Uni-T UT512 - Mètre

UT512 - Mètre Uni-T - Free user manual and instructions

Find the device manual for free UT512 Uni-T in PDF.

📄 2 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice Uni-T UT512 - page 1
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Insulation Resistance Tester
Brand Uni-T
Model UT512
Dimensions (H x W x L) 202 x 155 x 94 mm
Weight Approx 2 kg (including batteries)
Power Supply 8 × 1.5V LR14 (R14) batteries or optional power adaptor (230V AC, 50/60Hz, 50mA input, DC 14.5V, 600mA output)
Display LCD with 9999 counts and analog bar graph, backlight
Insulation Resistance Ranges 500V (0.5MΩ – 5.0GΩ), 1000V (2MΩ – 10.0GΩ), 2500V (5MΩ – 20.0GΩ), 5000V (10MΩ – 100GΩ)
Voltage Measurement DC: ±30V to ±600V, AC: 30V to 600V (50/60Hz)
Special Functions Timed measurement (up to 15 min), Polarization Index (PI), Compare (Pass/Fail) with adjustable limits, Data storage (18 readings), USB data transfer, Sleep mode (15 min auto-off)
Safety Rating CAT III 600V, Double Insulation, CE certified
Operating Temperature -10°C to 40°C
Storage Temperature -20°C to 60°C
Relative Humidity ≤85% at -10°C to 40°C, ≤90% at -20°C to 60°C
Computer Interface USB (with included cable and software)
Included Accessories Test leads (black, green, red), 8 × LR14 batteries, USB cable, software CD, tool box, manual
Maintenance Wipe with damp cloth and mild detergent; replace batteries when indicator shows empty; use only specified power adaptor
Spare Parts / Repairability Optional power adaptor (extra cost); contact dealer for missing or damaged items. No user-serviceable parts inside.
Standards Compliance EN 61010-1:2010, Pollution Degree 2, CAT III 600V, Double Insulation

Frequently Asked Questions - UT512 Uni-T

How do I measure insulation resistance with the UT512?
Press the IR button to select insulation resistance mode. Use the and buttons to choose a test voltage (500V, 1000V, 2500V, or 5000V). Connect the red test lead to the LINE terminals, the black to GUARD, and the green to EARTH. Attach the alligator clips to the circuit to be measured (make sure the circuit is de-energized). Press and hold the TEST button for 1 second to start the measurement. The high voltage output will begin, and the resistance value will be displayed.
How do I measure DC or AC voltage?
Press DCV or ACV to select voltage measurement. Insert the red test lead into the LINE terminal and the green into EARTH. Connect the probes to the circuit. The meter displays the voltage. Do not exceed 600V DC or 600V AC RMS.
How do I replace the batteries?
Turn off the meter and remove all test leads. Open the battery compartment by removing the screw at the back. Replace the 8 batteries with new 1.5V LR14 (R14) cells, observing correct polarity. Close the compartment and tighten the screw.
What is the Polarization Index (PI) function?
PI is the ratio of insulation resistance measured at two times (typically 1 minute and 10 minutes). The meter can automatically perform this test. Press TIME to enter timed mode, set TIME 1 and TIME 2 using and , then press TIME again to select PI. Press and hold TEST for 2 seconds. After the test, the PI value is displayed.
How do I use the Compare (Pass/Fail) function?
Press COMP to enable compare mode. Use and to set a threshold resistance value (e.g., 10MΩ). When performing an insulation test, the meter shows GOOD if the measured resistance is above the threshold, or NG if below.
How can I store and recall measurements?
To store a reading, press SAVE. The meter can store up to 18 readings. To recall, press LOAD, then use and to scroll through stored values. Press LOAD again to exit recall mode. To clear all stored data, press and hold CLEAR/※.
What does the battery indicator mean?
The battery indicator on the LCD shows four segments. When all segments are displayed, the battery is fully charged. When only one segment remains, the battery is nearly empty (10.7V–11.1V) and replacement is needed. If the indicator shows a hollow battery symbol, the voltage is below 5.9V and the meter should not be used.
Can I use the meter with a power adaptor?
Yes, an optional power adaptor (input 230V AC, 50/60Hz, output DC 14.5V, 600mA) is available. Connect it to the side input terminal. It is recommended to remove the batteries when using the adaptor to avoid potential leakage.
How do I transfer data to a computer?
Install the included software from the CD. Connect the meter to your computer via the USB cable. Press the USB button on the meter to start data transfer; the USB icon appears on the display. Press again to stop.
What safety precautions should I take?
Never apply more than 600V to the meter. Do not use in explosive or wet environments. Always discharge capacitive circuits after high voltage measurements. Keep fingers behind the guards on test leads. Use only specified test leads and power adaptor. Do not attempt to repair unless qualified.

User questions about UT512 Uni-T

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Mètre in PDF format for free! Find your manual UT512 - Uni-T and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. UT512 by Uni-T.

USER MANUAL UT512 Uni-T

Insulation Resistance Tester

Introduction

Uni-Trend Model UT512 insulation resistance tester (hereafter, "the Meter") is a handheld instrument designed primarily to make resistance/ insulation resistance measurement.

Unpacking the Meter

The Meter includes the following items:

able 1. Unpacking Inspection

ItemDescriptionQty
1English Operating Manual1 pc
2One-plug test lead to one alligator clip (Black)1 pc
3One plug test lead to one alligator clip (Green)1 pc
4Two-plug test lead to one alligator clip (Red)1 pc
51.5V Battery (LR14 / R14)8 pcs
6Tool Box1 pc
7USB Interface Cable1 pc
8Software1 pc
9Power adaptor (input voltage 230V, 50/60Hz, 50mA, output DC14.5V, 600mA) (optional, available at extra cost)1 pc

In the event you find any missing or damage, please contact your dealer immediately.

Safety Information

This Meter complies with EN 61010-1:2010 safety measurement requirement: Pollution Degree 2, measurement category CAT III 600V and Double Insulation.

CAT II (measurement category): Test and measuring circuits connected directly to utilization points (socket outlets and similar points) of the low-voltage MAINS installation.
CAT III (measurement category): Test and measuring circuits connected to the distribution part of the building's low-voltage MAINS installation.

Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired.

Danger identifies conditions and actions that pose hazard(s) to the user.
⚠ Warning alerts users to avoid electric shock.
Caution identifies conditions and actions that may damage the Meter and affect accurate measurement.
Operating Caution identifies conditions that user needs to take extra care during operating the Meter

Danger

Use of instrument in a manual not specified by the manufacturer may impair safety features/protection provided by the equipment. Read the following safety information carefully before using or servicing the instrument.

  • Do not apply more than 600V.
  • Do not use the Meter around explosive gas, vapor or dust.
  • Do not use the Meter in a wet environment.
  • When using the test leads, keep your figures away from the lead contacts. Keep your figures behind the finger guards on the leads.
  • Do not use the Meter with any parts or cover removed.
  • When carrying out insulation measurement, do not contact the circuit under test.

Warning

  • Do not use the Meter if it is damaged or metal part is exposed. Look for cracks or missing plastic.
  • Be careful when working above 33V rms, 46.7V ac rms or 70V DC. Such voltages pose a shock hazard.
  • Discharge all loading of circuit under test after measuring high voltage.
  • Do not change battery when the Meter is in wet environment.
  • Place test leads in proper input terminals. Make sure all the test leads are firmly connected to the Meter's input terminals.

● Make sure the Meter is turned off when opening the battery compartment.

Caution

  • When performing resistance tests, remove all power from the circuit to be measured and discharge all the power.
  • When servicing the Meter, use only only the test leads and power adaptor with the same model number or identical electrical specifications.
  • Do not use the Meter if the battery indicator (☐) shows a battery empty condition. Take the battery out from the Meter if it is not used for a long time.
  • Do not use or store the Meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after dampened.
  • Soft cloth and mild detergent should be used to clean the surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident.
  • Dry the Meter before storing if it is wet.

International Electrical Symbols

International symbols on the Meter and in this manual are explained in Table 2.

Table 2. International Electrical Symbols

Risk of electric shock
Equipment protected throughout by DOUBLE INSULATION or REINFORECD INSULATION
Direct current
Alternating current
Grounding
Caution
Empty of Built-In Battery
Conforms to Standards of European Union

Battery Saver (Sleep Mode)

The Meter enters the Sleep Mode and blanks the display if there is no button press for 15 minutes. This is done to conserve battery power. The Meter comes out of Sleep Mode when ON/OFF button is pressed and hold for 1 second.

Battery Indication

There is a battery indicator shown on the upper left corner of the display. Please refer to Table 3 for detailed explanation.

Table 3. Battery Indication

Battery IndicatorBattery Voltage
5.9V ~ 10.6V. It means the battery is empty, Do not use the Meter as it cannot guarantee accuracy.
10.7V ~ 11.1V. It means the battery is almost empty, replacing battery is necessary. At this status, the Meter can still output 500V and 1000V to measure, the measured accuracy will not be affected.
11.2V ~ 12.2V
12.2V or more

The Meter Structure

Below Figure 1 and Table 4 shows the Meter front structure and description

1 2 3 4 5 6 7 8 9 10-11 1 2 3 4 5 6 7 8 9 10-11 17 16 15 14 13 12 17 16 15 14 13 12 22 Green Black Red

Figure 1. The Meter Front Structure

Table 4. Meter Front Description

1LCD
2◀Arrow Button
3Emergency Stop
4Data Clear the Display Backlight Button
5▼Arrow Button
6On/Off Button
7Compare Button
8Insulation Resistance Button
9DC Voltage Measurement Button
10Timer Button.
11AC Voltage Measurement Button
12Test Button
13USB Button
14Data Store Button.
15Data Recall Button
16▶Arrow Button
17▲Arrow Button
18LINE: High voltage input terminal( Connected to two-plug red test lead )
19High voltage line shielding input terminal( Connected to two-plug red test lead )
20GUARD: Grounding protection input terminal( Connected to one-plug black test lead )
21EARTH: High resistance measurement input terminal ( Connected to one-plug green test lead )
22Testing leads:Two-plug red test lead to one alligator clip.One-plug black test lead to one alligator clip.One-plug green test lead to one alligator clip.

Below Figure 2 and Table 5 shows the Meter side structure and description

Uni-T UT512 - The Meter Structure - 2
Figure 2. The Meter Side Structure

Table 5. Meter Side Description

1Safety Shutter
2Power adaptor Input Terminal
3USB Port

Display

Table 6 and Figure 3 describe the display.
3 8 16 2 1 15 9 17 10 18 21 18 20 88.88VΩ 88.88VΩ 88.88VΩ 88.88VΩ 88.88VΩ 88.88VΩ 88.88VΩ 88.88VΩ 88.88VΩ 88.88VΩ 88.88VΩ 88.88VΩ 88.88VΩ 88,79 10 11 12 13 COMP Time1 Time2 SAVE LOAD F1 4 14 7 19 5 6 11 12 13

Figure 3. Display

Table 6. Display Description

NumberMeaning
1Analogue bar graph
2Risk of electric shock
3Battery life indicator
4Indicator for timer
5Timer 1 symbol
6Timer 2 symbo
7Indicates selected pass/fail compare value
8Indicator for power adaptor
9The continuity buzzer is on
10Indicator for data store full
11Data store is on
12Data recall is on
13Indicator for polarization index
14Step symbol
15Compare result: Pass
16Compare result: Fail
17Indicator for DC voltage
18Indicator for AC voltage
19Indicates negative reading
20Unit symbols
21Indicator for clearing

Key Functions

Table 7. Key Description

ON/OFFTurn on or off the Meter. Press and hold the button for 1 second to turn the Meter on. Press again to turn off the Meter.The Meter defaults at 500V range and under continuous measurement of insulation resistance when turned on.
E-STOPEmergency stop button. Press this button when the Meter crashes down and cannot turn off the power.
CLEAR / ※Short press to turn on or off the display backlight Long press to clear the stored data.
SAVEPress to store the current measurement value. The maximum number of stored reading is 18. When the stored readings memory is full, the Meter shows FULL and stop storing. Press and hold CLEAR / ※ to clear the stored value in order to store the next measurement value.
LOAD● Press once to recall the first stored value. ● Press again to exit Load feature. ● Load feature can only be used when there is no high voltage output.
● When the insulation resistance measurement has no testing voltage output, press to select previous voltage range. ● Under load mode: press to recall the previous stored value.
● When the insulation resistance measurement has no testing voltage output, press to select next voltage range. ● Under load mode: press to recall the next stored value.
● When setting the timer for the measurement of insulation resistance or polarization index, press to decrement the time. The maximum length of time is 15 minutes and 30 seconds, the Meter will automatically carry out measurement. ● When compare function is enabled for insulation resistance measurement, press to decrement a resistance comparing value. ● After polarization index measurement, press to display polarization index, TIME 2 and TIME 1 insulation resistance values in sequence.
● When setting the timer for the measurement of insulation resistance or polarization index, press to increment the time. The maximum length of time is 30 minutes and 30 seconds, the Meter will automatically carry out measurement. ● When compare function is enabled for insulation resistance measurement, press to increment a resistance comparing value. ● After polarization index measurement, press to display polarization index, TIME 2 and TIME 1 insulation resistance values in sequence.
USB● Press once to start the data transferring to the computer via USB, USB symbol shows on the display. ● Press again to stop the data transferring to the computer via USB, USB symbol disappears.
COMPSet a pass / fail limit for insulation tests. The default value is 10M Ω
TIMEPress to step through continuous, timed and polarization index measurements in sequence.
TESTPress to turn on/off the output of insulation resistance test voltage.
IRPress to initiate insulation resistance measurement
DCVPress to initiate DC voltage measurement
ACVPres to initiate AC voltage measurement

Measurement Operation

This section explains how to make measurements.

Press and hold ON/OFF to turn on the Meter, press again to turn off the Meter. The meter defaults at 500V range and continuous measurement of insulation resistance when turned on.

A. Measuring Voltage
Red Green V COM

Figure 4. Voltage Measurement

Operating Caution

● To avoid harm to you or damage to the Meter, please do not attempt to measure voltages higher than 600V or 600V rms, although readings may be obtained.
- Special care should be taken when measuring high voltage.

To measure voltage, set up the Meter as Figure 4 and do the following:

  1. Press DCV or ACV button to select DC voltage or AC voltage measurement

  2. Insert the red and green test leads into EARTH and two LINE terminals.

  3. When measuring DC voltage, if negative voltage is present on the red test lead, "-" symbol will show on the display.

Note

- When voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads away from the input terminals of the Meter.

B. Measuring Insulation Resistance
Red Green Black Groud

Figure 5. Insulation Resistance Measurement

Operating Caution

  • When performing insulation resistance tests, remove all power from the circuit to be measured and discharge all the power.
  • Operating the Meter must be very careful as it outputs dangerous voltage during measurement. Must make sure the tested object is firmed clipped, hands are away from the clips, then press TEST button to output high voltage.
  • Do not short circuit the testing leads during high voltages output or test insulation resistance after high voltages output. This kind of incorrect operating may cause sparking and fire, which damages the Meter and cause personal injury.
  • Do not measure over 10 seconds when: measuring resistance < 2M with use of 500V. measuring resistance < 5M with use of 1000V. measuring resistance < 8M with use of 1500V. measuring resistance < 10M with use of 2500V.

To measure insulation resistance, set up the Meter as Figure 5 and do the following:

  1. Press IR button to select insulation resistance measurement.
  2. When there is no testing voltage output, press ▲ and ▼ button to select 500V, 1000V, 2500V or 5000V voltage range.
  3. When performing insulation resistance tests, remove all power from the circuit to be measured and discharge all the power.
  4. Insert the red test lead into two LINE terminals, the black one into GUARD and the green one into EARTH.
  5. Connect the red and green alligator clip to the circuit to be measured, negative voltage outputs from LINE terminal.
  6. Choose one of insulation resistance measurement modes shown as below:

a) Continuous Measurement

  • Press TIME button to select continuous mode, there is no timer icon on the LCD
  • Press ◀ and ▶ hold TEST button for 1 second to begin and output insulation resistance test voltage, TEST button lights up, ⚠ blinks on every 0.5 seconds.
  • Press TEST button to turn off the voltage output, when measurement is completed. TEST button lights off, △ disappears. The LCD shows the current insulation resistance measurement value.

b) Timed Measurement

  • Press TIME button to select timed mode, the LCD displays TIME 1 and G symbols.
  • Press ◀ and ▶ buttons to set the time (00:10\~15:00). Within 1 minute, the time increment or decrement by every 5 seconds. Afterward, the time increment or decrement by every 30 seconds.
  • Then press and hold TEST button for 2 second to carry out timed measurement. TIME 1 and △ are displayed and blinked on the LCD on every 0.5 seconds.
  • When the set time is reached, the test voltage output will be turned off, and the measurement will be automatically stopped. The LCD displays the insulation resistance reading.

c) Polarization Index (PI) Measurement

  • Press TIME button to select timed mode, the LCD displays TIME 1 and Ⓧ symbols.
  • Press ◀ and ▶ buttons to set the time (00:10\~15:00). Within 1 minute, the time increment or decrement by every 5 seconds. Afterward, the time increment or decrement by every 30 seconds.
  • Press TIME button again. TIME 2, PI and ⏻ symbols appear on the LCD.

  • Press ◀ and ▶ buttons to set the time (00:15\~15:30). Within 1 minute, the time increment or decrement by every 10 seconds. Afterward, the time increment or decrement by every 30 seconds.

  • Then press and hold TEST button for 2 seconds to carry out the measurement.
  • TIME 1 and are displayed and blinked on the LCD on every 0.5 seconds before TIME 1 set time is reached.
  • TIME 2 and ▲ are displayed and blinked on the LCD on every 0.5 seconds before TIME 2 set time is reached.
  • When the two set time are reached, the test voltage output will be turned off and the measurement will be automatically stopped. The LCD displays the polarization index reading.
  • Press ◀▶ to step through the polarization index, TIME 2 and TIME 2 insulation resistance readings.

Calculation Tips:

PI = 3-minute \~ 10-minute resistance/30-second \~ 1-minute resistance

PI4 or more4~22.0~1.01.0 or less
StandardThe bestGoodWarningBad

d) Compare Function

  • Press COMP button to select compare feature. COMP symbol displays on the LCD.
  • Press and buttons to set the compare value

- You can choose the compare value from 10M Ω, 20MΩ, 30MΩ, 40MΩ, 50MΩ, 60MΩ, 70MΩ, 80MΩ, 90MΩ, 100MΩ, 200MΩ, 300MΩ, 400MΩ, 500MΩ, 600MΩ, 700MΩ, 800MΩ, 900MΩ, 1GΩ, 2GΩ, 3GΩ, 4GΩ, 5GΩ, 6GΩ, 7GΩ, 8GΩ, 9GΩ, 10GΩ, 20GΩ, 330GΩ, 40GΩ, 50GΩ, 60GΩ, 70GΩ, 80GΩ, 90GΩ, 100GΩ

● Press and hold TEST button for 2 seconds to carry out the measurement.
- The NG symbol will display if the insulation resistance value is smaller than compare value. Otherwise GOOD symbol will be displayed.

Using Power Adaptor

Refer to Figure 6 for the use of power adaptor.

Uni-T UT512 - Using Power Adaptor - 1

natural_image Line drawing of an electronic device with a power outlet and two cables (no text or symbols)

Figure 6. Using Power Adaptor

  1. Open the side safety shutter, then you will see there is a power adaptor input terminal.
  2. Make sure the Meter is power off and Insert the UT512 power adaptor to the input terminal.
  3. It is highly recommended to take out all the batteries when you are using the power adaptor.
  4. Make sure the Meter is power off when you
    disconnect the UT512 power adaptor from the Meter. 5. It is highly recommended to use Uni-Trend supplied UT512 power adaptor to avoid unnecessary danger

Connecting USB Interface

See Figure 7 for USB interface connection.

Uni-T UT512 - Connecting USB Interface - 1

natural_image Line drawing of a portable electronic device with attached cables and a power outlet (no text or symbols)

Figure 7. USB Interface Connection

  1. Install the included software, the installation guide can be seen from the CD.
  2. Open the side safety shutter, then you will see there is a USB port.
  3. Insert the included USB cable to the Meter's USB port and the other end to the computer.

Maintenance

This section provides basic maintenance information including battery replacement instruction.

Warning

Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information.

A. General Service

  • Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
  • To clean the terminals with cotton bar with detergent, as dirt or moisture in the terminals can affect readings.

● Turn the Meter to OFF when it is not in use.
● Take out the battery when it is not using for a long time
- Do not use or store the Meter in a place of humidity, high temperature, explosive, inflammable and strong magnetic field.
- If the Meter is wet, dry it before use.

B. Replacing the Battery
Uni-T UT512 - General Service - 1

natural_image Line drawing of a device with a lid and internal compartments (no text or symbols)

Figure 8. Battery Replacement

Warning

To avoid electric shock, remove all the test leads from the Meter when replacing the batteries.

Operating Caution

  • Do not mix to use old and new batteries.
  • Be careful the polarity is correct when installing batteries.
  • Do not use the Meter if the battery indicator (☐) shows.
  • Do you carry out measurement during the battery compartment is open.

Follow Figure 8 and proceed as follows to replace the battery:

● Turn the Meter to OFF and remove all connections from the terminals.
- Remove the screw from the battery compartment, and separate the battery compartment from the case bottom.
- Replace with 8pcs of new 1.5V (LR14) batteries. - Rejoin the case bottom and battery compartment and reinstall the screw.

Specifications

Safety and Compliances

CertificationCE
CompliancesIEC 61010 CAT.III 600V overvoltage and double insulation standard

General Specifications

Display (LCD)Digital: 9999 counts Analog bar graph.
Display BacklightBright backlight for clear readings in poorly lighted areas.
Computer connectionVia USB interface.
Data Logging and Recall18
AutorangeThe Meter automatically selects best range
Warning and red light will on.
Test VoltageAutomatically source the voltage.
COMP MeasurementUse the Compare function to set a pass/fail compare level for the insulation measurements.
PI MeasurementPolarization Index is the ratio of insulation resistance. Preset the timer for two points and the Meter will carry out the measurement automatically.
TIMETo carry out measurement by setting a specified time within 15 minutes.
OverloadingDisplay OL on insulation resistance range
Battery IndicatorDisplay
Icon DisplayEquips with function and battery indicator icons.
Current ConsumptionMaximum: around 600mAAverage: around 20mA
Operating Temperature-10 °C~40 °C14 ~F04 ) °F
Storage Temperature-20 °C~60 °C4 ~F52 ) °F
Relative Humidity ≤slant 85% @ -10°e40 below; ≤slant 90% @ -20°e60 °C
Battery Type8pcs of 1.5V (LR14) batteries or power adaptor (input voltage 230V, 50/60Hz, 50mA, input DC14.5V, 600mA).Power adaptor is optional at extra cost.
Dimensions (HxW xL)202 x 155 x 94 mm
WeightApprox 2kg (including battery)

Accuracy Specifications

Accuracy: ±([% of reading] + [number of least significant digits), guarantee for 1 year. Operating temperature: 18°C\~28°C Relative humidity: 45\~75%RH

A. Voltage Measurement

DC VoltageAC Voltage
Measurement Range±30 ~ ±600V30V~600V (50/60Hz)
Resolution1V
Accuracy ±(2%+5) <100V: ± 2%+8 ≥100V: ± 2%+5

B. Insulation Resistance Measurement

Output Voltage500V1000V1500V2500V
Display Range0.5MΩ~5.0GΩ2MΩ~10.0GΩ5MΩ~20.0GΩ10MΩ~100GΩ
Open Circuit VoltageDC 500V + 20%,-0%DC1000V + 20%,-0%DC 2500V + 20%,-0%DC2500V + 20%,-0%
Test Current1mA~1.2mA @500kΩ1mA~1.2mA @ 1MΩ1mA~1.2mA @ 1.5MΩ1mA~1.2mA @ 2.5MΩ
Short CircuitLess than 2.0mA
Accuracy100kΩ ~ 100MΩ (3%+5)100MΩ ~ 10GΩ (5%+5)10GΩ ~ 100GΩ (10%+5)

Operating Caution

At any output voltage, when the tested resistance is less than 10M the testing time cannot exceed 10 seconds continuously.

UNI-T®

UNI-TREND TECHNOLOGY (CHINA) CO., LTD.

No6, Gong Ye Bei 1st Road, Songshan Lake National High-Tech Industrial Development Zone, Dongguan City, Guangdong Province, China Tel: (86-769) 8572 3888 http://www.uni-trend.com

\*END\*