31.2001 - Sound Level Meter TFA - Free user manual and instructions
Find the device manual for free 31.2001 TFA in PDF.
| Product Type | Sound Level Meter (Sonimètre) |
| Brand | TFA |
| Model | 31.2001 |
| Standards Applied | IEC61672-1 Class2, ANSI S1.4 Type2 |
| Frequency Range | 31.5 Hz – 8 kHz |
| Measuring Range | 32 – 130 dB |
| Microphone | 1/2 inch electret condenser microphone |
| Display | 4-digit LCD with 0.1 dB resolution |
| Display Update Rate | 0.5 seconds |
| Frequency Weighting | A |
| Time Weighting | FAST |
| Accuracy | ±1.4 dB (94 dB @ 1 kHz reference) |
| Alarm Indication | OVER (above range) / UNDER (below range) |
| Power Supply | Two IEC R03 (AAA) batteries, 1.5V each |
| Battery Life | Approx. 50 hours (alkaline) |
| Operating Temperature | 0°C to 40°C |
| Operating Humidity | 10% to 90% RH |
| Storage Temperature | -10°C to 60°C |
| Storage Humidity | 10% to 75% RH |
| Dimensions (L x W x H) | 206 x 42 x 25 mm |
| Weight (with batteries) | 100 g (0.22 lbs) |
| Accessories Included | Two 1.5V batteries, instruction manual, screwdriver, windscreen |
| Special Functions | Autoranging, Max Hold recording, internal calibrator (94 dB @ 1 kHz) |
| Maintenance | Wipe case with dry cloth; do not use abrasives or solvents |
| Safety & Repairability | CE compliant; repairs only by qualified personnel using specified parts |
Frequently Asked Questions - 31.2001 TFA
User questions about 31.2001 TFA
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Sound Level Meter in PDF format for free! Find your manual 31.2001 - TFA and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. 31.2001 by TFA.
USER MANUAL 31.2001 TFA
Read the following safety information carefully before attempting to operate or service the meter.
Use the meter only as specified in this manual; otherwise, the protection provided by the meter may be impaired.
Environment conditions
β Altitude up to 2000 meters
β Relatively humidity 90% max.
β Operation Ambient 0 \~ 40°C
Maintenance & Clearing
β Repairs or servicing not covered in this manual should only be performed by qualified personnel.
Periodically wipe the case with a dry cloth. Do not use abrasives or solvents on this instruments.
Safety symbols
CE Comply with EMC
When servicing, use only specified replacement parts.
II. GENERAL DESCRIPTION
Thank you for using our Sound Level Meter. To ensure that you can get the most from it, we recommend that you read and follow the manual carefully before use.
This unit conforms to the IEC61672-1 Class2, ANSI S1.4 Type2 for Sound Level Meters.
This Sound Level Meter has been designed to meet the measurement requirements of safety Engineers, Health, Industrial safety offices and sound quality control in various environments.
III. FEATURES
β Mini size
β Easy to use
β High performance at a low cost
β Autoranging
β Maximum recording
β 4-digit LCD with a resolution of 0.1 dB
β Internal Calibrator (94dB@1kHz)
IV. SPECIFICATIONS
Standard applied: IEC61672-1 Class2, ANSI S1.4 Type2
Frequency range: 31.5Hz \~ 8KHz
Measuring level range: 32 \~ 130dB
Microphone: 1/2 inch electret condenser microphone
Display: LCD
Resolution: 0.1dB
Display Update: 0.5 sec.
Frequency Weighting: A
Time Weighting: FAST
Accuracy: ±1.4dB (under reference conditions, 94dB @ 1kHz)
Alarm function: "OVER" when input is more than upper limit of range.
"UNDER" when input is Less than lower limit of range.
Power supply: Two IEC R03 (size "AAA") batteries, 1.5V X2.
Battery Life: Approx. 50hrs (alkaline battery LR03)
Operation temperature: 0 to 40°C (32 to 104°F )
Operation humidity: 10 to 90%RH
Storage temperature: -10 to 60°C (14 to 140°F)
Storage humidity: 10 to 75%RH
Dimensions: 206 (L) × 42 (W) × 25 (H)mm
8. 1 ^ (L) × 1. 7 ^ (W) × 1. 0 ^ (H)
Weight: 100g (0.22 lbs) (including batteries)
Accessories: two 1,5V batteries, instruction manual, screwdriver, windscreen.
V. NAME AND FUNCTIONS

(1) Windscreen
When making outdoor measurements in windy weather, wind noise at he microphone can cause measurement errors. Such effects can be using the windscreen.
(2) Display

(3) Power On/Off & Max Hold button
Power On
Turn the unit on by press the "Max Hold" button.
Max recording mode
Press the “Max Hold” button to enter the Max Hold mode under the power-on. The LCD will show “MAX” symbol. The MAX recording mode captures and stores the maximum input values detected. When the unit detects an input that is above the recorded maximum, the unit recods the new maximum value.
Press the “Max Hold” button again to exit the MAX mode. Recorded values are erased, “MAX” symbol disappears.
Power-off
Turn the unit off by holding down the "Max Hold" button for at least 2 seconds. The LCD will show "OFF", release the button.
(4) Microphone
1/2 inch electret Condenser microphone
(5) Measurement / Calculation
Measurements or internal calibrator select switch
(6) Calibration adjustment
(7) Battery Cover
VI. MEASUREMENT PREPARATION
(1) Inserting the batteries
Lightly press the battery cover and slide it to the down. Insert two 1.5V batteries, paying attention to the polarity as indicated in the compartment. Replace the cover.
(2) Battery Replacement
When the battery voltage drops below the operating voltage, this
-+ symbol will appear.
Replace 1.5 Volt batteries.
V II. OPERATION PRECAUTIONS
(1) Wind blowing across the microphone will bring additional extraneous noise.
When using the instrument in the presence of wind, it is a must to mount the windscreen to not pick up undesired signals.
(2) Calibrate the instrument before the operation if the instrument was not in use for a long time or operated in bad environment.
(3) Do not store or operate the instrument at high temperature and high humidity environment.
(4) Keep microphone dry and avoid severe vibration.
(5) If the unit is not used for a long period of time, the batteries will have to be removed from the battery compartment, since it might leak.
VIII. CALIBRATION
Before starting a measurement, the unit must be calibrated. There are two types of calibration: internal electrical calibration and acoustic calibration using a sound calibrator. Normally, internal electrical calibration only is required.
Internal Calibrator (9dB @ 1kHz)
Lightly press the battery cover and slide it to the down. Slide the switch to the down. Verify that the Cal Mode is set to "Internal" CAL. LCD shows "CAL" approx. 1 sec. Activates the built-in oscillator (1 kHz, sine wave) for electrical calibration of the unit. When calibration is end, slide the switch to the up "MEAS" position.
IX. MEASUREMENT
(1) Turn the power ON to measure noise sound.
For general sound level measurements, A- - frequency weighting and F (FAST) time weighting (dynamic characteristics) the setting should be used.
(2) Make sure that slide the switch to the "MEAS" position on rear of the unit.
(3) Hold the instrument comfortably in hand and point the microphone at the suspected noise source.
(4) The numeric level indication shows the current measured sound level. The reading is updated twice every second.