MICROLIFE BP A10030 - Blood pressure monitor

BP A10030 - Blood pressure monitor MICROLIFE - Free user manual and instructions

Find the device manual for free BP A10030 MICROLIFE in PDF.

📄 103 pages English EN Download 💬 AI Question
Notice MICROLIFE BP A10030 - page 3
Pick your language and provide your email: we'll send you a specifically translated version.

User questions about BP A10030 MICROLIFE

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Blood pressure monitor in PDF format for free! Find your manual BP A10030 - MICROLIFE and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. BP A10030 by MICROLIFE.

USER MANUAL BP A10030 MICROLIFE

Europe / Middle-East / Africa

Microlite AG

Espenstrasse 139

9443 Winneau / Switzerland

Tel.+41/71/23/030

Fax-41/71/27/039

Email admin@microlife.ch

www.ni-----life.com

Asia

Microlide Corporation.

9F, 431, RuiGang Road, NaiHu

Taipei, 114, Taiwan, R.O.C.

Tel. +B86 2 B797-1288

Fax -B86 2 E797-1283

Email service@microlife.com.tw

www.micralife.com

North / Central / South America

Micralia USA, Inc

1617 Gulf to Bay Blvd., 2nd Floor Ste A

Clearwater, FL 33755 / USA

1cl.+1 727 442 5363

Fax-1 727 442 5377

Email msa@microlifeusa.com

www.microlife.com

C€0044

MICROLIFE BP A10030 - North / Central / South America - 1

natural_image Portrait of a woman with long hair against a dark blue background (no text or symbols visible)

Microlife BP A100
MICROLIFE BP A10030 - North / Central / South America - 2

text_image EN→1 FR→6 ES→12 PT→18 DE→24 RU→30 PL→36 HU→42 BG→48 RO→54 CZ→60 SK→66 SL→72 TR→78 GR→84 AR→90 FA→96

MICROLIFE BP A10030 - North / Central / South America - 3

natural_image Medical device with digital display showing blood pressure measurement (no visible text or symbols)

microlife

1BB A102-30V17-1 5215

Microlife BP A100

MICROLIFE BP A10030 - Microlife BP A100 - 1

text_image ①⑧⑨⑩⑪ ②⑥⑦③ ④⑤

MICROLIFE BP A10030 - Microlife BP A100 - 2

text_image Technical diagram of a blood pressure monitor with labeled parts and screen display

MICROLIFE BP A10030 - Microlife BP A100 - 3

text_image 129 78 65 10 11 17 16 14 13 M 12 12 15

MICROLIFE BP A10030 - Microlife BP A100 - 4

flowchart
graph TD
    A["Device 1: 100%"] --> B["Device 2: 75%"]
    A --> C["Device 3: 60%"]
    A --> D["Device 4: 50%"]
    A --> E["Device 5: 40%"]
    A --> F["Device 6: 30%"]
    A --> G["Device 7: 20%"]
    A --> H["Device 8: 10%"]
    A --> I["Device 9: 5%"]
    A --> J["Device 10: 0%"]

microlife

Guarantee Card BP A100

Name of Purchaser / Nom de l'achetsur / Nombre del comprador / Nome do comprador / Name des Kälfers / Ф.И.О. покупателя / Imię i nazwisko nabywycy / Väsárto neve / Име на купувача / Numele cumpäratorului / Jméno kupujícího / Meno zákaznika / Ime in primek kupca / Alicinn Adi / Оворотмінуvu оворотні

① ON/OFF button
② Display
③ Slot-in Card
④ Cuff Socket
⑤ Mains Adapter Socket
⑥ Cuff Compartment
⑦ Battery Compartment
⑧ Cuff
⑨ Cuff Connector
⑱ M-button (memory)

Display

⑩ Systolic Value
⑪ Diastolic Value
⑫ Pulse Rate
⑬ Heart Arrhythmia Indicator
⑭ Pulse
⑮ Battery Display
⑯ Stored Value
⑰ Memory Number

MICROLIFE BP A10030 - Display - 1

Read the instructions carefully before using this device.

MICROLIFE BP A10030 - Display - 2

Type BF applied part

MICROLIFE BP A10030 - Display - 3

Keep dry

Dear Customer,

Your new Microlife blood pressure monitor is a reliable medical device for taking measurements on the upper arm. It is simple to use, accurate and comes highly recommended for blood pressure monitoring in your home. This device was developed in collaboration with physicians and clinical tests carried out prove its measurement accuracy to be of a very high standard.*

Please read through these instructions carefully so that you understand all functions and safety information. We want you to be happy with your Microlife product. If you have any questions, problems or want to order spare parts please contact your local Microlife-Customer Service. Your dealer or pharmacy will be able to give you the address of the Microlife dealer in your country. Alternatively, visit the internet at www.microlife.com where you will find a wealth of invaluable information on our products.

Stay healthy – Microlife AG!

* This device uses the same measuring technology as the award winning «BP 3BTO-A» model tested according to the British Hypertension Society (BHS) protocol.

Table of Contents

  1. Important Facts about Blood Pressure and Self-Measurement

• How do I evaluate my blood pressure?

  1. Using the Device for the First Time

  2. Activating the fitted batteries

  3. Selecting the correct cuff

  4. Taking a Blood Pressure Measurement using this Device

  5. Appearance of the Heart Arrhythmia Indicator for early Detection
  6. Data Memory

• Viewing the stored values
- Memory full
- Clearing all values

  1. Replacing the Slot-in Card

  2. Battery Indicator and Battery change

  3. Low battery

  4. Flat battery – replacement
  5. Which batteries and which procedure?
    • Using rechargeable batteries

  6. Using a Mains Adapter

  7. Error Messages

10. Safety, Care, Accuracy Test and Disposal

  • Safety and protection
  • Device care
  • Cleaning the cuff
  • Accuracy test
  • Disposal

11. Guarantee

12. Technical Specifications

Guarantee Card (see Back Cover)

1. Important Facts about Blood Pressure and Self-Measurement

  • Blood pressure is the pressure of the blood flowing in the arteries generated by the pumping of the heart. Two values, the systolic (upper) value and the diastolic (lower) value, are always measured.
  • The device indicates the pulse rate (the number of times the heart beats in a minute).
  • Permanently high blood pressure values can damage your health and must be treated by your doctor!
  • Always discuss your values with your doctor and tell him/her if you have noticed anything unusual or feel unsure. Never rely on single blood pressure readings.
  • Enter your readings in the enclosed blood pressure diary. This will give your doctor a quick overview.
  • There are many causes of excessively high blood pressure values. Your doctor will explain them in more detail and offer treatment where appropriate. Besides medication, relaxation techniques, weight loss and exercise can also lower your blood pressure.
  • Under no circumstances should you alter the dosages of any drugs prescribed by your doctor!
  • Depending on physical exertion and condition, blood pressure is subject to wide fluctuations as the day progresses. You should therefore take your measurements in the same quiet conditions and when you feel relaxed! Take at least two measurements per day, one in the morning and one in the evening.
  • It is quite normal for two measurements taken in quick succession to produce significantly different results.
  • Deviations between measurements taken by your doctor or in the pharmacy and those taken at home are quite normal, as these situations are completely different.
  • Several measurements provide a much clearer picture than just one single measurement.

  • Leave a small break of at least 15 seconds between two measurements.

  • If you are pregnant, you should monitor your blood pressure very closely as it can change drastically during this time!
  • If you suffer from an irregular heartbeat (arrhythmia, see «Section 4.»), measurements taken with this device should only be evaluated after consultation with your doctor.

- The pulse display is not suitable for checking the frequency of heart pacemakers!

How do I evaluate my blood pressure?

Table for classifying blood pressure values in adults in accordance with the World Health Organisation (WHO) in 2003. Data in mmHg.

Range Systolic DiastolicRecommendation
blood pressure too low 100 60 Consult your doctor
1.blood pressure optimum100 - 12060 - 80Self-check
2.blood pressure normal120 - 13080 - 85Self-check
3.blood pressure slightly high130 - 14085 - 90Consult your doctor
4.blood pressure too high140 - 16090 - 100Seek medical advice
5.blood pressure far too high160 - 180100 - 110Seek medical advice
6.blood pressure dangerously high180 110 Urgently seek medical advice!

The higher value is the one that determines the evaluation.

Example: a readout value between 150/85 or 120/98 mmHg indicates «blood pressure too high».

The slot-in card ③ on the front of the device shows ranges 1-6 in the Table.

2. Using the Device for the First Time

Activating the fitted batteries

Pull out the protective strip from the battery compartment ⑦.

Selecting the correct cuff

Microlife offers 3 different cuff sizes: S, M and L. Select the cuff size to match the circumference of your upper arms (measured by close fitting in the centre of the upper arm). M is the correct size for most people.

Cuff size for circumference of upper arm
S 17 - 22 cm (6.75 - 8.75 inches)
M 22 - 32 cm (8.75 - 12.5 inches)
L 32 - 42 cm (12.5 - 16.5 inches)

Only use Microlife cuffs.

▶ Contact your local Microlife Service if the enclosed cuff ⑧ does not fit.
Connect the cuff to the device by inserting the cuff connector ⑨ into the cuff socket ④ as far as it will go.

3. Taking a Blood Pressure Measurement using this Device

Checklist for taking a reliable measurement

  1. Avoid activity, eating or smoking immediately before the measurement.
  2. Sit down for at least 5 minutes before the measurement and relax.
  3. Always measure on the same arm (normally left).
  4. Remove close-fitting garments from the upper arm. To avoid constriction, shirt sleeves should not be rolled up - they do not interfere with the cuff if they are laid flat.

  5. Always ensure that the correct cuff size is used (marking on the cuff).

  6. Fit the cuff closely, but not too tight.

  7. Make sure that the cuff is positioned 2 cm above the elbow.
  8. The artery mark located on the cuff (ca. 3 cm long bar) must lie over the artery which runs down the inner side of the arm.
    ●Support your arm so it is relaxed.
  9. Ensure that the cuff is at the same height as your heart.

  10. Press the ON/OFF button ① to start the measurement.

  11. The cuff will now pump up automatically. Relax, do not move and do not tense your arm muscles until the measurement result is displayed. Breathe normally and do not talk.

  12. When the correct pressure is reached, the pumping stops and the pressure falls gradually. If the required pressure was not reached, the device will automatically pump some more air into the cuff.

  13. During the measurement, the heart symbol ⑭ flashes in the display and a beep sounds every time a heartbeat is detected.

  14. The result, comprising the systolic ⑩ and the diastolic ⑪ blood pressure and the pulse rate ⑫ is displayed and a long beep is heard. Note also the explanations on further display symbols in this booklet.

  15. When the measurement has finished, remove the cuff and pack it into the device as shown in Fig. II.

  16. Enter the result in the enclosed blood pressure pass and switch off the device. (The monitor does switch off automatically after approx. 1 min.).

You can stop the measurement at any time by pressing the ON/OFF button (e.g. if you feel uneasy or an unpleasant pressure sensation).

4. Appearance of the Heart Arrhythmia Indicator for early Detection

This symbol ^13 indicates that certain pulse irregularities were detected during the measurement. In this case, the result may deviate from your normal blood pressure – repeat the measurement. In most cases, this is no cause for concern. However, if the symbol appears on a regular basis (e.g. several times a week with measurements taken daily) we advise you to tell your doctor. Please show your doctor the following explanation:

Information for the doctor on frequent appearance of the Arrhythmia indicator

This device is an oscillometric blood pressure monitor that also analyses pulse frequency during measurement. The device is clinically tested.

The arrhythmia symbol is displayed after the measurement, if pulse irregularities occur during measurement. If the symbol appears more frequently (e.g. several times per week on measurements performed daily) we recommend the patient to seek medical advice.

This device does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.

5. Data Memory

At the end of each measurement, this device automatically stores the result.

Viewing the stored values

Press the M-button ⑱ briefly, when the device is switched off. The display first shows the last stored result.

Pressing the M-button again displays the previous value. Pressing the M-button repeatedly enables you to move from one stored value to another.

Memory full

When the memory has stored 30 results, the memory is full From this point onwards, a new measured value is stored by overwriting the oldest value.

Clearing all values

All memory data are cleared when batteries are taken out of the device.

6. Replacing the Slot-in Card

You can replace the slot-in card ③ by pulling it out to the side, as shown in Fig. IV and replacing the paper insert.

It may be helpful to have your doctor note down the medication dosage or an emergency telephone number on the card. Extra cards are supplied with the device for this purpose.

7. Battery Indicator and Battery change

Low battery

When the batteries are approximately 34 empty the battery symbol ⑮ will flash as soon as the device is switched on (partly filled battery displayed). Although the device will continue to measure reliably, you should obtain replacement batteries.

Flat battery – replacement

When the batteries are flat, the battery symbol ⑮ will flash as soon as the device is switched on (flat battery displayed). You cannot take any further measurements and must replace the batteries.

  1. Open the battery compartment ⑦ at the back of the device by pushing inwards at the two arrows and pulling out the battery compartment cover.
  2. Replace the batteries – ensure correct polarity as shown by the symbols in the compartment.

Which batteries and which procedure?

Use 4 new, long-life 1.5 V, size AA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a prolonged period.

Using rechargeable batteries

You can also operate this device using rechargeable batteries.

Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat battery symbol appears. They should not remain inside the device as they may become damaged (total discharge as a result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor. Recharge batteries in an external charger and observe the information regarding charging, care and durability.

8. Using a Mains Adapter

You can operate this device using the Microlife mains adapter (DC 6V, 600 mA).

Only use the Microlife mains adapter available as an original accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are damaged.

  1. Plug the adapter cable into the mains adapter socket ⑤ in the blood pressure monitor.
  2. Plug the adapter plug into the wall socket.

When the mains adapter is connected, no battery current is consumed.

9. Error Messages

If an error occurs during the measurement, the measurement is interrupted and an error message, e.g. «ERR 3», is displayed.

ErrorDescriptionPotential cause and remedy
«ERR 1»Signal too weakThe pulse signals on the cuff are too weak. Re-position the cuff and repeat the measurement.*
«ERR 2»Error signal During the measurement, error signals were detected by the cuff, caused for instance by movement or muscle tension. Repeat the measurement, keeping your arm still.
«ERR 3»No pressure in the cuffAn adequate pressure cannot be generated in the cuff. A leak may have occurred. Check that the cuff is correctly connected and is not too loose. Replace the batteries if necessary. Repeat the measurement.
«ERR 5»Abnormal resultThe measuring signals are inaccurate and no result can therefore be displayed. Read through the checklist for performing reliable measurements and then repeat the measurement.*
«HI»Pulse or cuff pressure too highThe pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over 200 beats per minute). Relax for 5 minutes and repeat the measurement.*
«LO»Pulse too lowThe pulse is too low (less than 40 beats per minute). Repeat the measurement.*

* Please consult your doctor, if this or any other problem occurs repeatedly.
If you think the results are unusual, please read through the information in «Section 1.» carefully.

10. Safety, Care, Accuracy Test and Disposal

MICROLIFE BP A10030 - Safety, Care, Accuracy Test and Disposal - 1

Safety and protection

  • This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by incorrect application.
  • This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in the «Technical Specifications» section.

- Protect it from:

  • water and moisture
  • extreme temperatures
  • impact and dropping
  • contamination and dust
  • direct sunlight
  • heat and cold

  • The cuffs are sensitive and must be handled with care.

  • Only pump up the cuff once fitted.
  • The function of this device may be compromised when used close to strong electromagnetic fields such as mobile phones or radio installations and we recommend a distance of at least 1 m. In cases where you suspect this to be unavoidable, please verify if the device is working properly before use.
  • Do not use this device if you think it is damaged or notice anything unusual.
  • Never open this device.
  • If the device is not going to be used for a prolonged period the batteries should be removed.
  • Read the additional safety information provided within the individual sections of this instruction manual.

MICROLIFE BP A10030 - Safety and protection - 1

Ensure that children do not use this device unsupervised; some parts are small enough to be swallowed. Be aware of the risk of strangulation in case this device is supplied with cables or tubes.

Device care

Clean the device only with a soft, dry cloth.

Cleaning the cuff

You can machine wash the cuff cover at 30^ C (do not iron!).

MICROLIFE BP A10030 - Cleaning the cuff - 1

WARNING: Under no circumstances, however, may you wash the inner bladder! Always remove the sensitive bladder from the sleeve before washing and replace it carefully again afterwards.

Accuracy test

We recommend this device is tested for accuracy every 2 years or after mechanical impact (e.g. being dropped). Please contact your local Microlife-Service to arrange the test (see foreword).

Disposal

MICROLIFE BP A10030 - Disposal - 1

Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with domestic waste.

11.Guarantee

This device is covered by a 5 year guarantee from the date of purchase. The guarantee is valid only on presentation of the guarantee card completed by the dealer (see back) confirming date of purchase or the receipt.

  • Batteries, cuff and parts that become worn with use are not included.
  • Opening or altering the device invalidates the guarantee.
  • The guarantee does not cover damage caused by improper handling, discharged batteries, accidents or non-compliance with the operating instructions.

Please contact your local Microlife-Service (see foreword).

12. Technical Specifications

Operating conditions: 10 - 40 °C / 50 - 104 °F 15 - 95 % relative maximum humidity

Storage conditions: -20 - +55 °C / -4 - +131 °F 15 - 95 % relative maximum humidity

Weight: 610 g (including batteries)

Dimensions: 160 x 125 x 98 mm

Measuring procedure: oscillometric, corresponding to Korotkoff method: Phase I systolic, Phase V diastolic

Measurement range: 20 - 280 mmHg – blood pressure 40 - 200 beats per minute – pulse

Cuff pressure display

range: 0 - 299 mmHg

Resolution: 1 mmHg

Static accuracy: pressure within ± 3 mmHg

Pulse accuracy: ± 5 % of the readout value

Voltage source: 4 x 1.5 V alkaline batteries; size AA Mains adapter DC 6V, 600 mA (optional)

Battery lifetime: approx. 920 measurements (using new batteries)

IP Class: IP20

Reference to standards: EN 1060-1 /-3 /-4; IEC 60601-1; IEC 60601-1-2 (EMC); IEC 60601-1-11

Expected service life: Device: 5 years or 10000 measurements Accessories: 2 years

This device complies with the requirements of the Medical Device Directive 93/42/EEC.

Technical alterations reserved.

Microlife BP A100

FR

Poids: 610 g (piles incluses)

Dimensions: 160 x 125 x 98 mm

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : MICROLIFE

Model : BP A10030

Category : Blood pressure monitor