USER MANUAL RecoveryAir JetBoots THERABODY
- Product Overview
- What's in the Box
- Getting to Know the Device
- Using the Device
- How the Device Works
- Care, Cleaning, and Charging
EN 3-16
DE 17-30
FR 31-44
IT 45-58
ES 59-72
RU 73-86
FRCA 87-102
JetBoots Prime
1. Product Overview
Refresh and recover your legs with Therabody JetBoots Prime device—wireless, ultra-portable and uniquely designed to fold up like a pair of pants so you can take the science-backed benefits of pneumatic compression anywhere. Customize your treatment time and pressure to your body's needs, accelerating recovery, reducing muscle soreness and improving performance with maximum comfort and control. The device features our proprietary TruGrade Technology that delivers precise, sequential pressure to optimize circulation and remove metabolic waste more safely and effectively than competitors. So you can just sit back, relax and recharge your legs.
Intended Use
The Therabody JetBoots Prime device is an air compression therapy device intended to provide graduated pressure to the legs. The device is indicated for the temporary relief of minor muscle aches and pains, and for a temporary increase in blood circulation to the treated area in people who are in good health. The device simulates kneading and stroking of tissues by using an inflatable garment.
2.What's in the box
Two connected compression boots
2 meter USB-C to USB-C charging cable Note: A power adaptor is not included with this device. For best performance, use a USB-C power adaptor of at least 20W power capacity. Ensure it is from a certified manufacturer and has not suffered any structural damage.
Drawstring travel backpack
D Instruction manual



D

A
Figure 1. What's In the box

3. Getting To Know Your Device
Lead Compression Boot
Support Compression Boot
3 Console and Control Panel
a Navigation buttons (power, start/pause, pressure, time)
Indicators
USB-C Charging Port
Fabric Bridge
External pockets
6 Main Zippers
Pulling/Hanging Loops

Figure 2. the JetBoots Prime and control panel
4. Use the device
- Put on your boots. It is recommended that you wear shorts or thin pants while using the device. Reach a comfortable position, either sitting upright or lying down with your legs straight and feet in line with your waist. Unzip the compression boots completely and place each leg and foot inside each boot. Your legs should be in the center of each boot with your feet as close to the bottom as possible.
- Zip up the boots and keep them flat. Keep your legs straight and flat to prevent kinks and folds. Only hold the boots' edges when handling them. Avoid holding the device by the control panel. Zip both boots up.
"Do not use the device while standing or walking.




Figure 3. How to position oneself in the boots

- Tum on the device. To power on your device, press and hold the power button on the control panel. The LED indicators on the control panel will light up when the boots are on.
- Select your pressure intensity. Press the pressure button to toggle through four pressure intensity options: 25 mmHg, 50 mmHg, 75 mmHg, 100 mmHg. The pressure setting used in your last treatment will become the default setting for your next treatment. You can also add or subtract pressure during your treatment if desired.
- Select your treatment time. Press the time button to toggle through four time options: 20 minutes, 40 minutes, 60 minutes, or continuous. The time setting used in your last treatment will become the default setting for your next treatment.
- Start, pause, and exit treatment. Begin your treatment by pressing the start button. Press the pause button at any point during your treatment to pause and deflate the boots. Press the start button again to resume the treatment.
- Tum off the device. Press and hold the power button on the control panel to turn off the device. The device will automatically shut off after 10 minutes of inactivity, including completed or paused treatments. Unzip the compression boots fully and store.
- Lock the device. When not in use or cluring transport, lock the device to prevent turning it on accidentally. Turn off the device to activate the lock function. Quickly press the power button three times to initiate the lock function. Each LED light will flash sequentially to indicate that the device is locked. Quickly press the button three times again to unlock. Each LED light will flash sequentially (in the opposite direction) to indicate that the device is unlocked.
A Buttons Indicators
Power Button
Start/Pause/Stop Button
Time Button
Pressure Button



Pressure settings
Adjustable Pressure Range (25-100 mmHg)
in increments of 25 mmHg.

Time settings
4-time intervals (20, 40, 60, and continuous)

Figure 4. UI for settings
Battery Indicator
3 dots indicate full battery level
2 dots indicate mid battery level
1 dot indicates low battery level
5. How the Device Works
Pneumatic compression applies controlled pressure to the legs. The pressure applied through sequential inflation and deflation increases circulation, which enhances the body's natural healing processes, delivering more oxygen and nutrients to the muscles and transporting waste products from exercise away from the muscle. This increase in circulation also helps reduce swelling and fluid buildup, so you feel less stiff and sore, making this an effective method for recovery, particularly after physical activity.
JetBoots Prime TruGrade Technology
TruGrade Technology uses advanced, innovative technology to deliver precise pressure gradually up your legs, starting from your feet moving up to your thighs. Each of the JetBoots Prime compression boot's four air chambers overlap so pressure is delivered up the leg sequentially, without the risk of harmful pressure gaps. This optimizes circulation and removes metabolic waste more safely, and effectively. No matter the size of your leg, you will experience the same amount of pressure when selecting the same treatment setting as someone of a different size. Other systems deliver a set amount of pressure regardless of the leg size, which can lead to dangerous over-constriction and numbness. Our precise pressure control not only optimizes recovery, but provides a safer massage experience.
How the Sequential Cycle Works
During a sequential cycle, pressure is applied directionally, starting at the base (feet) and progressing upwards (torso). Once Chamber 1 inflates and reaches its preset pressure level, it holds the pressure and begins inflating Chamber 2. The device continues to hold pressure in each subsequent chamber until all four chambers are inflated. Once all four chambers are inflated, the device releases pressure.

Figure 5. Sequential cycle

6. Care, Cleaning, and Charging
Device Maintenance
The following maintenance instructions are important to ensure that your device continues to work as it was designed. Failure to follow these instructions may cause your device to stop working.
Cleaning the Device
The JetBoots Prime compression boots have non-porous medical grade material that helps prevent bacteria buildup and provides a resistant surface. The internal overlapping chambers provide a smooth surface to clean, meaning no chamber flaps where bacteria and odors can build up.
- Unzip the boots completely and lay them on a flat surface.
- Use a 70% isopropyl alcohol (IPA) cleaning solution to gently clean the inside of the boots. Never use oil-based cleaning products.
- Wipe down the device's exterior using a soft cloth that is free of debris to reduce surface contaminants.
- Ensure there is no remaining debris after fully cleaning the device.
- Once you have fully cleaned the device, fully pat dry the surface and hang to dry completely before continuing use.
Note: This product is NOT waterproof. Do not submerge the device in water or clean it under running water. Do not allow the device to come into contact with corrosive solutions, which can damage the device's appearance and function.
Charging the Device
- The device is USB-C enabled and comes with a 6.5 feet (2-meter) USB-C charging cable. The device can charge while the device is powered on or off.
- Connect one end of the charging cable to the USB-C charging port on the device. It is located on the side of the control panel. Plug the other side into a 20 W USB-C power adapter for optimal performance. Using a power adapter with less capacity may not allow you to use the device while charging.
- When the device is on, the LED indicators will show the battery status when the device is on or charging. Turn on the device to check the battery status.
1LED:Low battery
2 LEDs: Medium battery
3 LEDs: High battery
- The battery life of the device is approximately 180 minutes.
*Note: If using a charger from a third party, ensure that it is from a certified manufacturer and has not suffered any structural damage.
Store and Transport the Device
- Lay the device flat with the control panel facing up and allow compression boots to fully deflate.
- Fold the right boot over the left boot, like you would a pair of pants. Align all edges of the boots, so it looks like a single boot.
- Starting at the feet, fold them up 1/3 of the way up the leg.
- Fold once more, so that the fold reaches the top of the boots. DO NOT fold the boots too tight as this can damage the hoses and connection points.
- Slide the folded device into the drawstring bag. Pull the side strings outward to close the bag and wear as a backpack or pull the top strings to use as a carrying bag.
Ensure that the device is clean and dry before storing. To store and transport the device, keep the device in its drawstring travel backpack, and handle it gently. Store the device in a cool and dry place (Temperature: -4-158°F (-20°C - 70°C, Relative humidity: 10~93% RH), atmospheric pressure 190.0-1060 hPa. Do not store the device where temperatures may exceed 45°C/158°F, such as in direct sunlight or in a vehicle.






Figure 6. Folding Instructions

Troubleshooting
| SCENARIO POTENTIAL CAUSES POTENTIAL SOLUTIONS | ||
| Device does not turn on. | No electric power and no battery power in the device. | 1. Use the provided USB-C charging cable and a 20W USB-C power adapter to connect the device to a power outlet. Verify the device is charging or turning on. Note: using power adapters of lower power capacity will only allow the device to either charge or work, but not both simultaneously. 2. Check that the power adapter, charging cable, and the console are all properly connected. 3. If the device is not charging after verifying all of the above, the power adapter or the device may be damaged. If using another power adapter does not solve the issue, contact Therabody for further assistance. Visit https://www.therabody.com/us/en-us/support/support.html for the available contact methods. |
| Power adapter or USB-C charging cable damaged | ||
| Internal malfunction | ||
| Device automatically turns off and starts charging when connected to a USB-C power adapter. | The power adapter in use does not provide enough power to work and charge simultaneously | Check that the USB-C power adapter can provide at least 20W. If lower, allow the device to charge before using it. Contact Therabody for further assistance if the issue persists. Visit https://www.therabody.com/us/en-us/support/support.html for the available contact methods. |
| One or both compression boots do not fully inflate, and the device stops working. One yellow light is displayed. | Air leakage at the air hose connection or bladders. | 1. Inspect the air hose connections. Open the black zipper in the external pockets located in the internal side of the boots (under the "bridge"). 2. Check the air hoses are well connected to the air bladder nozzles. If loose, push and connect the air hose to the nozzle and check if the issue has been solved. 3. If the issue persists, the air bladders may be damaged. Contact Therabody for further assistance. Visit https://www.therabody.com/us/en-us/support/support.html for the available contact methods. |
| One or both boots do not fully inflate, and the device stops working. Two yellow lights are displayed. | Air cannot move through the air hoses or air bladders (kinking) | 1. Ensure both boots are worn properly and the legs are fully extended. Check that the device is not kinked, folded, or twisted. Note that if the size is too long, the device may kink and fold, leading to discomfort.2. Check the "bridge" area is not folded while in use. Please adjust the "bridge" or your sitting position to ensure it is not folded. This may lead to air hose kinks.3. Inspect the air hose connections for air hose kinks. Open the black zipper in the external pockets located on the internal side of the boots (under the "bridge").4. Contact Therabody for further assistance if the issue persists. Visit https://www.therabody.com/us/en-us/support/support.html for the available contact methods. |
| Device stops working and displays two yellow lights blinking. | Internal components are overheating | 1. Check the environmental conditions are within the specified range. Operating Temperature: 32-95°F (0-35°C); Relative Humidity: 30-85% RH; Atmospheric Pressure: 700-1060 hPa.2. Stop using the device and allow it to cool down for 15 to 20 mins.3. Contact Therabody for further assistance if the issue persists. Visit https://www.therabody.com/us/en-us/support/support.html for the available contact methods. |
| Device stops working and displays three yellow lights blinking. | Battery is overheating | |
| Device stops working and displays three yellow lights. | Internal component malfunction | Contact Therabody for further assistance. Visit https://www.therabody.com/us/en-us/support/support/support.html for the available contact methods. |
DO NOT attempt to troubleshoot by disassembling the device as this will void the warranty and may cause injury.
EN

Warnings and Guidance (Precautions and Contraindications)
Background
Therabody products are designed to unlock the body's natural ability to achieve health and well-being. Through science and technology, the Therabody portfolio of products allows people to access the therapeutic benefits of different natural phenomena to meet both their needs and preferences. There will be times when it is advisable to modify how devices are used (precautions) or times when it is not appropriate to use certain devices (contraindications). Read the following safety information for the device in its entirety prior to use.
Important Safety Information
General device Use
This device is intended for use by people in good health. This device is contraindicated against and should not be used by or on anyone with a history of epilepsy, seizures or cardiopathy.
The device is not recommended for anyone with an electronic implanted device (such as a pacemaker), cardiac arrhythmia, tumors, or acute episodes of inflammatory diseases. The device is not recommended for those who have arteriosclerosis, thromboses, or implants in the body region being treated.
Please consult your physician prior to using the device if you are pregnant and/or nursing. Immediately stop using the device at the first sign of discomfort. The device IS NOT a toy. If you have any medical considerations please consult your doctor before using the device.
Safety, Precautions, and Contraindications
Specific Therapies
These recommendations are derived from consultation with medical experts and published research regarding precautions and contraindications as of the printing date. For up-to-date information, please visit us online at https://www.therabody.com/us/en-us/precautions-and-contraindications.html.
Precautions:
Due care is required in these circumstances and device use may need to be modified. Consult with a medical professional if you currently have or suspect you may have any of the following conditions or if you have any questions.
Hypertension (controlled)
- Osteopenia
Osteoporosis
Pregnancy
Diabetes
Varicose veins
Bony prominences or regions
Abnormal sensations (e.g., numbness)
Sensitivity to pressure
Recent injury or surgery
- Scoliosis or spinal deformity
Medications that may alter client sensations
- Due to movement of fluids in the body when using the device, use with caution on patients with heart disease. High pressure is not recommended for patients who have peripheral occlusion disease.
Contraindications:
The following are circumstances where the potential risks may outweigh the benefits. Consult a medical professional before use.
- Skin rash, open wounds, blisters, local tissue inflammation, bruises, or tumors
Deep vein thrombosis, osteomyelitis
Bone fracture or myositis ossificans
Hypertension (uncontrolled)
Acute or severe cardiac, liver, or kidney disease
- Neurologic conditions resulting in loss or altered sensation
- Direct application to the face, throat, or genitalia
- Bleeding disorders
Recent surgery or Injury
Connective tissue disorders
- Peripheral vascular Insufficiency or disease
Medications that thin the blood or alter sensations
- Direct pressure over surgical site or hardware
- Direct pressure over eyes or throat
Extreme discomfort or pain felt by client
- Severe scoliosis or spinal deformity
- Pacemaker, ICD, or history of embolism
- Allergy to device material (Nylon)
Pregnancy/nursing
Abnormal sensations (ex. numbness)
Cancer/tumors
Epilepsy
Cardiopathy (heart disease
Limited Warranty and Customer Service
For full warranty information, please visit www.therabody.com/warranty.
Customers who are in need of product support should visit
https://www.therabody.com/us/en-us/support/support.html for the available contact methods.
FDA-Cleared
Limited Warranty Only With Authorized Retailer Purchase
© 2024 Therabody, Inc. All Rights Reserved.
Patents at www.therabody.com/patents
Manufactured and Distributed by:
Therabody, Inc.
1640 S Sepulveda, Suite 300
Los Angeles, CA 90025
In the Australian market, the JetBoots Prime warranty is 1 year.
For Australian Market Only: The purchase of this device is accompanied by a limited warranty. For
complete details about the rights afforded to you under this warranty, visit https://therabody.com/anz/en-au/warranty-support/.
Imported and Distributed By:
Therabody Australia Pty Ltd
Waterman Business Centre
175 Maroondah Highway
Suite 16, C307, Level 3
Ringwood Vic 3134 Australia
Importers:
Therabody International Limited
5th Floor, 40 Mesipil Road,
DO4C2N4 Dublin, Ireland
Therabody UK Limited
Portsoken House 155-157
The Minorities, London
United Kingdom, EC3N1LJ
UNITWARNINGS
READ ALL INSTRUCTIONS AND CAUTIONARY MARKINGS IN THIS MANUAL, ON THE CHARGER, AND ON THE DEVICE BEFORE USING OR CHARGING THE Therabody JetBoots Prime DEVICE AND ITS ACCESSORIES.
When using the device, the following basic precautions should always be adhered to:
- USE ONLY AS INSTRUCTED. Use the device as described in the Therabody JetBoots Prime User Manual only. The device is intended for over-the-counter use. The device is not intended to diagnose, cure, or prevent diseases. Use only Therabody recommended accessories and replacement parts. The device is an advanced mechanical tool with electric components. If the device and its accessories are
not used or maintained properly, there is a risk of fire, electric shock, or Injury. Failure to use and maintain the device and corresponding attachments in accordance with the instructions outlined in this manual will void the warranty and may result in product damage or physical Injury. The device is intended for commercial and home use.
-
NOT FOR CHILDREN. The device and its accessories are not intended for use by anyone under the age of 18. The device and accessories are not intended for use by individuals with reduced physical, sensory, or reasoning capabilities or lack of experience and knowledge, unless given supervision or instruction concerning use of the device by a person responsible for their safety. The device is not a toy. Advise children not to play with the device or accessories.
-
DO NOT STAND WHILE USING THE DEVICE. Do not attempt to stand up while using the device. Standing up while using the device can lead to discomfort, falling, and damage to the device's internal components.
-
CHARGING. If your device doesn't turn on or the battery indicator displays a low battery level, please charge before first use. Charge the battery using the USB-C cable (included). When you use the charging cable (included), make sure its USB-C connector is fully inserted into a compatible power adapter before you plug the adapter into a power outlet. You can also charge the battery with third party cables that are compliant with USB-C and with applicable country regulations and international and regional safety standards. Other adapters may not meet applicable safety standards, and charging with such adapters could pose a risk of death or injury. Do not force the cable into place. Fully charge the device at least once every six months to prevent device and battery damage. Unplug the power adapter and charging cable when not in use.
-
DO NOT OVERCHARGE. Do not leave the device connected to the power adapter for more than one hour after the battery has been fully charged. The battery includes a protection system to avoid the risk of overcharging. However, overcharging may reduce its life over time.
-
CHARGING LOCATIONS. Charge the device with a power adapter and charging cable. Avoid skin contact when the USB-C Is plugged into a power source as it may cause discomfort or injury. Do not sleep or sit on the USB-C connector. The device should be charged indoors in a well-ventilated, dry location. Do not charge the device outdoors, in a bathroom, or within 10 feet (3.1 meters) of a bathtub or pool. Do not use the device or charger on wet surfaces, and do not expose the charger to moisture, rain, or snow. Do not use the device or its compatible charger in the presence of explosive atmospheres (gaseous fumes, dust, grain, metal powders, or flammable materials). Sparks may be generated, possibly causing a fire. 7. UNPLugging THE CHARGER AND CARE. Pull the plug, not the cord, to reduce the risk of damage to the electrical plug and cord. Never carry the charger by its cord. Keep the cord away from heat, oil, and sharp edges. Do not stretch the charger cord or place the cord under strain. Do not handle the charger, terminals, or the device with wet hands. Always unplug this appliance from the electrical outlet Immediately after using and before cleaning. Store cable to ensure it is not stepped on, tripped over, or otherwise subjected to damage or stress. Do not use a power adapter or charging cable that has received a sharp blow, been dropped, run over, or damaged in any way. For long-term storage, store with a fully charged battery.
-
USAGE. Discontinue use of the device if you are feeling pain or discomfort. If at any point during the treatment you feel pain or discomfort beyond what's expected from a compression device or a massage, stop the treatment Immediately and remove the device. Discontinue use of the device if it overheats or becomes uncomfortably hot. Do not place any objects between your skin and the device.
-
DO NOT OPERATE UNDER BLANKET AND PILLOW. Excessive heating can occur and cause fire, electric shock, or injury.
-
DEVICE CARE AND SERVICE. Handle the device with care. The device contains electronic
components that can be damaged if dropped, burned, punctured, or crushed. Do not disassemble

the device or attempt to repair the device yourself. Disassembling the device may damage it and may cause injury. If the device is damaged or if malfunctions occur, contact Therabody Customer service. The warranty will be void if the device, batteries, or charger are disassembled or if any parts have been removed. Do not use if damaged as these may cause injury. Do not puncture or damage the device. Puncturing the device can lead to Incorrect operation. The device is NOT waterproof. The device is not machine washable. Do not place or store the device where it can fall or be pulled into a tub or sink. Do not place in or drop into water or other liquid. Do not reach for an appliance that has fallen into or come into contact with water. Unplug immediately. Clean the device according to the instructions found in the "Cleaning" section in the User Manual.
-
BATTERY CHEMICALS CAUSE SERIOUS BURNS. Never allow the internal battery to come into contact with the skin, eyes, or mouth. If a damaged battery leaks chemicals, use rubber or neoprene gloves to dispose of it. If skin is exposed to battery fluids, wash with soap and water and rinse with vinegar. If eyes are exposed to battery chemicals, immediately flush with water for 20 minutes and seek medical attention. Remove and dispose of contaminated clothing.
-
DO NOT SHORT CIRCUIT. A battery will short circuit if a metal object makes a connection between the positive and negative contacts on the battery or the 16V connector. Do not place a battery near anything that may cause a short circuit, such as coins, keys, or nails in your pocket. A short-circuited battery may cause fire and personal injury.
-
STORING THE DEVICE AND ACCESSORIES. Store in a cool, dry place. Only charge the device when the ambient temperature is between -4 - 158^ (-20^ - 70^) . Do not store the device, or charging cable where temperatures may exceed 158^(45^) , such as in direct sunlight, in a vehicle, or in a metal building during the summer. Avoid tightly rolling the device, as this may damage the device's internal components. 14. DEVICE DISPOSAL. This device contains a lithium-ion battery, and care must be taken upon disposal of the device. Before disposal of this device, please review your local laws and requirements surrounding Lithium-Ion Battery disposal. The preferred method of disposal is recycling the whole device.
Reporting adverse events to FDA
MedWatch is the Food and Drug Administration's (FDA) program for reporting serious reactions, product quality problems, therapeutic inequivalence/failure, and product use errors with human medical products, including drugs, biologic products, medical devices, dietary supplements, infant formula, and cosmetics. If you think you or someone in your family has experienced a serious reaction to a medical product, you are encouraged to take the reporting form to your doctor. Your health care provider can provide clinical information based on your medical record that can help the FDA evaluate your report. However, we understand that for a variety of reasons, you may not wish to have the form filled out by your healthcare provider, or your health care provider may choose not to complete the form. Your health care provider is not required to report to the FDA. In these situations, you may complete the Online Reporting Form yourself.
You will receive an acknowledgement from the FDA when your report is received. Reports are reviewed by FDA staff. You will be personally contacted only if we need additional information.
Submitting adverse event reports to FDA
Use one of the methods below to submit voluntary adverse event reports to the FDA:
- Report Online at: www.accessdata.fda.gov/scripts/medwatch/index.cfm?action=reporting.home
- Consumer Reporting Form FDA 3500B, Follow the Instructions on the form to either fax or mail it in for submission. For help filling out the form, see MedWatchLearn. The form is available at: www.fda.gov/downloads/aboutFDA/reportsmanualisforms/forms/ucm349464.pdf
-
Call FDA at 1-800-FDA-1088 to report by telephone.
-
Reporting Form FDA 3500 commonly used by health professionals. The form is available at: www.fda.gov/downloads/aboutFDA/reportmanualsforms/forms/ucm163919.pdf
GARMENT CARE
Meanings of symbols
Labels
SYMBOLS DESCRIPTION LOCATION
IP 22 Degree of protection against ingress of water On rating label

Read instructions before use On rating label

Level of protection type BF applied part On rating label

Therabody, Inc. 1640 S. Sepulveda Blvd. Suite 300 Los Angeles, CA 90025

Unique Device Identification (UDI)
On rating label

Separate collection for waste electrical and electronic equipment. Note: For more information about disposal of equipment, its parts and accessories, please contact your local distributor.
On rating label and packaging
This device is type BF applied equipment . It complies with Medical Electrical Safety Standards (IE(60601-1).
This device also complies with Medical EMC Standard (IEC 60601-1-2). The has been tested and found to comply with the electromagnetic compatibility (EMC) limits for medical devices to IEC 60601-1-2: 2014. These limits are designed to provide reasonable protection against harmful interference in atypical medical installation.
WARNING:
1' t t the other equipment should be observed to verify that they are operating normally.
2* WNG: Use of accessorles. transducers and cables other than those specifed or provided by the manufacturer this equipment could result in increased electromagnetic emissions or decreased electromagnetic immuity of this equipment and result in proper operation.
3*WARNING: Portable RF communications equipment (including peripherals such as antenna cables) should be used no closer than 30 cm (12 inches) to any part of the ME EQUIPMENT, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result
GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC EMISSION
| The JetBoots Prime is intended for use in the electromagnetic environment specified below. The customer or the user of JetBoots Prime should assure that it is used in such an environment. | ||
| EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT - GUIDANCE | ||
| RF emissions CISPR 11 | Group 1 | The JetBoots Prime uses RF energy only for its internal function. There for, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. |
| RF emissions CISPR 11 | Class B | The JetBoots Prime suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. |
| Harmonic emissions IEC 61000-3-2 | Class A | |
| Voltage function / flicker emissions IEC 61000-3-3 | Complies | |
GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
| The JetBoots Prime is intended for use in the electromagnetic environment specified below. The customer or the user of the JetBoots Prime should assure that it is used in such an environment. | |||
| IMMUNITY TEST IEC 60601 TEST LEVEL | COMPLIANCE LEVEL ELECTROMAGNETIC ENVIRONMENT - | GUIDANCE | |
| Electrostatic discharge (ESD)IEC 61000-4-2 | ±8 kV contact±2 kV, ±4 kV, ±8 kV, ±15 kV air | ±8 kV contact±2 kV, ±4 kV, ±8 kV, ±15 kV air | Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %. |
| Electrostatic transient / burstIEC 61000-4-4 | ±2kV for power supply lines±1kV for input/output lines | N/A | Mains power quality should be that of a typical commercial or home environment. |
| SurgeIEC 61000-4-5 | ±1kV differential mode±2kV common mode | N/A | Mains power quality should be that of a typical commercial or home environment. |
| Voltage dips, short interruptions and voltage variations on power supply input linesIEC 61000-4-11 | 0% UT: 0.5 cycle g) At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°0 % UT: 1 cycle and 70 % UT; 25/30 cycles Single phase: at 0°0 % UT: 250/300 cycle | N/A | Mains power quality should be that of a typical commercial or home environment. If the user of the JetBoots Prime requires continued operation during power mains interruptions, it is recommended that the JetBoots Prime be powered from an uninterruptible power supply or a battery. |
| Power frequency (50/60 Hz magnetic fieldIEC 61000-4-8 | 30 A/m 30 A/m Power frequency magnetic fields should be at levels | characteristic of a typical location in a typical commercial or home environment. | |
| NOTE U1is the a.c. mains voltage prior to application of the test level. | |||

The JetBoots Prime is intended for use in the electromagnetic environment specified below. The customer or the user of the JetBoots Prime should assure that it is used in such an environment.
| IMMUNITY TEST | IEC 60801 TEST LEVEL COMPLIANCE LEVEL ELECTROMAGNETIC ENVIRONMENT - GUIDANCE | ||
| Conducted RF IEC 61000-4-6 | 3 Vrms 150kHz to 80MHz6 V In ISM and amateur radio bands between 0,15 MHz and 80 MHz | 3 Vrms 150kHz to 80MHz6 V In ISM and amateur radio bands between 0,15 MHz and 80 MHz | Portable and mobile RF communications equipment should be used no closer to any part of the JetBoots Prime, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.Recommended separation distance: d=[3.5/V1]√P d=[12/V2]√P d=[3.5/E1]√P d=[7/E1]√P80 MHz to 800 MHz 800 MHz to 2.7 GHz |
| Radiated RF IEC 61000-4-3 | 10 V/m80MHz to 2.5GHz385MHz-5785MHzTest specifications for ENCLOSURE PORT IMMUNITY to RF wireless communication equipment (Refer to table 9 of IEC 60601-1,2,2014) | 10 V/m80MHz to 2.5GHz385MHz-5785MHzTest specifications for ENCLOSURE PORT IMMUNITY to RF wireless communication equipment (Refer to table 9 of IEC 60601-1,2,2014) | |
a The ISM industrial, scientific and technological standards between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 MHz, 14.0 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to 21.4 MHz, 24.99 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz. b Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy to assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the JetBoots Prime is used exceeds the applicable RF compliance level above, the JetBoots Prime should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the JetBoots Prime. c Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3W / m .
Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM - for EQUIPMENT and SYSTEMS
| Recommended separation distances between portable and mobile RF communications equipment and the JetBoots Prime | ||||
| The JetBoots Prime is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the JetBoots Prime can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the JetBoots Prime as recommended below, according to the maximum output power of the communications equipment | ||||
| Rated maximum output of transmitter W | Separation distance according to frequency of transmitter m | |||
| 150 kHz to 80 MHz outside ISM and amateur radio bands d=[3.5/V1]√P | 150 kHz to 80 MHz in ISM and amateur radio bands d=[12/V2]√P | 80 MHz to 800 MHz d=[12/V2]√P | 80 MHz to 800 MHz d=[7/E1]√P | |
| 0.01/0.12/0.20/0.035/0.07 | ||||
| 0.10/0.38/0.63/0.11/0.22 | ||||
| 1 | 12 | 2.00 | 0.35 | 0.70 |
| 10 | 3.8 | 6.32 | 110 | 2.21 |
| 100 | 12 | 20.00 | 35 | 70 |
JetBoots Prime
1. Produktübersicht
JetBoots Prime TruGrade Technology
Therabody International Limited
5th Floor, 40 Mespill Road,
DO4C2N4 Dublin, Ireland
Therabody UK Limited
Portsoken House 155-157
The Minorities, London
United Kingdom, EC3N1LJ
WARNHINWEISE
Unique Device Identification (UDI)

JetBoots Prime TruGrade Technology
Therabody International Limited
5e etage,40 Mespil Road
DO4C2N4 Dublin, Irland
Therabody UK Limited
Portsoken House 155-157
h 150 Mhz and 80 MHz are 3.5, 4.7, 6.9, 8.2, 10.3, 12.5, 14.7, 16.6, 18.7, 20.8, 22.9, 24.6, 26.8, 28.9, 30.4, 32.0, 34.0, 35.8, 37.6, 39.5, 41.4, 43.3, 45.2, 46.7, 48.8, 50.0, 51.6, 53.0, 54.0, 55.0, 56.0, 57.0, 58.0, 59.0, 60.0, 61.0, 62.0, 63.0, 64.0, 65.0, 66.0, 67.0, 68.0, 69.0, 70.0, 71.0, 72.0, 73.0, 74.0, 75.0, 76.0, 77.0, 78.0, 79.0, 80.0, 81.0, 82.0, 83.0, 84.0, 85.0, 86.0, 87.0, 88.0, 89.0, 90.0, 91.0, 92.0, 93.0, 94.0, 95.0, 96.0, 97.0, 98.0, 99.0, 100.0 and the same as for the previous example of the same frequency in the case of a single FM radio with a frequency of 1MHz (the frequency is set to 1MHz because the frequency is not given).
JetBoots Prime TruGrade Technology
Therabody International Limited
5th Floor, 40 Mespil Roa
DO4C2N4Dublin, Irlanda
Therabody UK Limited
Portsoken House 155-157
The Minorities, Londra,
JetBoots Prime TruGrade Technology
Therabody International Limited
5th Floor, 40 Mespil Road
DO4C2N4 Dublin, Irlanda
Therabody UK Limited
Portsoken House 155-157
The Minorities, Lc
HNNNNSAAGTURNTB
TOYOCPTPOCACTBA.1A0CTX6EHH HANHYHJN PON3BODHTENHOCTA
HOTOBYHIS NUTTER USB-C MPNPGTBOE NEHNE 20 B.
y6eTec,0000000000000000000000000000000000000000000000000
HHEHMETKAKX-ANHOOCTPXYPHNXNDBEHN



D

A
PcyHOK1.HTOBKopo6ke?

3.3HAKOMCTBO CYCTPOICTBOM
BeyuikomnpocHbcanor
2 BeDombI KOMPpeCIOHHbI Canor
3 KoHcIb nHaHEnb ynpaBHeHHa
KhoIkynpaBneHnB (BKnIOueHnE,CTapT/nay3a,DaNBeHnE,BpeM)
HnDnKaTOpbI
3apnnoPTUSB-C
4 TkaHeBaI nepeMbHka"MocT"
5 KapMaHbI
6 OCHOBHeIe 3aCTeKKn-MoHnHn
Tn(3aHHXMOXHO TAYHybNINNCNONB3OBaTBnnoDBuMBAHHYCTPOICTBA)

PucyHok 2.JetBoots Prime naneB ynpaBneHH
4.Испльбоване усторICTBa
- HadeBaHHe MAnKet. PekOMeHTyETCRNIOIb3OaBbYcTPOrCTBO HOBePbX 0oTbY IIN TOHKX 6pOJK. PImMITE yDooHoE HoIOJKeHHe, CJIg BepTKaIbHO IIN JIeKa, BblpIMBA HorN, CTynHnepeKOpTncom. MoHocTHo PaCCTeRMOHIO H2 KOMPiPeCCnOHbIX CaIorax N IOMecTe TKeHGY HcyTNHO BHyTpMbAHNKT. BaUN HOrN dOJIkHbI HAXOITbcB a CEHTpe KaKdO MaHKeTb, CTynHnDOJIkHb5BtIK KAK MOKIOB 6nIKke K HIXKHe NaCTh.
- 3acteHTHe moHHIO ha MaHKeTAX paCNOJokHTe Hx paBHOmePHo. IepKHTe HOrn pRmblIM, YTObI N36EkaTb nepeYIOB nCKlaoK MaHKeT. EcIn HyXHO - 6epHTecb TOnkO 3a Kpaar MaHKeT. He depKHTe yCTPOCTBO 3a NaHbIy npabEnHIN. IIOHCTbO 3acteHTHe moHHIO ha oboix canorax.
HeHCnONb3yIeYcTPOHCTBOCTOHNBOBpEMxXoDb6bl
- BkIOHHTyCTPOIcTBO. Yo6bI BKIOHHTyCTPOIcTBO, HAKMITE yIDepKINBAITe KHOKNY NITAHN Ha naHEIN ynpabNEHN. CBeToMDnHbE INHdkATOpbHa naHEIN ynpabNEHN 3aRopAOTc npBkIOHENyCTPOIcTBA.
- Bb6epTe HtEHCBHOBCTb DABeHnHa. HAKMITE KHOIky DABeHnA, YTObI IpeKIOHcyBc MEyDy TcBbPMBA BPaHATAM MHTHECHMBHOCTNa DABeHnA; 25 MM pTCT, 50 MM pTCT, 75 MM pTCT, 100 MM pTCT. HAcTPOKa DABeHnA, NcIbI3oBaBHnA BO BpEMI NocIeHne IpoUeDpybl, CTaHET HAcTPOKIO NO yMOUHmIO IyCNEuOe IpoUeDpybl. PIn KeJIahmN Bt TAKKe MOKeTe YEJIINHIVBAt NIMYMeHBsab DABeHnEBO Bv BPEMJeHnE.




Pcnyok3. PiojokHe Hor B canorax

- Bb6epTe dntenBhoctb npueDpybl. Haxmte Knohtky BpeMeHn, TcOsbpeKIOHTcM MeJy YteBpMa BapHaTAMn DnTentBHOCTM: 20 MHyT, 40 MHyT, 60 MHyT, Hn HnpepbBHO. BpeMa, NcIol3oBAHHO Bo BpeMa nocleHei npoeDpybl, ctaHet HAcTpoKoN ymoTHaHNO dra Cneyooue npoeDpybl.
H h 10000000000000000000000000000000000000000000000000000000000 - BkIIOHcHcyTPOTHBtHa HApKHTe HApEHPbAHTHe HA NaHEny TpABNHeHr, TOp6bI GoHnHbHTbYCTPOTHCTBO, YCTPOTHCTBO ABOTMAHHECKH BIKHOHTC Tcepe3 10 MHHyt 6e3eCTBIA, BIOHcHaeBepaehHHeHbnePiOeDpyh. IIOHbctoPbCteTHe KOMPeCCnOHHeHcAnf OHNrHa XpAEHeHs
- 5nKpOBKa yctpoCTBa. KOrla yctpoCTBO He MCnIb3eTcra Hr TpaHcnpTpyeTcra, 3aBnKpHyte erO, YTO6bI npodTbaTbE rO clyHae BkIOHeHE. BblIOHTeYCTPOCTBO, YTO6bI AKTBIOBAty FhNtIO bIOKpOBKn. BbCTPO HAKMITE KOHNY PNTAHN TPI pA3a, YTO6bI AKTNIOBAPbTb DNYHIO BLOKpOBKn. KaXDb CBTODNOH INHdKATOp 6yET NocIeDobateJIHO MIRATb, Yka3bBA HA To, YTO yCTPOCTBO 3aBnKoPbOBHa. BcCTPO HAKMITE KONIKY EEE TPI pA3a, YTO6bI pAsbNIOKpOBaTb YCTPOCTBO. KaXdb CBTQIOHIN HNDKATop 6yET MIRATb NocIeDobateJIHO (B I pOITIBNOJIOKHom HApBaNEHH), Yk3sBaBA Ha To, YTO yCTPOCTBO pAsbNIOKpOBaHO.

KHONKINHIDKAToPbI

KHONKa nHTaHn

KhONKa CTePT/Thay3a/CToN

Khonka BpeMeHH

KhoIIka daBneHna



HacTpoKn DaBHeHHa
Polyvinylpyrrolidinium dinitrobenzothio 12abbaa (25-100 MM pt. cr.l) C 12abm 25 MM pt.

Hactpoikn BpeMeHH
JetBoots Prime TruGrade Technology
TruGrade Tnece HcnpbEyet nepodbyy HnHOBaHIOHy ToxHOrHO NocIeBOBAtbHoro BbICOKOTOHOro DabEnHe, HauHHaOn CcoTTHN 3AkaHINBB6EpaMn. Kaedk H3eTpIXe B03duHbX KameK KompeccnHbX cantor JetBoots Prime HAcIbTaBaeTc HocoeHIn, NoToMdy ItoebeMe IIOaTeCBAeTBAH, 6es PNC4 BoHNHOBeHOTn oTneHbX OAnCTeY 63 daBHe 373333333333333333333333333333333333333333333333333333333333333
Kappa6oTaETnoCneIOBaTeIbHbIuNKl
BpEe 1000000000000000000000000000000000000000000000000000000000000000000000

Kamepa 4
KaMepa 3
KaMepa 2
Kamepa1
PcyHOK5.PocneoBAteBbHbIuK

6.yxod,nuctka n3apdka
06cnykBaHne yctpoiCTBa
BAnHO BnIOIaHTb CneIyoJIe HNCTpyKUIN IO TeXHHeckOMy yOxy, YTO6bI yCTpoIcTBO npdoJgApo pa6oTaB hAdJeXaUM m6p3aOM. HeBbIPOHNHeHne 3TNX INHCTpyKUIN MoXET pInBeCTN KIOIOMe KYCTPOCTBA.
OuNCTkaYcTpoIcTBA
MaHJTeBtI cHCTeMbI npeccOTepaNN JETBoots Prime h3rTObIeHbI H3 MATEpHaNA MeDINuMHcKOro KnLaccA, NOBepxHOCT bKOTOPORO NmHeH aNp, YTO nOMorAeT IpeodTbpATNb HAKONNEHb6 BakTePn I oBecneNbaet yctOyNUBOCTb K NsHocY. BlaROpadr CKbTyBm HAKNaIbIbAOUIMMCr Dpyr Ha dpyra KaMepaM NOBepxHOCT bNHeH aCkIaok B KOTOpbIX MORY T HAKANINBaTbCR 6akTePn I zanaxi.
- PONHOCTbIO pACCTeRHTe MOHINHO HA CANORAX I PA3NOKHTe IN HA POBHOI NOBEPXHOCTN.
- IcnoIb3yIte uCTaIe pactBop Ha OCHOBe 70% nOponpunIOBO cnpTa (IPA), YTO6bl AkkypaTHO OYNCHTb BHTpeHHIO NOBepxHOCb MaHKET. HMKoRda He IcnoIb3yIte UCTAIIme CpeCDTb Ha MaCLHaHO OCHOBe.
- PpOtpnTe yCTpoIcTBO CHApYKmMgKOuHCToTKaHbK, YTO6bIy6paTb 3aqrpa3HeHne NOBepxHOCTN.
4.Y6eNTecb,HTO NOcNE NOHNO OHCTKN YcTPONCTBA HE OCTaONCB OCTaKOB 3aqr3HeHH. - Nocne TOrO, KaK BbI NOHOCtBIO OCHNTe MaHKeTbI, BbITpTe IINHHIO Bnary I NOBecbTe Do NOLHO BcBixAHN, PpeXJe CmCNOJb30BaTb CHOBa.
PnmeH: 30T npOyKT HE ABnTcB ODOHnPOHmuaeMbIM. He norgpykaite yctpoiCTBO B BOY H Me MoTe ero noD npOTOH BOO. He donyckaiTe KOHTAKTa yCTPOICTBA C arpeccNBbIM pactBOPaM, KOTOpBE MOYr NOBPeNTB BHeUH N BIDN HfHKUHOAHbHOCTb YCTPOICTBa.
3apka yctpOCTBa
- UcpoIcTBo nOndepKINBaET 3apRy USB-C n noctabnAetc 3 aapdNbIM Ka6eIm USB-C dHIOH 2 MeTpa (6,5 yTOB). YcToPcTBo MoXeT aapRkAtbCBA BO BKIOueHHOM COCTOHN INN BO Bpemra60tbl.
- PdKIOHHTe OINH KOHeu zapHDOHO K zapHOMY NpO TY USB-C Ha yCTPOICTBE, OH pAcNtONoKEn c60ky OT NaHeHn UnpaBNeHnA. POnKIOHHTe dpyroH KOHeU K aDAnTepy NITAHm USB-C MoTHCtBO 20 BT (peKoMeHdyTeTcI dON TImMaJIbHO IN pOn3BDoHTeHocTn). McNoJIbOBAHne aAdaTpe PAITAHn MeHbSe MOnHoCTN MOKeT He NsOBoNTb IcNOJIbOBAbTa YcTPOIcBO BO BPEMzapdKn.
- CBeToIOIOHbIe HmIKaTOpBb6yT nKoA3bIBaTb CoTOrHHe 6aTapEn, KOrJa yCTpoiCtBO BkIoUcHcHnI IaJpaKaTeCBA. BkIoUcHTe yCTpoiCtBO, YTO6bI IpOBePntb CoTOrHHe aKKymIITopA.
1CBETONIOHn3Kn3apRb6aTapeu
2cBToOnoJa: cpeHn 3apJd 6aTapeN
3CBETOINoDA:Hn3Km3apJ6aTapeN
- Bpemr aBTOHMOHpoTb yCTpoNCTBa coCTaJIeT npIMepHO 180 MNHT.
**PpIMMeaHHe: PpI HcNpIb3OBAHm 3apJHOrO yCtPoIcTbA y6eIITceB, YTO OHO IpoIN3BeDEHo CepTnΦIuNPoBaHbIM IpoIN3BoIDITEeM He IMeET KaIX-Ni6oCTpyKTypbIX NOBpeXeHEnH.
XpaHeHne I TpaHCnOpTIpOBKa yCTpoIcTa
- POnoxte yctpoCTBO Ha poBHyIO NOBepxHocTb NaHeIbO ynpaBHeHn BBePcx I DaIeKOMnPecCCHOBHM CaNOrAM NOnHOCTbO CdyTbCra
2.CIOKHTe npabyu qacb MaHKeTbI NOBepx IeBoi, KAK ObUHc CKNaIbIaIOT 6pOKn. BbipOBHnTe BCE kpar MAHOKET, YTObI OHn POBHO Nekai n Dpyr Ha dpyre. - HaHHa co cTOn, cNoXnTe canOrn BBepx Ha 1/3 nInhbl CHN3y.
- Cnojnte epea TaK, TTo6bI H3r60ka3anCn Ha yOpBHe Bepx CanorOB. HE cKnAdbaiBaIte CanOr CNIIHKOM NIOTHO, TaK KaK 3TO MoKET NOBpeiTu MecTa IN NoKIOHcyHnA
KaeCTBe CymKn IypeHcOK.
Pepn xpaHHeM y6DnTeB, hOyCTPOCTBO HcyoE, IINXpAeHbN IN TPAHcONTPHOKbYCTPOCTBxApaHHTe ETO B TPAHcONTPHOBOHOM PHOKSAKE, OpbaaTaeB C HMM OCTOPOXHO. XpaHHTe YcPOCTBO N PcxNlHM CxmytceTmEPaHpyeT, -C0-70°C, OHOTCNIEHbNAkHOCTb: 10~93%, ATMOcΦepHOdANHEHcNE 1900-1060 rPa. He XpaHHTe YcPOCTBO TAM, rIe TEMNPaTpA MOKET PpeBbUaTb 70°C, HApNIpEM, nOJ PnPbIMM COJIHbHbIMN JIaAMM INB TPAHcONTPHOm CpeCTBSe.






PcCyHok 6. HnCtpyKmno cKnlaDbBaHHIO
YcTaPaHeHHeHEnoJaOK
https://www.therabody.com/us/en-us/support/support.html 66paeHHnB npoepkky.
OrpaHueHHa rapaHTn npn nokynke TOBko y ABTOp3ObaHHORo po3HUYHO rnpodabua
© 2024 Therabody, Inc. Bce npaba zauuueHbI.
INHOPMaun O nateHTax Ha caIeTwww.therabody.com/patents
PpOIN3BOIDCTBOI pacnpoocTpaHeHHe:
Therabody, Inc.
1640 S Sepulveda, Suite 300
Ioc-AHdkKeIec,KaIIΦopHn90025
UmmopTepoB:
Tepa6oINHTepH3uHnIIMNTeI
5-ntax, Mecnn Te3, 40
DO4C2N4Iy6nH,IpaH
Therabody UK Limited
NopTcOKeH Xayc 155-157
MeHbHnHCTBa,HOHdoH
Belenko6pTaHnE,EC3N1LJ
PNEUYPEXKDEHNAIPINUCIOJIb3OBAHNNYCTPOINCTBA
PPOHTHE BCE INCTPYKLIINI INPEIDYNTPEIDATEHBIEHADNHC BDAHHOM PYKOBIOCTBE,HA3APJHOM YCTPOINTBE HNA YCTPOINTBEPEPEDNCIOB3OBAHEM IIN3APJKOY YCTPOINTBTA Herabody JetBoots Prime n er okeccceypaob.
PnHcNoJIb3OBAHHYcTPOHCTBaBcERda CJeDyET PnIepeKINBaTbCra CJeDyHOxN XCHOBHbIX Mep npEOCTOPOKHOCTN:
- INCIOB3yIe TOIbKO B COOTBETCTBUN C INHCTPYKUIEH. IcnoJIb3yIte yctPoIcTB0 ToIbKO TaK, KAK OINcIaHO B PYKOBDCTBe TNOIbOBATeRThAbeoD. JetBoots Prime. YcTPOCTBO JetBoots Prime RABNTEC MednIHckmIM NpOaTe 6e3 peueTTa. YcTPOCTBo Tne pIepdAnHaHeNo DnIraHNOCTKnK, NeHEnHMI npOphiNAKTkn 3a2oBeNaHH. IcnoJIb3yIte ToIbKO peKOMEHDoBaHbte TheraBoab AKECCyaBPIn 3anaChbte NaCTs. 3To cneLAmH3npoBaHHbM EMAHHECKN HtCPmEtB C nEklTPrHcEKMM KOMPOnHeTAMc. EcnI yctPOCTBO nero AkCEccCyApBb HE NCIOBHTOCN HNI He oCBNYKBAOTCR DONJbHM OBaSM, CYUECTBY PNETK B03rOpAHy,
PNOJAEHIN 3NEKTHNNECKM TOKOM IINI TPAmbi, ECJIY NCTOPIEBO INI ERO YACHTI N AKECCyAPBHe HNCN0L3yETCn IINI HE OCBNYKBAeTc B COOTBcTcBN C PEKOMHEaJAMNI, HNIOXOEHBIMIN B DAHHOM PYKOBOCTBe, 3TO INPBEDED K aHNUYIPOBAAHO N PAOHTI N MOJEET PIPNEBCTK NOIBPEKDEHMOI PNOJCYTA INI FNNIECHKOIN TPAMe, UCTPOBJCTBO PEPNaHaHaHEn DO KOMMEpeckORIO N DOMAIIHero NCN0L3OBAHIA.
-
HE PEPDAH3HAHEHO DIIETEY, UCTPOICTBO N erO aecccayabI He npedHa3aHeHbI dN IcNOIb3oBAHINI MnAaIe 18 let. YcTPOCTBO H A KecccayAbE He pIepHa3aHeHbI dN IcNOIb3oBAHINI MnAaIe C OPAH4YHbIMI FAnHHckEMHM, CEHCOPBbIMI NIN NORVCEKMMN CnOC6HOCTMNI INC HEOCTaTOMbITaN 3AHHN, ECNI TOJIbKO MIIO, OTBeTCTBEHHOE 3X IN B6eONAcHObTb, He HAOHTTCIO NADHIOJDEHmN INI HE pONHCTPYKTKPTPOBAHO NO cNOIb3oBAHINIO UcTPOICTBA. YcTPOCTBO H E BIAETCR HpyuikO. IocOBetyteTeETM He nptBa c yctpoiCTBO M NIK AKECCyapAMN.
-
HE CTOTE BO BPEM NCIJIOJIb3OBAHINY ICTPOICTBA.He nIaTIEc bCTATb BO BPEM NCIJOIb3OBAHINY yCTPOICTBA.BCTaBaHIne BO BPEM NCIJOIb3OBAHINY yCTPOICTBA.MOKET pNIBECTK DNCOMFOPOT,PAHEHINO INOBPEKDEHNBOHYPTNEHHX KOMNIHOHTOB YCTPOICTBA.
4.3APJDKA VCTPOCTBA.EcnnBaue yctPOTBO HE BKNHOAECTCR INHINDMKATOP 6aIapen nokbaabaetnHNkyn ypoBEnb zapra, zaprAne Iero nepe npBbIMNIOIOBOAHmE.3aprAite akyyMHTOPC NOMOUBIO KABeJIY USB-C (BXoDITB KOMPINEKT).PpI INCOIbOBAHHN 3APdHORO kabeJIY (XoDITB KOMPINEKT)y6BeNTec,TO OER paoBEM USB-C NIOHOCTBO BCTABIEH B COBMCTMHbI aADTAPTEH NITAHN, PnEKe Yem NOKIIIOVAHbI aDAIETP K POSEK. Bt KaTke MOKeTe 3APRAtb AKKyMMHTOPC NOMOUBIO KABeJIY USB-C TOPOHNIH XPOB3BOHITEN, COOTBETCTBIOUIXx HOPMAm CPTAbI, a TAOKE MEKJHyAPoDBH IM PEPHOIAHBIM CTANDPAM 6eONACOHCTN. ADIATBPe bHEKOTOpbIX PNOIBMOITEJI MOrHT He COOTBETCTBOBAt BeJcDEBTIOUHM CTANDPAM 6eONACOHCTN, I 3APRAtb CnMOUIO ToXAPtAIDTAPOB 6PNETC K CMEPrTHI INPbAMAM. HbTaeTBC BCTABIV KABeJI B paaBem CUCVIMI. PNOHOCTBO ZAPRAKAE AKKyMUTTOP He Pekce OJHO rPA3B WIEcTB MeCEaE, YTObI PnpEOBATn TIOBPEKJHHe YcTOPTBA NkAkyMUTTOP. OTKIOUHATE aADTAPTE NITAHN I ZAPRbIb KABeJI, KOra OHNA EICNOBlyOTCR.
- HE PEPE3APJKAITE. He cTABJIeYteCTPOICTBO NOJIKIHOENHbIM K aadTIpepy NTANHIA 60NEe Yem Ha OINH TAC NocIE NOJIOH ZAPKnAKKyMylrTopa. AKKyMylrTop IMeET CHTcEMY 3aUITb, YTO6bl H6EkaB pNCKa pe3apZda. TEM He MeHee, C TeueHEm BpemeHn Pe3apZka MoKET COKpaNTb Ero cPOK cJyKb6b.
6.MECTAJ3AP4KIN.3ap4kJeTyeCTOBCnMOoJIHOaADIepeNITAHIN3ap4HOrO k6aeJI.HI3eGERaIKOTKAKTCKoKJI,KGDAUBC-NOKIOHKeH KTOCTOHNYIHTAAK,TAK KAK 8TO MOKETBbI3BaTbDcKOMFOpT INI TPA8U. He CnIte HAn He CaIDTeHBaP3EBM USCB-C.YcTPOICTBO cIeYET 3ap4KaTbBxOPOI NPOBETRBAeMOM,CXOMNOMEeEHIN.He 3ap4KaJIte YeCTPOICTBO HAOTKbIbTOMBQdXe,BBAHNKOKMHeTNEIBuPBDELnAX31,METPAOT BANHbIINII6accEHa. He NCnIOB3yTE YeCTPOCTBO HIN IIIap4HIOHe YcTPOCTBO HNA BLANKbIX NOBepxHOctAs,T Akoke He NpOBeBraTe YEAP4HIOHe YcTPOCTBO BO3dIECTBNO BnAfln,DOJHXNIINCHega. He IcNoB3yTe yCTPOICTBO HIN COMBEETMoe C HIM 3ap4HIOHe YcTPOCTBO B PNPCYTCTBMN B3bpblBOONaChbIX CpeI (rAooobpa3hie naPi,bNb,lBb,3e9O,MeTALNTNeHKe POpOKNiHIN NfETKOBOCnMaNEHBOUceM Mateo4HbI).MDOTBaNHOHHVbNcKbI,STO MOKETNIPBEETN KNOKApO
7.OTKJIIOUEHEME3AJPHNOFOYCTPOCTBAUYXOD.TAHNTe 3a BINky, a He 3a shHyp, yTO6bl CHH3MTb PNCKOBPEDEHNIAEKTPNHECKN KOHTAKTOB u Ka6Ena.HKmOg4HepeNEOHOCITE 3aepHOyUCPTOPBO3a UHP.DEPKeJTHUUPBdANOT UCHNYIKOB TENNA.MACNaI OcTbPsKPAeB. HPeactrVBAIEe KabeJb3apADHORo YcPTOCTBa n HpeMaEEuJIe C HATAKHEHM. HpeRnKAcaueTcck b KcTOPCTBc, 3APJHOyUCPTOCTBy INI PA3BeMAMMOKpbMMpyKaMM.BcerDAOTKIOHAteETOT pnp60bOT AEJKTPNEHCOY PO3eTKi CPa8y Nocne INCNoB3BOAHIN Npepd NCTOK. XpAHIeKABe9

Tak, Tc06bHa HeRo He HaCyIaIyI, He CnOtBkAicNcb H He NIOBBePraIINHbIM NOBpeKdEHNAM INI Haryp3kM, He IcNObl3yIteAanTeP nIaTHaIINuNn 3apAebHb, KOTOpBeNoYnHm pe3kn yUdp, yUnn, nepeexAnu kAm-Imo 6obpazOM nobepDInu. PnHnTeJIbHM xpaHeHn XpaHnTEyUCTOBCTBO C nONHOCTbO 3apKeHbIM AnKyMylrTOPM.
-
ICNIOJ3OBAHIE, INPKaPTIte IINIOJ3OBAHIE yUcTPOCTBA, eCN II YUBCTBYETe 60JI INI INCKOMoPT, ECN II B KAKOI-NIO MOMENT BO BPEM IPOUEyBbI bN OYCTBYETe 60JI INIKCKOMoPT, INPBEIIuAIOJIe OKIaEIAMeBTE OT KOMIPCECCNOHORO yTOpCTOBtI INI MACCKA, HMeDEIeNEHO INPKaPTIte NLeHHe I IN3BKeNTYE UCTPOCTBO, INPKaPTIte IINONLbOBAHIE yCTPOCTBA, ECIN OH NOpeRPeBaTeCIN CTAOHOBITc HekOMoPHTHO rOPAHIM. He pa3MeMaJIte KAKIn-1NIO pIeDmtBe MekdyKoKNe IMAHETOTyOCTOBtI.
-
HE ICSIOIb3YIe IOI OJ EIAJIOM IIOI OJYUKOJ. Ype3MEpHbI HArpeB MoKET PINBECTN K BO3rOHOPIPO, NOPAKHEINOJI AEKeTPIOeCKHM TOKOM IITIaPMAB.
10.YXQD N O6CTYKXBIAHVE YCTPOICTBA. 06paaHTBe c yctpoCTBOM OCTOPOHXo.
YcTPOCTBO COpeKJITALEKTPONHbIE KOMIOHNHeTI, KOtOpbe MOryT 6bITb NOpeKZbHbI pRn TApENHM,
OKOE, PpOKOE IINI PAzADLAMIBAHmE. He Pa36PAne YcTPOCTBO H n He NTaIeTc B pEMOHITPOBaTb
ero CamoTERTBHO. P3o6Pka YcTPOCTBA MOKeT PnBECTN K ero NOpeKDeHHIO TPAAMM.
ECIn YcTPOCTBO NOpeKZdEHO IINI BO3NHNI HEnCIPABHOCTN, OOBATTECB C bTyKbI NOpeKZk
KIIHETHOTherAbod. TaPAnTHIy 6bEt dAHyNJPOBAA B cTyAE N pAe36PObY cTOPTBc AkyKMUTOPa,
3aepHOrO YcTPOCTBA IINI UDAHNEH KAKIX-ILIO CHaeT. He MCNTLSyIte, ECIn OH NOBPeKZdEHI,
TAK KAI OH MOrY TnpBECTN K TpABMe. He NpokAlBaJIte H n He NpokEJaTe YcTPOCTBO.
IpOKoI YcTPOCTBA MOKeT PnBECTN K HEnPABUNHO B paOBe. YcTOPTBO HE ABIRETC BOHOHPOHUAeMbE.
YcTOPTBO HE NOIDEXHAT MAUINHO CTNPke. He Pa3MeuaJIte IHE XpAHITE
ycTOPTBO TAM, rDE OHO MOKeT NOACTB B BAHNl IINI PAKOBHJe. He NorpyKaIte I E POHJIe
ycTOPTBO B BOY IINI DpyHQY JIKXIOKT. He DpTaNABIECe DO YCTOPTBA, KOtOpoe YnANO B VOy
III INCIPOKCHNYOCB C BOiD. HemdeHIO HO OTKIOHCTY OCTBO CTcN. OHCTYTE CYPOTBC B COOTBTCTBN C IHCTPKUIMMI, PNBeDEHIO IMB PAnDENe I aPaDENe (OChTKA) B pyKOBODCTBE NOnbSOBaTeJIa.
13.XHMCKEHCNE BELEECTBA AKKYMVLYTROPA Bl3b1BAOT CEPBe3HbIE OXOGTI. HNKOTdA
He DoNyckaiye KOHTAKTA BCTPOEHNO AKKYMUYLTAPOA C KOKEJ, Ia3AAAIMI INI PTOM. B cTyALYE TyeKu
XMHNKATOB INI bPOPEKJHDO AKKYMUYLTAPOA IcNbLYte Pe3HIOHObI IIe HEnOPIeHOBe IepATKn
IaER oE TYNN3AUO. ECIn KOKA NpOBePTNaBc BOSdEETNo XIKDCKTEIH N AKKYMUTOPA, POMeHTe I EAOBc I cMbLOJ, A aTEM YKCYSCOM. ECIn Ira3a IIOpeBPrrNCs BOSdEETNIO XMMYNECKN BEECTB
IaKKYMUTOPA, HemdeJIeHNHG IOPMOTBE IX BOY B TEHeHME 20 MHyT IN o6bATNTeB K BPAY.
CHIMMTe NytININ4XyPIYe ZARAPREHHNO ODeKdy.
- HE DONTVCAJNTE KOPOTKTOO 3AmbkiHAYI, ECNI MetaJIINHECKN pRpDMET COeIMHIT PONOKHTBnH N OTRPuaTbHb KOHTA H AKKyMmYrTOpe IIN PAa3peME 16B, B AKKyMmYrTOpe PONIO3oET KOPOTKoe 3AmbikAHAE. He KLIaTIte AKkYMmYrTOp PAaOM C pRpEMTeAMM, KOtOpSE MoryT BIIBaBHb KOPOTKoe 3AmbikAHAE, TAKMIAN KAK MOHETb, KIKOYI INI TbOzdi N B ABAEM KAPMaHe. KOPOTKoe 3AmbikAHBe AKkYMmYrTOpe MOKET PnPIBECTN K N KApoxy T aPbMaAM.
13.XPAHEHIEYCTPOXICTBAVIKACECCYAPOB. XpAHnTB BYCXMPOXIAHDM Mecte. 3apaKaHTe YcTPOCTBO TOLIbKn PnI TemnePATyOpe KcykAIOeI cpeDnOT-4do 158^ (-20^) -70°C.He XpAHnTHeYcTPOCTBO HIN ZAPdHBN KaBeB TaM, Tqe TEmPEATyMOxET pReBbIaTb 158° 70^) , HAnPIMep, NOD PnAIMM COHTHeHBMIn LyAmi, B TPAHCNOPTHom CpeCTBE HIN B METALINIueCKOM 3aHNlNETo. He CKpyuBaIBMaHKeBTb YcTPOCTBA DnI xpaHeHIN, TAK KAK ETo MOket NOBPeNTB BYHTpeHHNE KOmHHeNTB yCTPOCTBA
- YTNIM3ALIMIYCTPOINCTBA. 3Oy cTPOINCTBO codePKMIT IYNTHNHOHBNAKKMYTARTOP, no3TOMY ero b6xodmIO npabnBHO yTNIM3POBATb. Pered yTNIM3AUcM 3TOrOYcTPOINCTBA 3UHTyHe MeOCHTE B3AOKHbI N TEOBAPBNK, kacaOLuCEeY aTNIM3AUcNl YTNIM3HOHb6aTaeR.
IpeIOHTeIbHbIM MeTODOM yTUN3aUIN ABnEeTcN nepepa60Ka Bcero ycTpoiCTBa.
YXOD 3A MAHJETAMN.
3HaueHnCIMBOJIOB
0603Haehn
CNMBONBI ONICAHNE PACNOJOXEHNIE
IP 22 Ctenenb 3aunTb ot nonanadnB oDbl Ha nacnopTho Ta6nueke

IpoHTTE INCTpyKUIO nepeiNcnoIbSObaHMe Ha nacnopTHe Ta6nHke

CTeeneHb3aunltBtTNBAFnpmehreMoHactnHaNacnOpTHoTaNIMHke

Therabody, Inc. 1640 S. Sepulveda Blvd.
JIoc300
Loo-AnJKeJeNec,KaIIINOpHnIg 90025
Ha nacnpoTHOa7nHKe

YHnKaIbHaN IeHTnФkaIaIyCtpoIcTBA (UDI)
Ha nactopnTHoT SaBnHke
ynakokake

PaanbHbIc b60cTOnxOobnAnEeKtpnHexeRcO n AnEeKtpnHOHO6OpDvBOAHa.
PIMMHeHMe: DnI NOnyChENr DOonNHTeBHOI HNOPMAUINo 8YTNIN3AUNo 06OpDIOBAHnE, ETO YACTeN INPIHADNEXHOCTe, NOKAnyTcA, CBXKNITecb C MeCTbHBM NCTPrN6bhotOpOM.
Ha nacnpthon Ta6nueke
3TOyctpOCTo 0THc0rTcKOB600B8HHKHeeCCsIII.THnBF.CoTeRtCByET CTaHpaTmMgHMHKOJI 9AeKcPb600SnHcOHTI [EC6061-]
yepaeetb taikc eoeefeteyfet aee
WTOHOTDIOPOPTIOPOHSIOI HIOHOIOOHTIOOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOIOOO
1 工 ^ 工 PEPYTEPEK/DEHNE: 1000000000000000000000000000000000000000000000000000000000000000
21PDEKHEHNEA
31TIPVITPVEHHPFIOEPOOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPOPO
PYKOBODCTBO IN DEKLAPAUINPON3BOIDTEI-3NEKTPOMARHHTHOE N3JYUEHNE
PYKOBODCTBO IN DEKJIAPAUNI PON3BODNTEN - 3NEKTPOMAHHTHAR YCTOHNBOCTb
| YctroiCTBO JeiBoots Prime IpeaHa3aHcHIOI IaI HcTIOB3OBAHnB B 3JIeKTPOMaHTHIO CpeDe, Yka3aHHIOI NmIK. 3aKa3aHcK IIaII PONbOAtelBJ JeiBoots Prime DoTJKNH y6eJHtCB T TOM, cTo YcTroiCTBO HcTIOB3OByET OIMHHO B TAKHX yCLOBIAx. | ||
| TECT COOTBTCTBNE 3JIeKTPOMAHTHTHAR CPEDA - YK3AHIN | ||
| PADIQUACTOTHBE H3JIyUHENI CISPR 11 | Гулла 1 | JetBoots Prime ИспочAstET радiochactOTHy IO HeDriTO ToIbI OIry AHyTPreHrE OунКИОHAlA. ПOSTOMY eTO pARDIOCHACTOTHBe H3JIyUHENI H3JIaHHTeIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbIbI b |
| PADIQUACTOTHBE H3JIyUHENI CISPR 11 | Клacc B | JetBoots Prime NOdIXoHIT DARE HcTIOB3OBAHnB BO BcCEX UyPeKHeHIXK, AkyIcHAr JIckIbe NOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe HOMeIe |
| FapMOHInueckme H3JIyUHENI IEC 61000-3-2 | HeprimMeHIMO | |
| Мерцalouche H3JIyUHENI IprI KcJIbB6aHIX HApTlKBeHnI RIE 61000-3-3 | HeprimMeHIMO | |
| YctpoctrBo JettBoots Prime npedhahnne Hdo nncnonbaaBnA B anekpmaHnHnO Cpe, ykaaahnHn Hnke 3akaaanik nI nonoosatoBte JetBoots Prime dpomkeny bcejntbC B ToTc yctpoctrBo nncnonbaeYcmeHHN B TAKINX YCNOH. | |||
| TECT HA UYCTOHUBOCTb TECTOBb YPOBEHB IEC 60601 YPOBEHB COOTBETCTBN 3NEKTPOMATHMHTHAR A CPEA - PEKOMEHDAUIM | |||
| 3noKtpoctatmuaocckn paarpa (ESD)IEC 61000-4-2 | +8 KB KOANTAK±2 KB, ±4 KB, ±8 KB,±15 KB NaipkAnBnB B BoaDxue | +8 KB KONTAKT±2 KB, ±4 KB, ±8 KB,±15 KB NaipkAnBnB B BoaDxue | Ponim doknkih bdyrepeabnnnhi, beotnni hmi nI xepamuecknmiI pIiAAM, Eclai noBnI pOktby CInITetHecknM MATEPAKOM,OTHOCHENHBA BlaXIOCTB BocDaXu doXkHa bOeHae 30% |
| 3noKtpoctatmuaoccknCKAHOKIEC 61000-4-4 | ±2 KB DAI MAIOHOCSTNIHN CHaSbKnHnI ±1 KB DAI PAHNI BxoJa/BaoXoa | H/II KaueCTBO anektpoHNTAHNA DoJIKNHO CoOCTBET | BoBATy KaueCTay TMnIHNOHOMKOMEPHECKON IIN DOMAUHKe CpeDb. |
| BbipocIEC 61000-4-5 | ±1 KB dVdFepenHnHnBn PeknHMM ±2 KB Ossui NpEKM | H/II KaueCTBO anektpoHNTAHNA DoJIKNHO CoOCTBET | BoBATy KaueCTay TMnIHNOHOMKOMEPHECKON IIN DOMAUHKe CpeDb. |
| CcknHn HaprakBeHn, koprtnkHe ppeblyaHn HkoIeBaHnHaprakBeHn HnA BxoDbXn PliHnXn HnKtPOnHITAHNIEIEC 61000-4-11 | 0% UT; 0.5 uKn gl pRn 0°, 45°, 90°, 135°, 180°, 225°, 270° u35°0% UT; 1 uKn Hn 70% UT; 25/30 uKn Hn OdoXoHaHoo: pRn 0°0% UT; 250/300 uKn | H/II KaueCTBO anektpoHNTAHNA DoJIKNHO CoOCTBET | BoBATy KaueCTay TMnIHNOHOMKOMEPHECKON IIN DOMAUHKe CpeDb. |
| NaCTOts cTeH (50/60 Ft) magnitnoe noileIEC 61000-4-8 | 30 A/m 30 A/m MaHnTHMe NoI pRmBnHnHnO TaCToTb DoJIKn HsCoTBeTCTBOBAtb | ypobHM, XapakertbnHm dm oBbHnOH Kompehckon IIN DOMAUHEn 06ctbHOBKn. | |
YcTPOCTBO JelBoo Prine npedhazahane no nncnlo30aHnB 6eKtpoMafHTHO Cpege, kaaahnn Hnke. 3a2aekn Knno hno bAton. JelBoo Prine daxhen ybogatcb a TOM, TTO yctpoDCTBO mno hno lybtaer c MmHNO B TAKHX YcNooHx.
| PRUEBA DE IMMUNIDAD | NIVEL DE PRUEBA IEC 60601 | NIVEL DE CONFORMIDAD ENTORNO ELECTROMAGNÉTICO - DIRECTRICES | |
| Повodимь RF IEC 61000-4-6 | 3 (срденихадраитунеско нарpondжени) | H/П | Передочью Мобличесо оборудане RF сбетаclлсдует саллуотерллостыг He bшиke liobшу чагу JelBoots Prime (ылочья кабел),ЧМ реком endуSpecупснше расстогене ydakleney, paskutanahno no уразиннoly, почilemeniMcy K chactotre nepedatrukhna. Реком endуSpecоп ор: PakomendуSpecоп ор: d=[3.5/V1]√P d=[12/V2]√P d=[3.5/E1]√P 80 MHz to 800 MHz d=[7/E1]√P 800 MHz to 2.7 GHz |
| Интуемь RF IEC 61000-4-3 | 10 B/M Ст 80 MЛ до 2,7 TПU 385MЛ-6785MЛ Норатиньш eнл'tаньш и STAIIHUIEHHOCTb ПОТА可КОТYCA OT odniuchstotTOHOr obspudobovnaya 6eopodobnoi CBsai (CM. tabluy 9 stan-daprta IEC 60601-1-2:2014) | 10 B/M Ст 80 MЛ до 2,7 TПU 385MЛ-6785MЛ Норатиньш eнл'tаньш и STAIIHUIEHHOCTb ПОТА可KOPOTYCA OT odniuchstotTOHOr obspudobovnaya 6eopodobnoi CBsai (CM. tabluy 9 stan-daprta IEC 60601-1-2:2014) | rde p - Маршимьнь having baxoDhar MOUCHSTbpeperaduTHa BA btaTt (Bt) no Дамнй Млрпсьнвдentes nepedatrukhna, a d - pekOMenduySpecор расстогене мexуду порадуSpecоп ор: PakomendуSpecоп ор: d = 3.5/E1√P 80 MHz to 800 MHz d=[7/E1]√P 800 MHz to 2.7 GHz |
IPYIMEAHHIEI.PB80MFGU800MFGPIMHEHETCA6oeeBbckmHdnaaohHactot.
PIMMEAHHE 2. DAnHbpeKoMeHdauu npHMHeMbI He Bo BcEs CmTuayraux. Ha 3MB bBnAorntnloaehne nOtpaxKeHne O T KOHCTpykun, npdeMToB HIOeJ.
150 Ml 40.7Ml-20.5Ml-40.5Ml-7.5Ml-10.5Ml
Hnepnne Hnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nnnnne nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee nannnee
PekomBeyembe pBCCTOAMM MEKdypeHNOCHbIMN M0bMbHM PdnoHOACTOBHMcpeCDTAMnCBaN JIeBoots Prime
JebBoots Prirte PnepaHbHaeH HIN NIOHOBBOHBA B 3NHTKPOPAMTHHTH CpeB, B KOPTOP 3NHAHTBHMR RF 80AMLHBNH HAKOHNCTIO KNOXHTOPKEM, 3AKHANH MNNONATOBH JEB Boots JMOKT CIOCTCBOBTBAD PNDPOTRAPBQeHHo HNOK TPOKPMATOHNH NOMX, NOEPAKDBMHN HMMHMAH NAOCTPCBHRMEKNOJQPATAHBM HMOBIMHMOOCBPOBDABHMRFCBAH (PepeDTAHKAM)JebBoots Prime, KAKPEKOMEHDOHOHNG, aCOATRTCTBVNCMAKCMHAFMBOHXDHO MOHCHTOHO CMOKHYIKHKUHCHOHO OPOCDyBDHBA.
| HOMA1HbtaHb MAX1HbtaHb Mouchouctb nepeatnka Bt | Раостockиke B завоньсстотучесы M | ||
| ot 150 кг,do 80 MФι Здрахедам ISM и постельскп разндалаллоанов | ot 150 кг,do 80 MФι ьдрахедам ISM и постельскп разндалаллоанов | ot 80 Мfι,do 800 Мfιо d=[3.5/V2]√P d=[3.5/E1]√P | |
| 0,01 0,12 0,20 0,035 0,07 | |||
| 0,1 0,38 0,63 0,11 0,22 | |||
| 11.2 | 2,00 0,35 | 0,70 | |
| 10 | 3,8 | 6,32 | 1,10 |
| 100 | 12 | 20,00 | 35 |
Длгпоредатногсь МБКСМВЛТНОВ BIXOДHON MOUHOCSTH,H9YK38HHOHNBILUE. pOKOMENJIWOB PACCSTOTAHN MOKUYI ПОЕРДATNKHAM K B MEXP(A)MOKERTB pACCHITAHNO C NOMOULYO YBPSHENHRI,ПIMMANMIMO K A SCOTATE ПОЕРDAHNTKA,TPE-P -MAKCHINHJIWA BIXOДHON MOUHOCCTb ПОЕРDAHNTKH B BBTAX (B) NO DSHNBM NPOINSOBDTA MEIPOEDATNIGH.
PIMIMEA/HVIE,INIPOCTAOXBOMTU80MTOPLNIMHRETOPAADNIOOCEEPOCTOBHEHDBIO6BCCOK24TOTHO10A0D33303
PIMMNEE2.DaHHepeKOMeHdauu npmmeHHMa He 80bcxCHTuHix. Ha pacnpcptpAHHe 3MB BAAIROT nortouae Hn O TpaekHe 0r KcHcTpykun, npedMetOB nnoo
JetBoots Prime
TruGrade Technology de JetBoots Prime
Therabody International Limited
5th Floor, 40 Mespil Road
DO4C2N4 Dublin, Ireland
Therabody UK Limited
Portsoken House 155-157
The Minorities. London
United Kingdom, EC3NILJ
MISES EN GARDE DE L'UNITÉ
LRE TOUTES LES INSTRUCTIONS ET LES MISES EN GARDE DANS CE MANUEL, SUR LE CHARGEUR ET SUR L'APPEARIL AVANT D'UTILISER OU DE CHARGER L'APPEARIL Therabody JetBoots Prime ET SES ACCESSOIRES.
www.fda.gov/downloads/aboutFDA/reportmanualsforms/forms/ucm163919.pdf