Sonel

MPI-520 - Measurement Sonel - Free user manual and instructions

Find the device manual for free MPI-520 Sonel in PDF.

📄 68 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice Sonel MPI-520 - page 7
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Multifunction Electrical Installation Tester
Brand Sonel
Model MPI-520
Measurement Functions Insulation resistance, loop impedance, RCD testing, earth resistance, continuity
Insulation Resistance Range Up to 1000 V, up to 2 GΩ
Loop Impedance Range 0.01 Ω - 1000 Ω
RCD Testing AC and A type, trip current and time
Earth Resistance 0.01 Ω - 10 kΩ
Continuity Testing Yes, with audible signal
Display Backlit graphic LCD
Power Supply Rechargeable battery pack or 4x AA batteries
Dimensions Approx. 220 x 100 x 60 mm
Weight Approx. 0.7 kg
Memory Internal storage for up to 990 measurements
Communication USB and Bluetooth (optional)
Safety Standards IEC 61557, CAT IV 300V
Operating Temperature -10°C to 40°C
Storage Temperature -20°C to 60°C
Ingress Protection IP20
Maintenance Clean with dry cloth; calibrate annually
Repairability Spare parts available from Sonel service centers

Frequently Asked Questions - MPI-520 Sonel

What is the MPI-520 used for?
The Sonel MPI-520 is a multifunction electrical installation tester used for measuring insulation resistance, loop impedance, RCD parameters, earth resistance, and continuity in electrical systems.
How do I power the MPI-520?
It can be powered by a rechargeable battery pack or four AA alkaline batteries. Use the provided charger for the rechargeable pack.
Can I test RCDs with the MPI-520?
Yes, it can test AC and A type RCDs, measuring trip current and trip time with selectable fault current and phase angle.
What is the maximum insulation resistance it can measure?
It measures insulation resistance up to 2 GΩ with test voltages up to 1000 V.
How do I connect the tester to a computer?
Use the USB port or optional Bluetooth adapter to transfer measurement data to a PC for analysis.
Is the MPI-520 safe to use on live circuits?
It is designed for safe use on live circuits up to CAT IV 300V. Always follow safety instructions in the manual.
How do I clean and maintain the device?
Clean with a soft, dry cloth. Avoid solvents. Have the device calibrated annually by an authorized service center.
What is the memory capacity?
The internal memory can store up to 990 measurement records, which can be viewed on the display or downloaded.
Can I replace the batteries myself?
Yes, the battery compartment is accessible. Use only AA alkaline batteries or the designated rechargeable pack.
Where can I find spare parts or repair service?
Contact Sonel authorized service centers for spare parts and repairs. The device is repairable with original components.

User questions about MPI-520 Sonel

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Measurement in PDF format for free! Find your manual MPI-520 - Sonel and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. MPI-520 by Sonel.

USER MANUAL MPI-520 Sonel

natural_image Line drawing of a medical device with control panel and buttons (no text or symbols)

USER MANUAL

METER FOR ELECTRICAL INSTALLATION PARAMETERS

MPI-520

MPI-520
Battery charger socket and USB port under sliding lid Measuring sockets Measuring clamp socket Battery charger socket USB port Initiation of measuring procedure Touch electrode Confirm selection ESC - back to the previous screen, exit the function Move / select: right / left, up / down START ENTER RCD MPI-S20 MEM DISPLAY ON/OFF button MENU - selection of additional meter settings Display backlight ON/OFF button FUNCTION SWITCH Selection of measuring functions: - AUTO - RCD: automatic test - I_A - RCD: measurement of initiation current - t_H-F - RCD: measurement of initiation period - R_E - measurement of earth resistance - R_DD - measurement of insulation resistance - U,I,P,Q S,f,cosφ - measurement of voltage, current, power, frequency, and φ - MEM - memory browsing and clearing, and data transmission - L_1 - phase sequence control - R_X R_{200mA} - measurement of protective and equipotential conductor resistance as well as low-voltage measurement of resistance - Z_L-PE RCD - measurement of fault loop impedance in L-PE circuit protected by residual current device (RCD) - Z_L-PE U_L-PE - measurement of fault loop impedance in L-PE circuit - Z_L-N,L_L U_L-N,L_L - measurement of fault loop impedance in L-N or L-L circuit

USER MANUAL

METER FOR ELECTRICAL INSTALLATION PARAMETERS MPI-520

CE

SONEL S. A.

Wokulskiego 11

58-100 Świdnica

Poland

The MPI-520 meter is a modern, easy and safe in use measuring device. Please acquaint yourself with the present manual in order to avoid measuring errors and prevent possible problems related to operation of the meter.

CONTENTS

1 Safety 5

2 Menu....6

2.1 Wireless transmission 6
2.2 Settings of measurements....7

2.2.1 Network voltage and frequency 7
2.2.2 Main result of short circuit loop impedance measurement 8
2.2.3 Measurement settings....8
2.2.4 Cell autoincrementing 8

2.3 Settings of the meter 9

2.3.1 LCD contrast....9
2.3.2 LCD backlight....10
2.3.3 Auto-OFF settings....10
2.3.4 Date and time....10
2.3.5 Factory (default) settings....11
2.3.6 Program update 11

2.4 Language selection.... 11
2.5 Information about manufacturer 11

3 Measurements....12

3.1 Measurement of alternating voltage and frequency 12
3.2 Checking correctness of PE (protective earth) connections.... 12
3.3 Measurement of current, active power, reactive power, apparent power and coefficient....13
3.4 Measurement of short circuit loop parameters 14

3.4.1 Measurement of short circuit loop parameters in L-N and L-L circuit 14
3.4.2 Measurement of short circuit loop parameters in L-PE circuit 16
3.4.3 Measurement of short circuit loop impedance in L-PE circuit protected with residual current device (RCD) 18
3.4.4 Prospective short-circuit current 19

3.5 Measurement of resistance-to-earth 20
3.6 Measurement of RCD parameters 23

3.6.1 Measurement of RCD disconnection current 23
3.6.2 Measurement of RCD disconnection time 25
3.6.3 Automatic measurement of RCD parameters 27

3.7 Measurement of insulation resistance.... 30

3.7.1 Double-lead measurement....30
3.7.2 Measurements with AutoISO-1000c adapter....33
3.7.3 Measurements by means of leads with UNI-Schuko outlet plug (WS-03 and WS-04).....34

3.8 Low-voltage measurement of resistance.... 36

3.8.1 Measurement of resistance of protective conductors and equipotential bonding with ±200 mA current....36
3.8.2 Measurement of resistance....38
3.8.3 Calibration of test leads....39

3.9 Checking sequence of phases 40

4 Memory of measurement result data....42

4.1 Recording measurement result data in the memory.... 42
4.2 Viewing memory data.... 43

4.3 Deleting memory data 46

5 Data transmission....47

5.1 Computer connection accessories 47
5.2 Data transmission with USB joint 47
5.3 Data transmission with OR-1 radio module 47

6 Power supply....49

6.1 Monitoring of the power supply voltage 49
6.2 Replacement of batteries (accumulators).... 50
6.3 Charging of accumulators 50
6.4 General principles regarding using Ni-MH accumulators 52

7 Cleaning and maintenance ....52

8 Storage....53

9 Dismantling and utilisation....53

10 Technical data....53

10.1 Basic data 53
10.2 Other technical data 60
10.3 Additional data 61

10.3.1 Additional uncertainties according to IEC 61557-2 (Rso)....61
10.3.2 Additional uncertainties according to IEC 61557-3 (Z)....61
10.3.3 Additional uncertainties according to IEC 61557-4 (R ±200 mA)....61
10.3.4 Additional uncertainties according to IEC 61557-5 ( R_E )....62
10.3.5 Additional uncertainties according to IEC 61557-6 (RCD) 63

11 Positions of the meter's cover 63

12 Manufacturer 63

1 Safety

MPI-520 meter is designed for performing check tests of protection against electric shock in mains systems. The meter is used for making the measurements the results of which determine safety of electrical installations. Therefore, in order to provide conditions for correct operation and the correctness of the obtained results, the following recommendations must be observed:

  • Before you proceed to operate the meter, acquaint yourself thoroughly with the present manual and observe the safety regulations and specifications determined by the producer.
  • The MPI-520 meter has been designed for the purpose of measurements of earth connection and equipotential bonding, ground resistivity, as well as clamps current measurements. Any application that differs from those specified in the present manual may result in a damage to the device and constitute a source of danger for the user.
  • The device must be operated solely by appropriately qualified personnel with relevant certificates to realise measurements of electric installation. Operation of the meter realised by unauthorised personnel may result in damage to the device and constitute a source of danger for the user.
  • Using this manual does not exclude the need to comply with occupational health and safety regulations and with other relevant fire regulations required during the performance of a particular type of work. Before starting the work with the device in special environments, e.g. potentially fire-risk/explosive environment, it is necessary to consult it with the person responsible for health and safety.
  • It is unacceptable to operate the following:

⇒ A damaged meter which is completely or partially out of order,
⇒ A meter with damaged test leads insulation,
A meter stored for an excessive period of time in disadvantageous conditions (e.g. excessive humidity). If the meter has been transferred from a cool to a warm environment of a high level of relative humidity, do not realise measurements until the meter has been warmed up to the ambient temperature (approximately 30 minutes).

- It should be remembered that BAT message appearing on the display indicates that supply voltage of the meter is too low. This message signals also that the batteries must be replaced or the accumulator charged. Measurements performed by means of the meter whose supply voltage is too low are burdened with additional errors that are impossible to be estimated by the user. Such measurements must not be relied on in order to state correctness of protection of a network tested.

- Battery spill and damage to the meter may occur if discharged batteries are left in the meter.

- Before measurements may commence, make sure the test leads are connected to the appropriate measurement sockets.

- Do not operate a meter with an open or incorrectly closed battery (accumulator) compartment on power it from other sources than those specified in the present manual.

- The R _ISO inputs of the meter are protected electronically from overload (e.g. due to having been connected to a live circuit) up to 440V rms for 60 seconds.

• Repairs may be realised solely by an authorised service point.

ATTENTION!

Only accessories for a given device should be used. Use of different accessories can lead to errors in the test connection and can introduce additional measurement uncertainties.

Attention:

Due to continuous development of the meter's software, the actual appearance of the display, in case of some of the functions, may slightly differ from the display presented in this operating manual.

Note:

An attempt to install drivers in 64-bit Windows 8 may result in displaying "Installation failed" message.

Cause: Windows 8 by default blocks drivers without a digital signature.

Solution: Disable the driver signature enforcement in Windows.

2 Menu

The Menu is accessible in each position of the rotary switch.

Sonel MPI-520 - Menu - 1

Press MENU push-button.

Wireless transmission Measurement settings Meter settings **Language choice** Manufacturer info

ENTER

Select a proper item by means of ▲ push-buttons. Enter a selected option by pressing ENTER.

2.1 Wireless transmission

See chapter 5.3.

2.2 Settings of measurements


Measurement settings 09:27 Mains voltage and frequency Loop main result Measurement settings Cell autoincrement Choose ENTER Edit ESC Exit


ENTER

Select a proper item by means of ▲ push-buttons. Enter a selected option by pressing ENTER.

2.2.1 Network voltage and frequency

Before measurements a nominal network voltage U_n (110/190V, 115/200V, 127/220V, 220/380V, 230/400V or 240/415V) should be selected that is valid in the area where measurements are made. This voltage value is used for calculating the values of prospective short-circuit current.

Determination of network frequency, that is the source of potential interferences, is necessary in order to select a proper measuring signal frequency in resistance-to-earth measurements. Only the measurement conducted with a properly selected frequency of measuring signal will ensure optimum filtration of interferences. The meter is designed for filtration of interferences that originate from 50 Hz and 60 Hz networks.


Mains voltage and freq. 09:27 Un voltage fn frequency □ 110/190V ✓ 50Hz □ 115/200V □ 60Hz □ 127/220V □ 220/380V ✓ 230/400V □ 240/415V Choose ENTER Approve ESC Exit


ENTER

By means of , push-buttons select a parameter to be changed, by means of , select network voltage and frequency. Confirm a choice made by means of ENTER push-button.

2.2.2 Main result of short circuit loop impedance measurement


Loop main result 09:27 Zs Ik Choose ENTER Approve ESC Exit


ENTER

By means of ▲ push-buttons select main result in the form of impedance Z_s or prospective short-circuit current I_k ; confirm a choice made by means of ENTER push-button.

2.2.3 Measurement settings

The setting enables activation/deactivation of the field displaying measurement settings. Show or hide the field with measurement settings by means of and push buttons, press ENTER push-button.

ZL-N,L-L, UL-N,L-L 09:27 L-N! f<45Hz U=0,0V I_K=--- R =--- X_L =--- U =--- f =--- L=1,2m N=1,2m START Meas. WIRE HELP

ZL-N,L-L, UL-N,L-L 09:27 L-N! f<45Hz U=0,0V I_K=--- R =--- X_L=--- U =--- f =--- START Meas. WIRE HELP

2.2.4 Cell autoincrementing


Cell autoincrement 09:27 ✓ On □ Off Choose ENTER Approve ESC Exit


ENTER

By means of ▲ push-buttons select the option of automatic incrementing of field number after its storing in the memory or the manual incrementing option (automatic incrementing is deactivated); confirm a choice made by means of ENTER push-button.

2.3 Settings of the meter


Meter settings 10:36 LCD Contrast LCD Backlight Auto-Off settings Date/time Default settings Software upgrade Choose ENTER Edit ESC Exit


ENTER

Select a suitable item by means of ▲, push-buttons; enter the edition of a selected option by means of ENTER push-button.

2.3.1 LCD contrast


LCD Contrast 09:27 Contrast 70 % Choose ENTER Accept ESC Exit


ENTER

Select contrast value by means of ▲, push-buttons; confirm a choice made by means of ENTER push-button.

2.3.2 LCD backlight


LCD Backlight 10:36 ✓ Always □ 30s □ 60s Choose ENTER Approve ESC Exit


ENTER

Use the ▲ and buttons to select the backlight off (with the

button) or time to Auto-off. Press ENTER to validate.

2.3.3 Auto-OFF settings

The setting defines time till automatic shutdown of idle meter.


Auto-Off settings ✓ AUTO-OFF disabled □ 5 minutes □ 15 minutes □ 30 minutes □ 60 minutes Choose ENTER Approve ESC Exit


ENTER

Set the time or deactivate Auto-OFF by means of ▲push-buttons; confirm a choice made by means of ENTER push-button.

2.3.4 Date and time

By means of ◀, push-buttons select the value to be changed (day, month, year, hour, minute). Set a required value by means of ▲, ▼ push-buttons. When required settings are made, press ENTER push-button.

2.3.5 Factory (default) settings

In order to introduce factory (default) settings, highlight YES by means of ◀, ▶ push-buttons and press ENTER push-button.

2.3.6 Program update

ATTENTION!

This function may be used only by the users who are fluent in operation of computer equipment.

The guarantee does not cover defective operation of the device resulting from wrong use of this function.

ATTENTION!

A new package of batteries should be installed before programming or the accumulator should be charged.

During programming the meter must not be switched off as well as the transmission cable must not be disconnected.

Before updating the program, download the program that is use for programming the meter from the manufacturer's website, install this program on your computer and connect the meter to the computer.

Select Software upgrade in the MENU and follow the instructions displayed by the program.

2.4 Language selection

  • By means of and ▲ push-buttons choose **Language choice** in the main MENU; press ENTER push-button.
  • Select a required language by means of ▲ and ▼ push-buttons; press ENTER push-button.

2.5 Information about manufacturer

By means of ▲ and ▼ push-buttons choose Manufacturer info in the main MENU; press ENTER push-button.

3 Measurements

Remarks:

  • A progress bar is displayed during long measurements.
  • The content of this chapter should be thoroughly familiarized with since it describes the meter circuits, the methods of measurements and basic principles concerning interpretation of measurement results.
  • Result of the latest measurement is remembered by the meter until a next measurement is started or measurement settings are changed or the measuring function is changed by means of the rotary switch or the meter is switched off. The result of the latest measurement is displayed on the screen for 20 seconds. It can be recalled by pressing ENTER push-button.

WARNING: During measurements (short circuit loop, RCD), earthed parts and parts accessible in the electrical installation being tested must not be touched.

WARNING: During a measurement, switching of the range switch is forbidden because it may damage the meter and pose a threat to the user.

3.1 Measurement of alternating voltage and frequency

The meter measures and displays alternating voltage and frequency of the network in all measuring functions except R_E , R_X , R±200mA , R_ISO -conductor. For U_L and R_ISO functions, only voltage is displayed. This voltage is measured for the frequencies within the range of 45..65 Hz as True RMS. If a frequency measured is outside the specified range, a proper message is displayed instead of the frequency value: f<45 Hz or f>65 Hz. Voltage is displayed as the main result only for U_L-N,L-L , Z_L-N,L-L , U_L-PE , Z_L-PE and U,I,P,Q,S,f,cosφ functions for Only U mode selected. The test leads should be connected as for a given measuring function.

3.2 Checking correctness of PE (protective earth) connections

L1 L2 L3 N PE or PE

When the meter is connected according to the drawing, touch the touch electrode with a finger and wait for about 1 second. When voltage if found on PE, the device displays PE! message (error in the installation; PE lead is connected to the phase lead) and generates a continuous audio signal. This possibility is available for all measuring functions that apply to residual current devices (RCD) and short circuit loop.

Remarks:

WARNING:

When phase voltage is detected on PE lead, measurements must be immediately stopped and a fault in the installation must be removed.

  • The person making a measurement must ensure that he/she is standing on a non-insulated floor during the measurement; otherwise the result of the measurement may be incorrect.
  • The threshold value, which triggers the signal of exceeded allowable voltage on PE conduit, is approximately 50 V.

3.3 Measurement of current, active power, reactive power, apparent power and cosφ coefficient


RPM OVA Zn2 Lr Zn1 Lr RPM OVA Zn2 Lr Zn1 Lr

Set the rotary switch of function selection at U,I,P,Q,S,f,cosφ position.


Sonel MPI-520 - Measurement of current, active power, reactive power, apparent power and cosφ coefficient - 2

flowchart
graph TD
    A["F1"] --> B["Arrow to F1"]
    B --> C["Arrow to ENTER"]
    C --> D["Arrow to ENTER"]

Press F1 push-button. Select "U,I,f,cosφ,P,Q,S" by means of and push-buttons and press ENTER push-button. (If you want to measure voltage or current only, select a proper position.)


Assemble the system according to the below drawing.
L1 L2 L3 N PE L N


Power measurement P =3,32kW UL-N =228,5V Q =1,41kvar I =15,8mA S =3,61kVA f =50,0Hz cos r=0,92 MODE HELP

Read out the results.

3.4 Measurement of short circuit loop parameters

Sonel MPI-520 - Measurement of short circuit loop parameters - 1

If there are residual current devices in the network tested, they should be bypassed by bridging for the period of impedance measurement. However, it should be remembered that the circuit tested is modified in this way and the results obtained may slightly differ from the actual results.

Each time after completion of measurements, modifications introduced to the installation for the period of measurements should be removed and operation of the residual current device should be checked.

The above remark does not apply to measurements of short circuit loop impedance with the use of Z_L-PE RCD function.

Sonel MPI-520 - Measurement of short circuit loop parameters - 2

Measurements of fault loop impedance performed downstream of inverters are ineffective and their results are unreliable. This is due to the instability of internal impedance in inverter circuits during its operation. The measurements of fault loop impedance should not be performed directly downstream of inverters.

3.4.1 Measurement of short circuit loop parameters in L-N and L-L circuit


Soda Umm ASYS Xod Ioh R R HESO HESO Umm

Set the rotary switch of function selection at Z_L-N,L-L/U_L-N,L-L position.


Sonel MPI-520 - Measurement of short circuit loop parameters in L-N and L-L circuit - 2

flowchart
graph TD
    A["F1"] --> B["Arrow to Top"]
    B --> C["Arrow to Bottom"]
    C --> D["ENTER"]

Press F1 push-button if lead length L needs to be selected. Select lead length by means of and push-buttons and press ENTER push-button.


Connect test leads according to the drawing a) for measurement in L-N circuit or b) for measurement in L-L circuit
L1 L2 L3 N PE a) b) L N lub L N


ZL-N,L-L, UL-N,L-L Ready! f=50,0Hz U_L-N = 239,5V L=1,2m N=1,2m 09:27 I_K=---- R =---- X_L=---- U =---- f =---- START Meas. WIRE HELP

The meter is ready for measurement.


Sonel MPI-520 - Measurement of short circuit loop parameters in L-N and L-L circuit - 5

Make measurement by pressing START push-button


ZL-N,L-L, UL-N,L-L Ready! U=242,7V ZL-N = 11,22Ω I_K=20,5A R =5,18Ω X_L=9,95Ω U =242,8V f =50,0Hz L=1,2m N=1,2m START Meas. ENTER Write ESC Exit WIRE HELP

Read out the result.

The result is displayed on the screen for 20s.

The result can be recalled by pressing ENTER push-button.

Remarks:

  • The result can be stored in the memory (see point 4.1).
  • When many measurements are made in short time intervals, the meter may emit a large amount of heat. As a result of this, the housing of the device may become hot. This is a normal phenomenon and the meter is equipped with the protection against excessive temperature. After approximately 15 consecutive measurements of short circuit loop, wait until the instrument cools down. This limitation results from the high current measurement and multi-functionality of the meter.
  • Minimum interval between successive measurements is 5 seconds. This minimum interval requirement is controlled by the meter. A next measurement can be made only when READY! message appears on the screen.

Additional information displayed by the meter

READY!The meter is ready for measurement.
L-N! U_L-N voltage is incorrect for making a measurement.
L-PE! U_L-PE voltage is incorrect for making a measurement.
N-PE! U_N-PE voltage exceeds allowable value of 50V.
Sonel MPI-520 - Additional information displayed by the meter - 1Phase connected to N terminal instead of L terminal (for example, exchange of L and N in the mains socket.
Temperature exceeded.
f!Network frequency is outside the range of 45...65 Hz.
Error during measureA correct result can not be displayed.
Loop circuit malfunction!The meter should be serviced.
No U_L-N !Lack of U_L-N voltage before the principal measurement.
U>500 V! and continuous audio signalBefore measurement, voltage at test terminals exceeds 500 V.

3.4.2 Measurement of short circuit loop parameters in L-PE circuit


Auto Icc R EM Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc Rc

Set the rotary switch of function selection at Z_L-PE/U_L-PE position.


Sonel MPI-520 - Measurement of short circuit loop parameters in L-PE circuit - 2

flowchart
graph TD
    A["F1"] --> B["Arrow to right"]
    B --> C["Arrow down"]
    C --> D["ENTER"]

Press F1 push-button if L lead length needs to be selected. Select a lead length by means of and push-buttons and press ENTER push-button.


Connect test leads according to one of the drawings.
L1 L2 L3 N PE L PE N or

Checking effectiveness of protection against electric shock of the meter housing in case of: a) TN network b) TT network

L1 L2 L3 PEN a) b) R0 L PE


ZL-PE, UL-PE Ready! f=50.0Hz U L-PE = 239,6V I_K =---- R =---- X_L =---- U L-PE =---- f =---- L=1,2m PE=1,2m START Meas. WIRE HELP

The meter is ready for measurement.


Sonel MPI-520 - Measurement of short circuit loop parameters in L-PE circuit - 6

Make measurement by pressing START push-button.


ZL-PE, UL-PE Ready! U=243,1V ZL-PE = 11,09Ω L=1,2m PE=1,2m 09:27 IK =20,7A R =5,25Ω XL =9,77Ω UL-PE=242,8V f =50,0Hz START Meas. ENTER Write ESC Exit WIRE HELP

Read out the result.

The result is displayed on the screen for 20s. The result can be recalled by pressing ENTER push-button.

Remarks:

  • Double lead measurement is possible when a test lead other that the lead with a mains socket is selected.
  • Remaining issues connected with the measurements as well as the messages displayed are the same as those described for measurements in L-N circuit or L-L circuit.

3.4.3 Measurement of short circuit loop impedance in L-PE circuit protected with residual current device (RCD)


R M H E R M H E R M H E R M H E R M H E R M H E R M H E R M H E R M H E R M H E R M H E R M H E R M H E R M

Set the rotary switch of function selection at Z_L-PE RCD position.


Sonel MPI-520 - Measurement of short circuit loop impedance in L-PE circuit protected with residual current device (RCD) - 2

flowchart
graph TD
    A["FL"] --> B["Directional Arrow"]
    B --> C["ENTER"]

Press F1 push-button if L lead length needs to be selected. Select a lead length by means of ▲ and push-buttons and press ENTER push-button.


Connect test leads according to one of the drawings.
RCD L1 L2 L3 N PE a L1 PE N b TN-S

Sonel MPI-520 - Measurement of short circuit loop impedance in L-PE circuit protected with residual current device (RCD) - 4

L1 L2 L3 N RCD PE L N TT R0

RCD L1 L2 L3 N PE PEN a L PE N b TN-C-S

Remarks:

  • Maximum measurement time is about 32 seconds. The measurement can be interrupted by pressing ESC push-button.
  • In the electrical installations in which 30 mA - rated residual current devices are used, it may happen that the sum of leakage currents of the installation and the test current will trigger RCD. In such a situation, one should try to reduce leakage current of the network being tested (for example, by disconnecting load points).
  • Remaining issues connected with the measurements as well as the messages displayed are the same as those described for measurements L-PE circuit.
  • The function works for residual current devices of nominal current ≥ 30 mA.

3.4.4 Prospective short-circuit current

The meter always measures impedance. The short-circuit current is calculated according to the following formula:

$$ I _ {\overline {{k}}} \frac {U _ {n}}{Z _ {s}} $$

where: U_n - nominal voltage of the network being tested selected in MENU, Z_s - impedance measured.

On the basis of U_n nominal voltage selected (point 2.1.1), the meter automatically recognizes the measurement at phase voltage or phase-to-phase voltage and takes it into account in the calculations. If the voltage of the network being tested is outside the tolerance range, the meter will not be able to determine a proper nominal voltage for the short-circuit current calculation. In such a case, horizontal dashes will be displayed instead a short-circuit current value. The following diagram shows voltage ranges for which short-circuit current value is calculated.

Sonel MPI-520 - Prospective short-circuit current - 1

bar | Voltage Range (V) | Current Value | | ----------------- | ------------- | | U=110V | 99 | | U=110V | 110 | | U=110V | 121 | | U=110V | 171 | | U=110V | 190 | | U=110V | 209 | | U=115V | 99 | | U=115V | 103 | | U=115V | 115 | | U=115V | 127 | | U=115V | 180 | | U=115V | 200 | | U=115V | 220 | | U=127V | 99 | | U=127V | 114 | | U=127V | 127 | | U=127V | 140 | | U=127V | 198 | | U=127V | 220 | | U=127V | 242 | | Un=220V | 99 | | Un=220V | 198 | | Un=220V | 220 | | Un=220V | 242 | | Un=230V | 99 | | Un=230V | 207 | | Un=230V | 230 | | Un=230V | 253 | | Un=230V | 360 | | Un=230V | 400 | | Un=230V | 440 | | Un=240V | 99 | | Un=240V | 216 | | Un=240V | 240 | | Un=240V | 264 | | Un=240V | 373 | | Un=240V | 415 | | Un=240V | 440 |

3.5 Measurement of resistance-to-earth

The three-pole measuring method is the basic type of resistance-to-earth measurement.

h elec- ted from of the

Disconnect earth electrode being tested from the installation of the building.

Auto L1 L2 R1 R2 R3 R4 R5 R6 R7 R8 R9 R10 R11 R12 R13 R14 R15 R16 R17 R18 R19 R20

Set the rotary switch of function selection at R_E position.

H E S

The current electrode (driven into earth) should be connected to H socket of the meter.

The voltage electrode (driven into earth) should be connected to S socket of the meter.

The earth electrode being tested should be connected to E socket of the meter.

The earth electrode being tested and the current electrode and the voltage electrode should be located in one line.

Re Ready! UN=0V RE = --- - Un=25V 09:27 RH=--- RS=--- δ=--- START Meas. Un HELP

The meter is ready for measurement.

Value of interference voltage U_N can be read on the display.

Sonel MPI-520 - ⑤ - 1

Press F1 push-button to change test voltage.

Measurement voltage 25V 50V Choose ENTER Accept ESC Exit


ENTER

Select test voltage by means of push-buttons and confirm by pressing ENTER.


Sonel MPI-520 - ⑤ - 4

Press START push-button to start the measurement.


Re Ready! UN=0V RE = 5 , 40Ω Un=50V 09:27 RH=1,53kΩ Rs=2,10kΩ δ=3% LIMIT! START Meas. ENTER Write Un HELP

Read out the result.

Resistance of current electrode Resistance of voltage electrode Value of additional uncertainty caused by resistance of the electrodes


S

Repeat the measurements (points 3, 7, 8) shifting the voltage electrode several metres: the electrode should be shifted farther and closer to the earth electrode being tested. If RE measurement results differ from one another by more than 3%, the distance of the current electrode from the earth electrode being tested should be considerably increased and the measurements should be repeated.

Remarks:

Measurement of resistance-to-earth may be carried out if voltage of interferences does not exceed 24 V. Voltage of interferences is measured up to the level of 100 V but above 50 V it is signalled as dangerous. The meter must not be connected to voltages exceeding 100 V.

- Particular attention should be paid to quality of connection between the object being tested and the test lead – the contact area must be free from paint, rust, etc. - If resistance of test probes is too high, Earth electrode measurement will be burdened with additional uncertainty. Particularly high uncertainty of measurement occurs when a small value of resistance-to-earth is measured with probes that have a weak contact with earth (such a situation occurs frequently when the earth electrode is well made and the upper soil layer is dry and slightly conductive). In such a case, the ratio between resistance of the probes and resistance of the earth electrode tested is very high and consequently, uncertainty of measurement that depends on this ratio is also very high. The uncertainty is displayed on the screen in the column of additional results. In order to reduce the uncertainty, it is recommended to improve the contact between the probe and soil, for example, by dampening with water the place where the probe is driven into earth, driving the probe into earth in a different place or using a 80 cm-long probe. Test leads should also be checked as follows: check whether their insulation is not defective and whether the lead – banana plug – probe contact areas are not corroded or loosened. In majority of cases the measurement accuracy achieved is satisfactory. However, one should always be aware of uncertainty value the measurement is burdened with.

Additional information displayed by the meter

RE>1,99 kΩMeasuring range is exceeded.
UN!Voltage at test terminals is higher than 24 V but lower than 50 V, measurement is blocked.
UN>50 V! and continuous audio signalVoltage at test terminals is higher than 50 V.
NOISE!Too low value of signal/noise ratio.
LIMIT!Error caused by resistance of electrodes >30%. (Measured values are used in calculation of uncertainty.)
Sonel MPI-520 - Additional information displayed by the meter - 1Interruption in measuring circuit or resistance of test probes is higher than 60 kΩ.
Electrode resistance >50 kΩResistance of electrodes within the range of 50...60 kΩ.
Aborted!Measurement has been interrupted with ESC key button.

3.6 Measurement of RCD parameters

Attention:

Measurement of U_B , R_E is always conducted with the use of sinusoidal current 0.4I_ n regardless of the settings concerning waveform and multiplication factor I_ n .

3.6.1 Measurement of RCD disconnection current


Auto NPN MEN NPN MEN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN MPN

Set the rotary switch of function selection at ▲ position.


RCD: I_A, U_B, R_C 09:27 L-PE! U=0,1V f<45Hz U_B =---- R_C =---- U_L-PE=---- I_A = ---- 300mA ~ □ U_L=25V ◄► Choose I_an ∧/ ∧ □/G/S HELP

Sonel MPI-520 - Measurement of RCD disconnection current - 3

Press F1 I_An push-button and move to I_An selection.

Press F2 push-button and move to selection of current waveform.

Press F3 ☐/G/S push-button and move to selection of RCD type.

Sonel MPI-520 - Measurement of RCD disconnection current - 4

flowchart
graph TD
    A["Top Circle"] --> B["Down Arrow"]
    C["Bottom Circle"] --> D["Down Arrow"]
    E["Center Circle"] --> F["Shaded Ring"]
    G["ENTER"] --> H["Arrow Down"]

Select an appropriate item by means of ▲ and ▼ push-buttons and confirm by pressing ENTER.


Sonel MPI-520 - Measurement of RCD disconnection current - 5

Move to selection of a second group of parameters by means of and push-buttons.


RCD: I_A, U_B, R_C L-PE! U=0,1V f<45Hz I_A = --- 300mA ~ □ U_L=25V 09:27 U_B =--- R_C =--- U_L-PE=--- Choose U_L MODE HELP

Sonel MPI-520 - Measurement of RCD disconnection current - 7

Press F1 _L push-button and move to selection of U_L .

Press F2 MODE push-button and move to selection of measurement mode.

ENTER

Select an appropriate item by means of ▲ and buttons and confirm by pressing ENTER.

push-


Connect the device to the installation according to the drawing.
RCD L1 L2 L3 N PE L N PE or


RCD: I_A, U_B, R_C Ready! U=242,2V f=50,0Hz I_A = --- 300mA ^ □ U_L=25V 09:27 U_B =--- R_C =--- U_L-PE=--- Choose START Meas. U_L MODE HELP

The meter is ready for measurement. Value of network voltage and frequency can be read on the display.


Sonel MPI-520 - Measurement of RCD disconnection current - 11

Press START to begin measurement.


RCD: I_A, U_B, R_C 09:27 L-PE! U=81,0V f=50,0Hz U_B =0,0V R_C =0Ω U_L-PE=242,7V I_A = 286mA 300mA ~ □ U_L=25V Choose ENTER Write U_L MODE HELP

Read out the result.

Remark:

- Measurement of t_AI disconnection time for selective RCD is not available.

Additional information displayed by the meter

U_B > U_L! The touch voltage exceeds a preset U_L threshold value.
!! placed on the right side of the result means that RCD is out of order
No U_L-N! Lack of neutral lead that is necessary for I _ n constant and pulsed with direct current offset

The remaining information is the same as for fault loop measurement (first 7 positions under point 3.4.1).

3.6.2 Measurement of RCD disconnection time

Auto Auto HEN HAN R1 R2 R3 R4 R5 R6 R7 R8 R9 R10 R11 R12 R13 R14 R15 R16 R17 R18 R19 R20

Set the rotary switch of function selection at position.

RCD: tA, UB, RE 09:27 L-PE! U=80,9V f=50,0Hz UB =---- RE =---- UL-PE=---- tA = --- 300mA ×1 ~ □ UL=25V Choose IΔn ×IΔn ∧/ ∧ HELP

Sonel MPI-520 - Measurement of RCD disconnection time - 3

Press F1 I_An push-button and move to selection of I_An .

Press F2 _n push-button and move to selection of l_an multiplication factor

Press F3 △/▽/▽ push-button and move to selection of current waveform.

ENTER

Select an appropriate item by means of ▲ and ▼ push-buttons and confirm by pressing ENTER.

Sonel MPI-520 - Measurement of RCD disconnection time - 5

Move to selection of a second group of parameters by means of and push-buttons.


RCD: tA, UB, RE 09:27 L-PE! U=81,2V f=50,0Hz UB =---- RE =---- UL-PE=---- tA = --- 300mA ×1 ~ □ UL=25V ◄► Choose UL □/G/S MODE HELP

Sonel MPI-520 - Measurement of RCD disconnection time - 7

Press F1 L push-button and move to selection of UL

Press F2 □/G/S push-button and move to selection of RCD type.

Press F3 MODE push-button and move to selection of measurement mode.

ENTER

Mark an appropriate position by means of ▲ and push-buttons and confirm by pressing ENTER.

Sonel MPI-520 - Measurement of RCD disconnection time - 9

Connect the device to the installation according to the drawing.

RCD L1 L2 L3 N PE L N PE or

RCD: tA, UB, RE Ready! U=233,1V f=50,0Hz tA = --- 300mA ×1 ~ □ UL=25V 09:27 | UB =---- RE =---- UL-PE=---- Choose START Meas. UL □/G/S MODE HELP

The meter is ready for measurement.

Value of network voltage and frequency can be read on the display.


Sonel MPI-520 - Measurement of RCD disconnection time - 12

Press START to begin measurement.


RCD: tA, Ue, Re L-PE! U=81,1V f=50,0Hz tA = 39ms 300mA ×1 ~ □ UL=25V ←→ Choose ENTER Write UL /G/S MODE HELP

Read out the result.
Remarks and information are the same as for IA measurement.

3.6.3 Automatic measurement of RCD parameters

The meter enables automatic measurement of the following: RCD disconnection times ( t_A ), disconnection current ( I_A ), touch voltage ( U_B ) and resistance-to-earth ( R_E ). Additionally, there is a possibility of automatic measurement of short circuit loop impedance Z_L-PE RCD in a manner described in point 3.4.3. In the automatic mode, it is not necessary to actuate a measurement each time by means of START push-button. The operator of the meter only has to initiate a measurement by pressing START push-button once and switch RCD on each time after it is triggered. The below table shows maximum number of parameters measured and sequence of measurements for preset value of rated current I_ n , selected current waveform, type of RCD (standard / selective / short-time delay) and U_L voltage.

No.Parameters measuredMeasurement conditions
I_ n multiplication factorInitial phase (polarization)
1. Z_L-PE
2. U_B R_E
3. t_A 0,5I_ n positive
4. t_A 0,5I_ n negative
5.* t_A 1I_ n positive
6.* t_A 1I_ n negative
7.* t_A 2I_ n positive
8.* t_A 2I_ n negative
9.* t_A 5I_ n positive
10.* t_A 5I_ n negative
11.* I_A positive
12.* I_A negative

* points in which an efficient RCD should disconnected


Auto NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM NAM

Set the rotary switch of function selection at AUTO position.


RCD Auto 09:27 L-PE! U=80,8V f=50,0Hz 300mA ~ □ UL=25V Choose Ian ∨/ ∧ □/G/S HELP

Sonel MPI-520 - Automatic measurement of RCD parameters - 3

Press F1 I_An push-button and move to I_An selection.

Press F2 △/▽ push-button and move to selection of current waveform.

Press F3 □/G/S push-button and move to selection of RCD type.

ENTER

Select an appropriate item by means of ▲ and ▼ push-buttons and confirm by pressing ENTER.


Sonel MPI-520 - Automatic measurement of RCD parameters - 5

Move to selection of a second group of parameters by means of and push-buttons.


RCD Auto 09:27 L-PE! U=81,0V f=50,0Hz 300mA ~ □ UL=25V Choose UL MODE HELP

Sonel MPI-520 - Automatic measurement of RCD parameters - 7

Press F1 _L push-button and move to selection of U_L .

Press F2 MODE push-button and move to selection of measurement mode.

Press F3 WIRE push-button and move to selection of L lead length (at Z_L-PE RCD measurement).

ENTER

Select an appropriate item by means of ▲ and ▼ push-buttons and confirm by pressing ENTER.

Connect the device to the installation according to the drawing.

RCD L1 L2 L3 N PE L N PE or


RCD Auto Ready! U=242,2V f=50,0Hz 300mA ~ □ UL=25V Choose START Meas. Ian ∧/ ∧ □/G/S HELP

The meter is ready for measurement. Value of network voltage and frequency can be read on the display.


Sonel MPI-520 - Automatic measurement of RCD parameters - 11

Press START push-button to start the measurement. If such measurements are selected that require triggering of RCD, operator of the meter should be in the vicinity of RCD and switch it on each time after it is triggered until the measurements are completed (a longer interruption may signify completion of the measurements).


RCD ZL-PE [RCD] Auto 09:27 ZL-PE= 88% 10% L=1,2m 300mA ~ □ UL=25V ESC Abort measure

Progress of measurement process is shown by progress bars: lower bar – total cycle; upper bar – measurement of Z_L-PE RCD and I_A .


RCD ZL-PE [RCD] Auto 09:27 I_K =20,6A R =5,31Ω X_L =9,85Ω U_L-PE=239,9V f <45Hz L=1,2m 300mA ~ □ U_L=25V ←1/2> ENTER Write ESC Exit ◀Screen Screen▶

Read out the result.

10
F3 F4

Groups of results displayed are changed by means of F3 and F4 push-buttons.

RCD ZL-PE [RCD] Auto 09:27 GOOD I_A =272mA+ =272mA- t_A(0.5I) >300ms+ >300ms- t_A(1I) =39ms+ =29ms- t_A(2I) =18ms+ =10ms- t_A(5I) =---- =---- L=1,2m 300mA ~ □ U_L=25V ←2/2> ENTER Write ESC Exit ◀Screen Screen▶

Remarks:

  • The measurement is interrupted, if during measurement of U_B/R_E , RCD has been triggered at 0.5I_ n current or if RCD has not been triggered in other cases or if a preset value of safe voltage U_L has been exceeded.
  • Store the result in the memory (see point 4.1) or press ESC push-button and display only network voltage and frequency.
  • Remaining remarks and information are the same as for I_A and Z_L-PE measurement.

3.7 Measurement of insulation resistance

WARNING: The object tested must not be live.

3.7.1 Double-lead measurement


Auto Iod Ax Ax Ax Ax Ax Auto Iod Ax Ax Auto Iod Ax Auto Iod

Set the rotary switch of function selection at R_ISO position.


Sonel MPI-520 - Double-lead measurement - 2

Press F1 push-button and move to selection of test voltage U_N .

ENTER

Select an appropriate item by means of ▲ and ▼ push-buttons and confirm by pressing ENTER.

Connect test leads according to the drawing.


Rx RISO+ RISO-


RISO Ready! U=0V== RISO = --- UN=500V START Meas. START CONT. meas. UN ENTER HELP

The meter is ready for measurement.


Sonel MPI-520 - Double-lead measurement - 6

flowchart
graph TD
    A["Start"] --> B["ENTER"]
    B --> C["START"]

Press and hold START push-button.

Measurement is performed continuously when the push-button is held in the pressed position.

In order to maintain the measurement, press ENTER push-button while holding START push-button in the pressed position. In order to interrupt the mesaurement, press START push-button again.


RISO UISO=521V RISO=1,40GΩ UN=500V 10:03 UISO= START +ENTER Cont. meas.

View of the screen during measurement performed with the use of ENTER push-button.

Sonel MPI-520 - Double-lead measurement - 8

Riso Ready! U=0V== Riso=1,40GΩ UN=500V 10:03 UISO=521V START Meas. ENTER Write START CONT. meas. UN HELP

Read out the result.

Remarks:

Sonel MPI-520 - Remarks: - 1

During measurements of insulation resistance, dangerous voltage up to 1 kV occurs at the ends of test leads of MPI-520 meter.

Sonel MPI-520 - Remarks: - 2

It is forbidden to disconnect test leads and to change the position of the function switch before completion of measurement. Failure to obey the above instruction will lead to high voltage electric shock and make it impossible to discharge the object tested.

  • The meter emits a continuous audio signal until test voltage reaches 90% of the preset value (and also when 110% of the preset value is exceeded).
  • After completion of measurement, the capacitance of the object tested is discharged by shorting R_ISO + and R_ISO -terminals with resistance of 100 kΩ.

Additional information displayed by the meter

Sonel MPI-520 - Additional information displayed by the meter - 1Test voltage is present on terminals of the meter.
NOISE!Interference voltage occurs on the object being tested. Measurement is possible but may be burdened with additional uncertainty.
LIMIT 1!Activation of current limit. The symbol displayed is accompanied by a continuous audio signal.
Sonel MPI-520 - Additional information displayed by the meter - 2Wrong accessory is connected to test jack (other than WS-03 or WS-04 or AutoISO-1000c).
Sonel MPI-520 - Additional information displayed by the meter - 3WS-03 or WS-04 lead for three-lead measurements is connected (measurement at L-PE, L-N and N-PE terminals).

3.7.2 Measurements with AutolISO-1000c adapter


Zeta Alto ASTO 3d Zeta Alto R R R R R HDD KVA Sparey

Set the rotary switch of function selection at R_iso position.


Adapter AutoISO 1000c

Connect AutoISO-1000c adapter. The meter detects this fact automatically and changes the appearance of the screen.


Riso: Cable 5 Ready! UN-PE = --- RL1-PE = --- UL1-N = --- RL1-PE = --- RL1-N = --- AutoISO-1000 U_N=500V 10:12 ←1/4> START Meas. UN MODE TIME HELP

Sonel MPI-520 - Measurements with AutolISO-1000c adapter - 4

Press F1 U_N push-button and move to selection of test voltage U_N .

Press F2 MODE push-button and move to selection of lead type (3-, 4- or 5-wire lead).

Press F3 TIME push-button and move to selection of a single measurement time.

ENTER

Select an appropriate item by means of ▲ and push-buttons and confirm by pressing ENTER.

Sonel MPI-520 - Measurements with AutolISO-1000c adapter - 6


Connect AutoISO-1000c adapter to the lead tested.
Adapter AutoISO 1000c


Sonel MPI-520 - Measurements with AutolISO-1000c adapter - 8

Press START push-buttons to start measurement. First, checking of voltages on particular pairs of wires is performed. If any of the voltages exceeds allowable voltage, the symbol of this voltage with “!” mark is displayed (e.g. _PE! ) and the measurement is interrupted.


RISO: Cable 5 10:31 UN-PE =521V UL1-PE=521V UL1-N =521V ←1/4> R N-PE = 497MΩ R L1-PE = 1, 19GΩ R L1-N = 696MΩ AutoISO-1000 UH=500V ENTER Write ESC Exit Screen Screen

Read out the results.

F3 F4

Groups of results displayed are changed by means of F3 and F4 push-buttons.

Remarks:

- Remarks and messages the same as in point 3.7.1.

3.7.3 Measurements by means of leads with UNI-Schuko outlet plug (WS-03 and WS-04)


Zeta Ora ATO 2nd Zeta Ora R1 R2 HEN EVA sperry

Set the rotary switch of function selection at R_ISO position.


Sonel MPI-520 - Measurements by means of leads with UNI-Schuko outlet plug (WS-03 and WS-04) - 2

natural_image Pure electrical connection diagram showing a power plug connected to an outdoor control panel (no text or symbols)

Connect WS-03 lead or WS-04 lead with UNI-Schuko outlet plug. The meter detects this fact automatically and changes the appearance of the screen.


Riso 10:33 Ready! UL-N=0V== UL-N =--- UL-PE=--- UN-PE=--- RL-N = ---- RL-PE = ---- RN-PE = ---- (L)(PE)(N) UN=500U START Meas. UN L ↔ N TIME HELP

Sonel MPI-520 - Measurements by means of leads with UNI-Schuko outlet plug (WS-03 and WS-04) - 4

flowchart
graph TD
    A["F1"] --> B["F2"]
    B --> C["F3"]

Press F1 U_N push-button and move to selection of test voltage U_N .

Press F2 L▶N push-button and move to selection of lead sequence: L, PE, N or N, PE, L.

Press F3 TIME push-button and move to selection of a single measurement time.

ENTER

Select an appropriate item by means of ▲ and push-buttons and confirm by pressing ENTER.

Sonel MPI-520 - Measurements by means of leads with UNI-Schuko outlet plug (WS-03 and WS-04) - 6

Remark: If it is known that L and N leads in the socket are exchanged, after pressing F2 push-button, the sequence of (N)(PE)(L) can be selected in order to ensure that the meter provides correct results of measurements.

Connect WS-03 lead or WS-04 lead to the socket tested.

WS-03


Sonel MPI-520 - Measurements by means of leads with UNI-Schuko outlet plug (WS-03 and WS-04) - 8

Press START push-buttons to start measurement. If any of the voltages exceeds allowable voltage value (50 V), Object under voltage message is displayed and the measurement is blocked.


RISO 10:36 UISO=284V UN-PE=--- LIMIT! RN-PE = 43% (L)(PE)(N) UN=500V ESC Abort measure

View of the screen during measurement. Symbol of the resistance being measured is displayed. Progress bar shows% of progress of total measurement.


Riso 10:36 UL-N =521U UL-PE=521U UN-PE=522U RL-N = 26 , 6MΩ RL-PE = 24 , 9MΩ RN-PE = 24 , 9MΩ ( L )(PE )(N ) UN=500V ENTER Write ESC Exit

Read out the results.

Remarks:

- Remarks and messages are the same as in point 3.7.1.

3.8 Low-voltage measurement of resistance

3.8.1 Measurement of resistance of protective conductors and equipotential bonding with ±200 mA current


Sonel MPI-520 - Measurement of resistance of protective conductors and equipotential bonding with ±200 mA current - 1

chemical Electrochemical cell diagram with labeled components including nucleus, mitochondria, and iron electrodes

Set the rotary switch of function selection at R_x R_±200 mA position.


Rx >1999Ω MODE AUTOZERO HELP

Sonel MPI-520 - Measurement of resistance of protective conductors and equipotential bonding with ±200 mA current - 3

Press F1 push-button and move to selection of measurement mode.


Mode select 10:39 Rx RCONT ±200mA Choose ENTER Accept ESC Exit

ENTER

Select R_CONT ± 200 mA item by means of ▲ and ▼ push-buttons and confirm by pressing ENTER.


Connect the meter to the object tested. Measurement starts automatically.
Diagram showing a circuit connection with labeled components L and N, and an attached display panel with four ports.


RCONT ±200mA Ready! R = 3, 57Ω 10:40 RF=0,03Ω RR=0,03Ω START Meas. ENTER Write to memory MODE AUTOZERO HELP

Read out the results.


START

Press START push-button in order to start a next measurement without disconnecting test leads from the object.

Remarks:

ATTENTION!

When "Object under voltage" message is displayed, the object tested is live. The measurement is blocked. The meter must be immediately disconnected from the object.

Additional information displayed by the meter

NOISE!Interference voltage occurs on the object tested. The measurement is possible however it will be burdened with additional uncertainty that is specified in the technical data.

3.8.2 Measurement of resistance


Senna OFTO Iod Ira Ri Rin GSA Shoe Iod

Set the rotary switch of function selection at R_x R_±200 mA position.


RCONT ±200mA Ready! R= --- 10:40 RF=---- RR=---- START Meas. ENTER Last meas. MODE AUTOZERO HELP

Sonel MPI-520 - Measurement of resistance - 3

Press F1 push-button and move to selection of measurement mode.


Mode select 10:41 Rx RCONT ±200mA Choose ENTER Accept ESC Exit

ENTER

Select R_x position by means of and push-buttons and confirm by pressing ENTER.

Sonel MPI-520 - Measurement of resistance - 6


Connect the meter to the object tested.
Diagram showing a circuit connection with two resistors and labeled terminals L and N, alongside an electronic device panel.


Rₓ = 3 , 5Ω MODE AUTOZERO HELP

Read out the result.

Remarks:

- Remarks and messages are the same as in point 3.8.1.

3.8.3 Calibration of test leads

In order to eliminate the impact of the resistance of test leads on measurement result, the compensation (autozeroing) of resistance may be performed. For this purpose, R_x and R_±200 mA functions have AUTOZERO sub-function.


RCONT ±200mA Ready! R = --- 10:44 RF=---- RR=---- START Meas. ENTER Last meas. MODE AUTOZERO HELP

Sonel MPI-520 - Calibration of test leads - 2

Press F2 push-button.

R AUTOZERO 10:44 Ready! To eliminate wire resistance short-circuit its endings and press START. START Meas. ESC Exit

Follow the instructions displayed on the screen.

START


RCONT ±200mA Ready! AUTOZERO R= --- 10:50 | RF=---- RR=---- START Meas. ENTER Last meas. MODE AUTOZERO HELP

AUTOZERO message appears that confirms completion of test leads calibration.

In order to remove the calibration made (return to default calibration), perform the above-mentioned activities with test leads open.

3.9 Checking sequence of phases


Auto Anti Air Air Air Air Air AIR Air Air

Set the rotary switch of function selection

at L_3 L_2 position.

Phase sequence 11:02 UL1-L2=0,3V UL1-L3=0,0V UL2-L3=0,3V HELP


Connect the meter to the installation according to the drawing.
L1 L2 L3 N PE RCD L1 L2 L3

The arrow rotates to the right: correct sequence of phases; the arrow rotates to the left: incorrect sequence of phases.

Phase sequence 11:02 U_L1-L2=399,00 U_L1-L3=400,20 U_L2-L3=399,50 Forward L1 L2 L3 Phase-to-phase voltages. Signalling the presence of individual phases. HELP

4 Memory of measurement result data

MPI-520 meters are equipped with the memory that can store 50,000 single measurement results. The whole memory is divided into 10 memory banks each of them containing 99 memory cells. Thanks to dynamic memory allocation, each of the memory cells can contain different quantity of single measurement results, depending on the needs. Optimal use of the memory can be ensured in this way. Each measurement result can be stored in a memory cell marked with a selected number and in a selected memory bank. Thanks to this, the user of the meter can, at his/her option, assign memory cell numbers to individual measurement points and the memory bank numbers to individual objects as well as the user can perform measurements in any sequence and repeat them without losing other data.

Memory of measurement result data is not deleted when the meter is switched off. Thanks to this, the data can be later read or sent to a computer. Also, the number of a current memory cell or memory bank is not changed.

Remarks:

  • Results of measurements performed for all measuring functions can be stored in one memory cell.
  • When autoincrementing of memory cell number is deactivated, a single result (group of results) stored into the memory does not increase automatically the number of the current memory cell in order to enable storing in this memory cell successive measurement results concerning a given measurement point (object). If series of measurements are made for one function, autoincrementing of memory cell number can be set in MENU. Such autoincrementing takes place after each case of data storing in the memory (activation of autoincrementing – point 2.1.5).
  • Only the results of the measurements started by pressing START key button can be stored in the memory (except autozeroing in low-voltage measurement of resistance).
  • Deletion of the memory is recommended after reading the data or before performing a new series of measurements that may be stored into the same memory cells as the previous ones.

4.1 Recording measurement result data in the memory


Sonel MPI-520 - Recording measurement result data in the memory - 1

Press ENTER after completion of measurement.

The frame indicates that at least one result is stored in the cell. The frame indicates at least one cell in the bank has measurement results stored in it. Type of measurement result to be stored Memory write 11:05 Meas. 4/99 Bank 1/10 RCONT ±200mA The cell is free for a given type of measurement. There are 6 results or one result consisting of 6 screens in the cell. Meas. Bank ENTER Save ESC Exit

Memory write 11:05 Meas. 3/99 Bank 1/10 RCONT ±200mA R=3,56Ω 11:05 RR=5,03Ω RF=2,09Ω 7/7 Meas. ▶ Bank ENTER Save ESC Exit Screen Screen

The cell is occupied for a given type of measurement.

Sonel MPI-520 - Recording measurement result data in the memory - 4

Measurement (memory cell) is selected by means of and push-buttons; memory bank is selected by means of and push-buttons.

Storing of data in the memory is performed by means of ENTER push-button.

Sonel MPI-520 - Recording measurement result data in the memory - 5

If you try to store data in an occupied memory cell, the following warning message will appear:

Memory write 11:06 Cell occupied. Overwrite? Yes No Choose ENTER Approve

Sonel MPI-520 - Recording measurement result data in the memory - 7

After selection of an option by means of and push-buttons, press ENTER push-button.

Remarks:

- In case of RCD the above warning message will appear also when an attempt is made to store a result of specific measurement (or result component) that has been made at different preset I_ n current or for a different preset type of RCD (standard / short-time delay / selective) than the measurements the results of which are already stored in this cell, despite the fact that the memory space designated for this result component may be free. When results of measurements made for a different type of RCD or a different I_ n current are stored, the results concerning a given RCD that have been stored previously will be lost. - Complete set of results (main result and supplementary results) for a given measuring function and preset measurement settings are stored in the memory.

4.2 Viewing memory data

Sonel MPI-520 - Viewing memory data - 1

Base Icc Lrr Lrr Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc Pc

Set the rotary switch of function selection at MEM position.


Sonel MPI-520 - Viewing memory data - 3

Select "Memory browsing" by means of and push-buttons.

Memory 11:09 Memory browsing Memory erasing Choose ENTER Approve


Sonel MPI-520 - Viewing memory data - 5

Press ENTER push-button.

Memory browsing Meas. 1/1 Bank 1/1 ZL-PE, UL-PE ZL-PE=12,92Ω 08:20 IK =17,80Ω R =9,19Ω XL =9,08Ω UL-PE=241,3V f =50,0Hz ←1/4> Meas. ← Bank ESG Exit ◀Screen Screen▶

First of the four results stored in this cell.

Select memory bank by means of and push-buttons; select a memory cell by means of and push-buttons; particular results or components of the result are selected by means of F3 and F4 push-buttons.

The following table specifies the sequence of data storing for individual measurement results.

No.Main resultSupplementary results
1 Z_L-PE or I_K k or Z_L-PE
R
XL
U_L-PE
f
U_B
2 t_A at 0.5I_ n , sinusoidal current, positive and negative initial phase R_E
U_-N
3 t_A at 1I_ n , sinusoidal current, positive and negative initial phase t_A at 2I_ n , sinusoidal current, positive and negative initial phase t_A at 5I_ n , sinusoidal current, positive and negative initial phase
4 I_A sinusoidal current, positive and negative initial phase
5-7as above for unidirectional pulsed current and positive and negative polarization
8-10as above for unidirectional pulsed current with direct current offset and positive and negative polarization
11-13as above for direct current and positive and negative polarization
14 Z_L-N (Z_L-I) or I_K k or Z_L-N (Z_L-L)
R
X_L
U_-N (U_L-L)
f
15 Z_L-PE or I_K k or Z_L-PE
R
X_L
U_-PE
f
16 R_E R_H
R_S
17 R_ISO U_ISO
[LIMIT I!]
[NOISE!]
or
18CABLE 3: R_ISO(N-PE) , R_ISO(L1-PE) , R_ISO(L1-N) , U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE]
or
19CABLE 4: R_ISO(L1-N) , R_ISO(L2-N) , R_ISO(L3-N) , U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE]
20CABLE 4: R_ISO(L1-L2) , R_ISO(L1-L3) , R_ISO(L2-L3) , U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE]
or
21CABLE 5: R_ISO(N-PE) , R_ISO(L1-PE) , R_ISO(L1-N) , U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE]
22CABLE 5: R_ISO(L2-N) , R_ISO(L3-N) , R_ISO(L1-L2) , U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE]
23CABLE 5: R_ISO (L1-L3), R_ISO (L2-L3), R_ISO (L2-PE), U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE] U_ISO , [LIMIT I], [NOISE]
24CABLE 5: R_ISO (L3-PE), U_ISO , [LIMIT I], [NOISE]
25R ±200 mA R_F
R_R
[NOISE !]

Remarks:

- While viewing the memory, empty measurements and memory banks are not accessible. "Measurement 1/20" signifies the first of the 20 measurements; measurements 21...99 are empty and inaccessible. The same principle applies to memory banks. If the memory is stored in a non-continuous manner, empty measurements and memory banks are omitted while the memory is being viewed.

4.3 Deleting memory data

Sonel MPI-520 - Deleting memory data - 1

Set the rotary switch of function selection at MEM position.

Select "Memory erasing" by means of ▲ and ▼ push-buttons.

Press ENTER push-button.

Memory erasing All memory erase Bank erase Cell erase Choose ENTER Approve ESC Exit ④ Select deletion of the whole memory, a memory bank or a measurement by means of ▲ and push-buttons. ⑤ Follow the instruction displayed by the meter.

5 Data transmission

Remarks:

  • Data transmission is not possible during the charging of accumulators.
  • Starting with firmware version 2.95, the support for data transmission via OR-1 module is disabled.

5.1 Computer connection accessories

What is necessary in order to operate the meter with a computer is additional accessories, namely a USB cable and appropriate software. If the required accessories such have not been purchased along with the meter, then they are available from the manufacturer or an authorised distributor.

The accessories may be used in case of many devices manufactured by SONEL S.A. which are equipped with the USB interface.

Detailed information regarding software is available from the manufacturer or an authorised distributor.

5.2 Data transmission with USB joint

  1. Set the rotational function selector at MEM.
  2. Connect the cable to the USB port of the computer and the USB socket of the meter.
  3. Start the programme.

5.3 Data transmission with OR-1 radio module

  1. Connect OR-1 module to the USB socket of the PC.
  2. Start data filing programme.
  3. Select Wireless transmission in the main MENU of the meter

Wireless transmission Measurement settings Meter settings **Language choice** Manufacturer info

or set the function switch to MEM and press F1.

Memory 11:09 Memory browsing Memory erasing Choose ENTER Approve

  1. If a PIN code change is necessary, select Modify PIN code.

Wireless transmission 00:00 Wireless transmission Modify PIN code Choose ENTER Edit ESC Exit

  1. Set the required code with the cursors.

PIN code modification 00:00 1 2 3 Choose ENTER Approve ESC Exit

The same code must be entered in the computer programme. It is used for securing transmission. 6. To start transmission, select Wireless transmission in the MENU or press F1 in the MEM position. The following messages will be displayed: Connecting and then Connection active. If it is impossible

to establish connection the message Wireless connection lost will appear. Once the connection is established, follow the programme manual for data filing.

Note:

Sonel MPI-520 - Note: - 1

Standard pin for OR-1 is the „123”. Settings in the meter according to section 2.2.

6 Power supply

6.1 Monitoring of the power supply voltage

The level of the charge of the batteries or accumulators is currently indicated by the symbol in the right upper corner of the display:

Sonel MPI-520 - Monitoring of the power supply voltage - 1

Bartery charged.

Sonel MPI-520 - Monitoring of the power supply voltage - 2

Bartery low.

Sonel MPI-520 - Monitoring of the power supply voltage - 3

Bartery fully discharged.

Battery empty! 09:30 BAT ON/OFF - Shutdown meter

Bartery fully discharged, measuring blocked.

Note:

  • The displayed BAT symbol means insufficient power supply voltage and the need to charge the accumulators,
  • Measurements realised with an insufficient meter power supply voltage are distorted with additional errors which are impossible to ascertain by the user and thus they cannot constitute a basis for a conclusion of correctness of the tested earthing system.

6.2 Replacement of batteries (accumulators)

MPI-520 meter is powered by 4 batteries (LR14). It can be also powered by the manufacturer's accumulator package (SONEL NiMH).

Battery charger is installed inside the meter and cooperates only with the manufacturer's accumulator package. The charger is powered by external power supply adapter. It can be also powered from the car cigarette lighter socket. The accumulator package as well as the power supply adapter belong to additional accessories and can be purchase separately.

WARNING:

If the test leads are left in the sockets during replacement of the batteries or the package of accumulators, there is a risk of electric shock with a dangerous voltage.

In order to replace the package of accumulators it is necessary to do the following:

  • Remove all the test leads from the sockets and turn the meter off,
  • Remove the four screws of the accumulators/batteries compartment (in the lower part of the casing),
  • Remove the compartment,
  • Remove the compartment cover and remove the accumulators,
  • Insert a new package of accumulators,
  • Insert (snap) the compartment cover,
  • Insert the compartment in the meter,
  • Replace the four screws of the accumulators/batteries compartment.

Sonel MPI-520 - WARNING: - 1

natural_image Technical line drawing of a mechanical device with internal components and mounting holes (no text or symbols)

NOTE!

Do not use the meter when the accumulator compartment is removed or open or power it from other sources than those mentioned in the present manual.

6.3 Charging of accumulators

Charging commences once the power supply has been connected to the meter regardless of the fact whether the meter is on or off. During charging the screen looks as it is presented in the following illustration. The accumulators are charged in accordance with the algorithm of „quick charge” – this process permits to reduce the duration of charging to approximately four hours. The end of the process

of charging is signalled by: Charging finished. In order to turn the device off, remove the power supply plug of the charger.

Operating mode Battery charging 13:13 Messages regarding the process of charging Charge in progress

Charging progress, the changing interior section symbolizes charging

Note:

- As a result of interferences in the network it is possible that the process of charging of accumulators will finish too fast. In the case too short a time of charging is detected it is necessary to remove the plug of the charger and start charging anew.

Additional informations displayed by the meter

MessageCauseProceeding
Battery connection error!Excessive voltage at the accumulator package during charging.Check the contacts of the accumulator package. Should the problem persist, replace the package.
No battery!No communication with the accumulator controller or batteries compartment put in.Check the contacts of the accumulator package. Should the problem persist, replace the package. Put the accumulators compartment instead of batteries.
Battery temperature too low!The ambient temperature is lower than 10°CIt is not possible to charge the accumulators correctly in such a temperature. Place the meter in a warm place and commence the charging mode anew.The present message may be displayed also in the case of deep discharging of the accumulators. It is then recommended to try to turn the charger repeatedly.
Precharge errorA damaged or deeply discharged accumulator packageThe message is displayed for a while and then the precharge process begins again. If after several attempts the message: Battery temperature too high! is displayed, replace the package.

6.4 General principles regarding using Ni-MH accumulators

  • If you do not use the device for a prolonged period of time, then it is recommended to remove the accumulators and store them separately.
  • Store the accumulators in a dry, cool and well ventilated place and protect them from direct sunlight. The temperature of the environment in the case of prolonged storage should not exceed 30^a C. If the accumulators are stored for a long time in a high temperature, then the occurring chemical processes may reduce their lifetime.
  • Accumulators NiMH resist normally 500-1000 charging cycles. The accumulators reach their maximum capacity after being formatted (2-3 charge and discharge cycles). The most important factor which influences the lifetime of an accumulator is the depth of discharge. The deeper the discharge of the accumulator, the shorter its lifetime.
  • The memory effect is limited in the case of NiMH accumulator. These accumulators may be charged at any point with no serious consequences. However, it is recommended to discharge them completely every few cycles.
  • During storage of Ni-MH accumulators they are discharged at the rate of approximately 30% per month. Keeping accumulators at high temperatures may accelerate this process even 100%. In order to prevent excessive discharge of accumulators, after which it would be necessary to format them, it is recommended to charge the accumulators from time to time (even if not in use).
  • Modern fast chargers detect both too low and too high a temperature of accumulators and react to the situation adequately. Too low a temperature should prevent the start of the process of charging, which might damage the accumulator irreparably. An increase of the temperature of the accumulator is a signal to stop charging and is a typical phenomenon. However charging at a high temperature of the environment apart from reducing the lifetime causes an accelerated increase of the temperature of the accumulator, which will be not charged to its full capacity.
  • Remember that in the case of quick charging accumulators are charged to approximately 80% of their capacity; better results may be obtained if the process of charging is continued: the charger goes then to the phase of charging with a low current and after next couple of hours the accumulators are charged to their full capacity.
  • Do not charge or use accumulators in extreme temperatures. Extreme temperatures reduce the lifetime of batteries and accumulators. Avoid placing devices powered from accumulators in very hot environments. The nominal working temperature must be absolutely observed.

7 Cleaning and maintenance

NOTE!

Apply solely the maintenance methods specified by the manufacturer within the present manual.

The casing of the meter may be cleaned with a soft, damp cloth using all-purpose detergents. Do not use any solvents or cleaning agents which might scratch the casing (powders, pastes, etc.).

Clean the probe with water and dry it. Before the probe is stored for a prolonged period of time it is recommended to grease it with any machine lubricant.

The reels and test leads should be cleaned with water and detergents, and then dried.

The electronic system of the meter does not require maintenance.

8 Storage

In the case of storage of the device, the following recommendations must be observed:

  • Disconnect all the test leads from the meter.
  • Clean the meter and all its accessories thoroughly.
    • Wind the long test leads onto the reels.

- In the case the meter is to be stored for a prolonged period of time, the batteries must be removed from the device.

- In order to prevent a total discharge of the accumulators in the case of a prolonged storage, charge them from time to time.

9 Dismantling and utilisation

Worn-out electric and electronic equipment should be gathered selectively, i.e. it must not be placed with waste of another kind.

Worn-out electronic equipment should be sent to a collection point in accordance with the law of worn-out electric and electronic equipment.

Before the equipment is sent to a collection point, do not dismantle any elements.

Observe the local regulations concerning disposal of packages, worn-out batteries and accumulators.

10 Technical data

10.1 Basic data

⇒ Abbreviation „m.v.” used in the specification of accuracy signifies standard measured value.

Measurement of alternating voltages (True RMS)

RangeResolutionAccuracy
0.0...299.9 V0.1 V±(2% m.v. + 6 digits)
300...500 V1 V±(2% m.v. + 2 digits)

• Frequency range: 45...65 Hz

Measurement of frequency

RangeResolutionAccuracy
45.0...65.0 Hz0.1 Hz±(0.1% m.v. + 1 digit)

• Voltage range: 50...500 V

Measurement of current (True RMS)

RangeResolutionAccuracy
0.0..99.9 mA0.1 mA±(5% m.v. + 3 digits)
100..999 mA1 mA
1.00..9.99 A0.01 A±(5% m.v. + 5 digits)
10.0..99.9 A0,1 A
100 ... 400 A1 A

- Nominal network frequency f_n : 50 Hz, 60 Hz

*) Error of clamp must be additionally taken into account

Range[W], [VA], [var]Resolution[W], [VA], [var]Accuracy(with regard to apparent power S) ^1
0.0..99.90,1±(7% m.v. + 3 digits)
100..9991
1.00..9.99 k0.01 k±(7% m.v. + 5 digits)
10,0..99,9 k0.1 k
100 ... 200 k1 k

• Voltage range: 0...500 V
- Current range: 0...400 A
- Nominal network frequency f_n : 50 Hz, 60 Hz
• Number of phases of the circuit tested: 1
• Range of cosφ display: 0.00..1.00 (resolution 0.01)
*) U: 50...500 V, I: 10 mA...400 A
Error of clamp must be additionally taken into account

Measurement of short circuit loop impedance Z_L-PE , Z_L-N , Z_L-L

Measurement of short circuit loop impedance Z_s

Test range according to IEC 61557:

Test leadTest range Zs
1.2 m0.13...1999 Ω
5 m0.17...1999 Ω
10 m0.21...1999 Ω
20 m0.29...1999 Ω
WS-03, WS-040.19...1999 Ω

Display range:

Display rangeResolutionAccuracy
0...19.99 Ω0.01 Ω±(5% m.v. + 3 digits)
20.0...199.9 Ω0.1 Ω
200...1999 Ω1 Ω
  • Nominal working voltage U nL-N √ U nL-L : 110/190 V, 115/200 V, 127/220 V, 220/380 V, 230/400 V, 240/415 V
    • Working range of voltage: 95...270 V (for Z_L-PE and Z_L-N ) and 95...440 V (for Z_L-L )
  • Nominal network frequency f_n : 50 Hz, 60 Hz
    • Working range of frequency: 45...65 Hz
    • Maximum test current (for 415 V): 41.5 A (10 ms)
    • Control of correctness of PE terminal connection by means of a touch electrode

Indications of short circuit loop resistance R_S and short circuit loop reactance X_S

Display rangeResolutionAccuracy
0..19.99 Ω0.01 Ω±(5% + 5 digits) of Z_S value

- Calculated and displayed for a value of Z_s<20

Indications of short-circuit current I_K

Test ranges according to IEC 61557 can be calculated on the basis of test ranges for Z_s and nominal voltages.

Display rangeResolutionAccuracy
0.055...1.999 A0.001 ACalculated on the basis of error for fault loop
2.00...19.99 A0.01 A
20.0...199.9 A0.1 A
200...1999 A1 A
2.00...19.99 kA0.01 kA
20.0...40.0 kA0.1 kA

- Prospective fault current calculated and displayed by the meter may slightly differ from the value calculated by the user with a calculator, basing on the displayed value of the impedance, because the meter calculates the current from unrounded value of fault loop impedance (which is used for displaying). As the correct value, consider current value, displayed by the meter or by firmware.

Measurement of short circuit loop impedance Z_L-PE RCD (without triggering of RCD)

Measurement of short circuit loop impedance Z_6

Test range according to IEC 61557: 0.50...1999 Ω for 1.2 m, WS-03 and WS-04 leads and 0.51...1999 Ω for 5 m, 10 m and 20 m leads

Display rangeResolutionAccuracy
0...19.99 Ω0.01 Ω±(6% m.v. + 10 digits)
20.0...199.9 Ω0.1 Ω±(6% m.v. + 5 digits)
200...1999 Ω1 Ω
  • It does not cause triggering of RCD of I n ≥ 30 mA
  • Nominal working voltage U_n : 110 V, 115 V, 127 V, 220 V, 230 V, 240 V
    • Working range of voltage: 95...270 V
  • Nominal network frequency f_n : 50 Hz, 60 Hz
    • Working range of frequency: 45...65 Hz
  • Control of correctness of PE terminal connection by means of a touch electrode

Indications of short circuit loop resistance R_S and short circuit loop reactance X_S

Display rangeResolutionAccuracy
0..19.99 Ω0.01 Ω±(6% + 10 digits) of Zsvalue
  • Calculated and displayed for a value of Z_s<20

Indications of short-circuit current I_K

Test range according to IEC 61557 can be calculated on the basis of test ranges for Z _s and nominal voltages.

Display rangeResolutionAccuracy
0.055...1.999 A0.001 ACalculated on the basis of accuracy for fault loop
2.00...19.99 A0.01 A
20,0...199.9 A0.1 A
200...1999 A1 A
2.00...19.99 kA0.01 kA
20.0...40.0 kA0.1 kA

- Prospective fault current calculated and displayed by the meter may slightly differ from the value calculated by the user with a calculator, basing on the displayed value of the impedance, because the meter calculates the current from unrounded value of fault loop impedance (which is used for displaying). As the correct value, consider k current value, displayed by the meter or by firmware.

Measurement of parameters of RCD

  • Nominal working voltage U _n : 110 V, 115 V, 127 V, 220 V, 230 V, 240 V
    • Working range of voltage: 95...270 V
  • Nominal network frequency f_n : 50 Hz, 60 Hz
    • Working range of frequency: 45...65 Hz

RCD trigger and response time test t_A (for t_A mode).

Test range according to IEC 61557: 0 ms ... to the upper limit of displayed value

Type of RCDMultiplication factor settingTest rangeResolutionAccuracy
Standard and short-time delay0.5 I_ n 0..300 ms1 ms± 2% m.v. ±2 digits ^1)
1 I_ n
2 I_ n 0..150 ms
5 I_ n 0..40 ms
Selective0.5 I_ n 0..500 ms
1 I_ n
2 I_ n 0..200 ms
5 I_ n 0..150 ms

1) for I_ n = 10 mA and 0,5 I_ n accuracy is ±2% m.v. ±3 digits
• Accuracy of differential current setting:

for 1*I An , 2*I An and 5*I _An ..... 0..8%

for 0,5* I_An -8..0%

Effective value of forced leakage current at measurement of RCD disconnection time

I_ n Multiplication factor setting
0.51
~^^^^==~^^^^==
1053,53,5510202020
301510,510,51530424260
10050353550100140140200
300150105105150300420420600
5002501751755007007001000*
10005001000
I_ n Multiplication factor setting
25
= =
102040404050100100100
30608484120150210210300
1002002802804005007007001000*
300600840840
5001000
1000

* - does not apply to U_n = 110 V, 115 V and 127

Measurement of resistance-to-earth R_E

Selected nominal current of RCDTest rangeResolutionTest currentAccuracy
10 mA0.01 kΩ...5.00 kΩ0.01 kΩ4 mA0..+10% m.v. ±8 digits
30 mA0.01 kΩ...1.66 kΩ12 mA0..+10% m.v. ±5 digits
100 mA1 Ω..500 Ω40 mA0..+5% m.v. ±5 digits
300 mA1 Ω..166 Ω120 mA
500 mA1 Ω..100 Ω200 mA
1000 mA1 Ω..50 Ω400 mA

Measurement of touch voltage U_B in relation to nominal differential current Test range according to IEC 61557: 10.0...99.9 V

Test rangeResolutionTest currentAccuracy
0..9.9 V0.1 V 0.4 × I_ n 0..10% m.v. ± 5 digits
10.0..99.9 V0..15% m.v.

Measurement of RCD disconnection current I_A for sinusoidal differential current Test range according to IEC 61557: (0,3...1,0) I_ n

Selected nomi- nal current of RCDTest rangeResolutionTest currentAccuracy
10 mA3.0..10.0 mA0.1 mA 0.3 × I_ n..1.0 × I_ n ±5% I_ n
30 mA9.0..30.0 mA
100 mA30..100 mA1 mA
300 mA90..300 mA
500 mA150..500 mA
1000 mA300..1000 mA

- it is possible to start the measurement from the positive of the negative half of forced leakage current

- test current passage time .... max. 3200 ms

Measurement of RCD disconnection current I_A for differential unidirectional pulsed current and unidirectional pulsed current with 6mA direct current offset

Test range according to IEC 61557: (0.35...1.4) I_ n for I_ n ≥ 30 mA and (0.35...2) I_ n for I_ n = 10 mA

Selected nomi- nal current of RCDTest rangeResolutionTest currentAccuracy
10 mA3.5..20.0 mA0.1 mA 0.35 × I_ n.2.0 × I_ n ± 10% I_ n
30 mA10.5..42.0 mA 0.35 × I_ n.1.4 × I_ n ± 10% I_ n
100 mA35..140 mA1 mA
300 mA105..420 mA
500 mA175..700 mA

• measurement can be performed for positive or negative half-periods of forced leakage current
- test current passage time .... max. 3200 ms

Measurement of RCD disconnection current I_A for differential direct current

Test range according to IEC 61557: (0.2...2)I _An

Selected nominal current of RCDTest rangeResolutionTest currentAccuracy
10 mA2.0..20.0 mA0.1 mA 0.2 × I_ n..2.0 × I_ n ± 10% I_ n
30 mA6..60 mA1 mA
100 mA20..200 mA
300 mA60..600 mA
500 mA100..1000 mA

• measurement can be performed for positive or negative forced leakage current
- test current passage time .... max. 5040 ms

Measurement of resistance-to-earth R_E

Test range according to IEC 61557-5: 0,5 Ω...1,99 kΩ for test voltage of 50 V and 0,56 Ω...1,99 kΩ for test voltage of 25 V

RangeResolutionAccuracy
0.00...9.99 Ω0.01 Ω±(2% m.v. + 4 digits)
10.0...99.9 Ω0.1 Ω±(2% m.v. + 3 digits)
100...999 Ω1 Ω
1.00...1.99 kΩ0.01 kΩ

• test voltage: 25 V or 50 V rms
- test current: 20 mA, sinusoidal rms 125 Hz (for f_n = 50 Hz) and 150 Hz (for f_n = 60 Hz)
- blocking of measurement at interference voltage of U N >24 V
• maximum measured voltage of interferences U N
max=100 V
• maximum resistance of auxiliary earth electrodes: 50 kΩ

Measurement of resistance of auxiliary earth electrodes R_H , R_S

Display rangeResolutionAccuracy
000...999 Ω1 Ω±(5% (RS + RE + RH) + 3 digits)
1.00...9.99 kΩ0.01 kΩ
10.0...50.0 kΩ0.1 kΩ

Measurement of interference voltages

Internal resistance: about 100 kΩ

RangeResolutionAccuracy
0...100 V1 V±(2% m.v. + 3 digits)

Low-voltage measurement of continuity of circuit and resistance

Measurement of continuity of protective conductors and equipotential bondings with ±200 mA current

Test range according to IEC 61557-4: 0,12...400 Ω

RangeResolutionAccuracy
0.00...19.99 Ω0.01 Ω±(2% m.v. + 3 digits)
20.0...199.9 Ω0.1 Ω
200...400 Ω1 Ω

• Voltage at open terminals: 4...9 V
- Output current at R<2 Ω: min. 200 mA (I sc: 200..250 mA)
• Compensation of test leads resistance
• Measurements for both current polarizations

Measurement of resistance with low current

RangeResolutionAccuracy
0.0...199.9 Ω0.1 Ω±(3% m.v. + 3 digits)
200...1999 Ω1 Ω

• Voltage at open terminals: 4...9 V
• Output current < 8 mA
• Audio signal for resistance being measured < 30 Ω±50%
• Compensation of test leads resistance

Measurement of insulation resistance

Test range according to IEC 61557-2 for U_N = 50 V: 50 kΩ...250 MΩ

Display range for U_N=50 V ResolutionAccuracy
0...1999 kΩ1 kΩ± (3% m.v. + 8 digits),[± (5% m.v. + 8 digits)] *
2.00...19.99 MΩ0.01 MΩ
20.0...199.9 MΩ0.1 MΩ
200...250 MΩ1 MΩ

* - for WS-03 and WS-04 leads

Test range according to IEC 61557-2 for U_N = 100 V: 100 kΩ...500 MΩ

Display range for UN = 100 VResolutionAccuracy
0...1999 kΩ1 kΩ± (3% m.v. + 8 digits)[± (5% m.v. + 8 digits)] *
2.00...19.99 MΩ0.01 MΩ
20.0...199.9 MΩ0.1 MΩ
200...500 MΩ1 MΩ

* - for WS-03 and WS-04 leads

Test range according to IEC 61557-2 for U_N = 250 V: 250 kΩ...999 MΩ

Display range for UN = 250 VResolutionAccuracy
0...1999 kΩ1 kΩ± (3% m.v. + 8 digits)[± (5% m.v. + 8 digits)] *
2.00...19.99 MΩ0,01 MΩ
20.0...199.9 MΩ0,1 MΩ
200...999 MΩ1 MΩ

* - for WS-03 and WS-04 leads

Test range according to IEC 61557-2 for U_N = 500 V: 500 kΩ...2.00 GΩ

Display range for UN = 500 VResolutionAccuracy
0...1999 kΩ1 kΩ± (3% m.v. + 8 digits)[± (5% m.v. + 8 digits)] *
2.00...19.99 MΩ0.01 MΩ
20.0...199.9 MΩ0.1 MΩ
200...999 MΩ1 MΩ
1.00...2.00 GΩ0.01 GΩ± (4% m.v. + 6 digits)[± (6% m.v. + 6 digits)] *

* - for WS-03 and WS-04 leads

Test range according to IEC 61557-2 for U_N = 1000 V: 1000 kΩ...3,00 GΩ

Display range for UN = 1000 VResolutionAccuracy
0...1999 kΩ1 kΩ± (3% m.v. + 8 digits)
2.00...19.99 MΩ0.01 MΩ
20.0...199.9 MΩ0.1 MΩ
200...999 MΩ1 MΩ
1.00...3.00 GΩ0.01 GΩ± (4% m.v. + 6 digits)

• Test voltages: 50 V, 100 V, 250 V, 500 V and 1000 V

• Accuracy of generated voltage (Robc [Ω] ≥ 1000*U _N [V]): -0+10% from the set value

• Detection of a dangerous voltage before commencing a measurement

- Discharging the object tested

- Measurement of insulation resistance with the use of UNI-Schuko plug (WS-03, WS-04) between all three terminals ( U_N=1000 V is not available)

- Insulation resistance measurement for multi-wire cables (max. 5) using an optional external adapter

- Measurement of voltage on terminals +R ISO , -R ISO within the range of: 0..440 V

• Test current < 2 mA

Remark:

With regard to measurements conducted with the use of WS-03 and WS-04 leads, if at least one out of three measurements has ended with current limitation (LIMIT \$ymbol is displayed), the results of remaining measurements may be burdened with additional uncertainty.

Phase sequence

  • Phase sequence indicator: correct, incorrect
  • Mains voltage range U _L-L : 95...500 V (45...65 Hz)
    • Display of phase-to-phase voltages

10.2 Other technical data

a) type of insulation acc. to EN 61010-1 and IEC 61557.....double
b) measurement category acc. to EN 61010-1 IV 300V (III 600V)
c) degree of housing protection acc. to EN 60529....IP54
d) power supply of the meter.... alkaline batteries 4x1,5 V LR14 (C) or accumulator package SONEL NiMH 4,8 V 4,2 Ah
e) parameters of AC adapter for the battery charge .... 100 V...240 V, 50 Hz...60 Hz
f) dimensions....288 x 223 x 75 mm
g) weight of the meter....about 2,2 kg
h) storage temperature -20+70^
i) working temperature....0...+50°C
j) temperature range suitable for initiating battery charging....+10°C to +40°C
k) temperatures at which loading is interrupted......below +5°C and above (or equal to) +50°C
I) humidity....20%..90%
m) nominal temperature....+23 ± 2°C
n) reference humidity....40%..60%
o) altitude (above sea level)....<2000m
p) time till automatic shutdown when idle (Auto-OFF) 120 seconds
q) number of measurements Z or RCD (for alkaline batteries) ...>3000 (2 measurements per minute)
r) number of measurements R_ISO or R (for alkaline batteries) .....>2000

s) display......LCD, segment-type
t) memory of measurement results....990 cells, 57,500 results
u) data transmission ....USB
v) quality standard development, design and manufacturing are ISO 9001, ISO 14001, ISO 45001 compliant
w) the device meets the requirements of IEC 61557 standard
x) the product meets EMC requirements (immunity for industrial environment) according to the following standards.... EN 61326-1 and EN 61326-2-2

10.3 Additional data

Data on additional uncertainties are useful mainly when the meter is used in non-standard conditions and for metrological laboratories for the purpose of calibration.

10.3.1 Additional uncertainties according to IEC 61557-2 (R _ISO )

Significant parameterDesignationAdditional uncertainty
Position E_1 0%
Supply voltage E_2 0% (BAT is not lit)
Temperature 0...35°C E_3 2%

10.3.2 Additional uncertainties according to IEC 61557-3 (Z)

Significant parameterDesignationAdditional uncertainty
Position E_1 0%
Supply voltage E_2 0% (BAT is not lit)
Temperature 0...35°C E_3 1,2 m lead – 0Ω5 m lead – 0.011Ω10 m lead – 0.019Ω20 m lead – 0.035ΩWS-03, WS-04 lead – 0.015Ω
Phase angle 0..30°C at the bot-tom of test range E_6.2 0.6%
Frequency 99%..101% E_7 0%
Network voltage 85%..110% E_8 0%
Harmonic E_9 0%
DC component E_10 0%

10.3.3 Additional uncertainties according to IEC 61557-4 (R ±200 mA)

Significant parameterDesignationAdditional uncertainty
Position E_1 0%
Supply voltage E_2 0.5% (BAT is not lit)
Temperature 0...35°C E_3 1.5%

10.3.4 Additional uncertainties according to IEC 61557-5 (R _E )

Significant parameterDesignationAdditional uncertainty
Position E_1 0%
Supply voltage E_2 0% (BAT is not lit)
Temperature 0...35°C E_3 ±0.25 digit/°C for 50 V±0.33 digit/°C for 25 V
Serial interference voltage E_4 1%, generally according to the be-low formulas
Resistance of electrodes E_5 2% generally according to the be-low formulas and the diagram
Frequency 99%..101% E_7 0%
Network voltage 85%..110% E_8 0%

Additional uncertainty caused by serial interference voltage

REAdditional uncertainty [Ω]
0.00...9.99 Ω±((0.01RE+ 0.012)UZ+ 0.003 UZ2)
10.0...99.9 Ω±((0.001RE+ 0.05)UZ+ 0.001 UZ2)
100 Ω...1.99 kΩ±((0.001RE+ 0.5)UZ+ 0.001 UZ2)

Additional uncertainty caused by resistance of electrodes:

_dod = ± ( _S100000 + R_S · 150 + _H · 004,0R_E + ·s^-8 · R_H^2 ) [%] Rs < 5 kΩ
_dod = ± ( 7,5 + _H · 004,0R_E + ·s^-8 · R_H^2 ) [%] Rs ≥ 5 kΩ

Additional error caused by resistance of electrodes [%]
Sonel MPI-520 - Additional uncertainties according to IEC 61557-5 (R _E ) - 1

line | Resistance of voltage electrode Rs[Ω] | R_H/R_E = 0.3 | R_H/R_E = 1 | R_H/R_E = 3 | R_H/R_E = 10 | R_H/R_E = 30 | R_H/R_E = 100 | R_H/R_E = 300 | R_H/R_E = 1000 | R_H/R_E = 3000 | | -------------------------------------- | ------------- | ----------- | ----------- | ------------ | ------------ | ------------- | ------------- | -------------- | -------------- | | 1 | ~0.001 | ~0.01 | ~0.1 | ~1 | ~10 | ~100 | ~1000 | ~10000 | ~10000 | | 10 | ~0.01 | ~0.1 | ~1 | ~1 | ~10 | ~100 | ~1000 | ~10000 | ~10000 | | 100 | ~0.1 | ~1 | ~10 | ~1 | ~10 | ~100 | ~1000 | ~10000 | ~10000 | | 1000 | ~1 | ~10 | ~10 | ~1 | ~10 | ~100 | ~1000 | ~10000 | ~10000 | | 10000 | ~1 | ~10 | ~10 | ~1 | ~10 | ~100 | ~1000 | ~10000 | ~10000 | | 100000 | ~1 | ~10 | ~10 | ~1 | ~10 | ~100 | ~1000 | ~10000 | ~10000 |

10.3.5 Additional uncertainties according to IEC 61557-6 (RCD)

I_A, t_A, U_B

Significant parameterDesignationAdditional uncertainty
Position E_1 0%
Supply voltage E_2 0% (BAT is not lit)
Temperature 0...35°C E_3 0%
Resistance of electrodes E_5 0%
Network voltage 85%..110% E_8 0%

11 Positions of the meter's cover

The movable cover enables using the meter in various positions.

1 2 3

1 – Cover as the bottom of the meter

2 – Cover used as a support

3 – Cover in the position that enables convenient use of the meter suspended on the neck by means of hanging straps

12 Manufacturer

The manufacturer of the device, which also provides guarantee and post-guarantee service is the following company:

SONEL S.A.

Wokulskiego 11

58-100 Świdnica

Poland

tel. +48 74 884 10 53 (Customer Service)

e-mail: customerservice@sonel.com

web page: www.sonel.com

Attention:

Service repairs must be realised solely by the manufacturer.

NOTES

WARNING AND GENERAL DATA DISPLAYED BY THE METER

WARNING!

The MPI-520 meter is designed for operation at rated phase voltages of 110V, 115V, 127V, 220V, 230V, and 240V as well as at phase-to-phase voltages of 190V, 200V, 220V, 380V, 400V, and 415V.

Any voltage that exceeds the admissible voltage must not be applied to any measuring terminals. Failure to observe this warning may result in damage to the device and cause danger to users.

L-N! U_L-N voltage is incompatible with measuring procedure.
L-PE! U_L-PE voltage is incompatible with measuring procedure.
N-PE! U_N-PE voltage exceeds the admissible level of 50V.
Phase connected to N terminal instead of L terminal.
Sonel MPI-520 - WARNING! - 1Exceeded temperature.
Dusn1Mains frequency exceeds the 45 ... 65Hz range.
Error measureCannot display correct result.
Sonel MPI-520 - WARNING! - 2Send meter to service centre.
Sonel MPI-520 - WARNING! - 3No U_L-N voltage prior to the main measurement.
Dusn2Measurement has been stopped with ESC key.
conti Sonel MPI-520 - WARNING! - 4Voltage on measuring terminals exceeds 500V prior to the measurement.
conti Sonel MPI-520 - WARNING! - 5Voltage on measuring terminals exceeds 50V; measurement of R_E is locked.
[H472]Voltage on measuring terminals exceeds 24V but does not reach 50V; measurement of R_E is locked.
[644Z]Uncertainty of R_E measurement from electrode resistance >30%.
[273D]Discontinuity in R_E measuring circuit or probe resistance exceeds 60kΩ.
and continuous beep signalVoltage between touch electrode and PE exceeds the admissible U_L threshold value.
Dusn3RCD failure if seen at the right-hand side of the result.
[K274]Presence of measuring voltage on meter terminals at measurement of R_ISO .
Dusn4Excessive signal interferences. Measurement may be distorted by additional variance.
Dusn5Initiation of current constraints at measurements of R_ISO .
[CKYW]Unsuitable accessories connected to measuring socket ( R_ISO ).
Sonel MPI-520 - WARNING! - 6WS-03 or WS-04 wire connected to three-conductor measurements of R_ISO .
Sonel MPI-520 - WARNING! - 7Condition of batteries or accumulators:Batteries or accumulators are chargedBatteries or accumulators are dischargedBatteries or accumulators are out of use
Sonel MPI-520 - WARNING! - 8Batteries or accumulators are out of use.Install new batteries or charge the accumulators.

SONEL S.A.

Wokulskiego 11

58-100 Świdnica

Poland

Customer Service

tel. +48 74 884 10 53

e-mail: customerservice@sonel.com

www.sonel.com

Table of contents Click a title to access it
Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Sonel

Model : MPI-520

Category : Measurement