on the scale or too high a load in one cornerRemove the load from the scale or distribute the weight more evenlySwitch off the scale, then switch it back on againIf the problem persists, inform seca Service

9 SERVICING

SECA 877 - SERVICING - 1

CAUTION!

Faulty measurements as a result of verification being performed poorly or not at all

Have verification performed only by authorized persons.
Always have verification performed if one or more quality seals are damaged.

seca recommends having your device serviced prior to verification.

SECA 877 - CAUTION! - 1

CAUTION!

Faulty measurements as a result of poor servicing

Have servicing and repairs carried out exclusively by seca Service or an authorized service partner.
You can find your local service partner at www.seca.com.

10 TECHNICAL DATA

10.1 General technical data

Dimensions • Depth • Width • Height356 mm 321 mm 60 mm
Net weight Approx. 4.0 kg/8.8 lbs
Ambient conditions, operation • Temperature • Air pressure • Humidity+10 °C to +40 °C (50 °F to 104 °F) 700 – 1060 hPa 30 % – 80 %, no condensation
Ambient conditions, storage/transport • Temperature • Air pressure • Humidity-10 °C to +65 °C (14 °F to 149 °F) 700 – 1060 hPa 0 % – 95 %, no condensation
Height of digits 20 mm
Power supply • Batteries - Supply voltage - Battery type9 V 6 x type AA, 1.5 V, not rechargeable
Medical device in accordance with Regulation (EU) 2017/745Class I with measuring function
EN 60601-1: Medical electrical equipment, type B
Application part in line with IEC 60601-1 Tread area
Type of protection IP20
Duty cycle Continuous duty

10.2 Weight measurement

Verification in line with Directive 2014/31/EUClass III
Maximum capacity 200 kg
Accuracy
• 0 to 50 kg± 50 g
• 50 kg to 150 kg± 100 g
• 150 kg to 200 kg± 200 g
Graduation
• 0 to 150 kg100 g
• 150 kg to 200 kg200 g

11 OPTIONAL ACCESSORIES

Accessory Article number
seca 415 carry case415 00 00 009
seca 421 carry case421 00 00 009

12 DISPOSAL

12.1 Disposing of the device

SECA 877 - Disposing of the device - 1

Do not dispose of the device in your household waste. The device must be properly disposed of as electronic scrap. Follow your respective national regulations. For more information, please contact seca Service or an authorized service partner.

12.2 Disposing of batteries and rechargeable batteries

SECA 877 - Disposing of batteries and rechargeable batteries - 1

Do not dispose of used batteries and rechargeable batteries in household waste, regardless of whether they contain harmful substances or not. As a consumer, you are legally obliged to dispose of batteries and rechargeable batteries via local authority collection points or trade collection points. Only dispose of batteries and rechargeable batteries once they are completely discharged.

13 WARRANTY

There is a two-year warranty period from delivery for defects attributable to poor materials or workmanship. All movable parts, e.g. batteries, cables, power supply units, rechargeable batteries etc. are exempt. Defects which come under the warranty will be repaired for the customer free of charge against proof of purchase. Additional claims cannot be considered. Costs of transport to and from seca are the responsibility of the customer if the device is located somewhere other than the customer's headquarters. In the event of transport damage, claims under warranty can only be made if the complete original packaging was used for transport and the device was secured and fastened in it according to its originally packaged condition. You should therefore keep all packaging parts.

The warranty will be voided if the device is opened by persons not expressly authorized by seca to do so.

In the event of a warranty issue, please contact your local seca office or the dealer from whom you ordered the product.

14 DECLARATION OF CONFORMITY

C E M24 0102 0123

seca gmbh & co. kg hereby declares that the product complies with the terms of the applicable European directives and regulations. The unabridged declaration of conformity can be found at www.seca.com.

TABLE DES MATIÈRES

1.4 Contre-indications

14 DÉCLARATION DE CONFORMITÉ

C E M24 0102 0123

4.2 Colocar as pilhas

SECA 877 - Colocar as pilhas - 1

ADVARSEL! Fare for at skride

INNEHÄLLSFÖRTECKNING

1 Apparatabeskrivning 170

11 VALINNAISET LISÄTARVIKKEET

1 BESCHRIJVING VAN HET APPARAAT

1.1 Toepassingsdoel

1.4 Contra-indications

WAARSCHUWING! Letsel

11 OPTIONELE ACCESSOIRES

Accessoires Artikelnummer
Draagtas seca 415415 00 00 009
Draagtas seca 421421 00 00 009

12 AFVALVERWIJDERING

Medical Measuring Systems and Scales since 1840

seca gmbh & co. kg

Hammer Steindamm 3-25

22089 Hamburg · Germany

T +49 40 20 00 00 0

F +49 40 20 00 00 50

E info@seca.com

seca operates worldwide with headquarters

in Germany and branches in

seca france

seca united kingdom

seca north america

seca schweiz

seca zhong guo

seca nihon

seca mexico

seca austria

seca polska

seca middle east

seca suomi

seca americalatina

seca asia pacific

seca danmark

seca benelux

seca lietuva

and with exclusive partners in

more than 110 countries.

All contact data at seca.com

Table of contents Cliquez un titre pour y accéder
Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : SECA

Model : 877

Category : Scale