Brecknell MS140300 - Scale

MS140300 - Scale Brecknell - Free user manual and instructions

Find the device manual for free MS140300 Brecknell in PDF.

📄 76 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice Brecknell MS140300 - page 5
View the manual : Français FR English EN
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Portable medical electronic scale with ramp
Brand Brecknell
Model MS140300
Capacity 300 kg / 660 lb / 660 lb:oz
Accuracy ± 3 divisions (0.3 kg / 0.6 lb / 15 oz)
Platform Dimensions 520 mm (L) x 520 mm (W) x 65 mm (H)
Indicator Dimensions 164 mm (L) x 76 mm (W) x 64 mm (H) (approx.)
Power Supply 4 AAA batteries (20 h battery life) or AC adapter 6 VDC, 500 mA
Display LCD 5 digits, 7 segments, 16 mm
Keypad 4 push buttons (ON/OFF/ZERO, TARE, UNIT, HOLD/PRINT)
Operating Temperature 5 °C to 35 °C
Humidity ≤ 95 % RH non-condensing
Communication RS-232 (COM1 port, TXD, RXD, GND), baud rate 1200-19200 bps
Main Functions Zero, tare, hold, print, unit change (kg, lb, lb:oz), adjustable auto-off
Calibration Software calibration with EEPROM storage, requires standard weight > 25% of capacity
Configuration 19 configurable parameters (resolution, units, delays, etc.)
Care and Cleaning Wipe exterior with damp cloth and mild detergent; do not spray directly on display
Safety Adjustable auto-off, ground fault protection (RCD/GFCI), EMC Class A compliance
Serviceability Service by qualified personnel only; adjustable zero potentiometer on PCB

Frequently Asked Questions - MS140300 Brecknell

How to turn the scale on or off?
Press the ON/OFF button to turn the indicator on or off.
How to perform zeroing?
Make sure the platform is empty and stable, then press the ZERO button. The scale can also auto-zero at power-on if configured.
How to use the Tare function?
Place an empty container on the scale, wait for stability, then press the TARE button. The display shows net weight at zero. To cancel tare, remove all load and press TARE again.
How to change the weighing unit?
Press the UNIT button to toggle between kg, lb, and lb:oz. Available units depend on configuration (parameter P11).
How to use the Hold function?
Press the HOLD/PRINT button to enter Hold mode. The scale can capture a fluctuating weight (live animal) or lock the display. Parameter P3 allows choosing the mode (max, auto, range).
How to calibrate the scale?
Set the config/cal switch to ON. In weighing mode, hold TARE and ON/OFF/ZERO until 'RL-' appears. Follow prompts for zero, then load a standard weight (≥25% of capacity) and enter its value using HOLD and UNIT buttons. Confirm with TARE.
What to do if the display shows 'O-----' or 'O_____'?
These messages indicate zero is out of the allowed range. Check that the platform is empty, perform a zero calibration, or refer to the Troubleshooting chapter.
How to configure the scale parameters?
Hold the ON/OFF/ZERO and UNIT buttons until 'SEEUP' appears. Use UNIT to change the flashing digit, HOLD/PRINT to move the cursor, and TARE to confirm and go to the next parameter. Parameters P1 to P19 are described in the manual.
How to use the RS-232 port?
Connect a DB9 cable to the COM1 port (pins 2 TXD, 3 RXD, 5 GND). Configure baud rate (P5) and protocol (P6) in the configuration menu. Use serial commands (W, S, Z, T, U, L, X) to query or control the scale.
What to do in case of error message 'CALEr'?
This message appears during incorrect calibration. Possible causes: wrong weight entered, unstable or non-linear load. Restart calibration following steps carefully, with a correct standard weight and stable load.

User questions about MS140300 Brecknell

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Scale in PDF format for free! Find your manual MS140300 - Brecknell and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. MS140300 by Brecknell.

USER MANUAL MS140300 Brecknell

Portable Medical Electronic Handrail Scale

Brecknell MS140300 - Portable Medical Electronic Handrail Scale - 1

User Instructions

Brecknell, LLC 2013. All rights reserved.

No part of this publication may be reproduced, stored in an electronic retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopying, recording or otherwise without the prior written consent of the copyright owner, or as permitted by law or under license. Full acknowledgment of the source must be given. Brecknell is a registered trade mark of the Brecknell, LLC. This publication was correct at the time of going to print however, Brecknell, LLC reserves the right to alter without notice the specification, design, price or conditions of supply of any product or service at any time.

All third party brands and product names used within this document are trademarks or registered trademarks of their respective holders.

Table of Contents

Chapter 1 General Information andWarnings 3

About this Manual 3

Text Conventions 3
Special Messages 3

Warnings 4

EMC Compliance 4
Routine Maintenance 5
Cleaning the Indicator 5
Sharp Objects 5

Chapter 2 Specifications 6

Scale Platform 6

Capacity 6
Accuracy 6
Dimension 6

Scale Indicator 6

Power Supply 6
Dimension 6
Display 7
Keypad 7

Environment 7

Load Cell Excitation 7
Communication 7

Chapter 3 Introduction 8

Indicator Mounting 8
Front Panel 8
Display 9
Keyboard 10

Navigating the Weigh Mode 11

Power the Indicator 11
Enter or Exit the HOLD Mode 11
Zero 11
Tare 11
Clear Tare Weight 11
Output Data (Print) 12
Change Weight Unit 12

Hold Function 14

Access the Hold Mode 14

p Mode 15

Entering the Setup Menu 15
Navigating the Setup Menu 15
Setup Menu Parameters 16
Relationship of Capacity and P7, P8 and P9 Settings 21
Exit the Setup Menu 24

Chapter 5 Calibration 25

Calibration Mode 25
Display ADC Code or Working Voltage Value 26

Chapter 6 Serial Communication 28

Com Port 1 28
Protocol 28
Transaction String 28

Commands and Response 30

Chapter 7 Troubleshooting 32

Display Characters 32

Display Symbols 33

Error Messages and Troubleshooting 33

1 General Information andWarnings

1.1 About this Manual

This manual is divided into chapters by the chapter number and the large text at the top of a page. Subsections are labeled as shown by the 1 and 1.1 headings shown above. The names of the chapter and the next subsection level appear at the top of alternating pages of the manual to remind you of where you are in the manual. The manual name and page numbers appear at the bottom of the pages.

1.1.1 Text Conventions

The keys used to interface with the PS-500 are located on the front panel of the indicator. The keystrokes are shown in BOLD incased between brackets. (e.g. [ZERO])

Displayed messages appear in LCD format (e.g. SELECTUP) type and reflect the case of the displayed message.

1.1.2 Special Messages

Examples of special messages you will see in this manual are defined below. The signal words have specific meanings to alert you to additional information or the relative level of hazard.

Brecknell MS140300 - Special Messages - 1

CAUTION!

This is a Caution symbol.

Cautions give information about procedures that, if not observed, could result in damage to equipment or corruption to and loss of data.

Brecknell MS140300 - Special Messages - 2

ELECTRICAL WARNING!

THIS IS AN ELECTRICAL WARNING SYMBOL.

ELECTRICAL WARNINGS MEAN THAT FAILURE TO FOLLOW

SPECIFIC PRACTICES OR PROCEDURES MAY RESULT IN

ELECTROCUTION, ARC BURNS, EXPLOSIONS OR OTHER HAZARDS

THAT MAY CAUSE INJURY OR DEATH.

Brecknell MS140300 - Special Messages - 3

NOTE: This is a Note symbol. Notes give additional and important information, hints and tips that help you to use your product.

1.2Warnings

  • Read all operating instructions carefully before use.
  • Avoid lengthy exposure to extreme heat or cold. Your scale works best when operated at normal room temperature. Always allow the scale to acclimate to a normal room temperature before use.
  • Allow sufficient warm up time. Turn the scale on and wait for a few minutes if possible. This will give the internal components a chance to stabilize before weighing.
    These electronic scales are precision instruments. Do not operate near an in-use cell phone, radio, computer or other electronic device. These devices emit RF and can cause unstable scale readings.
  • Avoid using in heavy vibration and airflow conditions.
  • Read the weight shortly after loading. The output of the loadcell and A/D may be a little influenced after weight sits for a long time.

Brecknell MS140300 - 1.2Warnings - 1

DANGER: FOR YOUR PROTECTION, ALL MAINS (110V OR 230V) EQUIPMENT USED WHERE DAMP OR WET CONDITIONS MAY OCCUR MUST BE SUPPLIED FROM A CORRECTLY FUSED SOURCE AND PROTECTED BY AN APPROVED GROUND FAULT PROTECTION DEVICE (RCD, GFCI ETC).

Brecknell MS140300 - 1.2Warnings - 2

DANGER: RISK OF ELECTRICAL SHOCK. BE SURE TO UPLUG THE INDICATOR BEFORE REMOVIDING THE COVER OR OPENING THE UNIT. REFER TO QUALIFIED SERVICE PERSONNEL FOR SERVICE.

1.3 EMC Compliance

The following warning may be applicable to your machine.

Brecknell MS140300 - EMC Compliance - 1

CAUTION!

This is a Class A product.

In a domestic environment this product may cause radio interference in which the user may be required to take adequate measures.

United States

This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

Canada

This digital apparatus does not exceed the Class A limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of Communications.

1.4 Routine Maintenance

Brecknell MS140300 - Routine Maintenance - 1

IMPORTANT: This equipment must be routinely checked for proper operation and calibration. Application and usage will determine the frequency of calibration required for safe operation.

Always turn off the machine and isolate from the power supply before starting any routine maintenance to avoid the possibility of electric shock.

1.5 Cleaning the Indicator

Brecknell MS140300 - Cleaning the Indicator - 1

Table 1.1 Cleaning DOs and DON'Ts

DO DO NOT
Wipe down the outside of standard products with a clean cloth, moistened with water and a small amount of mild detergentAttempt to clean the inside of the indicator
Use harsh abrasives, solvents, scouring cleaners or alkaline cleaning solutions
Spray the cloth when using a proprietary cleaning fluidSpray any liquid directly on to the display window

1.6 Sharp Objects

Do not use sharp objects such as screwdrivers or long fingernails to operate the keys.

2 Specifications

2.1 Scale Platform

2.1.1 Capacity

660 lb x 0.2 lb
-300 kg x 0.1 kg
660 lb x 5 oz

2.1.2 Accuracy

± 3 divisions (0.6 lb, 0.3 kg, 15 oz)

2.1.3 Dimension

20.5^ × 20.5^ × 2.5^ 0520mm(L)× 520mm(W)× 65mm(H)

2.2 Scale Indicator

  • Input signal range: 0mV - +30mV
    Sensitivity: >0.2 V / grad
  • Internal Resolution: Approximately 520,000 counts
    Display Resolution: can be selected between 500-100,000
    System Linearity: within 0.01% of Full Scale
  • Calibration Method: Software calibration with long-term storage in EEPROM

2.2.1 Power Supply

Alkaline batteries: 4 AAA size cells (20 hour battery life)
AC adapter: 6VDC, 500mA with central negative

Brecknell MS140300 - Power Supply - 1

Work current: .25mA
(when voltage in 5Vdc-8Vdc and not include load cell consumption)

2.2.2 Dimension

-6.5" L x 3.2" H x 1.2" W -164 mm (L) x 762 mm (W) x 64 mm (H)

2.2.3 Display

-5-digit,7-segment, 0.625" (16mm) LCD

2.2.4 Keypad

-4 push button keys

2.2.5 Environment

Working temperature 5° to35°C
Storage temperature -10°Cto 70°C
Humidity ≤95% RH without condensation

2.2.6 Load Cell Excitation

Because more than one load cell can be used on a scale, the following are required for the load cell set to be used with this indicator.

Voltage 4.4VDC
Max. Current 55mA (can power 4-350 ohm loadcells)
Signal connection 4 or 6 lead with sense leads
Max Sensitivity 0.3mV/V to +3mV/V (must be fit to >0.2μV /display grad)
Input Resistor ≥80 Ω
Output Resistor <10 KΩ

2.2.7 Communication

Mode Full-duplex oronly output mode can be selected
Baud Rate Selectable: 1200-2400-4800-9600-19200 bps
Data Format 7 databits, even or odd parity, 1 stop bit8 data bits, non-parity, 1 stop bit
Protocol7 selected protocols (one compatible with NCI standard SCP-01)
Output Datagross weight, net weight, tare weight, indicator displaying weight, weighing unit etc.

3 Introduction

This chapter introduces indicator mounting, display and keypad operation in normal weigh mode.

3.1 Indicator Mounting

The PS-500 is supplied with an ABS plastic bracket. Wall mount vertically or bench mount horizontally depending on your needs.

Brecknell MS140300 - Indicator Mounting - 1

Brecknell MS140300 - Indicator Mounting - 2

3.2 Front Panel

The front panel incorporates the display and keypad.

Brecknell MS140300 - Front Panel - 1

The PS-500 indicator utilizes a 7 segment LCD (Liquid Crystal Display) providing the weight and system information.

Arrows will illuminate to designate the current mode.

Brecknell MS140300 - Front Panel - 2
Figure 3.1 PS-500 Display

Table 3.1 LCD Display Annunciators and Definitions

Display Symbol Description
ZERO The scaleis at zero point and the gross weight is 0.
HOLD The scaleis in HOLD mode. The current live weight is being displayed when the arrow flashes. The current weight is locked when the arrow does not flash.
NET Indicates netnet mode and the tare weight is not 0.
lb Indicates the current unit of measure is lb.
oz Indicates the current unit of measure is oz.
kg Indicates the current unit of measure is kg.

The keyboard consists of four keys, some of which have multiple functions.

Brecknell MS140300 - Front Panel - 3
Figure 3.2 SBI-521 Keypad

Table 3.2 Function of the Keys (Normal Working Mode)

Key Function
Hold Print←If this key is only set for HOLD (P2=0), press to enter or exit the HOLD mode.
If this key is only set for PRINT (P2=1), press to output the data according to P4 setting.
If this key is only set for HOLD and PRINT (P2=2), press to output data according to P4 setting and hold the current weight.
Units↑Choose weighing units; lb, kg, lb:oz
Note: The weighing units that can be used are restricted by display division, and calibration weight unit (restricted by P8, P9, and P10)
For example, if the calibration unit is kg; calibration display resolution is 50kg (that means: P8=5, P9=0, P10=0). If the UNIT key is pressed to choose weighing units, lb or lb:oz are not allowed to choose since the display resolution of 100lb or 2000oz is not available to this indicator.
Tare←Tare function
This function can be activated only when the scale is in stable mode and the gross weight is not a negative value.
On/Off Zero→Power on
Zero function: When the weight is within zero range, it will activate as ZERO function and clear the tare weight. When the weight is not within the zero setting range (P13), the scale will show 0---- (zero point is over the setting range) or 0_______ (zero is below the setting range).
Power off
ignore modification
Prepare to exit from current working mode

3.5.1 Power the Indicator

Turn the indicator on or off with the [ON/OFF] key.

3.5.2 Enter or Exit the HOLD Mode

Press the [HOLD] key.

3.5.3 Zero

When the weight is stable and within the zero range, press the [ZERO] key to set a new zero point. Refer to Table 3.3 for zero limitations.

3.5.4 Tare

When the gross weight is larger than zero and the scale is stable, press the [TARE] key. The indicator will show a net weight of zero and the arrow to the right of NET will be illuminated. Refer to Table 3.3 for hare limitations.

3.5.5 Clear Tare Weight

Remove any weight on platform and wait until the scale is stable. Press the [TARE] key.

Table 3.3 Zero and Tare Limitations

StandardWeight on platformData in TARE memory unitkey function
Tare key Zero key
USA ≤0 no No action Zero
yes Clear the tared weight
>0 no Tare
yes
Canada ≤0 no No action Zero
yes Clear the tared weight
>0noTare
yesNo action
Europe≤0 no No action Zero
yes Clear the tared weight Zero andclear the tared weight
>0 no Tare Zeroclear the
yes Zero andtared weight
None (same with Europe)≤0 no No action Zero
yes Clear the tared weight Zero andclear the tared weight
>0 no Tare Zeroclear the
yes Zero andtared weight

3.5.6 Output Data (Print)

When scale is stable press the [PRINT] key.

3.5.7 Change Weight Unit

Press the [UNIT] key to select kg, lb or lb:oz. Note: under some conditions lb:oz is not available. Please refer the following tables.

Table 3.4 Use kg as Primary Unit

Calibration Division ValueDisplay Division Value
kg lblb:oz
0.0001kg0.0001kg0.0002lbNot available
0.001kg0.001kg0.002lbNot available
0.01kg0.01kg0.02lb0.5oz
0.1kg 0.1kg0.2lbNot available
1kg1kg 2lbNot available
10kg10kg20 lbNot available
0.0002kg0.0002kg0.0005 lbNot available
0.002kg0.002kg0.005 lb0.1 oz
0.02kg0.02kg0.05 lb1 oz
0.2kg0.2kg0.5 lbNot available
2kg2kg5 lbNot available
20kg20kg50 lbNot available
0.0005kg0.0005kg0.001 lbNot available
kg lb lb:oz
0.005kg 0.005kg 0.01 lb 0.2 oz
0.05kg 0.05kg 0.1 lb 2oz
0.5kg 0.5kg 1 lb Not available
5kg 5kg 10 lb Not available
50kg 50kg Not available Not available

Table 3.5 Use Ib as Primary Unit

Calibration Division ValueDisplay Division Value
kg lb lb:oz
0.0001lbNot available0.0001lbNot available
0.001 lb 0.0005 kg 0.001 lbNot available
0.01 lb 0.005 kg 0.01 lb0.2 oz
0.1 lb 0.05 kg 0.1 lb2 oz
1 lb 0.5 kg 1 lb Not available
10 lb 5 kg 10 lb Not available
0.0002 lb0.0001 kg0.0002 lbNot available
0.002 lb 0.001 kg 0.002 lbNot available
0.02 lb 0.01 kg 0.02 lb 0.5 oz
0.2 lb0.1 kg0.2 lbNot available
2 lb1 kg2 lb Not available
20 lb10 kg20 lb Not available
0.0005 lb0.0002 kg0.0005 lbNot available
0.005 lb 0.002 kg 0.005 lb0.1 oz
0.05 lb0.02 kg0.05 lb 1 oz
0.5 lb0.2 kg0.5 lbNot available
5 lb2 kg5 lb Not available
50 lb20 kg50 lb Not available

3.6 Hold Function

HOLD function can be used to freeze a displayed number. In this mode, the scale can capture a dynamic number, hold a stable number, or average a unstable number and then HOLD (freeze) this number temporarily for the user to view or record.

The HOLD function can be used in normal weighing mode, counting mode and percent weighing mode. After entering HOLD mode, the speed of A/D converter can be increased to 80Hz (if USER-HOLD-AD.H.SPD is set to YES) from original 10Hz for some dynamic weighing applications.

It is possible to weigh restless weighing samples such as live animals or moving objects within the hold function. The indicator provides special mode settings to accommodate weight movements.

3.6.1 Access the Hold Mode

To enter the HOLD mode, press the [HOLD] key while in the normal weighing mode.

4 Setup Mode

To set up the indicator, you must first enter the appropriate menu mode. The front panel keys become directional navigators to move around in the menus. See Table 4.1 for details.

4.1 Entering the Setup Menu

  1. Press and hold the [ON/OFF/ZERO] and [UNIT] keys unit LEUP is displayed.
    1a. This indicator offers 19 different setup parameters.

4.2 Navigating the Setup Menu

  1. Once SETUP is displayed, use the [UNIT] key to change the flashed digits and use the [HOLD/PRINT] key to shift the flashed position.

The parameter will be designated in the following format:

PN.x or PNN.xx - P is parameter; N or NN is the parameter number; x or xx is the choice within the parameter. Refer to Table 4.2 for parameters and the options for each parameter.

  1. Use the [TARE] key to confirm the entry, save the data and go to the next parameter.

Table 4.1 Key Functions in the Setup Mode

Key Function
Hold PrintMove cursor from right to left.
UnitsScroll selected digit (0 - 9).
TareCycles through the parameters. Confirm parameter choice selection.
On/Off ZeroExit from setup mode to normal work mode.

This section provides more detailed descriptions of the selections found in the Setup Menu.

The menu table shows the options and default parameter in LCD display format to coincide with the actual display.

Table 4.2 Parameter Choices and Explanations

SubMenu1 Option Default Parameter Description Comment
P1UU-15U5Auto off time00 = no auto off01-15 = auto off time in minutes. The scale will automatically turn off after specified time when there has been no fluctuation of weight.
P2U-22Hold key function0 = only HOLD function1 = only PRINT function2 = HOLD and PRINT function. Short press for Print function and long press for Hold function.
P3U-22Hold mode function0 = no hold function1 = hold larger weight reading2 = auto release hold function when weigh is below 10d and auto hold new stable weigh (more than 10d)
3-5U3-50 = unchangeable reading when the variety is within ±3-50d
P4U-12Output of print data0 = no RS232 function. Will not transmit or receive data.1 = output display data when PRINT is pressed and weight reading is stable.<LF><weight reading, minus sign, decimal point, unit><CR><EXT>2 = output gross, tare and net weight and weigh unit when PRINT pressed.<LF><Gross:weight reading, minus sign, decimal point, unit><CR><EXT>Tare:weight reading, decimal point, unit><CR><EXT><Net:weight reading, minus sign, decimal point, unit><CR><EXT>3 = continuously output displayed weight reading and unit. No data will be received.Output format same as choice 1.4 = continuously output gross, tare, net weight and weigh unit. No data will be received. Output formats same as choice 2.5 = output display data and weigh unit one time when scale is stable. Output format same as choice 1.6 = output gross, tare, net weight and weigh unit one time when scale become stable.Output formats same as choice 2.7 = Bio-RS232 data output, compatible to NCI-SP1
SubMenu1OptionDefaultParameter DescriptionComment
P50-43Baud rate for RS-2320 = 1200 bps1 = 2400 bps2 = 4800 bps3 = 9600 bps4 = 19200 bps
P60-20RS-232 protocol0 = 8NO1 = 7O12 = 7E1
P700-319Scale resolution selectRefer to Table 4.3 for resolution choices and values.
PB0-21Calibration division sizeRefer to Table 4.4 and Table 4.50 = 11 = 22 = 5
PG0-51Decimal point in calibrationRefer to Table 4.4 and Table 4.50 = 11 = 0.12 = 0.013 = 0.0014 = 0.00015 = 10
P100=11Calibration unitRefer to Table 4.4 and Table 4.50 = kg1 = lb
P110-65 Weigh units enable 0 = only kg1 = only lb2 = only lb:oz3 = kg or lb4 = kg or lb:oz5 = lb or lb:oz6 = kg, lb, or lb:oz
P120-73Power on zero range0 = calibration zero -point +1%FS1 = calibration zero -point +2%FS2 = calibration zero-point +5%FS3 = calibration zero-point +10%FS4 = calibration zero-point +20%FS5 = calibration zero-point +50%FS6 = calibration zero-point +100%FS7 = no limitation
P130-72Zero range for [ZERO] key0 = power-on zero-point +1%FS1 = power-on zero-point +2FS2 = power-on zero-point +5%FS3 = power-on zero-point +10%FS4 = power-on zero-point +20%FS5 = power-on zero-point +50%FS6 = power-on zero-point +100%FS7 = no limitation
P140-22 Weight signal within poweron zero point range0 = current weight1 = calibration zero-point2 = switch off zero point and power off tareweight as current tare weight
SubMenu1Option Default Parameter DescriptionComment
P15Weight signal not withinpower on zero point range0 = current weight1 = calibration zero-point2 = switch-off zero-point3 = continuously display 0----
P150-BBZero tracking rangeChoose the zero tracking range as per the stability of weighing system accuracy.0 = 0d, no tracking1 = +0.25d2 = +0.5d3 = +1d4 = +1.5d5 = +2d6 = +3d7 = +4d8 = +5d
P17Data filter intensity The larger the digit data filter intensity isis stronger and the speed of data updating is lower.0 = very weak1 = weak2 = middle3 = strong
P18Range of weight stabilitychecking0 = +0.5d1 = ±1d2 = +2d3 = +3d4 = +4d5 = +5d6 = +6d7 = +7d8 = +8d9 = +9d
P190-9/Overload limit range0 = FS+0d1 = FS+9d2 = 101%FS3 = 102%FS4 = 105%FS5 = 110%FS6 = 120%FS7 = 150%FS8 = 200%FS9 = no limitation

Table 4.3 Calibration Resolution (P7)

Parameter ChoiceCalibration ResolutionParameter ChoiceCalibration Resolution
00 500 167500
01 600 178000
02 750 1810000
03 800 1912000
04 1000 2015000
05 1200 2120000
06 1500 2225000
07 2000 2330000
08 2400 2435000
09 2500 2540000
10 3000 2650000
11 3500 2760000
12 4000 2870000
13 5000 2975000
14 6000 3080000
15 7000 31100000

Table 4.4 Calibration Units (kg) per P8, P9 and P10

Calibration Division ValueDisplay Division Value
kg lb lb:oz
0.0001kg 0.0001kg 0.0002lb Not available
0.001kg 0.001kg 0.002lb Not available
0.01kg 0.01kg 0.02lb 0.5oz
0.1kg0.1kg0.2lb Not available
1kg 1kg2lb Not available
10kg10kg20 lb Not available
0.0002kg 0.0002kg 0.0005 lbNot available
0.002kg 0.002kg 0.005 lb0.1 oz
0.02kg 0.02kg 0.05 lb1 oz
0.2kg0.2kg0.5 lb Not available
2kg 2kg5 lbNot available
20kg20kg50 lb Not available
0.0005kg 0.0005kg 0.001 lb Not available
kglb lb:oz
0.005kg 0.005kg 0.01 lb 0.2 oz
0.05kg 0.05kg 0.1 lb 2oz
0.5kg 0.5kg 1 lb Not available
5kg 5kg 10 lb Not available
50kg 50kg Not availableNot available

Table 4.5 Calibration Units (lb) per P8, P9 and P10

Calibration Division ValueDisplay Division Value
kglb lb:oz
0.0001lbNot available0.0001lbNot available
0.001 lb 0.0005 kg 0.001 lbNot available
0.01 lb 0.005 kg 0.01 lb0.2 oz
0.1 lb 0.05 kg 0.1 lb2 oz
1 lb 0.5 kg 1 lb Not available
10 lb 5 kg 10 lb Not available
0.0002 lb0.0001 kg0.0002 lbNot available
0.002 lb 0.001 kg 0.002 lbNot available
0.02 lb 0.01 kg 0.02 lb 0.5 oz
0.2 lb0.1 kg0.2 lbNot available
2 lb1 kg2 lb Not available
20 lb10 kg20 lb Not available
0.0005 lb0.0002 kg0.0005 lbNot available
0.005 lb 0.002 kg 0.005 lb0.1 oz
0.05 lb0.02 kg0.05 lb 1 oz
0.5 lb0.2 kg0.5 lbNot available
5 lb2 kg5 lb Not available
50 lb20 kg50 lb Not available

Table 4.6 Capacity Unit is kg or lb (Count by 1)

Resolution set by P7Division set by P8 (P8 = 0) and P9
0.00010.0010.010.1110
500 0.0500.5005.005005005000
600 0.0600.6006.006006006000
750 0.0750.7507.507507507500
800 0.0800.8008.008008008000
1000 0.1001.00010.001000100010000
1200 0.1201.20012.001200120012000
1500 0.1501.50015.001500150015000
2000 0.2002.00020.002000200020000
2400 0.2402.40024.002400240024000
2500 0.2502.50025.002500250025000
3000 0.3003.00030.003000300030000
3500 0.3503.50035.003500350035000
4000 0.4004.00040.004000400040000
5000 0.5005.00050.005000500050000
6000 0.6006.00060.006000600060000
7000 0.7007.00070.007000700070000
7500 0.7507.50075.007500750075000
8000 0.8008.00080.008000800080000
100001.000010.000100.00
120001.200012.000120.00
150001.500015.000150.00
200002.000020.000200.00
250002.500025.000250.00
300003.000030.000300.00
350003.500035.000350.00
400004.000040.000400.00
500005.000050.000500.00
600006.000060.000600.00
700007.000070.000700.00
750007.500075.000750.00
800008.000080.000800.00
10000010.0000100.0001000.00

Table 4.7 Capacity Unit is kg or lb (Count by 2)

Resolution set by P7Division set by \( \mathrm{P}8\left( {\mathrm{P}8 = 1}\right) \) and \( \mathrm{P}9 \)
0.0002 0.002 0.02 0.2 220
500 0.10001.000 10.00100.0 1000 10000
600 0.12001.200 12.00120.0 1200 12000
750 0.15001.500 15.00150.0 1500 15000
800 0.16001.600 16.00160.0 1600 16000
1000 0.20002.000 20.00200.0 2000 20000
1200 0.24002.400 24.00240.0 2400 24000
1500 0.30003.000 30.00300.0 3000 30000
2000 0.40004.000 40.00400.0 4000 40000
2400 0.48004.800 48.00480.0 4800 48000
2500 0.50005.000 50.00500.0 5000 50000
3000 0.60006.000 60.00600.0 6000 60000
3500 0.70007.000 70.00700.0 7000 70000
4000 0.80008.000 80.00800.0 8000 80000
5000 1.000010.000 100.01000.0 10000 100000
6000 1.200012.000 120.01200.0 12000 120000
7000 1.400014.000 140.01400.0 14000 140000
7500 1.500015.000 150.01500.0 15000 150000
8000 1.600016.000 160.01600.0 16000 160000
100002.000020.000200.0020000.020000200000
120002.4000 24.000240.0 2400 24000 240000
150003.0000 30.000300.0 3000 30000 30000
200004.0000 40.000400.0 40000 40000
250005.0000 50.000500.0 50000 50000
300006.0000 60.000600.0 60000 60000
350007.0000 70.000700.0 70000 70000
400008.0000 80.000800.0 80000 80000
5000010.0000100.0001000.0010000.0100000100000
6000012.0000120.0001200.0012000.0120000120000
7000014.0000140.0001400.0014000.0140000140000
7500015.0000150.0001500.0015000.0150000150000
8000016.0000160.0001600.0016000.0160000160000
10000020.0000200.0002000.0020000.0200000200000

Table 4.8 Capacity Unit is kg or lb (Count by 5)

Resolution set by P7Division set by P8 (P8 = 2) and P9
0.0005 0.005 0.05 0.5 50
500 0.25002.5000 25.00250.02500 2500
600 0.30003.000 30.00300.03000 30000
750 0.37503.750 37.50375.03750 37500
800 0.40004.000 40.00400.04000 40000
1000 0.50005.000 50.00500.05000 50000
1200 0.60006.000 60.00600.06000 60000
1500 0.75007.500 75.00750.07500 75000
2000 1.000010.000 100.001000.0100000
2400 1.200012.000 120.001200.0120000
2500 1.250012.500 125.001250.0125000
3000 1.500015.000 150.001500.0150000
3500 1.750017.500 175.0017500175000175000
4000 2.000020.000 200.0020000200000200000
5000 2.500025.000 250.002500.0250000
6000 3.000030.000 300.003000.0300000
7000 3.500035.00 350.003500.035000 350000
7500 375.0037.500 375.003750.0375000
8000 4.000040.000 400.004000.0400000
10000 5.000050.000 500.005000.0500000
12000 6.000060.000 600.006000.0600000
15000 7.500075.000 750.007500.0750000
2000010.0000100.0001000.0010000.010000001000000
2500012.5000125.0001250.0012500.012500001250000
3000015.0000150.0001500.0015000.015000001500000
3500017.5000175.0001750.0017500.017500001750000
4000020.0000200.0002000.0020000.020000002000000
5000025.0000250.0002500.0025000.025000002500000
6000030.0000300.0003000.0030000.03000003000000
7000035.0000350.0003500.0035000.03500003500000
7500037.5000375.0003750.0037500.03750003750000
8000040.0000400.0004000.0040000.04000004000000
10000050.0000500.0005000.0050000.05000005000000

4.5 Exit the Setup Menu

  1. Power off the indicator by pressing and holding the [ZERO/ON/OFF] key.
  2. Move the slide switch on the rear cover back to the left and replace the metal protective plate. Refer to section 4.2 for location.
  3. Turn the indicator back on by pressing the [ON] key. The display will go through a digit check, then settle into Normal Operating mode. All front panel keys will now return to their normal mode of operation.

5 Calibration

The config/cal switch must be set to the ON position in order to calibrate the indicator.

Brecknell MS140300 - Calibration - 1

NOTE: More than 25% of the full scale weight is needed for calibration.

5.1 Calibration Mode

Before calibration, prepare the standard weight (more than 25% of full scale).

  1. Remove any weight on the scale.
  2. When in normal weighing mode, press and hold down [TARE] and [ON/OFF/ ZERO] keys to enter calibration mode.
  3. When the indicator displays CRL^- , the scale is ready for calibration. Press [TARE] to confirm and go to next step or press [ON/OFF/ZERO] to exit the calibration mode.
  4. The indicator will display RP - - .
    4a. This means the following data is the full capacity according to your setting of display resolution (P7), display division value (P8), location of decimal point in calibration unit (P9) and capacity unit in calibration (P10).
    4b. If the setting of full scale is more than 199999 (regardless of decimal point and weight unit), the full scale capacity will be shown by first four digits and last four digits: Hxxx and Lxxx.
  5. Press [TARE] to go to next step directly, press [ON/OFF/ZERO] to exit the calibration mode or after a few seconds it will automatically to next step.
  6. The scale will automatically display the setting of division. Firstly it will display d . and then the data according to your setting of P8, P9 and P10. Refer to Table 5.1 to choose a division size.

Table 5.1 Divison Sizes

0.0001 kg/lb 0.002 kg/lb 0.0005 kg/lb
0.001 kg/lb 0.002 kg/lb 0.005 kg/lb
0.01 kg/lb 0.02 kg/lb 0.05 kg/lb
0.1 kg/lb 0.2 kg/lb 0.5 kg/lb
1 kg/lb 2 kg/lb 5 kg/lb
10 kg/lb 20 kg/lb 50 kg/lb
  1. Press [TARE] to go to next step directly, press [ON/OFF/ZERO] to exit the calibration mode or after a few seconds it will automatically to next step.
  2. CRL.PQ is displayed. This is the scale zero point. Remove weight from scale and press the [TARE] key.

  3. CRL.P will be displayed. This is the second calibration point. The default standard weight is 50% of full scale. Load weight from 12.5% to 100% of full scale.

  4. Use the [PRINT/HOLD] and [UNIT] keys to input the loaded weight.
    10a. If the input data is larger than 199999 it will be shown by first four digits and last four digits: Hxxx and Lxxx.
    10b. If the triangular symbol on the left bottom of LCD window appears, it means that the digit being changed is the displayed most significant bit which can only be 0/blank or 1.
  5. Press the [TARE] key to confirm your setting and the indicator will flash the input standard weight. Wait till the scale comes steady and input A/D data as per the standard weight.
  6. [HL]P will be displayed. This is the third calibration point. When xxxx kg (or lb) is displayed (100% FS is default), you can press [ON/OFF/ZERO] to exit the calibration mode or place a standard weight (must be in the range of 25%-100% FS, and equal or larger than that for the second calibration point; this is also the range of your input number) on the scale.
  7. Use the [PRINT/HOLD] and [UNIT] keys to input the loaded weight.
  8. Use [TARE] key to confirm the standard weight and input number are correct.
    14a. If the calibration weight for third point is same with that for second point and the calibration weight is more than 25% FS, input the standard calibration weight same as second point calibration and press [TARE] key to confirm the setting. The indicator will flash the input weight. If the indicator gets reasonable data (the input weight is correct, and the calibration weight of third calibration is more than equal to the calibration weight of second calibration), it will go to next step.
  9. [CALPO] will be displayed again. Press [TARE]. The display will blink. If the calibration is good the data will be saved. The display will perform an auto reset and return to the working mode.
    15a. If an error occurs during calibration, CRL.Er will be displayed. Start from step 1 and recalibrate.

5.2 Display ADC Code or Working Voltage Value

In this mode, you can examine the stability of weighing system and increment the ADC output code corresponding to the loaded weight.

Note:

  • The increment of ADC code for full scale weight must be larger or equal to 2 times of selected display division. Otherwise, the calibration cannot be properly completed.

e.g. The display division is 0.1kg . Load 100kg standard weight on the platform, the increment of ADC code should be at least more than 2× 100kg / 0.1kg = 2× 1000 = 2000 . In this case, the scale can be calibrated. Otherwise, smaller division needs to be chosen.

The data should be stable. Otherwise, the calibration cannot be properly completed.

In this mode, the proper ADC data at zero point can be viewed by examining the A/D data for loaded weight.

If the ADC increase for full capacity is NFS, the power-on zero range is set to Zp% FS (P12 setting) and zero key range is set to Zk% FS (P13 setting). Then proper ADC data of zero point is larger than (Zp% + Zk%) x NFS.

  1. From the weigh mode press and hold the [ON/OFF/ZERO] and [HOLD/PRINT] key until is shown. The indicator will show the firmware version.
  2. ADC increase for full capacity (NFS) can be made by loading the weight (W) on the platform and the ADC increase for W weight is Nw. The ADC increase for full capacity WFS is (NFS) = (Nw)x (WFS)/W.
    2a. It is possible a negative value will be displayed due to an error from the loadcell(s) or the position of the zero-point potentiometer on PCB. However, the software only deals with positive values. In this case the position of zero-point potentiometer will need to be adjusted to make the ADC data will be positive value and larger than (Zp% + Zk%)× NFS. Normally the indicator is factory-calibrated and adjustment is not necessary. Refer to Figure 5.1 for the zero point potentiometer location. Decrease ADC data by rotating clockwise, and increase ADC data by rotating counter-clockwise.

Brecknell MS140300 - Display ADC Code or Working Voltage Value - 1
Figure 5.1 Zero Point Potentiometer Location

  1. Press the [UNIT] key to select the ADC weight inner code or input the inner working voltage value. When Ux.xx is displayed, the displayed digits are the voltage value. The proper working power voltage is between 5V - 8V.
  2. Press the [TARE] key to display filtered or unfiltered weigh A/D data. When is on the data is filtered.
  3. Press the [ON/OFF/ZERO] key to exit this mode and return to the normal weighing mode.

6 Serial Communication

6.1 Com Port 1

COM1 is a RS-232 bi-directional port. Communication wires are connected to the RS-232 connector using TXD1, RXD1 and GND.

Table 6.1 PS-500 Com1 Port

Designation Description Pin
TXD transmit 2
RXD receive 3
GND ground/common 5
DSRdata set ready4
DTRdata terminal ready6
CTSclear to send7
RTSreturn to send8
no connection no connection1
no connection no connection9

NOTE: The indicator DB9 female connector has pin 4 and pin 6 shorted together. Pin 7 and pin 8 are also shorted together.

6.2 Protocol

The baud rate and byte format is fixed by P5 and P6 settings.

Responses to serial commands will be immediate or within one weight measure cycle of the scale. One second should be adequate for use as a time-out value by remote (controlling) device.

6.3 Transaction String

The length of the weight field will be 7 digits. One digit for minus sign, one for decimal point, 2 for unit of measure.

  • If the unit is lb:oz, another two for Ib and on for a space after Ib. Units of measure abbreviations are always lower case.
  • If the weight is over capacity, the scale will return nine "^" characters (the field of minus sign, decimal point, weight data is filled by "^").
  • If the weight is under capacity, it will return nine "" characters (the field of minus sign, decimal point, and weight data is filled by " ").
  • If the zero point is error, it will return nine "-" characters. The character will be "-" for negative weight or a space character for positive weight. Minus sign follow after the first digit.

Useless leading 0 before digits is suppressed.

Table 6.2 Symbols Used
Table 6.3 Bit Definition of H1H2H_3

<LF> Line Feed character (hex 0AH)
<CR> Carriage Return character (hex 0DH)
<ETX> End of Text character (hex 03H)
<SP> Space (hex 20H)
H1H2H3Four current status bytes
<P> Polarity character:"?" or " "
W1--W7Reading data, 1-7 bytes (seven digits)
<DP> Decimal point
U1U2Measure units, kg, lb, lb:oz, % or pcs; 2-5 bytes
BitByte 1 (H1) Byte2 (H 2) Byte3 (H 3)
0 0 = stable 0 = not under capacity 00 = compare disable
1 = not stable 1 = under capacity01 = lower limit10 = ok11= upper limit
1 0 = not at zero point 0 = not over capacity
1 = at zero point1 = over capacity
2 always 0 always 0 0 = gross weight
1 = net weight
3 0 = eeprom OK always 0 always 0
1 = eeprom error
4 always 1 always 1 always 1
5 always 1 always 1 always 1
6 always 0 always 1 always 0
7 parity parity parity

Command: W (57h 0dh), request current reading

Response
<LF>xxxxxxxxU1U2<CR><LF>H1H2H3<CR<ETX>---over capacity
<LF>__________U1U2<CR><LF>H1H2H3<CR<ETX>---under capacity
<LF>--------- U1U2<CR><LF>H1H2H3<CR<ETX>---zero-point error
Note: If the weigh unit is lb:oz, U1U2 oz in above responses
<LF><P=W1W2W3W4W5W6<DP=W7U1U2<CR><LF>H1H2H3<CR<ETX>---scale is stable and the current weigh unit is kg or lb. With or without decimal point and the position is as per the P9 setting.
<LF><P=W1W2W3lb<SP=W4W5lb<SP=W6W7oz<CR>H1H2H3<CR<ETX>
<LF><P=W1W2W3lb<SP=W4lb<SP=W5W6<DP=W7oz<CR>H1H2H3<CR<ETX>

Command: S (53h 0dh), request current status

Response
<H1>H2H3<CR><ETX>

Command: Z (5ah 0dh)

Response
Zero function is activated (simulate ZERO key) and it returns to current scale status.
<H1H2H3<CR><ETX>

If ZERO function cannot be activated, it will return to current scale status.

Command:T(54h0dh)

Response
TARE function is activated (simulate TARE key), and then returns scale status.
<H1>H2H3<CR><ETX>

If TARE function cannot be activated, it will return to current scale status.

Command: U (55h 0dh)

Response
Changes units of measure (simulate UNIT key) and return scale status with new units, The new measure unit should be allowed to use
<LF> U1U2 <CR><LF> H1H2H3 <CR><ETX>

Command:L(4ch 0dh)

Response
If Hold function can be activated, it will enable/disable hold function (simulate HOLD key) and returns scale status.
<H1H2H3<CR><ETX>

Command:X (58h 0dh)

Response
Power off the scale, just like press down the ON/OFF key to turn off the scale.

Command: all others
Table 6.4 Summary of Command and Response:

Response
Unrecognized command
<LF?> <CR></ETX>
Command Response
ASCII HEX
W<CR> 57 0d Read scale weight: <LF>xxxxxxxxU1U2 U3U4U5<CR><LF>H1H2H3 H4<CR><ETX>---over capacity ?<LF> U1U2U3U4U5<CR><LF>H1H2H3 H4<CR><ETX>---under capacity ?<LF>-------- U1U2 U3 U4U5<CR><LF>H1H2H3 H4<CR><ETX>---zero-point error ?<LF><p=W1W2W3W4W5<dp=W6U1U2U3U4U5<CR><LF>H1H2H3H4<CR><ETX>---normal data
S<CR> 53 0d <LF>H1H2H3H4<CR><ETX>; read scale status
Z<CR> 5a 0d <LF>H1H2H3H4<CR><ETX>; simulate ZERO key
T<CR> 54 0d <LF>H1H2H3H4<CR><ETX>; simulate TARE key
U<CR> 55 0d <LF> U1U2 U3 U4U5<CR><LF>H1H2H3H4<CR><ETX>; simulate UNIT key
L<CR> 4c 0d <LF>H1H2H3H4<CR><ETX>; simulate HOLD key
X<CR> 58 0d power off the scale, simulate OFF key
others <LF>? <CR><ETX>

7 Troubleshooting

This chapter gives explanations on commonly seen errors, display characters and display symbols.

7.1 Display Characters

ASCIIDisplay CharacterASCIIDisplay CharacterASCIIDisplay Character
08.A8.N8.
18.B8.O8.
28.C8.P8.
38.D8.Q8.
48.E8.R8.
58.F8.S8.
68.G8.T8.
78.H8.U8.
88.I8.V8.
98.J8.W8.
K8.X8.
L8.Y8.
M8.Z8.
Symbol Description
U----Zero is over the setting range
U----Zero point is below the setting range
Hd----Signal to ADC is over maximum range
Hd----Signal to ADC is below minimum range
--- Weight is over upper limitation or display data is over limitation
--- Weight is below lower limitation
EEPROM can't be accessed.
EEPROM The parameters are not the same as the backup data
EEPROM The setting parameter(s) is not in normal range
L bHE Battery voltage is below 4.7V
LHP.- - The setting full capacity will be displayed
LHL.PIT Calibration on point (x)
CALER Calibration error, maybe input data or loaded weight is incorrect, unstable, non-linear
d.- - The division will be displayed
PIT The x = parameter, y = parameter choice

7.3 Error Messages and Troubleshooting

Symptom Probable Cause Remedy
Hd- - - - - - Loadcell wires to indicator are incorrectly connected, shorted, opened, ADC or loadcell(s) are damaged.Make sure wires are ok and correctly connected. Replace loadcell or ADC chip, Service required.
D - - - - - -Weight reading exceeds Power On Zero limit.Make sure scale platform is empty. Perform zero calibration.
D - - - - - -Weight reading below Power On Zero limit.Install platform on scale. Perform zero calibration.
- - - - - - Weight reading exceeds Overload limit or the weight value cannot be displayed in the current unit of measure because it exceeds 6 digits.Reduce load on scale until weight value can be displayed. Use a more appropriate unit of measure.
- - - - - -Weight reading below under load limit.Install platform on scale. Perform zero calibration.
EEPE2Parameter is not correctly set.Re-set parameter
CRLFerCalibration error. Input data or loaded weight is too small, too big, unstable, non-linear.Input correct data, load correct weight onto platform, Service required.
Will not turn on.Power cord not plugged in or properly connected. Power outlet not supplying electricity. Battery discharged. Other failure.Check power cord connections. Make sure power cord is plugged into the power outlet. Check power source. Replace batteries. Service required.
Cannot zero the display or will not zero when turned on.Load on scale exceeds allowable limits. Load on scale is not stable. Load cell damage.Remove load on scale. Wait for load to become stable. Service required.

Brecknell

Brecknell USA

1000 Armstrong Dr.

Fairmont MN 56031

Tel:507-238-8702

Fax:507-238-8271

Email: sales@brecknellscales.com

http://www.brecknellbases.com

Brecknell UK

Foundry Lane,

Smethwick, West Midlands,

England B66 2LP

Tel:+44 (0) 8452 46 6717

Fax:+44 (0) 8452 46 6718

Email:

sales@brecknellbases.co.uk

http://www.brecknellbases.com

MS140/300

Smethwick, West Midlands,

England B66 2LP

Tél.: +44 (0) 8452 46 6717

Telc.:+44(0)8452466718

Courriel: sales@brecknellscales.co.uk

http://www.brecknellbases.com

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Brecknell

Model : MS140300

Category : Scale