HAYWARD Easy Chem - Pool controller

Easy Chem - Pool controller HAYWARD - Free user manual and instructions

Find the device manual for free Easy Chem HAYWARD in PDF.

📄 139 pages English EN Download 💬 AI Question 13 questions ⚙️ Specs
Notice HAYWARD Easy Chem - page 5
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Automatic pool controller for pH and ORP (Easy Chem)
Brand Hayward
Model Easy Chem (EASY-CH-DPHRXG / EASY-CH-DPHRXGS)
Dimensions (Length × Height × Depth) 212 × 303 × 113 mm
Weight 3.5 kg
Power supply 220-240 VAC, 50-60 Hz, 28 W
Measurement ranges pH: 0-14; ORP: -99 to 1000 mV; Temperature: 0 to +55 °C
Measurement accuracy ±0.02 pH; ±3 mV; ±0.5 °C
Flow rate of dosing pumps 1.5 l/h (for each pump)
Maximum back pressure 1.5 bar
Main functions Automatic pH and chlorine (ORP) regulation; proportional, timed or On/Off dosing; OFA alarms; automatic probe calibration
Inputs pH, ORP, temperature probes (PT100); Reed flow sensor; reservoir level (2 inputs); circulation pump trigger 220 VAC
Outputs 3 relays (250 VAC, 10 A): AUX1 (pH), AUX2 (ORP) and ALARM
Display Screen with button navigation (∧, ∨, OK, exit)
Calibration Automatic for pH, ORP and temperature; solutions provided (pH buffer 4 and 7, 465 mV solution)
Maintenance and cleaning Clean probes with soft brush and toothpaste; replace peristaltic tubes every 2 years (recommended annually)
Safety Disconnect before intervention; use a 30 mA differential circuit breaker; installation by qualified electrician; avoid freezing
Spare parts and repairability pH, ORP, temperature probes; peristaltic tubes; probe holder; valves; mounting kit; 3-year warranty (excluding wear parts)
General information Manual available in several languages; operating temperature: 0-55 °C; installation in ventilated technical room

Frequently Asked Questions - Easy Chem HAYWARD

How to install the pH and ORP probes?
The probes are delivered wet with a protective cap. Remove the cap and insert the probes into the double probe holder. Tighten by hand to ensure a seal. The tilt angle of the probe must never exceed 45° from vertical. Make sure the probes remain constantly in contact with the pool water.
How to perform pH calibration?
Access the calibration menu by pressing OK for 3 seconds. Select 'pH Calibration'. Follow the on-screen instructions: immerse the probe in pH 7 buffer solution, then in pH 4 (or a single point if configured). Calibration is automatic. Rinse the probe with demineralized water between solutions.
What does the 'OFA_1' or 'OFA_2' alarm mean?
OFA stands for 'Over Feed Alarm'. OFA_1 appears when the dosing duration exceeds 70% of the set time. OFA_2 at 100% blocks the pump. To reset, press OK for 3 seconds or use the exit button. Check the product levels and probe status.
How to prime the dosing pumps?
From the main menu, select 'Priming menu'. Use the ∧ or ∨ keys to choose the pH or ORP pump. Press OK to start priming. The pump will run continuously as long as you hold down the key. Release to stop.
What chemicals to use with the Easy Chem?
For pH, use an acid (pH-) or alkali (pH+) according to the configured dosing type. For ORP (chlorine), use a chlorinated product or a reducer depending on the mode (Low or High). Follow safety instructions: wear gloves and goggles, and place the reservoirs in a ventilated area.
How to clean the probes?
Turn off the device and disconnect the connectors. Unscrew the probe from the chamber. Clean the bulb with a soft toothbrush and toothpaste or a mild liquid detergent. Rinse with clean water. Dry and reassemble with new Teflon tape. Replace the probe if readings remain unstable.
The device does not turn on, what to do?
Check that the on/off switch is in the ON position. Make sure the power outlet is live (220-240 V). Check the connection cable between the screen and the unit. If the screen remains black, check the fuse (500 mA fast). If the problem persists, contact your installer.
How to set the pH set point?
From the main menu, select 'Configuration' then 'pH'. Use ∧ or ∨ to set the desired value (between 6 and 8 pH, default 7.4). Confirm with OK. You can also set the dosing type (Acid or Alkaline) and measurement alarms.
What is the lifespan of the probes and how to store them?
The probes have a lifespan of about 1 year. They must always remain wet: store them in the cap filled with water or in a KCl solution. Avoid freezing. In winter, remove the probes and keep them frost-free. Clean them regularly to extend their lifespan.
How to replace the peristaltic pump tube?
Open the pump cover. Release the left connector by pulling it upward. Turn the roller to free the tube. Place the new tube: insert the left connector, pass the tube under the guide, turn the roller to guide the tube to the right connector. Close the cover firmly. Replace the tube every year.

User questions about Easy Chem HAYWARD

3 questions about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

Is there an installation video available for the HAYWARD Easy Chem?
Frequently Asked Questions - 06/30/2026
Response Notice-Facile

Unfortunately, no installation video is available for the HAYWARD Easy Chem, but we offer the essential steps extracted from the official technical manual:

Installation of pH and ORP sensors

  • Remove the protective cap from both probes (pH and ORP)
  • Insert the probes into the double probe holder (model PSS3, 1/2" thread)
  • Maintain a maximum angle of 45° from the vertical
  • Ensure that the probes remain in constant contact with the water

Priming the dosing pumps

  • Go to the 6b4 - Priming menu
  • Select pH pump priming to prime the pH pump
  • Select ORP pump priming to prime the chlorine pump
  • Wait for the liquid to circulate in the dosing tubes before starting

Initial setup

  • Press and hold the OK button for 3 seconds to access the main menu
  • Set the pH to 7.4 (recommended default value)
  • Set the ORP to 700 mV (recommended default value)
  • Configure the pH alarms between 6.0 and 8.0
  • Configure the ORP alarms between 450 and 850 mV

Important: Electrical installation

The electrical connections (220-240V) must be carried out by a qualified electrician. The device must be protected by a 30mA residual current device.

For detailed diagrams and a complete step-by-step manual, refer to the full documentation of this manual.

Respond (be the first)
How do I adjust the pH and ORP on my HAYWARD Easy Chem controller?
Frequently Asked Questions - 06/12/2026
Response Notice-Facile

Here’s how to set up your HAYWARD Easy Chem controller to optimize pH and chlorine regulation:

pH Configuration

  1. Press the OK button for 3 seconds to access the Calibration menu
  2. Select “Configuration” then “pH”
  3. Use the up arrow (∧) or down arrow (∨) keys to set the desired value
  4. The allowed range is 6 to 8 pH (recommended default value: 7.4)
  5. Select the dosing mode: Acid or Alkaline as needed

Chlorine Regulation Configuration (ORP)

  1. Access the ORP Configuration menu
  2. Choose the dosing type: Proportional, Timed or On/Off
  3. Set the alarm thresholds according to your target range

Probe Calibration

Before commissioning, perform automatic calibration:

  • For pH: immerse the probe in the pH 7 buffer solution, then in the pH 4 solution
  • For ORP: use the provided 465 mV reference solution
  • Install the probes at an angle not exceeding 45° from vertical

Before First Use

  • Flush both dosing pumps until liquid flows through the pipes
  • Regularly clean the probes with a soft brush and mild detergent
  • Replace the peristaltic tubes annually to ensure optimal performance
Respond (be the first)
What does the error code OFA2 mean on my HAYWARD Easy Chem?
Frequently Asked Questions - 05/25/2026
Response Notice-Facile

OFA2 stands for "Over Feed Alarm". On your Hayward Easy Chem pH controller, OFA2 appears when the dosing pump operates at 100% of the set delay without reaching the desired setpoint (pH or ORP). This is a persistent situation and the pump is then automatically locked to prevent excessive dosing.

Difference with OFA1: OFA1 first appears at 70% of the delay (alert), OFA2 at 100% (lock). Unlike OFA1, the OFA2 alarm locks the pump and requires action.

Main causes on the Easy Chem:

  • The dosing pump is clogged or blocked (lime scale, chemical deposits)
  • The flow is insufficient due to a priming issue with the peristaltic pump
  • The configured OFA delay is too short compared to your pool's actual needs
  • The distribution piping is blocked or crushed
  • The chemical reservoir is empty or the product is blocked
  • The level probe indicates a lack of chemical product

To resolve OFA2 on your Hayward Easy Chem:

  1. Press OK for 3 seconds (or the escape key for 5 seconds) to reset the alarm
  2. Check that the chemical reservoir is not empty and that the product flows correctly
  3. Clean the peristaltic pump: open the cover, release the hose, and prime the pump (PRIMING menu: press the arrows to activate the pH or ORP pumps)
  4. Check that the suction and discharge pipes are not blocked or crushed
  5. Check the configured OFA delay in Setup pH or Setup ORP (parameter "OFA ALARM", range 0-60 minutes) - increase it if necessary
  6. Clean the valves and connectors (clean with a soft brush if lime scale is present)
  7. Test again with the priming function (PRIMING in the main menu)

Important: If OFA2 reappears immediately after resetting despite the above checks, or if the pump does not prime correctly, the peristaltic pump or an internal component needs to be replaced. The Santoprene tubes of the pump have a lifespan of 2 years (annual replacement recommended). Consult your installer or Hayward for a more in-depth diagnosis.

Respond (be the first)

Download the instructions for your Pool controller in PDF format for free! Find your manual Easy Chem - HAYWARD and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. Easy Chem by HAYWARD.

USER MANUAL Easy Chem HAYWARD

natural_image Abstract geometric logo with stylized letter H inside a circular frame (no text or symbols)

HAYWARD®

HAYWARD Easy Chem - HAYWARD® - 1

natural_image Two identical black-and-white industrial control units with buttons and dials, no visible text or symbols on the devices themselves.

CE EAC UK CA

HAYWARD Easy Chem - HAYWARD® - 2

natural_image Two line icons: one open with an information symbol, the other closed with an open book (no text or symbols present)

OWNER'S MANUAL

HAYWARD Easy Chem - HAYWARD® - 3

natural_image Abstract geometric logo with stylized letter H inside a circular frame (no text or symbols)

HAYWARD®

HAYWARD Easy Chem - HAYWARD® - 1

natural_image Two black industrial control panels with buttons and indicators, no visible text or symbols on the surfaces.

CE EAC UK CA

HAYWARD Easy Chem - HAYWARD® - 2

natural_image Two line icons: one open with an information symbol, the other closed with an open book (no text or symbols present)

EasyChem Double and Plug version | pH · ORP

OWNER'S MANUAL

PLEASE KEEP THIS MANUAL FOR FUTURE REFERENCE

⚠ WARNING – Carefully read the instructions that appear in this manual and on the device. Failure to comply with the instructions can cause injuries. This document must be given to every pool user, who should keep it in a safe place.

⚠ WARNING – Disconnect the equipment from the mains supply before any intervention.

⚠ WARNING – All electrical connections must be carried out by a qualified approved electrician in accordance with the standards currently in force in the country of installation or, failing this, in accordance with the international standard IEC 60334-7-702.

F NF C 15-100 GB BS7671:1992
D DIN VDE 0100-702 EW SIST HD 384-7-702.S2
A ÖVE 8001-4-702 H MSZ 2364-702:1994 / MSZ 10-533 1/1990
E UNE 20460-7-702 1993, REBT ITC-BT-31 2002 M MSA HD 384-7-702.S2
IRL IS HD 384-7-702 PL TS IEC 60364-7-702
I CEI 64-8/7 CZ CSN 33 2000 7-702
LUX 384-7.702 S2SK STN 33 2000-7-702
NL NEN 1010-7-702SLO SIST HD 384-7-702.S2
P RSIUEETR TS IEC 60364-7-702

⚠ WARNING – Check that the device is plugged into a power outlet that is protected against short-circuits. The device must also be powered via an isolating transformer or a residual current device (RCD) with a nominal operating residual current not exceeding 30 mA.

⚠ WARNING – Ensure that children cannot play with the device. Keep your hands and any foreign object away from openings and moving parts.

⚠ WARNING – Check that the supply voltage required by the product corresponds to the voltage of the distribution network and that the power supply cables are suitable for the product power supply.

⚠ WARNING – Chemicals can cause internal and external burns. To avoid death, serious injury and/or damage to equipment, wear personal protective equipment (gloves, goggles, mask, etc.) when servicing or maintaining this device. This device must be installed in an adequately ventilated place.

⚠ WARNING – The unit shall not be operated when there is no water flow in the cell.

⚠ WARNING – To reduce the risk of electric shock, do not use an extension cable to connect the device to the mains. Use a wall socket.

⚠ WARNING – Use, cleaning or maintenance of the device by children over 8 years of age or by people with impaired physical, sensory or mental capacities, or a lack of experience or expertise, should only take place once they have received appropriate instruction and under adequate supervision of an adult who is responsible for them, to ensure the device is handled safely and avoid all risk of danger.

⚠ WARNING – Use only original Hayward® parts.

⚠ WARNING – If the power supply cable is damaged, it must be replaced by the manufacturer, the after-sales service or similarly qualified persons to avoid danger.

⚠ WARNING – The device must not be used if the power cord is damaged. An electric shock could occur. A damaged power cord must be replaced by the after-sales service or similarly qualified persons to avoid danger.

INDEX

  1. General
  2. Pack contents
  3. Installation

3a. View of overall installation
3b. Wall-mounted installation
3c. Installation of pH and ORP probes, and acid dosing pump

  1. Electrical connections

4a. Electrical installation and wiring
4b. Electrical suppressor device
4c. Trigger input by circulation pump

  1. Specifications

  2. Setup and Operation

6a. Description of the home screen and defaults parameters
6b. Main menu
6c. View parameters
6d. Standby system
6e. Reset OFA Timer
6f. pH probe calibration
6g. ORP probe calibration
6h. Temperature probe calibration

  1. Dosing method
  2. Alarms
  3. Servicing
  4. Troubleshooting guide
  5. Environmental information

1. GENERAL

Before carrying out ANY work inside control panel of the EasyChem Double device, make sure you disconnect it from the power supply.

Failure to comply with the instructions contained in this manual could cause injury to people and/or damage to the appliance and the system.

2. PACK CONTENTS

HAYWARD Easy Chem - PACK CONTENTS - 1
A: PVC Crystal 4x6 suction hose (4 m)

HAYWARD Easy Chem - PACK CONTENTS - 2
B: Polyethylene delivery hose (5 m)

HAYWARD Easy Chem - PACK CONTENTS - 3
C: FPM Lip valve (3/8" GAS)

HAYWARD Easy Chem - PACK CONTENTS - 4
F: Reducer for injection valve (1/2" M to 3/8" F)

HAYWARD Easy Chem - PACK CONTENTS - 5
G: Foot filter (PP riser)

HAYWARD Easy Chem - PACK CONTENTS - 6
H: Mounting bracket kit (ø=6 mm screws)

HAYWARD Easy Chem - PACK CONTENTS - 7
L: Temperature sensor

HAYWARD Easy Chem - PACK CONTENTS - 8
M: pH probe

HAYWARD Easy Chem - PACK CONTENTS - 9
N: ORP probe

HAYWARD Easy Chem - PACK CONTENTS - 10
O: pH 4 Buffer solution

HAYWARD Easy Chem - PACK CONTENTS - 11
P: pH 7 Buffer solution

HAYWARD Easy Chem - PACK CONTENTS - 12
Q: 465 mv Calibration solution

HAYWARD Easy Chem - PACK CONTENTS - 13
R: Water S: Nuts for peristaltic tube

HAYWARD Easy Chem - PACK CONTENTS - 14

Item*\SystemEASY-CH-DPHRXG EASY-CH-DPHRXGS
A21
B21
C21
D22
E54
F21
G21
H11
L1
M11
N11
O11
P11
Q11
R11
S21

*NOTE: The values from the table represent the number of items inside the package.

3. INSTALLATION

3a. View of overall installation

HAYWARD Easy Chem - 3a. View of overall installation - 1

flowchart
graph TD
    A["①"] --> B["②"]
    B --> C["③"]
    C --> D["④"]
    D --> E["⑤"]
    E --> F["⑥"]
    F --> G["⑦"]
    G --> H["⑧"]
    H --> I["⑨"]
    I --> J["⑩"]
    J --> K["⑪"]
    K --> L["⑫"]
    L --> M["⑬"]
    M --> N["⑭"]
    N --> O["⑮"]
    O --> P["⑯"]
    P --> Q["⑰"]
    Q --> R["⑱"]
    R --> S["⑲"]
    S --> T["⑳"]
    T --> U["㉑"]
    U --> V["㉒"]
    V --> W["㉓"]
    W --> X["㉔"]
    X --> Y["㉕"]
    Y --> Z["㉖"]
  1. Pool
  2. Circulation pump
  3. Acid
  4. Chlorine
  5. Filter
  6. Heat exchanger
  7. pH probe
  8. ORP probe
  9. Temperature sensor

Warning!

Use with salt chlorinator:

For the pH systems, to prevent the risk of system malfunctioning or damage, observe the following instructions:

  1. Position the pH measuring probe prior to the chlorinator cell.
  2. To eliminate eddy currents, connect the pool water to an electrical ground point
  3. Position the product injection point after the chlorinator cell.

3b. Wall-mounted installation

Mount the control box on the wall. The box must be installed in the equipment room (dry, temperate, ventilated). Caution, acid vapours can cause irreversible damage to your device. Position the treatment product tanks accordingly.

Unplug the pool filter pump before you begin the installation. The installation must be performed in compliance with the regulations in effect in the country of installation.

HAYWARD Easy Chem - 3b. Wall-mounted installation - 1

natural_image Diagram showing four sequential steps of a mechanical or structural assembly with arrows indicating direction (no text or symbols present)

Warning! 2m

3c. Installation of pH and ORP probes, and acid dosing pump

The pH and ORP probes are «wet» packed and protected by a plastic cap. The probes must always remain wet. If the probes are allowed to dry, they will be permanently unusable (not covered by the warranty) and the pH-ORP test kit will be ineffective.

Remove the pH and ORP probes from their plastic protective caps and set the caps aside for later use (wintering). Insert the probes into the double probe holder and tighten to ensure they are watertight. Connect the probe holder to the valves screwed onto the saddle clamps and tighten by hand only. Check that the probes are watertight at startup. Seal with Teflon, if required.

After installation, check that the probes are constantly in contact with the water in the pool. When the filtration pump is not running (even for long periods), the water remaining in the chamber may be sufficient to protect the probes.

Probe positioning For optimum probe reading, position it perpendicular to the tubing. The probe's angle of inclination must never exceed 45° from vertical. ✓ ✓

4. ELECTRICAL CONNECTIONS

4a. Electrical installation and wiring

Connect the EasyChem Double to a permanent power outlet.

Open the peristaltic pump lid (1), unscrew four screws (2) and pull the peristaltic pump on the side (3), as illustrated in the pictures below:

HAYWARD Easy Chem - 4a. Electrical installation and wiring - 1

natural_image Three-panel diagram showing a device with control panel, internal components, and external casing (no text or symbols)

Label connection:
ORP probe pH probe PT100 NC Frequency INPUT GND LEV 1 LEV 2 Reed +5Vdc SIGN TRIGGER INPUT 220 Vac Relay pH Relay ORP Relay ALARM L N 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 000135964 R1-0

ClampDescriptionEASY-CH-DPHRXG / EASY-CH-DPHRXGSWire connection details
1 Input Probe pH pH Probe
2 Input Probe ORP ORP Probe
3 Input ProbeTEMPERATURE SENSOR (PT100)A= two wires sensor, probe included in the packageB= three wires sensor, please check the optional colors probeHAYWARD Easy Chem - 4a. Electrical installation and wiring - 3
4 Input Probe Not used---
5Frequency input by water meter pulse senderFlow Rate (Freq. Input)A= Mechanical reedB= Padwheel hall sensorHAYWARD Easy Chem - 4a. Electrical installation and wiring - 4
6 Level (product tank) pH Level probe Level probe for chemical tank
7 Level (product tank) Chlorine (ORP) level probe Level probe for chemical tank
8 Level (Reed sensor) Flow (External REED flow sensor) Flow sensor
9 Serial PortNot present---
10Trigger Input 220Vac (High Voltage)Circulation Pump (220Vac input)Line/Neutral wires
11Output Relay R1 (AUX1)RL1 AUX1 pHDry contact
12Output Relay R2 (AUX2)Easy-CH-DPHRXG:RL2 AUX2 OPREasy-CH-DPHRXGS:RL2 AUX2 OPR with Schuko socketDry contact
13Output Relay R3RL3 AlarmDry contact
14Earth connectorEarth---
15Power Supply220-240 Vac 50-60 Hz (F/N)---

4b. Electrical suppressor device

It is mandatory to install the coil suppressor, please see the instructions below:


HAYWARD Easy Chem - 4b. Electrical suppressor device - 1


HAYWARD Easy Chem - 4b. Electrical suppressor device - 2

natural_image Diagram of a mechanical assembly with a lever and pulley, no visible text or symbols

POWER SUPPLY CABLE


HAYWARD Easy Chem - 4b. Electrical suppressor device - 3

natural_image Simple line drawing of a mechanical device with a curved cable and a green cable (no text or symbols)


HAYWARD Easy Chem - 4b. Electrical suppressor device - 4
WRAP THE CABLE FOR A TURN

4c. Trigger input by circulation pump

220 Vac 50 Hz Trigger input cable

Circulation pump

Connect the Trigger cable as the example on the site with the traditional circulation pump, to check the flow is present in the pipe line.

220 Vac 50 Hz SVP Circulation pump Flow check sensor

Do not use the Trigger cable to the circulation pump if the SVP (speed variable pump) is present, but use the Flow check sensor, pay attention to disable or enable the function in the setting menu.

5. SPECIFICATIONS

Specifications EASY-CH-DPHRXG / EASY-CH-DPHRXGS
Range 0-14pH; -99 -1000mV; 0...+55°C
Accuracy ±0.02pH, ±3mV; ±0.5°C
Device precision ± 0.1 pH; ±10mV; ±1°C
Flow rate pump (l/h) 1.5 l/h
Flow rate pump (SPA model only) 1.5 l/h
Power supply 220-240 VAC 50-60 Hz
Consumption (W) 28 Watt
Probe calibration Automatic
Pump state Pause – Supply
Max. back-pressure 1.5bar
Relay contact (number 3) 250 Vac 10A (resistive load)
Fuse500 mA (fast)
Weight3.5 Kg
Dimensions (W–H–D)212 x 303 x 113 mm

HAYWARD Easy Chem - SPECIFICATIONS - 1

natural_image Back view of a black electronic device casing with mounting holes and internal grid patterns (no visible text or symbols)

HAYWARD Easy Chem - SPECIFICATIONS - 2

natural_image Simple black rectangular object with two vertical legs at the bottom, no visible text or symbols.

HAYWARD Easy Chem - SPECIFICATIONS - 3

natural_image Front view of a black industrial electrical control unit with two circular buttons and a digital display (no visible text or symbols)

HAYWARD Easy Chem - SPECIFICATIONS - 4

6. SETUP AND OPERATION

The device is designed to be connected to a protected outlet at all times. The EasyChem Double must not be disconnected unless the pool equipment is undergoing maintenance or the pool is to be closed (wintering).

Assuming that the chemical balance of the water is within the recommended ranges, the device can be started up.

6a. Description of the home screen and defaults parameters

HAYWARD Display V Button to decrease the value Escape button OK Enter button Button to increase the value Easy Chem -

Default parameters

n Item Value
1 Language EN
2 Set Point value 7.4 pH700 mV
3 Dosing method Acid (pH) Low (Redox)
4 OFA Time OFF
5 Calibration Full
6 Circulation pump ON
7 Dosing type PROP; Relay Aux1, Aux2, Aux3 are disabled
8 IN Freq.OFF
9 Flow Input NO (normally open) (Reed Sensor)
10 Power ON DelayOFF
11 Flow DelayOFF

Measure alarm ranges default parameters

nItemLimits
1Temp. Measure min+ 10°C
2Temp. Measure Max+ 38°C
3pH Measure min6 pH
4pH Measure Max8.0 pH
5ORP Measure min+450 mV
6ORP Measure Max+850 mV

To restore the default parameters, follow the steps below:

• Power off the device unit
- Keep V and ∧ pressed and power on the unit
• The unit will display Init.default
- Select the unit to reset - WiFi module or dosing system
- Press V or A to select Yes or no
- Select Yes and press OK to restore the default parameters.

6b. Main menu

Press OK 3 seconds to run the main menu and then use and select one of the following functions:

6b1. Calibration 6b2. Setup
HAYWARD Easy Chem - 6b. Main menu - 1

flowchart
graph LR
    A["Main Menu CALIBRATION"] --> B["Main Menu SETUP"]
    B --> C["Main Menu ADVANCED"]
    C --> D["Main Menu PRIMING"]

6b3. Advanced 6b4. Priming

Main menu - Press OK 3 seconds. At the entry of each menu item, the parameter can be directly modified by using ∧ and ∨.

Confirmation of the current setting and switching to the next item is done by pressing the OK button.

The menu has a circular structure: once you arrive at last item, the confirmation of the parameter set by pressing determines the return to the first menu item.

HAYWARD Easy Chem - 6b. Main menu - 2

6b1. Calibration menu

The Calibration menu consists of three (3) sub-menus:

• pH: Calibration routine
- ORP: Calibration routine
• Temperature: Calibration routine

Note: please check sections 6f, 6g and 6h for step by step calibration routine.

Main Menu CALIBRATION

CALIBRATION PH

CALIBRATION ORP

CALIBRATION TEMPERATURE

6b2. Setup menu

The Setup menu consists of two (2) sub-menus:

• pH
- ORP

Scroll through the menu using a single item and confirm with .

HAYWARD Easy Chem - 6b2. Setup menu - 1

Main Menu SETUP

SETUP PH

SETUP ORP

HAYWARD Easy Chem - 6b2. Setup menu - 2

HAYWARD®

Setup pH

The Setup pH menu consists of six (6) sub-menus:

  • Setpoint pH: 7.4pH (range 6-8pH) Set the pH value keep in the pool.
  • Setpoint Type: Acid (Acid/Alkaline) Type dosing product to regulate the pH value. Acid: add chemical Acid product, Alka: add chemical Alkaline product.
  • Temperature: 25°C (Automatic compensation by manual value or set Automatic to use an external temperature sensor)
  • OFA Alarm: Over Feed alarm value in minutes value (OFF...60'), it is a maximum relay time activation.
  • Alarm Min: Setting the minimum pH alarm value (0,0...6,0 pH)
    • Alarm Max: Setting the Maximum pH alarm value (8,0...14,0 pH)
    • *Time ON: range from 5" to 360" (Default Time ON: 180")
    • *Time OFF: range from 5" to 360" (Default Time OFF: 360")

\*Notes:

The items Time ON and Time OFF will be displayed if the Timed dosing method is set in the Advanced menu.

Time OFF must be equal to or greater than Time ON.

Scroll through the menu using

HAYWARD Easy Chem - \*Notes: - 1

select the item and confirm with

HAYWARD Easy Chem - \*Notes: - 2

Setup ORP

The Setup ORP menu consists of five (5) sub-menus:

  • Setpoint ORP: 700mV (range 400-850mV) Set the ORP value keep in the pool.
  • Setpoint Type: Low (LOW/HIGH) Type of dosing regulation. Low: add chemical chlorine, High: reduce chemical chlorine.
  • OFA Alarm: Over Feed alarm value in minutes value (OFF...60'), it is a maximum relay time activation.
  • Alarm Min: Setting the minimum ORP alarm value (0...600 mV)
    • Alarm Max: Setting the Maximum ORP alarm value (800...1000 mV)
    • *Time ON: range from 5" to 360" (Default Time ON: 180")
    • *Time OFF: range from 5" to 360" (Default Time OFF: 360")

\*Notes:

The items Time ON and Time OFF will be displayed if the Timed dosing method is set in the Advanced menu.

Time OFF must be equal to or greater than Time ON.

SETUP PH

SETPOINT PH 7.4pH

SETPOINT TYPE ACID (pH-)

TEMPERATURE 25°C

OFA ALARM OFF

ALARM MIN 6.0pH

ALARM MAX 8.0pH

TIME ON 180"

TIME OFF 360"

SETUP ORP

SETPOINT ORP 700mV

SETPOINT TYPE LOW

OFA ALARM OFF

Scroll through the menu using your item and confirm with .

HAYWARD Easy Chem - \*Notes: - 1

HAYWARD Easy Chem - \*Notes: - 2

6b3. Advanced menu

The Advanced menu consists of the following sub-menus:

• Language: Set language menu (EN, IT, ES, DE, FR)

- Circulating Pump: Enable/Disable input signal from pool circulation pump High voltage 230Vac.

- In Freq.: Flow rate function measure, Enable/Disable input frequency signal from water meter pulse sender, set Pulse/Liter or Liter/Pulse.

- Calib. pH: Double ore Single point and Reference calibration, Double point 7pH and 4pH, Single point 7pH, reference pH value.

- Calib. ORP: Single point and Reference calibration, Single point 465 mV, reference ORP mV value.

- Calib. Temp: Reference calibration by external hand held controller.

- Dosing Type pH: Proportional, Timed, On/Off, Off (disable pump)

- Dosing Type ORP: Proportional, Timed, On/Off, Off (disable pump)

- Aux Relay: Set AUX1: pH or disable; AUX2: ORP or Disable

- Password: Set value to enable password of setting menu. (0000 disable)

- Reset Cal: Menu to reset the last Calibration value, and reload the factory settings

- Reset all Param.: Menu to reset all parameter and reload the factory settings

- Control Panel: Menu to show the input raw signal, pH in mV raw signal, ORP in mV raw signal, Temperature in Ohm raw signal.

- Reed: set logic sensor NO/ NC (normally open/close)

- P.On Delay: Power ON delay function is a countdown timer Off...60' to delay dosing actions when switch on the device. During this period the system wait the chemical measure stabilizing the dosing pump are disable.

- Flow Delay: Flow delay function is a countdown timer Off...60' to delay dosing actions when the flow rate will be restable, and during this time the chemical measure will be stabilizing the dosing pump are disable.

ALARM MIN 600mV

ALARM MAX 800mV

TIME ON 180"

TIME OFF 360"

Main Menu ADVANCED

ADVANCED MENU LANGUAGE

Scroll through the menu using your home item and confirm with .

HAYWARD Easy Chem - 6b3. Advanced menu - 1

To exit the menu, press the instrument will display the question "save?";

confi rm with OK

For not saving, select NO using a confirm with .

HAYWARD Easy Chem - 6b3. Advanced menu - 2

6b4. Priming menu

Priming menu dosing pumps, please select the dosing pump to start the priming action.

Main Menu PRIMING

pH:Off ORP:Off

HAYWARD Easy Chem - 6b4. Priming menu - 1

flowchart
graph LR
    A["pH Pump press UP to run"] --> C["pHPump=0n Priming"]
    B["ORP Pump press DOWN to run"] --> C
    C --> D["OPR Pump=0n"]
    E["Priming menu"] --> C
    F["Empty spacing"] --> C

6c. View Parameters

HAYWARD Easy Chem - 6c. View Parameters - 1

flowchart
graph LR
    A["Empty spacing"] --> B["7.4 pH"]
    C["Temperature or Totalizer Flowrate measure, Moreover: <br>    • Circ. Pump OFF<br>    • Flow (Reed OFF)<br>    • Remote standby deletes all measures<br>    • Flow Delay with time countdown"] --> B
    B --> D["pH measure<br>    <br>    • <br>    • pH<br>    • OFA pH"]
    B --> E["ORP measure<br>    <br>    • <br>    • OFA ORP"]

6d. Standby system

Press (V) (e.g., us) the system sets in StandBy mode; all functions are disabled

6e. Reset OFA Timer

Press OK (seconds) to reset OFA Alarm or press to reset OFA Alarm.

6f. pH probe calibration

Always begin the calibration procedure with a calibration reset.

H₂O 1

HAYWARD Easy Chem - 6f. pH probe calibration - 2

natural_image Mechanical component diagram showing a piston and crankshaft assembly with an upward arrow (no text or symbols)

H₂O 3

7 pH 4

OK 5 s Set pH calibration

OK 7 pH calibration

HAYWARD Easy Chem - 6f. pH probe calibration - 7

natural_image Gray hourglass icon with '90 s' label and number 7 in top-right corner (no other text or symbols)

7 pH Quality probe 100 %

H₂O 9

10 4 pH

OK 4 pH calibration

90 s 12

4 pH Quality probe 100 %

H₂O 14

HAYWARD Easy Chem - 6f. pH probe calibration - 15

natural_image Mechanical assembly diagram showing a valve and bolt with a downward arrow indicating motion (no text or symbols)

16 OK Save and exit

Note: If you have selected the "1 point cal.", the calibration will be made only in 1 point using the 7pH buffer solution.

Reference calibration

CAL Reference 7.2 pH The unit will flash a pH value Set the value measured with the instrument Ex. 7.4 pH

CAL Reference 7.4 pH OK

6g. ORP probe calibration

Always begin the calibration procedure with a calibration reset.

H₂O 1

HAYWARD Easy Chem - Reference calibration - 4

natural_image Mechanical component diagram showing a shaft and gear assembly with an upward arrow (no text or symbols)

H₂O 3

4 465mV

OK 5 5 s Set Redox calibration

OK 465 mV calibration

7 90 s

465 mV Quality probe 100 %

H₂O 9

HAYWARD Easy Chem - Reference calibration - 12

natural_image Mechanical assembly diagram showing a pipe inserted into a gear with a downward arrow indicating motion (no text or symbols)

11 OK Save and exit

Reference calibration
CAL Reference 720 mV The unit will flash a redox value Set the value measured with the instrument Ex. 750 mV

CAL Reference 750 mV OK

6h. Temperature probe calibration

Always begin the calibration procedure with a calibration reset.

CAL Reference 26°C

The unit will flash a temperature value

Set the temperature value measured with the instrument Ex. 27°C

CAL Reference 27°C

HAYWARD Easy Chem - 6h. Temperature probe calibration - 1

7. DOSING METHOD

Setpoint = 7.4 pH

Dosing mode = Acid

Prop. Band= 1.0 pH (* Fixed value)

HAYWARD Easy Chem - DOSING METHOD - 1

line | Input pH measure | Dosing period 10 minutes | | ---------------- | ------------------------ | | 7.2 | 30% | | 7.48 | 30% | | 8.6 | 95% |

Setpoint = 700 mV

Dosing mode = Low

Prop. Band= 250mV (* Fixed value)

HAYWARD Easy Chem - DOSING METHOD - 2

line | Input ORP measure | Dosing period 10 minutes | | ----------------- | ------------------------ | | 400 | 95% | | 750 | 95% | | >750 | 30% |

Alarm for the pH/ORP Set Point

When the alarm band is set, a work window is created. If the allowed limits are exceeded the alarm relay closes

and remains closed until the measurement is reset or OK pressed to deactivate the alarm.

When the OFA time (Over Feed Alarm) is set, the dosing time of Set Point pH/ORP in time is controlled with two alarms:

  • First alarm at 70% of the time set is seen on the display, the alarm relay closes.
  • Second alarm at 100% of the time set is seen on the display and the alarm relay closes and the pH/ORP pump is blocked.

HAYWARD Easy Chem - Alarm for the pH/ORP Set Point - 1
seconds) to reset OFA Alarm or press

HAYWARD Easy Chem - Alarm for the pH/ORP Set Point - 2
o reset OFA Alarm.

HAYWARD Easy Chem - Alarm for the pH/ORP Set Point - 3

line | Parameter | Value | | --------------- | ----- | | ALR Band | High | | Set Point pH | Medium| | ALR Band | Low | | Pump pH | Low | | ALR Relay | High |

(*1 Measure Alarm Ranges - fixed values)

n Item Limits
1 Temp. Measure min + 10°C
2 Temp. Measure Max + 38°C

8. ALARMS

Alarm Display Actions to do
Level*only active measuresLev Low- Restore Product tank
Out of Range measureALARM BAND- Replace or check the measure probe- Restore measure
OFA First Alarm(time >70%)OFA_1 blinkingHAYWARD Easy Chem - ALARMS - 1- Press 3 seconds to reset or press for seconds to reset HAYWARD Easy Chem - ALARMS - 2 HAYWARD Easy Chem - ALARMS - 3
OFA Second Alarm(time 100%)OFA_2HAYWARD Easy Chem - ALARMS - 4- Press 3 seconds to reset or press for seconds to reset HAYWARD Easy Chem - ALARMS - 5 HAYWARD Easy Chem - ALARMS - 6
Flow RateFlow- Restore Flow Rate
Calibration FunctionCALIBRATION ERR- Restore Probe or Buff er solution and repeat calibration procedure
System ErrorParameter S errorOK- Press Restore Default parameter- Broken Unit
Alarm measure (*1)High Measure Low Measure- Adjust the chemical concentration

(*1 Measure alarm ranges default parameters)

n Item Limits
1 Temp. Measure min + 10°C
2 Temp. Measure Max + 38°C
3 pH Measure min 6 pH
4 pH Measure Max 8.0 pH
5 ORP Measure min +450 mV
6 ORP Measure Max +850 mV

9. SERVICING

During the first 10-15 days, your system will require more attention:

- Check that the pH remains at the ideal level (7.2 to 7.4).

- If the pH is exceptionally unstable and uses a lot of acid, check the alkalinity (see table).

If the balance is highly unstable, contact your pool installer/builder.

REMEMBER that the system needs a certain amount of time to adapt to your pool and will require additional chemicals during the first 3-5 days.

The pool must be regularly maintained and the skimmer baskets emptied whenever necessary.

Also check that your filter is not clogged.

DOSING PUMPS: Regularly check the acid level to ensure that the pump does not run dry. The dosing pump must be checked and serviced at regular intervals. The Santoprene tube of the peristaltic pump has a lifetime of 2 years. We recommend that you change it once a year.

Servicing the probe

The probe must be clean and free from oil, chemical deposits and contamination to function properly. As it is in continuous contact with the water in the pool, the probe may need to be cleaned weekly or monthly, depending on the number of bathers and other specific pool characteristics. A slow response, more frequent pH calibration and inconsistent readings indicate that the probe needs to be cleaned.

To clean the probe, turn off the power to the EasyChemDouble.

Unplug the probe connector from the control box, unscrew the probe and carefully remove it from the chamber. Clean the probe bulb with a soft toothbrush and regular toothpaste.

A household washing-up liquid detergent may also be used to remove any oil.

Rinse with fresh water, replace the Teflon tape on the threads, and reinstall the probe.

If the probe continues to give inconsistent readings or requires excessive calibration after it has been cleaned, it should be replaced. The lifetime of the probes is 1 year. We recommend that you calibrate them every month during the season the pool is in use.

Wintering

The EasyChemDouble, probe and pool piping run the risk of being damaged if the water freezes. In regions that experience long periods of cold weather, be sure to drain all the water from the pump and filter and from the supply and return pipes before winter. Do not remove the control box.

Probe storage

The end of the probe must always be in contact with water or a solution of KCl. If it is removed from the measuring chamber, it should be stored in the plastic cap provided (filled with water). If the storage cap has been mislaid, the probe should be stored separately in a small glass or plastic container with its end immersed in water.

The probe must always be in a frost-free environment.

Hose replacement procedure (valid for 2 and 3 rollers)

HAYWARD Easy Chem - Probe storage - 1

natural_image Cross-sectional diagram of a mechanical component with symmetrical arch and base (no text or symbols)

Open the pump's lid and release the hose by pulling the left connector upward.

HAYWARD Easy Chem - Probe storage - 2

natural_image Cross-sectional view of a mechanical component with no visible text or symbols

Position the roller at 7h05, turning it in the direction of the circular arrow.

HAYWARD Easy Chem - Probe storage - 3

natural_image Mechanical component diagram showing a housing with attached cable and mounting base (no text or symbols)

Completely release the left connector, holding it taut towards the outside, and turn the roller in the direction of the circular arrow so that the hose is freed up to the right connector.

HAYWARD Easy Chem - Probe storage - 4

natural_image 3D rendered mechanical component with central bore and mounting base (no text or symbols)

Position the roller at 7h05, turning it in the direction of the circular arrow.

HAYWARD Easy Chem - Probe storage - 5

natural_image Mechanical component diagram showing a pipe fitting and housing (no text or symbols)

Insert the left connector into the relative housing and pass the hose under the roller's guide. Turn the roller in the direction of the circular arrow, simultaneously accompanying the hose into the pump's head, until the right connector is reached.

HAYWARD Easy Chem - Probe storage - 6

natural_image Cross-sectional diagram of a mechanical component with no visible text or symbols

Close the pump's lid and press its surface hard so that it is properly locked into place.

Storing the pump after use

HAYWARD Easy Chem - Storing the pump after use - 1

natural_image Cross-sectional view of a mechanical component with no visible text or symbols

When the regulation device must be stored, clean water should be pumped through the hose in order to rinse it.

Then position the roller at 7h05, turning in the direction indicated by the circular arrow.

These two precautions will facilitate the subsequent reactivation of the unit.

10. TROUBLESHOOTING GUIDE

No display

Check that the On / Off switch is on.

Check the connection cable between the display and the control box.

Check the power supply: 210-230 V\~ 50 Hz.

Check the dosing method if the OFA Alarm show continuously.

Chech the lifetime probe by calibration method.

Chech the running pump by Priming function.

If the problem persists, contact your pool installer/builder.

11. ENVIRONMENTAL INFORMATION

Provision regarding professional waste from electrical and electronic equipment (WEEE). In compliance with directive 2012/19/EU regarding the management of waste from electrical and electronic equipment, this pump must be disposed of at a waste sorting site.

==> for more information contact your dealer.

Good management of waste from electrical and electronic equipment contributes to the prevention of damage to the environment and human health.

HAYWARD Easy Chem - ENVIRONMENTAL INFORMATION - 1

LIMITED WARRANTY

All HAYWARD products are covered for manufacturing defects or material defects for a warranty period of 3 years as of date of purchases. Any warranty claim should be accompanied by evidence of purchase, indicating date of purchase. We would therefore advise you to keep your invoice.

The HAYWARD warranty is limited to repair or replacement, as chosen by HAYWARD, of the faulty products, provided that they have been subjected to normal use, in compliance with the guidelines given in their user guides, provided that the products have not been altered in any way, and provided that they have been used exclusively with HAYWARD parts and components. The warranty does not cover damage due to frost and to chemicals. Any other costs (transport, labour, etc.) are excluded from the warranty.

HAYWARD may not be held liable for any direct or indirect damage resulting from incorrect installation, incorrect connection, or incorrect operation of a product.

In order to claim on a warranty and in order to request repair or replacement of an article, please ask your dealer.

No equipment returned to our factory will be accepted without our prior written approval.

Wearing parts are not covered by the warranty.

The wearing parts of the salt water chlorinator listed below must be maintained according to their estimated service life:

Seal set = 2 years

Santoprene tube = 2 years

Roller = 2 years

HAYWARD Easy Chem - Wearing parts are not covered by the warranty. - 1

natural_image Abstract geometric logo with stylized letter H inside a circular frame (no text or symbols)

HAYWARD®

HAYWARD Easy Chem - HAYWARD® - 1

natural_image Two black electronic devices with control buttons and indicator lights, no visible text or symbols on the devices themselves.

CE EAC UK CA

HAYWARD Easy Chem - HAYWARD® - 2

natural_image Two line icons: one open with an information symbol, the other closed with an open book (no text or symbols present)

EasyChem Double and Plug Version | pH - ORP

BETRIEBSANLEITUNG

BITTE BEWAHREN SIE DIESES HANDBUCH ZUM SPÄTEREN NACHSCHLAGEN AUF.

natural_image Diagram showing a mechanical assembly process with three stages: brick wall, vertical support, and final view (no text or symbols)

Achtung! 2m

natural_image Exterior view of a modern industrial air purifier unit with dual fans and control panel (no visible text or symbols)

HAYWARD Easy Chem - EasyChem Double and Plug Version | pH - ORP - 2

natural_image Close-up of a black industrial device with two metallic components and a digital display, showing a red arrow labeled 't' pointing to a component (no text or symbols on the device itself)

HAYWARD Easy Chem - EasyChem Double and Plug Version | pH - ORP - 3

natural_image Interior view of a black industrial machine with internal components and a red arrow indicating rotation (no visible text or symbols)

Anschluss des

Etiketts:

ORP probe pH probe PT100 NC Frequency INPUT GND LEV 1 LEV 2 Reed +5Vdc SIGN LEV 1 (pH) LEV 2 (ORP) TRIGGER INPUT 220 Vac Relay pH Relay ORP Relay ALARM L N 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 000135964 R1-0

natural_image Diagram of a cable being inserted into a device, showing cable routing and wire movement (no text or symbols)

STROMVERSORGUNGSKABEL

HAYWARD Easy Chem - Etiketts: - 2

natural_image Diagram of a mechanical device with a curved cable and labeled component (no text or symbols)

HAYWARD Easy Chem - Etiketts: - 3

natural_image Simple line drawing of a cable with a box and ring, no text or symbols present

WICKELN SIE DAS KABEL FÜR EINE UMDREHUNG

natural_image Front view of a black electronic device casing with mounting holes and internal circuitry (no visible text or symbols)

HAYWARD Easy Chem - Etiketts: - 4

natural_image Silhouette of two people standing in front of a dark rectangular object (no text or symbols visible)

HAYWARD Easy Chem - Etiketts: - 5

natural_image Front view of a black industrial electrical control unit with two circular components and a digital display (no visible text or symbols)

HAYWARD Easy Chem - Etiketts: - 6

natural_image Mechanical component diagram showing a shaft and gear assembly with an upward arrow (no text or symbols)

H₂O 3

7 pH

OK 5 s Set pH calibration

OK 7 pH calibration

7 90 s

7 pH Quality probe 100 %

H₂O 9

10 4 pH

OK 4 pH calibration

90 s 12

4 pH Quality probe 100 %

H₂O 14

HAYWARD Easy Chem - Etiketts: - 19

natural_image Mechanical assembly diagram showing a bolt and pipe assembly with a downward arrow indicating motion (no text or symbols)

16 OK Save and exit

CAL Reference 7.4 pH OK

natural_image Mechanical component diagram showing a shaft and gear assembly with an upward arrow (no text or symbols)

H₂O 3

465mV

OK 5 s Set Redox calibration

OK 465 mV calibration

7 90 s

465 mV Quality probe 100 %

H₂O 9

HAYWARD Easy Chem - Etiketts: - 29

natural_image Mechanical component diagram showing a lever and gear assembly (no text or symbols)

11 OK Save and exit

CAL Reference 750 mV OK

The unit will flash a temperature value

Set the temperature value measured with the instrument Ex. 27°C

CAL Reference 27°C

HAYWARD Easy Chem - Etiketts: - 32

7. DOSIERVERFAHREN

Sollwert = 7.4 pH

Dosiermodus = Säure

Prop. Band= 1,0 pH (* fester Wert)

Sollwert = 700 mV

natural_image Cross-sectional view of a mechanical component with symmetrical arms and mounting holes (no text or symbols visible)
natural_image Cross-sectional view of a mechanical component with no visible text or symbols
natural_image Mechanical component diagram showing a housing with a cable and housing assembly (no text or symbols visible)
natural_image 3D rendered mechanical component with central bore and mounting base (no text or symbols)
natural_image Mechanical assembly diagram showing a pipe fitting and housing component (no text or symbols)
natural_image Cross-sectional diagram of a mechanical component with symmetrical U-shaped housing and mounting points (no text or symbols)
natural_image Mechanical component diagram showing a curved housing with internal components and mounting holes (no text or symbols)
natural_image Abstract geometric logo with stylized letter H inside a dark circular frame (no text or symbols)

HAYWARD®

HAYWARD Easy Chem - HAYWARD® - 1

natural_image Two black rectangular electronic devices with control buttons and indicator lights, no visible text or symbols on the surfaces.

CE EAC UK CA

HAYWARD Easy Chem - HAYWARD® - 2

natural_image Two line icons: one open with an information symbol, the other closed with an open book (no text or symbols present)
natural_image Diagram showing four sequential steps of a mechanical or structural assembly on a brick wall, with no visible text or symbols.

HAYWARD Easy Chem - HAYWARD® - 3

natural_image Exterior view of a modern industrial air purifier unit with dual fans and control panel (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 4

natural_image Close-up of a black industrial device with attached components and a red arrow indicating rotation (no text or symbols visible)

HAYWARD Easy Chem - HAYWARD® - 5

natural_image Interior view of a black industrial machine with internal components and a red arrow indicating rotation (no text or symbols visible)
natural_image Diagram of a cable being inserted into a device, showing cable routing and cable attachment (no text or symbols)
natural_image Simple line drawing of a cable being inserted into a device, no text or symbols present

HAYWARD Easy Chem - HAYWARD® - 6

natural_image Simple line drawing of a cable with a box and loop, no text or symbols present

ENVUELVA EL CABLE UNA VUELTA

natural_image Front view of a black electronic device casing with mounting holes and internal circuitry (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 7

natural_image Exterior view of a modern office building (no signage)

HAYWARD Easy Chem - HAYWARD® - 8

natural_image Front view of a black industrial water heater unit with dual circular components and a digital display (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 9

Menú principal SETUP

HAYWARD Easy Chem - HAYWARD® - 10

SETUP pH

SETUP ORP

HAYWARD Easy Chem - HAYWARD® - 11

HAYWARD®

Setup pH

Menú principal AVANZADO

natural_image Mechanical component diagram showing a shaft and gear assembly with an upward arrow (no text or symbols)

H₂O 3

7 pH

OK 5 s Set pH calibration

OK 7 pH calibration

7 90 s

7 pH Quality probe 100 %

H₂O 9

10 4 pH

OK 4 pH calibration

90 s 12

4 pH Quality probe 100 %

H₂O 14

HAYWARD Easy Chem - HAYWARD® - 13

natural_image Mechanical assembly diagram showing a valve and gasket with a downward arrow indicating motion (no text or symbols)

16 OK Save and exit

CAL Reference 7.4 pH OK

natural_image Mechanical component diagram showing a shaft and gear assembly with an upward arrow (no text or symbols)

H₂O 3

465mV

OK 5 s Set Redox calibration

OK 465 mV calibration

7 90 s

465 mV Quality probe 100 %

H₂O 9

HAYWARD Easy Chem - HAYWARD® - 23

natural_image Mechanical assembly diagram showing a pipe inserted into a gear with a downward arrow indicating motion (no text or symbols)

OK 11 Save and exit

CAL Reference 750 mV OK

The unit will flash a temperature value

Set the temperature value measured with the instrument Ex. 27°C

CAL Reference 27°C

HAYWARD Easy Chem - HAYWARD® - 26

natural_image Cross-sectional diagram of a mechanical component with symmetrical supports and central hub (no text or symbols)
natural_image Cross-sectional view of a mechanical component with no visible text or symbols
natural_image Mechanical component diagram showing a housing with attached cable and mounting base (no text or symbols)
natural_image 3D rendered mechanical component with central bore and mounting base (no text or symbols)
natural_image Mechanical component diagram showing a shaft and housing assembly (no text or symbols)
natural_image Cross-sectional diagram of a mechanical component with symmetrical U-shaped housing and mounting base (no text or symbols)
natural_image Cross-sectional view of a mechanical component with no visible text or symbols
natural_image Abstract geometric logo with stylized letter H inside a circular frame (no text or symbols)

HAYWARD®

RATWAID Buy Chain RATWAID Buy Chain

CE EAC UK CA
HAYWARD Easy Chem - HAYWARD® - 2

natural_image Two line icons: one open with an information symbol, the other closed with an open book (no text or symbols present)

EasyChem Double and Plug version | pH · ORP

MANUEL DU PROPRIÉTAIRE

VEUILLEZ CONSERVER CE MANUEL POUR TOUTE RÉFÉRENCE ULTÉRIEURE

HAYWARD Easy Chem - EasyChem Double and Plug version | pH · ORP - 1

HAYWARD Easy Chem - EasyChem Double and Plug version | pH · ORP - 2

L'ÉQUIPEMENT EST DESTINÉ À ÊTRE UTILISÉ UNIQUEMENT DANS LES PISCINES

3b. Installation murale

natural_image Diagram showing four sequential steps of a mechanical or structural assembly on a brick wall, with no visible text or symbols.

Avertissement ! 2m

HAYWARD Easy Chem - 3b. Installation murale - 2

HAYWARD®

natural_image Exterior view of a modern industrial air purifier unit with dual fans and control panel (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 1

natural_image Close-up of a black industrial device with attached components and a red arrow indicating rotation (no text or symbols)

HAYWARD Easy Chem - HAYWARD® - 2

natural_image Interior view of a black industrial machine with internal components and a red arrow indicating rotation (no visible text or symbols)
natural_image Diagram of a mechanical assembly with a lever and conveyor belt, no text or symbols present

CÂBLE D'ALIMENTATION


HAYWARD Easy Chem - HAYWARD® - 3

natural_image Simple line drawing of a mechanical device with a curved arrow indicating motion (no text or symbols)


HAYWARD Easy Chem - HAYWARD® - 4
ENROULER LE CÂBLE POUR UN TOUR

natural_image Front view of a black electronic device casing with mounting holes and internal circuitry (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 5

natural_image Silhouette of two people standing in front of a dark rectangular object (no text or symbols visible)

HAYWARD Easy Chem - HAYWARD® - 6

natural_image Front view of a black industrial electrical control unit with two circular buttons and a digital display (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 7

HAYWARD Easy Chem - HAYWARD® - 8

HAYWARD®

6. CONFIGURATION ET FONCTIONNEMENT

flowchart
graph LR
    A["Menu principal ÉTALONNAGE"] --> B["Menu principal SETUP"]
    B --> C["Menu principal ADVANCED (AVANCÉ)"]
    C --> D["Menu principal AMORÇAGE"]

Menu principal SETUP

SETUP PH

SETUP ORP

HAYWARD Easy Chem - CONFIGURATION ET FONCTIONNEMENT - 1

HAYWARD®

Configuration pH

Menu principal AMORÇAGE

pH:Off ORP:Off

HAYWARD Easy Chem - Configuration pH - 1

natural_image Mechanical component diagram showing a shaft and gear assembly with an upward arrow (no text or symbols)

H₂O 3

7 pH

OK 5 s Set pH calibration

OK 7 pH calibration

7 90 s

7 pH Quality probe 100 %

H₂O 9

10 4 pH

OK 4 pH calibration

90 s 12

4 pH Quality probe 100 %

H₂O 14

HAYWARD Easy Chem - Configuration pH - 14

natural_image Mechanical component diagram showing a valve assembly with a bolt and pipe, no text or symbols present

16 OK Save and exit

CAL Reference 7.4 pH OK

natural_image Mechanical component diagram showing a shaft and gear assembly with an upward arrow (no text or symbols)

H₂O 3

465mV

OK 5 5 s Set Redox calibration

OK 465 mV calibration

7 90 s

465 mV Quality probe 100 %

H₂O 9

HAYWARD Easy Chem - Configuration pH - 24

natural_image Mechanical assembly diagram showing a pipe inserted into a gear with a downward arrow indicating motion (no text or symbols)

11 OK Save and exit

CAL Reference 750 mV OK

The unit will flash a temperature value

Set the temperature value measured with the instrument Ex. 27°C

CAL Reference 27°C

HAYWARD Easy Chem - Configuration pH - 27

7. MÉTHODE DE DOSAGE

Point de consigne = 7,4 pH

Mode de dosage = Acide

natural_image Cross-sectional diagram of a mechanical component with symmetrical arms and mounting holes (no text or symbols)
natural_image Cross-sectional view of a mechanical component with no visible text or symbols
natural_image Mechanical component diagram showing a housing with a curved pipe and mounting base (no text or symbols)
natural_image 3D rendered mechanical component with central bore and mounting base (no text or symbols)
natural_image Mechanical assembly diagram showing a connector with tubing and housing (no text or symbols)
natural_image Cross-sectional diagram of a mechanical component with symmetrical U-shaped housing and mounting base (no text or symbols)
natural_image Cross-sectional diagram of a mechanical component with no visible text or symbols
natural_image Abstract geometric logo with stylized letter H inside a circular frame (no text or symbols)

HAYWARD®

HAYWARD Easy Chem - HAYWARD® - 1

natural_image Two black rectangular electronic devices with control buttons and indicator lights, no visible text or symbols on the surfaces.

CE EAC UK CA

HAYWARD Easy Chem - HAYWARD® - 2

natural_image Two line icons: one open with an information symbol, the other closed with an open book (no text or symbols present)

EasyChem Double e Plug | pH · ORP

MANUALE D'USO

natural_image Diagram showing four sequential steps of a mechanical or structural assembly on a brick wall, with no visible text or symbols.

HAYWARD Easy Chem - HAYWARD® - 3

natural_image Exterior view of a modern industrial air purifier unit with dual fans and control panel (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 4

natural_image Close-up of a black industrial device with attached components and a red arrow indicating rotation (no text or symbols visible)

HAYWARD Easy Chem - HAYWARD® - 5

natural_image Interior view of a black industrial machine with internal components and a red arrow indicating rotation (no visible text or symbols)
natural_image Simple line drawing of a mechanical device with a lever and cable, no text or symbols present
natural_image Simple line drawing of a mechanical device with a curved arrow indicating motion (no text or symbols)


HAYWARD Easy Chem - HAYWARD® - 6
AVVOLGERE IL CAVO DI UN GIRO

natural_image Front view of a black electronic device casing with mounting holes and internal circuitry (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 7

natural_image Exterior view of a modern office building (no signage)

HAYWARD Easy Chem - HAYWARD® - 8

natural_image Front view of a black industrial electrical control unit with two circular components and a digital display (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 9

HAYWARD Easy Chem - HAYWARD® - 10

HAYWARD®

6. IMPOSTAZIONE E FUNZIONAMENTO

TEMPO DI ACCENSIONE 180"

TEMPO DI ACCENSIONE 180"

natural_image Mechanical component diagram showing a piston and crankshaft assembly with an upward arrow (no text or symbols)

H₂O 3

7 pH 4

OK 5 s Set pH calibration

OK 7 pH calibration

7 90 s

7 pH Quality probe 100 %

H₂O 9

10 4 pH

OK 4 pH calibration

90 s 12

4 pH Quality probe 100 %

H₂O 14

HAYWARD Easy Chem - TEMPO DI ACCENSIONE 180" - 13

natural_image Mechanical assembly diagram showing a valve and gasket with a downward arrow indicating motion (no text or symbols)

16 OK Save and exit

CAL Reference 7.4 pH OK

natural_image Mechanical component diagram showing a shaft and gear assembly with an upward arrow (no text or symbols)

H₂O 3

4 465mV

OK 5 5 s Set Redox calibration

OK 465 mV calibration

7 90 s

465 mV Quality probe 100 %

H₂O 9

HAYWARD Easy Chem - TEMPO DI ACCENSIONE 180" - 23

natural_image Mechanical assembly diagram showing a pipe with a valve and a gear base, no text or symbols present

11 OK Save and exit

CAL Reference 750 mV OK

The unit will flash a temperature value

Set the temperature value measured with the instrument Ex. 27°C

CAL Reference 27°C

HAYWARD Easy Chem - TEMPO DI ACCENSIONE 180" - 26

7. METODO DI DOSAGGIO

Setpoint= 7,4 pH

line | pH Level | Banda ALR | | -------------- | --------- | | Setpoint pH | 0 | | Banda ALR | 0 | | Pompa pH | 0 | | Banda ALR | 0 |
natural_image Cross-sectional view of a mechanical component with symmetrical U-shaped housing and mounting holes (no text or symbols)
natural_image Cross-sectional view of a mechanical component with no visible text or symbols
natural_image Mechanical component diagram showing a housing with a cable and housing assembly (no text or symbols visible)
natural_image 3D rendered mechanical component with central bore and mounting base (no text or symbols)
natural_image Mechanical component diagram showing a pipe fitting and housing (no text or symbols)
natural_image Cross-sectional diagram of a mechanical component with symmetrical U-shaped housing and mounting base (no text or symbols)
natural_image Cross-sectional diagram of a mechanical component with no visible text or symbols
natural_image Abstract geometric logo with stylized letter H inside a circular frame (no text or symbols)

HAYWARD®

HAYWARD Easy Chem - HAYWARD® - 1

natural_image Two black rectangular electronic devices with control buttons and indicator lights, no visible text or symbols on the surfaces.

CE EAC UK CA

HAYWARD Easy Chem - HAYWARD® - 2

natural_image Two line icons: one open with an information symbol, the other closed with an open book (no text or symbols present)
natural_image Diagram showing four sequential steps of a mechanical or structural assembly on a brick wall, with no visible text or symbols.

HAYWARD Easy Chem - HAYWARD® - 3

natural_image Exterior view of a modern industrial air purifier unit with dual fans and control panel (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 4

natural_image Close-up of a black industrial device with attached components and a red arrow indicating rotation (no text or symbols visible)

HAYWARD Easy Chem - HAYWARD® - 5

natural_image Interior view of a digital water dispenser with black casing and control panel, showing internal components and a red arrow indicating rotation (no text or symbols visible)
natural_image Diagram of a mechanical assembly with numbered label (2) and directional arrow, no readable text or symbols present.

CABO DE ALIMENTAÇÃO


HAYWARD Easy Chem - HAYWARD® - 6

natural_image Simple line drawing of a mechanical device with a curved arrow indicating motion (no text or symbols)


HAYWARD Easy Chem - HAYWARD® - 7
ENROLAR O CABO COM UMA VOLTA

natural_image Front view of a black electronic device casing with mounting holes and internal circuitry (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 8

natural_image Silhouette of a person standing in front of a dark rectangular object with a horizontal line above (no text or symbols visible)

HAYWARD Easy Chem - HAYWARD® - 9

natural_image Front view of a black industrial electrical control unit with two circular components and a digital display (no visible text or symbols)

HAYWARD Easy Chem - HAYWARD® - 10

HAYWARD Easy Chem - HAYWARD® - 11

HAYWARD®

flowchart
graph LR
    A["Menu principal AVANÇADO"] --> B["Menu principal PRIMING [ESCORVAMENTO"]]

HAYWARD Easy Chem - HAYWARD® - 1

HAYWARD Easy Chem - HAYWARD® - 2

natural_image Mechanical component diagram showing a piston and crankshaft assembly (no text or symbols)

H₂O 3

7 pH

OK 5 s Set pH calibration

OK 7 pH calibration

7 90 s

7 pH Quality probe 100 %

H₂O 9

10 4 pH

OK 4 pH calibration

12 90 s

4 pH Quality probe 100 %

H₂O 14

HAYWARD Easy Chem - HAYWARD® - 15

natural_image Mechanical assembly diagram showing a valve and gasket with a downward arrow indicating motion (no text or symbols)

16 OK Save and exit

CAL Reference 7.4 pH OK

natural_image Mechanical component diagram showing a valve and gasket assembly (no text or symbols)

H₂O 3

465mV

OK 5 s Set Redox calibration

OK 465 mV calibration

7 90 s

465 mV Quality probe 100 %

H₂O 9

HAYWARD Easy Chem - HAYWARD® - 25

natural_image Mechanical component diagram showing a lever and gear assembly (no text or symbols)

11 OK Save and exit

CAL Reference 750 mV OK

The unit will flash a temperature value

Set the temperature value measured with the instrument Ex. 27°C

CAL Reference 27°C

HAYWARD Easy Chem - HAYWARD® - 28

7. MÉTODO DE DOSAGEM

Setpoint = 7.4 pH

natural_image Cross-sectional view of a mechanical clamp or bracket component (no visible text or symbols)
natural_image Cross-sectional view of a mechanical component with no visible text or symbols
natural_image Mechanical component diagram showing a housing with attached cable and mounting holes (no text or symbols)
natural_image 3D rendered mechanical component with central bore and mounting base (no text or symbols)
natural_image Mechanical component diagram showing a shaft and housing assembly with no visible text or symbols
natural_image Mechanical component diagram showing a U-shaped housing with mounting feet and central bore (no text or symbols)
natural_image Cross-sectional view of a mechanical component with no visible text or symbols
Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : HAYWARD

Model : Easy Chem

Category : Pool controller