Geratherm Oxy One - Measuring equipment

Oxy One - Measuring equipment Geratherm - Free user manual and instructions

Find the device manual for free Oxy One Geratherm in PDF.

📄 148 pages English EN Download 💬 AI Question
Notice Geratherm Oxy One - page 29
Pick your language and provide your email: we'll send you a specifically translated version.

User questions about Oxy One Geratherm

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Measuring equipment in PDF format for free! Find your manual Oxy One - Geratherm and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. Oxy One by Geratherm.

USER MANUAL Oxy One Geratherm

Instructions for Use

Mode d'emploi

Note: The illustration used in this manual may differ slightly from the appearance of the actual product.

natural_image Simple line drawing of an open battery casing with internal circuitry (no text or symbols)

Geratherm Oxy One - 1

natural_image Hand placing a small square object into a device with a checkmark nearby (no text or symbols)

Geratherm Oxy One - 2

natural_image Hand inserting a blood glucose meter into a digital device, with a cross symbol nearby (no text or labels)

Anmerkungen:

scatter | Mittelwert von SpO₂ und SaO₂ | SpO₂ Abweichung | | ---------------------------- | --------------- | | 65 | -4 | | 70 | -2 | | 75 | 0 | | 80 | 2 | | 85 | 4 | | 90 | 6 | | 95 | 4 | | 100 | 2 | | 105 | 0 | | 110 | -2 | | 115 | -4 | | 120 | -6 | | 125 | -8 | | 130 | -6 | | 135 | -4 | | 140 | -2 | | 145 | 0 | | 150 | 2 | | 155 | 4 | | 160 | 6 | | 165 | 4 | | 170 | 2 | | 175 | 0 | | 180 | -2 | | 185 | -4 | | 190 | -6 | | 195 | -8 | | 200 | -6 | | 205 | -4 | | 210 | -2 | | 215 | 0 | | 220 | 2 | | 225 | 4 | | 230 | 6 | | 235 | 4 | | 240 | 2 | | 245 | 0 | | 250 | -2 | | 255 | -4 | | 260 | -6 | | 265 | -8 | | 270 | -6 | | 275 | -4 | | 280 | -2 | | 285 | 0 | | 290 | 2 | | 295 | 4 | | 300 | 6 | | 305 | 4 | | 310 | 2 | | 315 | 0 | | 320 | -2 | | 325 | -4 | | 330 | -6 | | 335 | -8 | | 340 | -6 | | 345 | -4 | | 350 | -2 | | 355 | 0 | | 360 | 2 | | 365 | 4 | | 370 | 6 | | 375 | 4 | | 380 | 2 | | 385 | 0 | | 390 | -2 | | 395 | -4 | | 400 | -6 | | 405 | -8 | | 410 | -6 | | 415 | -4 | | 420 | -2 | | 425 | 0 | | 430 | 2 | | 435 | 4 | | 440 | 6 | | 445 | 4 | | 450 | 2 | | 455 | 0 | | 460 | -2 | | 465 | -4 | | 470 | -6 | | 475 | -8 | | 480 | -6 | | 485 | -4 | | 490 | -2 | | 495 | 0 | | 500 | 2 | | 505 | 4 | | 510 | 6 | | 515 | 4 | | 520 | 2 | | 525 | 0 | | 530 | -2 | | 535 | -4 | | 540 | -6 | | 545 | -8 | | 550 | -6 | | 555 | -4 | | 560 | -2 | | 565 | 0 | | 570 | 2 | | 575 | 4 | | 580 | 6 | | 585 | 4 | | 590 | 2 | | 595 | 0 | | 600 | -2 | | 605 | -4 | | 610 | -6 | | 615 | -8 | | 620 | -6 | | 625 | -4 | | 630 | -2 | | 635 | 0 | | 640 | 2 | | 645 | 4 | | 650 | 6 | | 655 | 4 | | 660 | 2 | | 665 | 0 | | 670 | -2 | | 675 | -4 | | 680 | -6 | | 685 | -8 | | 690 | -6 | | 695 | -4 | | 700 | -2 | | 705 | 0 | | 710 | 2 | | 715 | 4 | | 720 | 6 | | 725 | 4 | | 730 | 2 | | 735 | 0 | | 740 | -2 | | 745 | -4 | | 750 | -6 | | 755 | -8 | | 760 | -6 | | 765 | -4 | | 770 | -2 | | 775 | 0 | | 780 | 2 | | 785 | 4 | | 790 | 6 | | 795 | 4 | | 800 | 2 | | 805 |< img src="boxmogrammato", "boxmogrammato" indicates the value of the boxplot on the x-axis. The y-axis is labeled as "SpO₂ Abweichung". The chart is saved as a PNG file named "Testmodell AOJ-70D".< img src="boxmogrammato", but it is not part of the original text in the image.< img src="boxmogrammato" indicates the original text in the image.< img src="boxmogrammato" indicates the original text in the image.< img src="boxmogrammato" indicates the original text in the image.< img src="boxmogrammato" indicates the original text in the image.< img src="boxmogrammato" indicates the original text in the image.< img src="boxmogrammato" indicates the original text in the image.< img src="boxmogrammate" indicates the original text in the image.< img src="boxmogrammate" indicates the original text in the image< img src="boxmogrammate" indicates the original text in the image< img src="boxmogrammate" indicates the original text in the image< img src="boxmogrammate" indicates the original text in the image< img src="boxmogrammate" indicates the original text in the image< img src="boxmogrammate" indicates the original text in the image< img src="boxmogrammate" indicates the original text in this image. The original text in this image is not included in the original data series. The original text in this image is not included in the original data series in this image. The original text in this image is not included in the original data series in this image. The original text in this image is not included in the original data series in this image. The original text in this image is not included in the original data series in this image. The original text in this image is not included in the original data series in this image. The original text in this image is not included in the original data series in this image. The original text in this image is not include an additional label for "boxmogrammato".< img src="boxmogrammato", but it is not part of the original text in this image.< img src="boxmogrammate" indicates that these values are estimated based on their original text and are not included in the original data series.< img src="boxmogrammate" indicates which only contains one of these values.< img src="boxmogrammate" indicates which only contains another of these values.< img src="boxmogrammate" indicates which only contains three or more times the original text. The original text in this image is estimated based on its original text and is not included in the original data series.< img src="boxmogrammate" indicates which only contains three or more times the original text in this image.< img src="boxmogrammate" indicates which only contains three or more times the original text in this image.< img src="boxmogrammate" indicates which only contains three or more times the original text in this image.< img src="boxmogrammate" indicates which only contains three or more times the original text in this image.< img src="boxmogrammate" indicates which only contains three or more times the original text in this image.< img src='boxmogrammate' indicates which only contains three or more times the original text in this image.< img src='boxmogrammate' indicates which only contains three or more times the original text in this image.< img src='boxmogrammate' indicates which only contains three or more times the original text in this image.< img src='boxmogrammate' indicates which only contains three or more times the original text in this image.< img src='boxmogrammate' indicates which only contain three or more times the original text in this image.< img src='boxmogrammate' indicates which only contains three or more times the original text in this image.< img src='boxmogrammate' indicates which only contains three or more times the original text in this image.< img src='boxmogrammate' indicates which only contains three or more times the original text in this image.< img src='boxmogrammate' indicates which only contains three or more time the original text in this image.< img src='boxmogrammate' indicates which only contains three or more time the original text in this image.< img src='boxmogrammate' indicates which only contains three or more time the original text in this image.< img src='boxmogrammate' indicates which only contains three or more time the original text in this image.< img src='boxmogrammate' indicates which only contains three or more time the original text

Pulse oximeter for measuring oxygen saturation

CONTENTS

Brief Description of Front Panel.... 23

General Description 23

Declaration 23

Measurement Principle 23

Safety Information 23

Contraindications 26

Components 26

Product Features 27

Intended Purpose 27

Operation Instructions 27

Product Accessories 27

Battery Installation 28

Installing the Lanyard 28

Cleaning and Disinfection 28

Maintenance and Storage 29

Specification 30

Graphical Plots of Data Points.... 31

Assessing the Validity of a SpO_2 Reading 33

Troubleshooting 34

Appendix 1 EMC Information.... 35

Possible Problems and Resolutions 39

Geratherm Oxy One - CONTENTS - 1

BRIEF DESCRIPTION OF FRONT PANEL

Geratherm Oxy One - BRIEF DESCRIPTION OF FRONT PANEL - 1

text_image Symbol of oxygen saturation 98 Bar graph Symbol of pulse rate 75 Power on Low-battery indicator Measured value of oxygen saturation Measured value of pulse rate

The pulse bar graph displays corresponding with the user's pulse beat. The height of the bar graph shows the user's pulse strength.

GENERAL DESCRIPTION

SpO_2 stands for peripheral capillary oxygen saturation. Oxygen saturation is defined as the ratio of oxyhemoglobin (HbO_2) to the total concentration of hemoglobin (i.e. Oxyhemoglobin + reduced hemoglobin) present in the blood. It is an important physiological parameter involved in respiration and circulation. The Pulse Oximeter feature herein is small, portable, non-invasive and easy to use. The user only needs to insert a finger into the chamber to measure his/her SpO_2 and pulse rate. The oximeter is suitable for home health monitoring.

Geratherm Oxy One - GENERAL DESCRIPTION - 1

DECLARATION

EMC of this product complies with IEC60601-1-2 standard. The materials which the user can come into contact with have no toxicity and no action on tissues comply with ISO10993-1, ISO10993-5 and ISO10993-10.

MEASUREMENT PRINCIPLE

Oxygenated blood absorbs light preferentially at 905nm (near infrared light), whereas deoxygenated blood absorbs light preferentially at 660nm (red light). A pulse oximeter works by passing a beam of red and infrared light through a pulsating capillary bed and then measure the amount of red and infrared light emerging from the tissues via a sensor. To improve accuracy, the Oximeter uses a proprietary algorithm to collect data from

pulsatile arterial blood and excludes local noise from the tissues. The relative absorption of light by oxyhemoglobin (HbO 2 ) and deoxyhemoglobin is then calculated according to the Beer-Lambert's law and a quantitative measurement of the users' oxyhemoglobin status i.e. oxygen saturation level (SpO 2 ) is derived.

Geratherm Oxy One - MEASUREMENT PRINCIPLE - 1

text_image Light beams sensor Light Emitting Diode

Due to the sensitivity of the pulse oximeter, finger should be kept stationary during measurement. It is recommended that you use this device for measurement before or after sports. Do not use for continuous monitoring.

SAFETY INFORMATION

DANGER

Indicates an imminent hazard that, if not avoided, will result in death or serious injury.

WARNING

Indicates a potential hazard or unsafe practice that, if not avoided, will result in death or serious injury.

Warnings

  1. Federal (U.S.) law restricts this device to sale by or on the order of a physician.
  2. Explosive hazard - DO NOT use the equipment in the environment with tinder such as anesthetic.
  3. DO NOT use the equipment while the patient is being scanned by MRI or CT.
  4. Don't near active HF surgical equipment and the RFshielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
  5. Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.

Geratherm Oxy One - Warnings - 1

SAFETY INFORMATION

  1. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

  2. The equipment is NOT intended for neonate and infant, and the patient's finger thickness should between 8 to 25.4 mm.

  3. It is recommended that the equipment should be inspected before use, when there is obvious damage, stop using the equipment.

  4. Discomfort or pain may appear if using the equipment ceaselessly, especially for microcirculation barrier patients, it is recommended that the equipment should not be used on the same finger more than 10 minutes.

  5. Please do not insert injured fingers into the pulse oximeter.

  6. The equipment is just a clinical diagnosis auxiliary equipment. The physiological data displayed on the equipment are for reference only and cannot be directly used for diagnostic interpretation.

  7. It is not recommended to use the equipment in high frequency environment such as electrosurgical equipment.

  8. Do not have the equipment immersed in liquid.

  9. Prevent children from swallowing the equipment or its accessories. Children must be accompanied by adult guardian when using the device.

  10. Please follow local ordinances and recycling instructions regarding disposal or recycling of the equipment and batteries.

  11. No maintenance or servicing the device when using.

  12. Users are NOT permitted to repair the equipment by themselves.

  13. There are NO replaceable components in the equipment

CAUTION

Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.

Geratherm Oxy One - CAUTION - 1

SAFETY INFORMATION

Cautions

  1. The equipment is designed to measure the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following:

• Excessive light, such as sunlight or direct home lighting
• Moisture in the equipment
• Finger is beyond recommended size range
• Poor pulse quality
• Venous pulsations
• Anemia or lowhemoglobin concentrations
• Cardio green and other intravascular dyes
• Carboxyhemoglobin
- Methemoglobin
• Dysfunctional hemoglobin
• Artificial nails or fingernail polish

  1. The light (the infrared is invisible) emitted from the equipment is harmful to the eyes so the user and the maintenance man should not look at the light.
  2. When the performance changes (such as inaccurate measurement or abnormal display), please stop using it immediately and contact the after-sales service personnel in time.

NOTE: Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components, including batteries.

CONTRAINDICATIONS

It is not intended to be used when the following happens: pulse oximeter should not be fixed on the tissue injury site; Not used for blood oxygen monitoring in hyperkinetic patients; Not used for blood oxygen measurement under hypoperfusion.

COMPONENTS

The oximeter consists of sensor, electronic circuits, display and plastic enclosures.

NOTES

• The sensor is placed in the middle of the device to which the finger insert.
• The sensor is the Applied Part of the equipment.

Geratherm Oxy One - NOTES - 1

PRODUCT FEATURES

• Lightweight, portable and easy to use.
• LED screen shows pulse rate, SpO_2
• Large font display.
• Low battery indicator.
• Auto shut down if no signal is detected within 30 seconds.

INTENDED PURPOSE

The pulse oximeter is a reusable device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients.

OPERATION INSTRUCTIONS

  1. Install two AAA batteries into battery compartment correctly.
  2. Insert one of your fingers into the finger chamber of the pulse oximeter.
    Note: The fingernail should be facing the top chamber (which contains the sensor). Finger should also be inserted completely into the chamber. Otherwise, measurement will be inaccurate.
  3. Press the power-on button to turn on the pulse oximeter.
  4. Finger and body should not tremble during measuring.
  5. Read correct data from displayed screen.
  6. The sound will alert the users when the blood oxygen levels fall below the 94 % threshold.

Notes:

  • When your finger is inserted into the oximeter, your nail surface must be upward.
  • The results may be wrong if you do not place your finger thoroughly in the oximeter.
  • Please use medical alcohol to clean the silicon which touches the finger inside of oximeter, and use alcohol to clean the test finger before and after each test. (The silicon inside of the oximeter belongs to medical silicon, which has no toxins and does no harm to the skin).
  • The sound function is only used to remind the user to check their blood oxygen level and is not used for any alarm purposes. Any time users have doubts about their blood oxygen saturation, they should check their blood oxygen measurement in time.

PRODUCT ACCESSORIES

  1. Pulse oximeter* 1 pc 3. Batteries* 2 pc
  2. User manual* 1 pc
  3. Lanyard* 1 pc

Geratherm Oxy One - PRODUCT ACCESSORIES - 1

BATTERY INSTALLATION

Put the two AAA batteries into battery compartment in correct polarities.

Geratherm Oxy One - BATTERY INSTALLATION - 1

text_image Illustration showing three-step steps of a blood glucose meter: battery charging, blood pressure testing, and digital display testing.

Notes:

  • Battery polarities should be correctly installed. Otherwise, damage may be caused to the device.
  • Please remove the batteries if the Oximeter will not be used for a long time.
  • This product should be kept away from children, because small parts such as battery cases, lanyards can cause children to choke.

INSTALLING THE LANYARD

Thread the thin end of the lanyard through the lanyard hole, and thread the coarse end of the lanyard through the thin end of the Lanyard, and tighten the lanyard.

Geratherm Oxy One - INSTALLING THE LANYARD - 1

NOTE

• Do not hang the oximeter on the cable.

CLEANING AND DISINFECTION

1. Cleaning

Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your place, the equipment should be cleaned more frequently. Recommended cleaning agents are:

- Ethanol (70 %)

To clean your equipment, follow these rules:

• Shut down the pulse oximeter.
- Clean the display screen using a soft, clean cloth dampened with a glass cleaner.

CLEANING AND DISINFECTION

  • Clean the exterior surface of the equipment and probe using a soft cloth dampened with the cleaner.
  • Wipe off all the cleaning solution with a dry cloth after cleaning if necessary.
    • Dry your equipment in a ventilated and cool place.

To avoid damage to the equipment, follow these rules:

CAUTIONS

• Always dilute according the manufacturer's instructions or use lowest possible concentration.
• Do not immerse the equipment in liquid.
• Do not pour liquid onto the equipment or accessories.
- Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or acetone-based cleaners).
- If you spill liquid onto the equipment, contact us or your service personnel.

2. Disinfection

Clean the pulse oximeter before disinfecting it. The recommended disinfectant is ethanol 70 %. Disinfection steps are the same as cleaning. Do not disinfect the instrument by using high-temperature/high-pressure disinfecting gas.

⚠️ Never use ETO or formaldehyde for disinfection.

Geratherm Oxy One - Disinfection - 1

MAINTENANCE AND STORAGE

  1. Replace the batteries in time when low battery indicator flashes.
  2. Clean the surface of the fingertip oximeter before use.
  3. Remove the batteries inside the battery cassette if the oximeter will not be operated for a long time.
  4. It is best to preserve the product in a place where the ambient temperatures is -20^ - 50^ and relative humidity is 10% - 95%
  5. It is recommended that the product should be kept in a dry environment anytime. A wet ambient might affect its lifetime and even might damage the product.
  6. Avoid exposure to direct sunlight.
  7. Avoid excessive radioactive infrared rays or ultraviolet rays.
  8. Please follow the law of the local government to deal with used battery.

SPECIFICATION

  1. Display Type: LED
  2. SpO_2 :

Measurement range: 0 % - 100 %

Display range: 35 % - 100 %

Resolution: 1 %

Accuracy: ±2 % for 70 % - 100 %; less than 70 % is unspecified.

  1. Pulse Rate:

Measurement range: 25 bpm - 250 bpm

Resolution: 1 bpm

Accuracy: ±2 bpm

Pulse Intensity: Bar graph indicator

  1. Power Requirements:

Two AAA alkaline Batteries

Power consumption: 30 mA (Normal)

Low battery indication: When the battery voltage is 1.9 V ± 0.2 V , the oximeter will shut down.

Battery Life: Two AAA 1.5 V, 600 mAh alkaline batteries could be continuously operated as long as 24 hours.

  1. Dimension: 63.5 mm (L) X 36 mm (W) X 34 mm (H)

Weight: 43 g (without battery)

  1. Environment Requirements: Temperature: 10 °C - 40 °C

Humidity (non-condensing): 15% – 95%

Atmospheric pressure: 70 kPa-106 kPa

  1. Storage/transportation conditions

Temperature: -20 °C - 50 °C

Humidity (non-condensing): 10% - 95%

Atmospheric pressure: 70 kPa - 106 kPa

  1. Measurement Performance in Low Perfusion Condition: 0.3%

  2. Safety Classification

Type of protection against electric shock: Internally powered equipment

Degree of protection against electric shock:

Type BF-Applied part (non-defibrillation proof)

Operating mode: Spot checking

Protection Against Harmful Ingress of Water or Particular Matter: IP22

Geratherm Oxy One - SPECIFICATION - 1

SPECIFICATION

  1. Range of the peak wave lengths: RED: 660 nm/IR: 905 nm
  2. Maximum optical output power: 1.2 mW
  3. Use-life: Two years

Note:

  1. SpO_2 and pulse rate accuracy are the essential requirements.
  2. There is no alarm that includes the capability to detect an Sp2Cor pulse rate physiological alarm condition is available.
  3. Data update duration is not applicable since it is not intended to be connected with other device to display the reading
  4. The oximeter is for spot checking, frequency of use is not applicable.
  5. There will be no reading if the oximeter is not well fitted.

GRAPHICAL PLOTS OF DATA POINTS
Geratherm Oxy One - Note: - 1

scatter | Mean of SpO₂ and SaO₂ | SpO₂ Daviation | | --------------------- | -------------- | | 65 | -4 | | 70 | -2 | | 75 | -3 | | 80 | -1 | | 85 | 0 | | 90 | 6 | | 95 | -2 | | 100 | 4 | | 105 | -1 | | 110 | 3 | | 115 | -5 | | 120 | 2 | | 125 | -3 | | 130 | 1 | | 135 | -2 | | 140 | 0 | | 145 | -1 | | 150 | -4 | | 155 | 3 | | 160 | -2 | | 165 | 1 | | 170 | -3 | | 175 | 2 | | 180 | -1 | | 185 | 0 | | 190 | -2 | | 195 | 3 | | 200 | -1 | | 205 | 2 | | 210 | -4 | | 215 | 1 | | 220 | -3 | | 225 | 0 | | 230 | -2 | | 235 | 3 | | 240 | -1 | | 245 | 2 | | 250 | -3 | | 255 | 1 | | 260 | -2 | | 265 | 0 | | 270 | -1 | | 275 | 3 | | 280 | -4 | | 285 | 2 | | 290 | -3 | | 295 | 1 | | 300 | -2 | | 305 | 0 | | 310 | -1 | | 315 | 3 | | 320 | -4 | | 325 | 2 | | 330 | -3 | | 335 | 1 | | 340 | -2 | | 345 | 0 | | 350 | -1 | | 355 | 3 | | 360 | -3 | | 365 | 2 | | 370 | -2 | | 375 | 1 | | 380 | -1 | | 385 | 0 | | 390 | -4 | | 395 | 3 | | 400 | -2 | | 405 | 1 | | 410 | -1 | | 415 | 0 | | 420 | -3 | | 425 | 3 | | 430 | -2 | | 435 | 2 | | 440 | -1 | | 445 | 1 | | 450 | -4 | | 455 | 0 | | 460 | -3 | | 465 | 3 | | 470 | -2 | | 475 | 2 | | 480 | -1 | | 485 | 0 | | 490 | -2 | | 495 | 3 | | 500 | -1 | | 505 | 2 | | 510 | -4 | | 515 | 1 | | 520 | -3 | | 525 | 0 | | 530 | -2 | | 535 | 3 | | 540 | -1 | | 545 | 2 | | 550 | -4 | | 555 | 1 | | 560 | -3 | | 565 | 0 | | 570 | -2 | | 575 | 3 | | 580 | -1 | | 585 | 2 | | 590 | -4 | | 595 | 1 | | 600 | -3 | | 605 | 0 | | 610 | -2 | | 615 | 3 | | 620 | -1 | | 625 | 2 | | 630 | -4 | | 635 | 1 | | 640 | -3 | | 645 | 0 | | 650 | -2 |

Bland-Altman graph for SpO_2-SaO_2

Geratherm Oxy One - Note: - 2

GRAPHICAL PLOTS OF DATA POINTS

Geratherm Oxy One - GRAPHICAL PLOTS OF DATA POINTS - 1

scatter | SaO₂% | SpO₂% | |-------|-------| | 80 | 68 | | 85 | 72 | | 90 | 75 | | 95 | 78 | | 100 | 82 | | 105 | 85 | | 110 | 88 | | 115 | 92 | | 120 | 95 | | 125 | 98 | | 130 | 100 |

The regression line of the measured data

Geratherm Oxy One - GRAPHICAL PLOTS OF DATA POINTS - 2

ASSESSING THE VALIDITY OF A SPO _2 READING

You can check the quality of the bar graph and the stability of the SpO_2 values on the Patient Monitor to assess whether the sensor functions properly and whether the SpO_2 readings are valid. Always use these two indications simultaneously to assess the validity of a SpO_2 reading.

NOTE:

  1. The SpO_2 accuracy has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, only about two-thirds of the measurements can be expected to fall within the specified accuracy compared to CO-oximeter measurements. The volunteer population in the studies composed of local healthy men and women from age 18 to 46, with variations of skin pigmentations. And the accuracy cannot be assessed by a function tester.
  2. The pulse rate accuracy is obtained by comparison to the pulse rate generated with an arterial oxygen simulator (also an electronic pulse simulator).
  3. Generally, the quality of the pulse rate bar graph reflects the quality of the light signals obtained by the sensor. A bar graph of poor quality manifests a decline of the signal validity. On the other hand, the stability of the SpO_2 values also reflects the signal quality. Different from varying SpO_2 readings caused by physiological factors, unstable SpO_2 readings are resulted from the sensor's receiving signals with interference. The problems mentioned above may be caused by patient movement, wrong sensor placement or sensor malfunction. To obtain valid SpO_2 readings, try to limit patient movement.

Geratherm Oxy One - NOTE: - 1

TROUBLESHOOTING

Trouble Possible Reason Solution
The equipment can't be turned on.The battery is drained away or almost drained away.Please replace battery.
The battery installation is incorrect.Install the battery over again.
The malfunction of the equipment.Please contact the service personnel.
The display is off suddenly.The equipment is set to shut down automatically in 30 seconds without any operations.Normal
The battery is almost drained away.Please replace battery.
The SpO_2 and Pulse Rate are not displayed stably.The thoroughfare from photo detector to light emitting diode was sheltered by some objects.Check and clean the inside of the probe especially the two windows of sensors.
The finger is shacking or the user is moving.The user needs to keep still.
The finger is not placed inside deep enough.Place the finger properly and try again.
The finger's size is too big or too small.Select the correct size finger to measure.
Excessive ambient light. Avoid the excessive ambient light irradiation.
Cyclical fluctuations in pulse rate values.If the measurement is proper, the user might has arrhythmia. Please consult with the doctor.
The SpO_2 and Pulse Rate can't be displayed normally.The finger is not properly positioned.Place the finger properly and try again.
The patient's SpO_2 is too low to be detected.Try again, go to a hospital for a diagnosis if you are sure the equipment works all right.

Geratherm Oxy One - NOTE: - 2

APPENDIX 1 EMC INFORMATION

Guidance and manufacturer's declaration-Electromagnetic emission
Pulse Oximeter is intended for use in the electromagnetic environment speci-fied below. The customer or the user of the Model AOJ-70D should assure that it is used in such an environment.
Emissions ComplianceElectromagneticenvironment - guidance
RF emissions CISPR 11Group 1 The Pulsese Oximeter uses RF energy only for its internal func-tion. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11Class B The PulseOximeter is suitable for use in all establishments, includ-ing domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC61000-3-2N.A. N.A.
Voltage fluctuations/flicker emissions IEC61000-3-3N.A.

Geratherm Oxy One - NOTE: - 3

Geratherm Oxy One - NOTE: - 4

Guidance and manufacturer's declaration - electromagnetic Immunity
The Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of Pulse Oximeter should assure that it is used in such an environment.
Radiated RF IEC 61000-4-3 (Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment)Test Frequency (MHz)Band (MHz)ServiceModulationMaximum Power (W)Distance (m)IEC 60601-1-2 Test Level (V/m)Compliance level (V/m)
385 380- 390 TETRA400Pulse modulation 18 Hz1.8 03 27 27
450 430- 470GMRS 460, FRS 460FM ± 5 kHz devi-ation on 1 kHz sine2 0.328 28
710704 - 787 LTE Band 13, 17Pulse modulation 217 Hz0.2 03 9 9
745
780
810800 - 960GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE-Band 5Pulse modulation 18 Hz2 0.328 28
870
930

APPENDIX 1 EMC INFORMATION

Radiated RF IEC 61000-4-3 (Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment)17201700 - 1990GSM 1800;CDMA 1900;GSM 1900;DECT;LTE Band 1, 3, 4, 25;UMTSPulse modulation217 Hz2 0.328 28
1845
1970
24502400 - 2570Bluetooth,WLAN,802.11 b/g/n,RFID 2450,LTE Band 7Pulse modulation217 Hz2 0.328 28
52405100 - 5800WLAN 802.11 a/nPulse modulation217 Hz0.2 03 9 95500
5785

Geratherm Oxy One - NOTE: - 5

APPENDIX 1 EMC INFORMATION

Guidance and manufacturer's declaration - Electromagnetic immunity
The Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of Pulse Oximeter should assure that it is used in such an environment.
Immunity test IEC 60601 test levelCompliance levelElectromagnetic environ-ment-guidance
Electrostatic discharge (ESD) IEC 61000-4-2±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV airFloors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electrical fast transient/burst IEC 61000-4-4Not appli-cableNot appli-cable
Surge IEC 61000-4-5Not appli-cableNot appli-cable
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11Not appli-cableNot appli-cable
Power frequency magnetic field IEC 61000-4-830 A/m 30 A/m 50Hz/60HzPower frequency 50Hz/60Hzmagnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Conducted RF IEC 61000-4-6Not appli-cableNot appli-cable
Radiated RF IEC 61000-4-310 V/m 80 MHz – 2.7 GHz 80 % AM at 1 kHz10 V/m 80 MHz – 2.7 GHz 80 % AM at 1 kHz
NOTE: UT is the a.c. mains voltage prior to application of the test level

Geratherm Oxy One - NOTE: - 6

POSSIBLE PROBLEMS AND RESOLUTIONS

Problems Possible Reason Solution
The Oximeter fails to display the blood oxygen saturation levels and/or pulse rate.1. The finger is not placed between the sensor and the Light Emitting Diode.2. The user's blood oxygen is too low to be detected.1. Make sure that the finger is placed right in between the sensor and the Light Emitting Diode.2. Make sure nothing is restricting your blood flow.
SpO_2 or PR is shown unstably placed between the1. The finger is not sensor and the Light Emitting Diode.2. The user is moving his/her finger and/or body.1. Make sure that the finger is placed right in between the sensor and the Light Emitting Diode.2. Try to stay still during measurement.
The oximeter can not be powered on.1. The batteries are drained.2. The batteries are incorrectly installed.3. The Oximeter is defective and/or damaged.1. Replace the batteries.2. Install the battery correctly.3. Contact the distributor.
The screen is suddenly off.1. Without use, the oximeter will be automatically shut down within 30 seconds.2. The batteries are drained.1. This is normal.Just turn on the pulse oximeter again.2. Replace the batteries.

There are no user-serviceable parts inside the oximeter. The cover should only be removed by qualified service personnel. Do not spray, pour, or spill any liquid on the oximeter, its accessories, connectors, switches, or openings in the enclosure as this may damage the oximeter.

Geratherm Oxy One - NOTE: - 7

Geratherm®

oxy one

Geratherm Oxy One - oxy one - 1

text_image Gerathem® $400 100 75 Min

MODE D'EMPLOI

Geratherm Oxy One - oxy one - 2

CARACTÉRISTIQUES DU PRODUIT

ACCESSOIRES POUR PRODUITS

natural_image Technical line drawing of a battery pack with internal compartments and external casing (no text or symbols)

Geratherm Oxy One - ACCESSOIRES POUR PRODUITS - 1

natural_image Hand placing a small square object onto a rectangular device with a checkmark nearby (no text or symbols)

Geratherm Oxy One - ACCESSOIRES POUR PRODUITS - 2

natural_image Hand inserting a blood glucose meter into a digital device, with a cross symbol indicating the catheter (no text or labels present)

REMARQUE

  1. Dimension : 63,5 mm (L) X 36mm (L) X 34mm (H)
text_image Illustration showing three-step battery installation process: battery, clean cell, and blood glucose meter with checkmark and cross symbols.

Nota:

63.5 mm (L) × 36 mm (L) × 34 mm (H)

MANUAL DE INSTRUÇÕES

POSSÍVEIS PROBLEMAS E RESOLUÇÕES

text_image Illustration showing three-step battery installation procedure: battery, clean cell, and blood glucose meter with checkmark and cross symbols.

Notes:

Shenzhen AOJ Medical Technology Co., Ltd.

Room 301&4F, Block A, Building A,

Jingfa Intelligent Manufacturing Park,

Xiaweiyuan, Gushu Community, Xixiang Street,

Bao'an District, 518126

Shenzhen

China

Geratherm Oxy One - Notes: - 1

Share Info GmbH

Heerdter Lohweg 83,

40549 Düsseldorf

Germany

Geratherm Oxy One - Notes: - 2

3rd Floor, Office C, Townend House, Park Street, Walsall,

West Midlands, WS1 1NS

United Kingdom

Geratherm Oxy One - Notes: - 3

Geratherm Medical GERMANY GmbH

Fahrenheitstraße 1

99331 Geratal

Germany

Phone: +49 36205 980

www.geratherm.com

PUOMOO0004V02

2025-07

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Geratherm

Model : Oxy One

Category : Measuring equipment