BEURER BM 55 - Sphygmomanometer

BM 55 - Sphygmomanometer BEURER - Free user manual and instructions

Find the device manual for free BM 55 BEURER in PDF.

📄 152 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice BEURER BM 55 - page 21
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Arm electronic blood pressure monitor
Brand Beurer
Model BM 55
Measurement method Oscillometric
Measurement zone Upper arm (circumference 22-36 cm)
Measurement range Pressure: 30-280 mmHg, Pulse: 40-200 beats/min
Accuracy Pressure: ±3 mmHg, Pulse: ±5%
Display Large LCD screen with date and time
Memory 2 x 60 measurements (for 2 users)
Main functions Arrhythmia detection, WHO classification, average calculation
Power supply 4 AAA batteries (1.5 V) or mains adapter (optional)
Dimensions (L x W x H) Approx. 150 x 100 x 50 mm
Weight Approx. 300 g (without batteries)
Battery life Approx. 200 measurements with new batteries
Care and cleaning Clean with a soft dry cloth; do not immerse
Safety Automatic shut-off, self-inflating cuff, clinical validation
Warranty 5 years (according to warranty conditions)
Included accessories Cuff, 4 AAA batteries, user manual

Frequently Asked Questions - BM 55 BEURER

How to use my Beurer BM 55 blood pressure monitor correctly?
Sit calmly, place your arm at heart level, put the cuff around your upper arm (22-36 cm) and press 'Start'. Stay still and do not talk during the measurement.
What does the flashing heart symbol or arrhythmia icon mean?
The symbol indicates a possible cardiac arrhythmia. If this symbol appears frequently, consult a doctor. It is not a diagnosis, but an indication.
Why does my blood pressure monitor show an error code (e.g. Error 2)?
Error 2 means the cuff is not correctly positioned. Check that the cuff is properly fitted and the tube is connected. See the manual for other codes.
How to clean and maintain my Beurer BM 55 blood pressure monitor?
Clean the device and cuff with a soft, dry cloth. Do not use abrasive products. Never immerse the device in water.
Can I use the device with a mains adapter?
Yes, you can use an optional mains adapter (6V DC, 600 mA, jack plug). Use only accessories recommended by Beurer.
How to interpret the results (WHO classification)?
The device displays a color code (green, yellow, orange, red) according to the WHO classification. Green: normal, Yellow: borderline high, Orange: moderate hypertension, Red: severe hypertension.
What to do if the cuff does not inflate properly?
Check that the tube is properly connected to the device and that the cuff is wrapped without folds. If the problem persists, replace the batteries or contact customer service.
How to delete the measurement memory?
Press the 'M' button to access the memory. Hold the 'M' button for 3 seconds to delete the data of the selected user.
What is the warranty of the Beurer BM 55?
The warranty is 5 years from the date of purchase, subject to proper use and presentation of the receipt.
Is the blood pressure monitor suitable for people with arrhythmia?
Yes, the BM 55 has an arrhythmia detector that identifies irregular heartbeats. It is suitable but does not replace medical advice.

User questions about BM 55 BEURER

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Sphygmomanometer in PDF format for free! Find your manual BM 55 - BEURER and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. BM 55 by BEURER.

USER MANUAL BM 55 BEURER

EN Blood pressure monitor Instructions for use....21

4 × 1,5V AAA (LR03)

BEURER BM 55 - 1

MIN OK MAX

Pure diagram of pipe connection with arrow indicating direction (no text or symbols)

Side profile illustration of a person using a medical device (no text or symbols)

Silhouette of a person sitting at a desk with crossed diagonal lines indicating crossed or no movement (no text or symbols)

Quelle: WHO, 1999 (World Health Organization)

Read these instructions for use carefully and keep them for later use, be sure to make them accessible to other users and observe the information they contain.

Contents

  1. Included in delivery....21
  2. Signs and symbols 22
  3. Intended use....22
  4. Warnings and safety notes....23
  5. Device description....25
  6. Initial use ......26
  7. Usage 28

Dear Customer,

thank you for choosing a product from our range. Our name stands for high-quality, thoroughly tested products for applications in the areas of heat, weight, blood pressure, body temperature, pulse, gentle therapy, massage, beauty and air. Please read these instructions for use carefully and keep them for later use, be sure to make them accessible to other users and observe the information they contain.

Best regards,

Your Beurer Team

  1. Cleaning and maintenance....33
  2. Accessories and replacement parts....33
  3. What if there are problems? 34
  4. Disposal 35
  5. Specifications....35
  6. Mains part.....36
  7. Warranty/service....37

1. Included in delivery

Check that the exterior of the cardboard delivery packaging is intact and make sure that all contents are present. Before use, ensure that there is no visible damage to the device or accessories and that all packaging material has been removed. If you have any doubts, do not use the device and contact your retailer or the specified Customer Service address.

• Blood pressure monitor
- Upper arm cuff
- 4 x LR03 AAA batteries
- Storage bag
- USB cable
- Instructions for use

2. Signs and symbols

The following symbols are used on the device, in these instructions for use, on the packaging and on the type plate for the device:

BEURER BM 55 - Signs and symbols - 1WARNINGWarning notice indicating a risk of injury or damage to health
BEURER BM 55 - Signs and symbols - 2IMPORTANTSafety note indicating possible damage to the device/accessory
BEURER BM 55 - Signs and symbols - 3NoteNote on important information
BEURER BM 55 - Signs and symbols - 4Observe the instructions for use
BEURER BM 55 - Signs and symbols - 5Application part, type BF
BEURER BM 55 - Signs and symbols - 6Direct current
BEURER BM 55 - Signs and symbols - 7Disposal in accordance with the Waste Electrical and Electronic Equipment EC Directive – WEEE
BEURER BM 55 - Signs and symbols - 8Do not dispose of batteries containing hazardous substances with household waste
BEURER BM 55 - Signs and symbols - 9Dispose of packaging in an environmentally friendly manner
BEURER BM 55 - Signs and symbols - 10Manufacturer
BEURER BM 55 - Signs and symbols - 11Temperature limitThe temperature limit values to which the medical device can safely be exposed are indicated.
BEURER BM 55 - Signs and symbols - 12Humidity, limitIndicates the humidity range to which the medical device can safely be exposed.
BEURER BM 55 - Signs and symbols - 13Atmospheric pressure, limitIndicates the range of atmospheric pressures to which the medical device can be safely exposed
BEURER BM 55 - Signs and symbols - 14Serial number
BEURER BM 55 - Signs and symbols - 15Device protected against foreign objects ≥ 12.5 mm and against water dripping at an angle
BEURER BM 55 - Signs and symbols - 16483This product satisfies the requirements of the applicable European and national directives.

3. Intended use

Intended use

The blood pressure monitor is intended for the fully automatic, non-invasive measurement of arterial blood pressure and pulse values on the upper arm.

Target group

It is designed for self-measurement by adults in the home environment and is suitable for users whose upper arm circumference is within the range printed on the cuff.

Indication/clinical benefits

The user can record their blood pressure and pulse values quickly and easily using the device. The recorded values are classified according to internationally applicable guidelines and evaluated graphically. Furthermore, the device can detect any irregular heart beats that occur during measurement and inform the user via a symbol in the display. The device saves the recorded measurements and can also output average values of previous measurements. The recorded data can provide healthcare service providers with support during the diagnosis and treatment of blood pressure problems, and therefore plays a part in the long-term monitoring of the user's health.

4. Warnings and safety notes

BEURER BM 55 - Warnings and safety notes - 1

Contraindications

  • Do not use the blood pressure monitor on newborns, children or pets.
  • People with restricted physical, sensory or mental skills should be supervised by a person responsible for their safety and receive instructions from this person on how to use the device.
  • If you have any of the following conditions, it is essential you consult your doctor before using the device: cardiac arrhythmia, circulatory problems, diabetes, pregnancy, pre-eclampsia, hypotension, chills, shaking.

  • People with pacemakers or other electrical implants should consult their doctor before using the device.

  • The blood pressure monitor must not be used in connection with a high-frequency surgical unit.
  • Do not use the cuff on people who have undergone a mastectomy.
  • Do not place the cuff over wounds as this may cause further injury.
  • Make sure that the cuff is not placed on an arm in which the arteries or veins are undergoing medical treatment, e.g. intravascular access or intravascular therapy, or an arteriovenous (AV) shunt.

BEURER BM 55 - Contraindications - 1

General warnings

  • The measured values taken by you are for your information only – they are no substitute for a medical examination. Discuss the measured values with your doctor and never make your own medical decisions based on them (e.g. regarding dosages of medicines).
  • The device is only intended for the purpose described in these instructions for use. The manufacturer is not liable for damage resulting from improper or incorrect use.
  • Using the blood pressure monitor outside your home environment or whilst on the move (e.g. whilst travelling in a car, ambulance or helicopter, or whilst undertaking physical activity such as playing sport) can influence the measurement accuracy and cause incorrect measurements.
  • Cardiovascular diseases may lead to incorrect measurements or have a detrimental effect on measurement accuracy.

  • Do not use the device at the same time as other medical electrical devices (ME equipment). This could lead to a malfunction of the device and/or an inaccurate measurement.

  • Do not use the device outside of the specified storage and operating conditions. This could lead to incorrect measurements.
  • Only use the cuffs included in delivery or cuffs described in these instructions for use for the device. Using another cuff may lead to measurement inaccuracies.
  • Please note that when inflating the cuff, the functions of the limb in question may be impaired.
  • Do not perform measurements more frequently than necessary. Due to the restriction of blood flow, some bruising may occur.
  • During the blood pressure measurement, the blood circulation must not be stopped for an unnecessarily long time. If the device malfunctions remove the cuff from the arm.
  • Place the cuff on your upper arm only. Do not place the cuff on other parts of the body.
  • Small parts may present a choking hazard for small children if swallowed. They should therefore always be supervised.

BEURER BM 55 - General warnings - 1

General precautions

  • The blood pressure monitor is made from precision and electronic components. The accuracy of the measurements and service life of the device depend on its careful handling.
  • Protect the device from impacts, humidity, dirt, marked temperature fluctuations and direct sunlight.

  • Ensure the device is at room temperature before measuring. If the measuring device has been stored close to the maximum or minimum storage and transport temperatures and is placed in an environment with a temperature of 20 °C, it is recommended that you wait approx. 2 hours before using the measuring device.

  • Do not drop the device.
  • Do not use the device in the vicinity of strong electromagnetic fields and keep it away from radio systems or mobile telephones.
  • We recommend that the batteries be removed if the device is not to be used for a prolonged period of time.
  • Avoid any mechanical restriction, compression or bending of the cuff line.
  • The temperature of the cuff can reach up to 42.1 °C during normal operation at an ambient temperature of 40 °C.

Measures for handling batteries

BEURER BM 55 - Measures for handling batteries - 1

  • If your skin or eyes come into contact with battery fluid, rinse the affected areas with water and seek medical assistance.
  • Choking hazard! Small children may swallow and choke on batteries. Therefore, store batteries out of the reach of small children.
  • Risk of explosion! Do not throw batteries into a fire.
  • If a battery has leaked, put on protective gloves and clean the battery compartment with a dry cloth.
  • Do not disassemble, open or crush the batteries.

BEURER BM 55 - Measures for handling batteries - 2

  • Observe the plus (+) and minus (-) polarity signs.
  • Protect the batteries from excessive heat.
  • Do not charge or short-circuit batteries.
  • If the device is not to be used for a relatively long period, take the batteries out of the battery compartment.
  • Use identical or equivalent battery types only.
    • Always replace all batteries at the same time.
  • Do not use rechargeable batteries!

BEURER BM 55 - Measures for handling batteries - 3

Notes on electromagnetic compatibility

  • The device is suitable for use in all environments listed in these instructions for use, including domestic environments.
  • The use of the device may be limited in the presence of electromagnetic disturbances. This could result in issues such as error messages or the failure of the display/device.
  • Avoid using this device directly next to other devices or stacked on top of other devices, as this could lead to faulty operation. If, however, it is necessary to use the device in the manner stated, this device as well as the other devices must be monitored to ensure they are working properly.
  • The use of accessories other than those specified or provided by the manufacturer of this device can lead to an increase in electromagnetic emissions or a decrease in the device's electromagnetic immunity; this can result in faulty operation.
  • Failure to comply with the above can impair the performance of the device.

5. Device description

Blood pressure monitor and cuff

1 2 3 beurer 4 10 9 5 LRA M1 M2 6 7 8

  1. Cuff
  2. Cuff line
  3. Cuff connector
  4. Connection for mains part and USB interface
  5. Display
  6. Memory buttons M1/M2
  7. Start/stop button ⓘ
  8. Resting indicator display
  9. Risk indicator
  10. Connection for cuff connector (left-hand side)

Display

  1. Date / time
  2. Systolic pressure
  3. Diastolic pressure
  4. Pulse value
  5. Pulse symbol ♥
  6. Release air (arrow)
  7. Number of memory space/memory display average value (A), morning (R), evening (P)
  8. Battery replacement
    symbol
  9. User memory
  10. Risk indicator

  11. Cardiac arrhythmia symbol

11 AM PM 28:88 28/88 10 9 388 1 388 8 388 M 88 188 7 6 5

Computer interface

The blood pressure monitor also allows you to transfer your measured values to the computer.

For this, you will need a USB cable (included in delivery) and the "beurer HealthManager" computer software.

The software can be downloaded free of charge from the download area under Service at www.beurer.com.

System requirements for the "beurer HealthManager" computer software

  • Windows 7 SP1 or higher
    • USB 2.0 (Type A) or higher

6. Initial use

Inserting the batteries

  • Remove the battery compartment lid on the rear of the device.
  • Insert four 1.5V AAA(alkaline type LR03) batteries. Make sure that the batteries are inserted the correct way round. Do not use rechargeable batteries.
  • Close the battery compartment lid again carefully.

4 x 1,5V AAA (LR03)

All display elements are briefly displayed, 24 h flashes in the display. Set the date and time as described below.

If the battery replacement symbol ☐ is permanently displayed, you can no longer perform any measurements and must replace all batteries. Once the batteries have been removed from the device, the date and time must be set again. Any saved measurements are retained.

Making settings

This menu allows you to set the following functions, one after another.

BEURER BM 55 - Making settings - 1

It is essential to set the date and time. Otherwise, you will not be able to save your measured values correctly with a date and time and access them again later.

If you press and hold the M1 or M2 memory button, you can set the values more quickly.

  • Press and hold the Start/stop button for 5 seconds.
  • Choose the desired hour format with the M1/M2 memory buttons and confirm with the Start/stop button Ⓞ

BEURER BM 55 - Making settings - 2

The year flashes on the display.

- Choose the desired year with the M1/M2 memory buttons and confirm with the Start/stop button Ⓞ

The month flashes on the display.

- Choose the desired month with the M1/M2 memory buttons and confirm with the Start/stop button ⓘ

The day flashes on the display.

- Choose the desired day with the M1/M2 memory buttons and confirm with the Start/stop button Ⓞ

i If the hour format is set as play sequence is reversed.

BEURER BM 55 - Making settings - 3

BEURER BM 55 - Making settings - 4

BEURER BM 55 - Making settings - 5

, the day/month dis-

Time

The hour flashes on the display.

  • Choose the desired hour with the M1/M2 memory buttons and confirm with the Start/stop button ⏻
    The minute flashes on the display.
  • Choose the desired minute with the M1/M2 memory buttons and confirm with the Start/stop button ⏻

BEURER BM 55 - Time - 1

BEURER BM 55 - Time - 2

Operation with the mains part

You can also operate this device with a mains part.

When doing so, there must not be any batteries in the battery compartment. The mains part can be obtained from specialist retailers or from the service address using order number 071.95.

  • To prevent possible damage to the device, the blood pressure monitor must only be used with the mains part described here.
  • Insert the mains part into the connection provided for this purpose on the blood pressure monitor. The mains part must only be connected to the mains voltage that is specified on the type plate.
  • Then insert the mains plug of the mains part into the mains socket.
  • After using the blood pressure monitor, unplug the mains part from the mains socket first and then disconnect it from the blood pressure monitor. As soon as you unplug the mains part, the blood pressure monitor loses the date and time setting but the saved measurements are retained.

7. Usage

General rules when measuring blood pressure yourself

  • In order to generate as informative a profile of the progression of your blood pressure as possible and ensure that the measured values can be compared, you should measure your blood pressure regularly and always at the same times of day. It is recommended that you measure your blood pressure twice a day: once in the morning after getting up and once in the evening.
  • You should always carry out the measurement when you are sufficiently physically rested. You should therefore avoid taking measurements during stressful periods.
  • Do not take a measurement within 30 minutes of eating, drinking, smoking or exercising.
  • Before the initial blood pressure measurement, make sure always to rest for 5 minutes.
  • Furthermore, if you want to take several measurements in succession, make sure always to wait for at least 1 minute between the individual measurements.
  • Repeat the measurement if you are unsure of the measured value.

Attaching the cuff

- Fundamentally, blood pressure can be measured on both arms. Certain deviations between the measured blood pressure on the right arm and left arm are due to physiological causes and completely normal. You should always perform the measurement on the arm with the highest blood pressure values. Before starting self-measurement, consult your doctor in this regard. From this point on, always take measurements on the same arm.

- The device may only be operated with one of the following cuffs. This should be selected in accordance with your upper arm circumference. The fit should be checked before measurement using the index mark described below.

Ref. no. Designation Armcircumference
163.946* Universal cuff 22-42 cm

* Included in standard delivery

  • Place the cuff onto the bare upper arm. The circulation of the arm must not be hindered by tight clothing or similar.
  • The cuff must be placed on the upper arm so that the bottom edge is positioned 2-3 cm above the elbow and over the artery. The line should point to the centre of the palm here.
  • Guide the end of the cuff that is sticking out through the metal ring, fold it back over the arm and close the cuff using the hook-and-loop fastener. The cuff should be fastened tightly, but not too tightly, so that two fingers can still fit under the closed cuff.

Illustration of a human wrist with a bandage device attached (no text or symbols)

Illustration of a person's leg with a medical device attached to the ankle (no text or symbols present)

Illustration of a hand holding a cable or wire, no text or symbols present

  • The cuff is suitable for you if the index mark (▼) is within the OK range after fitting the cuff.
  • Now insert the cuff line into the connection for the cuff connector.

MIN OK MAX

Pure diagram of pipe connection with arrow indicating direction (no text or symbols)

Adopting the correct posture

  • To carry out a blood pressure measurement, make sure you are sitting upright and comfortably. Lean back and place your arm on a surface. Do not cross your legs. Place your feet next to each other flat on the floor.
    • Always make sure that the cuff is at heart level.
  • To avoid distorting the measurement, you should remain as still as possible during the measurement and not speak.

Side profile illustration of a person sitting at a desk using a pressure monitor (no text or symbols)

Illustration of a person sitting at a desk with crossed-out lines indicating no text or symbols (no text or symbols present)

Performing the blood pressure measurement

Measurement

- To start the blood pressure monitor, press the Start/stop button ⓄAll display elements are briefly displayed.

The blood pressure monitor will begin the measurement automatically after 3 seconds.

The cuff automatically inflates.

i Measuring can be cancelled at any time by pressing the Start/stop button ①.

The cuff's air pressure is slowly released. If you already recognise a tendency for high blood pressure, you should reinflate the cuff and increase the cuff's pressure again. As soon as a pulse is found, the pulse symbol ♥ displayed.

  • Systolic pressure, diastolic pressure and pulse measurements are displayed.
  • A symbol at the bottom of the display also lights up to indicate whether you were sufficiently relaxed during the blood pressure measurement (green symbol = sufficiently at rest; red symbol = not at rest).

Observe the chapter on interpreting results/measuring the resting indicator in these instructions for use.

BEURER BM 55 - Measurement - 1

BEURER BM 55 - Measurement - 2

  • Er appears if the measurement could not be performed properly. Observe the chapter on error messages/troubleshooting in these instructions for use and repeat the measurement.
  • Now select the desired user memory by pressing the M1 or M2 memory buttons. If you do not select a user memory, the measurement is stored in the most recently used user memory. The relevant symbol for appears on the display.
  • Using the Start/stop button ① switch off the blood pressure monitor. The measurement is then stored in the selected user memory. If you forget to turn off the device, it will switch off automatically after approx. one minute. In this case too, the value is stored in the selected or most recently used user memory.
  • Wait for at least 1 minute before taking another measurement.

BEURER BM 55 - Measurement - 3

BEURER BM 55 - Measurement - 4

Evaluating results

General information about blood pressure

  • Blood pressure is the force with which the bloodstream presses against the arterial walls. Arterial blood pressure constantly changes in the course of a cardiac cycle.
    • Blood pressure is always stated in the form of two values:

  • The highest pressure in the cycle is called systolic blood pressure. This arises when the heart muscle contracts and blood is pumped into the blood vessels.

  • The lowest is diastolic blood pressure, which is when the heart muscle has completely stretched back out and the heart fills with blood.

- Fluctuations in blood pressure are normal. Even during repeat measurements, considerable differences between the measured values may occur. One-off or irregular measurements therefore do not provide reliable information about the actual blood pressure. Reliable assessment is only possible when you perform the measurement regularly under comparable conditions.

Cardiac arrhythmia

This device can identify potential disruptions of the heart rhythm when measuring and, if necessary, indicates this after the measurement with the symbol 🚙. This can be an indicator for arrhythmia. Arrhythmia is an illness in which the heart rhythm is abnormal because of flaws in the bioelectrical system that regulates the heartbeat. The symptoms (skipped or premature heart beats, pulse being slow or too fast) can be caused by factors such as heart disease, age, physical makeup, excess stimulants, stress or lack of sleep. Arrhythmia can only be determined through an examination by your doctor. If the symbol 🚙 appears frequently, please consult your doctor. Only they are able to determine any arrhythmia during an examination.

Risk indicator

The measurements can be classified and evaluated in accordance with the following table.

However, these standard values serve only as a general guideline, as the individual blood pressure varies in different people and different age groups etc.

It is important to consult your doctor regularly for advice. Your doctor will tell you your individual values for normal blood pressure as well as the value above which your blood pressure is classified as dangerous.

The bar chart on the display and the scale on the device show which category the recorded blood pressure values fall into. If the values of systole and diastole fall into two different categories (e.g. systole in the high normal category and diastole in the normal category), the graphical classification on the device always shows the higher category; for the example given this would be high normal.

Blood pressure value categorySystole (in mmHg)Diastole (in mmHg)Action
Level 3: severe hypertensionred ≥180≥110Seek medical attention
Level 2: moderate hypertensionorange160–179100–109Seek medical attention
Level 1: mild hypertensionyellow140–15990–99Regular monitoring by doctor
Blood pressure value categorySystole (in mmHg)Diastole (in mmHg)Action
High normalgreen 130-13985-89Regular monitoring by doctor
Normal green120-12980-84 Self-monitoring
Optimal green<120<80Self-monitoring

Source: WHO, 1999 (World Health Organization)

Measuring the resting indicator (using the HSD diagnosis)

The most frequent error made when measuring blood pressure is taking the measurement when not at rest (haemodynamic stability), which means that both the systolic and the diastolic blood pressures are distorted.

While measuring the blood pressure, the device automatically determines whether you are at rest or not.

BEURER BM 55 - Measuring the resting indicator (using the HSD diagnosis) - 1

GREEN: Haemodynamic stability

Measurement of the systolic and diastolic pressure is increased when the circulatory system is sufficiently at rest and is a very reliable indicator of resting blood pressure. The blood pressure measurement must be taken when the patient is physically and mentally rested, as it will be the basis for diagnosing the blood pressure level and regulating the patient's medical treatment. The blood pressure measurements taken when the circulatory system was at rest represent particularly reliable results.

BEURER BM 55 - GREEN: Haemodynamic stability - 1

RED: Lack of haemodynamic stability

It is very probable that the systolic and diastolic blood pressures have not been measured whilst the patient is at rest and the resting blood pressure measurement has therefore been distorted.

Repeat the measurement after a rest and relaxation period of at least five minutes. Go to a sufficiently quiet and comfortable spot and remain there calmly; close your eyes, breathe deeply and evenly and try to relax.

If the next measurement also shows insufficient stability, you can repeat the measurement after another resting period. If the measurements continue to show some instability, identify these blood pressure measurements as having been taken when the circulatory system had not been sufficiently rested.

In this case, nervousness or inner anxiety may be the cause and this cannot be cured by brief periods of rest. Existing cardiac arrhythmias may also prevent a stable blood pressure measurement. The accuracy of the resting blood pressure results is reduced in these users. Like any medical measurement method, the precision of the HSD diagnosis is limited and can lead to incorrect results in some cases.

Saving, displaying and deleting measured values

User memory

The results of every successful measurement are stored together with the date and time. If there are more than 60 measurements, the oldest measurements are lost.

  • To access memory recall mode, the blood pressure monitor must first be started. To do this press the Start/stop button ⓘ
  • Within 3 seconds of the full-screen display appearing, select the desired user memory (☐) with the M1 or M2 memory button.
  • To view the measurements for user memory press the M1 memory button.

User memory

- To view the measurements for user memory 🔒, press the M2 memory button.

Your last measurement will appear on the display.

BEURER BM 55 - User memory - 1

  • Press the relevant memory button (M1 or M2).
    If you have selected user memory1, the M1 memory button must be pressed. If you have selected user memory2, the M2 memory button must be pressed.

Flashes on the display. The average value of all saved measured values in this user memory is displayed.

BEURER BM 55 - User memory - 2

- Press the relevant memory button (M1 or M2).

Aflashes on the display. The average value of the morning measurements for the last 7 days is displayed (morning: 5.00 a.m. – 9.00 a.m.).

BEURER BM 55 - User memory - 3

- Press the relevant memory button (M1 or M2).

P#R# flashes on the display. The average value of the evening measurements for the last 7 days is displayed (evening: 6.00 p.m. - 8.00 p.m.).

BEURER BM 55 - User memory - 4

- When the relevant memory button (M1 or M2) is pressed again, the last individual measurement is displayed (in this example, measurement 03).

BEURER BM 55 - User memory - 5

  • When the relevant memory button (M1 or M2) is pressed again, you can view your individual measurements.
  • To switch the device off again, press the Start/stop button ①

You can exit the menu at any time by pressing the Start/stop button

  • To clear the memory of the relevant user memory, you must first select a user memory.
  • Start individual measurement access.
  • Press and hold the M1/M2 memory buttons for 5 seconds.

All the values in the current user memory are deleted.

BEURER BM 55 - User memory - 6

Transferring measurements

Connect the blood pressure monitor to your PC using the USB cable.

No data transfer may be launched whilst performing a measurement.

PCs shown on the display. Begin the data transfer in the "beurer HealthManager" PC software. During the data transfer, an animation is shown on the display. A successful data transfer is displayed as in figure 1. If the data transfer is unsuccessful, an error message appears as in figure 2. In this case, interrupt the PC connection and start the data transfer again.

After 30 seconds of not being in use or if communication with the PC is interrupted, the blood pressure monitor switches itself off automatically.

BEURER BM 55 - Transferring measurements - 1
figure 1

BEURER BM 55 - Transferring measurements - 2
figure 2

8. Cleaning and maintenance

  • Clean the device and cuff carefully using a slightly damp cloth only.
  • Do not use any cleaning agents or solvents.
  • Under no circumstances hold the device and cuff under water, as this can cause liquid to enter and damage the device and cuff.
  • If you store the device and cuff, do not place heavy objects on the device and cuff. Remove the batteries. The cuff line should not be bent sharply.

9. Accessories and replacement parts

Accessories and replacement parts are available from the corresponding service address (according to the service address list). Please state the corresponding order number.

Designation Item number and/ororder number
Universal cuff (22-42 cm) 163.946
Mains part (EU) 071.95
USB cable 163.484
  1. What if there are problems?
Error messagePossible cause Behebung
Er1The systolic pressure could not be recorded correctly.Repeat the measurement. Ensure that you do not move or speak.
Er2The diastolic pressure could not be recorded correctly.
HiThe systolic and diastolic pressures are above the measurement range.
LoThe systolic and diastolic pressures are below the measurement range.
Er3The cuff has been fastened too tightly.Repeat the measurement. Ensure that you do not move or speak.
Er4The cuff has been fastened too loosely.
Er5The inflation pressure is higher than 300 mmHg or the measured blood pressure is outside of the measurement range.
Er6Pumping up is taking longer than 160 seconds.
ErR, ErO, Er7 odEr Er8There is a system or device error.
The batteries are almost empty.Reinsert the batteries or replace with new batteries.
PcErUnable to send the data to the PC.Check there is a connection to the PC and repeat the process.

i Technical alarm - description

Should the recorded blood pressure (systolic or diastolic) lie outside the limits specified in the section "Technical specifications", the technical alarm will appear on the display indicating either "Hi" or "Lo". In such cases, you should seek medical assistance and check the accuracy of your procedure.

The limit values for the technical alarm are factory set and cannot be adjusted or deactivated. These alarm limit values are accorded second priority under the standard IEC 60601-1-8.

The technical alarm is a non-locking alarm and must not be reset. The signal shown on the display will disappear automatically after about 8 seconds.

11. Disposal

i Repairing and disposing of the device

  • Do not repair or adjust the device yourself. Proper operation can no longer be guaranteed in this case.
  • Do not open the device. Failure to comply will invalidate the warranty.
  • Repairs must only be carried out by Customer Services or authorised retailers. Before making a claim, please check the batteries first and replace them if necessary.
  • For environmental reasons, do not dispose of the device in household waste at the end of its service life.

Dispose of the device at a suitable local collection or recycling point in your country. Dispose of the device in

accordance with EC Directive – WEEE (Waste Electrical and Electronic Equipment). If you have any questions, please contact the local authorities responsible for waste disposal.

BEURER BM 55 - i Repairing and disposing of the device - 1

⚠ Disposing of the batteries

  • The empty, completely flat batteries must be disposed of through specially designated collection boxes, recycling points or electronics retailers. You are legally required to dispose of the batteries.
  • The codes below are printed on batteries containing harmful substances:

Pb = Battery contains lead,

Cd = Battery contains cadmium,

Hg = Battery contains mercury.

BEURER BM 55 - ⚠ Disposing of the batteries - 1

12. Specifications

Model no. BM 55

Type M1002
Measurement methodOscillometric, non-invasive blood pressure measurement on the upper arm
Measurement rangeCuff pressure 0 – 300 mmHg,systolic 60– 260 mmHg,diastolic 40 –199 mmHg,Pulse 40 –180 beats/minute
Display accuracySystolic ± 3 mmHg, diastolic ± 3 mmHg,pulse ± 5 % of the value shown
Measurement inaccuracyMax. permissible standard deviationaccording to clinical testing:systolic 8 mmHg /diastolic 8 mmHg
Memory 2 x 60 memory spaces
Dimensions L 186 mm x W 95 mm x H 56 mm
Weight Approx. 467 g (without batteries)
Cuff size 22 to 42 cm
Permissible operating conditions+10 °C to + 40 °C, ≤ 90 % relative air humidity (non-condensing)
Permissible storage and transport conditions- 20 °C to + 55 °C, ≤ 90 % relative air humidity, 800 -1050 hPa ambient pressure
Power supply4x1.5V —— AAA batteries
Battery life For approx.200 measurements, depending on the blood pressure level and/or pump pressure
Classification Internal supply, IPX0, no AP or APG, continuous operation, application part type BF

The serial number is located on the device or in the battery compartment.

Technical information is subject to change without notification to allow for updates.

  • This unit is in line with European Standard EN 60601-1-2 (In accordance with CISPR 11, IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6, IEC 61000-4-8, IEC 61000-4-11) and is subject to particular precautions with regard to electromagnetic compatibility (EMC). Please note that portable and mobile HF communication systems may interfere with this unit.
  • This device is in line with the EU Medical Devices Directive 93/42/EEC, the “Medizinproduktegesetz” (German Medical Devices Act) and the standards EN 1060-1 (non-invasive sphygmomanometers, Part 1: General requirements),

EN 1060-3 (non-invasive sphygmomanometers, Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems) and IEC 80601-2-30 (Medical electrical equipment – Part 2 – 30: Particular requirements for the safety and essential performance of automated non-invasive blood pressure monitors).

- The accuracy of this blood pressure monitor has been carefully checked and developed with regard to a long useful life. If using the device for commercial medical purposes, it must be regularly tested for accuracy by appropriate means. Precise instructions for checking accuracy may be requested from the service address.

13. Mains part

Model no. LXCP12-006060BEH

Input 100–240V,50–60Hz,0.5Amax
Output 6 V DC, 600 mA, only in connection with beurer blood pressure monitor.
Supplier Shenzhen longxc power supply co., Ltd
Protection This device is double insulated and protected against short circuit and overload by a primary thermal fuse.Make sure to take the batteries out of the compartment before using the mains adapter.
Polarity of the the DC voltage connection
Double insulated / equipment class 2
Enclosures and ProtectiveCoversEquipment enclosed to protect against contact with live parts, and with parts which can become live (finger, pin, hook test).The operator shall not contact the patient and the output plug of AC mains adapter simultaneously.

14. Warranty/service

Beurer GmbH, Söflinger Straße 218, 89077 Ulm, Germany (hereinafter referred to as "Beurer") provides a warranty for this product, subject to the requirements below and to the extent described as follows.

The warranty conditions below shall not affect the seller's statutory warranty obligations which ensue from the sales agreement with the buyer.

The warranty shall apply without prejudice to any mandatory statutory provisions on liability.

Beurer guarantees the perfect functionality and completeness of this product.

The worldwide warranty period is 5 years, commencing from the purchase of the new, unused product from the seller.

The warranty only applies to products purchased by the buyer as a consumer and used exclusively for personal purposes in the context of domestic use.

German law shall apply.

During the warranty period, should this product prove to be incomplete or defective in functionality in accordance with the following provisions, Beurer shall carry out a repair or a replacement delivery free of charge, in accordance with these warranty conditions.

If the buyer wishes to make a warranty claim, they should approach their local retailer in the first instance: see the attached “International Service” list of service addresses.

The buyer will then receive further information about the processing of the warranty claim, e.g. where they can send the product and what documentation is required.

A warranty claim shall only be considered if the buyer can provide Beurer, or an authorised Beurer partner, with

  • a copy of the invoice/purchase receipt, and
    – the original product.

The following are explicitly excluded from this warranty:

– deterioration due to normal use or consumption of the product;
- accessories supplied with this product which are worn out or used up through proper use (e.g. batteries, rechargeable batteries, cuffs, seals, electrodes, light sources, attachments and nebuliser accessories);
– products that are used, cleaned, stored or maintained improperly and/or contrary to the provisions of the instructions for use, as well as products that have been opened, repaired or modified by the buyer or by a service centre not authorised by Beurer;

– damage that arises during transport between manufacturer and customer, or between service centre and customer;
– products purchased as seconds or as used goods;
– consequential damage arising from a fault in this product (however, in this case, claims may exist arising from product liability or other compulsory statutory liability provisions).

Repairs or an exchange in full do not extend the warranty period under any circumstances.

FRANÇAIS

BEURER BM 55 - FRANÇAIS - 1

Mesures relatives aux piles

Diagram showing a battery connected to a motor, with a magnified inset illustrating the internal structure.

4 × 1.5V AAA (LR03)

MIN OK MAX

Pure diagram of pipe connection with arrow indicating direction (no text or symbols)

Side profile illustration of a person sitting at a desk using a pressure monitor (no text or symbols)

Illustration of a person sitting at a desk with crossed-out lines indicating no text or symbols (no text or symbols present)

Mesurer la tension artérielle

Source : WHO, 1999 (World Health Organization)

4 × 1.5V AAA (LR03)

Formato de hora → Fecha → Hora

BEURER BM 55 - Formato de hora → Fecha → Hora - 1

Side profile illustration of a person sitting at a desk using a computer (no text or symbols visible)

Simple line drawing of a person sitting at a desk with crossed-out lines indicating no text or symbols (no text or symbols present)

Fuente: WHO, 1999 (World Health Organization)

4 × 1.5V AAA (LR03)

BEURER BM 55 - Formato de hora → Fecha → Hora - 4

Illustration of a hand holding a cable with a circular component, no text or symbols present

Illustration of a person sitting at a desk with crossed-out lines indicating pressure or stress (no text or symbols)

Fonte: WHO, 1999 (World Health Organization)

4 × 1,5V AAA (LR03)

Illustration of a person sitting at a desk with crossed-out lines indicating no text or symbols (no text or symbols present)

Kaynak: WHO, 1999 (World Health Organization)

4 × 1.5V AAA (LR03)

Illustration of a hand holding a cable with a circular component, no text or symbols present

MIN OK MAX

Pure diagram of pipe connections without any text, numbers, or symbols

Side profile illustration of a person sitting at a desk using a pressure monitor (no text or symbols)

BEURER BM 55 - Formato de hora → Fecha → Hora - 12

Измерение

4 x 1.5V AAA (LR03)

MIN OK MAX ✓

Pure diagram of pipe connection with arrow indicating direction (no text or symbols)

Illustration of a person sitting at a desk with arms crossed out, no text or symbols present

Table of contents Click a title to access it
Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : BEURER

Model : BM 55

Category : Sphygmomanometer