Testo 175T3 - Measuring equipment

175T3 - Measuring equipment Testo - Free user manual and instructions

Find the device manual for free 175T3 Testo in PDF.

📄 164 pages English EN 💬 AI Question
Notice Testo 175T3 - page 34

User questions about 175T3 Testo

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

L'email reste privé : il sert seulement à vous prévenir si quelqu'un répond à votre question.

No questions yet. Be the first to ask one.

Download the instructions for your Measuring equipment in PDF format for free! Find your manual 175T3 - Testo and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. 175T3 by Testo.

USER MANUAL 175T3 Testo

Action: more steps, the sequence must be followed. > ... Action: a step or an optional step. - ... Result of an action. Menu Elements of the instrument, the instrument display or the program interface. [OK] Control keys of the instrument or buttons of the program interface. ... | ... Functions/paths within a menu. “...” Example entries2 Safety and the environment

Pos: 16 /TD/Überschr iften/3. Leistungs beschreibung @ 0\mod_11737 74791554_79.doc @ 301 @ 1 3 Specifications Pos: 17 /TD/Überschr iften/3.1 Verwendung @ 0\mod_1176211016437_ 79.doc @ 695 @ 2 3.1. Use Pos: 18 /TD/Leistungsb eschreibung/Ver wendung/testo 175 @ 6\mod_1 274950423499_79. doc @ 61883 @ 5 Data loggers testo 175 are used to store and to read out individual readings and measurement series. With testo 175 measurement values are measured, saved and transmitted to the PC via USB cable or SD card, where they can be read out and analysed with the software testo ComSoft. With the software the data loggers can also be individually programmed. Typical applications testo 175 T1 and testo 175 T2 are optimally suitable for temperature measurements in refrigerators, freezers, cold- storage rooms and cooling shelves. testo 175 T3 records two temperatures at the same time and is most suitable e.g. for monitoring the temperature spreading between feed and return flow in a heating system. testo 175 H1 controls the climatic conditions e.g. in warehouses, offices and in the production area. Pos: 19 /TD/Überschr iften/3.2 Technisc he Daten @ 0\mod_117621108843 7_79.doc @ 704 @4 Technical data

Please note that, according to EN 12830, this instrument must be regularly checked and calibrated as specified in EN 13486 (recommendation: every year) Contact us for more information.4 Technical data

Please note that, according to EN 12830, this instrument must be regularly checked and calibrated as specified in EN 13486 (recommendation: every year) Contact us for more information.4 Technical data

Sensor type NTC temperature sensor, capacitive humidity sensor Number of measuring channels 2x internal (stubs) Measuring ranges -20 to +55 °C -40 to +50 °Ctd 0 to 100 %rF (non-dewing atmosphere) System accuracy ±2%rF (2 to 98%rF) +0.03 %rF/K ± 1 digit ±0.4 °C (-20 to +55 °C) ± 1 digit Resolution 0.1 %rF, 0.1 °C Operating temperature -20 to +55 °C Storage temperature -20 to +55 °C Battery type 3x battery type AAA or Energizer L92 AAA-size cells Life 3 years (15 min. measuring cycle, +25 °C) Degree of protection IP 54 Dimensions in mm (LxWxH) 149 x 53 x 27 mm Weight 130g Housing ABS/PC Measuring cycle 10s - 24h (freely selectable) Interface Mini-USB, SD card slot Memory capacity 1 million readings Warranty 24 years, warranty conditions: see website www.testo.com/warranty EC Directive 2004/108/EC4 Technical data

Pos: 24 /TD/Leistungsb eschreibung/ Technische Daten/Hin weis Batteriestandzei t @ 6\mod_1274958762415_ 79.doc @ 62012 @ 5 Battery life The programming windows of the software provide you with typical guide values for the expected lifetime of the battery. This lifetime is calculated on the basis of the following factors:

  • Number of connected sensors Since the battery life depends on quite a few other factors, the calculated data can only serve as guide values. The following factors have a negative effect on the battery life:
  • longer flashing of the LEDs
  • frequent reading out (several times per day) via the SD-card
  • extreme fluctuations in operating temperature The following factors have a positive effect on the battery life:
  • display switched off The battery capacity reading in the display of the data logger is based on the calculated values. However, the data logger is switched off when a critical voltage level has been reached. It may therefore happen that:
  • readings are still recorded, even though the battery capacity reading says "empty".
  • the measurement program is stopped, even though the battery capacity reading just before indicated a still remaining battery capacity. In case of an empty battery or a battery change saved readings will not be lost.5 First steps

1. Open the lock with the key (1).

2. Remove the lock (2) from the locking pin.

3. Pull the locking pin (3) out of the holes in the wall

In order to reach the battery life under application temperatures below -10 °C you should use Energizer L92 AAA-size cells.

1. Place the data logger on its front.

2. Loosen the screws on the back of the data logger.

3. Remove the battery compartment cover.5 First steps

5. Place the battery compartment cover on the battery

6. Tighten the screws.

The instructions for the installation and operation of the software can be found in the testo ComSoft 5 Basic instruction manual, which can be downloaded together with the software. The software can be ordered on CD (article-no: 0572 0580), if the download from the Internet is not desired. For testo ComSoft Professional und testo ComSoft CFR: > Insert the CD into the CD-ROM drive.

1. Install the software testo ComSoft.

2. Connect the USB cable to a free USB port on the PC.

3. Loosen the screw on the right side of the data logger.

The display function can be switched on/off via the software testo ComSoft. Depending on the operating status, various information may be shown in the display. A detailed representation of the information that can be called up can be found under Menu overview, page 48. For technical reasons the display speed of liquid crystal displays becomes slower at temperatures below 0 °C (approx. 2 seconds at -10 °C, approx. 6 seconds at -20 °C). This has no influence on the measuring accuracy. testo 175 T1

1 Highest saved reading 2 Lowest saved reading 3 Reading 4 Units 5 Measurement program over 6 Measurement program running 7 Wait for start of measurement program 8 Start criterion Date/ Time programmed 9 Battery capacity6 Display and control elements

  • Flashes: programmed alarm value is shown
  • Lights: programmed alarm values were fallen short
  • Flashes: programmed alarm value is shown
  • Lights: programmed alarm values were exceeded testo 175 T2, testo 175 T3, testo 175 H1

1 Reading channel 1 2 Units channel 1 3 Reading channel 2 4 Units channel 26 Display and control elements

5 Measurement program over 6 Measurement program running 7 Wait for start of measurement program 8 Start criterion Date/ Time programmed 9 Battery capacity Icon Capacity >151 days <150 days <90 days <60 days <30 days > Read out data and change battery (see Reading out measurement data, page

  • Flashes: programmed alarm value is shown
  • Lights: programmed alarm values were fallen short
  • Flashes: programmed alarm value is shown
  • Lights: programmed alarm values were exceeded 12 Lowest saved reading 13 Highest saved reading 14 Lower limit value channel 1:
  • Flashes: programmed alarm value is shown
  • Lights: programmed alarm values were fallen short
  • Flashes: programmed alarm value is shown
  • Lights: programmed alarm values were exceeded6 Display and control elements

Representation Explanation Red LED flashes once every 10 seconds Remaining battery capacity has dropped below 30 days Red LED flashes twice every 10 seconds Remaining battery capacity has dropped below 10 days Red LED flashes three times every 10 seconds Battery is empty: Red LED flashes three times when pressing the button Limiting value was exceeded/fallen short of Yellow LED flashes three times Instrument changes from Wait- mode to Rec-mode. Yellow LED flashes three times when pressing the button Instrument is in Rec-mode Green and yellow LED flash three times when pressing the button. Instrument is in End-mode Green LED flashes three times when pressing the button Instrument is in Wait-mode Green LED flashes five times when pressing the button Long pressing of the GO button has set a time mark. Green, yellow and red LED flash in succession The battery has been changed.

6.3. Key functions A detailed representation of the display readings can be found under Menu overview, page 48. ✓ Instrument in operating status Wait and start criterion Button start programmed. > Press [GO] for approx. 3 seconds to start the measurement program. - The measurement program starts and Rec appears in the display. ✓ Instrument is in operating status Wait: > Press [GO] in order to change between displays of upper alarm value, lower alarm value, battery life and last reading. The displays appear in the specified sequence. ✓ Instrument is in operating status Rec or End: > Press [GO] in order to change between displays of highest saved reading, lowest saved reading, upper alarm value, lower alarm value, battery life and last reading. The displays appear in the specified sequence. Time mark The time mark function enables you to read out the memory contents starting from a defined time without reprogramming the logger. The readings starting with the start of recording are saved parallel to this. ✓ Instrument is in operating status REC: > Hold [GO] depressed for about 3 seconds to set a time mark.

Only one time mark can be set. When [GO] is pressed again, the existing time mark is deleted and a new one is set. - The green LED flashes five times. - The display only shows the readings starting from the set time mark.7 Using the product

Pos: 31 /TD/Überschr iften/6. Produkt ver wenden @ 0\mod_117377492855 4_79.doc @ 328 @ 1 7 Using the product Pos: 32 /TD/Erste Sc hritte/testo 175/Fühler anschließen @ 6\mod_12749 57942556_79.doc @ 61981 @ 7.1. Connecting a sensor Observe the following points when connecting sensors to data logger and measuring points. > Ensure correct polarity of the plugs. > Press the plugs firmly into the ports to ensure leak tightness. However, do not apply force! > Make sure that the plugs are firmly connected to the data logger or that the connections are closed with blanking plugs. > Ensure correct positioning of the sensor to avoid disturbing influences affecting the measurement. > testo 175 T3: Always make sure that you connect the sensor configured (via the software testo ComSoft) to the individual sockets. The numbers of the connections are printed on the housing. Pos: 33 /TD/Produkt verwenden/testo 174-neu/ Hinweis Datenlogger programmieren @ 5\mod_126 4769275750_79.doc @ 56853 @ 7.2. Programming data logger In order to adapt the programming of your data logger to your individual requirements, you require the testo ComSoft 5 Basic software. It is available in the Internet as a free download requiring registration www.testo.com, International, Service&Support | Download Center. The instructions for the installation and operation of the software can be found in the testo ComSoft 5 Basic instruction manual that is downloaded together with the software.

The menu overview shows exemplary display representations of the data logger testo 175-T2. The display must be switched on to be able to show the corresponding indications. This is accomplished with the software testo ComSoft. The indication in the display is updated according to the programmed measurement rate. Only readings from active channels are displayed. The channels are also activated via the software testo ComSoft. The symbols for upper or lower alarm value light up in operating states Rec and End, if the programmed alarm value has been exceeded or fallen short off. After 10 seconds without operating a key the display will return to its initial state.7 Using the product

Wait mode: Start criterion is programmed, but not yet fulfilled. ① Last reading

③ Lower alarm value ④ Battery capacity in days

Measurement value is not saved7 Using the product

Rec mode: Start criterion was fulfilled, data logger saves readings End mode: Measurement program finished (stop criterion reached – memory full or number of readings) depending on programming ① Last reading ② Highest reading

③ Lowest reading ④ Upper alarm value

⑤ Lower alarm value ⑥ Battery capacity in days

Last reading (see Fig. ①)

The scope of delivery does not include mounting materials (e.g. screws, wall plugs). ✓ The data logger has been removed from the wall bracket.

1. Position the wall bracket at the desired place.

2. Use a pen or something similar to mark the location for

the fastening screws.

3. Prepare the fastening location in accordance with the

fastening material (e.g. drill hole, insert wall plugs).

4. Fasten the wall bracket with suitable screws.

✓ The wall bracket has been mounted.

1. Slide the data logger into the wall bracket (1).

2. Push the locking pin (2) through the holes in the wall

3. Fasten the lock (3) on the locking pin.

7.6. Reading out measurement data

The measurement data remain stored in the data logger after they have been read out and can therefore be read out several times. The measurement data will only be deleted when the data logger is reprogrammed.

1. Connect the USB cable to a free USB port on the PC.

2. Loosen the screw on the right side of the data logger.

Use a coin for this purpose.

5. Read out the data logger and process the read out

data, see separate operating instructions testo ComSoft. Via SD card

1. Loosen the screw on the right side of the data logger.

Use a coin for this purpose.

3. Push the SD card into the SD card slot (2).

4. Hold [Go] depressed for longer than 2 seconds.

- The display shows CPY (testo 175 T1) or COPY (testo 175 T2, testo 175 T3, testo 175 H1). - The yellow LED lights during the copying process. - The green LED flashes twice and after the copying process the display shows OUT.

5. Remove the SD card.

6. Insert the SD card into the SD card slot on the PC.

7. Process the read out data, see separate operating

The battery change stops the currently running measurement program. However, stored measurement data are preserved.

1. Read out stored measurement data, see Reading out

measurement data, page 52. ✓ If it is no longer possible to read out the saved measurement data because the battery capacity is too low: > Change the batteries and read out the stored measurement data after.

2. Place the data logger on its front.

3. Loosen the screws on the back of the data logger.

4. Remove the battery compartment cover.8 Maintaining the product

5. Take the empty batteries out of the battery

6. Insert three new batteries (type AAA). Observe the

Only use new branded batteries. If a partially exhausted battery is inserted, the calculation of the battery capacity will not be performed correctly. In order to reach the battery life under application temperatures below -10 °C you should use Energizer L92 AAA-size cells.

7. Place the battery compartment cover on the battery

8. Tighten the screws.

The data logger needs to be reconfigured. For this purpose the software testo ComSoft must be installed on the computer and a connection to the data logger must be set up.

9. Connect the data logger to the PC with a USB cable.

10. Start the software testo ComSoft and set up a

connection to the data logger.

9 Tips and assistance Pos: 42 /TD/Überschr iften/8.1 Fragen und Ant worten @ 0\mod_117740201 7078_79.doc @ 1093 @ 9.1. Questions and answers Pos: 43 /TD/Tipps und Hilf e/Fragen und Antworten/ testo 175 @ 6\mod_12754710 58687_79.doc @ 62116 @ Question Possible causes / solution FULL appears in the display, the red LED flashes twice, out appears in the display. Insufficient memory capacity on SD card to save the data. > Remove the SD card, free up more memory space and copy data. Err appears in the display, the red LED flashes twice, out appears in the display. An error occurred while saving data to the SD card. > Remove the SD card, free up more memory space and copy data. nO dAtA appears in the display, the red LED flashes twice. The logger has not yet recorded any data and is in Wait mode. > Remove the SD card and wait until the logger is in Rec mode. rST appears in the display. The battery was changed. No data are recorded. > Reprogram the data logger via the software.

SD card to read out the data logger 175 0554 1700 Batteries (alkaline-manganese AAA- size cells) for applications down to -10 °C

Batteries (Energizer L92 AAA-size cells) for applications down to -10 °C

CD testo ComSoft 5 Basic (if free-of- charge, registered download from website is not wanted)

For further accessories and spare parts, please refer to the product catalogues and brochures or look up our website: www.testo.com1 Sommaire

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Testo

Model : 175T3

Category : Measuring equipment