5201 - Measuring equipment PeakTech - Free user manual and instructions
Find the device manual for free 5201 PeakTech in PDF.
User questions about 5201 PeakTech
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Measuring equipment in PDF format for free! Find your manual 5201 - PeakTech and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. 5201 by PeakTech.
USER MANUAL 5201 PeakTech
1. "Cd" means cadmium.
2. "Hg" means mercury.
3. "Pb" stands for lead.13
This product complies with the requirements of the following directives of the European Union for CE conformity: 2014/30/EU (electromagnetic compatibility), 2011/65/EU (RoHS). We herewith confirm that this product meets the essential protection standards, which are given in directions of council for adaptation of the administration regulations for UK of Electromagentic Compatibility Regulations 2016 and the Electrical Equipment (safety) regulations 2016. The following safety precautions must be observed before operation. Damages resulting from failure to observe these safety precautions are exempt from any legal claims whatever:
- Comply with the warning labels and other info on the equipment.
- Do not subject the equipment to direct sunlight or extreme temperatures, humidity or dampness.
- Do not subject the equipment to shocks or strong vibrations.
- Do not operate the equipment near strong magnetic fields (motors, transformers etc.).
- Keep hot soldering irons or guns away from the equipment.
- Allow the equipment to stabilize at room temperature before taking up measurement (important for exact measurements).
- Replace the battery as soon as the battery indicator “BAT” appears. With a low battery, the meter might produce false reading that can lead to electric shock and personal injury.
- Fetch out the battery when the meter will not be used for long period.
- Periodically wipe the cabinet with a damp cloth and mid detergent. Do not use abrasives or solvents.15
- Do not operate the meter before the cabinet has been closed and screwed safely as terminal can carry voltage.
- Do not store the meter in a place of explosive, inflammable substances.
- Do not modify the meter in any way.
- Opening the equipment and service- and repair work must only be performed by qualified service personnel.
- Measuring instruments don’t belong to children hands.
Cleaning the cabinet Clean only with a damp, soft cloth and a commercially available mild householder cleanser. Ensure that no water gets inside the equipment to prevent possible shorts and damage to the equipment. Introduction The PeakTech
5201 is a non-contact moisture meter with a high- frequency measurement method. It is used for the non-destructive testing of moisture in building materials of all kinds and for the detection of moisture distribution in walls, ceilings and floors. This practical tester is particularly suitable for pre-testing walls before further processing by plastering, papering or painting, as well as for the general assessment of the risk of mold in residential buildings or for the search for moisture damage.
Display LCD with automatic backlight Measurement Range
20 ~ 40 mm Low Battery Indication Symbol “ “ shown in the display Auto Power Off After 10 Minutes DATA-HOLD Holds measurement value in display Power Supply 3 x 1,5V AAA (UM-4) Operating Environment 0°C ~ 50°C; <80% RH Storage Environment -10°C ~ 50°C; <70% RH Size (WxHxD) 60 x 245 x 27mm Weight 220g17
5. Operating Instruction
Turn the Meter ON by pressing the button "ON", then press the MEAS button to start measuring. User press MEAS button for the first time when the meter turned on, please hold the meter up so that the ball sensor is in the air and not covered by your hand, any surface or object for 8 to 10 cm distance. The LCD will display "CAL" indication and then moisture value will be displayed in LCD. The value must below 0,5. If not, please turn the meter on over again and repeat hereinbefore operation. Then the meter start to measuring the object. Press the MEAS button again, the moisture measurement shall latch on the display for 30 sec. HOLD symbol will displayed. After 30 sec. the meter will turns off automatically. Note: The hand of the user has an influence on the measurement result. If you hold the device naturally in your hand and turn it on, calibration will be performed to zero (zeroing) to remove this effect. If you turn off the meter, it will be recalibrated after powering up again. If you change the position in the hand to a greater extent or if you change from one location to another, you should also perform a new calibration (zeroing) by switching it off and on again.20 Moisture Table of Comparison Construction Material Display (Unit) Moisture Status Gypsum <30 DRY 30-60 RISK >60 WET Cement <25 DRY 25-50 RISK >50 WET Wood <50 DRY 50-80 RISK >80 WET
Note: If the ball head in corners, it maintain a minimum distance of 8 to 10cm from corner areas. The ball head must be kept perpendicular to the measured material during the measurement, pressed firmly against the surface and not slanted. Metal parts in the surface will falsify the measurement values. Measure only in places that have the most homogeneous possible material composition In the case of material thickness of les than 20mm, there is the danger that moisture value. maybe the values is not correct. If “BAT” appears in the display, the battery should be replaced.21
Press SET button will enter the alarm set mode when the meter is already in data hold mode, the "RISK"-Icon appears on the LCD, You can use the UP and DOWN button to adjust the threshold. Press the "SET" button shall save the alarm setting. Then “WET” icon appears on the LCD, you can use the UP and DOWN button to adjust the threshold. Press the SET button shall save the alarm setting and then the meter back to the data hold Mode, if the moisture measurement is higher than the RISK alarm setting. It shall beep once every 2 sec. If moisture measurement exceeded the “WET” alarm setting, it shall beep four time every 1 sec. Note: The RISK value is adjust 0 to 50. The default value is 30. The WET value is adjust 50 to 100. The default value is 60.
Backlight Press UP button to switch the backlighting on or off.22
If the sign “BAT” appears on the LCD display, it indicates that the battery should be replaced. Remove the back cover and open the battery compartment. Replace the exhausted battery with new battery. Batteries, which are used up dispose duly. Used up batteries are hazardous and must be given in the - for this being supposed - collective container.
1. Keep the instrument dry.
2. Keep the probes clean.
3. Keep the instrument and battery out of reach of infant and
4. When the symbol ” ” appears, the battery is low and should
be replaced immediately. When you install battery, ensure the polarity connections are correct. If you will not use the instrument in a long period of time, remove the battery.23
7.1. Notification about the Battery Regulation The delivery of many devices includes batteries, which for example serve to operate the remote control. There also could be batteries or accumulators built into the device itself. In connection with the sale of these batteries or accumulators, we are obliged under the Battery Regulations to notify our customers of the following: Please dispose of old batteries at a council collection point or return them to a local shop at no cost. The disposal in domestic refuse is strictly forbidden according to the Battery Regulations. You can return used batteries obtained from us at no charge at the address on the last side in this manual or by posting with sufficient stamps. Contaminated batteries shall be marked with a symbol consisting of a crossed-out refuse bin and the chemical symbol (Cd, Hg or Pb) of the heavy metal which is responsible for the classification as pollutant:
1. "Cd" means cadmium.
2. "Hg" means mercury.
3. "Pb" stands for lead.24
All rights, also for translation, reprinting and copy of this manual or parts are reserved. Reproductions of all kinds (photocopy, microfilm or other) only by written permission of the publisher. This manual is according the latest technical knowing. Technical alterations reserved. We herewith confirm that the unit is calibrated by the factory according to the specifications as per the technical specifications. We recommend to calibrate the unit again, after one year. © PeakTech
EasyManual