DVM1500 - Multimeter VELLEMAN - Free user manual and instructions
Find the device manual for free DVM1500 VELLEMAN in PDF.
| Product Type | Digital Multimeter |
| Brand | Velleman |
| Model | DVM1500 |
| Dimensions (approx.) | 150 x 70 x 30 mm |
| Weight (approx.) | 300 g |
| Power Supply | 9V battery (type 6LR61) |
| Display | Digital LCD, 3 ½ digits |
| Measurement Functions | AC/DC voltage, AC/DC current, resistance, continuity, diode test |
| DC Voltage Range | 200 mV to 600 V |
| AC Voltage Range | 200 V to 600 V |
| DC Current Range | 200 µA to 10 A |
| Resistance Range | 200 Ω to 2 MΩ |
| Safety | Category CAT II 600 V, double insulation |
| Maintenance and Cleaning | Wipe with a dry cloth. Do not use solvents. |
| Spare Parts and Repairability | Test leads, battery, replacement fuse available |
| Included Accessories | Test leads, battery, user manual |
| Warranty | 2 years (according to Velleman conditions) |
Frequently Asked Questions - DVM1500 VELLEMAN
User questions about DVM1500 VELLEMAN
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Multimeter in PDF format for free! Find your manual DVM1500 - VELLEMAN and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. DVM1500 by VELLEMAN.
USER MANUAL DVM1500 VELLEMAN
DIGITAL MULTIMETER + NO CONTACT AC VOLTAGE DETECTOR
DIGITALE MULTIMETER + CONTACT'LOZE AC SPANNINGSDETECTOR
MULTIMETRE NUMÉRIQUE + DETECTEUR DE TENSION CA SANS CONTACT
MULTIMETRO DIGITAL + DETECTOR DE TENSION CA SIN CONTACTO
natural_image
Digital multimeter displaying a 0.000 millimeter with red and black probes, alongside a coiled wire (no visible text or symbols on main components)USER MANUAL 3
GEBRUIKERSHANDLEIDING 16
NOTICE D'EMPLOI 29
MANUAL DEL USUARIO 42
To all residents of the European Union
Important environmental information about this product


This symbol on the device or the package indicates that disposal of the device after its lifecycle could harm the environment. Do not dispose of the unit (or batteries) as unsorted municipal waste; it should be taken to a specialized company for recycling. This device should be returned to your distributor or to a local recycling service. Respect the local environmental rules.
If in doubt, contact your local waste disposal authorities.
Thank you for choosing Velleman! Please read the manual thoroughly before bringing this device into service. If the device was damaged in transit, don't install or use i: and contact your dealer.
Refer to the Velleman® Service and Quality Warranty on the last pages of this manual.
2. Us ed symbols












This symbol indicates: Read instructions
Not reading the instructions and manual can lead to damage, injury or death.
This symbol indicates: Danger
A hazardous condition or action that may result in injury or death
This symbol indicates: Risk of danger/damage
Risk of a hazardous condition or action that may result in damage, injury or death
This symbol indicates: Attention; important information ignoring this information can lead to hazardous situations.
AC (Alternating Current)
DC (Direct Current)
Both AC and DC
Double insulation (class II-protection)
Earth
Fuse
Capacitor
Diode
Continuity
| DVM1500 | Rev. 01 | |
| Backlight | ||
- Sa fety instructions
![]() | Follow the instructions below to guarantee a safe use of the meter and all its functionalities. |
![]() | During use of themeter, respect all directives concerning protection againstelectroshocks and misuse. Never exceed the indicated limits. |
![]() | WARNING:For safety reasons, please read the manual.Remark: this warning is also found on the back of the meter (top part). |
![]() | WARNING: To avoid electrical shock always disconnect the test leads prior to opening the housing. To prevent damage or injury, only use batteries and fuses with the same ratings as specified in this manual.Remark: refer to the warning on the back of the meter. |
![]() | |
| [IMAGE] | A void cold, heat and large temperature fluctuations. When the unit is moved from a cold to a warm location, leave it switched off until it has reached room temperature. This to avoid condensation and measuring errors. |
![]() | |
| [IMAGE] | Protect this device from shocks and abuse. Avoid brute force when operating. |
![]() | |
| [IMAGE] | Pollution degree 2-device. For indoor use only. Keep this device away from rain, moisture, splashing and dripping liquids. Not for industrial use.Refer to §5 Pollution degree. |
![]() | Keep the device away from children and unauthorised users. |
![]() | Risk of electric shock during operation. Be very careful when measuring live circuits. Use extreme caution when measuring voltages higher than 60Vdc or 30Vac rms. |
| Do not replace internal parts yourself. Replace damaged or lost accessories by identical ones with the same specifications. Order spare accessories e.g. test probes at your dealer. | |
![]() | This is an installation category QAT III 600V / QAT II 1000V measuring instrument. Never use this equipment in a higher category than indicated. Refer to §4 Overvoltage /installation category. |
| DVM1500 Rev. 01 | |
![]() | Read this manual thoroughly. Familia rise yourself with the functions of the device before actually using it. |
![]() | All modifications of the device are forbidden for safety reasons. Damage caused by user modifications to the device is not covered by the warranty. |
![]() | Only use the device for its intended purpose. Using the device in an unauthorized way will void the warranty. Damage caused by disregard of certain guidelines in this manual is not covered by the warranty and the dealer will not accept responsibility for any ensuing defects or problems. |
![]() | Make sure the meter is in the appropriate measuring range before connecting it to a test circuit. |
![]() | Do not measure circuits that may contain voltages >600V |
![]() | Do not measure current in circuits with voltages >250V |
![]() | Do not conduct resistance, diode- or continuity measurements on live circuits. |
![]() | Before each use, make sure the test probes are in good condition. Always place your fingers behind the protective edges of the test probes while measuring! Never touch free terminals when the meter is connected to a circuit. |
![]() | Switch off the meter and remove test probes prior to replacing the battery or fuses. |
![]() | When carrying out measurements on a TV set or switching power circuits, always be aware that high amplitude voltage pulses at the test points might damage the meter. |
4. Ov ervoltage/installation category
DMMs are categorized depending on the risk and severity of transient overvoltage that might occur at the point of test. Transients are: short-lived bursts of energy induced in a system, e.g. caused by lightning strike on a power line.
The existing categories according EN 61010-1 are:
| CAT I | A CAT I-rated meter is suitable for measurements on protected electronic circuits that are not directly connected to mains power, e.g. electronics circuits, control signals.. | |
| CAT II | A CAT II-rated meter is suitable for measurements in CAT I-environments and mono-phase appliances that are connected to the mains by means of a plug and circuits in a normal domestic environment, provided that the circuit is at least 10m apart from a CAT III- or 20m apart from a CAT IV-environment. E.g. household appliances, portable tools... | |
| DVM1500 | Rev. 01 | |
| CAT III | A CAT III-rated meter is suitable for measurements in CAT I- and CAT II-environments, as well as for measurements on (fixed) mono- or poly-phased appliances which are at least 10m apart from of a CAT IV-environment, and for measurements in or on distribution level equipment (low-voltage distribution boards, lighting circuits, electric ovens). | |
| CAT IV | A CAT IV-rated meter is suitable for measuring in CAT I-, CAT II- and CAT III-environments as well as on the primary supply level. Not that for all measurements on equipment for which the supply cables run outdoors (either overhead or underground) a CAT IV meter must be used. | |
Warning:
This device was designed in accordance with EN 61010-1 installation category CAT III 600V/ CAT II 1000V. This implies that certain restrictions in use apply that are related to voltages and voltage peaks which can occur within the environment of use. Refer to the table above.
This device is suitable for measurements up to 1000V:
- Protected electronic circuits that are not directly connected to mains power, e.g. electronics circuits, control signals, circuits behind isolating transformer...
- circuits that are directly connected to mains power, but limited to:
○ measurements on mon) -phase appliances that are connected to the mains by means of a plug
○ mono-phase appliances and circuits directly connected to the mains in a normal domestic environment, provided that the circuit is at least 10m apart from a CAT : II- or 20m apart from a CAT IV-environment. E.g. household appliances, portable tools, light circuits at more than 10m from a distribution board ...
This device is suitable for measurements up to 600V:
- measurements in/on low-voltage distribution boards (distribution boards behind meter box)
- measurements on (fixed) mono- or poly-phased appliances and circuits except in CAT IV-environments (e.g. mains outlets, electric ovens, lighting circuits, bus bars, low-voltage distribution boards and d circuit breakers).
This device is NOT suitable for:
• Voltage s above 1000V
- Measure ments on distribution equipment and outdoor installations including meter boxes and equipment/circuits outside or remote from the domestic environment e.g. circuits in sheds, garden houses and free-standing garages, or circuits using underground wiring e.g. garden lighting, pool-pump...

This device is only suitable for measurements up to 600V in CAT III and up to 1000V in CAT II.
5. Pol lution degree
IEC 61010-1 specifies different types of pollution environments, for which different protective measures are necessary to ensure safety. Harsher environments require more protection, and the protection against the pollution which is to be found in a certain environment depends mainly on the insulation and the enclosure properties. The pollution degree rating of the DVM indicates in which environment the device may be used.
| Pollution degree 1 | No pollution or only dry, nonconductive pollution occurs. The pollution has no influence. (only to be found in hermetically sealed enclosures) |
| Pollution degree 2 | Only nonconductive pollution occurs. Occasionally, temporary conductivity caused by condensation is to be expected.(home and office environments fall under this category) |
| Pollution degree 3 | Conductive pollution occurs, or dry nonconductive pollution occurs that becomes conductive due to condensation that is to be expected.(industrial environments and environments exposed to outside air - but not in contact with precipitation) |
| Pollution degree 4 | The pollution generates persistent conductivity caused by conductive dust or by rain or snow. (exposed outdoor environments and environments where high humidity levels or high concentrations of fine particles occur) |
Warning:
This device was designed in accordance with E N 61010-1 pollution degree 2. This implies that certain restrictions in use apply that are related to pollution which can occur within the: environment of use. Refer to the table above.

This device is only suitable for measurements in Pollution degree class 2 environments.
6. De scription
Refer to the illustrations on page 2 of this manual.
| ON/OFF | Switch the meter on or off using this key. | |
| FUNC. | Selection of the DC (default) or AC modes, and the ➕ (default) or ➞ modes. This key function is only available in the A and ➞ ( ➞) ranges. | |
| RANGE | Selection of the automatic (default) or manual modes, selection of manual or auto-ranging modes, selection of ranges in manual mode. This key function is only available in the V and Ω ranges. | |
| HOLD | Fixes the current value onto the display. Press again to return to normal mode. | |
| REL | Selection of the relative measurement mode. This key function is only available in the V, A, Ω, °C and CAP ranges. | |
| DVM1500 | Rev. 01 | |
| [7c56] | Switch the backlight on or off using this key. This key function is not available in the EF- ran ge. | |
| Rotary Switch | Use the rotary switch to select the function and the desired ranges. | |
| Terminals | VΩ +°CHz: Terminal receiving the red test lead for voltage, resistance, capacitance, diode test, temperature and frequency measurements.COM: Terminal receiving the black test lead as common reference.μAmA: Terminal receiving the red test lead for μA and mA measurements.10A: Terminal receiving the red test lead for 10A measurements. | |
7. Operating instructions


Risk of electric shock during operation. Be very careful when measuring live circuits.
To avoid damage, the meter features an alarm function. Check the connections, the range and the terminals as soon as the meter buzzes. However, this is only an indication and does not provide protection against damage or misuse!
- Never exceed the limit value for protection. This limit value is listed separately in the specifications for each range of measurement.
- Do not touch unused terminals when the meter is linked to a circuit which is being tested.
- Only use the meter in the indicated overvoltage/installation category. Never measure voltages that might exceed the indicated category values.
- Disconnect the test leads from the tested circuit before rotating the range selector in order to change functions..
- When carrying out measurements on a TV set or switching power circuits, always remember that high amplitude voltage pulses at the test points might damage the meter.
- Always be careful when working with voltages above 60Vdc or 30Vac rms. Keep your fingers behind the probe barriers at all times during measurement.
- Do not measure current in circuits with voltages > 250V
- Never perform resistance, diode or continuity measurements on live circuits. Make sure all capacitors in the circuit are depleted.
Non-contact AC voltage detector:


Remove the test probes when performing non-contact voltage measurements
This function only provides a quick indication whether AC voltages are present; it is not a fully reliable way to determine whether a circuit is safe to touch without risk of electrical shocks.
DVM1500
Rev. 01
- Set the rotary switch to the EF-range. T he green LED will switch on.
- Place the top of the multi-meter as close as possible to the lead or the mains outlet. The green LED will turn red and the multi-meter will beep an AC voltage is present.
Note: this function is only an indication whether an AC voltage is present, and does not guarantee that a circuit is voltage-free. It does not detect DC-voltages. When in doubt, perform a full voltage measurement.
Voltage measurements

Do not measure circuits where voltages > 600V CAT III or > 1000V CAT II may reside.

Always be careful when working with voltages above 60Vdc or 30Vac rms. Keep your fingers behind the probe barriers at all times during measurement. Do not touch unused terminals when the meter is linked to a circuit which is being tested.
- Connect the red test lead to the "VΩ +°C Hz" jack and the black lead to the "COM" jack.
- Set the rotary switch to V = for DC measurements or to V for AC measurements.
- Press the RANGE button to manually select the range.
- Connect the test probes to the source being measured.
- The measured value appears on the LCD display.
Note: for DC-measurements, when a negative polarity is present at the red test lead, the indicated value is preceded by a “-” sign.
Resistance measurements

Do not perform resistance measurements on live circuits.
- Connect the red test lead to the "VΩ -1°C Hz" jack and the black lead to the "COM" jack.
- Set the rotary switch to "Ω"
- Press the RANGE button to manually select the range.
- Connect the test probes to the source being measured.
- The measured value appears on the LCD display.
- Notes:
- Never perform resistance measurements on a live circuit and make sure all capacitors are completely depleted.
To increase accuracy when measuring low resistance values, first hold the tips of the measuring probes together to determine the resistance value of the test leads. Subtract this value from the measured value of the circuit.
- For resistance measurements above 1MΩ the meter needs a few seconds to stabilize the read-out.
- Should the measured resistance exceed the current range or in case of an open circuit, the display will show "OL".
Continuity measurements

Do not perform continuity measurements on live circuits.
- Connec i the red test lead to the "VΩ +°C Hz" jack and the black lead to the "COM" jack.
- Set the rotary switch to ✦.
- Press the FUNC button to select continuity test
- Connec t the test probes to the source being measured.
- When the measured resistance is less than 40 a continuous beep is produced and the resistance value is showed on the display. 'Should the measured resistance exceed the current range or in case of an open circuit, the display will show "OL".
Nota: Never perform continuity measurements on a live circuit and make sure all capacitors are completely depleted.
Diode measurements

Do not perform diode measurements on live circuits.
- Connect the red test lead to the "VΩ 10°C Hz" jack and the black lead to the "COM" jack.
- Set the rotary switch to ✦.
- Press the FUNC button to select diode test
- Connect the red test probe to the anode of the diode and the black test probe to the cathode.
- The meter shows the forward voltage drop of the diode. When the diode is connected with a reversed polarity, the display shows "OL".
Notes:
- Never perform diode measurements on a live circuit and make sure all capacitors are completely depleted.
○ Measuring diodes that are part of a circuit might produce faulty results. Consider unmounting from the circuit.
Capacitance measurements

Do not perform capacitance measurements on live circuits.
- Connect the red test lead to the "VΩ +°C Hz" jack and the black lead to the "COM" jack.
- Set the rotary switch to ± .
- Press the RANGE button to manually select the range.
- Connect the test probes to the capacitor and read its value from the display.
Notes:
- It takes a few seconds before the meter shows the result. This is normal behavior.
To increase accuracy when measuring low capacitance values (<4nF), first hold the tips of the measuring probes together to determine the capacitance value of the test leads. Subtract this value from the measured value of the circuit.
DVM1500
Rev. 01
- Never perform capacitance measurements on a live circuit and make sure all capacitors are completely depleted.
Frequency measurements

Do not perform frequency measurements in circuits with voltages > 250V.

Always be careful when working with voltages above 60Vdc or 30Vac rms. Keep your fingers behind the probe barriers at all times during measurement. Do not touch unused terminals when the meter is linked to a circuit which is being tested.
- Connect the red test lead to the "VΩ +°C Hz" jack and the black lead to the "COM" jack.
- Set the rotary switch to Hz.
- Press the RANGE button to manually select the range.
- Connect the test probes to the circuit and read the measured value from the LCD.
Notes:
- Accuracy of the reading is not guaranteed for amplitudes > 3Vrms.
- Use a shielded cable for measuring small signals in a noisy environment.
Temperature measurements

Do not touch any live parts with the temperature measuring probe.
- Set the rotary switch to the °C range. When no temperature measuring probe is connected, the current environmental temperature is displayed.
- Connect the black plug of the temperature probe to the COM jack and the red plug to the "VΩ ➕°CHz" jack (thermo couple type k).
- Hold the tip of the temperature probe against the object under test.
- Read the temperature on the LCD.
Current Measurements

Do not measure current in circuits with voltages > 250V

Current measuring jack AmA max. 400mA. For currents up to 10A use the 10A jack.

Always be careful when working with voltages above 60Vdc or 30Vac rms. Keep your fingers behind the probe barriers at all times during measurement.
- For measurements up to 400mA: connect the red test lead to the "μAmA" jack and the black lead to the "COM" jack.
- For measurements up to 10A: connect the red test lead to the "10A" jack and the black lead to the "COM" jack.
- Set the rotary switch to the 400 A range for measurements up to 400 A (only when test probe is connected to the AmA jack).
- Set the rotary switch to the 400mA range for measurements up to 400mA (only when test probe is connected to the μAmA jack).
DVM1500
Rev. 01
- Set the rotary switch to the 10A range for measurements up to 10A (only when test probe is connected to the 10A jack).
- Press the FUNC button to select DC current (DCA) or AC current (ACA) mode.
- Connect the test probes in series with the circuit.
- Read the measured value from the display.
Notes:
- for DC-current measurements, when a negative polarity is present at the red test lead, the indicated value is preceded by a “-” sign.
The μAmA-range is protected agains overcurrent with a F5 00mA 250V fuse, the 10A range is NOT PROTECTED!
8. Ma intenance

Do not replace internal parts yourself. Replace damaged or lost accessories by identical ones with the same specifications. Order spare accessories e.g. test probes at your dealer.

Disconnect the test leads from the test points and remove the test leads from the measuring jacks before replacing the batteries or fuses.
a. General maintenance
Wipe the device regularly with a moist, lint-free cloth. Do not use alcohol or solvents.
b. Fuse Replacement
- Switch off the multi-meter.
- Remove the 3 screws at the back and gently open the meter.
- Remove the fuse from the fuse holder and replace it with a new fuse of the same type and with the same specifications (F500mA/250V, ∅ 5 x 20mm).
- Close the meter carefully.
c. Battery Replacement

To avoid false readings, which could lead to possible electric shocks or personal injury, replace the battery as soon as the battery indicator appears.
- Switch off the multi-meter.
- Open the battery compartment using an appropriate screwdriver.
- Remove the batteries.
- Install 3 replacement batteries (3 x 1.5V A ∨ A batteries). Do not use rechargeable batteries.
- Close the battery compartment.
9. Technische specificaties
This device is not calibrated when purchased! Regulations concerning environment of use:
- Use t his meter only for measurements in CAT I and CAT II environments (see §4)
- Use t his meter only in a pollution degree 2 environment (see §5) Ideal working conditions include:
DVM1500
Rev. 01
• Temperature: 18 to 28°C (64°F to 82°F)
• relative humidity: max. 75%
• altitude: max. 2000m (6560ft)
| Overvoltage category | 1,000V CAT. II and 600V CAT. III |
| Pollution Degree | 2 |
| Operating Temperature | 0°C (32°F) ~ 40°C(122°F) (< 80% RH, < 10°C) |
| Storage Temperature | -10°C (14°F) ~ 60°C (140°F) (< 70% RH, remove batteries!) |
| Temperature Coefficient | 0.1x / C° (< 18°C or > 28°C) |
| Max. Voltage between Terminals and Earth | 750VAC RMS or 1,000VDC |
| Fuse Protection | μA and mA, F500mA / 250V, 5 x 20mm (10A range not protected) |
| Sample Rate | 3x/sec |
| Display | 3^3/4 digits LCD with automatic indication of functions and symbols |
| Over Range Indication | yes ("OL") |
| Low Battery Indication | yes ( ) |
| Polarity Indication | "- " displayed automatically |
| Data Hold | yes |
| Backlight | yes |
| Auto Power-Off | yes, after 15min. Buzzer sounds after 14min. as warning |
| Power Supply | 3 x 1.5V AAA batteries |
| Dimensions | 156 x 82 x 29mm |
| Weight | ± 220g (with batteries) |
| Accessories | manual, test probes, batteries, K-type thermocouple |
Non-contact AC voltage detector:
| Sensitivity | Frequency | Distance |
| > 50V | 50Hz | < 150 mm |
DC voltage
| Range | Resolution | Accuracy |
| 400mV | 0.1mV | ± (0.8% of rdg + 3 digits) |
| 4V | 1mV | |
| 40V | 10mV | |
| 400V | 100mV | |
| 100()V | 1V | ± (1.0% of rdg + 3 digits) |
Input impedance: 10MΩ
Max. input voltage: 1000V ---
AC voltage
| Range | Resolution | Accuracy |
| 4V | 1mV | ± (1.0% of rdg + 3 digits) |
| 40V | 10mV | |
| 400V | 100mV | |
| 750V | 1V | ± (1.2% of rdg + 3 digits) |
16.04.2010
DVM1500
Rev. 01
Input impedance: 10MΩ
Max. input voltage: 750V\~rms
Frequency range: 40Hz \~ 400Hz
DC current
| Range | Resolution | Accuracy |
| 4,000μA | 1μA | ± (1.2% of rdg + 3 digits) |
| 400μA | 0.1mA | |
| 10A | 10mA | ± (2.0% of rdg + 8 digits) |
Overload protection: F500mA/250V fuse for A and mA ranges; 10A range not protected
Max. input current: 400mA for μA and mA ranges, 10A for 10A range
Max. Circuit voltage for all ranges: 250V
AC current
| Range | Resolution | Accuracy |
| 4,000μA | 1μA | ± (1.5% of rdg + 5 digits) |
| 400μA | 0.1mA | |
| 10A | 10mA | ± (3.0% of rdg + 8 digits) |
Overload protection: F500mA/250V fuse for A and mA ranges; 10A range not protected
Max. input current: 400mA for μA and mA ranges, 10A for 10A range
Max. Circuit voltage for all ranges: 250V
Frequency range: 40Hz \~ 400Hz
Resistance
| Range | Resolution | Accuracy |
| 400Ω | 0.1Ω | ± (1.2% of rdg + 3 digits) |
| 4kΩ | 1Ω | |
| 40kΩ | 10Ω | |
| 400kΩ | 100Ω | |
| 4MΩ | 1kΩ | |
| 40MΩ | 10kΩ | ± (2.0% of rdg + 5 digits) |
Open circuit voltage: ± 0.25 V
Diode and audible continuity test
| Range | Description | Test Condition |
| Built-in buzzersounds if resistance < ± 40Ω | Open circuit voltage: ± 0.5V— | |
| Display reads approx. forward voltage of diode | Forward DC current ± 1mA Reversed DC voltage: ± 1.5V— |
Temperature
| Range | Resolution | Accuracy |
| -20°C ~ 0°C | 1°C | ± (5.0% of rdg + 4 digits) |
| 0°C ~ < 00°C | 1°C | ± (1.0% of rdg + 3 digits) |
| 400°C ~ 1,000°C | 1°C | ± (2.0% of rdg + 3 digits) |
DVM1500 Rev. 01
Capaciteitsmeting
| Range | Resolution | Accuracy |
| 4nf | 1pF | ± (5.0% of rdg + 5 digits) |
| 40nF | 10pF | ± (3.0% of rdg + 5 digits) |
| 400nF | 100pF | |
| 4μF | 1nF | |
| 40μF | 10nF | |
| 200μF | 100nF |
Frequency measurements
| Range | Resolution | Accuracy |
| 9.999Hz | 0.001Hz | ± (0.1% of rdg + 1 digit) |
| 99.99Hz | 0.01Hz | |
| 999.9Hz | 0.1Hz | |
| 9.999kHz | 1Hz | |
| 99.99kHz | 10Hz | |
| 199.99kHz | 100Hz | |
| >200kHz 100Hz Uns | pecified @ >200kHz | |
Amplitude vs. frequency measurement. (Input voltage): 0.6Vrms to max.250Vrms
For accurate measurements: amplitude between 1Vrms and 3Vrms
Use this device with original accessories only. Velleman nv cannot be held responsible in the event of damage or injury resulted from (incorrect) use of this device. For more info concerning this product and the latest version of this user manual, please visit our website www.velleman.eu. The information in this manual is subject to change without prior notice.
© COPYRIGHT NOTICE
This manual is copyrighted. The copyright to this manual is owned by Velleman nv. All worldwide rights reserved. No part of this manual may be copied, reproduced, translated or reduced to any electronic medium or otherwise without the prior written consent of the copyright holder.
GEBRUIKERSHANDLEIDING
1. Inl eiding
notice, thermocouple type K
Velleman® Service and Quality Warranty
Velleman® has over 35 years of experience in the electronics world and distributes its products in more than 85 countries.
All our products fulfil strict quality requirements and legal stipulations in the EU. In order to ensure the quality, our products regularly go through an extra quality check, both by an internal quality department and by specialized external organisations. If, all precautionary measures notwithstanding, problems should occur, please make appeal to our warranty (see guarantee conditions).
General Warranty Conditions Concerning Consumer Products (for EU):
- All consumer products are subject to a 24-month warranty on production flaws and defective material as from the original date of purchase.
- Velleman® can decide to replace an article with an equivalent article, or to refund the retail value totally or partially when the complaint is valid and a free repair or replacement of the article is impossible, or if the expenses are out of proportion.
You will be delivered a replacing article or a refund at the value of 100% of the purchase price in case of a flaw occurred in the first year after the date of purchase and delivery, or a replacing article at 50% of the purchase price or a refund at the value of 50% of the retail value in case of a flaw occurred in the second year after the date of purchase and delivery.
- Not covered by warranty:
- all direct or indirect damage caused after delivery to the article (e.g. by oxidation, shocks, falls, dust, dirt, humidity...), and by the article, as well as its contents (e.g. data loss), compensation for loss of profits;
- frequently replaced consumable goods, parts or accessories such as batteries, lamps, rubber parts, drive belts... (unlimited list);
- flaws resulting from fire, water damage, lightning, accident, natural disaster, etc. ...;
- flaws caused deliberately, negligently or resulting from improper handling, negligent maintenance, abusive use or use contrary to the manufacturer's instructions;
- damage caused by a commercial, professional or collective use of the article (the warranty validity will be reduced to six
(6) months when the article is used professionally);
- damage resulting from an inappropriate packing and shipping of the article;
- all damage caused by modification, repair or alteration performed by a third party without written permission by Velleman®.
- Articles to be repaired must be delivered to your Velleman® dealer, solidly packed (preferably in the original packaging), and be completed with the original receipt of purchase and a clear flaw description.
- Hint: In order to save on cost and time, please reread the manual and check if the flaw is caused by obvious causes prior to presenting the article for repair. Note that returning a non-defective article can also involve handling costs.
• Repairs occurring after warranty expiration are subject to shipping costs. - The above conditions are without prejudice to all commercial warranties.