USER MANUAL 877 SECA
Instructions for use
17-10-07-575-100c_2025-02S

Deutsch - Gebrauchsanweisung 3
English-Instructions for use 26
Francais - Mode d'emploi 49
Italiano - Istruzioni per l'uso 73
Espanol - Instrucciones de uso 96
Portugues - Manual de instruções de'utilisation 121
Dansk - Brugsanvising 145
Svenska - Bruksanvisning 169
Norsk - Bruksanvising 192
Suomi-Kayttöohje 215
Nederlands - Gebruiksaanwijzing 238
Elambdavika-Odeltaicxpnoan 261
Česky - Návod k použití 286
INHALTSVERZEICHNIS
1 Description of device 27
1.1 Intended use 27
1.2 User qualification 27
1.3 Patient target group 27
1.4 Contraindications 27
1.5 Clinical benefit 27
1.6 Functional description.... 27
2 Safety information 28
2.1 Safety information in these instructions for use. 28
2.2 Basic safety information... 28
Handling device. 28
Preventing electric shock 30
Preventing injuries and infections 30
Preventing damage to device 31
Handling measuring results 32
Handling packaging material. 32
Handling batteries and rechargeable batteries.. 33
3 Overview 34
3.1 Controls 34
3.2 Symbols in the display .... 34
3.3 Labels 35
4 Starting up device 38
4.1 Scope of delivery 38
4.2 Inserting batteries 38
4.3 Setting up and aligning the device 39
5 Operation 40
5.1 Switching on the device ... 40
5.2 Weighing 40
Weighing a patient. 40
Weighing babies/ toddlers (2 in 1) 41
6 Hygiene treatment 42
6.1 Cleaning 42
6.2 Disinfecting 42
6.3 Sterilizing 43
7 Function check 43
8 Troubleshooting 44
9 Servicing 45
10 Technical data. 45
10.1 General technical data.... 45
10.2 Weight measurement 46
11 Optional accessories. 46
12 Disposal 47
12.1 Disposing of the device .... 47
12.2 Disposing of batteries and rechargeable batteries.... 47
13 Warranty 47
14 Declaration of conformity 48
1 DESCRIPTION OF DEVICE
1.1 Intended use
The electronic flat scale supports physicians in decision-making regarding diagnosis or therapy, based on the patient's weight or weight changes.
1.2 User qualification
The device may only be used by persons with sufficient expertise.
1.3 Patient target group
The device is intended for persons of any age who can stand independently and who do not exceed the maximum weight capacity of the device. Babies and toddlers who cannot stand independently can be weighed using the 2 in 1 function whilst being held by an adult who can stand independently.
1.4 Contraindications
No contraindications are known.
1.5 Clinical benefit
The electronic flat scale supports physicians in decision-making regarding diagnosis or therapy, based on the measured weight.
By that, the electronic flat scale only provides an indirect clinical benefit.
The physician must perform additional examinations for an accurate diagnosis.
1.6 Functional description
Weight on this electronic personal scale is detected by four load cells.
The 2 in 1 function can be used to determine the weight of babies and toddlers. The child is held in an adult's arms during weighing to do this.
2 SAFETY INFORMATION
2.1 Safety information in these instructions for use

DANGER!
Used to identify an extremely hazardous situation. If you fail to take note of this information, serious irreversible or fatal injuries will occur.

WARNING!
Used to identify an extremely hazardous situation. If you fail to take note of this information, serious irreversible or fatal injuries may result.

CAUTION!
Used to identify a hazardous situation. If you fail to take note of this information, minor to moderate injuries may result.
NOTICE!
Used to identify possible incorrect usage of the device. If you fail to take note of this information, you may damage the device, or the measuring results may be incorrect.
NOTE
Includes additional information about use of the device.
2.2 Basic safety information
Handling device
- Please take note of the information in these instructions for use.
- Keep the instructions for use in a safe place. The instructions for use are a component of the device and must be available at all times.
English
In the interest of patient safety, you and your patients are obliged to report serious events that occur in connection with this product to the manufacturer and to the authority responsible in your country.

DANGER!
Risk of explosion
Do not use the device in an environment in which one of the following gases has accumulated:
-oxygen
- flammable anesthetics
- other flammable substances/air mixtures

CAUTION!
Patient hazard, damage to device
Have servicing carried out regularly as described in the relevant section of this document.
Technical modifications may not be made to the device. The device does not contain any user-serviceable parts. Only have servicing and repairs performed by an authorized seca service partner. You can find your local service partner at www.seca.com.
Use only seca original accessories and spare parts, otherwise seca will not grant any warranty.

CAUTION!
Patient hazard, malfunction
- Keep medical electrical devices such as high-frequency surgical equipment at a minimum distance of approx. 1 meter to prevent faulty measurements.
- Keep high-frequency equipment such as cellphones at a minimum distance of approx. 1 meter to prevent faulty measurements.
The actual transmission output of HF devices may require minimum distances of more than 1 meter. For details, go to www.seca.com.
Preventing electric shock

WARNING! Electric shock
- Set up devices which can be operated with a power supply unit so that the power supply socket is within easy reach and the power supply can be disconnected quickly.
- Ensure that your local electricity supply matches the details on the power supply unit.
- Never touch the power supply unit with wet hands.
Do not use extension cables and multiple outlets. - Ensure that cables are not crushed or damaged by sharp edges.
- Ensure that cables do not come into contact with hot objects.
Do not operate the device at an altitude of more than 3000m above sea level.
Preventing injuries and infections

WARNING! Injury from falling
Ensure that the device is steady and level.
- Route connector cables (if present) so that neither users nor the patient can trip over them.
Assist people with limited motor skills when they are getting up, e.g. from a wheelchair.
Ensure that the patient does not step directly onto or off the edges of the weighing platform.
- Ensure that the patient steps onto and off the weighing platform slowly and safely.

WARNING! Danger of slipping
- Ensure that the weighing platform is dry before the patient steps onto it.
- Ensure that the patient has dry feet before stepping onto the weighing platform.
- Ensure that the patient steps onto and off the weighing platform slowly and safely.

WARNING! Risk of infection
Before and after every measurement, wash your hands to reduce the risk of cross-contamination and nosocomial infections.
Should the patient have contagious diseases, subject the device to a hygiene treatment as described in the relevant section of these instructions for use immediately afterwards.
- Ensure that the patient does not have any open wounds or infectious skin alterations which may come into contact with the device.
Subject the device to a hygiene treatment at regular intervals.
Preventing damage to device
NOTICE! Damage to device
- Ensure that fluids never get inside the device. These can destroy the electronics.
For devices with power supply operation: Switch off the device before you disconnect the power supply unit from the power supply socket.
For devices with power supply operation: If the device is not to be used for an extended period, disconnect the power supply unit from the power supply socket. Only then is the device de-energized.
For devices with battery or rechargeable battery operation: If you are not using the device for an extended period of time, remove batteries or rechargeable batteries. Only then is the device de-energized.
Do not drop the device.
Do not subject the device to shocks or vibrations.
Perform a function check before each use as described in the corresponding section in this document. Do not operate the device if it is not working properly or is damaged.
Do not place the device in direct sunlight and ensure that it is not placed in the direct proximity of a heat source. The excessive temperatures could damage the electronics.
- Avoid rapid temperature fluctuations. If the device is transported so that a temperature difference of over 20^ occurs, the device must be left to stand for at least 2 hours before it is switched on, otherwise condensation may form; this may damage the electronics.
Use the device only in the intended ambient conditions outlined in the section entitled "Technical data".
- Store the device only in the intended storage conditions outlined in the section entitled "Technical data".
Use only cleaning agents and disinfectants which match the details in the section entitled "Hygiene treatment".
For scales: Ensure that maximum capacity is not exceeded.
Handling measuring results
NOTICE!
Inconsistent measuring results
Before you save and continue using measured values determined using this device (e.g. in seca software or in an EMR system), ensure that the measured values are plausible.
If measured values have been sent to seca software or to an EMR system, ensure before continuing to use them that the measured values are plausible and assigned to the correct patient.
Handling packaging material

WARNING!
Risk of suffocation
Packaging materials made of plastic film (bags) present a risk of suffocation.
- Store packaging material out of the reach of children.
If the original packaging material is no longer available, only use plastic bags with safety holes to reduce the risk of suffocation. Use recyclable materials if possible.
NOTE
Store the original packaging material for future use (e.g. returning for servicing).
Handling batteries and rechargeable batteries

WARNING! Injury
Batteries and rechargeable batteries contain harmful substances which may explode and so escape if not handled properly.
Do not try to recharge batteries.
Do not expose batteries/rechargeable batteries to heat.
Do not burn batteries/rechargeable batteries.
If harmful substances are leaking out, avoid contact with the skin, eyes, and mucous membranes. Rinse affected areas of the body with plenty of clean water and seek medical help at once.
NOTICE!
Damage to device and malfunction due to incorrect handling
Use only the type of battery/rechargeable battery quoted in this document.
Always replace all batteries/rechargeable batteries at the same time.
Do not short-circuit batteries/rechargeable batteries.
If you do not use the device for an extended period of time, remove the batteries/rechargeable batteries. This prevents acid from leaking into the device.
If harmful substances have penetrated the device, do not continue using it. Have the device checked by an authorized seca service partner and repaired if necessary.
3
OVERVIEW
3.1 Controls

| Item | Control Function | |
| 1 | 2 in 1 key | Starts the 2 in 1 function for weighing babies and tod- dlers |
| 2 | Foot screw 4 pcs, for precise alignment | |
| 3 | Battery compartment | Takes 6 AA-type batteries (AA/1.5 V) |
| 4 | Spirit level Indicates whether the device is horizontal | |
3.2 Symbols in the display

| Item Symbol Meaning | ||
| 1 | kg | Weight value in kilograms |
| 2 | NET Tare function | active |
English
| Item Symbol Meaning | ||
| 3 HOLD Hold function active | ||
| 4 | Non-verifiable function active | |
| 5 | Batteries are low | |
3.3 Labels
| Markings on the device and on the type plate | |
| Text/symbol | Meaning |
| Name and address of manufacturer, date of manufacture | |
| UDI | Unique Device Identification /product identification number as per Directive (EU) 2017/745) |
| REF | Article number |
| SN | Serial number |
| GAL | Value in m/s² (model-dependent) • Gravitational acceleration on earth • Depends on the intended location |
| ProDI | seca product identification number |
| Approval Type | Type designation of design approval |
| Mat.No. | Variant number |
| Follow instructions for use | |
| Medical electrical device, Type B | |
| IP20 | Type of protection to IEC 60529: • Protection against ingress of solid foreign bodies with a diameter of over 12.5 mm • Protection against access with fingers • No protection from ingress of water |
| e | Value in units of mass Used to classify and verify a scale |
| III | Class III scale according to Directive 2014/31/EU |
| CE M24 0102 0123 | Device complies with EU directives •M: Conformity label in compliance with directive 2014/31/EU for non-automatic scales (verified models) •24: (Example: 2024) Year in which declaration of conformity was implemented and the CE label was applied (verified models) •0102: Notified Body for Metrology (verified models) •0123: Notified Body for Medical Devices |
| MD | Medical device in accordance with Regulation (EU) 2017/745 |
| UK CA M24 xxx xxxx | Device complies with United Kingdom directives •M: Conformity label in compliance with UK directive SI 2016 no. 1152 for non-automatic scales (NAWIR) (verified models) •24: (Example: 2024) Year in which declaration of conformity was implemented and the UKCA label was applied (verified models) •xxxx: Notified Body for Medical Devices of the United Kingdom •yyyy: Notified Body for Metrology of the United Kingdom (verified models) |
| UK REP | Importer/representative in the United Kingdom: seca Ltd 40 Barn Street B5 5QB Birmingham United Kingdom |
| CH REP | Importer/representative in Switzerland: seca ag (schweiz) Medizinische Waagen und Messysteme Schönmitt Str. 2 CH-4153 REINACH |
| Do not dispose of device in household waste | |
English
| Markings on the packaging | |
| Symbol Meaning | |
| Protect from moisture | |
| Arrows indicate top of product Transport and store in an upright position | |
| Fragile, do not throw or drop | |
| Permitted min. and max. temperature for transport and storage | |
| Permitted min. and max. humidity for transport and storage | |
| Permitted min. and max. air pressure for transport and storage | |
| Open packaging here | |
| Packaging material can be disposed of through recycling programs | |
4 STARTING UP DEVICE
4.1 Scope of delivery

| Item | Component Pcs. | |
| a | Personal scale 1 | |
| b | Battery 6 | |
| - Instructions for use 1 | ||
4.2 Inserting batteries

You need 6 batteries (type AA/1.5 V, included in scope of delivery).
- Press on the latch of the battery compartment.
- Remove the cover of the battery compartment.
NOTICE!
Damage to device
Device may overheat and become damaged.
-
Ensure the correct polarity of the batteries (markings in the battery holder).
-
Insert the batteries.
NOTE
If bAtt appears in the display, you have inserted one of the batteries the wrong way round or the batteries have discharged.
- Close the battery compartment.
4.3 Setting up and aligning the device

CAUTION!
Faulty measurement as a result of devices being set up incorrectly
Soft floors (wooden boards, for example) give under the patient's weight and falsify the measuring result. Incorrect ambient conditions or lack of alignment lead to measurement errors.
- Select a setup location with a level, stable floor in order to achieve precise measuring results.
Use the device only in the intended ambient conditions outlined in the section entitled "Technical data".
Align the device before using it and every time it changes location.

- Place the device on a firm, level surface.
- Align the device by turning the foot screws so that the air bubble of the spirit level is in the exact center of the circle.
- Ensure that only the feet of the device are in contact with the ground.
5 OPERATION
CAUTION! Personal injury/faulty measurement
Before using the device each time, perform a function check as described in the corresponding section of these instructions for use.
5.1 Switching on the device

The scale has no load.
Tap the scale with your foot.
The scale is operational when 0.0 appears in the display.
The scale switches off automatically 20 seconds after it was last used.
5.2 Weighing
Weighing a patient
The scale has no load.
- If the scale has switched off automatically after the previous measurement, tap the scale with your foot.
NOTICE!
Damage to device
The display is sensitive and may be damaged under load
Make sure that the patient does not step on the display.


- Ask the patient to step onto the scale.
- Read off the measuring result.
English
Weighing babies/ toddlers (2 in 1)
The 2 in 1 function can be used to determine the weight of babies and toddlers. The child is held in an adult's arms during weighing to do this. Proceed as follows:
-
Ensure that there is no load on the scale.
-
Ask the adult to stand on the scale.
-
Read off the measuring result.
-
Keep the 2 in 1 key depressed until the message NET appears in the display.
The weight is saved.
0.0 appears in the display.

#
WARNING!
Faulty measurement as a result of initial weight changing
If the child is weighed together with a different adult, it will not be possible to determine the child's weight correctly.
- Ensure that the child is always weighed with the adult with whom the initial weight was determined.
- Ensure that the adult's weight does not change - as a result of their taking off a piece of clothing, for example.

- Ask the adult to get onto the scale holding the child.
The child's weight is displayed.
The symbol and the messages HOLD and NET are displayed.
- Read off the measuring result.
- Ask the adult to get off the scale with the child.
- You have the following options for deactivating the 2 in 1 function:
Press the 2 in 1 key until the symbol and the messages HOLD and NET are no longer displayed.
Switch off the device.
6 HYGIENE TREATMENT
WARNING! Electric shock
The device is not de-energized when the on/off button is pressed and the display goes out. Use of fluids on the device may cause an electric shock.
Before each hygiene treatment, ensure that the device is switched off.
Take the batteries/battery packs out of the device before every hygiene treatment.
Ensure that no fluids penetrate the device.
WARNING! Risk of infection
Subject the device to a hygiene treatment at regular intervals as described in this section.
NOTICE! Damage to device
Unsuitable cleaning agents and disinfectants may damage the sensitive surfaces of the device and lead to transparent components becoming clouded.
Do not use aggressive or abrasive cleaning agents.
Do not use organic solvents (e.g. white spirit or petroleum spirit).
6.1 Cleaning
If required, moisten a soft cloth with a mild soap solution and wipe the device over with it.
6.2 Disinfecting
- Disinfect the device at regular intervals with a disinfectant suitable for sensitive surfaces and acrylic glass (e.g. 70% ethanol).
- Follow the instructions for use of the disinfectant.
- Disinfect the device as follows:
English
Moisten a soft cloth with disinfectant and wipe down the device with it.
Comply with the intervals, see table:
| Interval Component | |
| Before and after every measure-ment with direct skin contact | Weighing plat-form |
| As required | Display Carry case |
6.3 Sterilizing
The device must not be sterilized.
7 FUNCTION CHECK
Perform a function check before each use.
A complete function check includes:
- A visual inspection for mechanical damage
- A test of the alignment of the device
- A visual and function check of the display elements
- A function check of all the controls shown in the section entitled "Overview"
- A function check of the optional accessories
If you find faults or deviations during the function check, first try to remedy the fault with the aid of the "Troubleshooting" section in this document.

CAUTION!
Personal injury
If you find faults or deviations during the function check which you are unable to remedy with the aid of the "Troubleshooting" section in this document, you must not use the device.
Have the device repaired by seca Service or an authorized service partner.
Follow the section entitled "Servicing" in this document.
8 TROUBLESHOOTING
If faults occur when operating the device, first attempt to remedy them yourself using the following table. If the fault persists, contact seca Service or an authorized service partner.
| Fault Cause Remedy | ||
| The scale does not display a weight although a load was placed on it | Scale is switched off or without batteries | Check that the scale is switched onCheck that batteries are inserted |
| The scale displays a weight following transport or a battery replacement even though no load has been placed on it | Scale has determined an incorrect zero point | Switch off the scale, then switch it back on again |
| 0.0 does not appear before weighing | Scale had load on it before being switched on | Remove the load from the scaleSwitch off the scale, then switch it back on again |
| ---- displayed before weighing | Scale had load on it before being switched on | Remove the load from the scaleSwitch off the scale, then switch it back on again |
| Single segment is illuminated continuously or not at all | Display faulty Inform seca Service | |
| Display: | Battery voltage is dropping Replace the batteries | |
| Display: bAtt | Batteries discharged or not inserted | Replace the batteriesInsert the batteries the right way round |
| Display: StOP | Maximum capacity ex- ceeded | Remove the load from the scale |
| Display: Er + figure | Too high a load on the scale or too high a load in one corner | Remove the load from the scale or distribute the weight more evenlySwitch off the scale, then switch it back on againIf the problem persists, inform seca Service |
9 SERVICING

CAUTION!
Faulty measurements as a result of verification being performed poorly or not at all
Have verification performed only by authorized persons.
Always have verification performed if one or more quality seals are damaged.
seca recommends having your device serviced prior to verification.

CAUTION!
Faulty measurements as a result of poor servicing
Have servicing and repairs carried out exclusively by seca Service or an authorized service partner.
You can find your local service partner at www.seca.com.
10 TECHNICAL DATA
10.1 General technical data
| Dimensions • Depth • Width • Height | 356 mm 321 mm 60 mm |
| Net weight Approx. 4.0 kg/8.8 lbs | |
| Ambient conditions, operation • Temperature • Air pressure • Humidity | +10 °C to +40 °C (50 °F to 104 °F) 700 – 1060 hPa 30 % – 80 %, no condensation |
| Ambient conditions, storage/transport • Temperature • Air pressure • Humidity | -10 °C to +65 °C (14 °F to 149 °F) 700 – 1060 hPa 0 % – 95 %, no condensation |
| Height of digits 20 mm | |
| Power supply • Batteries - Supply voltage - Battery type | 9 V 6 x type AA, 1.5 V, not rechargeable |
| Medical device in accordance with Regulation (EU) 2017/745 | Class I with measuring function |
| EN 60601-1: Medical electrical equipment, type B | |
| Application part in line with IEC 60601-1 Tread area | |
| Type of protection IP20 | |
| Duty cycle Continuous duty | |
10.2 Weight measurement
| Verification in line with Directive 2014/31/EU | Class III |
| Maximum capacity 200 kg | |
| Accuracy | |
| • 0 to 50 kg | ± 50 g |
| • 50 kg to 150 kg | ± 100 g |
| • 150 kg to 200 kg | ± 200 g |
| Graduation | |
| • 0 to 150 kg | 100 g |
| • 150 kg to 200 kg | 200 g |
11 OPTIONAL ACCESSORIES
| Accessory Article number | |
| seca 415 carry case | 415 00 00 009 |
| seca 421 carry case | 421 00 00 009 |
12 DISPOSAL
12.1 Disposing of the device

Do not dispose of the device in your household waste. The device must be properly disposed of as electronic scrap. Follow your respective national regulations. For more information, please contact seca Service or an authorized service partner.
12.2 Disposing of batteries and rechargeable batteries

Do not dispose of used batteries and rechargeable batteries in household waste, regardless of whether they contain harmful substances or not. As a consumer, you are legally obliged to dispose of batteries and rechargeable batteries via local authority collection points or trade collection points. Only dispose of batteries and rechargeable batteries once they are completely discharged.
13 WARRANTY
There is a two-year warranty period from delivery for defects attributable to poor materials or workmanship. All movable parts, e.g. batteries, cables, power supply units, rechargeable batteries etc. are exempt. Defects which come under the warranty will be repaired for the customer free of charge against proof of purchase. Additional claims cannot be considered. Costs of transport to and from seca are the responsibility of the customer if the device is located somewhere other than the customer's headquarters. In the event of transport damage, claims under warranty can only be made if the complete original packaging was used for transport and the device was secured and fastened in it according to its originally packaged condition. You should therefore keep all packaging parts.
The warranty will be voided if the device is opened by persons not expressly authorized by seca to do so.
In the event of a warranty issue, please contact your local seca office or the dealer from whom you ordered the product.
14 DECLARATION OF CONFORMITY
C E M24 0102 0123
seca gmbh & co. kg hereby declares that the product complies with the terms of the applicable European directives and regulations. The unabridged declaration of conformity can be found at www.seca.com.
TABLE DES MATIÈRES
1.4 Contre-indications
14 DÉCLARATION DE CONFORMITÉ
C E M24 0102 0123
4.2 Colocar as pilhas

ADVARSEL! Fare for at skride
INNEHÄLLSFÖRTECKNING
1 Apparatabeskrivning 170
11 VALINNAISET LISÄTARVIKKEET
1 BESCHRIJVING VAN HET APPARAAT
1.1 Toepassingsdoel
1.4 Contra-indications
WAARSCHUWING! Letsel
11 OPTIONELE ACCESSOIRES
| Accessoires Artikelnummer | |
| Draagtas seca 415 | 415 00 00 009 |
| Draagtas seca 421 | 421 00 00 009 |
12 AFVALVERWIJDERING
Medical Measuring Systems and Scales since 1840
seca gmbh & co. kg
Hammer Steindamm 3-25
22089 Hamburg · Germany
T +49 40 20 00 00 0
F +49 40 20 00 00 50
E info@seca.com
seca operates worldwide with headquarters
in Germany and branches in
seca france
seca united kingdom
seca north america
seca schweiz
seca zhong guo
seca nihon
seca mexico
seca austria
seca polska
seca middle east
seca suomi
seca americalatina
seca asia pacific
seca danmark
seca benelux
seca lietuva
and with exclusive partners in
more than 110 countries.
All contact data at seca.com
EasyManual