Uni-T UT533 - Mètre

UT533 - Mètre Uni-T - Free user manual and instructions

Find the device manual for free UT533 Uni-T in PDF.

📄 59 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice Uni-T UT533 - page 1
Pick your language and provide your email: we'll send you a specifically translated version.
Product TypeInsulation Multimeter (True RMS)
ModelUT533
Display6000 count LCD
AC Voltage Range6.000V to 1000V
DC Voltage Range6.000V to 1000V
DC Millivolt Range60.00mV to 600.0mV
AC Current Range60.00mA to 600.0mA
DC Current Range60.00mA to 600.0mA
Resistance Range600.0Ω to 40.00MΩ
Capacitance Range10.00nF to 100.0μF
Frequency RangeUp to 1MHz
Temperature Range (UT533 only)-40°C to 537°C (-40°F to 998°F)
Insulation Test Voltages50V, 100V, 250V, 500V, 1000V
Insulation Resistance Range0.1MΩ to 2.00GΩ
Safety RatingCAT III 1000V, CAT IV 600V
Power Supply6 × AA (AM3/LR6) batteries, ~2450mAh
Dimensions97mm × 202mm × 46mm
WeightApprox. 505g (including batteries)
Auto Power OffAfter 15 minutes of inactivity
Low Battery IndicationIcon on display
Fuse Protection1A/240V fast fuse, high breaking capacity
CleaningDamp cloth with mild detergent; no abrasives or solvents
Spare PartsTest leads, short leads, alligator clips, K-type thermocouple (UT533), batteries, fuse
RepairabilityBattery and fuse replaceable by user; internal circuit not alterable
General InformationTrue RMS measurement, insulation testing, data storage (up to 99 sets)

Frequently Asked Questions - UT533 Uni-T

How do I measure AC voltage with the UT533?
Insert red lead into VΩ→Hz terminal, black lead into COM. Set rotary switch to V~. Connect test leads across the circuit. The display shows True RMS value. Input impedance is 10MΩ.
What is the maximum voltage I can measure?
The UT533 can measure up to 1000V AC/DC. Do not exceed this rating to avoid electric shock or damage.
How do I perform an insulation test?
Set rotary switch to INSULATION. Press RANGE to select voltage (50V to 1000V). Connect short leads with alligator clips to INSULATION and mA terminals. Ensure circuit is dead. Press and hold TEST or use LOCK mode by pressing HOLD first.
What safety precautions should I take?
Always use proper terminals and range. Do not apply voltage above rated value. Disconnect power and discharge capacitors before resistance/continuity/diode tests. Use test leads with finger guards. Replace battery when low battery icon appears.
Can I measure temperature with the UT533?
Yes, the UT533 measures temperature from -40°C to 537°C using the included K-type thermocouple (rated to 260°C). Set rotary switch to mV, press SELECT for temperature, and RANGE to toggle between °C and °F.
How do I replace the battery or fuse?
Turn off Meter and remove leads. Remove two screws from battery door, open door. Replace 6 AA batteries. For fuse, also remove two screws from case bottom and one screw at back of battery. Replace with F1A 240V fast fuse only.
What does the LOCK function do in insulation testing?
Press HOLD to enable LOCK mode. Then press TEST once to start continuous insulation test. The Meter will keep applying voltage until you press HOLD or TEST again to release.
How do I store and recall measurements?
Press SAVE to store a single reading; press and hold SAVE for continuous storage (max 99 sets). Press LOAD to recall stored values, use ▲/▼ to navigate. Press CLR to clear data.
What is the accuracy for DC voltage measurement?
DC voltage accuracy is ±(0.3% + 4 digits) for all ranges (6.000V to 1000V). Input impedance is ≥10MΩ.
How should I clean the Meter?
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Keep terminals clean to avoid measurement errors.

User questions about UT533 Uni-T

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Mètre in PDF format for free! Find your manual UT533 - Uni-T and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. UT533 by Uni-T.

USER MANUAL UT533 Uni-T

P/N:110401103678X DATE:2018.06.26 REV.4

UNI-T®

0.000 00 TELA HAN A CE V2 +1Hz

UT531

UT532

UT533

Users Manual

Insulation

Resistance

Multimeters

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Table of Contents

Title Page

Introduction 5

Safety Information 5

Unpacking Inspection 9

The Meter Structure 10

Rotary Switch Positions 11

Buttons 13

Understanding the Display 17

Making Basic Measurements 20

Measuring AC Voltage 20

Measuring DC Voltage 23

Measuring DC Millivolts 25

Measuring Resistance and Capacitance 27

● Measuring Resistance 27

● Measuring Capacitance 29

Testing for Continuity 31

Testing Diodes 33

Measuring AC or DC Current 35

Measuring Frequency 38

Measuring Temperature 40

Testing Insulation 42

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Title Page

Cleaning 45

Replacing the Batteries 46

Replacing the Fuse 48

Specifications 49

Electrical Specifications 50

AC Voltage Measurement 50

DC Voltage Measurement 51

DC Millivolts Measurement 51

AC Current Measurement 51

DC Current Measurement 52

Ohms Measurement 52

Capacitance 52

Diode Test 53

Continuity Test 53

Temperature Measurement 53

Frequency Measurement 54

- At mV ≈ Range 54

- At DCV/ACV Range 55

Insulation Specifications 55

- UT531 55

- UT532 56

- UT533 57

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

List of Tables

TableTitlePage
1.Symbols8
2.Unpacking Inspection9
3.Rotary Switch Selections12
4.Buttons14
5.Display Indicators18
6.General Specifications49

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

List of Figures

Figure Title Page

  1. The Meter Structure — 10
  2. Rotary Switch 11
  3. Buttons 13
  4. Display Indicators----17
  5. Measuring AC Voltage 20
  6. Measuring DC Voltage 23
  7. Measuring DC Millivolts 25
  8. Measuring Resistance 27
  9. Measuring Capacitance 29
  10. Testing for Continuity 31
  11. Testing Diodes 33
  12. Measuring AC or DC Current 35
  13. Measuring Frequency 38
  14. Measuring Temperature 40
  15. Testing Insulation 42
  16. Replacing the Fuse and Battery 46

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Introduction

The UNI-T Models UT531, UT532 and UT533 are battery-powered, true RMS insulation multimeters (hereafter "the Meter") with 6000 count display.

The Meter measures or tests the following:

● AC/DC voltage and current
- Resistance
● Voltage and current frequency
Diode
- Continuity
- Capacitance
● Temperature (Model 532 and UT533 only)
• Insulation Testing

Safety Information

These Meters meet CAT III 1000V and CAT IV 600V IEC61010 standards: in pollution degree 2 and double injection.

CAT III: Distribution level, fixed installation, with smaller transient overvoltages than CAT IV.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

CAT IV: Primary supply level, overhead lines, cable systems etc.

Use the Meter only as specified in this users manual. Otherwise protection provided by the Meter may be impaired. See Table 1 for a list of symbols used on the Meter and in this manual.

A ⚠Warning identifies hazardous conditions and actions that could cause bodily harm or death.

A ⚠️ Caution identifies conditions and actions that could damage the Meter, the equipment under test, or cause permanent loss of data.

Warning

To avoid possible electric shock or personal inquiry, follow these guidelines:

- Use the Meter only as specified in this manual or the protection provided by the Meter might be impaired.

- Do not use the Meter or test leads if they appear damaged, or if the Meter is not operating properly. If in doubt, have the Meter serviced.

● Always use the proper terminal, switch position, and range for measurements before connecting Meter to circuit under test.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

  • Verify the Meter's operation by measuring a known voltage.
  • Do not apply more than the rated voltage as marked on the Meter, between the terminals or between any terminal and earth ground.
  • Use caution with voltages above 33V ac rms, 46.7V ac Peak, or 70V dc. These voltages pose a shock hazard.
  • Replace the battery as soon as the low battery indicative ( [+] ) appears.
  • Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance.
  • Do not use the Meter around explosive gas or vapor.
  • When using the test leads, keep your fingers behind the finger guards.
  • In order to comply with the safety standard, the Meter must be used along with the included test leads. Any damages of the test lead, must replace it with the same model number and specification.
  • Remove test leads from the Meter before opening the Meter case or battery door. Never operate the Meter with the cover removed or the battery door open.
  • Comply with local and national safety requirements when working in hazardous locations.
  • Under the influence of Radio-Frequency Electromagnetic Field & Conducted Radio-Frequency Electromagnetic Field phenomenon. The captioned model have a magnificent error in function measurement. It will be back to normal when the interference is removed.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

  • Use proper protective equipment, as required by local or national authorities when working in hazardous areas.
  • Avoid working alone.
  • User only the replacement fuse specified or the protection may be impaired.
  • Check the test leads for continuity before use. Do not use if the readings are high or noisy.
  • Do not use current terminals or current range to test voltage.

Table 1. Symbols

Double Insulated ± Earth Ground
~AC (Alternating Current) DC (Direct Current)
Diode + Battery (Low battery when shown on display)
···Continuity BuzzerAC or DC (Alternating Current or Direct Current)
CEConforms to Standard of European Union

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Unpacking Inspection

Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part:

Table 2. Unpacking Inspection

Users Manual1 piece
Test Lead1 pair
Short Lead1 pair
Alligator Clip1 pair
K type thermocouple (UT532 and UT533 only)1 piece
Batteries AA (AM3/LR6)6 pieces

In the event you find any missing or damage, please contact your dealer immediately.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

The Meter Structure

  1. Front Housing
  2. LCD Display
  3. Functional Buttons
  4. Rotary Switch
  5. Input Terminals

INSULATION MULTIMETER LNG-T UFS31 2500V 1000V -8888 mpFC mA MHz MnGΩ 2500V 1000V -1888V ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨ ⑩ ⑪ ⑫ ⑬ ⑭ ⑮ ⑯ ⑰ ⑱ ⑲ ⑳ ⑴ ⑵ ⑶ ⑦ ⑧ ⑨ ⑩ ⑪ ⑫ ⑬ ⑭ ⑮ ⑯ ⑰ ⑱ ⑲ ⑳ ④

Figure 1. The Meter Structure

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Rotary Switch Positions

Turn the Meter on by pressing the POWER button. Use the BLUE button to select any rotary switch alternate functions (labeled with blue letters). Rotary switch selections are shown in Figure 2 and described in Table 3.

mV HZ Ω-← V HZ mA ~V HZ 500V 1000V INSULATION

Figure 2. Rotary Switch

Table 3. Rotary Switch Selections

Switch PositionMeasurement Function
V~AC voltage from 60mV to 1000V.
V---DC voltage 1mV to 1000V.
mV---DC mV 0.01mv to 600mV
(UT532 and UT533 only)Temperature from -40°C to +537°C (-40°F to +998°F). Celsius is the default temperature measurement unit.
Ω-HOhms from 0.1 to 40M.Capacitance from 0.01nF to 100μF.
-))Continuity test. Beeper turns on at <30
→+Diode test. There is no ranging in this function.
mA≈DC mA from 0.01mA to 600mA (600mA overload for 2 minutes maximum)AC mA from 3mA to 600mA (600mA overload for 2 minutes maximum).
insulationPress to perform insulation testUT531: From 0.5M to 600M. Select 500V and 1000V testing voltage.UT532: From 0.2M to 2G. Select 250V, 500V and 1000V testing voltage.UT533: From 0.1M to 2G. Select 50V, 100V, 250V, 500V and 1000V testing voltage.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Buttons

Use the buttons to activate features that augment the function selected with the rotary switch. The buttons are shown in Figure 3 described in Table 4.

HOLD SAVE MAX/▲ RANGE /CLR LOAD MIN/▼ SELECT POWER TEST/Hz

Figure 3. Buttons

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Table 4. Buttons

ButtonDescription
POWERPress POWER to turn on the Meter. Press and hold POWER to turn off the Meter
HOLDIn insulation Test mode, this schedules a test lock the next time you press TEST on the Meter. The test lock acts to hold down the button until you press HOLD or TEST again to release the lock.In order measurement mode: Press to freeze the displayed value. Press again to release the display.
*/CLRPress and hold the button to clear the stored data. Press the button to turn the backlight on and off.
SAVEPress this button to store the single record. Press and hold this button to store data continuously. Press SAVE again to exit continuous data storing mode.The maximum number of data store is 99 sets. The Meter displays FULL when it reaches the maximum. Press CLR to clear the stored data in order to store next set of data

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Table 4. Buttons

ButtonDescription
LOADPress this button once to recall the stored value. Press or to recall the previous or next stored value. Press LOAD again to exit data recall mode.Press and hold this button to recall the stored data continuously. Press LOAD again to exit.After entering the Load mode, it is not allowed to turn the rotary switch to enter to measuring mode. You can only enter the measuring mode after exiting the Load mode.
MAX /▲In Measurement mode, except buzzer, diode and insulation test, press to start retaining maximum value.In Load, each press to go back to the previous stored reading
MIN /▼In Measurement mode, except buzzer, diode and insulation test, press to start retaining minimum value.In Load, each press to recall the next stored reading.
RANGEIn Insulation Test mode, press to select different testing voltage.In other mode, except buzzer and diode, press to change Ranging

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Table 4. Buttons

ButtonDescription
RANGEmode from Auto to Manual Ranging mode. Press and hold to return to Auto Ranging mode.
SELECTFunctions as a shift key. Press to access orange functions on the rotary switch.
Initiates an insulation test when the rotary switch is on the INSULTION position. Causes the Meter to source (output) a high voltage and measure insulation resistance.
InMeasurement mode, activate frequency measurement.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Understanding the Display

Display indicators are shown in Figure 4 and described in Table 5.

MAX MIN AVG PEAK H → DC AC (trms) •••• Auto Range Manual Range 61000 mV -8.8.8.8 nµ°F°C mAV Mk Hz MkGΩ MEM 88 FULL LOCK TEST 2500V -1888V MR CLR 1000V

Figure 4. Display Indicators

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Table 5. Display Indicators

IndicatorDescription
MAXIndicates maximum reading has been selected.
MINIndicates minimum reading has been selected.
HIndicates display hold is active
Diode test function is selected
DC ACAC / DC Voltage or Current measurement
trmsTrue RMS measurement
-»)Continuity test function is selected
Low battery. Indicates when it is time to replace the battery.⚠WarningTo avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the low battery indicator appears
High voltage warning
Auto RangeAuto ranging mode
Manual RangeManual ranging mode

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Table 5. Display Indicators

IndicatorDescription
61000mVDisplay range in use
Negative symbol
>Greater than symbol
8888Primary display
MEMData store symbol
MRData recall symbol
"88.The number of stored data
FULLWhen the number of stored data reaches 99, the display will show FULL.When the number of stored data is full, press CLR to clear up the stored data before storing another set of data
CLRClear up the stored data
Uni-T UT533 - UNI-T® - 1Indicates a test lock will be applied the next time you press TEST on the Meter. The test lock acts to hold down the button until you press HOLD or TEST again.
2500V1000VSource voltage rating for insulation test.UT531: 500V and 1000V ranges available.UT532: 250V, 500V and 1000V ranges available.UT533: 50V, 100V, 250V, 500V and 1000V ranges available.
·1888vSecondary Display
Measurement units

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Making Basic Measurements

Measuring AC Voltage (see figure 5)

UNIT U/TEUR INSULATION MULTIMETER 0Hz 200 MHz 10000 MHz Auto Range Rated Temp 10000 kV -8.88.8 >88 V LUX 2500V 1000V -1888V TORT-VOLTAGE RANGE G/A LOAD SMI/W SELECT TORT DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D- D+ VDHZ CE MILION mA COM VΩ φ + + + Hz

Figure 5. Measuring AC Voltage

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

To avoid harm to you or damages to the Meter from electric shock, please do not attempt to measure voltages higher than 1000V although readings may be obtained.

The AC Voltage ranges are: 6.000V, 60.00V, 600.0V and 1000V. If over the range, it displays"OL". To mesaure AC voltage, connect the Meter as follows:

  1. Insert the red test lead into the VΩ→Hz terminal and the black test lead into the COM terminal.
  2. Set the rotary switch to V\~.
  3. Connect the test leads across with the object being measured. The measured value shows on the display.
  4. The displayed value is True RMS.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Note:

  • If the value of voltage to be measured is unknown, set the Meter to autoranging or use the maximum measurement position 1000V and reduce the range step by step until a satisfactory reading is obtained.
  • In each range, the Meter has an input impedance of 10MΩ. This loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10kΩ, teh error is negligible (0.1% or less).
  • When DC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Measuring DC Voltage (see figure 6)

UNITED INSULATION MULTIMETER MAX 100V 750Ω 30kA (mm) Auto Range Wound Range 8100 mA -8.8.8.8 mV°C mA MHz MgΩ -86mA LOCK 250V 100V -1888V SET VOLTAGE RANGE /CLR LOAD SELECT POWER +/- + TEST ON +/- - DC +/- - CE INSOLUTON mA COM VΩ +/- +Hz -

Figure 6. Measuring DC Voltage

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

To avoid harm to you or damages to the Meter from electric shock, please do not attempt to measure voltages higher than 1000V although readings may be obtained.

The DC Voltage ranges are: 6.000V, 60.00V, 600.0V and 1000V. If over the range, it displays "OL". To measure DC voltage, connect the Meter as follows:

  1. Insert the red test lead into the VΩ -I(→) Hz terminal and the black test lead into the COM terminal.
  2. Set the rotary switch to V...
  3. Connect the test leads across with the object being measured. The measured value shows on the display.

Note:

  • If the value of voltage to be measured is unknown, set the Meter to autoranging or use the maximum measurement position 1000V and reduce the range step by step until a satisfactory reading is obtained.
  • In each range, the Meter has an input impedance of 10MΩ. This loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10kΩ, teh error is negligible (0.1% or less).
  • When DC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Measuring DC Millivolts (see figure 7)

UNI-T U7531 INSULATION MULTIMETER 8.8888 -86V LED LOAD BAND MUL/AL RANGE G/CLR LOAD MIN/WR RESET POWER VCC 0Ω - - - + - mV HZ CE INJALON mA COM V Ω-0 + - +Hz -

Figure 7. Measuring DC Millivolts

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

To avoid harm to you or damages to the Meter from electric shock, please do not attempt to measure voltages higher than 500V, otherwise you cannot obtain correct reading.

The DC Millivolts ranges are: 60.000mV and 600.0mV. If over the range, it displays "OL". To mesaure DC Millivolts, connect the Meter as follows:

  1. Insert the red test lead into the VΩ-I↔Hz terminal and the black test lead into the COM terminal.
  2. Set the rotary switch to mV...
  3. Connect the test leads across with the object being measured.
    The measured value shows on the display.

Note:

● The maximum input impedance is 4000MΩ.
- When DC Millivolts measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Measuring Resistance and Capacitance • Measuring Resistance (see figure 8)

UNI-T UTEX1 INSULATION MULTIMETER -88.88.88 -88V CLK 2500V 1000V TMT VOLTAGE LOAD SAVE MAX/A NAME C/CLR LOAD MIN/T SELECT POWER Vcc Vcc Vcc Vcc Vcc Ω*- SOLUTION mA COM VΩ+Hz A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3400 A3600 A3600 A3600 A3600 A3600 A3600 A3600 A3600 A3600 A3600 A3600 A3600 A3600 A3600 A3600 A3600 A3600 VCC 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125V 125W

Figure 8. Measuring Resistance

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

To avoid electric shock, injury, or damage to the Meter, disconnect circuit power and discharge all high-voltage capacitors before testing resistance.

The Resistance ranges are 600.0Ω, 6.000kΩ, 60.00kΩ, 600.0kΩ, 6.000MΩ and 40.00MΩ

To measure resistance, connect the Meter as follows:

  1. Insert the red test lead into the VΩ-H→Hz terminal and the black test lead into the COM terminal.
  2. Set the rotary switch to . Resistance measurement is default or press SELECT button to select measurement mode.
  3. Connect the test leads across with the object being measured.

The measured value shows on the display.

Note:

- When Resistance measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

● Measuring Capacitance (see figure 9)

UNIT UTS01 INSULATION MULTIMETER 0000 MHz 2250 0500 36.4V (Hz) Auto Range Manual Range 81000 kV -8.8.8.8 mJFt mA MHz MDD 88000 2500V -1888V LOCK 1900V SET CLK HOLD SAVE MAX/A RANKS /OLR LOAD MIN/T SELECT POWER W/NZ V/V2 Ω+ V/V2 TEST VOUT/MAN RANKS SELECT TEST IE 80V 100V MOUT/MAN CE INSULATION mA COM VΩ + K - Ω+

Figure 9. Measuring Capacitance

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

To avoid electric shock, injury, or damage to the Meter, disconnect circuit power and discharge all high-voltage capacitors before testing capacitance. Use the DC Voltage function to confirm that the capacitor is discharged.

The Capacitance ranges are 10.00nF, 100.0nF, 1000nF, 10.00μF and 100.0μF

To measure capacitance, connect the Meter as follows:

  1. Insert the red test lead or red short lead with alligator clip into the VΩ→Hz terminal and the black test lead or black short lead with alligator clip into the COM terminal. If the capacitor being tested has polarity, connect the positive into the VΩ→Hz terminal and the negative into the COM terminal
  2. Set the rotary switch to -IF. Press SELECT button to select capacitance measurement mode
  3. Connect the test leads across with the object being measured.

The measured value shows on the display.

Note:

- When Capacitance measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Testing for Continuity (see figure 10)

LINE-T UTE31 INSULATION MULTIMETER +8.888.8 mV°C -86°C 2000V -1888V LOAD BATE RANGE RESET HOLD BASE MAX/A RANGE SET/OUT LOAD MIN/T SELECT POWER TEST +50Hz +50Hz +50Hz +50Hz CE MILUSTRON mA COM VΩ + + + Hz -

Figure 10. Testing for Continuity

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

To avoid electric shock, injury, or damage to the Meter, disconnect circuit power and discharge all high-voltage capacitors before testing continuity.

To measure capacitance, connect the Meter as follows:

  1. Insert the red test lead into the VΩ-H→Hz terminal and the black test lead into the COM terminal.
  2. Set the rotary switch to .
  3. Continuous audible tone for test resistance below 30Ω

Note:

- When Continuity measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Testing Diodes (see figure 11)

UNI-T V7531 INSULATION MULTIMETER + -8.8888 -88mA LOCK 1000V -1888V HOLD SAVE MAX/A SADDLE E/OL LOAD MIN SELECT POWER TENS VCC TA 800V 1000V CE INSULATION mA COM VΩ -4 + - HZ 正向偏压

Figure 11. Testing Diodes

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

To avoid electric shock, injury, or damage to the Meter, disconnect circuit power and discharge all high-voltage capacitors before testing diodes.

To measure diode, connect the Meter as follows:

  1. Insert the red test lead into the VΩ-I(→Hz) terminal and the black test lead into the COM terminal.

  2. Set the rotary switch to ↗.

  3. For forward voltage drop readings on any semiconductor component, place the red test lead on the component's anode and place the black test lead on the component's cathode.

The measured value shows on the display.

Note:

  • In a circuit, a good diode should still produce a forward voltage drop reading of 0.5V to 0.8V; however, the reverse-voltage drop reading can vary depending on the resistance of other pathways between the probe tips.
  • Connect he test leads to the proper terminal as said above to avoid error display. The LCD will display OL indicating diode being tested is open or polarity is reversed.
    ● The unit of diode is Volt (V), displaying the forward voltage drop readings.
  • When Diode measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Measuring AC or DC Current (see figure 12)

LINS-T AUTOV REGULATION MULTIMETER MAX 2000 1500 1000 800 600 Auto Target Thermal Range 8100 mV -8.8.8.8 mV -380 100W 100W -100W SETVUNION RANGE LOAD RESET DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC DC AC mA RECLOSATION mA COM V Ω-φ + +Hz - 负极

Figure 12. Measuring AC or DC Current

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

To avoid personal injury or damage to the Meter:

  • Never attempt to make an in-circuit current measurement when the open-circuit potential to earth is greater than 600V.
  • Inspect Meter's fuses before testing. Broken fuses may damage to the Meter or hurt yourselves
  • Use the proper terminals, switch position, and range for your measurement.
  • Never place the probes in parallel with circuit or component when the leads are plugged into the current terminals.

The current ranges are 60.00mA and 600.0mA.

To measure current, connect the Meter as follows:

  1. Insert the red test lead into the mA terminal and the black test lead into the COM terminal.
  2. Set the rotary switch to mA ≈. DC Current is default. Press SELECT to toggle to AC current measurement mode.
  3. Connect the test leads across with the object being measured. The measured value shows on the display.
  4. AC Current displays True RMS value.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Note:

  • If the value of current to be measured is unknown, set the Meter to autoranging or use the maximum measurement position 600.0mA and reduce the range step by step until a satisfactory reading is obtained.
  • For safety sake, when measuring high current, each measurement time must be less than 10 seconds with an interval more than 15 minutes.
  • When Current measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Measuring Frequency (see figure 13)
U3521 INSULATION MULTIMETER mVHz VHz VHz CE INSULATION mA COM VΩ-+ mHz OFF: 0.0000000000000000000000000000000000000000000000000000000000000000000000000000000 8888 98 2566V~1888V TEST VOLTAGE START LOAD RESET SELECT TOUT V2 Q+ RIV 1888V~1888V V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2 V2

Figure 13. Measuring Frequency

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

  • Press Hz button to select frequency measurement mode at m V\~ or V\~ or V=Press Hz to exit frequency measurement mode and return to the previous meauremnet mode.
    ● The maximum range of frequency is 1MHz.
  • The signal attenuation is different from different measuring position and range. Therefore the required signal input amplitude and range may be different from different measuring position and range.
  • It is recommended to use mV...range which has higher input sensitivity.
  • If the input amplitude is higher than 1Vrms, it is required to use V== or V\~. to carry out the measurement.
  • When using current range to measure frequency, it is required to input singal more than 30% of full range. The reading obtained is only for reference.

Note:

- When Frequency measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Measuring Temperature (UT532 and UT533 only) (see figure 14)

WALUATION MULTICOMETER -88.88 °C F°C C MULUM mA CCM V °F °C F

Figure 14. Measuring Temperature

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

To avoid risk of shock do not connect thermocouple to electrically live circuits.

Caution

To avoid possible damage to the Meter or other equipment, remember that while the Meter is rated for -40^ to 537^ ( -40^ to 998^ ), the included K-type thermocouple is rated for 260^ ( 500^ ). For temperature out of that range, use a higher rated thermocouple.

To measure temperature, connect the Meter as follows:

  1. Insert the red thermocouple into the VΩH→ terminal and the black thermocouple into the COM terminal.
  2. Set the rotary switch to mV.
  3. Press SELECT to choose temperature range.
  4. Press RANGE to toggle between degrees Celsius (°C) or degrees Fahrenheit (°F).
  5. Place the thermocouple to the object being measured.
    The measured value shows on the display.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Testing Insulation (see figure 15)
UTS32 INSULATION MULTIMETER Auto Range Rated Range 0100V mV -8.8.88 mV°C mA V MHz MeGO 100V -1888V S LUX HCK3 BWE BAT/A RANGE CJ / CLR LOAD BAT / W SELECT POWER + - O- + - 50V 100V 250V 500V 1000V INSULATION CE INSULATION mA COM V Ω + Hz

Figure 15. Testing Insulation

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

Make sure the circuit to be tested is dead before taking measurement. Never measure the insulation resistance of live equipment or circuit. Never touch the circuit after measurement, as the capacitance to be saved on the circuit may cause electric shock.

Insulation tests should only performed on dead circuits. Check the fuse before testing.

To measure insulation resistance, set up the Meter as shown in Figure 15 amd follow the steps below:

  1. Insert the red short lead with alligator clip in the INSULATION input terminal and the black short lead with alligator clip in the mA input terminal.
  2. Turn the rotary switch to INSULATION position.
  3. Press RANGE to select he voltage.
  4. Connect the probes to the circuit to be measured. There are two methods to start the insulation test:

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

  • Press HOLD button to disable LOCK mode. Press and hold TEST to start the insulation test until TEST button is released
  • Press HOLD button to enter LOCK mode. Press TEST once to start the insulation test until pressing HOLD or TEST again to release LOCK mode.

  • Push and hold TEST button to start the test, the secondary display shows the test voltage applied to the circuit under test. The high voltage symbol along with a primary display showing the resistance in MΩ or GΩ. The TEST icon appears on the lower portion of the display.

  • When releasing the TEST button, LCD will not appear TEST icon. The resistance reading appears on the primary display until a new test is started or a different function or range is selected.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Uni-T UT533 - UNI-T® - 1

Caution

The following improper operating procedure may force the Meter to power off automatically. To avoid the danager of electric shock and damage to the Meter:

  • Replace batteries as soon as battery indicator is on. Initiating high voltage when the battery indicator is on, which may force the Meter to power off.
  • Must connect the probe to the circuit being measured before initiating high voltage.
  • When under high voltage mode, never turn the rotary switch to other position.
  • The testing time of small resistor or short circuit cannot exceed 20 seconds.
  • At insulation testing mode with testing voltage initiated, auto power off feature will be disabled until un-initiating the testing voltage.

Cleaning

  • Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Dirt or moisture in the terminals can affect readings.
    ● Turn the Meter off when it is not in use.
  • Do not use or store the Meter in a place of humidity, high temperature, explosive, inflammable and strong magnetic field.
  • The internal circuit of the Meter shall not be altered at will to avoid damage of the Meter and any accident.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Replacing the Batteries

Replace the batteries as shown in Figure 16. Follow the steps below to replace the batteries.

Uni-T UT533 - Replacing the Batteries - 1

natural_image Technical line drawing of a vehicle chassis with open and closed compartments (no text or symbols)

Figure 16. Replacing the Fuse and Battery

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Warning

To avoid shock, injury, or damage to the Meter:

  • To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as the battery indicator 📋 appears.
    ● Turn the Meter off and remove the testing leads from the terminals.

  • Remove the two screws from the battery door by using a standard screwdriver.

  • Open the battery door.
  • Remove and replace the batteries. 6pcs of AA (AM3/LR6), standard capacity around 2450mAh.
  • Rejoin the case bottom and battery compartment, and reinstall the screw.

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Replacing the Fuse

Replace the fuse as shown in Figure 16. Follow the steps below to replace the batteries.

Uni-T UT533 - Replacing the Fuse - 1

Warning

To avoid shock, injury, or damage to the Meter:

  • Use ONLY fuses with the amperage and voltage specified.
  • Turn the Meter off and remove the testing leads from the terminals.
  • Remove the two screws form the battery door, two screws from the case bottom, one screw at he back of the battery.
  • Open the battery door and the case bottom
  • Remove the fuse by gently prying one end loose, and take out the fuse from its bracket.
  • Install ONLY replacement fuses with identical amperage and voltage specified: F1A 240V
  • Rejoin the battery door and the case bottom, and reinstall the screws.
  • Do not operate the Meter with the case opened.

Replacement of the fuse is seldom required. Buring of a fuse always results from improper operation.

Specifications

Table 6. General Specifications

Maximum Voltage Applied to any Terminal1000V rms
mA Input Terminal1A / 240V Fast Fuse, High breaking capacity Fuse
Maximum Display6000, Updates 5 times per second
Operating Temperature0°C ~40°C (32°F~104°F)
Storage Temperature-10°C ~50°C (14°F~122°F)
Relative Humidity≤75% @ 0°C to below 31°C≤50% @ 31°C to above 40°C
AltitudeOperating: 2000 meterStorage: 10000 meter
BatteriesSix AA 1.5V batteries AM3/LR6 (standard capacity around 2450mAh)
Low batteryLCD display
Auto Power OffThe Meter will be automatically off when the Meter is not in use for more than 15 minutes
Size97mm x 202mm x 46mm
WeightAround 505 g (including batteries)

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Electrical Specifications

Accuracy: ±(a % of rdg + b digits), guarantee for 1 year.

Operating Temperature: 23°C±5°C

Relative Humidity: <75%

AC Voltage Measurement

RangeResolutionAccuracyInput Protection
50Hz ~60Hz60Hz ~ 1kHz
6.000V1mV± (1%+3)± (2%+3)1000V ≈
60.00V10mV
600.0V100mV± (2%+6)
1000V1V± (2%+3)

Remarks:

• Input Impedance: ≥10MΩ
● Displays effective value of sine wave
● Frequency Response: 50Hz\~1kHz

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

DC Voltage Measurement

RangeResolutionAccuracyInput Protection
6.000V1mV± (0.3%+4)1000V ≈
60.00V10mV
600.0V100mV
1000V1V

Remark:

• Input Impedance: ≥10MΩ

DC Millivolts Measurement

RangeResolutionAccuracyInput Protection
60.00mV10μV± (0.4%+4)600Vp
600.0mV100μV

Remark:

● Input Impedance: ≥4000MΩ

AC Current Measurement

RangeResolutionAccuracyInput Protection
60.00mA10μA± (1.5%+2)Fuse, Fast, 1A, 240V
600.0mA100μA

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Remark:

  • Displays True RMS value.
    ● Frequency Response 50Hz \~ 1kHz

DC Current Measurement

RangeResolutionAccuracyInput Protection
60.00mA10μA± (1.0%+2)Fuse, Fast, 1A, 240V
600.0mA100μA

Ohms Measurement

RangeResolutionAccuracyInput Protection
600.0Ω0.1Ω± (0.9%+2)600Vp
6.000kΩ
60.00kΩ10Ω
600.0kΩ100Ω
6.000MΩ1kΩ
40.00MΩ10kΩ± (1.5%+3)

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

Capacitance Measurement

RangeResolutionAccuracyInput Protection
10.00nF10pF±(3%+5)600Vp
100.0nF100pF
1000nF1nF
10.00μF10 nF
100.0μF100nF

Diode Test

• Resolution: 1mV
● Input Protection: 500Vp
- Open circuit voltage: around 3V
● Display approximate forward voltage drop nearest value.

Continuity Test

● Continuity Indication: Continuous audible tone for test resistance below 30Ω.
- Open Circuit Voltage: around 3V
● Input Protection: 500Vp
● Resolution: 0.1Ω

Temperature Measurement (UT532 and UT533 only)

RangeResolutionAccuracyInput Protection
-40°C~537°C1°C± (1%+10)600Vp
-40°F~998°F2°F± (1%+18)

Frequency Measurement

  • At mV... Range
RangeResolutionAccuracyInput Protection
60.00Hz0.01Hz± (0.1%+3)600Vp
600.0Hz0.1Hz
6.000kHz1Hz
60.00kHz10Hz
600.0kHz100Hz
1.000MHz1kHz

Remark:

- Input Sensitivity:

≤When 100kHz: ≤100mV rms

≥ 30mV rms

When 100kHz: ≥100mV rms

≤30V rms

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

- At DCV/ACV Range

RangeAccuracyInput Protection
10Hz~100kHz± (0.1%+3)600Vp

Remark:

● Input amplitude: ≥ 600mV rms

Insulation Specifications

• UT531

Testing VoltagesMeasurement RangeResolutionAccuracy
500V(0% to 20%)0.5MΩ ~99.9MΩ0.1MΩ± (3%+5)
100MΩ ~ 500MΩ1MΩ
1000V(0% to 20%)4.0MΩ ~ 99.9MΩ0.1MΩ
100MΩ ~ 600MΩ1MΩ

Remark:

● Short-Circuit Test Current less than 2.0mA

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

• UT532

Testing VoltagesMeasurement RangeResolutionAccuracy
250V(0% to 10%)0.2MΩ ~ 99.9MΩ0.1MΩ± (3%+5)
100MΩ ~ 250MΩ1MΩ
500V(0% to 20%)0.5MΩ ~ 99.9MΩ0.1MΩ
100MΩ ~ 500MΩ1MΩ
1000V(0% to 20%)4.0MΩ ~ 99.9MΩ0.1MΩ
100MΩ ~ 999MΩ1MΩ
1.00MΩ ~ 2.00GΩ10MΩ± (5%+5)

Remark:

● Short-Circuit Test Current less than 2.0mA

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

• UT533

Testing VoltagesMeasurement RangeResolutionAccuracy
50V(0% to 10%)0.1MΩ ~ 50.0MΩ0.1MΩ± (3%+5)
100V(0% to 10%)0.1~100.0MΩ0.1MΩ
250V(0% to 10%)0.2MΩ ~ 99.9MΩ0.1MΩ
100MΩ ~ 250MΩ1MΩ
500V(0% to 20%)0.5MΩ ~ 99.9MΩ0.1MΩ
100MΩ ~ 500MΩ1MΩ
1000V(0% to 20%)4.0MΩ ~ 99.9MΩ0.1MΩ
100MΩ ~ 999MΩ1MΩ
1.00GΩ ~ 2.00GΩ10MΩ± (5%+5)

Remark:
● Short-Circuit Test Current less than 2.0mA

* END **

UNI-T®

UT531/UT532/UT533 : USERS MANUAL

UNI-T.

UNI-TREND TECHNOLOGY (CHINA) CO., LTD.

No6. Gong Ye Bei 1st Road.

Songshan Lake National High-Tech Industrial

Development Zone, Dongguan City,

Guangdong Province, China

Tel: (86-769) 8572 3888

http://www.uni-trend.com

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Uni-T

Model : UT533

Category : Mètre