Source : WHO, 1999 (World Health Organization)
Electromagnetic compatibility Electromagnetic emissions Guidance and manufacturer’s declaration – electromagnetic emissions for all EQUIPMENT and SYSTEMS The Blood pressure monitor (BM 54) is intended for use in the electromagnetic environment specified below. The customer of the user of the Blood pressure monitor (BM 54) should assure that it is used in such and environment. Emission test
Electromagnetic environment – guidance
RF emissions CISPR 11
The Blood pressure monitor (BM 54) uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emission CISPR 11
Class B The Blood pressure monitor (BM 54) is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Electromagnetic immunity Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS The Blood pressure monitor (BM 54) is intended for use in the electromagnetic environment specified below. The customer of the user of the Blood pressure monitor (BM 54) should assure that it is used in such and environment. Immunity test
IEC 60601 test level
Electromagnetic environment – guidance
Electrostatic discharge (ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic material, the relative humidity should be at least 30%.
Power frequency (50Hz) magnetic field IEC 61000-4-8
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. 128
Guidance and manufacturer’s declaration – electromagnetic immunity for EQUIPMENT and SYSTEMS that are not LIFESUPPORTING The Blood pressure monitor (BM 54) is intended for use in the electromagnetic environment specified below. The customer or the user of Blood pressure monitor (BM 54) should assure that it is used in such an environment. Immunity test
IEC 60601 test level
Electromagnetic environment – guidance Portable and mobile RF communications equipment should be used no closer to any part of the Blood pressure monitor (BM 54), including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance.
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2.5 GHz
3 V/m 80 MHz to 800 MHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. a
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Blood pressure monitor (BM 54) is used exceeds the applicable RF compliance level above, the Blood pressure monitor (BM 54) should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Blood pressure monitor (BM 54). b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT or SYSTEM that are not LIFESUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the Blood pressure monitor (BM 54). The Blood pressure monitor (BM 54) is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Blood pressure monitor (BM 54) can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Blood pressure monitor (BM 54) as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter (W)
Separation distance according to frequency of transmitter (m) 80 MHz to 800 MHz
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. 131
Beurer GmbH • Söflinger Straße 218 • 89077 Ulm, Germany • www.beurer.com www.beurer-blutdruck.de • www.beurer-gesundheitsratgeber.com • www.beurer-healthguide.com 132
655.12_BM54_2018-02-27_01_IM1_BEU The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Beurer GmbH is under license. Other trademarks and trade names are those of their respective owners. Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc., registered in the U.S. and other countries. Google Play and the Google Play logo are trademarks of Google LLC. Android is a trademark of Google LLC.