BENNING CM E1 - Multimeter

CM E1 - Multimeter BENNING - Free user manual and instructions

Find the device manual for free CM E1 BENNING in PDF.

📄 38 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice BENNING CM E1 - page 14
View the manual : Français FR Deutsch DE English EN Nederlands NL
Pick your language and provide your email: we'll send you a specifically translated version.
Product typeEarth ground clamp meter
BrandBENNING
ModelCM E1
CategoryMultimeter
Dimensions (L x W x H)276 x 100 x 47 mm
WeightApprox. 750 g (with batteries)
Power supply9 V battery (IEC 6 LR 61)
Jaw opening38 mm
DisplayDigital LCD, 4 digits (9999 counts)
Measurement functionsEarth loop resistance (0.025 Ω - 1500 Ω), AC current / leakage current (0.300 mA - 35.00 A)
Measurement accuracy (resistance)± (1.5% to 20%) depending on range
Measurement accuracy (current)± (2.0% to 3.0%) depending on range
Special functionsData logging (116 values), HI/LO alarm thresholds, HOLD function, auto power off (APO), self-calibration
Overvoltage categoryCAT III 300 V
Pollution degree2
Protection ratingIP30
Operating temperature0 °C to 50 °C (humidity < 85%)
Storage temperature-20 °C to +60 °C (humidity < 75% without batteries)
StandardsDIN VDE 0411 Part 1/EN 61010-1, DIN VDE 0413 Part 5/EN 61557-5
MaintenanceCleaning with dry cloth and mild detergent; avoid solvents
RepairabilityRepair by qualified personnel only; calibration recommended annually
Package contentsDevice, carrying case, reference resistive loop, 9 V battery, instruction manual

Frequently Asked Questions - CM E1 BENNING

How to measure earth loop resistance with the CM E1?
Select the Ω or Ω+ function using the rotary switch. Wait for the self-calibration to complete (CAL 7...CAL 1). Open the clamp and clamp around the earth conductor. Close the clamp and read the value on the display.
What to do if the display shows 'OL'?
'OL' indicates over range. Ensure the measurement is within the specified range and the clamp is properly closed. If the 'NOISE' symbol appears, interference may be affecting the measurement.
How to replace the battery?
Turn off the device. Place it face down, unscrew the screws on the bottom case. Remove the old battery, disconnect the leads, then connect a new 9 V battery (IEC 6 LR 61). Replace the case and screw it back.
How to set the alarm thresholds?
In Ω function, press 'FUNC' repeatedly until 'HI' or 'LO' appears. Use the ▲ and ▼ buttons to adjust the value. Press 'FUNC' to confirm. Thresholds can be set from 0 to 1510 Ω.
How to use the data logging function?
Set the sampling rate (1 to 255 s) via 'FUNC' and the ▲/▼ buttons. Press 'REC' to start logging. Data is stored (max 116). To read, press 'FUNC' until 'NO.' appears and select the location with ▲/▼.
What does the 'NOISE' symbol mean?
The 'NOISE' symbol indicates the presence of interference signals (current >3 A or voltage >30 V) that can disturb the measurement. Move the device away from interference sources if possible.
What is the maximum allowed voltage?
The CM E1 is designed for CAT III 300 V circuits relative to earth. Do not exceed this voltage to avoid risk of electric shock.
How to disable auto power off?
Hold down the 'FUNC' button while turning the rotary switch from 'OFF' to a measurement position. The 'AP' symbol disappears from the display, indicating that APO is disabled.
How to clean the device?
Clean the case with a dry cloth and mild detergent. Never use solvents or abrasives. Ensure the battery compartment is clean and free of electrolyte.
Where to get the device calibrated?
To maintain accuracy, annual calibration is recommended. Contact Benning customer service at the address indicated in the manual to send the device.

User questions about CM E1 BENNING

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Multimeter in PDF format for free! Find your manual CM E1 - BENNING and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. CM E1 by BENNING.

USER MANUAL CM E1 BENNING

text_image J I G F E REC NO. NOISE H 8.8.8.8 Ω C A B Ω m VA AP K

BENNING CM E1 - 1

text_image Rx R1 R2 R3 ... RN

BENNING CM E1 - 2

text_image RI R₁ R₂ ... Rₙ Rₓ

BENNING CM E1 - 3

text_image Rx

Fig. 3: Earth loop resistance measurement

text_image Technical diagram showing a measurement device connected to an electric motor, with digital display and control buttons visible.

Bild 4: Wechselstrom-/ Ableitstrommessung
Fig. 4: AC current/leakage current measurement
Fig. 4: Mesure du courant alternatif / du courant de fuite
Fig. 4: Wisselstroom-/ lekstroommeting

BENNING CM E1 - 4

natural_image Technical line drawing of a handheld device with internal components and external casing (no text or symbols)

Bild 5: Batteriewechsel
Fig. 5: Battery replacement
Fig. 5: Remplacement de la pile
Fig. 5: Vervanging van de batteri

$$ R s = E / I \quad w o b e i g i l t: \quad R s = R x + (R 1 I I R 2 \dots I I R N) + R b + R I $$

Earth measuring clamp for

  • earth loop resistance measurement
  • AC current/ leakage current measurement

Table of contents

  1. User instructions
  2. Safety instructions
  3. Scope of delivery
  4. Device description
  5. General information
  6. Ambient conditions
  7. Electrical specifications
  8. Measuring with the BENNING CM E1
  9. Maintenance
  10. Environmental note

1. User instructions

This operating manual is intended for

  • skilled electricians and
  • electrotechnically trained personnel.

The BENNING CM E1 is intended for measurements under dry ambient conditions. It must not be used in electrical circuits with a nominal voltage higher than CAT III 300 V (see section 6 „Ambient conditions“ for details).

The following symbols are used in this operating manual and on the BENNING CM E1:

BENNING CM E1 - User instructions - 1

Application around and removal from HAZARDOUS LIVE conductors is permitted. Warning of electrical danger!

BENNING CM E1 - User instructions - 2

Indicates instructions which must be followed to avoid danger to persons. Attention! Must comply with documentation!

BENNING CM E1 - User instructions - 3

This symbol indicates that the information provided in the operating manual must be complied with in order to avoid risks.

BENNING CM E1 - User instructions - 4

CAT III Measuring category III is applicable to testing and measuring circuits connected to the distribution circuit of the low-voltage mains installation of a building.

BENNING CM E1 - User instructions - 5

This symbol on the BENNING CM E1 indicates that the BENNING CM E1 is equipped with protective insulation (protection class II).

BENNING CM E1 - User instructions - 6

Please observe the operating manual!

BENNING CM E1 - User instructions - 7

This symbol appears on the display to indicate a discharged battery.

BENNING CM E1 - User instructions - 8

(AC) Alternating voltage or current

BENNING CM E1 - User instructions - 9

Ground (voltage against ground)

2. Safety instructions

The instrument is built and tested in accordance with

DIN VDE 0411 Part 1/ EN 61010-1

DIN VDE 0411 Part 2-032/EN 61010-2-032

and has left the factory in perfectly safe technical condition.

To preserve this condition and to ensure safe operation of the device, the user must observe the notes and warnings given in these instructions at all times. Improper handling and non-observance of the warnings might involve severe injuries or danger to life.

BENNING CM E1 - Safety instructions - 1

WARNING! Be extremely careful when working with bare conductors or main line carrier! Contact with live conductors will cause an electric shock!

BENNING CM E1 - Safety instructions - 2

Before starting the current clamp multimeter, always check the device as well as all measuring leads for damages.

If it can be assumed that safe operation is no longer possible, switch the device off immediately and secure it against unintended operation.

Safe operation can be assumed to be no longer possible, if

  • the device exhibit visible damages,
  • the device no longer works,
  • the device has been stored under unfavourable conditions for a longer period of time,
  • the device was exposed to extraordinary stress during transport, or
  • if the device is are exposed to moisture.

BENNING CM E1 - Safety instructions - 3

Maintenance:

Do not open the multimeter, because it contains no components which can be repaired by the user. Repair and service must be carried out by qualified personnel only!

BENNING CM E1 - Maintenance: - 1

Cleaning:

Regularly wipe the housing by means of a dry cloth and cleaning agent. Do not use any polishing agents or solvents!

3. Scope of delivery

The scope of delivery of the BENNING CM E1 comprises:

3.1 One BENNING CM E1
3.2 One transport case with shoulder strap (10217859)
3.3 One reference earth resistance loop (10217860)
3.4 One 9 V block battery (IEC 6 LR 61)
3.5 One operating manual

Parts subject to wear:

- The BENNING CM E1 is supplied by means of one 9 V block battery (IEC 6 LR61).

4. Device description

see figure 1: Appliance front face

The display and operating elements shown in figure 1 are designated as follows:

① Measuring clamp, for clamping the earthing conductor/earth connection
② HOLD key, storage of the indicated measured value
3 Rotary switch, for selecting the measurement function
4 Digital display
⑤ REC key, for starting the data logger/storage
6 ▼ key, decreases the setting value
⑦ Opening lever, for opening and closing the current prongs
8 ▲ key, increases the setting value
9 FUNC key, function key for selecting alarm thresholds, sampling rate and storage location number

see figure 2: Digital display

The symbols shown in fig. 2 are designated as follows:

A Secondary display, for function selection and storage location number
B Digital display, for measured value, alarm threshold, sampling rate
© Ω Ohm, unit of loop resistance measurement
D mA, unit of current/ leakage current measurement
E ,hop resistance measurement with acoustic alarm function
F NOISE, detection of interfering signals, measurement might be affected
G, measuring clamp is not closed correctly
H battery status indication
① NO., storage location number
J REC, data logger is enabled
K AP, automatic switch-off is enabled (APO active)

5. Functions of the earth measuring clamp

5.1 General information

5.1.1 The digital display B is a 4-digit LC display with a font size of 11 mm and a decimal point. The highest numerical value to be displayed is 9999.

5.1.2 The range exceedance is indicated by „OL“. Attention, no other indication and warning in case of overload!

5.1.3 The rotary switch ③ is intended for selecting the measuring function. The selection of the measuring range is automatic.

5.1.4 HOLD key function: The measuring result can be stored by pressing the HOLD key ②. The "H" symbol simultaneously appears on the display ④.

5.1.5 REC key function: for starting the data logger or storing a measured value in the internal memory

5.1.6 FUNC function key: function key for selecting the alarm thresholds "HI" (high), "LO" (low), the sampling rate "SEC" (seconds) and storage location number "NO." (1-116)

5.1.7 The nominal measuring rate of the BENNING CM E1 is 2 measurements per second.

5.1.8 The BENNING CM E1 is switched on and off with the rotary switch ③. Shutdown position "OFF".

5.1.9 The BENNING CM E1 switches off automatically after approx. 4 min. to 6 min. (APO, Auto-Power-Off is enabled if the "AP" symbol K is shown on the display ④). It switches on again as soon as the rotary switch ③ is switched on again from switch position "OFF". Automatic switch-off can be deactivated by pressing the FUNC key and by simultaneously switching on the BENNING CM E1 from the switching position "OFF". The symbol AP K disappears from the display ④.

5.1.10 The BENNING CM E1 is supplied by a fitted 9 V block battery (IEC 6 LR 61).

5.1.11 If the battery voltage drops below the specified operating voltage of the BENNING CM E1, then a battery symbol Ⓗ appears in the display 4.

5.1.12 The battery life depends on the measuring function used and is intended for approx. 3000 measurements.

5.1.13 Temperature coefficient of the measured value: 0.1 x (stated measuring accuracy)/ °C < 18 °C or > 28 °C, related to the value for the reference temperature of 23 °C.

5.1.14 Dimensions of unit (length x width x height) = 276 x 100 x 47 mm Weight of unit: approx. 750 g (incl. battery)

5.1.15 Widest prong opening: 38 mm

5.2 Setting the alarm thresholds of the earth loop resistance

To measure the earth loop resistance, an upper (HI) and lower (LO) alarm threshold can be set. Use the rotary switch ③ to select the function Ω and press the FUNC key ⑨ until the "HI" symbol or "LO" symbol appears on the secondary display A. By pressing the ▼ key ⑥ and the ▲ key ⑧, the alarm threshold can be set from 0 ohms to 1510 ohms or "OL" respectively. After having set one alarm threshold or both, press the FUNC key ⑨ until the symbols on the secondary display A disappear.

As soon as the rotary switch position +· is selected, the earth measuring clamp compares the displayed value with the upper and lower alarm thresholds.

If the displayed value is higher than the upper alarm threshold, a pulsating acoustic signal is emitted and the "HI--" symbol is displayed.

If the displayed value is lower than the lower alarm threshold, a pulsating acoustic signal is emitted and

the "LO--" symbol is displayed.

BENNING CM E1 - Setting the alarm thresholds of the earth loop resistance - 1

Note:

  • To disable the alarm thresholds, set the upper alarm (HI) to "OL" and the lower alarm (LO) to "0".
  • The upper alarm (HI) cannot be lower than the lower alarm (LO), and the lower alarm (LO) cannot be higher than the upper alarm (HI).
  • When the data logger is enabled, the acoustic alarm function is disabled.
  • The alarm thresholds set remain stored until they are changed the next time.

5.3 Data logger function

The data logger function allows the automatic and manual storage of series of measurements (Funktion / ), mA/ A) with a predefined measuring interval (sampling rate) and up to 116 measured values. The measuring interval can be set from 1 s to 255 s. The measured values can be read out later by means of the display 4.

5.3.1 Setting the sampling rate

Press the FUNC key ⑨ several times until the “SEC” symbol appears on the secondary display A. The digital display B shows the sampling rate in seconds. Use the ▼ key ⑥ und the ▲ key ⑧ to set the sampling rate from 1 s to 255 s. Press and hold the keys to accelerate the setting process. Then, press the FUNC key ⑨ several times until the symbols on the secondary display A disappear.

Note:

To store only a single measured value, select a sampling rate of 0 s. Each time the REC key ⑤ is pressed, another measured value is stored in the internal memory and the storage location number is briefly shown on the secondary display ⚠.

BENNING CM E1 - Note: - 1

text_image SET 0001

5.3.2 Starting and stopping the data logger

Press the REC key ⑤ to start the data logger. At the same time, the REC symbol ① appears on the digital display ④. The measured values are stored in the internal memory taking into account the sampling rate set. The data logger can be stopped by pressing the REC key ⑤ and stops automatically as soon as the measured value memory is full. The REC icon ① on the digital display ④ disappears.

5.3.3 Calling measured values

Press the FUNC key ⑨ several times until the "NO." symbol ① appears on the digital display ④. By pressing the ▼ key ⑥ and the ▲ key ⑧, the storage location number is selected via the secondary display ④ and the corresponding measured value is shown on the digital display ⑧.

BENNING CM E1 - Calling measured values - 1

text_image NO. 0.000 NO. 10.00 Ω

5.3.4 Deleting the measured value memory

Delete the complete measured value memory by pressing the REC key ⑤ and by simultaneously switching on the BENNING CM E1 from the switching position "OFF". The "CL" symbol being shown on the display ④ means that the measured value memory has been deleted.

BENNING CM E1 - Deleting the measured value memory - 1

text_image 11

5.4 Functional principle of earth loop resistance measurement

Earth loop resistance measurement is one of the methods of measuring the earth resistance without using rods. It offers the advantage that no additional probes/auxiliary earth electrodes have to be installed and the earth connection itself does not have to be disconnected. The method is used, among other things, in electrical installations with several parallel earthing connections, such as overhead line networks with pole-type earth electrodes, lightning protection and street lighting systems.

The BENNING CM E1 has a specially shielded measuring clamp ① with integrated excitation winding and sensor winding. The excitation winding induces a current flow in the clamped earthing conductor/earth connection via a defined AC voltage E. The current flow I is measured via the sensor winding and the resistance of the entire earth loop Rs is calculated and displayed via the BENNING CM E1.

$$ R s = E / I \quad \text { provided that: } \quad R s = R x + (R 1 \text { II } R 2 \dots \text { II } R N) + R b + R l $$

$$ \text { with } (R 1 \text { II } R 2 \dots \text { II } R n) < < R x \text { and } (R b + R I) < R x \quad \text { results in } R s \approx R x $$

This means that the earth loop resistance Rs measured by the BENNING CM E1 is always higher than the searched earth loop resistance Rx. The higher the number of parallel earth connections, the more accurate the displayed value Rs corresponds to the searched earth resistance Rx. If the displayed resistance value is too high, check the earth connection.

Rs: Earth loop resistance (entire earth loop)

Rx: Searched earth resistance

R1 ... RN: Parallel earth connections, negligible the higher the number

Rb: Resistance of the soil, usually <1 Ω

RI: Resistance of the earthing lines, usually <1 Ω

see figure 3: Earth loop resistance measurement

6. Ambient conditions

  • The BENNING CM E1 is intended for measurements under dry ambient conditions
  • Maximum barometric height for measurements: 2000 m
  • Overvoltage category / installation category: IEC 60664/ IEC 61010 → 300 V category III
  • Contamination class: 2 according to EN 61010-1
  • Protection class: IP 30 (DIN VDE 0470-1, IEC/ EN 60529)

IP 30 means: Protection against access to dangerous parts and protection against solid impurities of a diameter > 2.5 mm, (3 - first index). No protection against water, (0 - second index).

- Operating temperature and relative humidity:

For operating temperatures from 0 °C to 50 °C: relative air humidity lower than 85 %, noncondensing.

- Storage temperature: The BENNING CM E1 can be stored at temperatures between - 20 °C and + 60 °C, at a relative air humidity lower than 75 % without batteries.

7. Electrical specifications

Note: The measuring accuracy is specified as the sum of:

- a relative part of the measured value and

- a number of digits (i.e. counting steps of the last digit).

This measuring accuracy applies to temperatures from 23 ^ ± 5 ^ and a relative air humidity lower than 80 %.

7.1 Loop resistance ranges

Measuring range Resolution Measurement accuracy*1
0.025 Ω - 0.250 Ω 0.002 Ω ± (1.5 % ± 0.05 Ω)
0.251 Ω - 1.000 Ω 0.02 Ω ± (1.5 % ± 0.05 Ω)
1.001 Ω -10.00 Ω 0.02 Ω ± (1.5 % ± 0.1 Ω)
10.01 Ω - 50.00 Ω 0.04 Ω ± (2.0 % ± 0.3 Ω)
50.01 Ω -100.0 Ω 0.04 Ω ± (2.0 % ± 0.5 Ω)
100.1 Ω - 200.0 Ω 0.4 Ω ± (3.0 % ± 1.0 Ω)
200.1 Ω - 400.0 Ω 2 Ω ± (5.0 % ± 5 Ω)
400.1 Ω - 600.0 Ω 5 Ω± (10 % ± 10 Ω)
600.1 Ω - 1500 Ω20 Ω± (20 %)

*1 valid for purely ohmic resistances, external magnetic field <30 A/m, external electric field <1 V/m Measuring frequency: 3.333 kHz
Accuracy of the reference resistance loop: approx. ± 1 %

Alarm threshold RangeResolution
Upper alarm (HI)0 - 1510 Ω1 Ω
Lower alarm (LO)0 - 1510 Ω1 Ω

7.2 AC current/ leakage current

The measured value is obtained and displayed as real r.m.s. value (True RMS, AC coupling). Its calibration is adapted to sinusoidal curves. In case of deviations from this curve shape, the accuracy of the displayed value decreases. Crest factor < 3.5

Measuring range ResolutionMeasurement accuracy in frequency range 50 Hz - 60 Hz
0.300 mA - 1.000 mA0.001 mA± (2.0 % ± 0.05 mA)
1.00 mA - 10.00 mA0.01 mA± (2.0 % ± 0.03 mA)
10.0 mA - 100.0 mA0.1 mA± (2.0 % ± 0.3 mA)
100 mA - 1000 mA1 mA± (2.0 % ± 3 mA)
0.200 A - 4.000 A0.001 A± (2.0 % ± 0.03 A)
4.00 A - 35.00 A0.01 A± (3.0 % ± 0.03 A)

Overload protection: 100 A

8. Measuring with the BENNING CM E1

8.1 Preparation for measuring

Operate and store the BENNING CM E1 at the specified storage and operating temperatures only! Do not permanently expose the device to sunlight.

- Strong sources of interference in the vicinity of the BENNING CM E1 might involve unstable readings and measuring errors.

8.2 Loop resistance measurement

BENNING CM E1 - Loop resistance measurement - 1

Do not exceed the maximum permitted voltage with respect to earth potential! Electrical danger!

  • Open the measuring clamp and check whether the metallic contact surfaces inside the clamp are free of dust and contamination.
  • Let the jaws of the measuring clamp snap together several times (open and close them) in order to establish a safe contact.
  • Use the rotary switch ③ to select the function (loop resistance) or ( + 1)) (loop resistance with alarm). Wait until the end of the self-calibration (CAL 7, CAL 6, ..., CAL 2, CAL 1) is confirmed by an acoustic signal and the ".OL" symbol appears on the digital display ⑧.

  • Open the measuring clamp and clamp the earthing conductor/earth connection to be tested. Let the jaws of the measuring clamp snap together again several times.

  • The measured value of the earth loop resistance can be read on the digital display B.

Note:

  • During self-calibration, do not place the measuring clamp ① around a conductor or open the measuring clamp.
  • If self-calibration does not stop, check whether the metallic contact surfaces inside the clamp are free of dust and contamination.
  • If interfering signals (current >3 A, voltage >30 V) are detected during the measurement, the "NOISE" symbol F appears on the digital display ④ and the measurement might be affected.
  • If the clamp was not closed correctly during the measurement, the C symbol G appears on the digital display 4.

see fig. 3: Earth loop resistance measurement

8.3 AC current/ leakage current measurement

BENNING CM E1 - AC current/ leakage current measurement - 1

Do not exceed the maximum permitted voltage with respect to earth potential! Electrical danger!

  • Use the rotary switch ③ of the BENNING CM E1 to select the desired function mA or A.
  • Open the measuring clamp ① and clamp the earthing conductor/earth connection to be tested.
  • The measured value of the leakage current can be read on the digital display Ⓑ.

see figure 4: AC current/ leakage current measurement

9. Maintenance

BENNING CM E1 - Maintenance - 1

Before opening the BENNING CM E1, strictly observe that the device is free of voltage! Electrical danger!

9.1 Securing the device

Under certain circumstances, safe operation of the BENNING CM E1 might no longer be ensured, e.g. in case of:

  • visible damage of the casing.
  • incorrect measuring results,
  • recognizable consequences of prolonged storage under inadmissible conditions and
  • recognizable consequences of extraordinary stress due to transport.

In such cases, immediately switch off the BENNING CM E1, disconnect it from the measuring points and secure it against further use.

9.2 Cleaning

Clean the exterior of the device with a clean dry cloth (exception: special cleaning wipers). Do not use any solvents and/or abrasives to clean the device. Make sure that the battery compartment and the battery contacts are not contaminated by leaking battery electrolyte.

If there are electrolyte contamination or white deposits in the area of the battery or the battery compartment, clean these areas as well by means of a dry cloth.

9.3 Battery replacement

BENNING CM E1 - Battery replacement - 1

Before opening the BENNING CM E1, strictly observe that the device is free of voltage! Electrical danger!

The BENNING CM E1 is fed by a 9 V block battery.

The battery must be replaced (see fig. 5) when the battery symbol ☐ appears in the display ④. When the BENNING CM E1 is switched on, a battery test is carried out.

Proceed as follows to replace the battery:

  • Switch off the BENNING CM E1.
  • Put the BENNING CM E1 face down and unscrew the screws of the bottom part of the housing.
  • Lift the bottom part of the housing off the front part.
  • Lift the discharged battery from the battery compartment and disconnect the battery supply lines from the battery.

  • The new battery have to be connected to the battery supply lines, and arrange these such that they are not crushed between the housing parts. Then, insert the new battery into the front part at the provided place.

  • Lock the bottom part of the housing into place on the front part and tighten the screws.

see figure 5: Battery replacement

BENNING CM E1 - Battery replacement - 2

Make your contribution for environmental protection! Do not dispose of discharged batteries via the household waste. Instead, return them to a collecting point for discharged batteries or spezial waste. Please look for information in your community's facilities.

9.4 Calibration

Benning guarantees compliance with the technical and accuracy specifications stated in the operating manual for the first 12 months after the delivery date.

To maintain accuracy of the measuring results, the device must be recalibrated in regular intervals by our factory service. We recommend recalibrating the device once a year. For this purpose, send the device to the following address:

10. Environmental note

BENNING CM E1 - Environmental note - 1

At the end of product life, dispose of the unserviceable device via appropriate collecting facilities provided in your community.

text_image NO. 0.001 10.00 Ω NO.
Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : BENNING

Model : CM E1

Category : Multimeter