SUH801 - Flow Meter IFM - Free user manual and instructions
Find the device manual for free SUH801 IFM in PDF.
User questions about SUH801 IFM
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Flow Meter in PDF format for free! Find your manual SUH801 - IFM and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. SUH801 by IFM.
USER MANUAL SUH801 IFM
Operating instructions
Ultrasonic flow meter without display
SUHxxx
Table of contents
1 Preliminary note.... 4
1.1 Symbols used.... 4
1.2 Warnings.... 4
1.3 Device overview 5
2 Safety instructions 6
2.1 Cybersecurity.... 6
3 Transport, handling and storage 7
4 Intended use 8
4.1 Application area 8
5 Function 9
5.1 Options for output OUT1 9
5.2 Options for output OUT2 9
5.3 IO-Link 10
6 Installation.... 11
6.1 Process connection.... 11
6.1.1 Clamp 11
6.2 Interference.... 12
6.3 Installation position.... 13
6.3.1 Recommended installation position 13
6.3.2 Non recommended installation position 13
6.4 Use in hygienic areas according to 3-A.... 14
7 Electrical connection 16
8 Operating and display elements.... 18
9 Set-up 19
10 Parameter setting 20
10.1 Adjustable parameters 20
10.2 Output configuration 22
10.2.1 Switching signal for limit value monitoring.... 22
10.2.2 Switching signal Diagnosis 23
10.2.2.1 Switching signal for flow direction.... 23
10.2.2.2 Switching signal for signal quality 23
10.2.3 Consumed quantity monitoring (totaliser function) 24
10.2.3.1 Switching signal totaliser 25
10.2.3.2 Pulse signal totaliser 25
10.2.4 Analogue signal.... 25
10.2.5 Frequency signal.... 28
10.2.6 Error behaviour of the outputs 29
10.2.7 Output off.... 30
10.3 Application configuration 31
10.3.1 Standard unit of measurement 31
10.3.2 Process value for OUT1 and OUT2 31
10.3.3 Damping 31
10.3.4 Output polarity 31
10.3.5 Low flow cut-off 32
10.3.6 Medium.... 32
10.3.7 Flow direction.... 32
10.3.8 Calibration 33
10.3.9 Zero calibration.... 33
10.3.10 Totaliser reset 34
10.3.11 Counting method of the totalisers 34
10.3.12 Reset the unit 35
10.4 Diagnostic functions 36
10.4.1 Read totaliser values 36
10.4.2 Memory 36
10.4.3 Operating hours counter 37
10.4.4 Internal temperature.... 37
10.4.5 Signal quality 37
10.5 Service functions.... 37
10.5.1 Device information 37
10.5.2 Simulation 38
11 Operation.... 39
12 Troubleshooting.... 40
12.1 Warning messages.... 40
12.2 Error messages 41
13 Maintenance, repair and disposal 42
14 Factory Settings 43
1 Preliminary note
You will find instructions, technical data, approvals and further information using the QR code on the unit / packaging or at documentation.ifm.com.
1.1 Symbols used
Requirement
Instruction
Reaction, result
bold Designation of keys, buttons or indications
→ Cross-reference without link
→ Cross-reference with link


Important note
Non-compliance may result in malfunction or interference
Information
Supplementary note
1.2 Warnings
Warnings indicate the possibility of personal injury and damage to property. This enables safe product handling. Warnings are graded as follows:

WARNING
Warning of serious personal injury
▷ If the warning is not observed, fatal and serious injuries are possible.

CAUTION
Warning of minor to moderate personal injury
▷ If the warning is not observed, minor to moderate injuries are possible.

ATTENTION
Warning of damage to property
▷ If the warning is not observed, damage to property is possible.
1.3 Device overview
| Article | Nominal width |
| SUH120 | DN15 |
| SUH220 | DN25 |
| SUH251 | DN25 |
| SUH301 | DN40 |
| SUH320 | DN40 |
| SUH420 | DN50 |
| SUH451 | DN50 |
| SUH501 | DN65 |
| SUH520 | DN65 |
| SUH601 | DN80 |
| SUH620 | DN80 |
| SUH701 | DN100 |
| SUH720 | DN100 |
| SUH801 | DN20 |
| SUH820 | DN20 |
2 Safety instructions
- The unit described is a subcomponent for integration into a system.
- The system architect is responsible for the safety of the system.
-
The system architect undertakes to perform a risk assessment and to create documentation in accordance with legal and normative requirements to be provided to the operator and user of the system. This documentation must contain all necessary information and safety instructions for the operator, the user and, if applicable, for any service personnel authorised by the architect of the system.
-
Read this document before setting up the product and keep it during the entire service life.
- The product must be suitable for the corresponding applications and environmental conditions without any restrictions.
- Only use the product for its intended purpose (→ Intended use).
- Only use the product for permissible media.
- If the operating instructions or the technical data are not adhered to, personal injury and/or damage to property may occur.
- The manufacturer assumes no liability or warranty for any consequences caused by tampering with the product or incorrect use by the operator.
- Installation, electrical connection, set-up, operation and maintenance of the product must be carried out by qualified personnel authorised by the machine operator.
- Protect units and cables against damage.
2.1 Cybersecurity
Installation
The device is suitable for operation in a secure environment according to IEC 62443-1-1.
The device was designed for operation behind a firewall.
▶ Carry out a risk assessment of the system according to IEC 62443-1-1.
▶ Take measures to ensure physical security.
Operation
▶ Observe the security functions described in the product documentation and the recommendations for their use.
Maintenance
▶ Back up system configuration and system data in accordance with your company's change management processes.
Decommissioning
▶ Ensure that no sensitive information can fall into unauthorised hands.
▶ Always reset the system settings to the factory settings before decommissioning the device.
3 Transport, handling and storage
▶ Store the device in its original packaging.
▶ When the device is to be stored again, use the original packaging.
▶ Otherwise, provide unused connections with either a mating connector or a protective cap and pack the device in suitable packaging.
▶ Observe the permissible ambient conditions for the device during storage (→ Technical data).
4 Intended use
The unit monitors liquid media.
The unit detects the flow velocity, the volume flow (volumetric flow quantity/time), the consumed quantity and the medium temperature.
4.1 Application area
Liquids with the following properties:
• Conductive water-based media with 90% water content
• Non-conductive water
• High-viscosity oils (viscosity: 30...68 mm 2 /s at 40 °C / 30...68 cSt at 104 °F)
• Examples of edible oils:
- Extra virgin olive oil
- Soya bean oil
- Sunflower oil
- Mustard oil
- Coconut oil
- Corn oil
- Peanut oil

Pressure Equipment Directive (PED):
The units comply with the Pressure Equipment Directive and are designed and manufactured for group 2 fluids in accordance with the sound engineering practice. Use of media from group 1 fluids on request.
5 Function
- The unit detects the volumetric flow based on the measuring principle of ultrasonic transit time difference.
- As additional process value the unit detects the medium temperature.
- The unit can be operated in SIO mode (standard input-output) or in IO-Link mode.
• The unit has many self-diagnostic options.
– Monitoring of the flow direction
– Monitoring of the signal quality
– Provision of warnings and error messages
- The device provides the results of its self-diagnostic via the outputs and the IO-Link interface.
- A green LED on the M12 connector indicates that the device is supplied with power.
- A simulation mode allows simplified set-up of the sensor.
5.1 Options for output OUT1
- Switching signal flow
- Switching signal temperature
- Switching signal diagnosis
– Direction of flow
– Signal quality
- Switching signal totaliser
- Pulse signal totaliser
• Frequency signal flow
• Frequency signal temperature - IO-Link
• OFF (output switched to high impedance)
5.2 Options for output OUT2
- Switching signal flow
- Switching signal temperature
- Switching signal diagnosis
– Direction of flow
– Signal quality
- Switching signal totaliser
- Pulse signal totaliser
- Analogue signal flow
• Analogue signal temperature
• Input for external totaliser reset
• OFF (output switched to high impedance)
5.3 IO-Link
IO-Link is a communication system for connecting intelligent sensors and actuators to automation systems. IO-Link is standardised in the IEC 61131-9 standard.

General information on IO-Link at io-link.ifm

Input Output Device Description (IODD) with all parameters, process data and detailed descriptions of the device at documentation.ifm.com
IO-Link offers the following advantages:
• Interference-free transmission of all data and process values
• Parameter setting in the running process or presetting outside the application
- Parameters for identifying the connected devices in the system
• Additional parameters and diagnostic functions
• Automatic backup and restore of parameter sets in case of device replacement (data storage)
- Logging of parameter sets, process values and events
• Device description file (IODD - Input Output Device Description) for easy project planning
• Standardised electrical connection
- Remote maintenance
6 Installation

CAUTION
If the medium temperature is above 50\ °C ( 122\ °F ), parts of the housing can increase in temperature to over 65\ °C ( 149\ °F ).
▷ Risk of burns.
▶ Protect the housing against contact with flammable substances and unintentional contact.

ATTENTION
No functional earthing when installed in an ungrounded pipe system (e.g. plastic pipes).
▷ Deficient operating function.
▶ Ground the device. Ground brackets for the M12 connector are available as accessories, see documentation.ifm.com.

After installation, air bubbles in the system can affect the measurement.
▶ Rinse the system after installation for ventilation.

▶ Ensure that the system is free of pressure during installation.
▶ The rules and regulations for the installation and operation of compressed air equipment must be observed.
6.1 Process connection
The SUHxxx device series has hygienic process connections located directly on the device.
6.1.1 Clamp
Depending on the design (→ Technical data at www.ifm.com), the devices have a clamp connection to DIN 32676 series A or C as process connection.
A suitable sealing ring and a hinge clamp or high-pressure clamp are required for installation. Sealing ring and clamp are not included in the scope of delivery.

Fig. 1: Clamp process connection
1: Clamp
2: Clamp connection of the pipe or adapter
3: Sealing ring 4: Clamp connection of the sensor
▶ If necessary, install the clamp adapter in the pipe.
▶ Insert the sealing rings and secure the device with a clamp. Observe the direction of flow ➞ 32.

▶ Avoid edge formation at the transition between the sensor and the pipe, as this can affect the flow profile and the measuring accuracy (→ Figure).
▶ Observe information on suitable pipe standards (→ documentation.ifm.com).


Fig. 2: Avoid edge formation in the process connection
6.2 Interference
Structures in the pipe, bends, valves, reducing pieces and the like affect the function of the unit.
▶ Adhere to the distances between sensor and interference.

Fig. 3: Interference
S: Interference
F: Direction of flow
A: Distance between interference and clamp at the inlet pipe (DN = external pipe diameter):
- 5 x DN: SUH1xxx / SUH8xx / SUH2xx / SUH3xx / SUH4xx
• 15 x DN: SUH5xx / SUH6xx / SUH7xx
B: Distance between interference and clamp at the outlet pipe (DN = external pipe diameter):
• 1 x DN: SUH1xxx / SUH8xx / SUH2xx / SUH3xx / SUH4xx
• 3 x DN: SUH5xx / SUH6xx / SUH7xx
6.3 Installation position
6.3.1 Recommended installation position
▶ Install the unit so that the measuring pipe is always completely filled.
▶ Install in front of or in a rising pipe.

If air bubbles can form in the pipe system:
▶ In case of horizontal installation, mount the sensor with the display on the side of the pipe (A).
A

B

C

Fig. 4: recommended installation position
F: direction of flow
A: horizontal installation, display on the side of the pipe.
B: horizontal installation, display on top of the pipe.
C: vertical installation.

The unit can be installed independently of the orientation if the following is ensured:
- No air bubbles can form in the pipe system.
- The pipes are always completely filled.
6.3.2 Non recommended installation position
• Directly in front of a falling pipe.
- In a falling pipe.
• Directly in front of the spout of a pipe.
• Directly in front of a valve.
• On the suction side of a pump.
- At the highest point of the pipe system.
6.4 Use in hygienic areas according to 3-A

The sensor is suited for CIP (clean in place) when installed correctly.
▶ Observe the application limits (temperature and material resistance) according to the data sheet.

Not suitable for systems that have to meet the criteria of E9.2 / 63-04 of the 3-A standard 63-04.

▶ For use according to 3-A, take note of the corresponding regulations for cleaning and maintenance.
▶ Ensure that the installation of the device in the system complies with 3-A guidelines.
▶ Use only process adapters with 3-A certification and marked with the 3-A symbol (→ Accessories at www.ifm.com).
▶ Secure clamp sensors with a suitable clamp.
▶ Use self-draining installation.
▶ To allow the medium to flow out of the process adapter, mount the device in the following installation position:
• Vertical installation in a rising pipe (A).
- or -
• Horizontal position with a slight gradient so that the medium does not come to a standstill (B).

Fig. 5: Process connection according to 3-A. 1: Minimum gradient; F = flow direction
| Type | Minimum gradient (DIN 32676 series A) | Type | Minimum gradient (DIN 32676 series C) |
| DN15 | 29° | 1/2" | --- |
| DN20 | 43° | 3/4" | 32° |
| DN 25 | 25° | 1" | 10° |
| DN40 | 49° | 1 1⁄2" | 42° |
| DN 50 | 16° | 2" | 12° |
| DN65 | 23° | 2 1⁄2" | 3° |
| DN80 | 30° | 3" | 3° |
| DN100 | 15° | 4" | 3° |
Tab. 1: Minimum gradient for use according to 3-A
7 Electrical connection

The unit must be connected by a qualified electrician.
Observe the national and international regulations for the installation of electrical equipment.
Voltage supply according to SELV, PELV.
▶ Disconnect power.
▶ Connect the unit as follows:


Fig. 6: Wiring diagram; MP: multifunction (IN, OUT, Data)
| Pin | Assignment |
| 1 | L+ |
| 3 | L- |
| 4 (MP1) | Switching signal flow Switching signal temperature Switching signal diagnosis Switching signal totaliser Pulse signal totaliser Frequency signal flow Frequency signal temperatureIO-LinkOFF (output switched to high impedance) |
| 2 (MP2) | Switching signal flow Switching signal temperature Switching signal diagnosis Switching signal totaliser Pulse signal totaliser Analogue signal flow Analogue signal temperature Input for external totaliser resetOFF (output switched to high impedance) |
Circuit examples:




1: 2 x positive switching
2: 2 x negative switching
3: 1 x positive switching / 1 x analogue
4: 1 x negative switching / 1 x analogue
8 Operating and display elements

Fig. 7: Operating status LED
The device has an LED on the M12 connector. When voltage is supplied, the LED lights green.
9 Set-up
After power on and expiry of the power-on delay time, the unit is in the normal operating mode. It carries out its measurement and evaluation functions and generates output signals according to the set parameters.
During the power-on delay time, the outputs are in the following status according to the set parameters:
• ON with normally open function (Hno / Fno)
• OFF with normally closed function (Hnc / Fnc)
• ON for detection of direction (dir. F)
• OFF for frequency output (FRQ)
• OFF for consumed quantity monitoring (ImP)
• 20 mA for current output (I)

When an IO-Link master is connected, the device automatically goes from SIO mode (standard input-output) into IO-Link mode if the port of the master is set to IO-Link mode.
10 Parameter setting
Parameters can be set before installation or during operation.

If you change parameters during operation, this will influence the function of the plant.
▶ Ensure that there will be no malfunctions in your plant.
During parameter setting the unit remains in the operating mode. It continues to monitor with the existing parameter until the parameter setting has been completed.
The device parameters can be set via the IO-Link interface in the following ways, for example:
- Parameter setting via a suitable parameter setting software, e.g. ifm moneo|configure
• Parameter setting via a PLC
• Parameter setting via an IIoT application
Requirements for parameter setting via the IO-Link interface:
√ The Input Output Device Description (IODD) for the device in case of parameter setting via a parameter setting software, see documentation.ifm.com
√ The IO-Link interface description (PDF) for the device in case of parameter setting via a PLC or IIoT application, see documentation.ifm.com
√ An IO-Link master
▶ Connect the IO-Link master to the parameter setting software, the PLC or the IIoT application.
▶ Connect the device to a suitable free port of the IO-Link master.
▶ Set the port of the IO-Link master to the IO-Link operating mode.
▷ The device changes to the IO-Link mode.
▶ Change the parameter settings in the software.
▶ Write the parameter settings to the device.

Support for system integration and parameter setting via IO-Link:
→ Manual of the parameter setting software (e.g. moneo)
Explanations and startup packages at ifm.com/cnt/io-link-system-integration.
10.1 Adjustable parameters
| Parameter | Explanation |
| AEP2 | Analogue end point for OUT2 = process value at which the output signal is 20 mA. |
| ASP2 | Analogue start point for OUT2 = process value at which the output signal is 4 mA. |
| APPL | Application reset (reset of application-specific parameter settings) |
| BtB | Back-to-Box reset (reset to factory settings) |
| CGA | Calibration factor in % for adapting the measured value curve to the application |
| coF | Correction factor for zero point calibration |
| dAP. F | Damping constant in seconds for flow (63 % rise time τ) |
| dFUx | Switching signal for diagnostic output OUTx: direction of flow (Fdir) or signal quality (Sig. Q) |
| DIn2 | Reset signal for external totaliser reset |
| Fdir | Direction of flow |
| FEP1 | Frequency end point for OUT1 = Upper measured value at which the frequency signal set under FrP1 is provided. |
| FHx | Upper limit for switching signal OUTx with window function |
| FLx | Lower limit for switching signal OUTx with window function |
| FOUx | Behaviour of output OUTx in case of an error |
| FProx | Counting method of the totaliser: consideration of the direction of flow |
| FrP1 | Frequency signal which is provided when the upper measured value (MEW or FEP1) is reached. |
| FSP1 | Frequency start point for OUT1 = Lower measured value from which a frequency signal is provided (only for temperature measurement). |
| Hi. F | Highest flow value measured |
| Hi. T | Maximum temperature value measured |
| ImPRx | Totaliser function: pulse signal (ImPR = YES) or switching signal (ImPR = NO) |
| ImPSx | Pulse value (= flow value at which 1 pulse is provided) |
| LFC | Low flow cut-off (= flow value below which flow is evaluated as standstill) |
| Lo. F | Lowest flow value measured |
| Lo. T | Minimum measured temperature value |
| MEdi | Selection of the medium to be monitored |
| oux | Output configuration for output OUTx (e.g. switching output with hysteresis function) |
| P-n | Output polarity for the switching outputs |
| rPx | Reset point for switching output OUTx with hysteresis function |
| rTox | Setting for the totaliser reset (manually or time-controlled) |
| SELx | Process value for output OUTx |
| S. FLW | Simulated flow value in simulation mode |
| S. On | Starts the simulation mode |
| SPx | Switch point for switching output OUTx with hysteresis function |
| S. Tim | Duration of the simulation in minutes |
| S. TMP | Simulated temperature value in simulation mode |
| uni. F | Standard unit of measurement for flow |
| uni. T | Standard unit of measurement for temperature |
| Vol.x | Current counter reading for totaliser Vol.x |
| Vol. L | Current counter reading for totaliser Vol. L over the whole lifetime |
10.2 Output configuration
This chapter describes the options for the output signals at OUT1 and OUT2.
10.2.1 Switching signal for limit value monitoring
A switching signal can be output for process value monitoring. OUTx changes its switching state when the set switching limits are exceeded or not reached. You can choose between hysteresis and window function.
Hysteresis function:

1: Process value
t: Time
SP: Set point
rP: Reset point
HY: Hysteresis
Hno: Hysteresis function NO (normally open)
Hnc: Hysteresis function NC (normally closed)
Fig. 8: Hysteresis function

When the hysteresis function is set, the set point SP and the reset point rP are set. The rP value must be lower than the SP value. The difference between SP and rP is at least 0.5 % of the final value of the measuring range (= hysteresis). If only the set point is changed, the reset point is changed automatically; the difference remains constant.
Window function:

1: Process value
t: Time
FH: Upper limit value
FL: Lower limit value
HY: Hysteresis
FE: Window area
Fno: Window function NO (normally open)
Fnc: Window function NC (normally closed)
Fig. 9: Window function

When set to the window function, the window high FH and the window low FL are set. The difference between FH and FL is at least 0.5% of the final value of the measuring range. FH and FL have a fixed hysteresis of 0.25% of the final value of the measuring range. This helps keep the switching status of the output stable if the flow rate varies slightly.
Parameters to be set:
oux= Hno, Hnc, Fno, Fnc; SPx; rPx; FHx; FLx
10.2.2 Switching signal Diagnosis
The unit features an integrated diagnostic function. When using the diagnostic function, the output is used exclusively for diagnostic message output, which it indicates by a switched signal.
The switching output is switched on in normal operation (normally closed). If the device detects a diagnostic case, the output will be switched off.
Diagnostic cases include:
- Reversal of the direction of flow ➕ 23
• Low signal quality / no signal ➞ 23
10.2.2.1 Switching signal for flow direction
A flow direction change can be monitored by providing a switching signal.
The output is switched on until the flow rate falls below the set minimum flow rate in negative flow direction (- LFC)(1).
Then the following applies:
• The output switches ON when + LFC is exceeded (2).
- The output switches OFF when - LFC is not reached (3).

LFC = Low flow cut-off → Low flow cut-off → 32.

Fig. 10: Monitoring of the flow direction by switching signals
+Q: Flow in positive flow direction
-Q: Flow in negative flow direction
+LFC: Minimum flow in positive flow direction
-LFC: Minimum flow in negative flow direction
An arrow with the text “flow direction” on the device indicates the positive flow direction. The direction of the flow measurement can be reversed using the parameter Fdir.
→ Flow direction → 32.
Parameters to be set:
oux = dOU ; dFUx = direction of flow
10.2.2.2 Switching signal for signal quality
The unit can provide a switching signal when the signal quality deviates from normal operation.
The signal quality of the sensor can be affected by irregularities in the medium (e.g. strong turbulences, air bubbles, particles or build-up).
The unit detects the signal quality in three stages:
| Signal quality | Explanation | Switching output |
| Normal | The device operates without restrictions (normal operation). | On |
| Low | The signal quality is disturbed, but the device is still working within its specifications. | OFF |
| No signal | No medium is present or no signal can be created. | OFF |
Parameters to be set:
oux = dOU ; dFUx = signal quality
10.2.3 Consumed quantity monitoring (totaliser function)
The unit has 3 internal quantity meters (totalisers Vol.1, Vol.2 and Vol. L). The totalisers continuously sum up the consumed quantity and provide this process value via the IO-Link interface.
| Totaliser | Process value | Read access via IO-Link |
| Vol.1 | Consumed quantity 1(This value is used for consumed quantity monitoring by switching or pulse signals) | Cyclic |
| Vol.2 | Consumed quantity 2 | Acyclic |
| Vol. L | Consumed quantity over the whole lifetime (lifetime totaliser) | Acyclic |
- The totalisers Vol.1 and Vol.2 can be reset. Totaliser Vol. L cannot be reset.
→ Totaliser reset → 34. - The totalisers Vol.1 and Vol.2 take account of the following parameter settings when totalising the consumed quantity:
→ Flow direction → 32.
→ Counting method of the totalisers → 34.
→ Low flow cut-off → 32.
- The Life Time Totalisator Vol. L totals all flow quantities regardless of the flow direction and counting method.
- When the detection range (cr. OL) is exceeded, the totalisers use the last valid flow rate value (measuring range end value) and continue counting with this value.
- In addition to the current consumed quantity, the value before the last reset is saved. This value and the time since the last reset can also be displayed.
Read totaliser values → 36

The totaliser saves the totalled consumed quantity at regular intervals. After a power failure this value is available as the current meter reading. If a time-controlled reset is set, the elapsed time of the set reset interval is also saved. This means that the possible data loss can amount to one minute.
- The accuracy of the consumed quantity measurement depends on the accuracy of the flow measurement.
- A switching signal or pulse signals can be provided for consumed quantity monitoring:
Switching signal totaliser ➕ 25
Switching signal totaliser ➕ 25
→ Pulse signal totaliser → 25

OUT1 and OUT2 cannot be used simultaneously for the consumed quantity monitoring.
10.2.3.1 Switching signal totaliser
A switching signal can be provided for consumed quantity monitoring.
When totaliser Vol.1 has totalled the flow quantity (pulse value) set under ImPS, the output provides a switching signal.
The flow direction is taken into account when totalling the flow quantity ➞ 34.
The output remains switched until a totaliser reset is carried out. When the totaliser has been reset, metering starts again.
The totaliser is reset automatically or manually.
The conditions for the totaliser reset and the switching signal can be set via the parameter rTo:
- rTo = OFF:
- The totaliser is only reset with a manual reset or after overflow.
- The output is switched when the totaliser has reached the flow quantity ImPS.
• rTo = ...h / d / w (hours / days / weeks):
- The totaliser is automatically reset after the set time.
- The output is only switched when the totaliser reaches the flow quantity ImPS by the set time.

The totalisers can be reset manually at any time via the rTox parameter. Totaliser Vol.1 can additionally be reset via an external signal at pin 2.
→ Totaliser reset → 34
Parameters to be set:
oux = ImP ; ImPSx ; ImPRx = no
10.2.3.2 Pulse signal totaliser
Pulse signals can be provided for consumed quantity monitoring.

Pulse signals are not available via the IO-Link interface.
The output provides a pulse signal each time totaliser Vol.1 has totalled the flow quantity (pulse value) set under ImPS.
The flow direction is taken into account when totalling the flow quantity → 34.
The pulse signal consists of a short switching on and off of the output.
Parameters to be set:
oux = ImP ; ImPSx ; ImPRx = yes
10.2.4 Analogue signal
The device provides an analogue signal proportional to the process value.
Within the measuring range the analogue signal is between 4...20 mA.
The measuring range can be scaled between -100% and 100% of the final value of the measuring range.

A negative flow value means flow against the flow direction set under [Fdir] ➕ 32.
- ASP2 determines at which measured value the output signal is 4 mA.
• AEP2 determines at which measured value the output signal is 20 mA.

Minimum distance between ASP2 and AEP2 = 20 % of the final value of the measuring range.
If the measured value is outside the measuring range or in the event of an internal error, the current signal indicated in the following figure is provided.

Fig. 11: Characteristic of the analogue output according to IEC 60947-5-7.
*For devices with nominal diameters DN15...50, the MEW is between 125...130°C (257...266°F), the temperature limits for OL and cr. OL shift accordingly.
A: Analogue signal MAW: Initial value of the measuring range
B: Process value MEW: Final value of the measuring range
1: Detection zone ASP: Analogue start point
2: Display range AEP: Analogue end point
3: Measuring range UL: Below the display range
4: Scaled measuring range cr. UL: Below the detection zone
Q: Flow OL: Above the display range
T: Temperature cr. OL: Above the detection zone

The analogue signal in case of a fault can be set via the parameter FOU: Error behaviour of the outputs ➕ 29.
Parameters to be set:
ou2 = 1 / 4...20 mA, ASP2; AEP2
10.2.5 Frequency signal
The device provides a frequency signal proportional to the process value.
The frequency signal is adjustable:
- FrP1 defines the frequency signal in Hz that is provided when the upper measured value is reached.
Setting range: 1 Hz...10 kHz.
The measuring range is scalable:
- FSP1 defines the lower measured value from which a frequency signal is provided.
• FEP1 defines the upper measured value at which the output signal has the frequency set under FrP1.

FSP1 is only available for temperature measurement. Minimum difference between FSP1 and FEP1 = 20 % of the final value of the measuring range.
If the measured value is outside the measuring range or in the event of an internal error, the frequency signal indicated in the following figure is provided.
Frequency signal for flow:

Fig. 12: Output characteristic of the frequency output, flow
| A: | Frequency signal | MAW: | Initial value of the measuring range |
| B: | Flow | MEW: | Final value of the measuring range |
| 1: | Indicating range | FEPx: | Frequency end point |
| 2: | Measuring range | FrPx: | Frequency signal (Hz) for upper measured value |
| 3: | Scaled measuring range | OL: | Above the indicating range |
| Err: | Error |
Frequency signal for temperature:

Fig. 13: Output characteristic of the frequency output, temperature
| A: | Frequency signal | MAW: | Initial value of the measuring range |
| B: | Temperature | MEW: | Final value of the measuring range |
| 1: | Indicating range | FSPx: | Frequency start point |
| 2: | Measuring range | FEPx: | Frequency end point |
| 3: | Scaled measuring range | FrPx: | Frequency signal (Hz) for upper measured value |
| Err: | Error | OL: | Above the indicating range |
| UL: | Below the indicating range |
Parameters to be set:
ou1 = FRQ; FrP1; FSP1 (only for TEMP); FEP1
10.2.6 Error behaviour of the outputs
The response of the OUTx output in case of a fault can be set via the parameter FOUx. Depending on the selected output function, the following signals are provided in case of a fault:
- Switching signal:
| FOUx | Process values SELx | Output signal | Explanation |
| On | All process values | The output switches ON in case of a fault. | As soon as a defective process value is present, the device sets all process values to invalid. |
| OFF | All process values | The output switches OFF in case of a fault. | |
| OU | Flow | The output switches OFF in case of a fault. | If the process value “Flow” is defective, the device continues to provide the process value “Temperature”. |
| Temperature | The output switches ON in case of a fault. | If the process value “Temperature” is defective, the device continues to provide the process value “Flow”. |
- Analogue signal:
| FOUx | Process values SELx | Output signal | Explanation |
| On | All process values | In case of an error the output goes to 21.5 mA. | As soon as a defective process value is present, the unit sets all process values to invalid. |
| OFF | All process values | In case of an error the output goes to 3.5 mA. | |
| OU | Flow | In case of an error the output goes to 3.5 mA. | If the process value “Flow” is defective, the unit continues to provide the process value “Temperature”. |
| Temperature | In case of an error the output goes to 21.5 mA. | If the process value “Temperature” is defective, the unit continues to provide the process value “Flow”. |
- Frequency signal:
| FOUx | Process values SELx | Output signal | Explanation |
| On | All process values | In case of an error the output goes to 130% of FrPx. | As soon as a defective process value is present, the device sets all process values to invalid. |
| OFF | All process values | In case of an error the output goes to 0 Hz. | |
| OU | Flow | In case of an error the output goes to 0 Hz. | If the process value “Flow” is defective, the device continues to provide the process value “Temperature”. |
| Temperature | In case of an error the output goes to 130% of FrPx. | If the process value “Temperature” is defective, the device continues to provide the process value “Flow”. |

The parameter FOU has no influence on the pulse signal, the diagnostic signals for flow direction and signal quality and the IO-Link process data transmission.
10.2.7 Output off
The output signal can be switched off via the parameter oux = OFF . The output then goes to high impedance.
Communication via the IO-Link interface on OUT1 remains active.
10.3 Application configuration
The chapter describes the setting options for adaptation to your specific application.
10.3.1 Standard unit of measurement
It is possible to set a unit of measurement for each process value, on which further parameter settings will be based.
Selectable values:
- Flow uni. F:
- SUHxx0: I/min; I/h; m³/h; m/s.
– SUHxx1: l/min; l/h; m3/h ; m/s; gal/min; gal/h; ft/s; oz/min.
• Temperature uni. T:
- SUHxx0: °C.
- SUHxx1: °C; °F.
10.3.2 Process value for OUT1 and OUT2
The process value to be output via OUTx can be selected using the parameter SELx.
Selectable values:
- FLOW: Flow
• TEMP: Temperature
10.3.3 Damping
The set damping constant stabilises the output signals. Abrupt changes in the physical process values are smoothed out.
This concerns the outputs and the process value transmission via the IO-Link interface.
The damping constant dAP is used to set after how many seconds the output signal reaches 63 % of the final value if the measured value changes suddenly.
The damping constant is added to the response time of the sensor (→ Technical data).
The UL and OL signals are defined under consideration of the damping time.

Measured value damping only has an effect on the process value flow.
10.3.4 Output polarity
The output polarity is set via the parameter P-n.
The setting affects both switching outputs.
- PnP: The switching output is positive switching.
- nPn: The switching output is negative switching.
10.3.5 Low flow cut-off
Low flow quantities can be ignored using the parameter LFC (Low flow cut-off). Flow below the LFC value is evaluated by the sensor as standstill (Q = 0).
The LFC value influences:
• The switching signal for flow
• the analogue signal for flow
• The frequency signal for flow
• the consumed quantity monitoring (switching or pulse signal for flow)
- The totalisation of the consumed quantity by the totaliser.
• the memory values for minimum and maximum flow

Fig. 14: Low flow cut-off
+LFC: Minimum flow in positive flow direction
-LFC: Minimum flow in negative flow direction
1: Flow which is evaluated as standstill
10.3.6 Medium
The sensor provides various characteristic curves for the respective media. They can be selected via the MEdl parameter.
Selectable values:
- H2O: Water
- OIL46: High-viscosity oils (viscosity: 30...68 mm 2 /s at 40 °C / 30...68 cSt at 104 °F)
10.3.7 Flow direction
The positive flow direction can be defined by the user. This setting affects the following functions:
→ Consumed quantity monitoring (totaliser function) → 24
Switching signal for flow direction 23
→ Analogue signal → 25
An arrow with the text “flow direction” on the device indicates the positive flow direction (factory setting). The direction of the flow rate measurement can be reversed using the parameter Fdir:
| Fdir | Direction of flow |
| + | Flow direction in case of factory setting |
| - | Flow direction contrary to the factory setting |
10.3.8 Calibration
The calibration factor CGA is used to adjust the temperature-viscosity compensation of the sensor to the characteristics of the medium used. The calibration factor influences the slope of the measurement characteristic of the flow measurement.

The slope modification of the measurement characteristic is indicated in percent. The factory setting is CGA = 100%. After a change the calibration can be reset to factory setting.

Fig. 15: calibration of the measurement characteristic
1: calibration factor CGA
2: process value
MW: measured value
V0: measurement characteristic at factory setting
V1: measurement characteristic 1 after calibration
V2: measurement characteristic 2 after calibration
10.3.9 Zero calibration
If there is a systematic deviation between the measured value and the actual process value, this measurement inaccuracy can be corrected using the correction factor cOF.

The unit for coF corresponds to the set standard unit of measurement for flow rate.
The internal zero point is shifted by the set value.

Fig. 16: Zero-point calibration (calibration offset)
t: Time
MEW: Final value of the measuring range
V0: Curve of measured values at factory setting
V1: Curve of measured values after offset
V2: Curve of measured values after offset
Setting range:
coF - FLOW = - 10 % ... +10 %

The parameter is reset to the factory setting both via an application reset and a back-to-box reset.
10.3.10 Totaliser reset
The totalisers Vol.1 and Vol.2 can be reset in different ways:
| Type of reset | Parameter | |
| 1. | Manual reset | Command Reset Totaliser x |
| 2. | Time-controlled reset | Reset Totaliser x =... h (hours)... d (days)... w (weeks) |
| 3. | Reset via external signal | ou2 = In. DDln2:+EDG = reset for rising edge-EDG = reset for falling edgeHIGH = reset for high signalLOW = reset for low signal |
| 4. | Reset via overflow(maximum display range is reached) | Reset Totaliser x = OFF |
Totaliser Vol. L cannot be reset.
If totaliser Vol.1 is reset in one of the above ways, the output is also reset in the case of consumed quantity monitoring.
→ Switching signal totaliser → 25.
10.3.11 Counting method of the totalisers
The totalisers Vol.1 and Vol.2 take account of the flow direction when totalising the consumed quantity. The following counting methods can be defined via the parameter FProx:
| FProx | Counting method |
| 0+ | Negative volumetric flow values (against the marked flow direction) are not taken into account for totalling. |
| -0 | Positive volumetric flow values (corresponding to the marked flow direction) are not taken into account for totalling. |
| -+ | Negative flow values are subtracted from the consumed quantity. |
| ++ | All volumetric flow values are totalled irrespective of the volumetric flow direction. |
Tab. 2: Counting method of the totalisers
The counting method of Vol. L cannot be set. The lifetime totaliser totals all volumetric flow quantities irrespective of the flow direction.
The counting method affects the output signals for consumed quantity monitoring.
→ Consumed quantity monitoring (totaliser function) → 24.

Fig. 17: Taking into account the flow direction when totalling the consumed quantity
+Q: Flow quantity in positive direction
-Q: Flow quantity in negative direction
V: Flow quantity absolute (= sum of negative and positive flow)
1: Flow changes to negative direction
2: Flow changes to positive direction
3: Flow taken into account for totalisation
When the direction of flow is changed, the minimum flow quantity LFC is taken into account.
Low flow cut-off 32.
10.3.12 Reset the unit
The unit can be reset in 2 ways:
• APPL (application reset): reset of the parameter settings. The following is reset:
– All changed application-specific parameters

If IO-Link data storage is activated, this triggers a parameter update in the master. This writes the parameters configured in the master to the device again. An application reset may therefore be ineffective.
- BtB (Back to Box): reset to factory settings. The following is reset:
– All changed application-specific parameters
- All writeable unit identification parameters such as Application Specific Tag, Function Tag or Location Tag.
– Diagnostic parameters, status parameters, events.

After the Back to Box reset, the sensor suspends communication and measurement operation until the voltage is interrupted. The IO-Link data storage is not triggered.

We recommend documenting your own settings in the chapter Factory setting before carrying out a reset.
10.4 Diagnostic functions
The device offers a range of diagnostic functions.
Diagnostic messages can be provided via an output signal:
Switching signal Diagnosis → 23.
In addition, the diagnostic information described below is available via the IO-Link interface.
10.4.1 Read totaliser values
For the totalisers, the following values can be read at any time via the IO-Link interface:
Totaliser values Vol.1 and Vol.2
- Current flow quantity (= consumed quantity since the last totaliser reset)
• Value before the last totaliser reset
• Time since the last totaliser reset
Lifetime totaliser (for the entire operating time)
- Flow quantity in preferred direction (= positive direction of flow)
- Flow quantity in non-preferred direction (= negative direction of flow)
→ Flow direction → 32
10.4.2 Memory
The unit stores the maximum and minimum measured process values.
The current value can be read via the IO-Link interface.
Selectable values:
• Lo. F: Minimum value memory for volumetric flow
• Hi. F: Maximum value memory for volumetric flow
• Lo. T: Minimum value memory for temperature
• Hi. T: Maximum value memory for temperature

It makes sense to delete the memories as soon as the unit operates under normal operating conditions for the first time.
10.4.3 Operating hours counter
The operating hours since the first set-up are stored by the unit.
The current value can be read via the IO-Link interface.
The counter cannot be reset.
10.4.4 Internal temperature
The sensor measures the internal temperature.
The current value can be read via the IO-Link interface.
10.4.5 Signal quality
The signal quality of the sensor can be affected by irregularities in the medium (e.g. strong turbulences, air bubbles, particles or build-up).
The unit detects the signal quality in three stages:
| Signal quality | Explanation |
| Normal | The unit operates without restrictions (normal operation). |
| Low | The signal quality is disturbed, but the unit is still working within its specifications. |
| No signal | No medium is present or no signal can be created. |
The current value can be read via the IO-Link interface.
In addition, the signal quality can be indicated via a switching signal.
Switching signal for signal quality ➕ 23
10.5 Service functions
10.5.1 Device information
Unalterable device information is stored on the unit. This includes:
- Product name
- Product family
- Manufacturer
- Manufacturer ID
- Device ID
- Serial number
• Hardware / firmware revision - Description
In addition, further freely definable tags with a maximum length of 32 characters can be assigned to the unit via the IO-Link interface using suitable parameter setting software. This includes:
- application-specific tag
- function tag
- location tag
10.5.2 Simulation
With this function, process values are simulated and their signal path is checked.
Process values that lead to an error message or warning can be simulated (e.g. OL).
When the simulation is started, the values of the totaliser are frozen and the simulated totaliser is set to 0. The simulated flow value then has an effect on the simulated totaliser. When the simulation is ended, the initial totaliser values are restored.
During the simulation:
- The simulation has no effect on the current process values. The outputs operate as previously set.
- The original totaliser value remains saved without any changes even if there is a real flow.
- No error messages of the current application are available. They are suppressed by the simulation.
The following values can be simulated:
• process values for flow and temperature
- process values outside the measuring range (cr. UL, UL, OL, cr. OL)
Parameters to be set:
S. Tim; S. FLW; S. TMP
11 Operation
After power on and expiry of the power-on delay time, the unit is in the normal operating mode. It carries out its measurement and evaluation functions and generates output signals according to the set parameters.
12 Troubleshooting
The device has many self-diagnostic options. It monitors itself automatically during operation.
Warning and error messages are output as an event via the IO-Link interface.
The status signals are classified according to NAMUR recommendation NE107.
If the measured temperature value fails, the process value for flow rate is still available.

Additional diagnostic functions are available via IO-Link ➕ IO-Link interface description at documentation.ifm.com.
12.1 Warning messages
| IO-Link event Name / code | Problem | Corrective measures |
| Short circuit0x771030480d | Short circuit on output OUT1 and / or OUT2. | ▶ Check output for short circuit or excessive current. |
| Temperature exceeded0x421016912d | Admissible internal device temperature exceeded. | ▶ Eliminate heat sources. |
| Temperature not reached0x422016928d | Admissible internal device temperature not reached. | ▶ Insulate device. |
| Component malfunction0x501020496d | A process value is erroneous. | ▶ Repair or replace the device. |
| Process value belowthe valid range0x8C3035888d | Below the display range (UL). Process value uncertain. | ▶ Check measuring range. |
| Process value abovethe valid range0x8C1035856d | Above the display range (OL). Process value uncertain. | ▶ Check measuring range. |
| Override active. Device status = 20x8CDC36060d | A process value differs from the measured value. PV is set to “0” while over-ride bit is set in PDOut. | ▶ Deactivate PDOut override. |
| Signal quality low0x8CBF36031d | Signal quality low. | ▶ Remove the device and check for deposits. ▶ Check application for interference (air bubbles/particles). |
| Simulation active0x8C0135841d | Simulation active. | ▶ End simulation. |

In the event of a warning, the outputs react according to the setting under FOU = OU. Exception: Short circuit.
12.2 Error messages
| IO-Link event Name / code | Problem | Corrective measures |
| Hardware fault in the device0x500020480d | Device faulty / malfunction. | ► Replace the device. |
| No media detected0x8CC536037d | No medium present or signal quality too low due to interference in the pipe length. | ► Check whether medium is present in the sensor tube.► Remove the device and check for deposits.► Check application for interference (air bubbles/particles). |
| Parameter error0x632025376d | Parameter setting outside the valid range. | ► Check parameter setting.► Perform a back-to-box reset. |
| Measuring range exceeded0x8C2035872d | Above the detection range (cr. OL). | ► Check the measuring range. |
| Measuring range not reached0x8C2035872d | Below the detection range (cr. UL). | ► Check the measuring range. |

In the event of an error, the outputs react according to the setting under FOU.
13 Maintenance, repair and disposal
The operation of the unit is maintenance-free.
Only the manufacturer is allowed to repair the unit.
▶ After use dispose of the device in an environmentally friendly way in accordance with the applicable national regulations.
14 Factory Settings
SUHxx0:
| Parameter | SUH120 | SUH820 | SUH220 | SUH320 | SUH420 | SUH520 | SUH620 | SUH720 |
| SP1 / FH1 | 13l/min | 15l/min | 48l/min | 75l/min | 200l/min | 480l/min | 720l/min | 1200l/min |
| rP1 / FL1 | 12.3l/min | 14.2l/min | 45.5l/min | 71.1l/min | 189.6l/min | 455l/min | 682.6l/min | 1137.6l/min |
| SP2 / FH2 | 26l/min | 30l/min | 96l/min | 150.0l/min | 400l/min | 960l/min | 1440l/min | 2400l/min |
| rP2 / FL2 | 25.3l/min | 29.2l/min | 93.5l/min | 146.1l/min | 389.6l/min | 935l/min | 1402.6l/min | 2337.6l/min |
| FSP1 | -40 °C | -40 °C | -40 °C | -40 °C | -40 °C | -40 °C | -40 °C | -40 °C |
| FEP1 | 65l/min | 75l/min | 240l/min | 375.0l/min | 1000l/min | 2400l/min | 3600l/min | 6000l/min |
| FrP1 | 1000 Hz | 1000 Hz | 1000 Hz | 1000 Hz | 1000 Hz | 1000 Hz | 1000 Hz | 1000 Hz |
| ImPS1 | 0.1 L | 0.1 L | 0.1 L | 0.1 L | 0.1 L | 0.1 L | 0.1 L | 0.1 L |
| ImPR1 | YES | YES | YES | YES | YES | YES | YES | YES |
| ImPS2 | 0.1 L | 0.1 L | 0.1 L | 0.1 L | 0.1 L | 0.1 L | 0.1 L | 0.1 L |
| ImPR2 | YES | YES | YES | YES | YES | YES | YES | YES |
| ASP2 | 0l/min | 0l/min | 0l/min | 0l/min | 0l/min | 0l/min | 0l/min | 0l/min |
| AEP2 | 65l/min | 75l/min | 240l/min | 375.0l/min | 1000l/min | 2400l/min | 3600l/min | 6000l/min |
| Dln2 | +EDG | +EDG | +EDG | +EDG | +EDG | +EDG | +EDG | +EDG |
| SEL1 | FLOW | FLOW | FLOW | FLOW | FLOW | FLOW | FLOW | FLOW |
| ou1 | HNO | HNO | HNO | HNO | HNO | HNO | HNO | HNO |
| dOU1 | Fdir | Fdir | Fdir | Fdir | Fdir | Fdir | Fdir | Fdir |
| FOU1 | OFF | OFF | OFF | OFF | OFF | OFF | OFF | OFF |
| SEL2 | FLOW | FLOW | FLOW | FLOW | FLOW | FLOW | FLOW | FLOW |
| ou2 | I | I | I | I | I | I | I | I |
| dOU2 | Fdir | Fdir | Fdir | Fdir | Fdir | Fdir | Fdir | Fdir |
| FOU2 | OFF | OFF | OFF | OFF | OFF | OFF | OFF | OFF |
| uni. F | l/min | l/min | l/min | l/min | l/min | l/min | l/min | l/min |
| uni. T | °C | °C | °C | °C | °C | °C | °C | °C |
| dAP | 0.6 s | 0.6 s | 0.6 s | 0.6 s | 0.6 s | 0.6 s | 0.6 s | 0.6 s |
| P-n | PnP | PnP | PnP | PnP | PnP | PnP | PnP | PnP |
| LFC | 1.0l/min | 1.0l/min | 1.0l/min | 3.0l/min | 5.0l/min | 20l/min | 25l/min | 45l/min |
| MEdl | H2O | H2O | H2O | H2O | H2O | H2O | H2O | H2O |
| Fdir | + | + | + | + | + | + | + | + |
| CGA | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
| cOF | 0 l/min | 0 l/min | 0 l/min | 0 l/min | 0 l/min | 0 l/min | 0 l/min | 0 l/min |
| rTo1 | OFF | OFF | OFF | OFF | OFF | OFF | OFF | OFF |
| rTo2 | OFF | OFF | OFF | OFF | OFF | OFF | OFF | OFF |
| FPro1 | 0+ | 0+ | 0+ | 0+ | 0+ | 0+ | 0+ | 0+ |
| FPro2 | 0+ | 0+ | 0+ | 0+ | 0+ | 0+ | 0+ | 0+ |
| S. FLW | 32.5l/min | 37.5l/min | 120l/min | 187.5l/min | 500l/min | 1200l/min | 1800l/min | 3000l/min |
| S. TMP | 50 °C | 50 °C | 50 °C | 50 °C | 50 °C | 50 °C | 50 °C | 50 °C |
| S. Tim | 3 min | 3 min | 3 min | 3 min | 3 min | 3 min | 3 min | 3 min |
| S. On | OFF | OFF | OFF | OFF | OFF | OFF | OFF | OFF |
SUHxx1:
| Parameter | SUH801 | SUH251 | SUH301 | SUH451 | SUH501 | SUH601 | SUH701 |
| SP1 / FH1 | 3.96gal/min | 12.68gal/min | 19.81gal/min | 52.83gal/min | 126.8gal/min | 190.2gal/min | 317gal/min |
| rP1 / FL1 | 3.76gal/min | 12.02gal/min | 18.78gal/min | 50.09gal/min | 120.2gal/min | 180.3gal/min | 300.5gal/min |
| SP2 / FH2 | 7.93gal/min | 25.36gal/min | 39.63gal/min | 105.67gal/min | 253.6gal/min | 380.4gal/min | 634gal/min |
| rP2 / FL2 | 7.72gal/min | 24.7gal/min | 38.60gal/min | 102.93gal/min | 247gal/min | 370.5gal/min | 617.5gal/min |
| FSP1 | -40 °C | -40 °C | -40 °C | -40 °C | -40 °C | -40 °C | -40 °C |
| FEP1 | 19.81gal/min | 63.4gal/min | 99.06gal/min | 264.17gal/min | 634gal/min | 951gal/min | 1585gal/min |
| FrP1 | 1000 Hz | 1000 Hz | 1000 Hz | 1000 Hz | 1000 Hz | 1000 Hz | 1000 Hz |
| ImPS1 | 0.1 gal | 0.1 gal | 0.1 gal | 0.1 gal | 0.1 gal | 0.1 gal | 0.1 gal |
| ImPR1 | YES | YES | YES | YES | YES | YES | YES |
| ImPS2 | 0.1 gal | 0.1 gal | 0.1 gal | 0.1 gal | 0.1 gal | 0.1 gal | 0.1 gal |
| ImPR2 | YES | YES | YES | YES | YES | YES | YES |
| ASP2 | 0gal/min | 0gal/min | 0gal/min | 0gal/min | 0gal/min | 0gal/min | 0gal/min |
| AEP2 | 19.81gal/min | 63.4gal/min | 99.06gal/min | 264.17gal/min | 634gal/min | 951gal/min | 1585gal/min |
| Dln2 | +EDG | +EDG | +EDG | +EDG | +EDG | +EDG | +EDG |
| SEL1 | FLOW | FLOW | FLOW | FLOW | FLOW | FLOW | FLOW |
| ou1 | HNO | HNO | HNO | HNO | HNO | HNO | HNO |
| dOU1 | Fdir | Fdir | Fdir | Fdir | Fdir | Fdir | Fdir |
| FOU1 | OFF | OFF | OFF | OFF | OFF | OFF | OFF |
| SEL2 | FLOW | FLOW | FLOW | FLOW | FLOW | FLOW | FLOW |
| ou2 | I | I | I | I | I | I | I |
| dOU2 | Fdir | Fdir | Fdir | Fdir | Fdir | Fdir | Fdir |
| FOU2 | OFF | OFF | OFF | OFF | OFF | OFF | OFF |
| uni. F | gal/min | gal/min | gal/min | gal/min | gal/min | gal/min | gal/min |
| uni. T | °F | °F | °F | °F | °F | °F | °F |
| dAP | 0.6 s | 0.6 s | 0.6 s | 0.6 s | 0.6 s | 0.6 s | 0.6 s |
| P-n | PnP | PnP | PnP | PnP | PnP | PnP | PnP |
| LFC | 0.26gal/min | 0.26gal/min | 0.79gal/min | 1.32gal/min | 5.3gal/min | 6.6gal/min | 11.9gal/min |
| MEdl | H2O | H2O | H2O | H2O | H2O | H2O | H2O |
| Fdir | + | + | + | + | + | + | + |
| CGA | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
| cOF | 0 gal/min | 0 gal/min | 0 gal/min | 0 gal/min | 0 gal/min | 0 gal/min | 0 gal/min |
| rTo1 | OFF | OFF | OFF | OFF | OFF | OFF | OFF |
| rTo2 | OFF | OFF | OFF | OFF | OFF | OFF | OFF |
| FPro1 | 0+ | 0+ | 0+ | 0+ | 0+ | 0+ | 0+ |
| FPro2 | 0+ | 0+ | 0+ | 0+ | 0+ | 0+ | 0+ |
| S. FLW | 9.91gal/min | 31.7gal/min | 49.53gal/min | 132.09gal/min | 317gal/min | 475.5gal/min | 792.5gal/min |
| S. TMP | 122 °F | 122 °F | 122 °F | 122 °F | 122 °F | 122 °F | 122 °F |
| S. Tim | 3 min | 3 min | 3 min | 3 min | 3 min | 3 min | 3 min |
| S. On | OFF | OFF | OFF | OFF | OFF | OFF | OFF |