TECHNOLINE WL 1035 - Climate control

WL 1035 - Climate control TECHNOLINE - Free user manual and instructions

Find the device manual for free WL 1035 TECHNOLINE in PDF.

📄 143 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice TECHNOLINE WL 1035 - page 22
Pick your language and provide your email: we'll send you a specifically translated version.
Product type Indoor air quality monitoring device
Brand TECHNOLINE
Model WL 1035
Power supply 5 V DC, 1.2 A power adapter (model KA06E-0501200EU), 1.8 m USB cable included
Display type Triple digital display for PM2.5, CO2, VOC, with moving graphic indicator
Sensors NDIR for CO2, MEMS for VOC, laser for PM2.5
CO2 measurement range 400 – 5,000 ppm (accuracy ±5% + 50 ppm up to 2,000 ppm)
PM2.5 measurement range 0 – 999 μg/m³ (accuracy <100 μg/m³: ±15 μg/m³, >100 μg/m³: ±15%)
VOC measurement range 0.2 – 9.99 mg/m³ (accuracy ≤0.7 mg/m³: ±0.35 mg/m³, >0.7 mg/m³: ±35%)
Temperature range 0 °C to 50 °C / 32 °F to 122 °F (resolution 0.1 °C)
Relative humidity range 20 – 95% (resolution 1%, accuracy ±5% between 30-70%, ±7% outside range)
Memory Recording of MAX, MIN, AVG values over selectable 8, 12, 24 or 36 hours
Adjustable alerts High and low thresholds for CO2, PM2.5, VOC and relative humidity, with optional audible alarm
Calibration Automatic (over 7 days) and manual possible
Backlight Automatic brightness adjustment (high/medium/low) via optical sensor
Additional indicators Overall air quality level (colors), ventilation recommendation, mold alert (>70% RH) and dryness alert (<30% RH)
Maintenance and cleaning Clean with a soft cloth dampened with water and mild soap; unplug before cleaning; do not use solvents
Safety Do not open the housing; use only the provided adapter; avoid extreme temperatures, direct light, dust, immersion; do not dispose of in fire
Operating temperature 0 to 50 °C, <95% RH non-condensing
Included accessories Power adapter, USB cable, table stand
Reparability No user-serviceable parts; manual calibration recommended every 12 months by a qualified laboratory

Frequently Asked Questions - WL 1035 TECHNOLINE

How to turn on the WL 1035 device?
Plug the power adapter into a wall outlet, then connect the USB cable to the back of the device. The screen turns on automatically showing temperature and humidity. PM2.5, CO2 and VOC measurements appear after about 3 minutes of warm-up.
What to do if the screen stays off?
Check that the power adapter is properly plugged into the outlet and the back of the device. Try another wall outlet. If the problem persists, contact customer service.
How to set the CO2 alert?
Press and hold the ALERT SET button until CO2 flashes. Use the + or - buttons to set the low then high threshold. Press ALERT SET to confirm. The default low threshold is 400 ppm, high is 2000 ppm.
How to activate the audible alarm?
Press the ALARM button once to activate the alarm. The icon appears. When the set high threshold is exceeded, the device emits a continuous beep (faster if the level is very high). Press ALARM again to turn off the sound.
Where to place the device for accurate measurements?
Place the device in a well-ventilated area, away from walls, electronic devices (computer, TV, printer) and direct sources of heat or humidity. Avoid continuously highly polluted environments. Ensure at least 6 hours of ventilation per day for automatic calibration.
What do the colored bars on the screen mean?
Each measurement (PM2.5, CO2, VOC) displays a colored bar graph indicating quality: GOOD (green), NORMAL (yellow), MEDIUM (orange), BAD (red), VERY BAD (purple). The overall level is shown on the right with ventilation recommendation.
How to perform manual calibration?
Place the device in clean air (ideally 400 ppm CO2, 0.2 mg/m³ VOC, <4 μg/m³ PM2.5). Turn on and wait 3 minutes. Press and hold SELECT SENSOR to choose the sensor, then hold CAL until 'CAL' flashes. Wait about 10 minutes until 'CAL' turns off.
How to clean the device?
Unplug the device. Use a soft cloth slightly dampened with water and mild soap. Never use benzene, thinners or abrasive cleaners. Do not immerse the device in water and do not allow moisture to enter the openings.
What is the accuracy of CO2 measurements?
CO2 accuracy is ±5% + 50 ppm up to 2000 ppm. The measurement range is 400 to 5000 ppm with a resolution of 1 ppm. An NDIR sensor is used.
What to do if the CO2 display shows 'Err'?
Unplug the power adapter for about 10 seconds, then plug it back in. If 'Err' persists, the CO2 sensor may be faulty. Contact technical support.

User questions about WL 1035 TECHNOLINE

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Climate control in PDF format for free! Find your manual WL 1035 - TECHNOLINE and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. WL 1035 by TECHNOLINE.

USER MANUAL WL 1035 TECHNOLINE

B4 – “ALERT SET” -Taste

B5 – “ALARM” -Taste

B6 - “+ / MEM” -Taste

C – Rückseite

WL 1035 - Instruction manual Air quality monitor with PM2.5 / CO₂ / TVOC, temperature and relative humidity

Overview
TECHNOLINE WL 1035 - WL 1035 - Instruction manual Air quality monitor with PM2.5 / CO₂ / TVOC, temperature and relative humidity - 1

text_image A1 A2 A3 A4 A5 A6 A7 A8 A9 88°C A10 NORWEL GONNA A11 PEACHY MAX A12 DEAL MIN A13 RED CHEDER AVG. A16 REPOZEN 88 Feinstud nach NRLG MAX BGN RVG. PM2.5 888 CO 8888 PPM 1400 Flichtige Organische Verbindungen TVOC 18.88 Lufung erfender sich Lufung empfehlern LUPTIGUALERT SEAR SCHLECHT SCHLECHT METTEL NORMAL GUT A14 A15 A16 A17 A18 A19

TECHNOLINE WL 1035 - WL 1035 - Instruction manual Air quality monitor with PM2.5 / CO₂ / TVOC, temperature and relative humidity - 2

text_image B1 B2 B3 B4 B5 B6

TECHNOLINE WL 1035 - WL 1035 - Instruction manual Air quality monitor with PM2.5 / CO₂ / TVOC, temperature and relative humidity - 3

text_image techno Technoline WL 1035 Technoprolde Import-Export GmbH Gewerbepark 10 10/45 Willesz. Germany www.technolins.de Verwendende Sie nur dso in der Bodenungsanleitung se fgeformte Neutel / Use only the power adapter based in the user manual / Utilienz unigemert l'adaptaur secteur membrane dans le manuel d'utilisation. C2 C3 C4 C5 C6

A – Front side / Display

A1 – PM2.5 display

A2 - MIN / MAX / AVG icon

A3 - CO _2 display

A4 – High/Low alert icons

A5 – Sound alarm icon

A6 - TVOC display

A7 – Moving chart indicator

A8 – Warning icon (CO _2 > 2000 ppm)

A9 – Temperature display

A10 – Mold alert icon

A11 – Comfort level display

A12 – Dryness alert icon

A13 – Humidity display

A14 – Ventilation recommendation indicator

A15 – Overall air quality indicator

A16 – Humidity high/low alert icons
A17 – PM2.5 concentration level
A18 - CO _2 concentration level
A19 – TVOC concentration level

B – Upper side / Buttons

B1 – Light sensor

B2 - “- / CAL” button

B3 – “SELECT SENSOR” button

B4 – “ALERT SET” button

B5 – “ALARM” button

B6 - “+ / MEM” button

C - Back side

C1 – Sensor vents

C2 – Ventilation vents (for heat release)

C3 – Insert DC jack to the back case

C4 - USB-cable

C5 - AC/DC adapter

C6 – Insert into AC wall socket

Features

  • Large triple display of PM2.5/ CO 2 /TVOC level with moving chart indicator
    • Non-dispersive infrared (NDIR) gas sensor for CO
    2 detection
  • TVOC (Microelectro Mechanical System - MEMS) sensor module to test the content of harmful substance in the air
  • PM2.5 Laser Particle Sensor module for monitor the dust in the air
  • Wide CO2 measuring range from 400 to 5000 PPM
  • PM 2.5 measuring range from 0 to 999~ug / m^3
    • TVOC measuring range from 0.2 to 9.99 mg/m ^3
  • Overall air quality level and ventilation recommendation on basis of PM 2.5/ CO2/ TVOC concentration in the air (coloured display elements)
    • Maximum (MAX), minimum (MIN) and average (AVG) value memory of last 8, 12, 24 or 36 hours selectable

  • Adjustable PM 2.5 / CO2 / TVOC high or low value alert with optional alarm sound

  • Adjustable alert for high or low humidity
    • Automatic baseline calibration
  • Backlight with automatic adjustable brightness levels by light sensor
    • Temperature and relative humidity
  • Mold alert and dryness indicator
    • Built-in memory chip to save settings when power is off
  • Table standing
  • 1.8 m USB cable and 5V\~1.2A adaptor KA06E-0501200EU as power supply (included)

Installation

Plug in the enclosed 5V-1.2A AC/DC adaptor to an AC wall outlet and then the DC jack of the USB cable to the back of your air quality monitor. Connect the free ends of the USB cable with AC/DC power adapter with each other. The display will come on with the relevant temperature and relative humidity. Your PM 2.5/CO 2 /TVOC readings flash with “----”. The relevant data will be displayed in around 180 seconds. It takes 1-2 hours to have all the data of PM 2.5/CO 2 /TVOC, temperature and relative humidity readings initially stabilized.

If the AC power is interrupted, the monitor does not function, the display and the backlight are turned off. The settings of your unit will be kept by the internal memory chip so your monitor will function again after the AC power resumes.

Note:

- For first time use, put the monitor in a well ventilated environment (around lowest possible value of TVOC at 0.22 mg/m ^3 , CO2 at 400 PPM, PM2.5 at 5 ug/m ^3 ) for at least 12

hours to avoid the influence of micro intervening gas inside the package.

Where to place your air quality monitor

  • Make sure the place you put your air quality monitor at is well ventilated for at least 6 hours everyday, or you move the monitor to a well ventilated room overnight to perform Automatic Baseline Calibration.
  • Do not place your monitor in a highly polluted environment for a long time, the accuracy of the sensors will be reduced or damaged. (i.e. avoid TVOC concentration always over 1 mg / m^3 , CO_2 over 1000 ppm, PM2.5 over 100 ug / m^3 ).
  • Please keep the air quality monitor away from a wall surface or any electronics i.e. computer / TV / printer / photocopier as too close to those will affect the readings of PM2.5/CO _2 /TVOC.
  • Avoid contact with organic solvents that include silicon gel and other adhesives, paintings, drugs, oil, and high concentration gases.

Colored bar indication charts

PM 2.5/ CO₂/ TVOC concentration level in the air

Different PM2.5/CO _2 /TVOC concentration level is indicated in coloured bar chart under each reading :

TECHNOLINE WL 1035 - PM 2.5/ CO₂/ TVOC concentration level in the air - 1

bar | CO2 | PPM | PM2.5 | |---|---|---| | GUT | 400~600 | 0~15 | | NORMAL | 601~1000 | 16~35 | | MITTEL | 1001~1500 | 36~54 | | SCHLECHT | 1501~2000 | 55~250 | | SEHR SCHLECHT | over 2000 | over 250 |

TECHNOLINE WL 1035 - PM 2.5/ CO₂/ TVOC concentration level in the air - 2

bar TVOC | mg/m³ | GUT | NORMAL | MITTEL | SCHLECHT | SEHR SCHLECHT | |---|---|---|---|---|---| | | 0.2 ~ 0.3 | 0.31~ 1 | 1.01 ~ 3 | 3.1 ~ 7.0 | over 7.0 | A!

The readings are displayed in the steps: GUT (good), NORMAL (normal), MITTEL (middle), SCHLECHT (bad) and SEHR SCHLECHT (very bad).

The overall air quality concentration level

The overall air quality level is indicated by the coloured chart on the right side of the display together with related ventilation recommendation on the basis of PM2.5/CO _2 /TVOC readings:-

TECHNOLINE WL 1035 - The overall air quality concentration level - 1

The readings are displayed in the steps: GUT (good), NORMAL (normal), MITTEL (middle), SCHLECHT (bad) and SEHR SCHLECHT (very bad).

It also shows the ventilation icon “” if ventilation is recommended “Lüftung empfohlen” or ventilation is needed “Lüftung erforderlich”

The humidity concentration level

The relative humidity is displayed in \% with coloured recommendations:

TECHNOLINE WL 1035 - The humidity concentration level - 1

bar | Category | SEHR FEUCHT (%) | FEUCHT (%) | IDEAL (%) | TROCKEN (%) | SEHR TROCKEN (%) | SCHIMMEL-GEFAHR (Circular Marker) | BEFEUCHTEN (Upward Arrow) | | :--- | :--- | :--- | :--- | :--- | :--- | :--- | :--- | | over 70% | | | | | | | | | 60-70% | | | | | | | | | 40-60% | | | | | | | | | 30-40% | | | | | | | | | <30% | | | | | | | |

The readings are displayed in the steps: SEHR TROCKEN (very dry), TROCKEN (dry), IDEAL (ideal), FEUCHT (wet) and SEHR FEUCHT (very wet)

Mold alert icon “SCHIMMELGEFAHR 📋” will be displayed when the humidity is over 70%.

Dryness alert icon “BEFEUCHTEN 🔊” will be displayed when the humidity is below 30%.

To set high/low alert of CO₂ / PM2.5 / TVOC / relative humidity

- Press and hold the button, the CO

_2 display flashes.

Press the or the button to select PM 2.5 or CO or TVOC or relative humidity display.

2

- When CO _2 display is selected, press the button once, digits “400” (default) flash and appears.

- Press the or the button to adjust the low alert level of CO_2 . Press once to adjust +/- 10 PPM each time. Press and hold to accelerate setting in fast speed. Lowest alert level is 400 PPM.

- Press the button once again, the CO

2 digits "2000"

(default) flash and appears. Press the or the button to adjust the high alert level of CO_2 . Press once to adjust +/- 10 PPM. Press and hold to accelerate setting in fast speed. The highest CO_2 alert level is 5000 PPM.

- Press the button once again to confirm your settings (or no button is pressed in approximate 10 seconds). Your air quality monitor returns to the normal display.

- Repeat the same of above by selecting another sensor to set the high-low alert value :-

PM2.5 : +/- 5 ug/m ^3 (default low alert 0 ug/m ^3 , high alert 250 ug/m ^3 , highest 550 ug/m ^3 )

TVOC : +/- 0.1 mg/m ^3 (default low alert 0.2 mg/m ^3 , high alert 7 mg/m ^3 , highest 9.99 mg/m ^3 )

Relative humidity: +/- 1% (default low alert 35%, high alert 70%, highest 95%)

When the measured PM 2.5 / CO₂ / TVOC / RH% reaches the set high / low alert level, the related ▶ or ▽ icon appears and flashes to remind you.

To use the sound alarm (for high alert)

TECHNOLINE WL 1035 - To use the sound alarm (for high alert) - 1

- Press the button once to turn on the alarm. When the

alarm is on, the icon appears under the CO _2 reading in the middle of the display.

- When the measured PM 2.5, CO _2 or TVOC reaches the set

high alert level, both the 7 and icons flash and an audible beep sound will be given continuously. The beep sound will accelerate when the measured PM 2.5, CO₂ or or TVOC level exceeds more over the set alert level.

If this happens, please open the windows, adjust the air ventilation and avoid too many people gathering together to improve the air quality of your environment.

- Press the button once to disable the voice alarm.

TECHNOLINE WL 1035 - To use the sound alarm (for high alert) - 2

remains flashing if the PM 2.5, CO_2 or TVOC level exceeds the set high alert.

To set the maximum / minimum / average memory time period (default at 12 HRS)

Press and hold the button, digits "12" HRS flash. Press the

TECHNOLINE WL 1035 - To set the maximum / minimum / average memory time period (default at 12 HRS) - 1

or the

TECHNOLINE WL 1035 - To set the maximum / minimum / average memory time period (default at 12 HRS) - 2

button to select 8 HRS, 12 HRS, 24 HRS or

36 HRS duration of memory. The selected duration appears under MEMORY (between the PM2.5 and CO _2 display).

To recall the maximum, minimum, time-weighted average value memory

Press the button one at a time to view the Maximum => Minimum => Average record of the PM 2.5/ CO₂/ TVOC, temperature and RH% (the MAXON icon appears).

After setting the memory duration, the MAX/MIN/AVG. memory data is reset (all digits on the display become “----”) and your monitor will start to store data for the next 24-hour period.

Note:

- You need to power on your monitor for at least 12 hours to display the relevant last 12 hours weighted average of PM 2.5/CO₂/ TVOC/Temperature and humidity level (and also power on 8 hours to display the last 8-hour weight average)

To calibrate the PM2.5/CO2/TVOC sensors

This monitor is factory calibrated to a 400 PPM CO _2 and 0.2mg/m ^3 TVOC as standard concentration.

Automatic baseline calibration

Your monitor is equipped with a smart chip so that it will automatically calibrate itself to reduce the drift of the CO_2 /TVOC sensor for long time usage. It will calibrate itself automatically at the

minimum CO_2 / TVOC reading detected during 7 days of continuous monitoring.

Make sure the place you put your monitor at is well ventilated (with a CO_2 level near 400 PPM or TVOC near 0.2 mg/m ^3 ) for at least an hour or so everyday, or you move the monitor to such a well ventilated room for an hour or so everyday to perform Automatic Baseline Calibration.

Note:

- Please try not to place your monitor in an environment with a TVOC concentration higher than 1mg/m^3 for a long time. Otherwise, the sensitivity of the TVOC sensor will be reduced or the TVOC sensor will be damaged.

Manual calibration of the PM2.5/CO _2 /TVOC sensors

It is recommended to undergo manual calibration performed by a qualified laboratory every 12 months.

This must be done in a well ventilated room or an environment with clean air (ideally in a room showing lowest value of CO_2 at 400 PPM, TVOC at 0.2mg/m^3 , PM2.5 below 4 ug/m^3 ).

Caution : Do not calibrate the monitor in an environment of unknown CO_2 concentration.

  • Power on your air quality monitor it on and wait for 3 minutes until the first PM2.5/CO _2 /TVOC readings are displayed.
  • Then press and hold the button, only CO 2 display flashes. Press the or the button repeatedly to select. PM2.5, CO 2 or TVOC is displayed. Then press and hold the CAL button until “CAL” flashes.

- In around 10 minutes, the flashing “CAL” will stop and the calibration is completed. Your air quality monitor will return to normal display mode.

- To stop the manual calibration anytime, press the button once.

TECHNOLINE WL 1035 - Manual calibration of the PM2.5/CO _2 /TVOC sensors - 1

Light Sensor

There is a light sensor at the top left side of your air quality monitor. The brightness of the display is automatically adjusted to HI/ MED/LO level according to the surrounding light environment.

Note:

- Do not place any object on the top of the light sensor or cover it in any way, as this would affect the light sensing function.

Troubleshooting

No display or functions

Check the AC/DC adaptor is properly connected.

Response time

Ensure the air flow channels at the back of the monitor are not blocked or covered.

CO _2 reading shows “Err”

The CO_2 sensor is not working properly. Unplug the AC/DC adapter (C8) for 10 seconds to insert it to your monitor again.

Care and maintenance

  • Place your monitor on a stable surface, away from sources of direct sunlight or excessive heat or moisture.
  • Clean your monitor with a soft cloth moistened only with mild soap and water. Stronger agents such as Benzine, thinner or similar materials can damage the surface of the item. Make sure the monitor is unplugged before cleaning.
  • Do not open the back case or tamper with any components of this monitor.

- Do not plug in any other AC/AC or AC/DC adaptor with incorrect specifications or voltage.

CO _2 levels and guidelines

Below are some CO_2 reference levels in general:

400\~600 ppm Background (normal) outdoor air level.

600\~1,000 ppm Typical level for occupied spaces with good air exchange.

1,000 \~1,500 ppm Start ventilation of the room.

1,500\~2,000 ppm Level associated with complaints of drowsiness and poor air. Ventilation recommended!

2,000\~5,000 ppm Level associated with headaches, sleepiness, and stagnant, stale, stuffy air. Poor concentration, loss of attention, increased heart rate and slight nausea may also be present. Ventilation strongly recommended!

5,000 ppm Exposure may lead to serious oxygen deprivation resulting in permanent brain damage, coma, and death. Requires immediate ventilation!

PM2.5 air quality guideline

Fine particulate matter (PM2.5) is an air pollutant that is a concern for people's health when levels in air are high. PM2.5 are tiny particles in the air that reduce visibility and cause the air to appear hazy when levels are elevated.

The WHO updated the new air quality guideline (AQG) levels in 2021 from evidence-based data and information. The highest recommended average annual emission level for PM2.5 is 5 g/m^3 and the 24-hour level is 15 g/m^3 .

Total volatile organic compounds (TVOC) levels

The TVOC covers a large range of different organic gases that may be chemically similar and difficult to distinguish (e.g. Benzene,

Formaldehyde, Napthalene, Styrene, Trichloroethylene, Tetrachloroethylene, Toluene).

The TVOC is considered an important indicator for indoor hygiene and indoor air quality (IAQ). In addition to serious health concerns, there is the psychological aspect: homes, offices, and other environments that smell clean typically seem more welcoming than areas with foul odors caused by organic compounds.

Reference LevelAir informationActionTVOC (mg/m3)Air quality
Level 1Clean hygienic air (target value)No action required.< 0.3Very good
Level 2Good air qualityVentilation recommended0.3 – 1.0Normal
Level 3Noticeable comfort concerns (Not recommended for exposure > 12 months)Ventilation recommended. Identify sources.1.0 – 3.0Medium
Level 4Significant comfort issues (Not recommended for exposure > 1 month)Refresh air whenever possible. Increase ventilation. Search for source is required.3.0 – 7.0Poor
Level 5Unacceptable conditions (not recommended)Breath only if unavoidable and only for short periods.> 7.0Bad

Declaration of conformity

Hereby, Technotrade declares that this product WL 1035 is in compliance with the essential requirements and other relevant

provisions of Directive 2014/30/EU and ROHS 2011/65/EC. The original EU declaration of conformity may be found at: www.technoline.de/doc/4029665410357

Specifications

CO_2

Measurement range : 400 – 5000 PPM

Resolution : 1 PPM

Accuracy : +/-5% + 50 PPM (up to 2000 PPM)

Warm up time : 3 minutes

PM 2.5

Measurement range : 0 – 999 ug/m ^3

Resolution : 1 ug/m ^3

Accuracy : <100 ug/m

100 ug/m ^3 (+/- 15 ug/m ^3 ), ^3 (+/- 15% reading)

Warm up time : 3 minutes

TVOC

Measurement range : 0.2 - 9.99 mg/m ^3

Resolution : 0.01 mg/m

Accuracy : ≤0.7mg/m

0.7 mg/m ^3 +/-0.35 mg/m ^3 , ^3 +/-35% (up to 3mg/m ^3 )

Warm up time : 3 minutes

Temperature range : 0°C to 50°C / 32°F to 122°F

Temperature resolution : 0.1°C / 0.2°F

Relative Humidity range : 20% - 95%

Relative Humidity resolution : 1%

Accuracy : 30% \~ 70% +/-5%; out of these ranges

+/-7% RH

Operating temperature : 0°C to 50°C / 32°F to 122°F,

< 95% RH non-condensing

Out of range display

CO_2 : 400 (below 400 PPM) /

HI (over 5000 PPM)

RH% alert : Low - 35%, High - 70%

Voice alarm : Off

Temperature : °C

Max-Min-Average value

Memory duration : 12 HRS

Precautions

  • The item is intended to be used only indoors.
  • Do not subject the item to excessive force or shock.
  • Do not expose the item to extreme temperatures, direct sunlight, dust or humidity.
  • Do not immerse in water.
  • Avoid contact with any corrosive materials.
  • Do not dispose this item in a fire as it may explode.
  • Do not open the inner back case or tamper with any components of this item.

Use the product only for its intended purpose!

Consideration of duty according to the law of electrical devices

TECHNOLINE WL 1035 - Precautions - 1

This symbol means that you must dispose of electrical devices separated from the General household waste when it reaches the end of its useful life. Take your item to your local waste collection point or recycling centre. This applies to all countries

of the European Union, and to other European countries with a separate waste collection system.

natural_image Symbolic icon of a person walking with arrows, enclosed in a circle (no text or symbols)
natural_image Three black warning symbols: triangular warning triangle, house with circular pattern, and triangular warning triangle (no text or labels)

B4 – “ALERT SET” toets

B5 – “ALARM” toets

B6 - “+ / MEM” toets

C – Achterzijde

C1 – Sensoropeningen

Total volatile organic compounds (TVOC) niveaus

A4 – Ikony alertów High/Low

B6 – Przycisk "+ / MEM"

C - Tył

bar | CO2 | GUT | NORMAL | MITTEL | SCHLECHT | SEHR SCHLECHT | |---|---|---|---|---|---| | PPM | 400~600 | 601~1000 | 1001~1500 | 1501~2000 | over 2000 | Error bar indicates a significant deviation from the main data point. An exclamation mark symbol appears near the top right of the chart. The label 'P' likely denotes an error or warning. The number 'I' is shown at the end of the chart.

TECHNOLINE WL 1035 - Total volatile organic compounds (TVOC) niveaus - 1

bar | Group | Condition | Value | | :--- | :--- | :--- | | PM2.5 ug/m³ | GUT | 0 ~ 15 | | PM2.5 ug/m³ | NORMAL | 16 ~ 35 | | PM2.5 ug/m³ | MITTEL | 36 ~ 54 | | PM2.5 ug/m³ | SCHLECHT | 55 ~ 250 | | PM2.5 ug/m³ | SEHR SCHLECHT | over 250 | | TVOC mg/m³ | GUT | 0.2 ~ 0.3 | | TVOC mg/m³ | NORMAL | 0.31 ~ 1 | | TVOC mg/m³ | MITTEL | 1.01 ~ 3 | | TVOC mg/m³ | SCHLECHT | 3.1 ~ 7.0 | | TVOC mg/m³ | SEHR SCHLECHT | over 7.0 |
Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : TECHNOLINE

Model : WL 1035

Category : Climate control