Testo 625 - Measuring equipment

625 - Measuring equipment Testo - Free user manual and instructions

Find the device manual for free 625 Testo in PDF.

📄 56 pages English EN 💬 AI Question
Notice Testo 625 - page 20
View the manual : Français FR Deutsch DE English EN

User questions about 625 Testo

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

L'email reste privé : il sert seulement à vous prévenir si quelqu'un répond à votre question.

No questions yet. Be the first to ask one.

Download the instructions for your Measuring equipment in PDF format for free! Find your manual 625 - Testo and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. 625 by Testo.

USER MANUAL 625 Testo

testo 625 Humidity/temperature measuring instrumentGeneral notes20 General notes This chapter provides important advice on using this documentation. The documentation contains information that must be applied if the product is to be used safely and efficiently. Please read this documentation through carefully and familiarise yourself with the operation of the product before putting it to use. Keep this document to hand so that you can refer to it when necessary. Identification Representation Meaning CommentsNote Offers helpful tips and information. ±, 1, 2 Objective Denotes the objective that is to be achieved via the steps described. Wheresteps are numbered, you must alwaysfollow the order given! Condition A condition that must be met if an actionis to be carried out as described. i, 1, 2, ... Step Carry out steps. Where steps are numbered, you must always follow the order given! Text Display text Text appears on the instrument display. Control button Press the button. - Result Denotes the result of a previous step. º Cross-reference Refers to more extensive or detailed information. Instruments wwith rradio mmodule 9915.00MHz FFSK: NNOOTTEE:: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful inter- ference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is needed.
  • Consult the dealer or an experienced radio/TV technician for help. Button1. Safety advice 21

This chapter gives general rules which must be followed and observed if the product is to be handled safely. Avoid personal injury/damage to equipment i Do not use the measuring instrument and probes to measure on or near live parts. i Never store the measuring instrument/probes together with solvents and do not use any desiccants. Product safety/preserving warranty claims i Operate the measuring instrument only within the parameters specified in the Technical data. i Always use the measuring instrument properly and for its intended purpose. Do not use force. i Do not expose handles and feed lines to temperatures in excess of 70 °C unless they are expressly permitted for higher temperatures. Temperatures given on probes / sensors relate only to the measuring range of the sensors. i Open the instrument only when this is expressly described in the documentation for maintenance and repair purposes. Carry out only the maintenance and repair work that is described in the documentation. Follow the prescribed steps when doing so. For safety reasons, use only original spare parts from Testo. Ensure correct disposal i Take faulty rechargeable batteries/spent batteries to the collection points provided for them. i Send the product back to Testo at the end of its useful life. We will ensure that it is disposed of in an environmentally friendly manner.

This chapter gives the areas of application for which the product is intended. Use the product only for those applications for which it was designed. Ask Testo if you are in any doubt. testo 625 is a compact measuring instrument for measuring humidity and temperatures by means of a plug- in humidity/temperature probe (humidity module) and/or a humidity/temperature probe with radio transmission (radio module accessory part required). The product was designed for the following tasks/applications:

  • Measuring room climate
  • Buildings, offices, warehouses The product should not be used in the following areas:
  • Areas at risk of explosion.
  • Diagnostic measurements for medical purposes3. Production description 23

3. Product description

This chapter provides an overview of the components of the product and their functions.

3.1 Display and control

elements Overview Probe socket Display Control buttons Battery compartment (rear) Radio module and service compartment (rear) Button functions Button FunctionsSwitch instrument on; switch instrument off (press and hold)Switch display light on / offKeep reading, displaymaximum/minimum valueOpen/leave configuration mode (pressand hold); In configuration mode:Confirm inputIn configuration mode:Increase value, select optionIn configuration mode:Reduce value, select optionChange between displaying relativehumidity, dew point and wetbulbtemperature.Change between displaying connectedprobe and radio probe ( lit).

  • 4 segments in the battery symbol are lit: Instrument battery is fully charged
  • No segments in the battery symbol are lit: Battery is almost spent Measurement channel: Radio probe (the number of “radio wave” segments shown indicates the strength of the signal) Battery capacity of radio probe (above the radio probe symbol : Radio probe battery is almost spent

Probe socket A plug-in measuring probe can be connected via the probe socket on the head of the instrument. Radio module (accessory part) Radio probes may only be used in countries in which they have been Type Approved (see application infor- mation of the radio probe). A radio measuring probe can be connected via the radio module.

Voltage is supplied by means of a 9V monobloc battery (included in delivery) or rechargeable battery. It is not possible to run the instrument from the mains supply or charge a rechargeable battery in the instrument.4. Commissioning 25

This chapter describes the steps required to commission the product.

Inserting aa rradio mmodule ((accessory ppart): Radio probes may only be used in countries in which they have been Type Approved (see application infor- mation of the radio probe). The instrument is switched off. 1 To open the radio module compartment on the rear of the instrument, push the clip lock downwards and remove the lid of the radio module compartment. 2 Insert the radio module. 3 To close the radio module compartment, replace the radio module compartment lid and close it.

Inserting aa bbattery/rechargeable bbattery: 1 To open the battery compartment on the rear of the instrument, push the lid of the battery compartment in the direction of the arrow and remove it. 2 Insert a battery/rechargeable battery (9V monobloc). Observe the polarity! 3 To close the battery compartment, replace the lid of the battery compartment in position and push it against the direction of the arrow.

This chapter describes the steps that have to be executed frequently when using the product.

5.1 Connecting a probe

Plug-in probes Plug-in probes must be connected before the measuring instrument is switched on so that they are recognised by the instrument. i Insert the connector of the probe into the probe socket. Radio probes Radio probes may only be used in countries in which they have been Type Approved (see application infor- mation of the radio probe). A radio module (accessory part) is required for the use of radio probes. The radio module must be connected before the measuring instrument is switched on so that it is recognised by the instrument. Each radio probe has a probe ID (identification number). This must be set in configuration mode. º See the chapter PERFORMING SETTINGS.

Switching tthe iinstrument oon: i Press . - Measurement view is opened: The current reading is displayed, or ----- lights up if no reading is available.5. Operation 27

Switching tthe iinstrument ooff: i Press and hold (for approx. 2s) until the display goes out.

5.4 Performing settings

To oopen cconfiguration mmode: The instrument is switched on and is in measurement view. Hold, Max or Min are not activated. i Press and hold (for approx. 2s) until the display changes. - The instrument is now in configuration mode. You can change to the next function with . You can leave configuration mode at any time. To do so, press and hold (for approx. 2s) until the instrument has changed to measurement view. Any changes that have already been made in configuration mode will be saved.

Carrying oout aa hhumidity ccalibration: A 2-point calibration of connected humidity probes can be carried out (11.3%RH and 75.3%RH). Configuration mode is opened, CAL is lit. 1 Select the desired option with / and confirm with :

  • oFF: Humidity calibration is not carried out.
  • on: Humidity calibration is carried out. oFF was selected: º Continue with objective TO REGISTER THE RADIO PROBE.

fresitptsvnl????5. Operation28 on was selected: 2 Put the humidity probe into the reference medium and wait for the equalisation period to elapse. - The current humidity reading and the calibration point (nominal value) are displayed. 3 Start the calibration menu with . 4 Select the desired option with / and confirm with :

  • no: Humidity value is not calibrated.
  • YES: Humidity value is calibrated. no was selected: º Continue with objective TO REGISTER THE RADIO PROBE. YES was selected: - The calibration is carried out. 5 Repeat steps 2 to 4 for the 2

calibration point. - When calibration has been completed, the instrument changes to the next setting function.

To rregister tthe rradio pprobe: Radio probes may only be used in countries in which they have been Type Approved (see application infor- mation of the radio probe). The setting function for radio probes is only available if a radio module (accessory part) is inserted into the measuring instrument. º See the chapter COMMISSIONING. If no radio module is inserted: º Continue with the objective TO SET AUTO OFF. Each radio probe has a probe ID (RF ID). This consists of the last 3 digits of the serial no. and the position of the slide switch in the radio probe (

Configuration mode is opened, is flashing and AUTO is lit. The radio probe is switched on and the transfer rate is set to 2 readings per second (see the advice on using the radio probe).5. Operation 29

Select the desired option with and confirm with

  • YES: Switches automatic probe detection on (recommended).
  • no: Switches automatic probe detection off. no was selected: 2 Use / to set the probe ID manually and confirm with . º Continue with objective TO SET AUTO OFF. YES was selected: - Automatic probe detection is started. - Once a radio probe is found, the probe ID is displayed. If no probe is found, NONE lights up. If more than one radio probe is switched on: i If necessary, select the desired probe with / . Possible causes why probes are not found:
  • The radio probe is not switched on or the battery of the radio probe is spent.
  • The radio probe is outside the range of the measuring instrument.
  • Sources of interference are influencing the radio transmission (e.g. reinforced concrete, metal objects, walls or other barriers between transmitter and receiver, other transmitters of the same frequency, strong electromagnetic fields). i If necessary, rectify the possible causes for the disruption to the radio transmission and start configuration mode again. 2 Press to change to the next function.

To sset AAuto OOff: Configuration mode is opened, AutoOff is flashing. i Select the desired option with / and confirm with :

  • on: The measuring instrument switches off automatically if no button is pressed for 10min (Hold or Auto Hold is lit).
  • oFF: The measuring instrument does not switch itself off automatically.

To sset tthe uunit oof mmeasurement: Configuration mode is opened, UNIT is lit. i Select the desired unit of measurement with / and confirm with .

To rreset: Configuration mode is opened, RESET is lit. i Select the desired option with / and confirm with :

  • no: Instrument is not reset.
  • Yes: Instrument is reset. The instrument is reset to the factory settings. - The instrument returns to measurement view.6. Measuring 31

This chapter describes the steps that are required to perform measurements with the product.

Taking aa mmeasurement: The instrument is switched on and is in measurement view. i Put the probe in position and read off the readings.

Changing tthe mmeasurement cchannel ddisplay: i To change between displaying the connected probe and the radio probe ( ): Press . i To change between displaying relative humidity (%), the calculated dew point temperature (td °C, below 0°Ctd/32°Ftd frost point temperatures are displayed) and the calculated wetbulb temperature (wetbulb): Press .

Holding tthe rreading, ddisplaying tthe mmaximum/minimum value: The current reading can be recorded. The maximum and minimum values (since the instrument was last switched on) can be displayed. i Press several times until the desired value is displayed. - The following are displayed in turn:

  • Hold: the recorded reading

Resetting tthe mmaximum/minimum vvalues: The maximum/minimum values of all channels can be reset to the current reading. 1 Press several times until Max or Min lights up. 2 Press and hold (approx. 2s). - All maximum or minimum values are reset to the current reading.

7. Care and maintenance

This chapter describes the steps that help to maintain the functionality of the product and extend its service life.

Cleaning tthe hhousing: i Clean the housing with a moist cloth (soap suds) if it is dirty. Do not use aggressive cleaning agents or solvents!

Changing tthe bbattery/rechargeable bbattery: The instrument is switched off. 1 To open the battery compartment on the rear of the instrument, push the lid of the battery compartment in the direction of the arrow and remove it. 2 Remove the spent battery/rechargeable battery and insert a new battery/rechargeable battery (9 V monobloc). Observe the polarity! 3 To close the battery compartment, replace the lid of the battery compartment in position and push it against the direction of the arrow.8. Questions and answers 33

8. Questions and answers

This chapter gives answers to frequently asked questions. Question Possible causes Possible solution is lit (bottom right · Instrument battery is · Replace instrument in display). almost spent. battery. is lit (above · Radio probe battery · Replace radio probe symbol). is almost spent. battery. Instrument switches · Auto Off function · Switch function off. itself off automatically. is switched on.

  • Residual capacity · Replace battery. of battery is too low. Display: ----- · Probe is not plugged in. · Switch instrument off, connect probe and switch instrument back on again.
  • Registered probe · Register radio probe was not found. again, see chapter PERFORMING SETTINGS, objective TO REGISTER
  • Probe break. · Please contact your dealer or Testo Customer Service. Display reacts slowly · Ambient temperature · Raise ambient is very low. temperature. Display: uuuuu · Permitted measuring · Keep to permitted range was undershot. measuring range. Display: ooooo · Permitted measuring · Keep to permitted range was exceeded. measuring range. If we are unable to answer your question, please contact your dealer or Testo Customer Service. Contact details can be found on the guarantee card or on the Internet under www.testo.com

10. Accessories/spare

Radio probes may only be used in countries in which they have been Type Approved (see application information of the radio probe). For a complete list of all accessories and spare parts, please refer to the product catalogues and brochures or look up our website: www.testo.com

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Testo

Model : 625

Category : Measuring equipment