USER MANUAL CM P1 BENNING
TRUE RMS Digital Current Clamp Meter for - AC current measurement (BENNING CM P1) - AC/ DC current measurement (BENNING CM P2)Table of contents1. User notes2. Safety note3. Scope of delivery4. Unit description5. General information6. Ambient conditions7. Electrical specifications8. Measuring with the BENNING CM P1/ P29. Maintenance10. Environmental note1. User notes These operating instructions are intended for - qualified electricians and - electrotechnically trained persons. The BENNING CM P1/ P2 is intended for making measurements in dry environment. It must not be used in power circuits with a nominal voltage higher than CAT III 600 V (More details in Section 6. “Ambient conditions”). The following symbols are used in these operating instructions and on the BENNING CM P1/ P2: Application around and removal from HAZARDOUS LIVE conductors is permitted. This symbol indicates an electrical hazard. This symbol indicates sources of danger when using the BENNING CM P1/ P2 (see documentation). This symbol on the BENNING CM P1/ P2 indicates that the unit is protection insulated (safety class II).This symbol appears in the display for a discharged batteryDC voltage or current AC current or voltage Earth (voltage to earth)11/ 2018
The instrument is built and tested in accordance with DIN VDE 0411 part 1/ EN 61010-1 DIN VDE 0411 part 2-032/ EN 61010-2-032 and has left the factory in perfectly safe technical condition. To maintain this condition and to ensure safe operation of the unit, the user must observe the notes and warnings given in these instructions at all times. Improper handling and non-obser- vance of the warnings might involve severe injuries or danger to life.
WARNING! Be extremely careful when working with bare conductors or main line carrier! Contact with live conductors will cause an electric shock!
Before starting the unit, always check it for signs of damage. Should it appear that safe operation of the unit is no longer possible, it should be shut down immediately and secured to prevent that it is switched on accidentally. It may be assumed that safe operation is no longer possible: - if the device exhibit visible damages, - if the unit no longer works, - after long periods of storage under unfavourable conditions, - after being subject to rough transportation, or - if the device is exposed to moisture.
3. Scope of delivery
The following items make up the standard BENNING CM P1/ P2 package:
3.3 Two 1.5 V micro-batteries (IEC LR03/ AAA)
3.4 One Operating Manual
Note on consumable parts: - The BENNING CM P1/ P2 is supplied by two 1.5 V micro-batteries (IEC LR 03/ AAA)
See figure 1: Appliance front face The display and operating elements shown in fig. 1 are as follows:
Measurement prongs, for inserting and gripping the single conductor containing current
NCV sensor, detector of the voltage indicator,
Prong guard, protects user from accidental contact with conductor
Opening lever, for opening and closing the current prongs
H/ INRUSH key (blue), storage of the displayed measured value (HOLD), inrush current measurements (A AC)
NCV key, voltage indicator for determining the AC voltage to earth,
A key, for measuring the AC current (BENNING CM P1) and the AC/ DC current (BENNING CM P2),
ON/OFF key, for switching the current clamp on/off, enables the digital display illumination,
Digital display, for displaying the measured value and range exceedance
Battery compartment cover, at the rear of the housing11/ 2018
is designed as a 4 digit liquid-crystal indicator with 10 mm digit height and decimal point. The highest value displayed is 4200.
works automatically. Only a polarity con- trary to the technical current direction is indicated with “-” (see arrow on the measuring clamp).
5.1.3 In case of a range exceedance (overflow), “0L.” is displayed.
Attention: no display or warning by complete overload.
5.1.4 The BENNING CM P1/P2 is switched on or off by pressing the ON/OFF key
. To switch it off, press and hold the key
for approx. 2 seconds. If the BENNING CM P1/ P2 is switched on, the display illumination can be enabled by pressing the key
. It is switched off by pressing the key again or automatically after approx. 10 seconds.
5.1.5 The selection of the measuring range is automatic.
has two functions: Pressing the H/INRUSH key
stores the measuring result. The “HOLD” symbol simul- taneously appears on the display
again to switch the device back to measuring mode. In measuring mode “A AC”, press the H/INRUSH key
for approx. 2 seconds to enable the inrush current measurement. At the same time, the digital display
shows the “INRUSH” symbol. Here, the measuring process is initiated for 100 milliseconds after a current (> 5 A) is applied. The averaged value of this time period is displayed. Press the key
for approx. 2 seconds again to switch the device back to normal measuring mode.
has two functions: Pressing the ∆/LPF key
enables the low-pass filter with a limiting frequency of approx. 160 Hz. At the same time, the digital display
shows the “LPF” symbol. Press the key
again to switch the device back to the normal mode. In measuring mode “A DC” (BENNING CM P2), press the ∆/LPF key
for approx. 2 seconds to carry out a null balance. It can also be used for differential measurement (null balance is possible for any value!). This is shown by means of the “∆” symbol on the digital display
again to switch the device back to the normal mode.
Pressing the NCV key
enables the voltage indicator function (localization of AC volt- ages to earth). At the same time, the digital display
shows the „ “ and “EF” symbols (see 8.3). Press the A key
to switch the device back to the current measuring range.
to select the current measuring range. With the BENNING CM P2, you can toggle between the current types “DC” and “AC” by pressing the key
5.1.10 The measuring rate of the BENNING CM P1/ P2 amounts nominally to 2 measurements
per second for the digital display.
5.1.11 The BENNING CM P1/ P2 switches off automatically after approx. 15 minutes (APO,
Auto-Power-Off is activated, if the symbol is shown on the display
). The auto- matic switch-off can be disabled by pressing the H/INRUSH key
and by switching on the BENNING CM P1/P2 via the ON/OFF key
. The symbol disappears from the display
5.1.12 The BENNING CM P1/ P2 is supplied by two 1.5 V batteries (IEC LR03/ AAA/ micro).
5.1.13 If the battery voltage drops below the specified operating voltage of the BENNING CM
P1/ P2, then a battery symbol appears in the display
5.1.14 The battery life depends on the measuring function used and is approx. 200 hours for
the BENNING CM P1 and approx. 30 hours for the BENNING CM P2 without using the11/ 2018
background lighting.
5.1.15 Temperature coefficient of the measured value: 0.1 x (stated measuring accuracy)/ °C
< 18 °C or > 28 °C, related to the value for the reference temperature of 23 °C
5.1.16 Dimensions of unit (length x width x height) = 149 x 59 x 27,5 mm.
Weight of unit: 140 g (incl. batteries)
5.1.17 Widest prong opening: 23 mm
6. Ambient conditions
- The BENNING CM P1/ P2 is intended for making measurements in dry environment. - Maximum barometric elevation for making measurements: 2000 m, - Overvoltage category: IEC 60664/ IEC 61010 → 600 V category III - Contamination class: 2 (EN 61010-1), - Protection class: IP 30 (DIN VDE 0470-1, IEC/ EN 60529) IP 30 means: Protection against access to dangerous parts and protection against solid impurities of a diameter > 2.5 mm, (3 - first index). No protection against water, (0 - second index). - Operating temperature and relative humidity: For operating temperatures from 0 °C to 40 °C: relative air humidity lower than 80 %, non- condensing - Storage temperature: The BENNING CM P1/ P2 can be stored at temperatures between - 10 °C and + 60 °C, at a relative air humidity lower than 70 % without batteries.
7. Electrical specifications
Note: The measuring precision is specified as the sum of - a relative fraction of the measured value and - a number of digits (counting steps of the least significant digit). This specified measuring precision is valid for temperatures in the range from 23 °C ± 5 °C and relative humidity less than 80 %.
7.1 AC current range
The measured value is obtained and displayed as real r.m.s. value (True RMS, AC coupling). Its calibration is adapted to sinusoidal curves. In case of deviations from this curve shape, the accuracy of the displayed value decreases. Crest factor < 1.6 up to 100 % of each final measuring range value Crest factor < 3.2 up to 50 % of each final measuring range value Low-pass filter deactivated: Measuring range Resolution Measurement accuracy (50 Hz - 60 Hz) Measurement accuracy (45 Hz - 400 Hz) Overload protection
Low-pass filter activated: Measuring range Resolution Measurement accuracy (50 Hz - 60 Hz) Overload protection
Inrush current: Measuring range Resolution Measurement accuracy (50 Hz - 60 Hz) Overload protection
5.0 A - 400 A 0.1 A ± (3.5 % of reading + 8 digits) 420 A eff.
Integration time: 100 ms Threshold value of the trigger current: 5 A
7.2 DC current range (BENNING CM P2)
Measuring range Resolution Measurement accuracy Overload protection
8. Measuring with the BENNING CM P1/ P2
8.1 Preparing the measurement
Operate and store the BENNING CM P1/ P2 at the specified storage and operating tempera- tures only! Do not permanently expose the device to sunlight. - Strong sources of interference in the vicinity of the BENNING CM P1/ P2 might involve unstable readings and measuring errors.
8.2 Current measurement
Do not exceed the maximum permitted voltage with respect to earth potential! Electrical danger!
For direct current (DC) measurements (BENNING CM P2), observe correct polarity! The arrow on the measuring clamp indicates the technical current direction. + → – - After switching it on, the BENNING CM P1/P2 is automatically in function A. Otherwise, press the A key
to switch the BENNING CM P2 to the initial mode. - Operate opening lever
single wire live conductor centrally by means of the BENNING CM P1/ P2 current probe. - Read off the digital display unit
See figure 2: Current measurement
The voltage indicator function is not intended for testing the absence of volt- age. Even without an indication or acoustic signal, a dangerous contact volt- age might be applied. Electrical danger! The voltage indicator function is intended for non-contact detection of an alternating field. The detector
is located on the measuring clamp within the NCV marking. Switch the BENNING CM P1/P2 on and select the voltage indicator function by pressing the NCV key
shows the „ “ and “EF” symbols. If a phase voltage is localized, an acoustic signal will be emitted and the intensity of the alternating field will be indicated on the digital display
by a maximum of 4 bars “- - - -”. An indication is made in earthed AC current networks only!11/ 2018
Practical hint: Interruptions (cable breaks) in cables lying around openly such as e.g. cable reels, fairy lights etc. can be traced from the feeding point (phase) to the point of interruption. Functional range: ≥ 230 V See figure 3: Voltage indicator with buzzer
Before opening the BENNING CM P1/ P2, make sure that it is free of voltage! Electrical danger!
9.1 Securing the instrument
Under certain circumstances safe operation of the BENNING CM P1/ P2 is no longer ensured, for example in the case of: - Visible damage of the casing. - Incorrect measurement results. - Recognisable consequences of prolonged storage under improper conditions. - Recognisable consequences of extraordinary transportation stress. In such cases the BENNING CM P1/ P2 must be switched off immediately, disconnected from the measuring points and secured to prevent further utilisation.
Clean the casing externally with a clean dry cloth (exception: special cleaning wipers). Avoid using solvents and/ or scouring agents for cleaning the instrument. It is important to make sure that the battery compartment and battery contacts are not contaminated by leaking electrolyte. If electrolyte contamination or white deposits are present in the region of the batteries or battery casing, clean them too with a dry cloth.
9.3 Battery replacement
Before opening the BENNING CM P1/ P2, make sure that it is free of voltage! Electrical danger! The BENNING CM P1/ P2 is supplied by means of two 1.5 V batteries of type AAA (IEC LR03). Battery replacement is required, if the battery symbol appears on the display
. When the BENNING CM P1/P2 is switched on, a battery test is performed. Proceed as follows to replace the batteries: - Switch the BENNING CM P1/P2 off. - Lay the BENNING CM P1/ P2 face down and release the screws of the battery compartment cover. - Lift the battery compartment lid (in the housing recess area) from the bottom section. - Remove the discharged batteries from the battery compartment. - Insert the new batteries into the battery compartment observing correct polarity. - Place the battery compartment cover onto the bottom part and tighten the screw. See figure 4: Battery replacement
Make your contribution to environmental protection! Do not dispose of dis- charged batteries in the household garbage. Instead, take them to a collect- ing point for discharged batteries and special waste material. Please inform yourself in your community.11/ 2018
Benning guarantees compliance with the technical and accuracy specifications stated in the operating manual for the first 12 months after the delivery date. To maintain accuracy of the measuring results, the device must be recalibrated in regular intervals by our factory service. We recommend recalibrating the device once a year. For this purpose, send the device to the following address: Benning Elektrotechnik & Elektronik GmbH & Co. KG Service Center Robert-Bosch-Str. 20 D - 46397 Bocholt