USER MANUAL Pro Health 5572AQ REVITIVE
natural_image
Close-up of a black and white striped electronic device with a power button icon (no text or symbols visible)
Please read the user's manual Carefully before using this product. USE ONLY AS DIRECTED.
If symptoms persist, consult your healthcare professional.
natural_image
Illustration of a seated person on a chair with feet highlighted in an inset showing foot positioning (no text or symbols)How-to videos/Tutoriels vidéo/Anleitungsvideos/Video esplicativi/Vídeos instructivos/ Instructievideo's

text_image
QR code image containing encoded data, no visible human-readable textFor videos of setup and using your Revitive system, go to: support.revitive.com
What's in the box? 4
Parts and Controls 5-6
Introduction to Revitive 7
Operator profile 7
Indications for use 7
Important Safeguards 8-11
Who should NOT use Revitive 8
Warnings and Cautions 9-10
Safety Precautions 11
Adverse Reactions 11
Instructions for use 12-15
Step 1: Setting up Revitive for use 12
Step 2: Using the SoftTouch foot-pads 13-15
Cleaning and storing 15
Troubleshooting
Technical Specifications
Warranty
What's in the box?
On opening the carton, please check that the following components are provided. If you think anything is missing, please contact us using the helpline numbers on the back of this booklet.
1 AC/DC power adaptor Cable length 180cm

text_image
REVITIVE2 Revitive ProHealth

natural_image
Circular diagram with concentric rings and a central indicator light, no text or symbols present.3 Remote control

natural_image
Illustration of a remote control with a black rotary knob and 'REVITIVE' text below (no additional symbols or text)4 AAA batteries x 2


Parts and controls
Control panel

text_image
30 Time remaining Intensity Control -/+ 40 Intensity levelRemote control

text_image
trol Intensity Control - /+ Power ON/OFF REVITIVEOperator Profile
The intended operator profile is based on users in the home environment who may suffer from immobility, poor circulation and related issues such as pain, leg stiffness and swelling in the legs, ankles and feet.
Revitive is intended for use by the end user in a non-clinical setting and without the supervision or intervention of a clinician during use.
A typical user may be of any adult age range and is not specific to any demographic of gender, ethnicity or educational background.
Indications for use
When using the foot-pads, Revitive it is intended to:
- Help maintain leg vein health - by increasing circulation, delivering more oxygenated blood and reducing swelling (oedema) in legs, feet and ankles.
- Improve circulation to reduce or prevent blood-pooling (stasis) - caused by poor circulation (Chronic Venous Insufficiency/varicose veins)
- Reduce Pain and discomfort in the legs/ankles/feet - caused by poor circulation (Peripheral Arterial Disease)
- Improve symptoms associated with varicose veins/ Chronic Venous Insufficiency
- Improve circulation in the legs - caused by Peripheral Arterial Disease
- Increase walking distance before the onset of claudication symptoms (pain) – caused by Peripheral Arterial Disease
If you are otherwise healthy and have a sedentary lifestyle or spend long periods inactive, Revitive may help to:
- Alleviate tired, aching & heavy legs, including cramp
- Reduce swollen feet & ankles
• Help maintain leg vein health - Actively increase circulation
The above Indications for Use are certified under the Medical Devices Directive 93/42/EEC.
Visit www.revitive.com for more information
Read all instructions before use.
Save these instructions.
Revitive should not be used by some people.

Do not use if (contraindications):
- You are fitted with an electronic implanted device such as a heart pacemaker or Automatic Implantable Cardioverter Defibrillator (AICD)
- You are pregnant
- You are being treated for, or have the symptoms of, an existing Deep Vein Thrombosis ("DVT")
Long periods of inactivity can put you at greater risk of developing Deep Vein Thrombosis (DVT). DVT is a blood clot and usually occurs in a deep leg vein. If part of the DVT breaks off it may lead to potentially life-threatening complications such as pulmonary embolism.
If you have been inactive for prolonged periods and suspect you may have a DVT, consult your doctor immediately. To prevent dislodging the clot do not use Revitive.
In some cases of DVT there may be no symptoms. However it is important to be aware of the symptoms that may include:
- Pain, swelling and tenderness in one of your legs (usually your calf)
• A heavy ache in the affected area
• Warm skin in the area of the clot
- Redness of your skin, particularly at the back of your leg, below the knee
Consult your doctor as soon as possible if you show any signs of the above symptoms.

Warnings
Consult with your doctor before using this device if:
• You are in the care of a doctor
- You have a history of heart problems
- You have had medical or physical treatment for your pain
- You have suspected or diagnosed heart disease
• You have suspected or diagnosed epilepsy
• You are unsure about the suitability of Revitive for you
- You are unsure about the cause of your symptoms
If you have a metallic implant:
- If you have a metallic implant, you may experience pain or discomfort near the implant when applying electrical stimulation. If this should occur discontinue use and seek advice from your doctor
- The electrical stimulation may feel more intense close to a metallic implant. It is safe to continue use provided no pain is experienced. You may need to adjust the intensity to a comfortable level
When using foot-pads:
- Only apply to areas with intact/unbroken skin
• Do not apply foot-pads directly or adjacent to these areas:
• on cut or broken skin, open wounds or rashes; swollen, red, infected or inflamed areas; or skin eruptions (such as phlebitis, thrombophlebitis, varicose veins, cellulitis). Check under your feet and toes before using the device.
• on, or close to, malignant tumours
• on areas treated with radiotherapy (within the past 6 months) - near the thorax (chest) may increase the risk of cardiac fibrillation
• on your head, face, neck or chest - Ensure that any water based moisturiser is evenly applied and thoroughly absorbed into the skin before applying stimulation. There is a chance that uneven application could increase the risk of skin irritation or burn, when using the electrical stimulation
- Symptoms may worsen during the initial treatment phase before getting better. This may occur if the body has not fully adjusted to increased muscular activity and blood circulation. If this occurs, reduce the intensity and the duration of treatment which will reduce the initial symptoms. If symptoms persist consult with your doctor
When not to use Revitive:
• There are times you should not use Revitive. Do not use Revitive:
• In the presence of electronic monitoring equipment
• Together with a life-supporting medical electronic device
- When you are in the bath or shower
• While you are sleeping
• In conjunction with a brace or cast without first consulting with your doctor
- Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally

Warnings
Additional Warnings:
- Use of cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation
- Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Revitive System, including cables specified by the manufacturer

Cautions
When applying stimulation:
- Be careful when applying stimulation over areas of skin that lack normal sensation. It may cause skin irritation due to the inability to feel stimulation until the intensity is too high. Use a low intensity to achieve a gentle muscle contraction, and/or use for a shorter time, to avoid over-stimulation. Check for signs of skin irritation (redness), bruising or pain. If in doubt consult your doctor
- This product is not intended for use by persons with reduced physical, sensory or mental capabilities, unless they are supervised by a person responsible for their safety
- Be careful when applying stimulation:
• After recent surgical procedures (within the past 6 months) as stimulation may disrupt your healing process - If your tissues are likely to bleed following an injury such as a muscle tear - it is recommended not to stimulate the immediate area within the first 12 hours after sustaining the injury. Use a low intensity and/or shorter time to avoid over-stimulation
• After a long period of immobility or inactivity – use a low intensity and shorter time to avoid over-stimulation or muscle fatigue
Additional Cautions:
- The long term effects of electrical stimulation are unknown
- Electrical stimulation may not work for every user, please seek advice from your doctor
Safety Precautions
- Do not stand on the machine. Use only when seated
- Do not position Revitive so that it is difficult to disconnect or turn off
- Use Revitive only for its intended purpose
- Do not expose Revitive to extreme heat
- Do not spill liquid on Revitive or its accessories
- Do not overload the electrical socket
- Keep device out of the reach of children
- Keep power cords and cables out of the reach of children to prevent risk of strangulation
- Revitive foot-pads may be used by multiple persons. Ensure device is cleaned after each use
- During use, do not touch foot-pads with your hands until Revitive has powered off
- Use Revitive only with the accessories supplied by, or purchased from, the manufacturer
- Check cables periodically for damage
- Do not open Revitive or repair it yourself. This will invalidate your warranty and may cause serious harm
- In the unlikely event your Revitive malfunctions, disconnect it from the power source and contact your nearest authorised agent
• Revitive has passed the required tests for Electromagnetic Interference (EMI); it may still be affected by excessive emissions and/or may interfere with more sensitive equipment
- It has been reported that some universal remote control devices (e.g. for TV etc.) can change the settings on Revitive if used during a treatment. If this should occur, simply adjust the time or intensity settings on Revitive back to where you want them using the Revitive Remote Control, or using the device Control Panel
- After any exposure to hot or cold temperatures outside the specified operating range of 10 – 40°C allow the product to re-adjust to the recommended operating temperatures to ensure continued product performance
Adverse Reactions
- If you experience adverse reactions, stop using Revitive and consult with your doctor immediately.
Instructions for use
Step 1: Setting up Revitive for use
a Remove all parts from the box:
• AC/DC power adaptor
• Revitive ProHealth device
- Remote control
- AAA batteries x 2
See "What's in the box?" on page 4 to help identify the parts.

natural_image
Illustration of a hand holding a pen and ruler, no text or symbols presentb Locate the remote control and insert the batteries.

natural_image
Simple line drawing of a magnifying glass over a curved surface, no text or symbols presentC Locate the AC/DC power adaptor.
Plug the small end of the power cord into the adaptor socket on the side of the device.

text_image
REVITIVEd Plug the power adaptor into the nearest electrical socket.
Before you use Revitive for the first time, read the Important Safeguards on pages 9-12. If in doubt, consult your doctor before using the product.
Step 2: Using the SoftTouch foot-pads
For best results:
• Use Revitive foot-pads for at least 30 to 60 minutes each day, 7 days a week. If you have varicose veins (CVI), use Revitive foot-pads for 60 minutes per day.
- It is important to use Revitive at a high enough intensity to give you a strong comfortable muscle contraction. Most people achieve a strong muscle contraction over intensity level 40 (intensity range 1-99).
• If you are diagnosed with a long term medical condition such as High Blood Pressure, High Cholesterol, Diabetes, Osteoarthritis, Chronic Venous Insufficiency (CVI), Peripheral Arterial Disease (PAD) or COPD, it can take up to 8 weeks to help reduce symptoms associated with these chronic conditions.
Do not use Revitive, using EMS stimulation, for more than 6 sessions of 30 minutes (or the equivalent) per day.
a It is important that you are well hydrated. We recommend that you apply a moisturiser to the soles of the feet to help hydrate the skin and improve the delivery of electrical stimulation.
Please use a water based moisturiser such as the Revitive Foot and Leg Cream.
We also recommend drinking a glass of water before using Revitive.
Sit with both bare feet on the foot-pads

natural_image
Illustration of a person seated on a chair with feet resting on the floor, showing foot positioning (no text or symbols)b Get into a comfortable seated position.
For best results sit with your knees at a 90 degree angle (right angle).
c Place Revitive on the floor in front of you.
Revitive is designed to be used while seated. Never stand on the Revitive unit.
d Place both bare feet on the foot-pads. Make sure you remove all footwear, including socks/stockings.

natural_image
Diagram showing a device with three circular components and a diagonal line, no text or symbols present.e Press the power button on the device or on the remote control to turn Revitive on.
You will hear a beep and the display will read ⏻ 30 minutes.
To turn off Revitive at any time during operation, press the power button on the device or on the remote control.

natural_image
Diagram showing a device with two circular elements and a central circular component, divided by a diagonal line (no text or symbols)f Press the (+) Intensity Control one time to start stimulation.
You can use the buttons on the control panel or remote control. Ensure both feet are on the foot-pads when increasing the intensity.
Press the (+) button to increase the level.
Press the (−) button to decrease the level.
g Ensure you familiarise yourself with the feeling of the electrical muscle stimulation. Use on a low intensity for 30-60 minutes once per day for a few days.
Set the intensity at a level where you can feel the stimulation in your feet which creates a calf muscle contraction.
It is normal to feel tingling or varying sensations in your feet and calf muscles.
Revitive does not vibrate.
h After a few days of familiarisation, use a higher intensity that provides you with strong but comfortable calf muscle contractions.
Most people experience a strong contraction at intensity level 40 or over.
The IsoRocker® enables Revitive to rock up and down once muscle contractions are nearing the desired level. The rocking is caused by muscle contractions in the leg, and the resulting ankle movement allowed by the IsoRocker is a good indicator that the intensity level is sufficient to provide the therapy. Increasing the intensity above this level will enhance the therapy, but should not be allowed to become uncomfortable.
You can set the intensity between 1-99.
Please note that the intensity required may vary from day to day.
If you feel pain or discomfort:
- Remove one or both feet from the foot-pads to stop the stimulation
- Lower the intensity of the foot-pads using the control, before replacing your feet on Revitive

natural_image
Diagram of a battery cell with positive and negative terminals, no text or symbols presentWhen you have finished setting the intensity, the display reverts to the Time Display.
The ① Time Display will count down will count down from 30 minutes as Revitive cycles through its program.
Each time you adjust the intensity, the intensity level shows in the display. Again, when you have finished adjusting the intensity, the display reverts to the Minutes-counter, continuing to count down the session.

natural_image
Diagram of a device with wheels and a central circular component, no text or symbols presentj When Revitive times out, the ⓄTime Display reads 00 and you will hear three beeps.
k Revitive switches itself off automatically.
Cleaning and storing Revitive
a Ensure Revitive is switched off.
b Wipe down the foot-pads with a soft damp cloth.
Do
not
clean
with
chemicals.
Do not immerse Revitive in water.
c Store Revitive in a cool, dry and dust-free location.
Store out of direct sunlight.
Replacing the Batteries
The remote control requires two AAA 1.5V batteries to operate.
Troubleshooting
Problem Possible Cause
Solution
| Revitive is on (lights illuminated on the LED display) but I cannot feel the electrical stimulation through the foot-pads. | Not placing both bare feet on the foot-pads at the same time. | Ensure that your feet are bare and each foot is placed on each of the foot-pads at the same time. Keep increasing the intensity up to a maximum of 99 until you feel the stimulation. |
| Your feet may be dry. Moisturise the | soles of your feet to improve conductivity and stimulation and try the procedure again. You may also have to increase the intensity level. | |
| You may be dehydrated. Drink plenty of water before and after using the device. The device uses your body to create the electrical circuit. Water is an excellent conductor of electricity and if your body is less hydrated (below 60%) then the stimulation may be less, therefore it is important to always remain well hydrated. | ||
| The intensity level may be on too low a setting. | This is a very safe device. Keep increasing the intensity level towards 99 until you feel the stimulation. You may find that you have to increase the intensity level as you get used to the therapy. The aim is not to get to 99 but to find a setting that produces strong muscle contractions in your calves and is comfortable for you. | |
| If, having tried the solutions above, you still cannot feel the stimulation: | Test the device by placing one hand across both foot-pads at the same time (The heel end of the foot-pads is easiest). With your other hand, and starting from zero, increase the intensity level until you can feel the stimulation. If you can feel the stimulation through your hand then the device is working. If on 99 you still cannot feel the stimulation then please contact your authorised dealer. | |
| No power or lights to Revitive when it is switched on. | AC Adaptor not switched on at the electrical socket or AC Adaptor not plugged into device properly. | Check electrical socket power is switched on and the AC Adaptor is plugged into the device correctly. If still not working – contact your authorised dealer. |
| Revitive is not vibrating. | REVITIVE IS NOT DESIGNED TO VIBRATE. | |
| The IsoRocker® is not "rocking". | It is the muscles in your legs that cause the "rocking", the IsoRocker® is a pivot. | The IsoRocker® will only rock when increased to an intensity which causes sufficient calf muscle contraction. It may be that you cannot comfortably increase the intensity high enough until you get used to the sensation. It is important that intensity is adjusted to a level that is comfortable. |
| The device is too far in front of you. | Sit with your knees at a 90 degree angle. | |
| When using the IsoRocker® on a hard floor it makes a tapping noise. | Incorrect positioning of the device or too high an intensity level. | Adjust the positioning of the device or lower the intensity level to reduce the device tapping. Alternatively use a floor mat under the device to cushion the sound. |
Problem Possible Cause Solution
| My legs are aching after treatment. | You may have the intensity on too high a setting and your muscles are being overworked. | Leave adequate time after each treatment to allow the muscles to recover (just like after vigorous exercise!). On your next session start on a lower setting (where you can feel the mild electrical stimulation and it is comfortable) and reduce the duration until your muscles have acclimatised to the stimulation. |
For more information including Frequently Asked Questions, please visit www.revitive.com
Technical Specifications
| Name of product Revitive ProHealth | |
| Model 5572AQ | |
| Frequency (+/- 10%) 20Hz - 44Hz | |
| Output current Max 15mA | |
| Weight | 1.12 Kg +/- 0.5Kg |
| Dimensions 360mm (∅) x 70.9mm (D) | |
| Power consumption 5W | |
| AC adaptor | CE Approved |
| Power sourceInput (adaptor used) ➕Output ➕ | 100-240V AC ( ➕), 50/60Hz, 0.18A5V ( ➕) DC ⋯,1.0A |
| Applied partsParts of Revitive that in normal use come into physical contact with the user. | SoftTouch Pads (Foot-Pads) - 195cm2 |
| Durable Period (Service Life) of Device and Remote control | 4 years |
The Remote Control replicates the controls found on the device
| Operating Frequencies 38 KHz | |
| Operating Range Distance | 0 to 25m at horizontal0 to 18m at +/- 30 degree angle from the horizontal. |
Output Specifications for Electrical Muscle Stimulation (EMS):
| Waveform Biphasic | |
| Shape Square symmetrical with polarity reversal | |
| Maximum Output Voltage (+/-15%) @500Ω 35Vp | |
| Pulse Duration (+/-10%) 450-970μs | |
| Net Charge @ 500Ω [0.001]mC | |
| Maximum Power Density @ 500Ω 0.68 mW/cm2 | |
| ON Time (+/-10%) | 1.90 - 8.30s |
| OFF Time (+/-10%) | 1.00 - 1.50s |
The values of PULSE DURATIONS, PULSE repetition frequencies and amplitudes, including any d.c. component, shall not deviate by more than ±20% when measured with a load resistance within the range specified.
If confirmation is required that the Revitive works within its Essential Performance after a certain period of time, contact the manufacturer
Symbols
| Complies with European Medical Devices Directive (93/42/EEC) | ![]() |
| UK Conformity AssessedProduct conforms to all applicable U.K. legislative requirements. | ![]() |
| Device serial number including year (YYYY) and month (MM) of manufacture can be found on the box and back of unit | #YYYYMMXXXXX |
| Item number | ![]() |
| Indicates the unique device identification reference | ![]() |
| Contraindications, Warnings and CautionsMake sure you understand these before using Revitive | ![]() |
| Power | ![]() |
| Time Remaining | ![]() |
| Intensity Level | [HKSD] |
| Center Positive Polarity | [HS24] |
| Class II medical electrical equipment double insulated | ![]() |
| Type BF medical electrical equipment | ![]() |
| Legal manufacturer of the device | ![]() |
| EU/EC European Authorised Representative | ![]() |
| Consult instructions for use | ![]() |
| Ingress Protection Rating | [IGK2] |
| FCC markCertification mark employed on electronic products manufactured or sold in the United States which certifies that the electromagnetic interference from Revitive is under limits approved by the Federal Communications Commission.Actegy Ltd complies with all applicable FCC rules. | ![]() |
| The Waste Electrical and Electronic Equipment Directive (WEEE Directive).At the end of the product lifecycle, do not throw this product into normal household garbage, but take it to a collection point for the recycling of electronic equipment | ![]() |
| Use-by date YYYY MM DD | |
| Humidity, temperature and air pressure limit for storage and transport | ![]() 1060 hPa500 hPa |
| Humidity, temperature and air pressure limit for operating conditionsAfter any exposure to hot or cold temperatures outside the specified operating range of 10 - 40°C allow the product to re-adjust to the recommended operating temperatures to ensure continued product performance. | ![]() 1060 hPa700 hPa |
| Indoor Use Only | |
| Medical device does not contain natural rubber latex | ![]() |
| Do not disassemble |
Your 2-year warranty
It is important to retain the retailer's receipt as proof of purchase. Staple your receipt to this back cover for future reference.
Please quote the following information if the product develops a fault. These numbers can be found on the base of the product:
Model no:
Lot no:
All Revitive devices are individually tested before leaving the factory. In the unlikely event of any device proving to be faulty within 30 days of purchase, it should be returned to the place of purchase for it to be replaced.
If the fault develops after 30 days and within 24 months of original purchase, you should contact your local distributor quoting model number and LOT number on the product, or write to your local distributor at the address shown.
You will be asked to return the product (in secure, adequate packaging) to the address shown with a copy of proof of purchase.
Subject to the exclusions set out below (see Exclusions) the faulty device will then be repaired or replaced and dispatched, usually within 14 working days of receipt.
If, for any reason, this item is replaced during the 2-year guarantee period, the guarantee on the new item will be calculated from the original purchase date. Therefore, it is vital to retain your original till receipt or invoice to indicate the date of initial purchase.
To qualify for the 2-year guarantee, the device must have been used according to the manufacturer's instructions supplied.
Exclusions
1 Actegy, manufacturer of Revitive devices, shall not be liable to replace the goods under the terms of the guarantee where:
- The fault has been caused or is attributable to accidental use, misuse, negligent use or used contrary to the manufacturer's recommendations or where the fault has been caused by power surges or damage caused in transit.
- The device has been used on a voltage supply other than that stated on the product or used with a power adaptor other than the one supplied with the product.
• Repairs have been attempted by persons other than our service staff (or authorised dealer). - The device has been used for hire purposes or non-domestic use.
• The device is second hand.
2 Actegy are not liable to carry out any type of servicing work, under the guarantee.
3 Accessories such as Electrode Body Pads and bags are not covered by the guarantee.
4 Remote control batteries and any damage from leakage are not covered by the guarantee.
5 This guarantee does not confer any rights other than those expressly set out above and does not cover any claims for consequential loss or damage. This guarantee is offered as an additional benefit and does not affect your statutory rights as a consumer.
To activate your free 2-year warranty please register your device at: support.revitive.com
Sommaire
natural_image
Circular diagram with concentric rings and a central indicator light, no text or symbols present.3
Télécommande

natural_image
Illustration of a remote control device with buttons and power symbol (no text or labels)4
Piles AAA x 2

Pièces et commandes

text_image
Repose-pieds SoftTouch Panneau de commande/écran Bouton de MARCHE/ARRÊT
natural_image
Illustration of a hand holding a pen and ruler, no text or symbols presentnatural_image
Simple line drawing of a magnifying glass over a light bulb (no text or symbols)natural_image
Illustration of a seated person on a chair with feet placed on the floor, showing foot positioning (no text or symbols)natural_image
Abstract diagram with circular elements and a diagonal division, no text or symbols presentnatural_image
Diagram showing a device with two circular components and a central fan-like structure, divided by a diagonal line (no text or symbols)natural_image
Digital display icon showing 30 and plus signs, surrounded by curved dark bands (no text or symbols)natural_image
Pure diagram of a circular symbol with plus and minus signs, no readable text or labels2
Revitive ProHealth

natural_image
Circular diagram with concentric rings and a central white bar, featuring a '90' indicator and a power button symbol (no readable text or labels)3
Fernbedienung

natural_image
Line drawing of a remote control with buttons and power symbol (no text or labels)natural_image
Illustration of a hand holding a pen and ruler, no text or symbols presentnatural_image
Simple line drawing of a magnifying glass over a curved surface (no text or symbols)natural_image
Illustration of a seated person on a chair with feet resting on the floor, shown in two views: top view (no text or symbols)natural_image
Diagram showing a device with two circular components and a diagonal line, no text or symbols present.Bedienungsanleitung

natural_image
Diagram showing a car interior with two circular elements and a central circular component, divided by diagonal lines (no text or symbols)natural_image
Diagram of a battery with a circular display showing 30V, surrounded by two side blades (no text or symbols)natural_image
Diagram showing a circular symbol with plus and minus signs, surrounded by diagonal striped borders (no readable text or labels)2 Revitive ProSalute

natural_image
Circular diagram with concentric rings and a central indicator light, no text or symbols present.3 Telecomando

natural_image
Illustration of a remote control with a rotary knob and power button, labeled 'REVITIVE' at the bottom (no other text or symbols)4 Due batterie AAA

Componenti e comandi

Visit www.revitive.com for more information
natural_image
Illustration of a hand holding a pen and ruler, no text or symbols presentnatural_image
Simple line drawing of a magnifying glass over a surface, no text or symbols presentnatural_image
Illustration of a person seated on a chair with feet resting on the floor, showing foot positioning and posture (no text or symbols)natural_image
Diagram showing two circular components with a diagonal line and a central circle, no text or symbols present.
natural_image
Diagram showing a car interior with a circular component and a magnified view of the dashboard (no text or symbols)natural_image
Simple diagram with a central circular display showing '30' and plus/minus signs, surrounded by diagonal striped borders (no text or symbols)natural_image
Diagram of a car wheel with tire tracks and a circular dial indicator (no text or symbols)natural_image
Circular diagram with concentric rings and a central '40' button, no text or symbols present3
Mando a distancia

text_image
REVITIVE4
2 pilas AAA

Piezas y mandos

Visit www.revitive.com for more information
natural_image
Illustration of a hand holding a pen or tool, no text or symbols presentnatural_image
Simple line drawing of a magnifying glass hovering over a surface, no text or symbols presentnatural_image
Illustration of a person seated on a chair with feet resting on the floor, showing foot positioning and posture (no text or symbols)natural_image
Diagram showing two circular components with power and current indicators, separated by a diagonal line (no text or symbols)natural_image
Diagram showing a device divided into two sections with circular annotations and a central button, no readable text or symbols present.natural_image
Diagram of a battery with three leads and a central display showing 30 ohms (no text or symbols)natural_image
Diagram of a battery cell with 40Ω and polarity indicators, no text or symbols present2
Revitive ProHealth

natural_image
Circular diagram with concentric rings and a central '40' button, no text or symbols present3
Afstandsbediening

natural_image
Illustration of a remote control with a black rotary knob and 'REVITIVE' label (no additional text or symbols)4
AAA - batterijen x2

natural_image
Illustration of a hand holding a pen and ruler, no text or symbols presentnatural_image
Simple line drawing of a magnifying glass over a surface, no text or symbols presentC Pak de AC/DC-adapter.
natural_image
Illustration of a person seated on a chair with feet resting on the floor, showing foot positioning and body positioning (no text or symbols)natural_image
Diagram showing a split electrical circuit with two circular components and a power button, no text or symbols present.natural_image
Diagram showing a device with a circular component and a stylized eye, divided by a diagonal line (no text or symbols)natural_image
Simple diagram with a central circular display showing '30' and plus/minus signs, surrounded by dark textured vertical bars (no text or symbols)natural_image
Diagram of a speedometer with tire pattern and central display (no text or symbols)health powered by science
inspired by you"

Actegy Ltd
1 West Point
Western Road
Bracknell, RG12 1HJ
United Kingdom
Tel: +44 (0)800 014 6377

MDSS
Schiffgraben 41
30175 Hannover, Germany.
Distributor/Distributeur/Vertrieb/Distributore/Distribuidor
UK:
Imported by and
UK Responsible Person:
Actegy Ltd
1 West Point
Western Road
Bracknell, RG12 1HJ
United Kingdom
Tel: +44 (0)800 014 6377
info.uk@actegy.com
FR/IT/ES/NL/BE:
Revitive SAS
Kundenservice@revitive.de