TEKTRONIX FCA3000 - Spectrum Analyzer

FCA3000 - Spectrum Analyzer TEKTRONIX - Free user manual and instructions

Find the device manual for free FCA3000 TEKTRONIX in PDF.

📄 8 pages English EN Download 💬 AI Question
Notice TEKTRONIX FCA3000 - page 5
View the manual : Français FR English EN
Pick your language and provide your email: we'll send you a specifically translated version.

User questions about FCA3000 TEKTRONIX

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Spectrum Analyzer in PDF format for free! Find your manual FCA3000 - TEKTRONIX and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. FCA3000 by TEKTRONIX.

USER MANUAL FCA3000 TEKTRONIX

FeatureFCA3100 SeriesFCA3000 Series
Frequency Resolution12 digit/s12 digit/s
Time Resolution50 ps100 ps
Voltage Resolution1 mV3 mV
Meas. Speed to Internal Memory250k measurement/s250k measurement/s
3.5M results750k results
Talker-only Output (GPIB/USB)4k measurement/sNo
Individually Triggered Measurements650/s500/s
Block Transfer Speed15k measurement/s5k measurement/s
Frequency/Period, Time, Phase, Volt, Duty Cycle, Pulse, Rise TimeYesYes
Graphic Display of Trend, Histogram, Modulation DomainYesYes
Totalize, TIEYesNo
Programmable Pulse OutputYesNo
Continuous MeasurementsYesNo

All measurements are displayed with a large main parameter value and smaller auxiliary parameter values (with less resolution). Some measurements are only available as auxiliary parameters.

Frequency A, B, C

CharacteristicDescription
ModeNormal, Back-to-Back (FCA3100 Series)
Range
Input A, B0.001 Hz to 300 MHz
Input C3 or 20 GHz
Resolution12 digits in 1 s measuring time (normal) 11 digits in 1 s measuring time (back-to-back)
Aux ParametersVmax, Vmin, Vp-p

Frequency Burst A, B, C

(FCA3020 and FCA3120 - 20 GHz Only)

Frequency and PRF of repetitive burst signals can be measured without an external control signal and with selectable-start arming delay.

CharacteristicDescription
FunctionsFrequency in burst (in Hz) PRF (in Hz)
RangeInput A, B, C: See Frequency spec.
Minimum Burst DurationDown to 40 ns
Minimum Pulses in Burst
Input A, B3 (6 above 160 MHz)
Input C3 × prescaler factor
PRF Range0.5 Hz to 1 MHz
Start Delay10 ns to 2 s, 10 ns resolution
Aux ParametersPRF

Period A,B,C

CharacteristicDescription
ModeSingle, Average, Back-to-Back (FCA3100 Series)
Range
Input A, B3.3 ns to 1000 s (single, average) 4.0 μs to 1000 s (back-to-back)
Input C10 ns down to 50 ps
Resolution100 ps (single); 12 digit/s avg. (FCA3000 Series) 50 ps (single); 12 digit/s avg. (FCA3100 Series)
Aux ParametersVmax, Vmin, Vp-p

Ratio A/B, B/A, C/A, C/B

CharacteristicDescription
Range(10-9) to 1011
Input Frequency
Input A, B0.1 Hz to 300 MHz
Input C3 or 20 GHz
Aux ParametersFreq 1, Freq 2

Time Interval A to B, B to A, A to A, B to B

CharacteristicDescription
RangeNormal calculation: 0 ns to +106s Smart calculation: -106s to +106s
Resolution100 ps single (FCA3000 Series) 50 ps single (FCA3100 Series)
Min Pulse Width1.6 ns
Smart CalculationSmart Time Interval to determine sign (A before B or A after B)

Positive and Negative Pulse Width A, B

CharacteristicDescription
Range2.3 ns to 106s
Min Pulse Width2.3 ns
Aux ParametersVmax, Vmin, Vp-p

Rise and Fall Time A, B

CharacteristicDescription
Range1.5 ns to 106s
Trigger Levels10% and 90% of signal amplitude
Min Pulse Width1.6 ns
Aux ParametersSlew rate, Vmax, Vmin

Time Interval Error (TIE) A, B

Normalized period back-to-back measurements, calculated as TIE(k) = k *

T_REF - T_i , when T_i = Individual Period Back-to-Back and T_REF = Reference

Period Value.

Positive and Negative Duty Factor A, B

CharacteristicDescription
Range0.000001 to 0.999999
Frequency Range0.1 Hz to 300 MHz
Aux ParametersPeriod, pulse width

Phase A Relative B, B Relative A

CharacteristicDescription
Range-180° to +360°
ResolutionSingle cycle: 0.001° to 10 kHz, decreasing to 1° >10 MHz. Resolution can be improved by averaging (statistics)
Frequency RangeUp to 160 MHz
Aux ParametersFreq (A), Va/Vb (in dB)

Totalize A, B (FCA3100 Series)

CharacteristicDescription
ModeTot A, Tot B, Tot A+B, Tot A-B, Tot A/B
Range1 to 1010 counts
Frequency RangeUp to 160 MHz
Start ControlManual, start arming
Stop ControlManual, stop arming, timed
Aux ParametersOther Totalize functions

Vmax,Vmin,Vp-pA,B

CharacteristicDescription
Range-50 V to +50 V, -5 V to +5 VRange is limited by the specification for max input voltage without damage (see input A, B)
Frequency RangeDC, 1 Hz to 300 MHz
ModeVmax, Vmin, Vp-p
Resolution3 mV (FCA3000 Series)1 mV (FCA3100 Series)
Uncertainty (5 V range, typical)
DC, 1 Hz to 1 kHz1% + 15 mV
1 kHz to 20 MHz3% + 15 mV
20 to 100 MHz10% + 15 mV
100 to 300 MHz30% + 15 mV
Aux ParametersVmin, Vmax, Vp-p

Time Stamping A, B, C

Raw time-stamp data together with pulse counts on inputs A, B, or C, accessible through GPIB or USB only.

CharacteristicDescription
Max Sample SpeedSee GPIB specifications
Max Frequency160 MHz
Time-stamp Resolution100 ps (FCA3000 Series)50 ps (FCA3100 Series)

Input and Output Specifications

Inputs A and B

CharacteristicDescription
Frequency RangeDC Coupled: DC to 300 MHz AC Coupled: 10 Hz to 300 MHz
Impedance1 MΩ / 20 pF or 50 Ω (VSWR ≤ 2:1)
Trigger SlopePositive or negative
Max Channel Timing Difference500 ps
Sensitivity15 mVRMS (DC-200 MHz) 25 mVRMS (200-300 MHz)
AttenuationX1, X10
Dynamic Range (X1)30 mVp-p to 10 Vp-p within ±5 V window
Trigger LevelReadout on display
ResolutionFCA3000 Series: 3 mV FCA3100 Series: 1 mV
Uncertainty (X1)±(15 mV + 1% of trigger level)
AUTO trigger levelTrigger level is automatically set to 50% point of input signal (10% and 90% for rise/fall time)
Auto Hysteresis
TimeMin hysteresis window (hysteresis compensation)
FrequencyOne-third of input signal amplitude
Analog LP FilterNominal 100 kHz, RC type
Digital LP Filter1 Hz to 50 MHz cutoff frequency
Max Voltage without Damage
1 MΩ350 V (DC + AC peak) to 440 Hz, falling to 12 VRMS (X1) at 1 MHz
50 Ω12 VRMS
ConnectorBNC

Input C - 3 GHz (FCA3003 and FCA3103 Products)

CharacteristicDescription
Operating Input Voltage Range
100 to 300 MHz20 mVRMS to 12 VRMS
0.3 to 2.5 GHz10 mVRMS to 12 VRMS
2.5 to 2.7 GHz20 mVRMS to 12 VRMS
2.7 to 3.0 GHz40 mVRMS to 12 VRMS
Prescaler Factor16
Impedance50 Ω nominal, VSWR < 2.5:1
Max Voltage without Damage12 VRMS, pin-diode protected
ConnectorType-N Female

Input C - 20 GHz (FCA3020 and FCA3120 Products)

CharacteristicDescription
Frequency Range0.25 to 20 GHz
Operating Input Voltage Range
250 to 500 MHz-21 to +27 dBm
0.5 to 14 GHz-27 to +27 dBm
14 to 18 GHz-27 to +27 dBm
18 to 20 GHz-21 to +27 dBm
Prescaler Factor128
Impedance50 Ω nominal, VSWR < 2.0:1
AM Tolerance>90% within sensitivity range
Max Voltage without Damage+27 dBm
ConnectorType Precision-N Female

Rear Panel Inputs and Outputs

CharacteristicDescription
Reference Input1, 5, or 10 MHz; 0.1 to 5 VRMS sine; impedance ≥1 kΩ
Reference Output10 MHz; >1 VRMS sine into 50 Ω
Arming InputArming of all measuring functions
ImpedanceApprox. 1 kΩ
Frequency rangeDC to 80 MHz
Pulse Output (FCA3100 Series)Programmable through front GPIB/USB
ModePulse Out, Gate Open, Alarm Out
Period20 ns - 2 s, in 10 ns increments
Pulse width10 ns - 2 s, in 10 ns increments
OutputTTL levels in 50 Ω, rise time 2 ns
Rear-panel Measurement InputsA, B, C (Option RP only)
Impedance1 MΩ / 50 pF or 50 Ω (VSWR ≤ 2:1)
ConnectorsSMA female for rear input C BNC for all other inputs/outputs

Auxiliary Functions

Trigger Holdoff

CharacteristicDescription
Time Delay Range20 ns to 2 s, 10 ns resolution

External Start and Stop Arming

CharacteristicDescription
ModesStart, Stop, Start and Stop Arming
Input ChannelsA, B, or (rear panel) E
Max Rep. Rate for Arming Signal
Channel A, B160 MHz
Channel E80 MHz
Start-time Delay Range20 ns to 2 s, 10 ns resolution

Statistics

CharacteristicDescription
FunctionsMaximum, Minimum, Mean, ΔMax-Min, Standard Deviation, and Allan Deviation
DisplayNumeric, histograms, or trend plots
Sample Size2 to 2 × 109samples
Limit QualifierOff, or capture values above, below, inside, or outside limits
Measurement PacingPacing Time Range: 4 μs to 500 s

Mathematics

CharacteristicDescription
Functions(K*X+L)/M and (K/X+L)/M. X is current reading and K, L, and M are constants; set using the keyboard or as frozen reference value (X0)

Other Functions

CharacteristicDescription
Measuring Time20 ns to 1000 s for frequency, burst, and period average. Single cycle for other measuring functions
Time-base ReferenceInternal, external, or automatic
Display HoldFreezes the result, until a new measurement is initiated through a restart
Limit AlarmGraphical indication on front panel and/or SRQ through GPIB, plus pulse output connector (FCA3100 Series)
Limit ValuesLower limit, upper limit
SettingsOff, or alarm if value is above, below, inside, or outside limits
On AlarmStop or Continue
DisplayNumeric + Graphic
Stored Instrument Setups20. Instrument setups can be saved/recalled from internal nonvolatile memory. 10 can be user protected
DisplayBacklit LCD graphics screen for menu control, numerical readout, and status information
Number of digits14 digits in Numerical mode
Resolution320 × 97 pixels

GPIB Interface

CharacteristicDescription
CompatibilityIEEE 488.2-1987, SCPI 199953131A Compatibility mode
Interface FunctionsSH1, AH1, T6, L4, SR1, RL1, DC1, DT1, E2
Max Measurement Rate
GPIB15k/5k readings/s (Block mode)4k/- readings/s (Talker Only mode)650/500 readings/s (individual GET triggered)
To internal memory250k readings/s
Internal Memory Size750k readings (FCA3000 Series)3.75M readings (FCA3100 Series)

USB Interface
Calibration

CharacteristicDescription
USB Version2.0 full speed (11 Mb/s)
CharacteristicDescription
ModeClosed case, menu controlled
Calibration Frequencies0.1, 1, 5, 10, 1.544, and 2.048 MHz

General Specifications

Environmental Data

CharacteristicDescription
ClassMIL-PRF-28800F, Class 3
Operating Temp0 °C to +50 °C
Storage Temp-40 °C to +71 °C
Humidity5-95% (10-30 °C) 5-75% (30-40 °C) 5-45% (40-50 °C)
AltitudeOperating: 2,000 m Storage: 12,000 m
SafetyDirective 2006/95/EC, EN61010-1, UL61010-1, CAN/CSA C22.2 No. 61010-1
EMCEU Directive 2004/108/EC, EN61326-1, EN61326-2-1, Class A

Power Requirements

CharacteristicDescription
Basic Version90 to 265 VRMS, 45 to 440 Hz, <40 W

Time-base Options

CharacteristicStandardMedium Stability (MS)High Stability (HS)
Time-base TypeTCXOOCXOOCXO
Uncertainty Due to -
Aging
Per 24hNA<5×10-9*1<5×10-10*1
Per month<5×10-7<6×10-8<1×10-8
Per year<5×10-6<2×10-7<5×10-8
Temperature variation (typ. values)
0-50 °C<1×10-5<5×10-8<5×10-9
20-26 °C<3×10-6<2×10-8<1×10-9
Short-term Stability: t = 1 sNot specified<1×10-10<1×10-11
Root Allan Variance: t = 10 sNot specified<1×10-10<1×10-11
Power-on StabilityNA<1×10-7<1×10-8
Deviation versus final value after 24h ON time, after a warm-up time of:30 min30 min10 min
Total Uncertainty, for Operating Temperature 20 °C to 26 °C, at 2σ (95%)
Confidence Interval
1 year after calibration<7×10-6<2.4×10-7<0.6×10-7
2 years after calibration<1.2×10-5<4.6×10-7<1.2×10-7

*1 After 1 month of continuous operation.

Physical

Dimensionmmin.
Height903.6
Width2108.25
Depth39515.6
Weightkglb.
Net2.75.8
Shipping3.57.5

Ordering Information

Models

ModelDescription
FCA3000Timer/Counter/Analyzer 300 MHz / 100 ps
FCA3003Timer/Counter/Analyzer 3 GHz / 100 ps
FCA3020Timer/Counter/Analyzer 20 GHz / 100 ps
FCA3100Timer/Counter/Analyzer 300 MHz / 50 ps
FCA3103Timer/Counter/Analyzer 3 GHz / 50 ps
FCA3120Timer/Counter/Analyzer 20 GHz / 50 ps

FCA3000/3100 Series Include: Timer/Counter, line cord, calibration certificate, Quick Start User Manual, CD-ROM with user manual (English, French, German, Spanish, Simplified Chinese, Traditional Chinese, Korean, Russian, Japanese), Programmer's Guide, Technical Specifications, Trial version of TimeView™ Software, and CD-ROM with National Instruments LabVIEW SignalExpress™ Tektronix Edition, Limited Edition Software.

Please specify power plug when ordering.

Instrument Options

OptionDescription
MSMedium-stability Oven Time Base
HSHigh-stability Oven Time Base
RPRear-panel Connectors

Power Plug Options

OptionDescription
A0North America
A1Universal Euro
A2United Kingdom
A3Australia
A5Switzerland
A6Japan
A10China
A11India
E1UK and Euro (FCA3000 and FCA3100 only)

Service Options

OptionDescription
CA1Provides a single calibration event or coverage for the designated calibration interval, whichever comes first
C3Calibration Service 3 Years
C5Calibration Service 5 Years
D1Calibration Data Report
R5Repair Service 5 Years

Recommended Accessories and Software

AccessoryDescription
HCTEK4321Hard Carrying Case
AC4000Soft Carrying Case
174-4401-xxUSB Host to Device Cable, 3 ft.
012-0991-xxGPIO Cable, Double Shielded
012-1256-xxBNC Male to BNC Male, Cable Shielded, 9 ft., 50 Ω
012-0482-xxBNC Male to BNC Male, Cable Shielded, 3 ft., 50 Ω
SIGEXPTENational Instruments SignalExpress™ Tektronix Edition Interactive Measurement Software – Professional Version
TVA3000TimeView™ Modulation Domain Analysis Software

TEKTRONIX FCA3000 - Ordering Information - 1

TEKTRONIX FCA3000 - Ordering Information - 2

ISO 9001

THE OUTCOME ANALYSIS FOR OVERTAINING

Product(s) are manufactured in ISO registered facilities.

TEKTRONIX FCA3000 - Ordering Information - 3

Product(s) complies with IEEE Standard 488.1-1987, RS-232-C, and with Tektronix

Standard Codes and Formats.

Contact Tektronix:

ASEAN/Australasia (65) 6356 3900

Austria 00800 2255 4835*

Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777

Belgium 00800 2255 4835*

Brazil +55 (11) 3759 7600

Canada 1800 833 9200

Central East Europe, Ukraine and the Baltics +41 52 675 3777

Central Europe & Greece +41 52 675 3777

Denmark +45 80 88 1401

Finland +41 52 675 3777

France 00800 2255 4835*

Germany 00800 2255 4835*

Hong Kong 400 820 5835

India 000 800 650 1835

Italy 00800 2255 4835*

Japan 81 (3) 6714 3010

Luxembourg +41 52 675 3777

Mexico, Central/South America & Caribbean (52) 56 04 50 90

Middle East, Asia and North Africa +41 52 675 3777

The Netherlands 00800 2255 4835*

Norway 800 16098

People's Republic of China 400 820 5835

Poland +41 52 675 3777

Portugal 80 08 12370

Republic of Korea 001 800 8255 2835

Russia & CIS +7 (495) 7484900

South Africa +41 52 675 3777

Spain 00800 2255 4835*

Sweden 00800 2255 4835*

Switzerland 00800 2255 4835*

Taiwan 886 (2) 2722 9622

United Kingdom & Ireland 00800 2255 4835*

USA 1800 833 9200

  • European toll-free number. If not accessible, call: +41 52 675 3777

Contact List Updated 25 May 2010

For Further Information

Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visit www.tektronix.com

TEKTRONIX FCA3000 - For Further Information - 1

Copyright © 2010, Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks or registered trademarks of their respective companies.