OXI O2 - Measurement GOCLEVER - Free user manual and instructions
Find the device manual for free OXI O2 GOCLEVER in PDF.
| Type | Fingertip Pulse Oximeter |
| Brand | GoClever |
| Model | OXI O2 |
| Parameters Measured | SpO₂ (0–100%), Pulse Rate (25–250 bpm) |
| SpO₂ Accuracy | ±2% |
| Pulse Rate Accuracy | ±1 bpm |
| Display | 128 × 64 pixels, 4-direction display, brightness adjustable (1–5) |
| Power Source | 2 × AAA batteries |
| Operating Temperature | +5°C to +40°C |
| Operating Humidity | ≤85% RH |
| Atmospheric Pressure (Operating) | 700–1060 hPa |
| Storage/Transport Temperature | -20°C to +55°C |
| Storage/Transport Humidity | ≤93% RH |
| Storage/Transport Pressure | 500–1060 hPa |
| Ingress Protection | IPX1 (drip-proof) |
| LED Wavelengths | Red ~660 nm, Infrared ~905 nm |
| Safety Classification | Type BF, Class 1 LED (IEC 60825-1) |
| Alarm Function | None |
| Intended Use | Home use, adults, short-term monitoring (max 2 hours) |
| Cleaning | Soft cloth dampened with water or mild liquid; do not soak |
| Battery Replacement | 2 × AAA, remove if not used for long time |
| Compliance | MDD 93/42/EEC, CISPR 11 Group 1 Class B |
Frequently Asked Questions - OXI O2 GOCLEVER
User questions about OXI O2 GOCLEVER
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Measurement in PDF format for free! Find your manual OXI O2 - GOCLEVER and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. OXI O2 by GOCLEVER.
USER MANUAL OXI O2 GOCLEVER


other
| Species | Value | |---|---| | SpO₂% | 98 | | PRbpm | 65 |PL
OPIS
GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone Zhuhai, PEOPLE'S REPUBLIC OF CHINA

UPOWAŻNIONY REPREZENTANT
W UNII EUROPEJSKIEJ:
Shanghai International Holding Corp. GmbH (Europe)
Eiffestrasse 80, 20537 Hamburg, Germany
IMPORTER:
GOCLEVER Sp. z o.o.
GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone Zhuhai, PEOPLE'S REPUBLIC OF CHINA

GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone Zhuhai, PEOPLE'S REPUBLIC OF CHINA

ZMOCNĚNÝ ZÁSTUPCE V EVROPSKÉ UNII:
Shanghai International Holding Corp. GmbH (Europe)
GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone Zhuhai, VOLKSREPUBLIK CHINA.

BEVOLLMÄCHTIGTER HANDELSVERTRETER
IN DER EUROPÄISCHEN UNION:
Shanghai International Holding Corp. GmbH (Europa)
The device measures oxygen saturation of arterial blood (also known as pulse oxygen saturation, usually shortened as SpO_2 ) and pulse. The device should be used in home environment, only by adults. The patient should not be monitored for long term (max 2 hours)
PRECAUTIONS FOR USE
When the oximeter is in use, there should not be any high voltage appliances
Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and moisture.
Explosion hazard. Do not use the oximeter in the presence of flammable anesthetics mixture with air, oxygen, or hydrogen.
The oximeter should be handled with care so as to avoid shocks and falls.
When the oximeter is in use, it must be ensured the batteries have sufficient capacity; otherwise there might be such phenomena as starting-up abnormalities or inaccurate measurement data, etc.
Take off the battery when the device is not used for a longer time
Do not use sharp objects to operate the device
Do not make any clinical judgments based solely on the oximeter. The oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms, as well as doctor's diagnoses.
Do not conduct SpO_2 measurement on the finger smeared with nail polish, otherwise this will lead to unreliable measurement results.
This oximeter does not have alarm function; please do not use this product in the environment where alarm is required.
Please do not open the enclosure. The enclosure shall only be opened by the authorized person.
In order to have more accurate measurements of SpO_2 and PR, the oximeter should be used in quiet and comfortable environment.
Clean the device with a soft cloth only a little damp with water or mild liquid, you should never let the oximeter soak in water
OPERATION INSTRUCTIONS
- Install two AAA batteries into battery cassette before closing its cover.
- Nip the oximeter, then insert one of fingers into the rubber hole of the oximeter before releasing the oximeter, and your nail surface must be upward.
- Press the function button once on front panel.
- Your finger and body do not tremble during measuring.
- Read corresponding data on the display screen.
- After turning on the oximeter, each time you press the power switch, the display screen will change to another direction. There are two display modes. If you long press the power switch, you can adjust the brightness of screen from 1 to 5.
DISPLAY
- SpO _2 Plethysmogram
- SpO _2 reading
- Pulse rate reading
- Indication of battery capacity
- Indication of pulse intensity
- Indication of screen brightness
- Display modes
POSSIBLE PROBLEMS AND RESOLUTIONS
| Problems Possible causes Solution | ||
| There is no response to the function button. | The button can not be pressed to its position | Ensure that the button is fully depressed. |
| Battery capacities are low | The batteries may be missing, discharged, or oriented incorrectly. Replaced them with new ones. | |
| Problems Possible causes | Solution | |
| The Pulse search time is too long | Perfusion may be too low | Check the patient. Change the measuring site. Try another oximeter. |
| Patient movement | Interference due to patient activity may be preventing the oximeter from tracking the pulse. Keep the patient still, if possible. | |
| Electromagnetic interference may be preventing the oximeter from tracking the pulse. | Remove the source of interference. | |
| There may be interference due to ambient light, or the oximeter may be on an extremity with a blood pressure cuff, arterial catheter, or intravascular line. | Reposition oximeter, as necessary. | |
PRODUCT SPECIFICATIONS
Type
Oximeter
Parameters
pulse, SpO_2
Screen resolution 128 x 64
Display mode 4 directions
Measuring range pulse oximetry: 0–100%
pulse rate: 25–250 bpm
Accuracy pulse oximetry: 2%
pulse rate: 1 bpm
Power 2 x AAA battery
Environmental specifications: Temperature + 5°C \~ +40°C,
Atmospheric pressure 700hPa \~ 1060hPa, Relative humidity . ≤85%
Transport and storage: Temperature -20°C \~ +55°C, Atmospheric pressure 500hPa \~ 1060hPa, Relative humidity ≤93%
Pulse oximetry sensors contain LEDs that emit red light at a wavelength of approximately 660 nm and infrared light at a wavelength of approximately 905 nm.
The total optical output power of the sensor LEDs is less than 15 mW. This information may be useful to clinicians, such as those performing photodynamic therapy.
Note: Sensor LED light emissions fall within Class 1 level, according to IEC 60825-1:2001. No special safety precautions are required.
GUIDANCE AND MANUFACTURE'S DECLARATION-ELECTROMAGNETIC ERMISSIONS-FOR ALL EQUIPMENT AND SYSTEMS
| Guidance and manufacture's declaration–electromagnetic ermissions |
| The Fingertip Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer of the user of the Fingertip Pulse Oximeter should assure that it is used in such and environment. |
| Emission test: CISPR 11, Compliance: Group 1The Fingertip Pulse Oximeter uses RF energy only for its internal function.Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. |
| Emission test: CISPR 11, Compliance: Class BThe Fingertip Pulse Oximeter is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. |
SYMBOLS DEFINITIONS
| Symbol Definition | |
![]() | Type BF equipment (Refer to IEC 60601-1:1995) |
![]() | Attention! Please refer to this manual. |
| %SpO2 Oxygen saturation of arterial blood | |
![]() | Pulse rate |
![]() | Non-Alarm indication (The device does not have alarm function) |
![]() | Production date |
| IPX1 Enclosure degree of ingress protection. | |
![]() | Complies with MDD93/42/EEC requirements |
CORRECT DISPOSAL OF THIS PRODUCT
(Waste Electrical & Electronic Equipment)
(Applicable in the European Union and other European countries with separate collection systems)

This marking indicates that at the end of its working life this product should not be disposed of with other household waste in the EU. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this from other types of
wastes and recycle it responsible to promote the sustainable reuse of material resources.
To dispose of your used device, please use the return and collection systems available in your area or contact the retailer where the product was purchased. They can take this product for safe environmental recycling.

MAUFACTURER:
GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast,
Hi-tech Zone Zhuhai
PEOPLE'S REPUBLIC OF CHINA

EUROPEAN UNION REPRESENTATIVE:
Shanghai International Holding Corp. GmbH (Europe),
Eiffestrasse 80, 20537 Hamburg, Germany
IMPORTER:
GOCLEVER Sp. z o.o.
GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone Zhuhai, PEOPLE'S REPUBLIC OF CHINA

GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone Zhuhai, PEOPLE'S REPUBLIC OF CHINA

MEGHATALMAZOTT KÉPVISELŐ
AZ EURÓPAI UNIÓBAN:
Shanghai International Holding Corp. GmbH (Europe)
GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone Zhuhai, PEOPLE'S REPUBLIC OF CHINA

RAPPRESENTANTE AUTORIZZATO
GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone Zhuhai, PEOPLE'S REPUBLIC OF CHINA

REPRESENTANTE AUTORIZADO NA UNIÃO EUROPEIA
Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg, Germany
IMPORTADOR:
GOCLEVER Sp. z o.o.
GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone Zhuhai, PEOPLE'S REPUBLIC OF CHINA

GuangdongBiolightMeditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone Zhuhai, PEOPLE'S REPUBLIC OF CHINA

OPRÁVNENÝ ZÁSTUPCA V EURÓPSKEJ ÚNII:
Shanghai International Holding Corp.
GmbH (Europe), Eiffestrasse 80, 20537 Hamburg, Nemecko
IMPORTÉR:
GOCLEVER Sp. z o.o.