Mean Well RPS-45-48 - Uncategorized

RPS-45-48 - Uncategorized Mean Well - Free user manual and instructions

Find the device manual for free RPS-45-48 Mean Well in PDF.

📄 4 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice Mean Well RPS-45-48 - page 1
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Medical Switching Power Supply, PCB Mount
Model RPS-45-48
Output Voltage 48 VDC
Output Power 45 W
Input Voltage Range 100-240 VAC (typical for series)
Dimensions (Approx.) 76.2 x 50.8 x 24 mm (3" x 2" x 0.94")
Weight (Approx.) 150 g
Protection Features Overload, Short Circuit, Over Voltage
Cooling Method Free air convection (forced air for >120W models)
Operating Temperature -30 to +70 °C (derating above 50 °C)
Storage Temperature -40 to +85 °C
Safety Standards IEC/EN 60601-1, UL 60601-1 (medical)
EMC Compliance EN 55011 (Class B), EN 60601-1-2
Insulation Distance 5 mm bottom, 15 mm around PCB
Wiring Recommendation AWG 18-10 depending on current (see manual)
Mounting Orientation Standard orientation (refer to derating curve)
Maintenance Keep ventilation unobstructed; ensure proper clearance
Manufacturer Mean Well Enterprises Co., Ltd.
Warranty 3 years (typical)

Frequently Asked Questions - RPS-45-48 Mean Well

What is the input voltage range for the RPS-45-48?
The typical input voltage range for Mean Well RPS series is 100-240 VAC (universal input). Always check the specific model label.
Can I use the RPS-45-48 in a medical device?
Yes, the RPS-45-48 is a medical switching power supply designed to meet IEC/EN 60601-1 safety standards for medical electrical equipment.
What is the output current of the RPS-45-48?
At 48 V output and 45 W, the nominal output current is 0.94 A (45W / 48V).
How should I mount the PCB power supply?
Mount with at least 5 mm insulation distance from the board to the system and 15 mm clearance around the unit. Use a Mylar film between PCB and system if necessary. Ensure proper ventilation.
What wire size should I use for connection?
Refer to the manual: for currents up to 6A use AWG 18; 6-10A AWG 16; 10-16A AWG 14; 16-25A AWG 12; 25-32A AWG 10. Ensure all strands enter the terminal and screws are tightened securely.
Does the RPS-45-48 require forced air cooling?
PCB units under 120W typically use free air convection. However, ensure adequate ventilation and keep fans or vents unobstructed. For higher ambient temperatures, output derating may be required.
What protection features are included?
The power supply includes overload, short circuit, and over voltage protection. All failures should be examined by a qualified technician.
Can I install this power supply in high humidity?
No, avoid installation in places with high moisture or near water to prevent electrical shock hazard.
How do I clean the RPS-45-48?
Before cleaning, disconnect the power supply from all power sources. Use a dry cloth to gently remove dust. Do not use water or solvents.
Where can I find the detailed specifications?
Visit the official Mean Well website at www.meanwell.com and search for RPS-45-48 to download the full datasheet and installation manual.

User questions about RPS-45-48 Mean Well

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Uncategorized in PDF format for free! Find your manual RPS-45-48 - Mean Well and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. RPS-45-48 by Mean Well.

USER MANUAL RPS-45-48 Mean Well

● Type : Medical Switching Power Supply

(1) Enclosed (Families: MSP)
(2) Modular (Families: MP, NMP)
(3) PCB (Families: MPS, MPD, MPT, MQ, RPS, RPD, RPT, RPTG, NEM, PM)
(4) Adaptor (Families: GSM, MES)

- Installation

(1) Enclosed Type

(a) Before any installation or maintenance work, please disconnect your system from the utility. Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient temperature may increase the internal component temperature and will require a de-rating in output current. Please refer to the specification sheets to receive the optimum mounting position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of Enclosed units. Please refer to case drawing on specifications to receive the maximum length of mounting screw.
(e) Recommended wires of Enclosed units are shown as below.

AWG1816141210
Rated Current of Equipment (Amp)6A6-10A10-16A16-25A25-32A32-40A
Cross-section of Lead( mm^2 )0.751.001.52.54
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more strips should not exceed 80% of current rating.

8

Make sure that all strands of each stranded wire enter the terminal connection and the screw terminals are securely fixed to prevent poor contact.

(f) For other information about the products, please refer to www.meanwell.com for details.

(2) Modular Type

(a) Before any installation or maintenance work, please disconnect your system from the utility. Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient temperature may increase the internal component temperature and will require a de-rating in output current. Please refer to the specification sheets to receive the optimum mounting position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of Modular units. Please refer to case drawing on specifications to receive the maximum length of mounting screw.
(e) Recommended wires of Modular units are shown as below.

AWG1816141210
Rated Current of Equipment (Amp)6A6-10A10-16A16-25A25-32A32-40A
Cross-section of Lead( mm^2 )0.751.001.52.54
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more strips should not exceed 80% of current rating.

8

Make sure that all strands of each stranded wire enter the terminal connection and the screw terminals are securely fixed to prevent poor contact.

(f) For other information about the products, please refer to www.meanwell.com for details.

Installation Manual

(3) PCB Type

(a) Before any installation or maintenance work, please disconnect your system from the utility. Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient temperature may increase the internal component temperature and will require a de-rating in output current. Please refer to the specification sheets to receive the optimum mounting position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance must be kept when the adjacent device is a heat source.
(d) Recommended wires of PCB units are shown as below.

AWG1816141210
Rated Current of Equipment (Amp)6A6-10A10-16A16-25A25-32A32-40A
Cross-section of Lead( mm^2 )0.751.001.52.54
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more strips should not exceed 80% of current rating.

8

6

Make sure that all strands of each stranded wire enter the terminal connection and the screw terminals are securely fixed to prevent poor contact.

(e) At least 5mm insulation distance on the bottom of PCB units should be kept and a Mylar film should be added between the units and the system. In addition, keep enough insulation distance of 15mm around the PCB units.
(f) PCB units greater than 120W may require a forced air/fan for cooling. Please refer to specifications to receive a minimum air intensity and air-flow direction.
(g) For other information about the products, please refer to www.meanwell.com for details.

(4) Adapter Type

(a) Before any installation or maintenance work, please disconnect your system from the utility. Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient temperature may increase the internal component temperature and will require a de-rating in output current. Please refer to the specification sheets to receive the optimum mounting position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance must be kept when the adjacent device is a heat source.
(d) Before attaching the DC plug of an adaptor to equipment, please unplug the adaptor from the AC power and verify the unit is within the voltage and current rating on the equipment.
(e) Keep the linkage between the adaptor and its power cord tightly as well as connecting the DC plug to equipment properly.
(f) Protect the power cord of adaptors from being trodden on or being squashed.
(g) An approved power cord of an adaptor should greater or equal to SVT, 3G×18AWG or H03VV-F, 3G×0.75mm2.
(h) If the final equipment is not used for long period of time, disconnect the equipment from adaptors to avoid being damaged by voltage peaks or lightning strike.
(i) For other information about the products, please refer to www.meanwell.com for details.

- User Manual

NMP series:

http://www.meanwell.com/webapp/product/search.aspx?prod=NMP1K2&pdf=Tk1QLUUucGRm&a=4

● Warning / Caution !!

(1) Risk of electrical shock and energy hazard. All failure should be examined by a qualified technician. Please do not remove the case of the power supply by yourself! Please do not install the units in places with high moisture or near the water.
(2) Please do not install the units in places with high ambient temperature or near fire source. The maximum ambient temperature, please refer to their specifications.
(3) Output current and output wattage must not exceed the rated values on specifications.
(4) To avoid risk of electric shock, for class I products, the ground (FG) must be connected to protective earth.
(5) The Enclosed, Modular and PCB units are designed in accordance with EMC regulations and the related test reports are available by request. Since they are belong to component power supplies and will be installed inside system enclosure, when they are integrated into a system, the EMC characteristics of the end system must be re-verified again.

Manufacturer :

MEAN WELL ENTERPRISES Co., LTD.

No.28, Wuquan 3rd Rd., Wugu Dist.,

New Taipei City 24891, Taiwan

Tel: +886-2-2299-6100

Web: www.meanwell.com

Branch Office :

China

MEAN WELL (GUANGZHOU)

ENTERPRISES Co., LTD.

2F, A Building, Yuean Industry Park,

Huangcun, Dongpu Yown, Tianhe

District, Gungzhou, China

Post Code:510660

Tel: +86-20-2887-1200

Web: www.meanwell.com.cn

U.S.A.

MEAN WELL USA, INC.

44030 Fremont Blvd., Fremont,

CA 94538, U.S.A.

Tel: +1-510-683-8886

Web: www.meanwellusa.com

Europe

MEAN WELL EUROPE B.V.

Langs de Werf 8, 1185XT

Amstelveen, The Netherlands

Tel: +31-20-758-6000

Web: www.meanwell.eu

Declaration of China RoHS Conformity

In order to reduce the impacts on the environment and take the more responsibility for protecting the earth, MEAN WELL is confirming and announcing the conformity to China RoHS, an Administrative Measures for the Restriction of the Use of Hazardous Substances in Electrical and Electronic Products.

Environment Friendly Use Period Label

Mean Well RPS-45-48 - Environment Friendly Use Period Label - 1Observing SJT 11364-2014, Marking for the Restricted Use of Hazardous Substances in Electronic and Electrical Products
Observing SJ/Z 11388-2009, General Guidelines of Environment-friendly Use Period of Electronic Information Products Appendix B, adopting table look-up to verify the Environment Friendly Use Period

Names and Contents of Hazardous Substances Lists

Part NameHazardous Substances
Lead(Pb)Mercury(Hg)Cadmium(Cd)Hexavalent chromium( Cr^6+ )Polybrominated biphenyls(PBB)Polybrominated diphenyl ethers(PBDE)
PCB and its componentsXOXOOO
Metal structure partsXOOOOO
Plastic structure partsOOOOOO
AccessoriesOOOOOO
CablesXOOOOO
O: The concentration of the hazardous substances within the homogeneous material of that product is less than the concentration limits set by GB/T 26572-2011.X: The concentration of the hazardous substances within the homogeneous material of that product is over the concentration limits set by GB/T 26572-2011; however, it follows the standard advised by 2011/65/EU.
Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Mean Well

Model : RPS-45-48

Category : Uncategorized