HDS4202M-N - Measurement Owon - Free user manual and instructions

Find the device manual for free HDS4202M-N Owon in PDF.

📄 84 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice Owon HDS4202M-N - page 2
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Handheld Digital Storage Oscilloscope & Multimeter
Model Owon HDS4202M-N
Dimensions 18 cm x 11.5 cm x 4 cm
Weight 645 g
Display 3.7-inch color LCD, 640 x 480 pixels, 65536 colors
Power Supply AC adapter 100-240 V, 50/60 Hz; Built-in 7.4 V Li-ion battery
Oscilloscope Bandwidth 200 MHz (typical)
Sampling Rate Up to 1 GS/s (real-time)
Record Length 6,000 points per channel
Multimeter Resolution 3 3/4 digits (4000 counts)
Multimeter Functions DC/AC voltage, DC/AC current, resistance, diode, capacitance, continuity
Max Input Voltage (Oscilloscope) 400 V peak (with 10:1 probe)
Max Input Voltage (Multimeter) 1000 V DC or AC peak (750 V AC RMS)
Trigger Modes Edge, Video (NTSC/PAL/SECAM)
Automatic Measurements 18 types: frequency, period, mean, peak-to-peak, RMS, Vmax, Vmin, etc.
Connectivity USB (mass storage and communication), RS-232
Warranty 3 years (1 year for accessories)
Safety Compliance CE (EN61010-1, EN61326-1), CAT II
Maintenance Clean with soft dry cloth; do not spray liquids; charge battery every 4 hours
Repairability Battery replaceable by qualified personnel; use only specified lithium battery

Frequently Asked Questions - HDS4202M-N Owon

How do I switch between oscilloscope and multimeter modes?
Press the DMM/OSC key to toggle between oscilloscope (DSO) and multimeter (DMM) modes.
What is the maximum input voltage for the oscilloscope?
The maximum input voltage is 400 V peak (with a 10:1 probe). For direct input, do not exceed 42 V peak (30 Vrms) to avoid electric shock.
How can I save waveform data?
Insert a USB mass storage device (FAT32, allocation unit size 4096) into the USB port and press the COPY key. The waveform is saved as a .bin or .bmp file.
The oscilloscope does not power on. What should I do?
Charge the battery using the AC adapter for at least 15 minutes before turning on. If it still does not start, contact Owon service.
How do I perform an auto-calibration?
Press MENU, select FUNCTION, then press F2 (Auto calibration). Remove all probes and cables, then press F2 again to start. This is recommended when temperature changes exceed 5°C.
Can I measure AC voltage with the multimeter?
Yes. Press the V key, then press SET to select ACV. The range is up to 750 V RMS, 40-400 Hz.
How do I use the cursor measurement feature?
Press MENU, select CURS MEAS, then choose Voltage or Time. Use the OPTION keys to move the cursors and read values at the bottom of the screen.
What does the 'Stop' indicator mean?
It means the screen is frozen. Press the RUN/STOP key again to resume live measurement.
How often should I clean the device?
Clean the exterior with a soft, dry cloth as needed. Avoid using abrasive chemicals or spraying liquids. Ensure the device is dry before use.
Where can I find replacement probes or accessories?
Contact Owon or an authorized dealer. Standard accessories include oscilloscope probes, multimeter test leads, AC adapter, USB cable, and a carrying case.

User questions about HDS4202M-N Owon

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Measurement in PDF format for free! Find your manual HDS4202M-N - Owon and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. HDS4202M-N by Owon.

USER MANUAL HDS4202M-N Owon

July 2015 edition Ver1.0.0

Copyright © Lilliput Company. All rights reserved.

The Lilliput's products are under the protection of the patent rights, including ones which have already obtained the patent rights and those which are applying for. The information in this manual will replace all that in the materials published originally.

The information in this manual was correct at the time of printing. However, Lilliput will continue to improve products and reserves the rights to change specification at any time without notice.

owon is the registered trademark of the Lilliput Company.

General Warranty

Lilliput warrants that the product will be free from defects in materials and workmanship for a period of 3 years from the date of purchase of the product by the original purchaser from the Lilliput Company. And the warranty period of accessories such as probe, battery, adapter is one year. This warranty only applies to the original purchaser and is not transferable to the third party. If the product proves defective during the warranty period, Lilliput either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Lilliput for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Lilliput.

In order to obtain service under this warranty, Customer must notify Lilliput of the defect before the expiration of the warranty period. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Lilliput, and with a copy of customer proof of purchase.

This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Lilliput shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Lilliput representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Lilliput supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.

Please contact the nearest Lilliput's Sales and Service Offices for services or a complete copy of the warranty statement.

For better after-sales service, please visit our website and register the purchased product online.

Excepting the after-sales services provided in this summary or the applicable warranty statements, Lilliput will not offer any guarantee for maintenance definitely declared or hinted, including but not limited to the implied guarantee for marketability and special-purpose acceptability. Lilliput should not take any responsibilities for any indirect, special or consequent damages.

Table of Contents

1. DECLARATION OF CONFORMITY ...... 1

1.1 Declaration of Conformity 1

1.2 Package Contents ...... 2

2. SAFETY INFORMATION......3

2.1 Safety Symbols and Terms.... 3

2.1.1 Safety Symbols 3

2.1.2 Safety Terms.... 3

2.2 General Safety Information 3

3. GENERAL CHARACTERISTICS......6

4. PERFORMING THE GENERAL INSPECTION ...... 7

4.1 Performing the General Inspection 7

4.1.1 Check whether there is any Damage on it Due to Transportation....7

4.1.2 Make a Check on Accessories....7

4.1.3 Make a Check on the Complete Instrument....7

5. INPUT CONNECTIONS 8

5.1 Input Connections 8

5.1.1 Input connections 8

5.1.2 The connection of 1 kHz/5 V Square-wave test signal 9

5.2 Front Panel and Keys Overview 9

6. USING THE OSCILLOSCOPE 12

6.1 About this Chapter 12

6.2 Powering Up the Oscilloscope....12

6.3 Oscilloscope Operation Window....12

6.4 Menu Description....14

6.5 Manually Setting the Vertical System, Horizontal System and Trigger Position ……14

6.5.1 Setting the Vertical System....14

6.5.2 Setting the Horizontal System and Trigger Position 15

6.6 Recall factory settings....16

6.7 Input Connections 17

6.8 Displaying an Unknown Signal with Auto Set....17

6.9 Automatic Zero-returning of Trigger Horizontal Position and Trigger Level Position ·18

6.10 Automatic Measurements 18

6.11 Freezing the Screen 19

6.12 Using Average for Smoothing Waveforms....19

6.13 Using Persistence to Display Waveforms....20
6.14 Using Peak Detection to Display Glitches 21
6.15 Selecting AC-coupling 22
6.16 Reversing the Polarity of the Displayed Waveform 23
6.17 Use USB mass storage device to save waveform data 23

7. USING THE MULTIMETER 29

7.1 About this Chapter 29
7.2 Making Meter Connections 29
7.3 Multimeter Operation Window 29
7.4 Making Multimeter Measurements 30

7.4.1 Measuring Resistance Values 30
7.4.2 Measuring Diode 31
7.4.3 On-off Test 32
7.4.4 Measuring Capacitance 32
7.4.5 Measuring DC Voltage 33
7.4.6 Measuring AC Voltage 34
7.4.7 Measuring DC Current 35
7.4.8 Measuring AC Current 36

7.5 Freezing the Readings....37
7.6 Taking a Relative Measurement 38
7.7 Selecting Automatic/ManualRange Adjustment 39

8. ADVANCED FUNCTION OF OSCILLOSCOPE 40

8.1 About this Chapter 40
8.2 Setting the Vertical Channel 40

8.2.1 Setting the Channel Coupling 41
8.2.2 Open and Close Settings on Channel 42
8.2.3 Setting the probe attenuation 42
8.2.4 Setting of Inverted Waveform 43

8.3 Setting the Trigger System 43
8.4 Triggering Control 43

8.4.1 Edge Trigger....44
8.4.2 Video Trigger 44

8.5 Setting the Acquiring Mode......47
8.6 Display Setting....47

8.6.1 Display Style....47
8.6.2 Persistence 48
8.6.3 Cymometer....48

8.7 Waveform Saving Setups 49

8.7.1 Waveform Saving Setups in normal mode 49
8.7.2 Waveform Saving Setups in FFT mode 50

8.8 Function Setting Menu ....51

8.9 Making Automatic Measurements ....52
8.10 Setting the Cursor Measurements ....53

8.10.1 Setting the cursor measurement in normal mode ....53
8.10.2 Setting the cursor measurement in FFT mode 56

8.11 Autoscale 57
8.12 Record 58
8.13 Using FFT 61
8.14 System State Menu 67

8.14.1 Real time clock 68

8.15 Setting of Time Base Mode 68
8.16 Data Transmission....70

9. TROUBLE SHOOTING.... 71

10. APPENDIX 72

10.1 Appendix A: Specifications....72

10.1.1 Oscilloscope 72
10.1.2 Meter 74
10.1.3 General Specifications ......77

10.2 Appendix B: Maintenance and Cleaning 78

10.2.1 Maintenance 78
10.2.2 Storage of Oscilloscope....78
10.2.3 Replacing the Lithium Battery Unit 78

1. Declaration of Conformity

1.1 Declaration of Conformity

HDS-N Series Handheld Digital Storage Oscilloscope & Multimeter

Manufacturer: Fujian Lilliput Optoelectronic Technology Co., Ltd.

Statement of Conformity

Base on the result using appropriate standards, the product is in conformity with

Electromagnetic Compatibility Directive 2004/108/EC

Low Voltage Directive 2006/95/EC

Sample tests

Standards used:

EN61010-1: 2001 (2nd edition)

Safety Requirements for Electrical Equipment for Measurement, Control, and

Laboratory use-part 1: General Requirements

EN61326-1: 2006

Electrical Equipment for Measurement, Control and Laboratory use-EMC

Requirements-part 1: General Requirements

EN61000-3-2: 2000+A2: 2005

Electromagnetic Compatibility (EMC)-part 3: Limits-Section 2: Limits for Harmonic

Current Emissions (Equipment Input Current less than/equal to 16A per phase)

EN61000-3-3: 1995+A1: 2001

Electromagnetic Compatibility (EMC)-part 3: Limits-Section 3 Limitation of Voltage

Fluctuations and Flicker in Low-Voltage Supply systems for Equipment with Rated

Current less than or equal to 16A

The tests have been performed in typical configuration.

This conformity is indicated by the symbol CE, i.e. "Conformite Europeenne".

1.2 Package Contents

See the picture below. This picture is only for your reference. If any discrepancy found between certain picture and product, please look the actual product.

#DescriptionStandardOptional
1Oscilloscope and Battery
2AC-DC adapter
3Oscilloscope Probe × 1 (grey)
4Multimeter test lead × 2 (black and red)
5USB Mass Storage connect cable
6Extension module for capacitance measurement
7Probe adjustment tools
8USB communication cable or RS-232C communication cable
9User Manual
10CD-ROM (software)
11Metal carrying case
12Soft carrying case
13the output terminal of 1 kHz 5 V square-wave test signal

Labeled diagram of electronic testing equipment including a multimeter, capacitors, and a case with accessories

Figure 1: Digital Oscilloscope Parts

2. Safety Information

In order to ensure the correct using and the best efficient service, please carefully read this user's manual.

2.1 Safety Symbols and Terms

2.1.1 Safety Symbols

These symbols may appear in this manual or on the instrument.

Owon HDS4202M-N - Safety Symbols - 1

Warning: "Warning" identifies conditions and actions that pose hazards to the users.

Owon HDS4202M-N - Safety Symbols - 2

Caution: "Caution" identifies conditions and actions that may damage the product or other properties.

DANGER:

High Voltage

Refer to the Manual

Protective Conductor Terminal

Chassis Ground

Earth (ground) Terminal

Owon HDS4202M-N - Safety Symbols - 3

Owon HDS4202M-N - Safety Symbols - 4

Owon HDS4202M-N - Safety Symbols - 5

Owon HDS4202M-N - Safety Symbols - 6

Owon HDS4202M-N - Safety Symbols - 7

2.1.2 Safety Terms

The following terms may appear on the instrument:

Danger: The term "Danger" is used in this manual to indicate that when you read this mark, personal injury may be caused to you immediately.

Warning: The term "Warning" is used in this manual to indicate that when you read this mark, personal injury may not be caused to you immediately, but you need to be cautionary.

Notice: The term "Notice" is used in this manual to indicate that damages may be caused on this product or other properties.

2.2 General Safety Information

Carefully read the following safety information in order to avoid any personal injury and damage on this product or any products connected with it. This product can only be used in the specified

applications to prevent any possible dangers.

Owon HDS4202M-N - General Safety Information - 1

Warning:

To avoid fire or electrical shock, please use proper power adapter. Use only the power adapter appointed by the manufacturer and subject to approval of being used in the user's country.

Owon HDS4202M-N - Warning: - 1

Warning:

The two channels of the oscilloscope are non-isolated electrically. The channels should adopt common basis during measuring. To prevent short circuits, the 2 probe ground must not be connected to 2 different non-isolated DC level.

Owon HDS4202M-N - Warning: - 1

Warning:

The channels should adopt common basis during measuring. To prevent short circuits, the 2 probe ground must not be connected to 2 different non-isolated DC level.

The diagram of the oscilloscope ground wire connection:

Owon HDS4202M-N - Warning: - 1

flowchart
graph LR
    A["Signal Input"] --> B["Probe"]
    B --> C["Oscilloscope"]
    C --> D["AC"]
    D --> E["Adapter"]
    E --> F["Electrical Outlet"]
    G["Ground Clip"] --> B

The diagram of the ground wire connection when the battery-powered oscilloscope is connected to the AC-powered PC through the ports:

Owon HDS4202M-N - Warning: - 2

flowchart
graph LR
    A["Signal Input"] --> B["Probe"]
    B --> C["Oscilloscope (Battery-power)"]
    C --> D["PC"]
    D --> E["Electrical Outlet"]
    F["Ground Clip"] --> B
    G["USB/COM Cable"] --> C

It is not allowed to measure AC power when the oscilloscope is AC powered through the adapter, or when the battery-powered oscilloscope is connected to the AC-powered PC through the ports.

Owon HDS4202M-N - Warning: - 3

Warning:

To avoid fire or electrical shock if a test tool input is connected to more 42V peak (30Vrms) or on circuits of more than 4800VA:

  • Use only insulated voltage probes, test leads and adapter supplied with the test tool, or indicated by OWON as suitable for the Oscilloscope & Multimeter.
  • Before use, inspect voltage probes, test leads and accessories for mechanical damage and replace when damaged.
  • Remove all probes, test leads and accessories that are not in use.

● Always connect the power adapter first to the AC outlet before connecting it to the Oscilloscope & Multimeter.
- Do not apply voltages that differ more than 400 V from earth ground to any input when measuring in a CAT II environment.
- Do not apply voltages that differ more than 400 V from each other to the isolated inputs when measuring in a CAT II environment.
- Do not apply input voltages above the rating of the instrument Use caution when using 1:1 test leads because the probe tip voltage will be directly transmitted to the Oscilloscope & Multimeter.
- Do not use exposed metal BNC or banana plug connectors.
- Do not insert metal objects into connectors.
● Always use the Oscilloscope & Multimeter only in the manner specified.
- Voltage ratings that are mentioned in the warning are given as limits for "working voltage". The represent V ac rms (50-60Hz) for AC sine wave applications and as V dc for DC applications. Overvoltage Category II refers to local level, which is applicable for appliances and portable equipment.

Only qualified technical personnel are permitted to perform maintenance.

Pay attention to the nominal values of all terminals: To avoid fire or electric shock, please keep a watchful eye on all nominal values and marks specified for this product. Before any connection performed on this product, carefully read the user's manual of the product for further information of nominal values.

No operation is allowed without the instrument cover plate: If the cover plate or panel has been removed, do not perform any operation on this product.

No touch is allowed on bare conductors: When the product is powered on, do not touch any bare joints or parts of the scope meter.

Operation is prohibited in case of any undetermined failure: When in doubt any damage on this product, consult the qualified personnel for checking on it.

Keep ventilation in good condition: Refer to the user manual for detail installation instructions in order to fix this product correctly and provide it with good ventilation conditions.

No operation is allowed under a humid environment.

No operation is allowed under an explosive environment.

Keep clean and dry on the product surface.

Using the Oscilloscope & Multimeter in a manner not specified may impair the protection provided by the equipment.

3. General Characteristics

Oscilloscope

■ 2 in 1 (Multimeter function support);
■ Record length of 6,000 points for each channel;
■ Reading-out with the cursor;
■ Eighteen automatic measurement functions;
■ Autoscale function;
■ Color liquid crystal display of high resolution and high contrast with adjustable back light;
■ Storage and call-out of waveforms;
■ Automatic setting function provided capable of fast setting;
■ FFT function
■ Implementation of detecting the average and peak values of the waveform;
■ Edge and video triggering function;
■ RS232 or USB communication ports;
■ Multiple Language User Interface.

Multimeter

■ 3 and 3/4 digits;
■ Volts, Amps, Ohms, Diode, Capacitance, Continuity measurement;
■ 10A maximum amplitude;
■ Isolated inputs between oscilloscope and multimeter.

4. Performing the General Inspection

4.1 Performing the General Inspection

When you have got a new HDS-N series oscilloscope, it is suggested that you should perform a general inspection on the instrument according to the following steps.

4.1.1 Check whether there is any Damage on it Due to Transportation

If the packing boxes or foam cushions are found in serous damage, keep them in a proper place until the complete instrument and accessories have passed the electrical and mechanical tests.

4.1.2 Make a Check on Accessories

The accessory list has been described in the picture of Digital Oscilloscope Parts of this manual. You can make a check and find whether there is any accessory loss with reference to the Appendix. In case of any accessory loss or damage, consult the OWON dealer responsible for such a business or the local office of OWON.

4.1.3 Make a Check on the Complete Instrument

If the instrument is damaged in its appearance or it fails in normal operation or performance test, consult the OWON dealer responsible for such a business or the local office of OWON. If the instrument is damaged due to transportation, keep the packing in a proper place and consult the transportation department and the OWON dealer responsible for such business, who will provide an instrument replacement or maintenance.

5. Input Connections

5.1 Input Connections

5.1.1 Input connections

See the following Figure 2:

Oeon 1 2 3 4 5 6

Figure 2: Input connections

Description:

  1. The power adapter is supplied for AC power supply and battery recharging.

  2. Multimeter test lead.

  3. Multimeter input jacks, including four circular banana jacks. The first jack measuring the current 2 A-10A, the second jack on the current Measurement, the third COM ground input and the fourth measuring voltage, resistance, Capacitance input.

  4. Oscilloscope probe.

  5. Oscilloscope channel input.

  6. The output terminal of 1 kHz/5 V square-wave test signal.

5.1.2 The connection of 1 kHz/5 V Square-wave test signal

At the left side of the oscilloscope, it is a port for testing 1 kHz/5 V square-wave signal which is used to adjust the probe, shown as Fig.3

owon HDS1021M-N F1 F2 F3 F4 F5 AUTO SET SHIFT VERTICAL ZERO CHANNEL SETTING OPTION RUN STOP A V R SET 10A≡ mA≡ COM V/μC 10A≡ mA≡ COM V/μC

Fig.3. The connection of Square-wave test signal

5.2 Front Panel and Keys Overview

See the following Figure 4:

owon 1 2 3 4 F1 F2 F3 F4 F5 6 7 5 12 9 13 10 14 11 15 25 8 16 17 22 18 21 19 20 24 26 27 28 29 AUTO SET COPY CARM DISC + + VERTICAL ZERO CHANNEL SETTING TYPICAL MENS A V R SET 10 AM mA COM V/Ω/C MAX MAX MAX MAX

Figure 4: Front Panel

Description:

  1. AC adapter Port
  2. RS-232C Port.
  3. USB Port.
  4. USB Mass Storage Port.
  5. Power switch.
  6. F1 - F5: Switch or Adjust options for each menu.

  7. AUTO SET: Under DSO mode, automatically selects the horizontal scales, vertical scale, and trigger level according to the input signal.

  8. COPY: Press to save the waveform data into the USB Mass Storage Device.

  9. ▲ (red): Adjust vertical position in Channel.

  10. VERTICAL ZERO (red): Adjust vertical position to zero.

  11. ▼ (red): Adjust vertical position.

    • (red): Adjust voltage scale.
  12. CHANNEL SETTING (red): Enter Channel settings directly.

    • (red): Adjust voltage scale.
  13. RUN/STOP: key for running or stopping the operation

  14. LIGHT: backlight control.

  15. DMM/OSC: Operation mode switching key between oscilloscope and multimeter.

  16. MENU ▲: Choose the upper item on the menu list.

  17. MENU: Show / Hide the menu

  18. MENU ▼: Choose the lower item on the menu list.

  19. OPTION: the key for DSO setting and combines with four arrow keys. Then to set main time base, trigger horizontal position and trigger vertical position. Furthermore adjust cursor 1 (V1 or T1) and cursor 2 (V2 or T2) position during cursor measurement.

  20. ▲ (OPTION): Oscilloscope display upward adjustment key.

  21. ▼ (OPTION): Oscilloscope display downward adjustment key.

  22. ▶ (OPTION): Oscilloscope right-direction adjustment key.

  23. ◀ (OPTION): Oscilloscope left-direction adjustment key.

  24. A: Selects DMM current measurement

  25. V: Selects DMM voltage measurement

  26. R: Selects DMM impedance, diode, continuity and capacitance measurement

  27. SET: convert AC and DC during measure current or voltage in Multi-meter; convert resistance, diode, on-off and capacitance measure during resistance measuring.

6. Using the Oscilloscope

6.1 About this Chapter

This chapter provides a step-by-step introduction to the scope functions. The introduction does not cover all of the capabilities of the scope functions but gives basic examples to show how to use the menus and perform basic operations.

6.2 Powering Up the Oscilloscope

Connect oscilloscope to AC power via AC-DC adapter as shown in Figure 1. (The oscilloscope may still work with built-in Li-ion battery even without AC power supply).

Turn the oscilloscope on by pressing down the power on/off key

The instrument then performs self checking after power on. A greeting window and a sentence "Press any key to continue." will display on the screen when the system finishes self checking.

The users can press any key to enter the measuring function.

The oscilloscope is powered up in its last setup configuration.

6.3 Oscilloscope Operation Window

See the following figure 5:

1 2 3 4 5 6 F 1.000KHz Vp 122.0v H -240.0us Trig'd 16 14 7 15 - Time Base - Trig 50v- Edge Trig mode Pising Slope DC Coupling Auto Mode Holdoff 13 12 11 10 8 7 9

Figure 5: Oscilloscope Operation Window

Description:

  1. Battery electric quantity indicating symbols.
  2. Auto measurement window 1, in which "F" means frequency, "T" means cycle, "V" means the average value, "Vp" the peak-peak value, "Vr" the root-mean-square value, "Ma" the maximum amplitude value, "Mi" the minimum amplitude value, "Vt" the Voltage value of the waveform's flat top value, "Vb" the Voltage value of the waveform's flat base, "Va" the amplitude value, "Os" the overshoot value, "Ps" the Preshoot value, "RT" the rise time value, "FT" the fall time value, "PW" the +width value, "NW" the -Width value, "+D" the +Duty value, "-D" the -Duty value.
  3. Auto measurement window 2.
  4. The pointer indicates the horizontal triggering position.
  5. This reading gives the Time Difference between the horizontal triggering position and the screen centerline. It reads zero when the pointer is in the center of the screen.
  6. The trigger state indicates the following information.

Auto: The oscilloscope is working in the automatic mode and displaying the waveform under the non-trigger state.

Trig'd: The oscilloscope has detected a trigger and collecting the information generated after the trigger.

Ready: All pre-triggered data have been captured and the oscilloscope has been ready to receive trigger signals.

Scan: The oscilloscope can gather and display the waveform data continuously in scanning mode.

Stop: The oscilloscope has stopped collecting the waveform data.

  1. Green pointer indicates vertical trigger position when edge trigger, and no green pointer when video trigger.
  2. A hidden-style menu: With the MENU key pressed, you can view or hide the menu.
  3. Menu setting options: There are different setting options for different menus.
  4. It reads the value of trigger voltage level.
  5. The reading gives the value of primary time base.
  6. These graphics present the coupling modes of channel. The graphic "\~" indicates AC, the graphic "-" indicates DC, the graphic " 1= " indicates GND.
  7. This reading shows the vertical Voltage Unit Scale of channel.
  8. The red pointer gives the grounding datum point of the waveform on channel, which is the zero position of channel. No display of this pointer indicates that the channel has not been

opened.

  1. OPTION operation prompt: There are different prompts for different OPTION operations.
  2. Waveform display area, and the wave shows red.

6.4 Menu Description

The following example shows how to use the tool's menus to select a function, as shown in the following figure.

  1. Press the MENU key to display the Function Menu on the right of the screen and the corresponding optional settings on the bottom. Press MENU again will hide the Function Menu.
  2. Press the MENU ▲ or MENU ▼ key to select different function menus.
  3. Choose one key from F1 to F5 and press it to change function setting.

See the following figure 6:

F 1.000KHz Vp124.0v M 860.0us Trig'd TRIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 50v- M 500ms 0.00ms Edge Trig mode Bising Slope DC Coupling Auto Mode Holdoff

Figure 6: the Tool's Menus

6.5 Manually Setting the Vertical System, Horizontal System and Trigger Position

6.5.1 Setting the Vertical System

To set the vertical position of channel:

(1) The key ▲ (red) and ▼ (red) can be used to adjust the vertical position of channel; the relevant information will be displayed at the left bottom of the screen.
(2) The key + (red) and - (red) can be used to adjust the voltage scale of channel; the relevant information will be displayed at the left bottom of the screen.

Note: Press VERTICAL ZERO (red) key to set the vertical position of channel to zero.

6.5.2 Setting the Horizontal System and Trigger Position

OPTION key is a multiple setting key to set trigger vertical position, main time base and horizontal position (trigger horizontal position) during edge trigger and video trigger.

The following example shows how to use OPTION key to make a setting.

The following guidance is for the operation on the mode of edge triggering and video triggering.

  1. Press once the OPTION key; the following is displayed at the bottom left side of the screen, as shown in the figure below.

◀/▶ - Time Base

▲/▼ - Trig

See the following figure 7:

Owon HDS4202M-N - The following guidance is for the operation on the mode of edge triggering and video triggering. - 1

line | Time (μs) | Value | |-----------|-----------| | 0 | 1.00001 | | 500 | 1.00001 | | 1000 | 1.00001 | | 1500 | 1.00001 | | 2000 | 1.00001 |

Figure 7

  1. Press ◀ (OPTION) or ▶ (OPTION) to adjust the main time base; press ▲ (OPTION) or ▼ (OPTION) to adjust trigger level position, press R key to "zero" the trigger level position.

  2. Press OPTION again and left bottom side display as

◀/▶ - Time

▲/▼ - Trig

See the following figure 8:

Owon HDS4202M-N - The following guidance is for the operation on the mode of edge triggering and video triggering. - 2

line | Time (μs) | Voltage (Vp) | | --------- | ------------ | | 0 | 1.00001 | | 500 | 1.00001 | | 1000 | 1.00001 | | 1500 | 1.00001 | | 2000 | 1.00001 |

Figure 8

  1. Press ◀(OPTION) or ▶(OPTION) to adjust time base horizontal position, press V key to "zero" the horizontal position; press ▲(OPTION) or ▼(OPTION) to adjust trigger level position, press R key to "zero" the trigger level position.

  2. Press OPTION again to back up operation 1.

Term interpretation

  • Vertical scale factor: It stands for the voltage amplitude represented by a division in the vertical direction of the display area, through the adjustment of which you can amplify or attenuate the signal and thus regulate the signal amplitude into the expected measurement range.
  • Vertical zero position: It is referred to as the grounding datum point, through the adjustment of which you can regulate the display position of the waveform on the screen.
  • Main time base: It means the time values represented by a division in the horizontal direction of the display area.
  • Trigger horizontal position: It means the time deviation between the actual trigger point and the screen central line, which will be displayed as 0 at the center point of the screen.
  • Trigger level position: It represents the voltage deviation between the actual trigger level and the zero position of the triggering signal source channel.

6.6 Recall factory settings

If you want to reset the Oscilloscope to the factory settings, do the following:

  1. Press MENU key and the function menu appears on the right side of the screen
  2. Press the MENU ▲ or MENU ▼ key to select FUNCTION setting and three options are visible at the bottom of the screen.

  3. Press F1 key to select "Recall Factory" to recall the factory settings.

  4. Press F2 key to select "Auto calibration". If the ambient temperature variation is up to or larger than 5 Celsius degree, the Auto calibration function should be performed. But this will not affect the using and capability by the heat from LCD and electronic component

See the following Figure 9:

F ? Vp800.0mv M 0.000ms AUTO TRIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 20v- M 500ms 0.00ms Recall Factory Auto ca libration English Language

Figure 9: Reset the Oscilloscope

6.7 Input Connections

See Figure 2 on P8.

Look at the bottom and the right of the Oscilloscope. The Oscilloscope has six signal inputs: one safety BNC jack inputs (CH1 and CH2) for scope measurements, four safety 4-mm banana jack inputs for Multimeter R, V and A measurements.

Isolated input allows independent floating measurements between Multimeters and Scopes.

6.8 Displaying an Unknown Signal with Auto Set

The Auto-Set feature lets the Oscilloscope display and measure unknown signals automatically. This function optimizes the position, range, time base and triggering, assures a stable display of virtually any waveform. This feature is especially useful for quickly checking several signals.

To enable the Auto-Set feature, do the following:

  1. Connect the test probe to the tested signals.

  2. Press the AUTO SET key and the Oscilloscope is under the automatic measurement condition. The tested signals appear on the screen.

6.9 Automatic Zero-returning of Trigger Horizontal Position and Trigger Level Position

When we adjust the trigger horizontal position and trigger level position to be maximal to make it off the screen center remotely, then we perform the following steps to make trigger horizontal position and trigger level position return to zero automatically.

  1. Press V key, the trigger horizontal position automatically returns to zero.
  2. Press R key, the trigger level position automatically returns to zero.

6.10 Automatic Measurements

The Oscilloscope offers 18 ranges of automatic scope measurements. Your can display two numeric readings: measurement 1 and measurement 2. These readings are selectable independently, and the measurements can be done on the input waveform.

To choose a frequency measurement, do the following:

  1. Press MENU key and the function menu appears on the right side of the screen.
  2. Press MENU ▲ or MENU ▼ key to select MEAS SET1. Four items selectable are visible at the bottom of the screen.
  3. Press F1 key to select Freq measurement. The measurement 1 window shows the frequency for input waveform.

To choose a Peak-Peak measurement, do the following:

  1. Press MENU key and the function menu are displayed on the right side of the screen.
  2. Press MENU ▲ or MENU ▼ key and select MEAS SET2, with 4 items selectable displayed at the bottom of the screen.
  3. Press F4 key to select PK-PK measurement. The measurement 2 window shows the peak-peak value for input waveform.

See the following figure 10:

F 1.000KHz Vp122.0v N -240.0us Trig'd TRIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 50v- N 500us -4.00V Freq Period Mean PR-PK 1/5

Figure 10: Automatic Scope Measurements

6.11 Freezing the Screen

You can freeze the screen (all readings and waveforms)

  1. Press the RUN/STOP key to freeze the screen and STOP appears at top right side of the screen.
  2. Press the RUN/STOP key once more to resume your measurement.

See the following figure 11:

Owon HDS4202M-N - Freezing the Screen - 1

line | Time (ns) | Amplitude (Hz) | | --------- | -------------- | | 0.00 | 1.00001 |

Figure 11: Freezing the Screen

6.12 Using Average for Smoothing Waveforms

Using the Average acquisition mode, you can smooth out the displayed waveform by averaging multiple data samples. The number of averaging is selectable from 4, 16, 64, and 128.

Note: In order for average mode to work in the best way, the waveform must be repetitive; As the

number of averaging increases, the slower the waveform update becomes.

  1. Press the MENU key and the function menu appears on the right side of the screen.
  2. Press MENU ▲ or MENU ▼ key to select ACQU MODE, with four items selectable displayed at the bottom of the screen.
  3. Press the F3 key to select Average, then, press F4 key to jump to Averages 64 item. This averages the outcomes of 64 acquisitions and shows the final averaging result on the screen, shown as the following figures.

See the following figure 12:

Owon HDS4202M-N - Using Average for Smoothing Waveforms - 1

line | Parameter | Value | | --------------- | --------- | | Sample | Peak Detect | | Average | 62.40V | | Averages | 62.40V | | Trig'd | 0.000ns |

Figure 12: Average Factor Sampling Mode

6.13 Using Persistence to Display Waveforms

You can use Persistence to observe dynamic signals.

  1. Press MENU key and the function menu appears on the right side of the screen.
  2. Press MENU ▲ or MENU ▼ key to select DISP SET.
  3. Press F2 key to select Persist 1 sec, 2 sec, 5 sec, Infinite or OFF. In this case, jump to Infinite and the observed dynamic is kept on the screen continuously. When the item OFF is selected, the Persistence function is closed.

Look at the display, a screen like the following figure 13 can be shown.

Owon HDS4202M-N - Using Persistence to Display Waveforms - 1

line | Time (ms) | Amplitude | | --------- | --------- | | 0 | 0 | | 546.1 | 10000 | | 122.0 | 0 | | 1.76 | 10000 | | 25.0 | 0 |

Figure 13: Persistence to Observe Dynamic Signals

6.14 Using Peak Detection to Display Glitches

You can use this function to display events (glitches or other asynchronous waveforms) of 50 ns or wider.

  1. Press MENU key and the function menu appear at the right side of the screen.
  2. Press MENU ▲ or MENU ▼ key to select the ACQU MODE. Four items selectable are displayed at the bottom of the screen.
  3. Press F2 key and jump to Peak Detect. In this case, you can test the glitch.

Now, you can see a screen that looks like the following figure 14.

V 1.970v Vp4.080v M 118.0us Trig'd TPIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET -Time Base -CH1 Volts/Div 10v- M 10us 2.36v Sample Peak Average 16 Averages the glitch

Figure 14: Peak Detection

Term interpretation

Collecting mode: The oscilloscope transforms the collected analog data into a digital form after they are gathered in the following three different modes, that is, sampling, peak value detection and averaging values.

Sampling: The oscilloscope takes samples from the signal at a equal time interval to reconstruct the waveform in this mode, by which the analog signal can be expressed correctly in most cases, yet, the rapid changes can not be collected between two sampling time intervals, causing the confusion and loss the narrow pulse in the signal probably.

Peak value detection: The oscilloscope takes samples from the maximum and minimum of signals in each sampling interval and shows the waveform with the sampled data in this mode, thus, by which the oscilloscope may collect the possibly lost narrow pulse in the sampling mode but the noise is obvious.

Averaging values: the oscilloscope collects several waveforms and average over them, and displays the averaged waveform in this mode, by which the random noise can be reduced.

Duration time: When a new waveform is displayed, the previous waveform shown on the screen does not disappear immediately only to be displayed for a period of time, that is, the duration time, by setting which, the waveform can be displayed more continuously and thus a display similar to that shown by the analog oscilloscope can be achieved.

Roll scan mode: The oscilloscope updates the waveform sampling points by scrolling display through a screen from left to right in this mode, which is only applicable to the primary time base setting of above 50ms.

6.15 Selecting AC-coupling

After a reset, the Oscilloscope is dc-coupled so that ac and dc voltages appear on the screen. Use ac-coupling when you wish to observe a small ac signal that rides on a dc signal. To select ac-coupling, do the following:

  1. Press MENU key and the function menu appear at the right side of the screen.
  2. Press MENU ▲ or MENU ▼ key to select CH1 SETUP. Four items selectable are visible at the bottom of the screen.
  3. Press the F1 key and jump to AC. The bottom left side of the screen displays the ac-coupling icon.

Now, you can see a screen that looks like the following figure 15.

F 546.0Hz Vp122.0v M -1.760ms Trig'd TRIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 50v~ M 50Vms 25.6v AC ON 10 X OFF Coupling Channel Probe Inverted

Figure 15: AC-Coupling

6.16 Reversing the Polarity of the Displayed Waveform

To invert the input waveform, do the following:

  1. Press the MENU key and the function menu appears at the right side of the screen.
  2. Press the MENU ▲ or MENU ▼ key to select CH1 SETUP. Four items selectable are displayed at the bottom of the screen.
  3. Press F4 key to jump to Inverted, set it as ON. The inverted waveform of CH1 is displayed on the screen.

Now, you can see a screen that looks like the following figure 16.

F 546.0Hz Vp122.0v N -1.760ms Trig'd TPIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 50v~ N 500ms 25.6v AC ON Coupling Channel 10 X Probe ON Inverted

Figure 16: Inverted On

6.17 Use USB mass storage device to save waveform data

Insert the USB mass storage device into the USB port and press copy key, then the current

waveform data will be saved into the USB mass. The save waveform has two format, one is vector format and the other is bitmap format, according to the display settings from the choice of the communication settings. Name the file with current date and suffix by .bin or .bmp. Then connect the USB mass storage device with the computer after data is saved, and open the data of vector format with OWON analysis software, or direct open the bitmap.

Note: It has clew in a series of process USB operation storage, such as "USB save successfully".

USB disk Requirements

The supported format of the USB disk: FAT32 file system, the allocation unit size cannot exceed 4K, mass storage USB disk is also supported. If the USB disk doesn't work properly, format it into the supported format and try again. Follow any of the following two methods to format the USB disk: using system-provided function and using the formatting tools. (The USB disk of 8 G or 8 G above can only be formatted using the second method – using the formatting tools.)

Use system-provided function to format the USB disk

  1. Connect the USB disk to the computer.
  2. Right click Computer→Manage to enter Computer Management interface.
  3. Click Disk Management menu, and information about the USB disk will display on the right side with red mark 1 and 2.

Computer Management File Action View Help Computer Management (Local System Tools Task Scheduler Event Viewer Shared Folders Local Users and Groups Performance Device Manager Storage Disk Management Services and Applications Click "Disk Management" Volume Layout Type File System Status file (G): Simple Basic NTFS Healthy (Logical Drive) KINGSTON (L): Simple Basic FAT32 Healthy (Active, Primary Partition) Life (H): Simple Basic NTFS Healthy (Logical Drive) Software (F): Simple Basic NTFS Healthy (Logical Drive) win7 (C): Simple Basic NTFS Healthy (System, Boot, Page File, Active, Crash Dump, Primary win8 (E): Simple Basic NTFS Healthy (Logical Drive) work (L): Simple Basic NTFS Healthy (Logical Drive) exp (D): Simple Basic NTFS Healthy (Logical Drive) Disk 0 Basic 461.29 GB Online win7 (C): xp (D): win8 (E): software file (G): life (H): work (L) 40.00 GB 40.01 GB 80.01 GB 80.01 GB 80.01 GB 101.22 GB Healthy ( Healthy ( Healthy ( Healthy ( Healthy ( Healthy ( L Disk 1 Removable KINGSTON (L) 7.21 GB FAT32 Healthy (Active, Primary Partition) CD-ROM 0 DVD (J.) No Media Unallocated Primary partition Extended partition Free space Logical drive

Disk Management of computer

  1. Right click 1 or 2 red mark area, choose Format. And system will pop up a warning message, click Yes.

Disk Management This is the active partition on this disk. All data on the partition will be lost. Are you sure you want to format this partition? Yes No

Format the USB disk warning

  1. Set File System as FAT32, Allocation unit size 4096. Check "Perform a quick format" to execute a quick format. Click OK, and then click Yes on the warning message.

Format L: Volume label: KINGSTON File system: FAT32 Allocation unit size: 4096 ✓ Perform a quick format □ Enable file and folder compression OK Cancel

Formatting the USB disk setting

  1. Formatting process.

Computer Management File Action View Help Volume Layout Type File System Status □ File (G) Simple Basic NTPS Healthy (Logical Drive) KINGSTON (L:) Simple Basic FAT32 Formatting : (2%) □ Make (H) Simple Basic NTPS Healthy (Logical Drive) □ software (F:) Simple Basic NTPS Healthy (Logical Drive) □ win7 (C:) Simple Basic NTPS Healthy (System, Boot, Page File, Active, Crash Dump, Primary) □ win8 (E:) Simple Basic NTPS Healthy (Logical Drive) □ work (J:) Simple Basic NTPS Healthy (Logical Drive) □ exp (D:) Simple Basic NTPS Healthy (Logical Drive) Disk 0 Basic 481.29 GB Online win7 (C): XP (D:) win8 (E): software file (G) life (H) work (L) 40.00 GB 40.01 GB 80.01 GB 80.01 GB 101.22 GB Healthy Healthy Healthy Healthy Healthy Healthy Healthy(L) Disk 1 Removable 7.21 GB Online KINGSTON (L:) 7.21 GB Formatting : (2%) CD-ROM 0 DVD (L:) No Media ■ Unallocated ■ Primary partition ■ Extended partition ■ Free space ■ Logical drive Actions Disk Management More Actions

Formatting the USB disk

  1. Check whether the USB disk is FAT32 with allocation unit size 4096 after formatting.

Use Minitool Partition Wizard to format

Download URL: http://www.partitionwizard.com/free-partition-manager.html

Tip: There are many tools for the USB disk formatting on the market, just take Minitool Partition Wizard for example here.

  1. Connect the USB disk to the computer.
  2. Open the software Minitool Partition Wizard.
  3. Click Reload Disk on the pull-down menu at the top left or press keyboard F5, and information about the USB disk will display on the right side with red mark 1 and 2.

MiniTool Partition Wizard Home Edition v0.0 General View Disk Partition Dynamic Disk Wizard Help Reload Disk FS Apply Changes Ctrl+A Undo Last Changes Ctrl+Z Discard All Changes Ctrl+O Edit Copy Partition Wizard Copy Disk Wizard Partition Recovery Wizard Basic (KINGSTONNTFS) 7.2 GB (Used IR) 1 Operations Format Partition Delete Partition Move/Passion Partition Merge Partition Split Partition Copy Partition Set Label Align Partition Hide Partition Set Inactive Change Drive Letter Change Cluster Size Change Serial Number Change Partition Type ID Surface Test Vize Partition Show Partition Properties 3 Operations Pending GPT/Primary Logical Simple Spanned Striped Removed RAID5 Unallocated Reload Disk Information Disk 1 47.03 MB 8.91 MB 38.13 MB FAT Primary Hidden C:\WIN/P 19.53 GB 14.51 GB 5.02 GB NTF5 Primary Active & Boot & Syntt D 45.02 GB 34.75 GB 14.27 GB FAT32 Logical None E 68.36 GB 39.39 GB 26.97 GB NTF5 Logical None F 79.10 GB 49.72 GB 29.30 GB FAT32 Logical None G ※线浏览 62.03 GB 4.03 GB 78.00 GB FAT32 Logical None Disk 2 (KEWSTON) 7.2 GB ICE7MB FATGB MSX Primary Action

Reload Disk

  1. Right click 1 or 2 red mark area, choose Format.

Move/Resize Extend Merge Split Copy Explore Create Delete Align Format Label Change Letter

Choose format

  1. Set File System FAT32, Cluster size 4096. Click OK.

Format Partition Formatting the partition will destroy the data on the partition! Partition Label: KINGSTON File System: FAT32 Cluster Size: 4 KB OK Cancel

Format setting

  1. Click Apply at the top left of the menu. Then click Yes on the pop-up warning to begin formatting.

MiniTool Partition Wizard Home Edition v8.0 Apply Undo Discard Move/Resize Merge

Apply Changes? Apply pending changes? NOTE: We strongly recommend you turn off Power Saving feature on your computer and close all other applications before you apply the pending changes. Yes No

Apply setting

7. Formatting process

Apply Pending Operation(s)... 1 of 1 pending operations. Format Partition Partition Letter: 1 Partition Label: NEW VOLUME Disk Index: 2 Partition Index: 1 Total 100% Format partition: 100% Format partition 100% Cancel

Format process

8. Format the USB disk successfully

Successful Apply all the pending changes successfully. OK

Format successfully

7. Using the Multimeter

7.1 About this Chapter

This chapter provides a step-by-step introduction to the multi-meter functions of the test tool hereafter. The introduction gives basic examples to show how to use the menus and perform basic operations.

7.2 Making Meter Connections

Use the four 4-mm safety banana jack inputs for the Meter functions: 10A, mA, COM, V/Ω/C.

See figure 2 on P8 for the connections.

7.3 Multimeter Operation Window

1 2 3 4 5 MANUAL DCV 12.06V RUN -2 0 2 4 V × 10 5.25V 12 7 8 Manual II / Δ Auto 11 10 9

Figure 17: Multimeter Operation Window

Description

  1. Battery electric quantity indicator.

  2. Manual/Auto range indicators, among which the MANUAL means measuring range in manual operation mode and AUTO refers to the measuring range in automatic operation mode.

  3. Measurement mode indicators: DCV:

Direct voltage measurement ACV:

Alternating voltage measurement DCA:

Direct current measurement ACA:

Alternating current measurement R:

Resistance measurement

Owon HDS4202M-N - Description - 1

: Diode measurement

Owon HDS4202M-N - Description - 2

: On/Off measurement

C: Capacitance measurement

  1. The relative magnitude measurement indicator.
  2. Running state indicators, among which RUN expresses continuous update and STOP represents the screen locking.
  3. The reference value of the relative magnitude measurement.
  4. The multiplying power of the dial indication. To multiply the reading of dial pointer by multiplying power will get the measurement result.
  5. The main reading of measurement
  6. Automatic control measuring range.
  7. Absolute/ relative magnitude measuring control: The sign "||" expresses the absolute magnitude measuring control and "△" represents the relative magnitude measuring control.

Manually measuring range control.

  1. Manual measurement control.

  2. Test lead indicated the scale of test reading; different test modes display different colors.

7.4 Making Multimeter Measurements

Press DMM/OSC key, the oscilloscope will switch to the multimeter measure; the screen will display the multimeter windows.

7.4.1 Measuring Resistance Values

To measure a resistance, do the following:

  1. Press the R key and R appears at the top of the screen.

  2. Insert the black lead into the COM banana jack input and the red lead into the V/Ω/C banana jack input.

  3. Connect the red and black test leads to the resistor. The resistor value readings are shown on the screen in Ohm.

Now, you can see a screen that looks like the following figure 18.

AUTO R RUN 4 3 2 1 0 Ω ×100 120.1Ω Manual II/Δ Auto

Figure 18: Resistance Measurement

7.4.2 Measuring Diode

To make a measurement on the diode, do the following:

Owon HDS4202M-N - Measuring Diode - 1

  1. Press the R key and R appears at the top of the screen.
  2. Press SET key until the following is displayed on the screen
  3. Insert the black lead into the COM banana jack input and the red lead into the V/Ω/C banana jack input.
  4. Connect the red and black leads to the resistor and the diode resistor readings are displayed on the screen in V.

Now, you can see a screen that looks like the following figure 19.

Figure 19: Diode Measurement

7.4.3 On-off Test

To perform an On-off test, do the following:

  1. Press the R key and R appears on the top of the screen.
  2. Press the SET key until the following is shown on the screen.
  3. Insert the black lead into the COM banana jack input and the red lead into the V/Ω/C banana jack input.
  4. Connect the red and black leads to the test point. If the resistance value of the tested point is less than 50 , you will hear beep sound from the test tool.

Now, you can see a screen that looks like the following figure 20.

Figure 20: On-off Test

7.4.4 Measuring Capacitance

To measure a capacitance, do the following:

  1. Press the R key and R appears on the top of the screen
  2. Press the SET key until C appears at the top of the screen.
  3. Insert the capacitance measurer of this multimeter to COM jack and V/Ω/C jack.
  4. Insert the capacitance to the capacitance measurer, then screen shows the capacitance reading.

Notice: when measured value is less than 5 nF capacitance, please use relative value measuring mode to improve measuring precision. It will take about 30 seconds if capacitance measurement is larger than 40 uF.

Now, you can see a screen that looks like the following figure 21.

AUTO C 10.05nF C × 10 10.07 nF Manual II/Δ Auto

Figure 21: Capacitance Measurement

7.4.5 Measuring DC Voltage

To measure a DC voltage, do the following:

  1. Press the V key and DCV appears at the top of the screen.
  2. Insert the black lead into the COM banana jack input and the red lead into the V/Ω/C banana jack input.
  3. Connect the red and black leads to the measured point and the measured point voltage value is displayed on the screen.

Now, you can see a screen that looks like the following figure 22.

DCV AUTO RUN -4 -2 0 2 4 V ×10 17.26 V Manual II / Δ Auto

Figure 22: DC Voltage Measurement

7.4.6 Measuring AC Voltage

To measure the AC voltage, do the following:

  1. Press the V key and DCV appears at the top of the screen.
  2. Press the SET key and ACV appears at the top of the screen.
  3. Insert the black lead into the COM banana jack input and the red lead into the V/Ω/C banana jack input.
  4. Connect the red and black leads to the measured points and the AC voltage values of measured points will be displayed on the screen.

Look at the display; you can see a screen that looks like the following figure 23.

AUTO ACV RUN 0 1 2 3 4 V ×10 9.44V Manual II/Δ Auto

Figure 23: AC Voltage Measurement

7.4.7 Measuring DC Current

To measure a DC current which is less than 400 mA, do the following:

  1. Press the A key and DCA appears at the top of the screen. The unit on the main reading screen is mA. mA and 10A will display on the right bottom of screen, press F4 and F5 to switch the measurement between mA and 10A. 400mA is acquiescently.
  2. Insert the black lead into the COM banana jack input and the red lead into the mA banana jack input.
  3. Connect the red and black leads to the measured points and the DC current values of measured points will be displayed on the screen.

Look at the display; you can see a screen that looks like the following figure 24.

DCA 0 -2 -4 mA x 10 2.37mA Manual II/Δ Auto mA 10A

Figure 24: DC Current Measurement for 400 mA

To measure a DC current which is larger than 400 mA, do the following:

  1. Press the A key and DCA appears at the top of the screen. The unit on the main reading screen is mA.
  2. Press F5 key change to 10A measurement, the unit on the main reading screen is A.
  3. Insert the black lead into the COM banana jack input and the red lead into the 10A banana jack input.
  4. Connect the red and black leads to the measured point and the DC current value of the measured point will be displayed on the screen.
  5. Press F4 return to 400 mA measure.

Look at the display; you can see a screen that looks like the following figure 25.

DCA -1 -0.5 0 0.5 -1 10A ×10 -0.40A Manual II / Δ Auto mA 10A

Figure 25: DC Current Measurement for 10A

7.4.8 Measuring AC Current

To measure an AC current which is less than 400 mA, do the following:

  1. Press the A key and DCA appears at the top of the screen. The unit on the main reading screen is mA. mA and 10A will display on the right bottom of screen, press F4 and F5 to switch the measurement between mA and 10A. 400mA is acquiescently.
  2. Press the SET key once and ACA is visible at the top of the screen.
  3. Insert the black lead into the COM banana jack input and the red lead into the mA banana jack input.
  4. Connect the red and black leads to the measured point and the AC current value of the measured point will be displayed on the screen.

Look at the display; you can see a screen that looks like the following figure 26.

AUTO ACA RUN 0 1 2 3 4 mA x 10 2.75mA Manual II/Δ Auto mA 10A

Figure 26: AC Current Measurement for 400 mA

To measure an AC current which is larger than 400 mA, do the following:

  1. Press the A key, an instruction prompt you to plug the multimeter test leads in the right position. After check, press any key to enter into multimeter measure and DCA appears at the top of the screen. The unit on the main reading screen is mA.
  2. Press the SET key once and ACA is visible at the top of the screen.
  3. Press F5 to select 10A measure, the unit of main reading window is A.
  4. Insert the black lead into the COM banana jack input and the red lead into the 10A banana jack input.
  5. Connect the red and black leads to the measured point and the AC current value of the measured point will be displayed on the screen.
  6. Press F4 return to 400mA measure.

Look at the display, you can see a screen that looks like the following figure 27.

MANUAL ACA RUN 0 0.25 0.5 0.75 1 10A x 10 4.89A Manual II / Δ Auto mA 10A

Figure 27: AC Current Measurement for 10A

7.5 Freezing the Readings

You can freeze the displayed readings at any time.

  1. Press the RUN /STOP key to freeze the screen and STOP will be displayed at the top right of the screen.
  2. Again, press the RUN /STOP key, you can resume your measurement.

Look at the display; you can see a screen that looks like the following figure 28.

AUTO DCV STOP -4 -2 0 2 4 V × 10 12.01 V Manual II / Δ Auto

Figure 28: Freezing the Readings

7.6 Taking a Relative Measurement

A currently measured result relative to the defined reference value is displayed in a relative measurement.

The following example shows how to take a relative measurement. At first, it is required to acquire a reference value.

  1. Press R key and R is displayed on the top side of the screen.
  2. Press the SET key until C appears at the top of the screen.
  3. Plug capacitance extended module in capacitance measure jack.
  4. When the reading leveling off, press F2 key and is displayed on the top side of the screen. The saved reference value is displayed below .
  5. Plug capacitance, the displayed major reading on the screen is actual the capacitance value.
    Look at the display; you can see a screen that looks like the following figure 29.

AUTO C 10.05nF 2 3 4 5 0 C × 10 10.07 nF Manual II / Δ Auto

Figure 29: Relative Measurement

7.7 Selecting Automatic/ManualRange Adjustment

The defaulted range mode of the instrument is automatic range. To switch to the manual range, perform the following steps:

  1. Press F1 key and MANUAL is displayed on the top left side of the screen to enter the manual range mode.
  2. Under the manual range mode, the measuring range is increased by a stage when pressing F1 key each time, and when reaching the highest stage, it jumps to the lowest stage by pressing F1 key once again. To multiply the reading of dial pointer by multiplying power and the unit of main reading on the screen will get the measurement result.
  3. Press F3 key and AUTO is displayed on the top left side of the screen to switch back to the automatic range mode.

Look at the display; you can see a screen that looks like the following figure 30.

MANUAL DCV -2 0 2 4 V x 10 12.01 V Manual II/Δ Auto

Figure 30: Automatic/Manual Range Adjustment

Attention: capacitance measurement without manual range mode.

8. Advanced Function of Oscilloscope

8.1 About this Chapter

This chapter will detail the oscilloscope function of the test tool.

8.2 Setting the Vertical Channel

Each channel has its own independent vertical menu and each item can be set respectively based on the specific channel.

To make vertical Channel settings, do the following:

  1. Press the MENU key and the function menu appears at the right of the screen.
  2. Press the MENU ▲ or MENU ▼ key to jump to CH SETUP and 4 options appears at the bottom of the screen.
  3. Select key from F1 to F4 to make different settings.

Now, you can find a screen that looks like the following figure 31.

F 546.0Hz Vp124.0v H -60.00us Trig'd TPIC MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 50v~ H 500us 24.0v AC ON Coupling Channel 10 X Probe ON Inverted

Figure 31: Setting the Vertical

The following Table describes the Vertical Channel menu:

Function menuSettingDescription
CouplingACThe dc component in the input signal is blocked.
DCThe ac and dc components of the input signal are allowed.
GroundInput signal is interrupted.
ChannelOFFClose the channel.
ONOpen the channel.
Probe1 XSelect one according to the probe attenuation level to ensure a correct vertical scale reading.
10 X
100 X
1000 X
InvertedOFFWaveform is displayed normally.
ONOpen the Invert function of the waveform setting.

8.2.1 Setting the Channel Coupling

Take CH1 for example:

Press F1 key, set Coupling as AC to make an AC coupling setting. The DC component contained in the tested signal is blocked.

Press F1 key, set Coupling as DC to make a DC coupling setting. Both DC and AC components contained in the tested signal are permitted.

Press F1 key, set Coupling as Ground to make a Ground coupling setting. Input signal is interrupted.

The waveform is displayed as the following figure 32, figure 33, and figure 34.

F 546.0Hz Vp124.0v H -60.00us Trig'd TRIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 50v~ M 500us 24.0v AC ON Coupling Channel 10 X Probe ON Inverted

Figure 32: AC Coupling

F 546.0Hz Vp122.0v M -60.00us Trig'd TRIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 50v- M 500us 24.0v DC ON Channel 10 X Probe ON Inverted

Figure 33: DC Coupling

F ? Vp0.000mv M -60.00us Trig'd TRIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 50v÷ M 500us 24.00 Ground Coupling ON Channel 10 X Probe ON Inverted

Figure 34: Ground Coupling

8.2.2 Open and Close Settings on Channel

Press F2 Channel key first, and then OFF to make a Close setting on channle.

Press F2 Channel key first, and then ON to make an Open setting on chandle.

8.2.3 Setting the probe attenuation

To prevent excessive input voltage, we recommend you to set the probe attenuation level to the 10X position to prevent excessive voltage.

After attenuating the probe level by 10:1, you also need to magnify the display level by 10 X to match the displayed amplitude with the real amplitude.

Press F3 Probe to adjust the probe attenuation level.

Table: Probe attenuation level and the corresponding menu setting

Probe attenuation levelCorresponding Menu Setting
1:11 X
10:110 X
100:1100 X
1000:11000 X

8.2.4 Setting of Inverted Waveform

Inverted waveform: The displayed signal reverses 180 degrees relatively to the ground potential.

Press F4 to set Inverted as ON to start Invert; as OFF to close Invert.

8.3 Setting the Trigger System

The Trigger defines the time when the acquisition of data and display of waveform. If it is set correctly, the trigger can turn an unstable display into a significant waveform.

When starting the acquisition of data, the oscilloscope collects sufficient data to draw the waveform at the left side of the triggering point. With waiting for the triggering condition, the oscilloscope is gathering data continuously. After a trigger is detected, the oscilloscope gathers enough data continuously to draw the waveform at the right side of the triggering point.

To make a trigger mode setting, do the following:

  1. Press the MENU key and the function menu appears at the right of the screen.
  2. Press the MENU ▲ or MENU ▼ key to select TRIG MODE and five options are displayed at the bottom of the screen.
  3. Select from F1 to F5 key to make a different setting.
  4. Press OPTION and left bottom display as below during edge trigger and video trigger:

Owon HDS4202M-N - Setting the Trigger System - 1

Owon HDS4202M-N - Setting the Trigger System - 2

Owon HDS4202M-N - Setting the Trigger System - 3

Owon HDS4202M-N - Setting the Trigger System - 4

  1. Press ▲ (OPTION) or ▼ (OPTION) to adjust trigger vertical position, Press ◀ (OPTION) or ▶ (OPTION) to adjust time base horizontal position or adjust horizontal position.

8.4 Triggering Control

There are three trigger modes including Edge trigger and Video trigger. Each trigger mode is set by different function menu.

Edge trigger: The edge trigger type triggers on the incoming signal edge. Use the edge trigger for all signals except for video related ones.

Video trigger: Perform video field trigger or line trigger on the standard video signals.

The following describes Edge trigger and Video trigger menus respectively.

8.4.1 Edge Trigger

The Edge trigger is a mode by which trigger occurs at the triggering threshold value of the input signal edge. With the Edge trigger selected, the trigger happens on the rise or fall edge of the input signal, shown as the figure 35.

F 546.0Hz Vp122.0v M -60.00us Trig d TRIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 50v- M 500us 24 0v Edge Rising DC Auto Trig mode Slope Coupling Mode Holdoff

The Edge trigger menu is described in the following table.

Function menuSettingsDescription
SlopeRisingTriggering on the rise edge of the signal.
FallingTriggering on the fall edge of the signal.
Trig modeAutoAcquisition of waveforms is possible even if there is no triggering condition detected.
NormalAcquisition of waveforms can only be done when the triggering condition is satisfied.
SingleSampling on a waveform when one trigger is detected, and then stop sampling.
CouplingACWith this mode selected, the DC component is prevented from passing-though.
DCAll components are allowed.
HoldoffTo go to holdoff menu

8.4.2 Video Trigger

The video trigger type is designed to capture the video signal format, NTSC, PAL or SECAM. For any other signal type, use the edge trigger.

Figure 36 is the Video Odd Field Trigger. Figure 37 and figure 38 is the Video Line trigger.

Owon HDS4202M-N - Video Trigger - 1

line | Time (ms) | Voltage (Volts/Div) | | --------- | ------------------- | | 0 | 0 | | 1 | 1 | | 2 | 0 | | 3 | 1 | | 4 | 0 | | 5 | 1 | | 6 | 0 | | 7 | 1 | | 8 | 0 | | 9 | 1 | | 10 | 0 | | 11 | 1 | | 12 | 0 | | 13 | 1 | | 14 | 0 | | 15 | 1 | | 16 | 0 | | 17 | 1 | | 18 | 0 | | 19 | 1 | | 20 | 0 | | 21 | 1 | | 22 | 0 | | 23 | 1 | | 24 | 0 | | 25 | 1 | | 26 | 0 | | 27 | 1 | | 28 | 0 | | 29 | 1 | | 30 | 0 | | 31 | 1 | | 32 | 0 | | 33 | 1 | | 34 | 0 | | 35 | 1 | | 36 | 0 | | 37 | 1 | | 38 | 0 | | 39 | 1 | | 40 | 0 | | 41 | 1 | | 42 | 0 | | 43 | 1 | | 44 | 0 | | 45 | 1 | | 46 | 0 | | 47 | 1 | | 48 | 0 | | 49 | 1 | | 50 | 0 | | 51 | 1 | | 52 | 0 | | 53 | 1 | | 54 | 0 | | 55 | 1 | | 56 | 0 | | 57 | 1 | | 58 | 0 | | 59 | 1 | | 60 | 0 | | 61 | 1 | | 62 | 0 | | 63 | 1 | | 64 | 0 | | 65 | 1 | | 66 | 0 | | 67 | 1 | | 68 | 0 | | 69 | 1 | | 70 | 0 | | 71 | 1 | | 72 | 0 | | 73 | 1 | | 74 | 0 | | 75 | 1 | | 76 | 0 | | 77 | 1 | | 78 | 0 | | 79 | 1 | | 80 | 0 | | 81 | 1 | | 82 | 0 | | 83 | 1 | | 84 | 0 | | 85 | 1 | | 86 | 0 | | 87 | 1 | | 88 | 0 | | 89 | 1 | | 90 | 0 | | 91 | 1 | | 92 | 0 | | 93 | 1 | | 94 | 0 | | 95 | 1 | | 96 | 0 | | 97 | 1 | | 98 | 0 | | 99 | 1 | | Note: The actual values in the CSV data are not provided in the code. The code does not output any data from the image. The values are estimated based on the given code format. The numbers in the table above are the corresponding values in the 'Video' tab. There is no label for the data series.

Figure 36: Video Odd Field Trigger
F 25.91KHz Vp7.000v N 0.000ns Trig d TPIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET So- N Sous Video Line NUM NTSC Trig mode Sync MODE Holdoff 1/2

F 25.91KHz Vp7.200v N 0.000ns Trig'd TPIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 5v- N Steps Increase Line 2 Line No. --> 2/2

Figure 37: Video Designed Line trigger (Page1) Figure 38: Video Designed Line trigger (Page2)

The Video trigger menu is described in the following table (First page):

Function menuSettingsDescription
SyncLineMake a video line trigger synchronization setting.
Field OddMake a video field trigger synchronization setting.
Field EvenMake a video odd field trigger synchronization setting.
Field LineMake a video even field trigger synchronization setting.
NUMMake a video designed line synchronization setting.
MODUNTSCPAL/SECAMVideo format setting
HoldoffTo go to holdoff menu
--->1/2Back to previous menu (When Sync selected as Line NUM)
  1. when the sync is Line NUM, the second page menu is shown as bellow.
Line Increase Set the line value to increase
DecreaseSet the line value to decrease
Line No.Set and Show the line valve
2/2Back to previous menu

When you go to the holdoff menu, you can see a screen in the following figure 39.

F 29.62KHz Vp6.720v? M 0.000ns Trig'd TPIG MODE CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 2v- M 50ms 100ns Off Time Increase Reset Back Time Off Time Off Time

Figure 39: Trigger Holdoff

the Holdoff menu is described in the following table:

Function menuSettingsDescription
Off TimeSet time slot before another trigger event.
TimeIncreaseSet the off time to increase.
DecreaseSet the off time to decrease.
ResetOff TimeReset Holdoff time to100ns.
BackBack to previous menu.

Note:

Trigger Holdoff can stabilize complex waveform, such as the pulse range. Holdoff time is the oscilloscope's waiting period before starting a new trigger. During Holdoff, oscilloscope will not trigger until Holdoff ends.

Term interpretation

Holdoff: Set time slot before another trigger event.

Trigger modes: There are three kinds of trigger modes available for this oscilloscope, they are, Auto (acquires signal continuously), Normal (acquires signal when trigger conditions are met) and Single (manually triggers the signal).

Automatic trigger mode: The oscilloscope can acquire the waveform without any triggering condition detected in this mode, in which it will be triggered compulsively when waiting for a specified period of time without any triggering condition ignited. When an invalid trigger is enforced, the oscilloscope can not keep the waveform in phase.

Normal trigger mode: In this mode, the oscilloscope cannot acquire the waveform until it is

triggered. When there is not any trigger, the oscilloscope will display the original waveform without new waveforms captured.

Single mode: In this mode, the oscilloscope will detect a trigger and capture a waveform at each time when the customer presses the RUN/STOP key.

8.5 Setting the Acquiring Mode

The ACQU MODE menu is described in the list shown as below.

Function menuSettingsDescription
SampleThe waveform data is sampled at an equal time interval. The sample mode accurately reconstructs the waveform, but cannot respond to rapid changes and sudden peaks.
Peak DetectThe maximum and minimum data in the sampling interval are picked up. The peak detect mode captures rapid changes and sudden peaks, but the waveform becomes noisy.
AverageMultiple samples are averaged together. The average mode reduces the noise level, but the waveform must be repetitive.
Averages4, 16, 64 or 128Select the average number.

8.6 Display Setting

The DISP SET menu is described in the following table.

Function menuSettingsDescription
TypeVectorsThe vector drawing mode shows the waveform as a smooth line, connecting each data point.
DotsThe dot drawing mode shows the waveform as a collection of independent data points.
PersistOFFThe persistence setting sets how long the old waveforms remain in the display, useful for observing the waveform variations.
1sec
2sec
5sec
Infinite
CarryBitmapThe data transmitted in communication are bitmaps.
VectorsThe data transmitted in communication are vectors.
CymometerONto set up cymometer limit to "ON" status.
OFFto set up cymometer limit to "OFF" status.

8.6.1 Display Style

The display style includes Vectors and Dots displays, shown as the following figure 40, figure 41.

Owon HDS4202M-N - Display Style - 1

line | Time (ns) | Voltage (V) | | --------- | ----------- | | 0 | 546.0 | | 500 | 122.0 | | 1000 | 4.00 | | 1500 | 4.00 | | 2000 | 4.00 | | 2500 | 4.00 | | 3000 | 4.00 | | 3500 | 4.00 | | 4000 | 4.00 | | 4500 | 4.00 | | 5000 | 4.00 | | 5500 | 4.00 | | 6000 | 4.00 | | 6500 | 4.00 | | 7000 | 4.00 | | 7500 | 4.00 | | 8000 | 4.00 | | 8500 | 4.00 | | 9000 | 4.00 | | 9500 | 4.00 | | 10000 | 4.00 | | 10500 | 4.00 | | 11000 | 4.00 | | 11500 | 4.00 | | 12000 | 4.00 | | 12500 | 4.00 | | 13000 | 4.00 | | 13500 | 4.00 | | 14000 | 4.00 | | 14500 | 4.00 | | 15000 | 4.00 | | 15500 | 4.00 | | 16000 | 4.00 | | 16500 | 4.00 | | 17000 | 4.00 | | 17500 | 4.00 | | 18000 | 4.00 | | 18500 | 4.00 | | 19000 | 4.00 | | 19500 | 4.00 | | 20000 | 4.00 | | 20500 | 4.00 | | 21000 | 4.00 | | 21500 | 4.00 | | 22000 | 4.00 | | 22500 | 4.00 | | 23000 | 4.00 | | 23500 | 4.00 | | 24000 | 4.00 | | 24500 | 4.00 | | 25000 | 4.00 | | 25500 | 4.00 | | 26000 | 4.0 | | 26500 | 4.3 | | 27522 | - | | 28566 | - | | 39777 | - | | 49777 | - | | 59777 | - | | 69777 | - | | 79777 | - | | 89777 | - | | 99777 | - | | 199888 | - | | 299888 | - | | 399888 | - | | 499888 | - | | 599888 | - | | Note: The frequency values are not explicitly labeled in the code provided in the code above the grid lines, so they are calculated based on the given code number (F). The frequency values are estimated based on the given code number (Vp122.ov). The frequency values are estimated based on the given code number (Trig'd). There is no label for the data series.

Figure 40: Dots Style

F 545.9Hz Vp122.0v M 4.000us Trig'd TRIG MODE CH SETUP MEAS SET1 MEAS SET2 CURS MEAS ACQU MODE TIME MODE FUNCTION DISP SET 50v- OFF Persist Bitmap Carry ON cymometer 4.000u Vectors Type

Figure 41: Vectors Style

8.6.2 Persistence

With Persist function selected, the displayed saved original data gradually decay in color and the new data are bright in color; with infinite persistence mode selected, the oscilloscope keeps all past traces of the displayed waveform.

8.6.3 Cymometer

It is a 6 digits cymometer. Its measurement range of frequency is 2Hz to full bandwidth.

Set up cymometer limit to "ON" status.

To make a cymometer setting, do the following:

If you want to measure frequency of the Channel, you can do as the follows:

  1. Press MENU, the function menu will appear on the right of the screen.
  2. Press MENU ▲ or MENU ▼ and choose DISP SET, five options will show at the bottom

of the screen.

3. Press F4 and choose ON.

See the following figure 42:

Owon HDS4202M-N - Press F4 and choose ON. - 1

line | Time (ns) | Voltage (Vp) | | --------- | ------------ | | 0 | 122.0 | | 546.0 | 122.0 | | 100 | 4.000 | | 150 | 4.000 | | 200 | 4.000 | | 250 | 4.000 | | 300 | 4.000 | | 350 | 4.000 | | 400 | 4.000 | | 450 | 4.000 | | 500 | 4.000 | | 550 | 4.000 | | 600 | 4.000 | | 650 | 4.000 | | 700 | 4.000 | | 750 | 4.000 | | 800 | 4.000 | | 850 | 4.000 | | 900 | 4.000 | | 950 | 4.000 | | 1000 | 4.000 | | 1050 | 4.000 | | 1100 | 4.000 | | 1150 | 4.000 | | 1200 | 4.000 | | 1250 | 4.000 | | 1300 | 4.000 | | 1350 | 4.000 | | 1400 | 4.000 | | 1450 | 4.000 | | 1500 | 4.000 | | 1550 | 4.000 | | 1600 | 4.000 | | 1650 | 4.000 | | 1700 | 4.000 | | 1750 | 4.000 | | 1800 | 4.000 | | 1850 | 4.000 | | 1900 | 4.000 | | 1950 | 4.000 | | 2000 | 4.000 | | 2050 | 4.000 | | 2100 | 4.000 | | 2150 | 4.000 | | 2200 | 4.000 | | 2250 | 4.000 | | 2300 | 4.000 | | 2350 | 4.000 | | 2400 | 4.000 | | 2450 | 4.000 | | 2500 | 4.000 | | 2550 | 4.000 | | 2600 | 4.000 | | 2650 | 4.000 | | 2700 | 4.000 | | 2750 | 4.000 | | 2800 | 4.000 | | 2850 | 4.000 | | 2900 | 4.000 | | 2950 | 4.0 | | 365 | - | | - | - | | - | - | | - | - | | - | - | | - | - | | - | - | | - | - | | - | - | | - | - | | - | - | | - | - | | - | - | | - | - | | - | - | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ | | - | ~ |

Figure 42 Cymometer

8.7 Waveform Saving Setups

The oscilloscope can save 4 waveforms, which can be displayed on the screen with the present waveform. The recalled waveform saved in the memory cannot be adjusted.

8.7.1 Waveform Saving Setups in normal mode

The WAVE SAVE menu is described in the following list.

Function menuSetupsDescription
WAVEA, B, C and DSelect the address for saving or recalling a waveform.
SaveStore the waveform of a selected signal source into the selected address.
ShowONOFFClose or start displaying the waveforms stored in address A, B, C or D.

To save a waveform in address A, do the following:

  1. Press the MENU key and the function menu appears at the right of the screen.
  2. Press the MENU ▲ or MENU ▼ key to select the WAVE SAVE. Three options are displayed at the bottom of the screen.
  3. Press the F1 key to select the address A.
  4. Press the F2 key to save the waveform in address A.

To display the saved waveform on the screen, do the following:

  1. Press the F3 key to select ON for the address A. The waveform saved in address A will be displayed on the screen in green color.

The display color is green, and the zero point of waveform, voltage and time is purple

Now, you can see a screen that looks like the following figure 43.

F 546.0Hz Vp122.0v N 0.000ns Trig'd Mem: A 50v 500us CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET WAVE SAVE 50v- M 500us 14.0v A Wave Save ON Show

Figure 43: Waveform Saving

8.7.2 Waveform Saving Setups in FFT mode

FFT being on, the WAVE SAVE is described in the following list.

Function menuSetupsDescription
WAVEA, B, C and DSelect the address for saving or recalling a waveform.
SaveStore the waveform of a selected signal source into the selected address.
ShowONOFFClose or start displaying the waveforms stored in address A,B, C or D.

To save a waveform on CH1 in address A, do the following:

  1. Press the MENU key and the function menu appears at the right of the screen.

  2. Press the MENU ▲ or MENU ▼ key to select the WAVE SAVE. Three options are displayed at the bottom of the screen.

  3. Press the F1 key to select the address A.

  4. Press the F2 key to save the waveform in address A

To display the saved waveform on the screen, do the following:

  1. Press the F3 key to select ON for the address A. The waveform saved in address A will be displayed on the screen in green color and the zero point of waveform, Vamp and Freq is purple.

Now, you can see a screen that looks like the following figure44

FFT Mem: A FFT Hamming 20dB 250Hz Time Base 20dB 250Hz (5KHz/μ) Trig'd CURS MEAS ACQU MODE TIME MODE FUNCTION DISP SET WAVE SAVE AUTOSCALE PECOPD FFT MODE

Figure 44: Waveform Saving in FFT Mode

8.8 Function Setting Menu

The FUNCTION menu is described in the following list.

Function menuSettingDescription
Recall FactoryResume the instrument to its factory settings.
Auto calibrationPerform the Auto-calibration procedure.
LanguageChineseEnglishOtherSelect the display language.

Auto-calibration

The Auto-calibration function automatically configures internal parameters to maintain the accuracy. Run the Auto-calibration in the following cases:

The temperature fluctuates more than 5 degrees Celsius during operations.

Operate the oscilloscope in a new bench top or field environment.

Procedure:

  1. Press the MENU key and select the FUNCTION menu using MENU ▲ or MENU ▼ key.

  2. Press F2 (Auto calibration). A message appears on the display, asking you to remove all cables and probes from oscilloscope.

  3. After removing all cables, press F2 (Auto calibration) again. The Auto-calibration automatically starts and a message appears, showing that the calibration is ongoing.

To interrupt calibration, press any key during the calibration.

8.9 Making Automatic Measurements

The oscilloscope can perform 18 types automatic measurements such as frequency, cycle, average value, peak-to-peak value, root mean square value, Vmax, Vmin, Vtop, Vbase, Vamp, Overshoot, Preshoot, Rise Time, Fall Time, +Width, -Width, +Duty, -Duty; and gives two kinds of measurement results simultaneously on the screen.

The function menu for automatic measurements is described in the following list.

Function menuDescription
FreqMeasure the frequency of waveform.
PeriodMeasure the period of waveform.
MeanMeasure the average value of waveform.
PK-PKMeasure the peak-to-peak value of waveform.
Cyc RMSMeasure Root Mean Square (RMS) value of waveform.
VmaxMeasure the Vmax of waveform
VminMeasure the Vmin of waveform
VtopMeasure the Vtop of waveform
VbaseMeasure the Vbase of waveform
VampMeasure the Vamp of waveform
OvershootMeasure the Overshoot of waveform
PreshootMeasure the Preshoot of waveform
Rise TimeMeasure the Rise Time of waveform
Fall TimeMeasure the Fall Time of waveform
+WidthMeasure the +Width of waveform
-WidthMeasure the -Width of waveform
+DutyMeasure the +Duty of waveform
-DutyMeasure the -Duty of waveform

To measure the frequency of waveform with Measurement 1 and the peak-to-peak value of waveform with Measurement 2, do the following:

  1. Press the MENU key and the function menu is shown at the right of the screen.
  2. Press the MENU ▲ or MENU ▼ key to select MEAS SET 1. Four options appear at the bottom of the screen.
  3. Press the F1 key to select the frequency measurement. The measurement window on the screen shows the frequency of waveform.
  4. Press the MENU ▲ or MENU ▼ key to select MEAS SET 2. Five options appear at the bottom of the screen.
  5. Press the F4 key to jump to the PK-PK measurement. The measurement window on the screen shows the peak-to-peak value of waveform.

Now, you can see a screen that looks like the following figure 45.

Owon HDS4202M-N - Making Automatic Measurements - 1

line | Parameter | Value | | --------------- | --------- | | F | 546.0 Hz | | Vp | 122.0 V | | M | 0.000ns | | Wave | -4.00V | | Frequency | 50V | | Period | 1.0ms | | Mean | 1.0ms | | PR-PK | 1/5 |

Figure 45: Automatic Measurements

8.10 Setting the Cursor Measurements

This oscilloscope allows you to make manual cursor measurements on time and voltage.

8.10.1 Setting the cursor measurement in normal mode

The cursor measurement menus are listed and described in the following table.

Function menusSettingsDescription
TypeOFFVoltageTimeClose the cursor measurement.Display the voltage measurement cursor and menu.Display the time measurement cursor and menu.
DeltaDisplay the difference of measured value between the two channels
Cur1Display the relating measured value of Cursor 1
Cur2Display the relating measured value of Cursor 2

To make a voltage measurement on CH1, doing the following:

  1. Press the MENU key and the function menus are displayed at the right of the screen.

  2. Press the MENU ▲ or MENU ▼ key to select CURS MEAS. Five options are shown at the bottom of the screen.

  3. Press F1 key to select the measurement type Voltage. Two purple crossing dashed lines V1 and V2 are shown on the screen.

  4. Press OPTION and display as

Owon HDS4202M-N - Setting the cursor measurement in normal mode - 1

Owon HDS4202M-N - Setting the cursor measurement in normal mode - 2

When setting OPTION ▲ or OPTION ▼, V2 will move up and down and the relating voltage value to the zero position of Channel 1 will show at the bottom of the screen. Setting OPTION ◀ or OPTION ▶, V1 will move up and down and the relating voltage value to the zero position of Channel 1 will show at the bottom of the screen, at the same time the increment shows the absolute value of V1-V2 shown as Fig 46:

F 546.0Hz Vp122.0v M 0.000ms Trig'd CH SETUP MEAS SET1 MEAS SET2 CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET WAVE SAVE 50v- M 1.0ms -4.00v Voltage Type 122.00v Delta -66.00v Curs1 56.00v Curs2

Figure 46: Use the Cursor for a Voltage Measurement

When log out submenu, every value will show on the right bottom of the screen as figure 47:

Owon HDS4202M-N - Setting the cursor measurement in normal mode - 4

line | Parameter | Value | | --------- | --------- | | Delta | 122.00 v | | Curs1 | -66.00 v | | Curs2 | 56.00 v | | Value | 546.036 Hz |

Figure 47

To use the cursor for a time measurement on CH1, do the following:

  1. Press the MENU key and the function menus are displayed at the right of the screen.

  2. Press the MENU ▲ or MENU ▼ key to select CURS MEAS.

  3. Press the F1 key to select the measurement type as Time. Two vertical dashed lines T1 and T2 appear on the screen.

  4. Press OPTION and display as

Owon HDS4202M-N - Setting the cursor measurement in normal mode - 5

Owon HDS4202M-N - Setting the cursor measurement in normal mode - 6

Press ▲(OPTION) or ▼(OPTION) and see T1 move left and right and LCD display the time value match T1 to point position in panel; press ◀(OPTION) or ▶(OPTION) and see T2 move left and right and LCD display the time value match T2 to point position in panel.

Now, you can see a screen that looks like the following figure 48.

Owon HDS4202M-N - Setting the cursor measurement in normal mode - 7

line | Time Type | Delta (ms) | Curs1 (ms) | Curs2 (ms) | |-----------|------------|------------|------------| | F | 1.88 | 1.44 | -440.00 | | Vp | 122.0 | 1.44 | -440.00 | | M | 0.00 | 1.44 | -440.00 | | Trig'd | | | |

Figure 48: Cursor time measurement

When log out submenu, every value will show on the right bottom of the screen as figure 49:

Owon HDS4202M-N - Setting the cursor measurement in normal mode - 8

line | Parameter | Value | | --------- | --------- | | Delta | 1.88 ms | | Curs1 | 1.44 ms | | Curs2 | -440.00 us| | Time | 546.036 Hz|

Figure 49: Use the Cursor for a Time Measurement

8.10.2 Setting the cursor measurement in FFT mode

FFT being on, the cursor measurement menus are listed and described in the following table.

Function menusSettingsDescription
TypeOFFClose the cursor measurement.
RangeDisplay the vamp measurement cursor and menu.
FrequencyDisplay the freq measurement cursor and menu.
DeltaDisplay the difference of measured value between the two channels
Curs1Display the relating measured value of Cursor 1
Curs2Display the relating measured value of Cursor 2

To make a range measurement on CH1, doing the following:

  1. Press the MENU key and the function menus are displayed at the right of the screen.
  2. Press the MENU ▲ or MENU ▼ key to select CURS MEAS. Five options are shown at the bottom of the screen.
  3. Press F1 key to select the measurement type Range. Two purple crossing dashed lines V1 and V2 are shown on the screen.
  4. Press OPTION and display as

Owon HDS4202M-N - Setting the cursor measurement in FFT mode - 1

Owon HDS4202M-N - Setting the cursor measurement in FFT mode - 2

Press ▲ (OPTION) or ▼ (OPTION) to see V1 move up and down and the panel will display the voltage value match V1 and Channel 1 Zero; Press ◀ (OPTION) or ▶ (OPTION) to see V2 move up and down and panel will display the voltage value match V2 to Channel 1 Zero. LCD panel also displays the absolute value of V1-V2.

Now, you can see a screen that looks like the following figure 50.

Owon HDS4202M-N - Setting the cursor measurement in FFT mode - 3

line | Range Type | Value | | ---------- | --------- | | Delta | 62.40dB | | Ours1 | -47.20dB | | Ours2 | 15.20dB |

Figure 50: Cursor range measurement

8.11 Autoscale

The function is applied to follow-up signals automatically even if the signals change at any time. Autoscale enables the instrument to set up trigger mode, voltage division and time scale automatically according to the type, amplitude and frequency of the signals.
The menu is as follows:

Function menusSettingsDescription
AutoscaleOFFONTurn off AutoscaleTurn on Autoscale
ModeHorizontalVerticalHORI—VERTOnly adjust vertical scale according to input signal.Only adjust horizontal scale according to input signal.Adjust both vertical and horizontal scales according to input signal.
Owon HDS4202M-N - Autoscale - 1Display only one or two cycles in waveform.Display Multi-cycles in waveform.

If you want to measure voltage of Channel 1, you can do as the follows:

  1. Press MENU, the function menu will appear on the right of the screen.
  2. Press MENU ▲ or MENU ▼ and choose AUTOSCALE, three options will show at the bottom of the screen.
  3. Press F1 and choose ON.

  4. Press F2 and choose HORI-VERT.

  5. Press F3 and visual displays on the screen as figure 51:

Owon HDS4202M-N - Autoscale - 2

F 546.0Hz Vp122.0v N 0.000ns Trig'd CUPS MEAS ACQU MODE TIME MODE FUNCTION DISP SET WAVE SAVE AUTOSCALE RECORD FFT MODE 50v- M 1.0ms -4.00v ON Autoscale HOP-VERT Mode

Figure 51: Autoscale Horizontal- Vertical multi-period waveforms

Note: Things you should be ware when turning on Autoscale.

  1. The symbol Ⓐ is flickering on the top left corner of screen every half second.

  2. The oscilloscope is setting Triggers automatically for the incoming signal based on its best approach. Access from front panel to trigger controls is disabled. A warning message will be displayed on the bottom left corner of the screen when trying to change submenus of Trigger controls.

  3. Oscilloscope is set as DC coupling with AUTO Triggering. A warning message is displayed when trying to change these settings.

  4. The oscilloscope will turn off Autoscale if user is trying to adjust vertical position, voltage division, trigger level or time scale. To back to Autoscale, Press AUTOSET.

  5. The oscilloscope is always on main time base.

  6. The oscilloscope is turned into Peak Detection mode if it is in Average mode.

  7. When video triggering, the horizontal time scale is 50us. If one channel is showing edge signal, the other channel is showing video one, the time scale refers to 50us as video one as standard.

8.12 Record

Waveform record: The function provides you to set the time interval between the frames and record the waveform up to 1000 numbers and get the better analysis effect more than before

through Playback and Save functions.

Four record modes: Record, Play back, Storage, OFF.

Record: Record the waveforms up until the set number of frames at specified time intervals.

Record menu list:

MenuSettingInstruction
ModeRecordSelect record mode.
Play backSelect play back mode.
StorageSelect storage mode.
OFFTurn off all recorder functions.
End frame1-1000Set number of record frames.
OperatePlayPress to start recording.
StopPress to stop recording.
Interval1 ms - 1000sSet time interval between record frames.
--->1/2To next menu
RefreshedONThe waveforms are in the state of refreshing when recording.
OFFThe waveforms stop refreshing when recording.
DirectionIncreaseThe value increases from that of end frame to time interval
DecreaseThe value decreases from that of end frame to time interval
--->1/2Back to previous menu.

Note: Both the waveforms are recorded at record mode. If one channel is off when recording, the channel data is not available at Playback mode.

Play back: Play back the recorded waveforms.

Play back menu list:

MenuSettingInstruction
Play modeLoopSet repeat play mode.
SingleSet single time play mode.
OperatePlayPress to start playback.
StopPress to stop playing.
Interval1ms-20sSet up interval value between frames.

OWON HDS-N Series Single Channel Handheld DSO & DMM of Oscilloscope
8-Advanced Function

1/2To next menu
Start Frame1-1000Set start frame.
Cur frame1-1000Select current frame to be played.
End frame1-1000Set End frame.
DirectionIncreaseDecreaseThe value increases at the mode of setting start frame, cur frame and end frameThe value decreases at the mode of setting start frame, cur frame and end frame
2/2Back to previous menu (Only when Sync is set as Line NUM).

Note: the Run/Stop button can also replay or continue the waveform display.

Storage: Store recorded waveforms in non-volatile memory according to the setup frames.

Storage menu list:

MenuSettingInstruction
Start Frame1-1000Set first frame to be saved.
End frame1-1000Set last frame to be saved.
DirectionIncreaseDecreaseThe value increases from the set start frame to the end frameThe value decreases from the set start frame to the end frame
1/2To next menu
SaveSave recorded waveform to internal memory location.
LoadRecall recorded waveform from internal memory location.
2/2Back to previous menu.

Record the waveform as follows:

  1. Press the MENU key and the function menu appears at the right of the screen.
  2. Press the MENU ▲ or MENU ▼ key to select RECORD and five options are displayed at the bottom of the screen.
  3. Press the F1 key to select the Record mode.
  4. Press F2 and set the End frame at 300 (select the direction as Increase or Decrease in the second menu before step4).
  5. Press F4 and set the time Interval to 1.00ms.
  6. Press F5 to enter into the second menu.

  7. Press F1 and select OFF to stop refreshing waveforms.

  8. Press F5 to back up to the first menu.
  9. Press F3 and select Play to operate.

Shown as Fig.52 and Fig.53
Owon HDS4202M-N - Record the waveform as follows: - 1

line | Parameter | Value | | --------------- | --------- | | F 546.0Hz | 546.0Hz | | Vp122.0v | Vp122.0v | | M 0.000ns | 0.000ns | | Trig'd | (no label)| | Record: 49 | 49 | | CUPS MEAS | (no label)| | ACQU MODE | (no label)| | TIME MODE | (no label)| | FUNCTION | (no label)| | DISP SET | (no label)| | WAVE SAVE | (no label)| | AUTOSCALE | (no label)| | RECORD | (no label)| | FFT MODE | (no label)| | -4.00V | -4.00V | | Record Mode | 300 | | End frame | End frame| | Play Operate | Operate | | 1.00 ms Interval| 1.00 ms | | --> 1/2 | 1/2 |

Fig.52: the first page of the record

Owon HDS4202M-N - Record the waveform as follows: - 2

line | Time (ms) | Value | |-----------|-------| | 0 | 0 | | 546.0 | 175 | | 122.0 | 0 | | 0 | -1.2 | | 50 | 0 | | 100 | 1.2 | | 175 | 0 | | 200 | -1.2 |

Fig.53: the second page of the record

8.13 Using FFT

An FFT breaks down signals into component frequencies, which the oscilloscope uses to display a graph of the frequency domain of a signal, as opposed to the oscilloscope's standard time domain graph. You can match these frequencies with known system frequencies, such as system clocks, oscillators, or power supplies.

FFT in this oscilloscope can transform 2048 points of the time-domain signal into its frequency

components and the final frequency contains 1024 points ranging from 0Hz to Nyquist frequency.

The following table describes the FFT menu:

Function MenuSettingInstruction
FFTONTurn on FFT function
OFFTurn off FFT function
WindowRectangleSelect window for FFT.
Blackman
Hanning
Hamming
FormatdBSet Vrms as vertical scale unit
VrmsSet dBVrms as vertical scale unit
Zoom×1Set multiple ×1
×2Set multiple ×2
×5Set multiple ×5
×10Set multiple ×10

Taking the FFT operation for example, the operation steps are as follows:

  1. Press MENU key and the function menu appear on the right side of the screen.
  2. Press MENU ▲ or MENU ▼ key to select FFT MODE. Five items selectable are available at the bottom of the screen.
  3. Press F1 to set ON. The green waveform F is shown in the screen after calculation.
  4. Press F2 to switch over Window including Hamming, Rectangle, Blackman and Hanning.
  5. Press F3 to switch over Format dB and Vrms.
  6. Press F4 to zoom in or out the waveform, the magnification includes ×1 , ×2 , ×5 , ×10 .
  7. Press CHANNEL SETTING (red),

- One of the following three prompts circularly will show at the bottom-left of the screen when the format is dB.

▲/▼—CH Volts/Div

▲/▼ — FFT dB Scale

- If left bottom of the screen displays "▲/▼—FFT dB Scale", press + (red) and - (red) button to adjust dB value of each DIV, including 1dB, 2dB, 5dB, 10dB, 20dB;

- If left bottom of the screen displays "▲/▼—CH Volts/Div", press + (red) and - (red) button to adjust voltage of channel, and "2v\~" shows in left bottom of the screen;

- If left bottom of the screen displays "▲/▼—FFT Zero", press ▲ (red) and ▼ (red) button to adjust the position of the waveform along the vertical position such as "FFT 1.20 divs".

  1. Press OPTION (OPTION), the following prompts shows on the left bottom on the screen.

◀/▶— Time Base

▲/▼—Trig

Or

◀/▶— Time

▲/▼—Trig

When "◀/▶—Time" is displayed, press OPTION ◀ and OPTION ▶ menu button to adjust the position of the waveform along the horizontal position. "FFT

-2.00divs" shown on the left bottom means the section start of the waveform departures from the Origin for 2.00DIV. The shown frequency of M is the exact frequency of the cursor point in the middle of spectrum. Shown as Figure.54.

When "◀/▶—Time Base" is displayed, press OPTION ◀ and OPTION ▶ to adjust horizontal base on horizontal base, such as "250 Hz (5 kS/s)" will be shown on the bottom status bar.

Owon HDS4202M-N - Using FFT - 1

line | Time (s) | Amplitude | |----------|---------| | 250 | Peak |

Figure 54 The Shown Frequency of M

Selecting an FFT Window

■ The FFT feature provides four windows. Each is a trade-off between frequency resolution and magnitude accuracy. What you want to measure and your source signal characteristics help determine which window to use. Use the following guidelines to select the best window.

TypeDescriptionWindow
RectangleThis is the best type of window for resolving frequencies that are very close to the same value but worst for accurately measuring the amplitude of those frequencies. It is the best type for measuring the frequency spectrum of nonrepetitive signals and measuring frequency components near DC.Use rectangle for measuring transients or bursts where the signal level before and after the event are nearly equal. Also, use this window for equal-amplitude sine waves with frequencies that are very close and for broadband random noise with a relatively slow varying spectrum.Owon HDS4202M-N - Selecting an FFT Window - 1
HammingThis is a very good window for resolving frequencies that are very close to the same value with somewhat improved amplitude accuracy over the rectangle window. It has a slightly better frequency resolution than the Hanning.Use Hamming for measuring sine, periodic and narrow band random noise. This window works on transients or bursts where the signal levels before and after the event are significantly different.Owon HDS4202M-N - Selecting an FFT Window - 2
HanningThis is a very good window for measuring amplitude accuracy but less so for resolving frequencies.Use Hanning for measuring sine, periodic and narrow band random noise. This window works on transients or bursts where the signal levels before and after the event are significantly different.Owon HDS4202M-N - Selecting an FFT Window - 3
BlackmanThis is the best window for measuring the amplitude of frequencies but worst at resolving frequencies.Use Blackman-Harris for measuring predominantly single frequency waveforms to look for higher order harmonics.Owon HDS4202M-N - Selecting an FFT Window - 4

Fig55, 56, 57, 58 show four kinds of window function referring to sine wave of 1kHz .

Owon HDS4202M-N - Selecting an FFT Window - 5

line | Time (GHz) | Amplitude (1.000kHz) | |------------|----------------------| | 250 | Peak |

Fig 55: Rectangle window
Owon HDS4202M-N - Selecting an FFT Window - 6

line | Time (GHz) | Amplitude (1.000kHz) | |------------|----------------------| | 20 | ~0 | | 250 | ~1.0 |

Fig 56: Blackman window

Owon HDS4202M-N - Selecting an FFT Window - 7

line | Time (GHz) | Amplitude (1.000KHz) | |------------|----------------------| | 20 | ~0 | | 250 | ~1.0 |

Fig.57. Hanning window
Owon HDS4202M-N - Selecting an FFT Window - 8

line | Time (GHz) | Amplitude (1.000KHz) | |------------|----------------------| | 0 | 0 | | 1.0 | 0 | | 2.0 | 0 | | 3.0 | 0 | | 4.0 | 0 | | 5.0 | 0 | | 6.0 | 0 | | 7.0 | 0 | | 8.0 | 0 | | 9.0 | 0 | | 10.0 | 0 | | 11.0 | 0 | | 12.0 | 0 | | 13.0 | 0 | | 14.0 | 0 | | 15.0 | 0 | | 16.0 | 0 | | 17.0 | 0 | | 18.0 | 0 | | 19.0 | 0 | | 20.0 | 0 | | 21.0 | 0 | | 22.0 | 0 | | 23.0 | 0 | | 24.0 | 0 | | 25.0 | 0 | | 26.0 | 0 | | 27.0 | 0 | | 28.0 | 0 | | 29.0 | 0 | | 30.0 | 0 | | 31.0 | 0 | | 32.0 | 0 | | 33.0 | 0 | | 34.0 | 0 | | 35.0 | 0 | | 36.0 | 0 | | 37.0 | 0 | | 38.0 | 0 | | 39.0 | 0 | | 40.0 | 0 | | 41.0 | 0 | | 42.0 | 0 | | 43.0 | 0 | | 44.0 | 0 | | 45.0 | 0 | | 46.0 | 0 | | 47.0 | 0 | | 48.0 | 0 | | 49.0 | 0 | | 50.0 | 0 | | 51.0 | 0 | | 52.0 | 0 | | 53.0 | 0 | | 54.0 | 0 | | 55.0 | 0 | | 56.0 | 0 | | 57.0 | 0 | | 58.0 | 0 | | 59.0 | 0 | | 60.0 | 0 | | 61.0 | 0 | | 62.0 | 0 | | 63.0 | 0 | | 64.0 | 0 | | 65.0 | 0 | | 66.0 | 0 | | 67.0 | 0 | | 68.0 | 0 | | 69.0 | 0 | | 70.0 | 0 | | 71.0 | 0 | | 72.0 | 0 | | 73.0 | 0 | | 74.0 | 0 | | 75.0 | 0 | | 76.0 | 0 | | 77.0 | 0 | | 78.0 | 0 | | 79.0 | 0 | | 80.0 | 0 | | 81.0 | 0 | | 82.0 | 0 | | 83.0 | 0 | | 84.0 | 0 | | 85.0 | 0 | | 86.0 | 0 | | 87.0 | 0 | | 88.0 | 0 | | 89.0 | 0 | | 90.0 | 0 | | 91.0 | 0 | | 92.0 | 0 | | 93.0 | 0 | | 94.0 | 0 | | 95.0 | 0 | | 96.0 | 0 | | 97.0 | 0 | | 98.0 | 0 | | 99.0 | 0 | | Note: The actual values may vary due to the random nature of the data generation process.

Fig.58. Hamming window

Quick Tips

If desired, use the zoom feature to magnify the FFT waveform.
■ Use the default dB scale to see a detailed view of multiple frequencies, even if they have very different amplitudes. Use the Vrms scale to see an overall view of how all frequencies compare to each other.
■ Signals that have a DC component or offset can cause incorrect FFT waveform component magnitude values. To minimize the DC component, choose AC Coupling on the source signal.
■ To reduce random noise and aliased components in repetitive or single-shot events, set the oscilloscope acquisition mode to average.

Term interpretation

Nyquist frequency: The highest frequency that any Real Time Digital Oscilloscope can measure

is exactly half of the sampling rate under the condition of no mistakes, which is called Nyquist frequency. If under-sampling occurs when the frequency sampled is higher than Nyquist frequency, "False Wave" phenomenon will appear. So pay more attention to the relation between the frequency being sampled and measured.

NOTE:

While FFT mode is turned on, DO NOT USE THESE SETTINGS:

1) Horizontal window setting.
2) Trigger control "SET 50%"
3) Auto-scale function
4) Wave record;
5) Measure 1 and Measure 2.

8.14 System State Menu

The system state menu is used to display information about the present horizontal system, vertical system, trigger system and others. The operation steps are shown as below.

  1. Press the MENU key and the function menu is displayed at the right of the screen.
  2. Press the MENU ▲ or MENU ▼ key to select the SYS STAT. Four options appear at the bottom of the screen.
  3. Sequentially press F1 to F4 key and the corresponding state information will be shown on the screen.
  4. The screen that looks like the following figure 59 will be displayed.

F 1.000KHz Vp130.0v M -240.0us Trig'd HORIZONTAL SYSTEM STATUS TIME BASE MAIN TIME BASE MAIN SCALE 500us WINDOW SCALE 20us MAIN POSITION -240.0us WINDOW POSITION -100.00us ACQUIPE MODE SAMPLE Fourier analysis Window Hamming Format dB ACQU MODE TIME MODE FUNCTION DISP SET MAVE SAVE AUTOSCALE RECORD FFT MODE SYS STAT 50v- M 500us -4.00v Horizontal Vertical Trigger Misc

Figure 59: System State

8.14.1 Real time clock

Function: Be synchronous with the U disk storage time.

Do below steps to set up the real time clock:

  1. Press MENU key and the function menu is displayed on the right side of the screen.
  2. Press MENU ▲ or MENU ▼ key and select SYS STAT and there will be 4 items displayed at the bottom of the screen.
  3. Press F4 key and select Misc to display the system time, see Fig.60.

F 999.8Hz Vp134.0v N -240.0us Trig'd MISC Series Number HDS1021M N1505060001 V1.0.0 Checksum F61A2EF9 2009-03-25 00:44:29 ACQU MODE TIME MODE FUNCTION DISP SET MAVE SAVE AUTOSCALE PECORD FFT MODE SYS STAT 50v- M 500us -4.00v Horizontal Vertical Trigger Misc

Fig.60: System Time

  1. Press OPTION to enter into timing mode and if that the background color of the year, month, date, hour, minute, second which are in need of correct be deepened, the value can be adjusted.
  2. Press OPTION▲ or OPTION▼ to change the corresponding value of the year, month, date, hour, minute and second effectively.
  3. Press OPTION ◀ or OPTION ▶ to switch to the year, month, date, hour, minute and second that are in need of adjustment.
  4. Press OPTION to enter into normal time mode after timing.
  5. If need to timing, please kindly repeat the above steps.
    Note: The digital clock stops working when in time calculating mode.

8.15 Setting of Time Base Mode

The time base mode menu is explained as the following table.

Function menuSettingExplanation
Main TimeBaseHorizontal main time base is used to wave display
Zone WindowUse two cursors to define a window area
WindowExpand the defined window to full-screen display

For the operation of window extension, please execute the following steps:

  1. Press MENU key, display the function menu on the right side of the screen.
  2. Press MENU ▲ or MENU ▼ key to select TIME MODE, display three options at the bottom.
  3. Press F2 key to select Zone Window.
  4. Press OPTION key to pop up TIME BASE, at this time, then press ◀ (OPTION) and ▶ (OPTION) key to adjust the time base window area defined by two cursors, the window size will vary.
  5. Press OPTION key and call TIME, at this time, press ◀(OPTION) and ▶(OPTION) to adjust the window position defined by two cursors, the window position is the time difference of the window center to main time base's horizontal pointer.
  6. Press F3 key, select Window, the defined window extends into the full-screen display.

The screen that looks like the following figure 61, 62 will be displayed.

Owon HDS4202M-N - Setting of Time Base Mode - 1

line | Time (ms) | Value | |-----------|-----------| | 0 | -240.0 | | 50 | -100.0 | | 100 | -100.0 | | 150 | -100.0 | | 200 | -100.0 | | 250 | -100.0 | | 300 | -100.0 | | 350 | -100.0 | | 400 | -100.0 | | 450 | -100.0 | | 500 | -100.0 | | 550 | -100.0 | | 600 | -100.0 | | 650 | -100.0 | | 700 | -100.0 | | 750 | -100.0 | | 800 | -100.0 | | 850 | -100.0 | | 900 | -100.0 | | 950 | -100.0 | | 1000 | -100.0 | | 1050 | -100.0 | | 1100 | -100.0 | | 1150 | -100.0 | | 1200 | -100.0 | | 1250 | -100.0 | | 1300 | -100.0 | | 1350 | -100.0 | | 1400 | -100.0 | | 1450 | -100.0 | | 1500 | -100.0 | | 1550 | -100.0 | | 1600 | -100.0 | | 1650 | -100.0 | | 1700 | -100.0 | | 1750 | -100.0 | | 1800 | -100.0 | | 1850 | -100.0 | | 1900 | -100.0 | | 1950 | -100.0 | | 2000 | -100.0 | | 2050 | -100.0 | | 2100 | -100.0 | | 2150 | -100.0 | | 2200 | -100.0 | | 2250 | -100.0 | | 2300 | -100.0 | | 2350 | -100.0 | | 2400 | -100.0 | | 2450 | -100.0 | | 2500 | -100.0 | | 2550 | -100.0 | | 2600 | -100.0 | | 2650 | -100.0 | | 2700 | -100.0 | | 2750 | -100.0 | | 2800 | -100.0 | | 2850 | -100.0 | | 2900 | -100.0 | | 2950 | -10Ou | | 3D | -4.4Ou | | Full Width| Main TimeBase| | Zone Window| Zone Window| | Window | Window | | Zone window| Zone window|

Figure 61: Zone Window
Owon HDS4202M-N - Setting of Time Base Mode - 2

line | Time (ms) | Value | |-----------|-----------| | 0 | 100.0 | | 50 | -100.0 |

Figure 62: Window

8.16 Data Transmission

For the operation of data transmission, please execute the following steps:

  1. Press MENU key, display the function menu on the right side of the screen.
  2. Press MENU ▲ or MENU ▼ key to select DISP SET.
  3. Press F4 key, select the Bitmap or Vectors for data transmission.
  4. Use a data line to connect the oscilloscope and the PC.
  5. Open the OWON software that had been installed completely.
  6. Setting the parameter under the use's manual, and then starts data transmission.

9. Trouble Shooting

1. The oscilloscope does not power up

It may be caused by the dead battery. At this time, the oscilloscope will not start even if it is powered by the battery charger. Firstly, charge the battery and supply the oscilloscope with power through the battery charger. However, don't switch on the oscilloscope until waiting for 15 minutes. If the oscilloscope still cannot start, please contact OWON for service.

2. The oscilloscope stops working only after operating for several seconds

Probably the battery is dead; Check the battery symbol at the upper left of the screen. The symbol indicates that the battery has been run out of power and must be charged.

3. The measurement type displays ERR when you switch to the Multimeter mode.

Probably you do not select the measuring mode. In this case, push down any key of the three keys A, V or R. Thus, the relative measuring mode will be shown on the screen. If the ERR is still displayed, restart the oscilloscope.

4. The measured voltage amplitude value is 10 times larger or smaller than the real value under the oscilloscope mode.

Check whether the channel attenuation level matches the actual probe attenuation ratio.

5. The waveform is displayed on the screen but is not stable under the oscilloscope mode.

  • Check whether the signal item in the trigger mode menu matches the actually used signal channel.
  • Check on the trigger mode: The edge trigger mode is applicable to the universal single and the video trigger mode applicable to the video signal. Only when the proper trigger mode is applied, can the waveform be in stable.
  • Try to change the trigger coupling into the HF rejection and LF rejection to filter the HF or LF noise trigger by the interference. (Only for the models that have this function.)

6. There is no display on the screen when you push down RUN/STOP key under the oscilloscope mode.

Check whether the trigger mode in the trigger mode menu is in normal or single shot and whether the trigger level is out of the waveform range.

In such condition, adjust and make the trigger level in the middle or select the non-auto trigger mode. Additionally, press the AUTO SET key and complete the above setting

7. When select the average sampling in the sampling mode or select a longer display time in the display mode under the oscilloscope, the display speed is slow.

It is normal in the above case.

10. Appendix

10.1 Appendix A: Specifications

10.1.1 Oscilloscope

Only if other instructions are provided, are all technical specifications applicable to the probe with the 10X attenuation switch setting and the HDS-N series digital type oscilloscope. In order to be up to these specifications, the oscilloscope should meet the following requirement.

The instrument should operate continuously for more than 30 minutes under the specified operating temperature.
If the operating temperature is up to or larger than 5 Celsius degrees, the system function menu must be opened to make the system perform a "Auto- calibration" procedure (see "calibration" on P51).

Except those specifications marked with the word Typical, all specifications can be up to.

Sampling

Sampling modesNormal samplingPeak detectionAverage value
Sampling rateHDS10201-N100 MS/s
HDS2061M-NMax 500 MS/s
HDS3101M-NMax 1 GS/s

Input

Input couplingDC, AC, Ground
Input impedanceHDS1021M-N1 MΩ±2% connected in parallel with 20 pF±5 pF
HDS2061M-N1 MΩ±2% connected in parallel with 15 pF±5 pF
HDS3101M-N
Probe attenuation coefficient1X, 10X, 100X, 1000X
Max. Input voltage400 V (peak)
Channel delay time (typical)150 ps

Horizontal

Sampling rate rangeHDS1021M-N0.25 S/s - 100 MS/s
HDS2061M-N0.25 S/s - 500 MS/s
HDS3101M-N0.25 S/s - 1 GS/s
Waveform interpolation(sin x) /x
Record length6k points on each channel
Scanning speed range (S/div)HDS1021M-NHDS2061M-NHDS3101M-N5ns/div - 100s/div, stepping in the "1-2-5" mode
Sampling rate and delay time accuracy±100ppm(any time interval which is equal to or larger than 1ms)
Time interval (△T)measurement accuracy (full bandwidth)Single:±(1 sampling interval time +100ppm×reading+0.6ns)>average 16 :±(1 sampling interval time +100ppm×reading+0.4ns)

Vertical

Analog digital converter (A/D)With the resolution of 8 bits, make sampling on both channels synchronously.
Displacement rangeHDS1021M-N±2 V (5 mV/div - 200 mV/div)
HDS2061M-N±2 V (5 mV/div - 200 mV/div), ±50 V (500 mV/div - 5 V /div)
HDS3101M-N
Analog bandwidthHDS1021M-N20 MHz
HDS2061M-N60 MHz
HDS3101M-N100 MHz
Single bandwidthFull bandwidth
Low frequency response (AC coupling, -3dB)≥10 Hz (at the BNC)
Rise time (typical one at the BNC)HDS1021M-N≤17.5 ns
HDS2061M-N≤ 5.8 ns
HDS3101M-N≤ 3.5 ns
DC gain accuracy±3%
DC measurement accuracy (average value sampling mode)The voltage difference (△V) between any two points on the waveform after averaging the captured waveforms more than 16: ±(5% reading + 0.05 divisions).

Trigger

Triggering level range±6 divisions from the screen center
Triggering level accuracy (typical) which is applicable to the signal with rise and fall time equal to or longer than 20ns±0.3 divisions
Trigger displacement655 divisions for pre-triggering and 4 divisions for post-triggering.
Trigger Holdoff range100 ns – 10 s
Make a 50% level setting (Typical).Operation with the input signal frequency equal to or larger than 50Hz.
Signal system and line/field frequency (Video triggering mode)Support the NTSC, PAL and SECAM broadcasting systems of any field or line frequency.
Cymometer
Readout resolution6 digits
Frequency rangeAC coupling, 2 Hz to full bandwidth
Single source● when the triggering mode is edge triggering, it is a one channel Cymometer.● When the triggering mode is video triggering, the cymometer will not work..

Measurement

Cursor measurementVoltage difference ( V ) and time difference ( T ) between cursors
Auto measurementPeak-to-peak value, average value, root mean square value, frequency , period, Vmax, Vmin, Vtop, Vbase, Vamp, Overshoot, Preshoot, Rise Time, Fall Time, +Width, -Width, +Duty, -Duty.

Probe

1X position10X position
BandwidthUp to 6 MHz (DC)Up to full bandwidth (DC)
Attenuation rate1: 110: 1
Compensation range15 pf – 35 pf
Input resistance1 MΩ ± 2%10 MΩ ± 2%
Input Capacitance85 pf – 115 pf14.5 pf - 17.5 pf
Input voltage<200 V DC + Peak AC<600 V DC + Peak AC

10.1.2 Meter

Voltage (VDC)

Input Impedance: 10 MΩ.

Max. Input Voltage: 1000 V (DC or AC peak value)

RangeAccuracyResolution
400.0 mv±1%±2 digits100 uV
4.000 V1 mV
40.00 V10 mV
400.0 V100 mV
1000.0 V1 V

Voltage (VAC)

Input Impedance: 10 MΩ

Max Input Voltage: 750 V (AC, RMS value)

Frequency range: from 40 Hz to 400 Hz

Display: RMS value of the sine wave

RangeAccuracyResolution
4.000 V±1%±3 digits1 mV
40.00 V10 mV
400.0 V100 mV
750.0 V±1.5%±3 digit1 V

Direct Current (DC)

RangeAccuracyResolution
40.00 mA±1.5%±1 digit10 uA
400.0 mA±1.5%±1 digit100 uA
10 A±3%±3 digits10 mA

Alternating Current (AC)

RangeAccuracyResolution
40.00 mA±1.5%±3 digits10 uA
400.0 mA±2%±1 digit100 uA
10 A±5%±3digits10 mA

Resistance

RangeAccuracyResolution
400.0 Ω±1% ±3 digits0.1 Ω
4.000 kΩ±1% ±1 digit1 Ω
40.00 kΩ10 Ω
400.0 kΩ100 Ω
4.000 MΩ1 kΩ
40.00 MΩ±1.5%±3 digits10 kΩ

Capacitance

RangeAccuracyResolution
51.20 nF±3%±3 digits10 pF
512.0 nF100 pF
5.120 uF1 nF
51.20 uF10 nF
100 uF100 nF

Diode

Voltage reading: 0 V - 1.5 V.

On-off Test

There is a beep sound when the on-resistance is less than 50 .

10.1.3 General Specifications

Basic parameter

Mechanical dimension18 cm×11.5 cm×4 cm
Weight645 g
Power consumptionHDS1021M-N< 6 W
HDS2061M-N< 7 W
HDS3101M-N
Display type3.7" color liquid crystal display
Display resolution640 (horizontal) ×480 (vertical) pixels
Display color65536 colors

Power Adapter

Power supply100 - 240 V AC 50/60Hz
Voltage outputHDS1021M-N8.5 VDC
HDS2061M-NHDS3101M-N9 VDC
Current outputHDS1021M-N1500 mA
HDS2061M-NHDS3101M-N3000 mA

Battery type: 7.4V built-in Li-ion battery.

Working environment

Temperature

OperationUsed battery0 to 50 °C (32 to 122 °F)
Power adapter0 to 40 °C (32 to 104 °F)
Storage-20 to 60 °C (-4 to 140 °F)

Humidity

Operation0 to 10 °C (32 to 50 °F)no condensation
10 to 30 °C (50 to 86 °F)95 %
30 to 40 °C (86 to 104 °F)75 %
40 to 50 °C (104 to 122 °F)45 %
Storage-20 to 60 °C (-4 to 140 °F)no condensation

10.2 Appendix B: Maintenance and Cleaning

10.2.1 Maintenance

Do not store or place the instrument in locations where the liquid crystal display (LCD) may be directly exposed to the sunshine for a long time.

Be careful: Do not spray liquid into the instrument.

Cleaning

Inspect the instrument and the probe frequently in accordance with operating conditions. Clean the outer surface of the instrument according to the following steps:

  1. Wipe off the floating dust outside of the instrument and the probe by using soft cloth. When cleaning the LCD, do not scuff the transparent LCD protective screen.
  2. Wipe the instrument in power off status by soft cloth that is moist but not dripping. It is allowed to be cleaned by soft detergent or fresh water. Do not use any abrasive chemical detergent to avoid the instrument or the probe being damaged.

Owon HDS4202M-N - Cleaning - 1

Warning: Before running, please confirm that the instrument has been dried out so as to avoid electrical short circuit and personal injury caused by moisture.

10.2.2 Storage of Oscilloscope

If the test tool is to be stored for a long time, it is required to charge the lithium battery before storage.

Charging the oscilloscope

The lithium battery is possibly not charged when delivery. To make the battery with enough electric quantity, it must be charged for 4 hours (the test tool must be turned off during charging). The battery can supply power for 4 hours after being charged completely.

When supplying power by using the battery, a battery indicator is displayed on the top of the screen to show the consumption condition of electric quantity. ☐ shows that the battery can only be used for about 5 minutes. To charge the battery and power the instrument, connect the oscilloscope using a power adapter according to Figure 1 to charge the battery. The charging speed can be increased by turning off the test tool.

Notice

To avoid superheat of battery during charging, the environment temperature is not allowed to exceed the permissible value given in technical specification.

Note:

No hazard will occur even connecting the charger for a long time, e.g. during a whole weekend. The instrument can automatically switch to slowly charging status.

10.2.3 Replacing the Lithium Battery Unit

It is usually not required to replace the battery unit. But when it is required to replace it, only qualified personnel can carry out this operation, and only use the same specification lithium battery.