BRAUN YK-81CEU - Pulse oximeter

YK-81CEU - Pulse oximeter BRAUN - Free user manual and instructions

Find the device manual for free YK-81CEU BRAUN in PDF.

📄 84 pages English EN Download 💬 AI Question
Notice BRAUN YK-81CEU - page 3
Pick your language and provide your email: we'll send you a specifically translated version.

User questions about YK-81CEU BRAUN

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Pulse oximeter in PDF format for free! Find your manual YK-81CEU - BRAUN and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. YK-81CEU by BRAUN.

USER MANUAL YK-81CEU BRAUN

Not made with natural rubber latex

Enthält keinen

Naturkautschuk

Ikke fremstilte at
naturgimmulfatd.

Thank you for purchasing the Braun Pulse oximeter, YK-81C. This pulse oximeter is a high-quality product tested in accordance with international ISO safety and performance standards. This device uses red and infrared light at set wavelengths along with a photodetector to measure oxygen saturation in blood and the associated pulse rate.

The Braun Pulse oximeter is intended to be used by consumers in a home-use setting (such as home setting or places of business).

Please read these instructions carefully before using this product and keep the instructions and the pulse oximeter in a safe place.

Indications for use

The Braun Pulse oximeter (YK-81C) is a non-sterile, reusable device intended for spot checking or intermittent determination of functional oxygen saturation in arterial blood, as well as the associated pulse rate by using the tip of the finger as the measurement site. The device is only intended for adults, in a home-use environment (such as a home setting or places of business).

The Pulse oximeter displays the % SpO2 , pulse rate, as well as the pulse amplitude using a vertical bar graph.

This device is factory calibrated and no further calibration is required by the user.

BRAUN YK-81CEU - Indications for use - 1

WARNING AND PRECAUTIONS

Please read the manual carefully before use.

Device isn't suitable for continuous monitoring.

Pulse oximeters are sensitive to motion. Keep hands still while taking a reading.

Pulse oximeters require sufficient blood flow to obtain proper readings. Poor blood circulation can result in inaccurate readings. If your hands are cold or if you have poor circulation, warm your hands by rubbing them together or use another warming method before attempting to obtain a reading. A tourniquet, blood pressure cuff, other devices or clothing that cause blood flow restrictions may also result in inaccurate readings.

Fingernail polish or acrylic nails obstruct the light transmission through the fingertip and may also result in inaccurate readings.

Your finger must be clean for a proper reading.

The Pulse oximeter should be clean in order to get a proper reading. Please ensure the finger clip area is cleaned, following the cleaning instruction in the manual.

If a reading is difficult to obtain, switch to another finger or to the other hand. The index or middle finger is recommended for taking measurements.

There are a number of other conditions which may lead to an inaccurate reading; including but not limited to, recent medical tests that included an injection of dyes, use of arterial catheters, a weak pulse, low levels of hemoglobin in the blood, low perfusion (the quality of your pulse), elevated levels of dysfunctional hemoglobin, the strength and type of light that you are in while using the pulse oximeter, and the existence of cell phones, radios, and fixed transmitters within certain ranges of the pulse oximeter during use.

This device does not provide a low SpO2 alarm.

DO NOT use on the same finger continuously for over 2 hours.

The device is not appropriate for the constant monitoring of the user. Prolonged and continuous monitoring may increase the potential of unexpected changes of dermal conditions such as putrescence, etc. A feeling of discomfort or pain may occur if using the device continuously, especially for those users who experience poor perfusion or immature dermographia by light collimation.

In addition to the items described in the Warnings and precautions section, inaccurate measurements may be caused by factors including but not limited to:

  • Immersing the device in liquid.
    Significant levels of dysfunctional hemoglobin (such as carboxyhemoglobin or methemoglobin).
  • Intravascular dyes such as indocyanine green or methylene blue.
  • Measurements may be adversely affected in the presence of high ambient light such as direct sunlight. In bright light conditions, cover the sensor area if necessary.
  • Excessive user movement.
    High-frequency electrosurgical interference and defibrillators.
    Venous pulsations.
  • Hypotension, severe vasoconstriction, severe anemia, or hypothermia.

DO NOT modify this equipment without the authorization of the manufacturer.

Fingertip pulse oximetry

General description

Oxygen mixes with blood in the lungs and is transported in blood mainly by hemoglobin. Chemically the oxygen molecule combines loosely and reversibly with the heme portion of the hemoglobin. The amount of oxygen that combines with the hemoglobin is determined by the partial pressure of oxygen. The amount of oxygen in blood, expressed as a percentage of hemoglobin saturation is a measure of oxygen (SaO2) in the artery and is estimated by a pulse oximeter as an SpO2 reading.

The amount of oxygen in blood (as measured by the pulse oximeter SpO2 reading) is a very important physiological parameter in respiratory and circulatory systems. Many respiratory diseases reduce the oxygen-hemoglobin saturation in human blood. Moreover, factors such as Automatic Organic Regulation Malfunction caused by anesthesia, trauma resulting from major surgery or medical examination, or other diseases due to virus or bacteria can also cause a reduction in oxygen hemoglobin saturation in arterial blood which can be detected by the pulse oximeter through a low SpO2 reading. This can result in symptoms such as depression, vomiting, and asthenia in patients.

The fingertip pulse oximeter is a compact, low-power consumption, portable device. When the index finger is inserted into the fingertip photoelectric sensor, the display screen of the Braun Pulse oximeter will show the measured value of % SpO2 and pulse rate as well as a vertical bar graph showing the pulse amplitude.

Measurement principle

Pulse oximetry is based on the difference in the selective absorption of red light by oxygenated blood or oxyhemoglobin (HbO2, which is bright red) as compared to deoxyhemoglobin (Hb, which is dull red) or de-oxygenated blood. This selective absorption is based on the Beer Lambert Law. A red (660nm) and infrared LED (940nm) on the top portion of the prospective clamp finger-type sensor emits light at these two frequencies. The red light transmitted through the finger is selectively absorbed by the oxygenated arterial blood and the remaining light is detected by a photodetector on the lower portion of the clamp finger. The infrared light then passes through the finger without any selective absorption and is detected by the photodetector on the lower side of the clamp. The transmitted light captured by the photodetector is converted into a voltage or current signal and then passed through signal conditioning and filtering circuits using a microprocessor. % SpO2 is then derived by analyzing the changes in the amplitude of these pulsatile signals derived from the photodetector and calibrating them using pulse scanning and recording technology. The pulse rate is then derived from these associated waveforms.

Product description

  1. OLED display
  2. Power/Menu button

  3. Finger clip

  4. Lanyard

How to use your Braun Pulse oximeter

Installing and replacing batteries (initial set up)

Push on the edge of the battery cover behind the arrows to slide it from the device.

BRAUN YK-81CEU - Installing and replacing batteries (initial set up) - 1

Install 2 AAA batteries.

BRAUN YK-81CEU - Installing and replacing batteries (initial set up) - 2

Replace battery cover.

BRAUN YK-81CEU - Installing and replacing batteries (initial set up) - 3

Taking a measurement

Press on top and bottom housing to open rubber finger clip.

BRAUN YK-81CEU - Taking a measurement - 1

Insert finger with nail facing up into rubber finger clip.

It is recommended that you use the oximeter on your index finger.

BRAUN YK-81CEU - Taking a measurement - 2

Make sure you finger is fully inserted into the device and not pressed hard into a surface.

Press Power/Menu button on the top of the device.

BRAUN YK-81CEU - Taking a measurement - 3

Keep finger steady while taking a measurement.

BRAUN YK-81CEU - Taking a measurement - 4

Rotating the display

If you want to change the display direction, press and release (approx. 1 second) the Power/Menu button on the front panel. The display can be viewed in 6 different orientations.

BRAUN YK-81CEU - Rotating the display - 1

Care and cleaning

Use 70% alcohol to clean the pulse oximeter finger insert area and your finger, before and after each test.

BRAUN YK-81CEU - Care and cleaning - 1

Parameter set up

Press the Power/Menu button to turn on the device. With the device turned on, press and hold the Power/Menu button to change the display mode.

Press and hold the Power/Menu button (approx. > 3seconds) until you

BRAUN YK-81CEU - Parameter set up - 1

enter the parameter setup menu 1 and hold down again (approx. 3 seconds) to enter screen 2.

BRAUN YK-81CEU - Parameter set up - 2

BRAUN YK-81CEU - Parameter set up - 3

  1. In menu 1 or 2 press the Power/Menu button to cycle through the various parameters. You will see the * move to each parameter as you cycle through. When you are at the parameter you wish to change, press the Power/Menu button for 3 seconds to change the parameter setting.
  2. If you wish to restore to factory settings, move the * to the restore selection on menu 1 and hold the Power/Menu button down for approximately 3 seconds.
  3. To exit the menus and return to the measurement screen, move the *to the exit selection on menu 1 or 2 and hold the Power/Menu button for 3 seconds, or simply wait 10-12 seconds without interacting with device and it will return to the measurement screen.

The measure Attention

If using device outdoors or under strong light, please increase the screen brightness to improve legibility.

Choose a lower screen brightness to conserve battery power.

Limited 2 year warranty

A. This 2 year limited warranty applies to repair or replacement of product found to be defective in material or workmanship. This warranty does not apply to damage resulting from commercial, abusive, unreasonable use or supplemental damage. Defects that are the result of normal wear and tear will not be considered manufacturing defects under this warranty. KAZ IS NOT LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY NATURE, ANY IMPLIED WARRANTY OF MERCHANT ABILITIES OR FITNESS FOR A PARTICULAR PURPOSE ON THIS PRODUCT IS LIMITED IN DURATION TO THE DURATION OF THIS WARRANTY. Some jurisdictions do not allow the exclusion or limitation of incidental or consequential damages or limitations on how long an implied warranty lasts, so the above limitations or exclusions may not apply to you. This warranty gives you specific legal rights and you may also have other rights which vary from jurisdiction to jurisdiction. This warranty applies only to the original purchaser of this product from the original date of purchase. B. At its option, Kaz will repair or replace this product if it is found to be defective in material or workmanship. C. This warranty does not cover damage resulting from any unauthorized attempts to repair or from any use not in accordance with the instruction manual. Refer to page 158 for list of European Contact numbers. Please be sure to specify a model number.

NOTE: IF YOU EXPERIENCE A PROBLEM, PLEASE CONTACT CONSUMER RELATIONS FIRST OR SEE YOUR WARRANTY. DO NOT RETURN THE PRODUCT TO THE ORIGINAL PLACE OF PURCHASE. DO NOT ATTEMPT TO OPEN THE DEVICE HOUSING YOURSELF, DOING SO MAY VOID YOUR WARRANTY AND CAUSE DAMAGE TO THE PRODUCT OR PERSONAL INJURY.

Errors and troubleshooting

Error messageSituationSolution
SpO2 or PR can not be shown normallyFinger is not fully inserted into deviceRetry by correctly inserting the finger
User's oxyhemoglobin value is too low to be measuredTry a few more timesIf you can make sure that there is no problem in the product, please go to the hospital in a timely manner for exact diagnosis
SpO2 or PR reading is unsteadyFinger is not fully inserted into device User is moving or finger is shakingRetry by correctly inserting the finger Remain as still as possible
The oximeter can not be turned onInadequate power or power off Batteries are installed incorrectly The oximeter might be damagedReplace the batteries Please reinstall the batteries Please contact the customer service center
Unit turned offThe product automatically shuts off when no signal can be detected within 8 seconds Inadequate powerNormal Replace the batteries
Low battery displayedBatteries need to be replacedReplace the batteries

Product specifications

Type: Braun Pulse oximeter, YK-81C

Display Type: OLED display

SpO2 Measurement range: 70% - 100%

Accuracy: 80% - 100% ± 2%

70% - 79% ± 3%

Resolution: 1%

Pulse Rate

Measurement range: 30BPM\~254 BPM

Accuracy: ≤ 100± 1 BPM

> 100 ± 2BPM

Resolution: 1 BPM

Power Supply: two AAA 1.5V alkaline batteries

Power Consumption: below 30mAh

Automatically Power-off: The product automatically shuts off when no signal can be detected within 8 seconds

Dimension: Approx. 58mm× 35mm× 30mm

Operation Temperature: 5^ 40^ (41^ 104^)

Storage Temperature: -10°C~40°C (14°F ~ 104°F)

Operation Humidity: 15%~80%

Storage Humidity: 10%~95%

Air Pressure: 70kPa~106kPa

This device conforms to the following standards: IEC 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety, and essential performance. IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests.

IEC 60601-1-11: Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

ISO 80601-2-61: Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.

ISO 10993-1 Biological evaluation of medical devices- Part 1:

Evaluation and testing within a risk management process.

ISO 10993-5: Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity.

ISO 10993-10: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

ISO 15223-1: Medical devices - Symbols to be used with medical device labels, labeling and information to be supplied - Part 1: General requirements.

ISO 14971 - Medical Devices - Application Of Risk Management To Medical Devices

IEC 62366 - Medical devices - Part 1: Application of usability engineering to medical devices

Maintenance and storage

  1. Replace the batteries when low voltage symbol is lit.
  2. Clean the surface of the oximeter before it is used to take a measurement.
  3. Remove the batteries if you will not use the oximeter for an extended amount of time.
  4. For product preservation store the device in -10 40^ (14-104°F) and 10% -95% humidity.
  5. It is recommended that the product should be kept dry at all times. Excessive humidity might affect its lifetime and even damage the product.
  6. This device is factory calibrated and no further calibration is required by the user.

BRAUN YK-81CEU - Maintenance and storage - 1

To protect the environment, dispose of empty batteries at appropriate collection sites according to national or local regulations.

Box content

One lanyard

One user manual

2 AAA 1.5V alkaline batteries

Symbols and definitions

BRAUN YK-81CEU - Symbols and definitions - 1

Type BF Applied Part

BRAUN YK-81CEU - Symbols and definitions - 2

Operating Humidity

BRAUN YK-81CEU - Symbols and definitions - 3

Operating Temperature

BRAUN YK-81CEU - Symbols and definitions - 4

Consult Instructions for Use

BRAUN YK-81CEU - Symbols and definitions - 5

Batch Code

BRAUN YK-81CEU - Symbols and definitions - 6

Serial Number

BRAUN YK-81CEU - Symbols and definitions - 7

Medica Device

BRAUN YK-81CEU - Symbols and definitions - 8

Catalogue Number

BRAUN YK-81CEU - Symbols and definitions - 9

Manufacturer

BRAUN YK-81CEU - Symbols and definitions - 10

Voltage

BRAUN YK-81CEU - Symbols and definitions - 11

Storage Humidity

BRAUN YK-81CEU - Symbols and definitions - 12

Storage temperature

BRAUN YK-81CEU - Symbols and definitions - 13

Authorized representative in the European Community

IP22

IP22: Protected against solid foreign objects greater than 12.5mm in diameter and dripping water when tilted at 15^

DE Deutschland

Rotation of displayet

IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests. (Sahköset laäketieteelliset laitteet - Osa 1-2: Yleiset

ISO 10993-10: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization (ISO 10993-10:

WAARSCHUWINGEN EN VOORZORGSGMAATREGELEN

aill is yolssil jus sll uaiu wabi.2

jaiy jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii jaii

.

104-14) 40\~10- jj jjj jjj jjj jjj.4

4.8% 95 -%.10%(

a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a

ailll 1i iii i jll lglc ilil

j 10000000000000000000000000000000000000000000000

aaiiaaiiie 1

aoll

S APQ = S AQP + S_ PQQ

AAA1.5

jzglg jzgl

BRAUN YK-81CEU - jzglg jzgl - 1

EF 5

BRAUN YK-81CEU - jzglg jzgl - 2

jill 1

BRAUN YK-81CEU - jzglg jzgl - 3

jssssssssssssssssssssss

BRAUN YK-81CEU - jzglg jzgl - 4

a#do

BRAUN YK-81CEU - jzglg jzgl - 5

Jlalll

BRAUN YK-81CEU - jzglg jzgl - 6

BRAUN YK-81CEU - jzglg jzgl - 1

COS < AN = 12

BRAUN YK-81CEU - jzglg jzgl - 2

aasss

BRAUN YK-81CEU - jzglg jzgl - 3

m = 311

BRAUN YK-81CEU - jzglg jzgl - 4

jzjjzaiy

BRAUN YK-81CEU - jzglg jzgl - 5

j 1

BRAUN YK-81CEU - jzglg jzgl - 6

(1) a + b = 1 (2) a + b = 3 (3) a + b = 4

IP22

iip22

ailll 1e all 12.5 le la 12

YK-81C, 000000000000000000000000000000000000000

OLED

auii 5

SpO2": 70% - 100%"

2± 100 - 80

7.3±79-7.70

%1

a254\~a30

jaiill Joo joi jbi

a 100

+ = 2

a2+100<

a 1

AA 1.5

a + b = 1

yj jie jie jixi loic jilali jlll lai

:Jtai Jtaiill

8J

30×35×58

2k + 1 = 3

\~41) 40-5

| x - 2| = 1| x| · | r1^2 + r2^2| = 1| x| · | r1| · | r2| < 35r1r2

104

\~(14)a240~a240-10

:j

104

7.80 7.15 : 社会经济

95\~10:

JSSU 106\~JSU 70:glb

aaiiiee aiee iieae

aLd aal dall:1eJ-aaJySlaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaIaI

a2111c1211a2111

aallll:2-1e jj-aaee 60601-60601-1

  • aayablaaegsll yjaiy1: jayssll jayall - pjjll 1

adalll:11-1jgj-aaallll:EN60601-1-11

aulll lalal jaiy cllb: gail jall - pJdI y aalw alal

aJglll aaglll 5e jall - aalb s jauu uuuuauuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuuUU

.

  1. jj- 4a 1000000000000000000000000000000000000000000000000000000000

jzlllclalao golodilai gill jzll - aialll jzll:ISO 15223-1

alalll1jll - 1a jgi gill clogale, ciaalall, iiall

aalb jyj1 100 - aalb jyj1 - ISO 14971

aulls jyj yle pduuul yauuua uus - aulls jyj - IEC 62366

ailll lalall jll aiall /aiall jge biaol 21 aiall .1

iic jsiLioic. iDx Jiaill eii jiae Jg Jg sJg sJg

j3aalallalall jle baiiai iyy

Aaleal 120jy

aaii i 1

j1j3aiaiallaii 2j1iaial

aaii 10-12 aaii jaiii

.

gbsjy j g j y aeg eel i ci j jllal gall gall plaiaal all g

a

AaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaBaaa

juiuui jiuai jiuai jiuai

gill glll lglal g 100e aae e ae aee aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aae aee

aalalalalalalalalalalalalalalalalalal

a a a a a a a a a a a a a a a a a a a

dIgai jiaiiaizao 2020 Uaiaaa jia gao aaiyai yjia jiai iiaii gaiiu

aalalal alalal y aalalal y aalalal

aannnnnne aennnnnne nnnnnnne annnnnne

jlll jlll jlll jlll jlll jlll jlll jlll

BRAUN YK-81CEU - juiuui jiuai jiuai jiuai - 1

jaii jaiy gaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i jaiy i

al jall aall yaiy iiaa aiaaa aaii aaii aaii jay jaij
jia cial llaa gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy gaiy

Cullill 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 125 12

Jilu w jilun

Jy 1000000000000000000000000000000000000000000

Juslllall 3j j j j j j j j j j j j j j j j j j j j j j j j j j j j j j

jlll jil 1200000000000000000000000000000000000000000

Abjellipssusill 5s

.

j j 1111111111111111111111

Aaiinll 5d jn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nannn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn nnnn nn

e 1

#

a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a a

.5j 1111111111111111111111111111

e 154 154 154 154 154 154 154 154 154 154 154 154 154 154 154 154 154 154

SE

Svenska

Support contact numbers

AustriaIreland (The Republic Of)Slovakia
π +43 (0)1 360 277 1225π +353 (0)1 525 1809π +421 2 5011 2115
BahrainItalySlovenia
π +973 17582250π +39 02 3859 1183π +386 (01) 888 86 74
Belgium/LuxemburgJordanSouth Africa
π +32 (0)2 620 01 01π +962 6 582 0112-3-4-5π +27 (0) 11 844 6190
BosniaKuwaitSpain
π +387 (33) 902 911π +965 24833274π +34 913 754 176
BulgariaLebanonSweden
π +35 929 358 684π +961 (01) 512002π +46 (0)8 5199 3097
CroatiaNetherlandsSwitzerland
π +385 (01) 3444 856π +31 (0) 78 201 8001π +41 (0) 22 567 5200
CyprusNorwayTunisia
π +357 22575016π +47 23 5001 20π +216 71716880
Cyprus Northern regionPolandTurkey
π +0392 22 72 367π +48 22 512 39 02π +90 216 337 22 55
Czech RepublicPortugalUAE
π +420 (0) 2 25 43 97 69π +351 21 060 8045π +971 4 353 4506
DenmarkQatarUnited Kingdom/ Northern Ireland
π +45 35 15 80 40π +974 4 4075048
π +974 4 4075000π +974 4 4075000π +44 (0) 207 949 0115
Finland
π +358 (0)9 81 71 00 14Romania
Franceπ +40 214 255 566
π +33 (0)1 85 14 80 95Saudi-Arabia
Germanyπ +966 3 8692244
π +49 (0)21 173 749 003Western Region - Jeddah
Greeceπ +966 (0) 22565555
π +30 211 180 94 56Central Region - Riyadh
Hungaryπ +966 (0) 12886808
π +36 (06)1 429 2216Eastern Region - Al-Khobar
Icelandπ +966 (0) 38940555
π +354 555 3100South Region - Abha
Israelπ +966 (0) 72376062
π +1 800250221

BRAUN YK-81CEU - Support contact numbers - 1

Xuzhou Yongkang Electronic Science Technology CO., Ltd

4F C8 Building, No. 40 Jingshan Road Economic Development District, Xuzhou, Jiangsu, China

BRAUN YK-81CEU - Support contact numbers - 2

BRAUN YK-81CEU - Support contact numbers - 3

Prolix GmbH

Brehmstr.56,40239

Duesseldorf-Germany

BRAUN YK-81CEU - Support contact numbers - 4

© 2021 All Rights Reserved.

Imported and distributed by:

Kaz Europe Sarl

A Helen of Troy Company

Place Chaudron 18

CH-1003 Lausanne - Switzerland

www.helenofroy.com/emea-en/

BRAUN YK-81CEU - Support contact numbers - 5

Printed in China

© 2021 All rights reserved.

REF YK-81C

P/N:A005511R1

04FEB21

Helen of Troy Creative Department Artwork Specifications
Brand: Braun Category: Oximeter Model: YK-81CEU
Artwork Part #: A005511R1
Die Line Part #: NA
Subject: Owners Manual Region: EMEA
Flat Size: MM: W160 x H154
Folded Size: MM: W80 x H154
Scale: 1/1 Material: 50# matte text
Page count: 164 Revision: 3 Date: 04FEB21
Release Date: 15DEC20 Rerelease Date: 05FEB21
Colors: Dielines (Do not print) Spot Colors Cyan 0% Magenta 0% Yellow 0% Black 100% GFE-63-1-4U PMS PMS XXXX
Special Instructions:
Quality Requirement of Artwork and Quality Clarification Process of Artwork Printing: Meet Eng-QS-06&02
Helen of Troy Creative Services Marlborough, MA 01752 USA +1508 490 7000
Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : BRAUN

Model : YK-81CEU

Category : Pulse oximeter