Product Type Medical Switching Power Supply (PCB Type)
Model RPT-60D
Brand Mean Well
Series RPT-60
Output Power 60W
Input Voltage Range 85-264VAC
Efficiency Typ. >85%
Protection Features Short circuit, overload, overvoltage
Cooling Method Free air convection
Operating Temperature -20°C to +70°C (derating)
Safety Standards UL 60601-1, EN 60601-1
EMC Standards EN55011, EN55022
Installation Type PCB mount
Maintenance No user-serviceable parts; disconnect power before servicing
Warranty Contact manufacturer

Frequently Asked Questions - RPT-60D Mean Well

What is the RPT-60D?
The RPT-60D is a medical switching power supply from Mean Well, designed for PCB mounting and delivering 60W output.
What input voltage does the RPT-60D accept?
It accepts a wide input voltage range of 85-264VAC, suitable for global use.
How should I install the RPT-60D?
Installation must follow the PCB type guidelines: disconnect power before work, ensure proper insulation distance (at least 5mm on bottom, 15mm around), and use recommended wire gauges (AWG 18-10 depending on current).
What wire gauge should I use for the RPT-60D?
Recommended wire gauges vary by current: for up to 6A use AWG18, 6-10A use AWG16, 10-16A use AWG14, 16-25A use AWG12, 25-32A use AWG10. For 5 or more strips, use 80% of rating.
Does the RPT-60D require forced cooling?
No, the RPT-60D is designed for free air convection cooling. However, if ambient temperature is high or output exceeds 120W (this unit is 60W), forced air may be needed.
What safety certifications does the RPT-60D have?
It complies with medical safety standards UL 60601-1 and EN 60601-1, ensuring safe operation in medical equipment.
Can I mount the RPT-60D in any orientation?
Mounting orientations other than standard may require output current derating. Refer to the specification sheet for optimum mounting position and derating curve.
How do I clean the RPT-60D?
Cleaning is not required. Keep ventilation holes free from obstructions and avoid moisture. Do not open the unit.
What should I do if the RPT-60D fails?
Risk of electrical shock! All failures should be examined by a qualified technician. Do not remove the case yourself.
Where can I find the full specifications for the RPT-60D?
Visit www.meanwell.com for detailed product specifications, including output voltage options, dimensions, and derating curves.

User questions about RPT-60D Mean Well

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Uncategorized in PDF format for free! Find your manual RPT-60D - Mean Well and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. RPT-60D by Mean Well.

USER MANUAL RPT-60D Mean Well

● Type : Medical Switching Power Supply

(1) Enclosed (Families: MSP)
(2) Modular (Families: MP, NMP)
(3) PCB (Families: MPS, MPD, MPT, MQ, RPS, RPD, RPT, RPTG, NEM, PM)
(4) Adaptor (Families: GSM, MES)

- Installation

(1) Enclosed Type

(a) Before any installation or maintenance work, please disconnect your system from the utility. Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient temperature may increase the internal component temperature and will require a de-rating in output current. Please refer to the specification sheets to receive the optimum mounting position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of Enclosed units. Please refer to case drawing on specifications to receive the maximum length of mounting screw.
(e) Recommended wires of Enclosed units are shown as below.

AWG1816141210
Rated Current of Equipment (Amp)6A6-10A10-16A16-25A25-32A32-40A
Cross-section of Lead( mm^2 )0.751.001.52.54
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more strips should not exceed 80% of current rating.

8

Make sure that all strands of each stranded wire enter the terminal connection and the screw terminals are securely fixed to prevent poor contact.

(f) For other information about the products, please refer to www.meanwell.com for details.

(2) Modular Type

(a) Before any installation or maintenance work, please disconnect your system from the utility. Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient temperature may increase the internal component temperature and will require a de-rating in output current. Please refer to the specification sheets to receive the optimum mounting position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance must be kept when the adjacent device is a heat source.
(d) Keep enough insulation distance between mounting screws and internal components of Modular units. Please refer to case drawing on specifications to receive the maximum length of mounting screw.
(e) Recommended wires of Modular units are shown as below.

AWG1816141210
Rated Current of Equipment (Amp)6A6-10A10-16A16-25A25-32A32-40A
Cross-section of Lead( mm^2 )0.751.001.52.54
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more strips should not exceed 80% of current rating.

8

Make sure that all strands of each stranded wire enter the terminal connection and the screw terminals are securely fixed to prevent poor contact.

(f) For other information about the products, please refer to www.meanwell.com for details.

Installation Manual

(3) PCB Type

(a) Before any installation or maintenance work, please disconnect your system from the utility. Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient temperature may increase the internal component temperature and will require a de-rating in output current. Please refer to the specification sheets to receive the optimum mounting position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance must be kept when the adjacent device is a heat source.
(d) Recommended wires of PCB units are shown as below.

AWG1816141210
Rated Current of Equipment (Amp)6A6-10A10-16A16-25A25-32A32-40A
Cross-section of Lead( mm^2 )0.751.001.52.54
Note: The suggested maximum current above only suit for 1-4 strips, 5 or more strips should not exceed 80% of current rating.

8

6

Make sure that all strands of each stranded wire enter the terminal connection and the screw terminals are securely fixed to prevent poor contact.

(e) At least 5mm insulation distance on the bottom of PCB units should be kept and a Mylar film should be added between the units and the system. In addition, keep enough insulation distance of 15mm around the PCB units.
(f) PCB units greater than 120W may require a forced air/fan for cooling. Please refer to specifications to receive a minimum air intensity and air-flow direction.
(g) For other information about the products, please refer to www.meanwell.com for details.

(4) Adapter Type

(a) Before any installation or maintenance work, please disconnect your system from the utility. Ensure that it cannot be re-connected inadvertently!
(b) Mounting orientations other than standard orientation or operate under high ambient temperature may increase the internal component temperature and will require a de-rating in output current. Please refer to the specification sheets to receive the optimum mounting position and information about the de-rating curve.
(c) Fans or ventilation holes must be kept free from any obstructions. Also a 10-15 cm clearance must be kept when the adjacent device is a heat source.
(d) Before attaching the DC plug of an adaptor to equipment, please unplug the adaptor from the AC power and verify the unit is within the voltage and current rating on the equipment.
(e) Keep the linkage between the adaptor and its power cord tightly as well as connecting the DC plug to equipment properly.
(f) Protect the power cord of adaptors from being trodden on or being squashed.
(g) An approved power cord of an adaptor should greater or equal to SVT, 3G×18AWG or H03VV-F, 3G×0.75mm2.
(h) If the final equipment is not used for long period of time, disconnect the equipment from adaptors to avoid being damaged by voltage peaks or lightning strike.
(i) For other information about the products, please refer to www.meanwell.com for details.

- User Manual

NMP series:

http://www.meanwell.com/webapp/product/search.aspx?prod=NMP1K2&pdf=Tk1QLUUucGRm&a=4

● Warning / Caution !!

(1) Risk of electrical shock and energy hazard. All failure should be examined by a qualified technician. Please do not remove the case of the power supply by yourself! Please do not install the units in places with high moisture or near the water.
(2) Please do not install the units in places with high ambient temperature or near fire source. The maximum ambient temperature, please refer to their specifications.
(3) Output current and output wattage must not exceed the rated values on specifications.
(4) To avoid risk of electric shock, for class I products, the ground (FG) must be connected to protective earth.
(5) The Enclosed, Modular and PCB units are designed in accordance with EMC regulations and the related test reports are available by request. Since they are belong to component power supplies and will be installed inside system enclosure, when they are integrated into a system, the EMC characteristics of the end system must be re-verified again.

Manufacturer :

MEAN WELL ENTERPRISES Co., LTD.

No.28, Wuquan 3rd Rd., Wugu Dist.,

New Taipei City 24891, Taiwan

Tel: +886-2-2299-6100

Web: www.meanwell.com

Branch Office :

China

MEAN WELL (GUANGZHOU)

ENTERPRISES Co., LTD.

2F, A Building, Yuean Industry Park,

Huangcun, Dongpu Yown, Tianhe

District, Gungzhou, China

Post Code:510660

Tel: +86-20-2887-1200

Web: www.meanwell.com.cn

U.S.A.

MEAN WELL USA, INC.

44030 Fremont Blvd., Fremont,

CA 94538, U.S.A.

Tel: +1-510-683-8886

Web: www.meanwellusa.com

Europe

MEAN WELL EUROPE B.V.

Langs de Werf 8, 1185XT

Amstelveen, The Netherlands

Tel: +31-20-758-6000

Web: www.meanwell.eu

Declaration of China RoHS Conformity

In order to reduce the impacts on the environment and take the more responsibility for protecting the earth, MEAN WELL is confirming and announcing the conformity to China RoHS, an Administrative Measures for the Restriction of the Use of Hazardous Substances in Electrical and Electronic Products.

Environment Friendly Use Period Label

Mean Well RPT-60D - Environment Friendly Use Period Label - 1Observing SJT 11364-2014, Marking for the Restricted Use of Hazardous Substances in Electronic and Electrical Products
Observing SJ/Z 11388-2009, General Guidelines of Environment-friendly Use Period of Electronic Information Products Appendix B, adopting table look-up to verify the Environment Friendly Use Period

Names and Contents of Hazardous Substances Lists

Part NameHazardous Substances
Lead(Pb)Mercury(Hg)Cadmium(Cd)Hexavalent chromium( Cr^6+ )Polybrominated biphenyls(PBB)Polybrominated diphenyl ethers(PBDE)
PCB and its componentsXOXOOO
Metal structure partsXOOOOO
Plastic structure partsOOOOOO
AccessoriesOOOOOO
CablesXOOOOO
O: The concentration of the hazardous substances within the homogeneous material of that product is less than the concentration limits set by GB/T 26572-2011.X: The concentration of the hazardous substances within the homogeneous material of that product is over the concentration limits set by GB/T 26572-2011; however, it follows the standard advised by 2011/65/EU.