USER MANUAL Comfort 4 Aponorm
text_image
mIoN20 36.8°C IBCD
natural_image
Partial view of a dark gray circular shape with no visible text or symbolsIR200
Ear Thermometer

Microlife IR 200





text_image
microlife Pur Mintell MintellMicroscopy
Eccerelnaise 139
0442 W. drau/ Switzerland
www.incrbc.com
- - ModuleLAB
F. Linson, 52
0022 Virus: Uhuang
C€0044
071200-6711(2)
Microlife IR 200 IR 200 Guarantee Card
































microlife
Microlife Ear Thermometer IR 200
① Measuring sensor
② Display
① N-bulon (memory) ② NBTB bulon
④ NOOL DUMI ⑤ Befesa processing team
(4) START(K)
② All segments displayed
① Memory
②. Pretty or reasoning
④ Canei poion in acson d². Neuromand ognolis
- Body mode
① Object node
34 In correct location ind
⑲ Error function display
- Lowshibey industrial
(2) Charging: Hellen Lasso and Pannan de Bocalinde
de Recalha
20 Messure temperature
② Pressure temperature too low
② Ambient temperature too high
(3) AnteTher,anina
Elkndesky
② Pathology
② DomTus
② Beaper incisioning
④ Reporting history (4) Protocols can
(3) 田志华

Read the refractionally before using the device.
Type RE applied part
This 'Mand'la hemiceleria is a high quality product incorporating
The latest technology and tested in accordance with International standards, 42th by using technology, this device can provide a
stable, heat-inference-free reading with each measurement.
The advice performs a self-test every time it is switched on its
always guarantee the extracted accuracy of any measurement. The May 10th Fair Thermopower is planned for the period.
the W-CAE LTR HOMMELK is normal of the period's measure normal and monitoring of human body temperature. It is
Intended to use on people of all ages.
This dichometer has been clinically tested and proven to be safe and accurate when used in accordance to the operating
Instruction manual.
Please read through these instructions carefully in order for you to understand all details and risks.
(1) 2016 年 10 月 1 日
Table of Contents
- The Advantages of this Thermometer
• Measurement in second • Development process
• Múltice uase la de range of meesuremente
• Prob: cover free
- Prola LED - Accuracy of
- Cante and easy to
• Multiple readings recall
• Safe and hygienic
• Fover 25mm 2. Important Sails
- Important safety instructions
-
How this Thermometer means
-
To send an instructional assessment
-
Control Displays and Symbols
-
Selling Date, Time and Beoper Functions
-
Clearing between Body and Child Blind
-
Changing between body and object mass
- Directions for Use
- Meaning in hot
- Measuring in oct mode
- Changing between Celsius and Fahrenheit
-
How to recall 32 readings in Memory Mode
-
Error Messages
-
Cleaning and D
microble
Name of Dardner / Sam de Schelzer / Norecht
The following table is in English:
Наименование
Date of Purchase / Date of order / Date of ann. stov
Ko#d#m/S#n#s##/B#e#es###/
- Battery Replacement
- Guarantee
- Technical Specifications
- www.microlife.com Guarantee Card (see Back Cover)
1. The Advantages of this Thermometer
Measurement in 1 second
The innovative infrared technology allows measurement of ear temperature in only 1 second.
Correct position indication
ACCUsens guidance system confirms the right position in the ear with «good» displayed on the LCD display and a beep.
Multiple uses (wide range of measurement)
This thermometer offers a wide range of measurement from 0.1 - 99.9 °C / 32.2 - 211.8 °F, meaning the unit can be used to measure body temperature or it also has a feature allowing it to be used to measure surface temperature of the following examples:
- Milk surface temperature in a baby's bottle
• Surface temperature of a baby's bath - Ambient temperature
Probe cover free
This thermometer is more user-friendly and more cost-effective since a probe cover is not required.
Probe LED
This thermometer includes a probe LED light which enables the user to find the correct ear position in the dark and is used as a reminder (flashing LED and probe light) to clean the thermometer after each temperature measurement.
Accurate and reliable
The unique probe assembly construction incorporates an advanced infrared sensor, ensuring that each measurement is accurate and reliable.
Gentle and easy to use
- The ergonomic design enables simple and easy use of the thermometer.
- This thermometer can even be used on a sleeping child without causing any interruption.
- This thermometer is quick, therefore child-friendly.
Multiple readings recall
Users will be able to recall the last30 readings with a record of both time and date when entering the recall mode, enabling efficient tracking of temperature variations.
Safe and hygienic
- No risk of broken glass or mercury ingestion.
• Completely safe for use on children. - Cleaning of the probe can be done with an alcohol-moistened cotton tissue, making this thermometer completely hygienic for use by the whole family.
Fever alarm
10 short beeps and a red LCD backlight alert the patient that he/she may have a temperature equal to or higher than 37.5 °C.
2. Important Safety Instructions
- Follow instructions for use. This document provides important product operation and safety information regarding this device. Please read this document thoroughly before using the device and keep for future reference.
- This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by incorrect application.
- Never immerse this device in water or other liquids. For cleaning please follow the instructions in the «Cleaning and Disinfecting» section.
- Do not use this device if you think it is damaged or notice anything unusual.
- Never open this device.
- Earwax in the ear canal may cause a lower temperature reading. Therefore it is important to ensure the subject's ear canal is clean.
- If the measurement result is not consistent with the patient's finding or unusually low, repeat the measurement every 15 minutes or double check the result by another core body temperature measurement.
- This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in the «Technical Specifications» section.
- Ensure that children do not use this device unsupervised; some parts are small enough to be swallowed.
- Do not use this device close to strong electromagnetic fields such as mobile telephones or radio installations. Keep a minimum distance of 3.3 m from such devices when using this device.
- Protect it from:
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
- If the device is not going to be used for a prolonged period the battery should be removed.

WARNING: The measurement results given by this device is not a diagnosis. It is not replacing the need for the consultation of a physician, especially if not matching the patient's symptoms. Do not rely on the measurement result only, always consider other potentially occurring symptoms and the patient's feedback. Calling a doctor or an ambulance is advised if needed.
3. How this Thermometer measures Temperature
This thermometer measures infrared energy radiated from the eardrum and the surrounding tissue. This energy is collected through the lens and converted to a temperature value. The measured reading obtained directly from the eardrum (Tympanic Membrane) ensures the most accurate ear temperature. Measurements taken from the surrounding tissue of the ear canal generate lower readings and may result in misdiagnosis of fever.
To avoid an inaccurate measurement
- Switch on the thermometer by pressing the START/IO button ⑥.
- After one beep is heard (and the temperature scale icon is flashing), straighten the ear canal by gently pulling the middle of the ear back and up.
- Place the probe ① firmly into the ear canal. «Good» will be displayed with a short beep to confirm the device has detected the correct position. Press the START/IO button ⑥ and keep the probe in the ear until the thermometer beeps to signal the completion of the measurement.
4. Control Displays and Symbols
- All segments displayed ⑦: Press the START/IO button ⑥ to turn on the unit; all segments will be shown for 1 second.
-
Ready for measurement ⑨: When the unit is ready for measurement, the «°C» or «°F» icon will keep flashing. The probe LED light is activated and will keep flashing.
-
Correct position indication ⑩: The probe LED light will stop flashing (stays illuminated) and «good» will be displayed on the LCD, when the measuring sensor detects an appropriate position.
- Measurement complete ⑪: The reading will be shown on the display ② with the «°C» or «°F» icon; the unit is ready for the next measurement, when the «°C» or «°F» icon is flashing again.
- Low battery indicator ⑯: When the unit is turned on, the «battery» icon will keep flashing to remind the user to replace the battery.
5. Setting Date, Time and Beeper Functions
Setting the date and time
- After the new battery is fitted, the year number flashes in the display ②7. You can set the year by pressing the M-button ③. To confirm and then set the month, press the MODE button ④.
- Press the M-button ③ to set the month. Press the MODE button ④ to confirm and then set the day.
- Follow the previously mentioned instructions to set the day, hours and minutes.
- Once you have set the minutes and pressed the START/IO button ⑥, the date and time are set and the time is displayed.

If no button is pressed for 20 seconds, the device automatically switches to ready for measuring ⑨.

Cancel time setup: Press the START/IO button ⑥ during time setup. The LCD will show Date/Time icons with «--:--». After that press the START/IO button ⑥ to start the measurement. If no further action is taken within 60 seconds, the device will automatically turn off.

Change current date and time: Press and hold the MODE button ④ for approx. 8 seconds until the year number starts to flash ②7. Now you can enter the new values as described above.
Setting the beeper
- Press and hold the MODE button ④ for 3 seconds to set the beeper ②8.
- Press the M-button ③ to either turn the beeper on or off. The beeper is activated when the beeper icon ②8 is shown without a cross.

When the beeper setting has been chosen, press the START/IO button ⑥ to enter the «ready for measuring» mode; otherwise the device automatically switches to ready for measuring after 10 seconds ⑨.
6. Changing between Body and Object Mode
- Press the START/IO button ⑥. The display ② is activated to show all segments for 1 second.
- The default mode is body mode ⑫. Press the MODE button ④ to switch to object mode ⑬. For switching back to body mode, press the MODE button again.
7. Directions for Use
Always remove the protective cap ③0 before use.
Measuring in body mode
- Press the START/IO button ⑥. The display ② is activated to show all segments for 1 second.
- When the «°C» or «°F» icon is flashing, a beep sound is heard and the thermometer is ready for measuring ⑨.
- The probe LED light is activated and will keep flashing.
- Straighten the ear canal by pulling the ear up and back to give a clear view of the eardrum.
- For children under 1 year; pull the ear straight back.
-
For children of 1 year to adult; pull the ear up and back.
Also refer to the short instructions at the front. -
While gently pulling the ear, insert the probe snugly into the ear canal.
-
The probe LED light will stop flashing (stays illuminated) and «good» will be displayed on the LCD, when the measuring sensor detects an appropriate position.
-
Immediately press the START/IO button ⑥. Release the button and wait for the beep sound. This indicates the end of the measurement.
-
Remove the thermometer from the ear canal. The display shows the measured temperature ^⑪ .
-
For the next measurement wait until the «°C»/«°F» icon is flashing and follow steps 5-7 above.
-
Press and hold the START/IO button ⑥ for 3 seconds to turn off the device; otherwise the device will automatically switch off after approx. 60 seconds.
Measuring in object mode
- Press the START/IO button ⑥. The display ② is activated to show all segments for 1 second.
- Press the MODE button ④ to switch to object mode.
- Aim the thermometer at the center of the object you want to measure with a distance of no more than 5 cm. Press the
START/IO button ⑥. After 1 second a long beep will verify the completion of measurement.
-
Read the recorded temperature from the LCD display.
-
For the next measurement wait until the «°C»/«°F» icon is flashing and follow steps 3-4 above.
NOTE:
• Patients and thermometer should stay in similar room condition for at least 30 minutes.
- To ensure accurate readings, wait at least 30 sec. after 3-5 continuous measurements.
- Accumulation of ear wax on the probe can result in less accurate temperature readings or cross infection between users.
- It is essential that the probe is cleaned after every measurement. Therefore, this device reminds the user to clean the probe when turning off the device. «CLEAN ME» 24 is displayed and the probe LED light will flash for 3 seconds. For cleaning, follow the instructions in the «Cleaning and Disinfecting» section.
- After cleaning the measuring sensor ① with alcohol, wait 5 minutes before taking the next measurement, in order to allow the thermometer to reach its operating reference temperature.
- 10 short beeps and a red LCD backlight alert the patient that he/she may have a temperature equal to or higher than 37.5^ .
- For an infant, it is best to have the child lying flat with his/her head sideways so the ear is facing upwards. For an older child or adult, it is best to stand behind and slightly to the side of the patient.
- Always take the temperature in the same ear, since the temperature readings may be different from ear to ear.
- In the following situations it is recommended that three temperatures in the same ear be taken and the highest one taken as the reading:
-
New born infants in the first 100 days.
-
Children under three years of age with a compromised immune system and for whom the presence or absence of fever is critical.
-
When the user is learning how to use the thermometer for the first time until he/she has familiarized himself/herself with the device and obtains consistent readings.
-
If the measurement is surprisingly low.
- Don't take a measurement while or immediately after nursing a baby.
- Don't use the thermometer in high humidity environments.
- Patients should not drink, eat or exercise before/while taking the measurement.
- Doctors recommend rectal measurement for newborn infants within the first 6 months, as all other measuring methods might lead to ambiguous results.
- Readings from different measuring sites should not be compared as the normal body temperature varies by measuring site and time of day, being highest in the evening and lowest about one hour before waking up. Normal body temperature ranges:
- Axillar: 34.7 - 37.3 °C / 94.5 - 99.1 °F
- Oral: 35.5 - 37.5 °C / 95.9 - 99.5 °F
- Rectal: 36.6 - 38.0 °C / 97.9 - 100.4 °F
- Microlife IR 200: 35.4 - 37.4 °C / 95.7 - 99.3 °F
8. Changing between Celsius and Fahrenheit
This thermometer can display temperature readings in either Fahrenheit or Celsius. To switch the display between °C and °F, press and hold the MODE button ④ for 3 seconds; the beeper icon is shown on the display. Press the MODE button again; the current measurement scale («°C» or «°F» icon) will be shown on the display ⑰. Change the measurement scale between °C and °F by pressing the M-button ③. When the measurement scale has been chosen, press the START/IO button ⑥ to enter the «ready for measuring» mode; otherwise the device automatically switches to ready for measuring after 10 seconds ⑨.
9. How to recall 30 readings in Memory Mode
This thermometer can recall the last 30 readings with a record of both time and date.
- Recall mode ⑱: Press the M-button ③ to enter recall mode when the power is off. The memory icon «M» will flash.
- Reading 1 - the last reading ⑲: Press and release the M-button ③ to recall the last reading. Number «1» and a flashing «M» are displayed.
Pressing and releasing the M-button ③ after the last 30 readings have been recalled will resume the above sequence from reading 1.
10.Error Messages
- Measured temperature too high ⑳: Displays «H» when measured temperature is higher than 43 °C / 109.4 °F in body mode or 99.9 °C / 211.8 °F in object mode.
- Measured temperature too low ②: Displays «L» when measured temperature is lower than 32 °C / 89.6 °F in body mode or 0.1 °C / 32.2 °F in object mode.
- Ambient temperature too high ②: Displays «AH» when ambient temperature is higher than 40 °C / 104 °F.
- Ambient temperature too low ②3: Displays «AL» when ambient temperature is lower than 10 °C / 50 °F in body mode or lower than 5,0 °C in object mode.
- Incorrect location indication ⑭: The probe is not correctly inserted in the ear canal. Please insert the probe as described in this manual.
- Error function display ⑮: The system has a malfunction.
- Blank display ⑲: Check if the battery has been loaded correctly. Also check polarity (<+> and <->) of the battery.
- Flat battery indication ②6: If only the «battery» icon is shown on the display, the battery should be replaced immediately.
11.Cleaning and Disinfecting
Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl) to clean the thermometer casing and the measuring sensor. Ensure that no liquid enters the interior of the device. Never use abrasive cleaning agents, thinners or benzene for cleaning and never immerse the device in water or other cleaning liquids. Take care not to scratch the surface of the sensor lens and the display.
12.Battery Replacement
This device is supplied with one lithium battery, type CR2032. The battery needs replacing when only the «battery» icon ②6 is shown on the display.
Remove the battery cover ⑲ by sliding it in the direction shown. Insert the new battery with the + at the top.

Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with domestic waste.
13. Guarantee
This device is covered by a 5 year guarantee from the date of purchase. During this guarantee period, at our discretion, Microlife will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
- Transport costs and risks of transport.
- Damage caused by incorrect application or non-compliance with the instructions for use.
- Damage caused by leaking batteries.
- Damage caused by accident or misuse.
- Packaging/storage material and instructions for use.
- Regular checks and maintenance (calibration).
- Accessories and wearing parts: Batteries, Probe cover (optional).
Should guarantee service be required, please contact the dealer from where the product was purchased, or your local Microlife service. You may contact your local Microlife service through our website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee will be granted if the complete product is returned with the original invoice. Repair or replacement within guarantee does not prolong or renew the guarantee period. The legal claims and rights of consumers are not limited by this guarantee.
- Technical Specifications
| Type: | Adjusted mode Ear Thermometer IR 200 |
| Mode type: | Adjust mode |
| Measuring site: | Ear |
| Reference body site: | Sublingual |
| Measurement range: | Body mode: 32.0-43.0 °C / 89.6 - 109.4 °FObject mode: 0.1 - 99.9 °C / 32.2 - 211.8 °F |
| Resolution: | 0.1 °C / °F |
| Measurement accuracy (Laboratory): | Body mode: ±0.2 °C, 35.0 ~ 42.0 °C / ±0.4 °F, 95.0 ~ 107.6 °F±0.3 °C, 32.0 ~ 34.9 °C and 42.1 ~ 43.0 °C /±0.5 °F, 89.6 ~ 94.8 °F and 107.8 ~ 109.4 °FObject mode:±1.0 °C, 0.1 ~ 99.9 °C / ±2 °F, 32.2 ~ 211.8 °F |
| Clinical results: | Repeatability: 0.19 °CBias: 0.03 °CLimits of agreement: 1.33 °C |
| Display: | Liquid Crystal Display, 4 digits plus special icons |
Acoustic: The unit is turned ON and ready for the measurement: 1 short beep. Complete the measurement: 1 long beep. System error or malfunction: 3 short beeps. Fever alarm: 10 short beeps.
Memory: 30 readings recall in the memory mode with a record of both time and date.
Backlight: The display light will be GREEN for 1 second, when the unit is turned ON. The display light will be GREEN for 5 seconds, when a measurement is completed with a reading less than 37.5 °C / 99.5 °F. The display light will be RED for 5 seconds, when a measurement is completed with a reading equal to or higher than 37.5 °C / 99.5 °F.
Operating 10 - 40 °C / 50.0 - 104 °F
conditions: 15 - 95 % relative maximum humidity
Storage -25 - +55 °C / -13 - +131 °F
conditions: 15 - 95 % relative maximum humidity
Automatic Approx. 1 minute after last measurement has been
Switch-off: taken.
Battery: 1 x CR2032 battery 3V
Battery
lifetime: approx. 1000 measurements (using a new battery)
Dimensions: 159 x 43 x 60 mm
Weight: 60 g (with battery), 58 g (w/o battery)
IP Class: IP22
Reference to ISO 80601-2-56; IEC 60601-1; IEC 60601-1-2
standards: (EMC); IEC 60601-1-11
Expected
service life: 5 years or 12000 measurements
This device complies with the requirements of the Medical Device Directive 93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act a biennial technical inspection is recommended for professional users. Please observe the applicable disposal regulations.
15.www.microlife.com
Detailed user information about our thermometers and blood pressure monitors as well as services can be found atwww.microlife.com.
www.microlife.com/support
www.microlife.com/support
www.microlife.com/support
www.microlife.com/support
www.microlife.com/support
:IPIP22
IEC 60601-1-2 (EMC); IEC 60601-1-11
www.microlife.com .15
FA



MODE

«L»
«AL»
www.microlife.com/support
IP22
IEC 60601-1-2 (EMC); IEC 60601-1-11