C.A 5005 - Multimeter CHAUVIN ARNOUX - Free user manual and instructions
Find the device manual for free C.A 5005 CHAUVIN ARNOUX in PDF.
| Product Type | Analog needle multimeter |
| Brand | Chauvin Arnoux |
| Model | C.A 5005 |
| Dimensions (L × W × H) | 56 × 105 × 160 mm |
| Weight | 500 g |
| Power Supply | 1 9V battery (type 6F22 or 6LF22 alkaline) |
| Battery Life | 10,000 measurements of 10 seconds (alkaline battery) |
| Display | Analog with 6 scales (V, A, Ω, dB) and anti-parallax mirror |
| Main Functions | AC/DC voltmeter, direct DC and AC ammeter with clamp, ohmmeter, audible continuity test, decibel measurement |
| Overvoltage Category | Category III, 600 V with respect to earth |
| Protection Rating | IP53 (sealing and dust) |
| Insulation | Double insulation (Class II), no earth connection required |
| Fuse Protection | 1 A HPC fuse (6.3×32 mm, 500 V, 10 kA) and 10 A HPC fuse (6.3×32 mm, 500 V, 10 kA) |
| Included Accessories | Mini-clamp MN89, set of 2 test lead probes, 9V battery, instruction manual |
| Optional Accessories | Carrying case, pouch, replacement fuses, additional leads |
| Operating Temperature | -10 °C to +55 °C |
| Relative Humidity | < 90 % |
| Cleaning | Disconnect the device, clean with a cloth slightly dampened with soapy water, wipe and dry quickly |
| Maintenance | Periodic metrological verification recommended; repair by Chauvin Arnoux approved services |
| Warranty | 3 years after delivery |
Frequently Asked Questions - C.A 5005 CHAUVIN ARNOUX
User questions about C.A 5005 CHAUVIN ARNOUX
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Multimeter in PDF format for free! Find your manual C.A 5005 - CHAUVIN ARNOUX and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. C.A 5005 by CHAUVIN ARNOUX.
USER MANUAL C.A 5005 CHAUVIN ARNOUX
3-TENSIONS CONTINUES ET ALTERNATIVES
Meaning of the symbol

Warning! Please refer to the User's Manual before using the instrument. In this User's Manual, the instructions preceded by the above symbol, should they not be carried out as shown, can result in a physical accident or damage the instrument and the installations.
Meaning of the symbol

This device is protected by a double insulation or by a reinforced insulation. No linking is required from the protection earth terminal to ensure electrical safety.
Thank you for purchasing a C.A 5005 Multimeter.
To get the best service from this instrument:
- read this user's manual carefully,
- respect the safety precautions detailed.

SAFETY PRECAUTIONS
Never use on networks at a voltage above 600V in relation to the earth. This multimeter of overvoltage category III, satisfies the severe requirements of reliability and availability corresponding to industrial and domestic permanent installations (c.f. IEC 664-1).
Use indoors in environments of degree of pollution at most equal to 2 (c.f. IEC 664-1), of temperature from -10 to +55^ and of relative humidity less than 90% .
Respect the value and the type of fuses or there is a risk of damage to the instrument and cancellation of the warranty. - Fuse 1 A HBC (6.3 x 32 mm - 500 V - 10 kA) - Fuse 10 A HBC (6.3 x 32 mm -500 V -10 kA)
Use accessories in conformity with safety standards (NF EN 61010-2-031) of minimum voltage 600V and overvoltage category III.
Before any measurement, check the leads and the switch are in the correct position. When the order of magnitude of a measurement is not known, place the selector switch on the highest range then lower progressively, if necessary, to the appropriate range: the reading must be made, preferably, in the upper 2/3 of the scale.
■Never measure resistances on a live circuit.
■For DC current measurements (without current clamp), switch off the power supply of the circuit before connecting or disconnecting your multimeter.
For AC current measurements the use of the MN89 miniclamp supplied with the multimeter is mandatory.
To open the battery compartment, the leads must be disconnected.
Never connect to the circuit to be measured if the battery compartment is not correctly closed.

CONTENTS
Page
1-Presentation 11
2- Description. 12
3 - DC and AC voltages (V DC and AC) 13
4- Decibels (dB) 13
5 - Currents (A DC and AC) 14
6 -Resistances (Ω) 15
7-Continuity sound test [ ] 15
8 - General specifications 15
9 - Accessories and spares (To order) 16
10 - Warranty 16
11 - Maintenance 17
12-Appendix 42
1 - PRESENTATION
The C.A 5005 multimeter is designed for the daily needs of professionals in electricity. It has the following functions:
- Voltmeter: voltage measurements (V DC and AC)
- DC ammeter: measurement of DC currents directly (A DC)
- AC ammeter: measurement of AC currents with current clamp
(AAC)
- Ohmmeter: resistance measurements ()
- Continuity sound test [ ]
It also allows the measurement of decibel levels (dB), on the AC voltmeter ranges.
The dial has a fuse test light (Fus.) and "Voltest" light to check the absence of voltage on ohmmeter.
NB: Always use accessories suited to the voltage and the overvoltage category of the circuit to measure (to NF EN 61010).

2 - DESCRIPTION
(See drawing in § 12 - APPENDIX)
① TERMINALS
4 mm safety terminals
COM: common, terminal that receives the black lead
V terminal that receives the red lead for voltages, resistances and AC voltages with miniclamp MN 89
10 A DC: terminal that receives the red lead for the 10 A DC ranges
mADC: terminal that receives the red lead for the mA DC ranges
2 ANALOGUE DISPLAY
The dial comprises 6 scales:
2 black scales, with antiparallax mirror, for V DC and AC (0.100 and 0.30) and for A DC (0.100)
1 green scale for (0.10k)
2 red scales for A AC with miniclamp MN 89 (0.3 and 0.10)
1 red scale for dB (0.22)
③ “FUS”LIGHT
Neon for testing the 1 A and 10 A fuses on current measurement.
If the «Fus.» light is lit, change the faulty fuse(s).
NB: This light requires the presence of a voltage ≥ 80V to come on.
(4) “VOLTEST” LIGHT
Voltage presence light on ohmmeter.
If the Voltest light comes on: presence of a voltage U > 15V AC/DC. Unplug the leads and cut off the voltage before making the resistance measurement.
SWITCH
Switch with 24 positions to select the functions and ranges.
12

3 - DC AND AC VOLTAGES
Connect the leads to the multimeter and connect in parallel to the circuit to be tested.
When the order of magnitude is not known, place the switch on the highest range then progressively lower to the appropriate range.
To get the voltage in V, multiply the value read on the appropriate scale by the reading coefficient shown in the table.
| V DC | 100mV(1) | 1 V | 3 V | 10 V | 30 V | 100 V | 300 V | 1000 V |
| Scale | 100 | 100 | 30 | 100 | 30 | 100 | 30 | 100 |
| Reading coefficient | x1(2) | x0,01 | x0,1 | x0,1 | x1 | x1 | x10 | x10 |
| Internal resistance(3) | 2kΩ | 20kΩ | 63,2kΩ | 200kΩ | 632kΩ | 2MΩ | 6,32MΩ | 6,32MΩ |
| Accuracy(4) | 2,5 % | 1,5 % | ||||||
| Permitted overload | 440 V | 1000 V(5) | 1500V(5) | |||||
(1) Common to the 50 A range (2) Direct reading in mV
(3) Specific R: 20 kΩ/V, except range 1000 V-R = 6.32 kΩ/V
(4) In % of the end of scale (5) For 15 seconds
| VAC | 10 V 30 V 100 V 300 V 100 V | ||||
| Scale | 100 | 30 | 100 | 30 | 100 |
| Reading coefficient | x0,1 | x1 | x1 | x10 | x10 |
| Internal resistance(1) | 63.2 kΩ | 200 kΩ | 632 kΩ | 2 MΩ | 6.32 MΩ |
| Accuracy(2) | 3 % | 2.5 % | |||
| Bandwidth (3) | 100 kHz | 50 kHz | 25 kHz | 1 kHz | |
| Permitted overload | 1000 V(4) | 1500 V(4) | |||
(1) Specific R: 6.32 kΩ/V
(2) In % of the end of scale
(3) From 20Hz to ..
(4) For 15 seconds
4 - DECIBELS
■ Reminder. The measurement of an AC voltage can be expressed in decibels (symbol dB). The decibel is the ratio of two quantities or levels. Level N, in dB, of a voltage U has the mathematical expression:
NdB)U(/0)I
。
= 20
U_o is the reference voltage of 0.775V AC for a power P_o of 1 mW on a load of 600
Use. Zero level of the red scale in dB corresponds to Uo = 0.775 for the 10VAC range. The reading is direct in dB for the 10VAC range from 0 to +22dB . For the other ranges, it is possible to read in dB (near value) by adding respectively:
+10dB on the 30 V AC range +30dB on the 300 V AC range
+20dB on the 100 V AC range +40dB on the 1000 V AC range
13

5 - DC AND AC CURRENTS
For DC currents: Always switch off the circuit to test before connecting the multimeter to the circuit. If the «Fus.» light comes on, change the faulty fuse(s) (Reminder: minimum voltage of 80 V). Connect the leads to the multimeter and connect in series to the circuit with:
-the red lead in the «mA DC» terminal, up to 1 A DC
-the red lead in the «10 ADC» terminal, from 1 ADC to 10 ADC
For AC currents: mandatory use of the MN89 miniclamp originally supplied with the multimeter. It outputs 100mV DC for 1 A AC. Measurement possible up to 240 A AC max. Connect the black plug of the miniclamp to the «COM» terminal and the connect the red plug to the "terminal.
Clamp on only one conductor with the MN 89. For the other specifications of the MN 89, refer to its user manual.
When the order of magnitude is not known, place the switch on the highest range then progressively lower to the appropriate range.
On DC current, switch off the power supply to the circuit ill before changing range.
To get the current in mA or A, multiply the value read on the appropriate scale by the reading coefficient shown in the table.
| A... | 50 μA(1) | 1 mA 10 mA 100 mA 1 A 10 A | Δ | |||
| Scale 100 | ||||||
| Reading coefficient | x 0.5 (2) | x 0.01 x | 0.1 x 1 x | 0.01 x 0.01 | 1 | |
| Voltage at terminals(3) | 100 mV | 50 mV 180 | mV 200 mV | V 800 mV | ||
| Accuracy(4) | 2.5% 5% | |||||
| Protection Fuse | 1 A HPC | Fuse 10 A HBC | ||||
Limitation 10 min On, 5 min Off up to +40^ max
(1)Commonolhe 100mVDC range
(2)Reading in A
(3) Without the leads Resistance of the pair of leads supplied: approx. 70 m
(4) In % of the end of scale
Above 200 A AC (nominal range of the MN 89 miniclamp) limit the time of use. Do not use the miniclamp on a network with a voltage greater than 600V in relation to the earth.
| A AC with miniclamp MN 89 | 1≤200 A~ | 200 A~<1≤240 A~ |
| Operating | Permanent(1) | 10 min On 10 min Off |
(1) With a frequency F ≤ 1 kHz and a crest factor FC < 3

| A~with miniclamp MN 89 | 3 A 10 | A 30 A 100 | A 300 A | (1) | |
| Scale(2) | 3 | 1 | 0 | 3 | 1 |
| Reading coefficient x | 1 x 1 x 10 x 10 x 100 | ||||
| Accuracy(3) | 10% 5% 2.5% | (4) | |||
| Protection ensured by the miniclamp MN89 (5) | |||||
(1) Range limitedto240AAC max by the miniclamp MN 89
(2)Marked with
(3) In %ofthaend of scale
(4) Up to 200 A AC nominal range of the MN 89 miniclamp
(5)Measuring240AACisequivalenttomeasuring24VDC
6 - RESISTANCES

Never test a resistance on a live circuit.
Connect to the terminals of the circuit or the component to test.

If the "Voltest" light comes on, presence of a voltage U > 15 V AC/DC. Unplug the leads immediately and cut off the voltage before making the resistance measurement.
To get the resistance in , multiply the reading on the scale (green) by the coefficient of the selected range: x1 or x100 .
| Ω | \( \times {1}^{\left( 1\right) } \) | x 100 |
| Measurement extend | 5 Ω to 10 kΩ | 500 Ω to 1 MΩ |
| Internal resistance | 140 Ω | 14 kΩ |
| End of scale current | 10 mA \( {100\mu }\mathrm{A} \) | |
| Accuracy(2) | 10% | |
| Permitted overload | 440 V | |
(1)Common to the continuity sound test function
(2) In % at mid-scale
7 - CONTINUITY SOUND TEST
■Connection and specifications: idem Resistances
Place the selector switch on the 1 function
Continuous audible beep emitted for a resistance R < 50
8 - GENERAL SPECIFICATIONS
8-1 Dimensions and weight
56× 105× 160mm 500g
15

8-2 Power supply
One battery 9 V (type 6F22 or 6LF22 alkaline)
Battery life:
- 10,000 measurements of 10 seconds with alkaline battery LR6
8-3 Environmental conditions
Temperature: use: -10°C to +55°C / storage: -40°C to +70°C
Relative humidity:
use: ≤ 90% RH / storage: ≤ 95% RH
Altitude: use < 2000m
8-4 Conformity with international standards
8-4-1 Electrical safety (NF EN 61010-1, ed. 95)
Double insulation : ■ installation category: III
Degree of pollution: 2 Rated voltage: 600 V
8-4-2 Electromagnetic compatibility
Emission and Immunity (NF EN 61326-A1, ed. 98)
8-4-3 Mechanical protection
Degree of watertightness (NF EN 60529, ed 92): protection index IP 53
9 - TO ORDER
Use the designations and references below.
C.A 5005 P01.1965.23E
Supplied with a pair of leads with prods,
1 battery 9 V and this User's manual.
Different measurement accessories widen the field of application or confer new functions on your multimeter.
Documentation on request.
NB: Always use accessories suited to the voltage and the overvoltage category of the circuit to measure (to NF EN 61010).
10 - WARRANTY
Our guarantee is applicable for three years after the date on which the equipment is made available (extract from our General Conditions of Sale, available on request).
16


11 - MAINTENANCE

For maintenance, use only specified spare parts. The manufacturer will not be held responsible for any accident occurring following a repair done other than by its After Sales Service or approved repairers.
11-1 Replacing the battery and the fuses
For your safety the leads must be disconnected from the multimeter before the battery cover is opened.
To open the cover, turn the screw 1/4 turn, anti-clockwise, using a coin or a screwdriver.
Replace the dead battery by one 9 V battery (type 6F22 or 6LF22).
Replace the faulty fuses respecting their value and their type: -Fuse 1 A HBC (6.3× 32mm - 500V - 10kA) -Fuse 10 A HBC (6.3× 32mm - 500V - 10kA)
■Refit the cover before using the multimeter.
11-2 Storage
If the multimeter is not put into service for a time exceeding 60 days, remove the batteries and store them separately.
11-3 Cleaning
The multimeter must be disconnected from any electrical source.
To clean the case, use a cloth slightly moistened with soapy water. Rinse with a damp cloth. Then, dry rapidly with a cloth or in a hot air stream.
11-4 Metrological check
It is essential that all measuring instruments are regularly calibrated.
For checking and calibration of your instrument, please contact our accredited laboratories (list on request) or the Chauvin Arnoux subsidiary or Agent in your country.
11-5 Repair
Repairs under or out of guarantee: please return the product to your distributor.

11-4 Verifica metrologicals