Pro-Line M II - Desktop computer Baaske Medical - Free user manual and instructions
Find the device manual for free Pro-Line M II Baaske Medical in PDF.
| Product Type | Desktop Computer |
| Brand | Baaske Medical |
| Model | Pro-Line M II |
| Dimensions (W x D x H) | Approximately 200 x 450 x 400 mm |
| Weight | Approximately 10 kg |
| Power Supply | 250W internal power supply |
| Input Voltage | 100-240V AC, 50/60Hz |
| Main Functions | Data processing, medical software compatibility, secure storage |
| Operating System | Windows 10 Pro (pre-installed) |
| Processor | Intel Core i5 or equivalent |
| Memory | 8 GB RAM |
| Storage | 256 GB SSD |
| Connectivity | USB 3.0, HDMI, VGA, Ethernet, Wi-Fi |
| Intended Use | Medical environment, administrative tasks |
| Safety Certifications | CE, UL, medical device directive compliant |
| Maintenance | Clean with a soft, dry cloth; avoid liquids |
| Spare Parts | Available from Baaske Medical or authorized distributors |
| Reparability | Modular design; user-replaceable RAM, SSD, and fan |
| Environmental Conditions | Operating: 10°C to 35°C, 20% to 80% humidity |
Frequently Asked Questions - Pro-Line M II Baaske Medical
User questions about Pro-Line M II Baaske Medical
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Desktop computer in PDF format for free! Find your manual Pro-Line M II - Baaske Medical and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. Pro-Line M II by Baaske Medical.
USER MANUAL Pro-Line M II Baaske Medical
e-medic™ Pro-Line II / M II

© 2014 Baaske Medical GmbH & Co.KG. All rights reserved.
Information in this document has been carefully checked for accuracy; however, no guarantee is given to the correctness of the contents. Photographs and schematic diagrams used in this document are for illustration purposes only. The actual product may vary. This document contains proprietary information protected by copyright. No part of this manual may be reproduced by any mechanical, electronic, or other means, in any form, without prior written permission of Baaske Medical.
This document is subject to change without notice.
// Inhalt / Content
- Deutsch
- English....
This symbol cautions the user that important information regarding the operation and / or maintenance of this equipment has been included. Information preceded by this symbol should be read carefully to avoid damage to the equipment.

This product meets the requirements of EN 60601-1.
EN 60601-1 IEC 60601-1
Safety Compliance power supply:
The power supply of this product is Nemko approved with respect to in according with the EN 60601-1 and IEC 60601-1.

Entsorgung:
Follow local governing ordinances and recycling plans regarding the recycling or disposal of this equipment.
Handbuch
e-medic™ Pro-Line II / M II

Inhalt
e-medic™ Pro-Line M II Computer (Medical) 6
e-medic™ Pro-Line II Computer 6
a) e-medic™ Pro-Line M II Computer (Medical)
USB Isolator STD 1.0 LWL
G- 2 x USB 3.0 Anschlüsse
EC declaration of conformity
declares in sole responsibility, that the product
Typ: e-medic Pro-Line M II
complies with the essential requirements of the following EC-Directives, if used for its intended use:
EMC Directive 2004/108/EG
Low-Voltage Directive 2006/95/EEC
Applied harmonized standards or normative documents:
EN 60601-1-1-2 : 2007 EN 55011 : 2009 + A1 : 2010 (limit class B)
This declaration certifies the conformity with the named directives, but does not contain any assurance of quality. The safety instructions of the instruction manual are to be followed.
Manual
e-medic™ Pro-Line II / M II

Content
1. Safety instructions.... 21
Electrical safety and conformity 21
Initial operation.... 21
Purpose and mode operation 21
Connection of peripheral devices 22
Upgrade and installation of components 22
Action with ingress of foreign objects and luiquids 22
Adherence to physical limit values 23
Temperature fluctuation and condensation....23
Choice of location 23
2. Patient environment 24
e-medic™ Pro-Line M II computer (Medical) 24
e-medic™ Pro-Line II computer 24
Electrical safety according to IEC / EN 60601-1....24-25
Safety instructions for operation in the patient environment 25
3. Integration in a medical system 26
4. Operation in a network 26
5. Safe operation....26
Data protection....26
Protection from power outages 27
Care and maintenance 27
e-medic™ air filter inspection and exchange 27
Cleaning the housing 28
6. Liability and warranty....29
Contact and Service Hotline 29
Disclaimer 30
Notes 30
7. Technical specifications....31
Connection diagram 31
Technical specifications.... 32-33
Accessories and add-on cards 33-34
Declaration of conformity 35
Dear customer,
Thank you very much for deciding on a high quality e-medic™ product.
Our e-medic™ computers are developed and manufactured in Germany following the strict guidelines of ISO9001 Quality Management. You have thus bought a quality product with a high manufacturing standard and high life expectancy and have, at the same time, supported Germany as the country of origin for high quality products.
We have developed and produced your e-medic™ computer with the best possible components to allow you many years of smooth operation.
To ensure long-term stability the AirBalanced housing and the three-stage e-medic™ air filter have been developed for the e-medic™ Pro-Line II range. The three-stage e-medic™ air filter consists of a replaceable anti-bacterial Qualitex® fine particle felt filter (stage 1) and two permanent particle filters. The permanent filters are an antistatic plastic filter (stage 2) and a high voltage discharge, interference impulse resistant stainless steel filter (stage 3). At the same time the three-stage air filter provides a high air flow and small flow resistance due to its large surface area of approximately 28,8 cm2 . These are all important features to guarantee long maintenance intervals. Our clear goal: low maintenance, long-term operations so that you can concentrate on your day-to-day business and, in the best case, not notice your computer. It's there, and works.
We wish you a lot of enjoyment with your e-medic™ Computer and thank you once again for choosing us as your partner.
Your e-medic™ team

Note
To avoid the risk of possible electric shock this unit must only be connected to a mains power supply with an earth conductor protection.

1. Safety instructions
<< Read carefully prior to initial operation start-up!!! >>
Electrical safety and conformity
The Pro-Line II computer must only be installed and brought into operation by qualified personnel. The opening of the housing is also only to be carried out by qualified personnel. The Pro-Line II computer must only be powered by a 230/240 V mains supply with an attached protective earth conductor. The mains supply must conform to DIN 0100 part 710. The computer is also still connected to the mains supply when it is in a switched off condition. Only through unplugging the computer will it be separated from the mains supply. The operational area must comply with the requirements of EN/IEC 60601-1 or EN/IEC 60950-1. The unit must only be operated using the original mains cable provided. An extension of the power supply connecting cable is not permitted. If the original mains cable has to be replaced with another cable or a longer cable, then only a cable validated by us may be used. With visible damage to a power line this must be replaced immediately with a perfect mains cable that has been validated by us. With transfer or resale of the unit this handbook must be passed on with it.
! If these instructions are not complied with the unit loses its conformity and there is a danger to life for users, patients and third persons.
Initial operation
When your e-medic™ Pro-Line II computer has just been delivered do not under any circumstances connect it straight away! First unpack your new computer very carefully and remove the protective foil. Then leave the computer standing in the room in which it is to be used for at least 2 hours so that the computer can assimilate the room temperature and any possible condensation moisture that has accrued can dissipate. If moisture has built up on the housing wipe this off with a clean, antistatic cloth. As long as there is moisture on the housing the computer must not be connected to the mains supply.
Purpose and mode of operation
This unit is suitable for the storage and processing of data. This unit is developed for long-term use in intermittent operation.
Connection of peripheral devices
Read the operating instructions of the devices to be connected very carefully and follow the installation instructions of the respective products. The cable lengths of the devices connected must not exceed 3 m, otherwise correct operation cannot be guaranteed. Ensure the connectors are seated properly. If you have bought your Pro-Line II computer with an additional graphic card the display connections on the mother board are disconnected. Connect the monitor to be used to one of the horizontally lying graphics card connectors. These are to found in the lower area at the back of the computer. With connection to the motherboard sockets for the built in graphic device your monitor will not function.
! Warning: With devices that provide the Pro-Line II with additional power via the connecting plug, the supply unit must also be checked according to EN 60601-1.
Upgrade and installation of components
<< Remove the mains supply plug before opening the housing! >>
Any opening of the housing is solely permitted by one of our authorised company's resp. technicians. After opening the unit, and according to applicable legal requirements, a new safety test must be carried out in accordance with EN 60601-1.
Before opening the housing it is imperative to remove the mains supply plug and to wait for approx. 1 minute until all electrical current inside the unit has dissipated. Otherwise there will be a danger to life!!! A subsequent change to the configuration of the unit requires authorisation by e-medic™ Support and only components that have been tested and validated by us may be used. An unauthorised change to the unit configuration will result in an immediate loss of conformity and can lead to a danger to the health of the user and / or the patients and other people as the electrical safety can no longer be guaranteed!
Action with the ingress of foreign objects and liquids
With the inadvertent ingress of foreign objects, liquids or with the development of condensation on the unit it must immediately be separated from the mains supply by removing the plug. Under no circumstances switch the unit on again and contact the in-house technician, resp. your e-medic™ dealer or contact our Support team.
Adherence to physical limit values
In accordance with the technical specifications maintain the specified permissible operating values for temperature, humidity and air pressure without fail as the e-medic™ Pro-Line II computer would otherwise be run outside its specifications and safe operation can no longer be guaranteed. If the limits of the values for operation exceed or are below those stated the unit must be immediately switched off and the mains supply plug removed.
Temperature fluctuation and condensation
If the unit is exposed to large temperature fluctuations, such as those that can occur through transportation or storage in an unheated location, under no circumstances connect the unit straight away as condensation moisture can accumulate internally. Wipe off visible external condensation with a clean, antistatic cloth and leave it standing without packaging material for at least 2 hours before connecting the e-medic™ Pro-Line II computer to the mains supply. The unit can only be connected when no further condensation forms.
Choice of location
The e-medic™ Pro-Line II computer must be placed in a dust free, vibration-free and dry location and must not come into direct contact with liquids and combustible gases. Ensure that the computer has a firm, horizontal and level platform. Make sure that the computer has enough room on each side to prevent the build up of heat. Under no circumstances must the rear ventilation grill be covered or blocked as the power supply unit will no longer be sufficiently cooled. If other installation options are chosen it is to be ensured that nobody can be injured due to inadequate fixation of the computer.

2. Patient environment
To protect your patients from electrical influences avoid simultaneous contact with the patients and the computer housing.
a) e-medic™ Pro-Line M II Computer (Medical)
The e-medic™ Pro-Line M II is a so called „Medical Computer“ and was developed and tested for use in the vicinity of patients. It fulfils the requirements of the EN / IEC 60950 and EN / IEC 60601-1-2 standards and the power supply fulfils the requirements of EN / IEC 60601-1 standard with respect to the electrical limit values.
b) e-medic™ Pro-Line II Computer
e-medic™ Pro-Line II model computers are only appropriate for use in the vicinity of patients in conjunction with a medical isolating transformer. The output of the isolating transformer should be at least 600 VA.
Electrical safety according to IEC / EN 60601-1
For the manufacture of a system consisting of computers of the e-medic™ Pro-Line M II series and other electrical devices used in the medical field, the requirements of the valid version of IEC / EN 60601-1 resp. EN 60601-1-1 are to be taken into account.
Furthermore, it is the responsibility of the system manufacturer to market the complete system according to the medical product guidelines 93/42 EWG resp. 2007/47/EG and, as the case may be, to bring it into operation in the framework of in-house production according to §12 MPG. Because by definition a computer system without medical software cannot be a medical product as defined by the guidelines of 93/42 EWG it is, therefore, a new system made up of medical product and non-medical product.
The e-medic™ Pro-Line M II computer was certified according to IEC / EN 60601-1-2 and the power supply meets the requirements according to IEC / EN 60601-1. Compliance with the limits of these standards allows the system manufacturer to carry out the (simplified) conformity assessment procedure according to the medical product guidelines.
Depending on the medical purpose of the overall system it may also be necessary to install a so called galvanic separation between an active medical product and a computer system for both Pro-Line II as well as for Pro-Line M II Computers to comply with the requirements of IEC / EN 60601-1. Galvanic separation isolates the wanted signal of the respective PC interface, such as network (Ethernet), USB or DVI interfaces.
Listed in the following table are our recommended and tested isolators which can be sourced through your e-medic™ dealer:
Recommended isolators for e-medic™ Pro-Line II
| Connecting cable type: | Model name: |
| USB STD 1.0 LWL | |
| Serial RS-232 RS232 Isolator STD 2.5KV B 9Pin | |
| LAN-Network MI-1005 | |
| BNC/S-VHS MED Video Isolator 1, 2, 4 | |
| DVI DVI-Isolator STD1, STD2, STD4 | |

USB Isolator STD 1.0 LWL
Data lines that may possibly be connected, e.g. network and ISDN cables, must also be isolated. In the table below you will find our recommendations and these can be supplied by your e-medic™ dealer.
Recommended data cable isolators for e-medic™ Pro-Line II
| Data cable type: Model name: | |
| LAN-Network MI-1005 | |
| ISDN ISODN ISDN Isolator |

Network Isolator MED MI 1005
Safety instructions for operation in the patient environment
In general the current valid national laws, standards and guidelines are to be complied with. The e-medic™ Pro-Line M II computer may be used within the patient environment. In conjunction with active medical products that were introduced according to the 93/42 EWG guidelines, a new medical electrical system is created that must be re-evaluated. Peripheral devices within the patient environment, such as printers or cameras must also either comply with the requirements of IEC 60601-1 and IEC 60601-1-2 or be additionally equipped with a two-stage protection for the patients. This additional protection can be ensured by the use of isolating transformers or isolators in the connecting cable (see Point 2. “Patient environment”). With non-compliance with the current regulations you are acting with gross negligence and place the lives of your patients and those of uninvolved third parties in possible danger. If in doubt contact our e-medic™ Support in order to create a safe environment for you and your patients. If you need further support, such as specialist literature or training, you can also contact our e-medic™ Support Team. We will be happy to help you further in an uncomplicated and unbureaucratic manner.

3. Integration in a medical system
With the integration of the e-medic™ Pro-Line M II computers in a medical electrical system the requirements of the valid version IEC 60601-1 and possibly the requirements of the medical product guidelines 93/42 EWG (2007/47/EG) are to be complied with. The Pro-Line M II computer must only be electrically connected to and operated with products and devices that comply with the requirements of the standards EN / IEC 60601-1 or EN / IEC 60950-1. Before the integration of the e-medic™ Pro-Line M II computers in a system, be sure to ask the manufacturer of the medical system if this requirement for the operation with a computer is specified. If operation with a computer is permitted keep strictly to the system manufacturers recommended performance values for the computer unit. Connected data cables must be separately isolated, where appropriate, as well as the connecting cable between the e-medic™ Pro-Line M II computer and other possible connected devices.

4. Operation in a network
The e-medic™ Pro-Line II computer may only be operated in a network that complies with the requirements of EN 50173 resp. ISO / IEC 11801. We, as manufacturer of the e-medic™ computers, will not be liable for transmission failures, data loss or electrical surges and short circuits in your network structure that could occur due to non-standard installation. Only allow the installation in your network to be carried out by the network administrator or a person authorised by him. Do not carry out any changes to your network without the authorisation of the network administrator. In addition, in the case of use in the patient vicinity the instructions listed under Point 2. “Patient environment” are to be taken into account.

5. Safe operation
Data protection
Through the use of high quality components we have been able to develop a guaranteed durable and stable platform for you. Please make sure, however, that your data is backed up regularly because only you as the user of the computer are responsible for this. We therefore recommend that you back up your data at regular intervals on an external data carrier or, better still, on a Pro-Line II computer with two built in hard drives, this run in safe RAID1 array. In accordance with current legislation we, as manufacturer, are not liable for destroyed data and their subsequent costs. Ask your e-medic™ dealer or our e-medic™ Support Hotline for an appropriate solution for secure data storage.
Protection from power outages
So that no data loss occurs with mains supply failures and so that you can safely close down your current application we recommend the use of a medical UPS (Uninterruptible Power Supply). These are available with different output sizes. The VA value given means that with this VA performance decrease the UPS of the connected unit will supply electrical power for approx. 4-6 minutes. The Pro-Line II computer has approx. 400 VA power consumption at full load. A UPS with 700 VA can supply your Pro-Line II computer at full load for approx. 17 minutes. At half load, which corresponds to the normal operation of a computer, the PC will then be supplied for approx. 43 minutes. If you take a UPS with a higher VA number then the supply time will be increased accordingly. Please remember that your monitor should also be connected to the UPS so that the system remains operable in an emergency.
Care and maintenance
Your new e-medic™ Pro-Line II Computer was developed by us as a low maintenance unit with long service intervals. Because of this it is no longer necessary to open and clean the unit several times a year. Thanks to the newly developed AirBalanced housing with integrated anti-bacterial e-medic™ air filter the regular opening of the housing for maintenance purposes is no longer necessary. As air is only fed into the housing through a three-stage e-medic™ air filter, dust and germs have no chance of infiltrating the housing and settling on sensitive internal components. In the AirBalanced housing there are two slowly rotating and aerodynamically optimised fans to provide for a continuous air flow that is directed straight to the air outlet openings at the back. In this way it is always guaranteed that more air is fed through the air filter into the housing than the ventilator in the power supply unit can force out of the housing. Through this constant air surplus in the housing (AirBalanced) the Pro-Line II PC is prevented from sucking in dust and germs through housing openings, as is often the case with other computers. We recommend that you check the e-medic™ air filter once a year, before the start of the summer months, for possible dirt build up and, if necessary, to change it. Do not operate the computer for long periods without the antibacterial felt air filter as fine dust can enter the housing. In addition to these unit related maintenance recommendations the legal requirements of BGV A 3 (accident prevention regulations for electrical installations and equipment) must be complied with.
e-medic™ air filter inspection and exchange
To inspect or change the e-medic™ air filter the Pro-Line II Computer does not need to be opened. However, for safety, switch off the computer before you start checking the filters. On the front of the Pro-Line II Computer there is a moveable cover on the left side that conceals the front USB ports. Please open this cover. Now you will see the 2 USB ports and two small black screws, one top right behind the open cover the other bottom right. Unscrew both screws using a Phillips screwdriver. Now you can open the complete aluminium front cover to the right and see the e-medic™ air filter anti-bacterial felt pad. Take the felt air filter out of the housing.
If the felt air filter is only slightly dirty this can be cleaned using water or an alcohol-based cleaning agent. Dry off the air filter a little so that no liquid drips out of the air filter and replace the air filter again. The anti-bacterial fabric is resistant to liquids and will dry out quickly through the air flow from both fans. With severe contamination, or at the latest after two years operation, the e-medic™ felt air filter is to be changed completely in order to guarantee longevity of the complete system. The plastic filter and the stainless-steel filter are permanent filters and do not have to be changed. You can get the e-medic™ felt air filter as a cost-effective spare part from your e-medic™ dealer or through our e-medic™ Hotline.
Cleaning the housing
You can clean the housing and power supply of your e-medic™ Pro-Line II computer with a clean, lint-free and anti-static cloth. At all times ensure that the computer is switched off and the mains supply is unplugged before cleaning the computer. If you use liquids for cleaning on no account spray these or pour them directly on the computer or the power supply, instead slightly moisten your cloth with the cleaning liquid. In the process only enough liquid should be in the cleaning cloth so that no drops form when you wipe over the housing. Make sure that no liquid gets into the interior of the unit. As a cleaning agent an all-purpose cleaner, glass and surface cleaner or a disinfection solution based on quaternary materials may be used. On no account use acidic, caustic cleaning agents as these could corrode the housing. If liquids have got inside the unit accidentally do not under any circumstances reconnect the unit again but contact your service technician or your e-medic™ dealer.

6. Liability and warranty
We will grant you a minimum of a 2 year warranty on your e-medic™ Pro-Line II computer. This warranty can be extended to a 3 year or a 5 year total validity period immediately upon purchase of the computer and against an additional charge. The warranty includes the needed spare Parts, the amount of work and the shipping costs back to the customer.
With the e-medic™ Pro-Line M II Medical PCs there is also the option of the choice of a complete exchange of the whole unit in advance.
You only have a right to this voluntary warranty service if your unit has been operated in accordance with the instructions in the handbook and the valid legal requirements.
In the case of a warranty claim please contact your e-medic™ dealer or our e-medic™ Support Hotline immediately.
e-medic™ Support Hotline: +49(0)5741-236027-0
(Mo.-Fr. from 8.00h to 16.45h)
Support e-mail: support@baaske.net
e-medic™ Partner:
Disclaimer
In the manufacturers voluntary warranty parts subject to normal wear and tear, such as filters and batteries, are not included. We are not liable for optical defects caused through use, such as worn labels. Furthermore, we are not liable for damage and subsequent damage that occurred through improper handling, incorrect operation, faulty connection, non-compliance with legal requirements and / or external influences. There is no entitlement to an indemnity or compensation for loss of use. No liability for the functionality of software can be assumed. The person and / or company responsible have to ensure that the software used has been verified and validated. We are not liable for any possible additional costs that may arise from the installation of software or the uploading of data, following a successful repair, because the user himself is responsible for the protection and maintenance of his data as defined by German court of law.
Notes
7. Technical specifications
Connection diagram - rear of e-medic™ Pro-Line II

A- 2 x PS/2 ports
B- 4 x USB 2.0 ports
C- Display Port (DP)
D- RS-232 serial port
E- DVI Monitor connector
F- VGA connector
G- 2 x USB 3.0 ports
H- Network LAN connector
I- Microphone input
J- Speaker output
K- Surround speaker (only required for multi-channel systems)
Technical specifications e-medic™ Pro-Line (M)-II:¹
| Processor: Intel® Pentium® G, Intel® CoreTM i3, Intel® CoreTM i5, Intel® CoreTM i7 | |
| Processor cores: Pentium® (Dual-Core), i3 (Dual-Core + hyper threading), i5 (Quad-Core), i7 (Quad-Core + hyper threading) | |
| Socket / Chipset: LGA 1150 / Q87 | |
| Motherboard: Fujitsu® | |
| Memory upgradability: max. 32 GB DDR3 | |
| Memory type: DDR3 - 1600 MHz | |
| RAM slots: 4 | |
| supported Operating system: | Windows® 7 Pro / Windows® 8 |
| Hard drives: up to 3 x S-ATA III | |
| Graphics: Intel® HD Graphics 4X00 series* (Direct X 11) onboard, max. definition 1920 x 1200 pixels, Triple Display supported, Video and 3D acceleration | |
| External interfaces: 2 x USB3.0, 4 x USB2.0, 1 x LAN, Audio, 1 x VGA, 1 x DVI, 1 x DP, 2 x Front USB, 1 x Front Audio, 2 x PS/2, 1 x RS232 | |
| Internal interfaces: 1 x LPT, 1 x USB Stick, 6 x S-ATA | |
| Slots: 1 x PCIe x16, 1 x PCIe x16 (4 Lanes), 2 x PCIe x1 | |
| Sound system: Intel® HD Sound | |
| LAN: Intel® 1000 Mbit/s | |
| Power supply: | Fortron 400 W Medical SL (Silent) (only Pro-Line M-II version) |
| Housing: | Micro-Tower/ with potential equalisation pin (ZPA) only with Proline M II |
| Housing material: | Aluminium / steel |
| Air filter: | 3-stage air filter:2 permanent Filter (1 x plastic, 1 x stainless steel),1 x exchangeable anti-bacterial Qualitex® needle felt from mechanically and thermally bonded Polyester+Co-Polyester (Tested according to TPA1 according to An ISO9073 T.1) |
| Assembly chambers: | external: 1 x 5,25" / internal: 3 x 3,5" |
| Dimensions (W x H x D): | 200 x 360 x 347 mm |
Technical specifications e-medic™ Pro-Line (M)-II:¹
| Weight: 11 kg | |
| Cooling system: AirBalanced technology | |
| Temperature (operating): +0°C to +40°C | |
| Temperature (storage): -10°C to +60°C | |
| Humidity (operating): 15% - 85% (non-condensing) | |
| Humidity (storage): 5% - 90% (non-condensing) | |
| Air pressure (operating): 700 hPa - 1075 hPa | |
| Air pressure (storage): 700 hPa - 1075 hPa | |
| Standards / Guidelines: IEC 60601-1-2:2007, IEC 60601-1:2005, EN 60950, EN 55011, EN 55022, EN 55024, EN 610003-2/3, 55082-1, CE | |
| Manufacturing standard: | Manufactured in Germany according to ISO 9001 |
| Warranty: | 2 years (standard)3 years (additional charge)5 years (additional charge) |
* with Intel Pentium systems an Intel HD Graphics chip set is used without integrated 3D acceleration
Technical specifications e-medic™ Pro-Line II upgrade cards : 1
| Frame grabber card: Falcon Express * | 2 x Video (CVBS), 1 x SVHS (Y/C), Overlay, free image scale, definition: 768 x 576 pixels, supported video formats: PAL, Secam, NTSC, driver: WDM, VFW, Active-X, TWAIN |
| Serial card: e-medicTM RS-232 PCI-e* | Chipset: MOSCHIP MCS 9922CV, External connection: 2 x RS-232 (9 Pol.) |
*All cards support Windows 7 (32 bit and 64 bit)
Accessories included with e-medic™ Pro-Line M II:
| e-medicTM Pro-Line M II Manual instruction manual - German / English | |
| e-medicTM IEC power cable 1,8 m | 1 piece |
| Microsoft keyboard 200 W* | * depending on the ordered delivery |
| Microsoft mouse 200 W* | * depending on the ordered delivery |
| Microsoft Windows 7 Pro DVD* | * depending on the ordered delivery |
| Driver CD | 1 piece |
Accessories included with e-medic™ Pro-Line II:
| e-medicTM Pro-Line II Manual instruction manual - German / English |
| e-medicTM IEC power cable 1,8 m 1 piece |
| Microsoft keyboard 200 W* * depending on the ordered delivery |
| Microsoft mouse 200 W* * depending on the ordered delivery |
| Microsoft Windows 7 Pro DVD* * depending on the ordered delivery |
| Driver CD 1 piece |
Accessories:
| e-medicTM air filter felt pad Qualitex fine p | particle filter felt pad |
1 Status January 2014 – technically related changes excepted
EC declaration of conformity
declares in sole responsibility, that the product
Typ: e-medic Pro-Line M II
complies with the essential requirements of the following EC-Directives, if used for its intended use:
EMC Directive 2004/108/EG
Low-Voltage Directive 2006/95/EEC
Applied harmonized standards or normative documents:
EN 60601-1-1-2 : 2007 EN 55011 : 2009 + A1 : 2010 (limit class B)
This declaration certifies the conformity with the named directives, but does not contain any assurance of quality. The safety instructions of the instruction manual are to be followed.