Sonel

MRU-11 - Measurement Sonel - Free user manual and instructions

Find the device manual for free MRU-11 Sonel in PDF.

📄 36 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice Sonel MRU-11 - page 6
Pick your language and provide your email: we'll send you a specifically translated version.
Product TypeEarth Resistance Meter
ModelMRU-11
BrandSonel
Dimensions221 x 102 x 62 mm (without test leads)
Weight660 g (with batteries)
Power Supply4 x AA alkaline or NiMH rechargeable batteries
Measurement Methods2-pole, 3-pole, 4-pole (Wenner) earth resistance; soil resistivity
Measuring Range (Earth Resistance)0.01 Ω to 9999 Ω (depending on method)
Soil Resistivity Range0.00 Ωm to 999 kΩm (with distance L 1-50 m or 1-150 ft)
Interference Voltage Measurement0 to 100 V RMS, accuracy ±(10% + 1 digit)
Test Voltage25 V or 50 V (selectable)
Test Frequency125 Hz (50 Hz networks) / 150 Hz (60 Hz networks)
Measuring Current>20 mA (short circuit)
Auto-Off Time300 s, 600 s, 900 s, or off (selectable)
Ingress ProtectionIP67
Safety CategoryCAT III 300 V (up to 2000 m), double insulation
Operating Temperature-10°C to +50°C
Storage Temperature-20°C to +60°C
DisplayLCD with backlight
Battery MonitoringSymbol indication; low battery warning
MaintenanceClean with damp cloth; no user-serviceable parts inside
Standards MetEN 61010-1, EN 61557-1/-5, EN 61326-1/-2-2
Accessories IncludedTest leads, auxiliary electrodes, reels (per package)
WarrantyManufacturer guarantee (standard terms)

Frequently Asked Questions - MRU-11 Sonel

How do I measure earth resistance with the 3-pole method?
Disconnect the tested earthing from installation. Connect the current electrode (H) to H socket, voltage electrode (S) to S socket, and the earthing under test to E socket. Press START. The main result R_E appears at bottom; use UP button to view R_H, R_S, ER, and U_N values.
What do the symbols 'NOISE!' and 'LIMIT!' mean?
NOISE! indicates high interference voltage (>24 V but <40 V) affecting accuracy. LIMIT! indicates resistance of auxiliary electrodes >30% (ER shown) or >19.9 kΩ (RH/RS shown), making measurement unreliable.
How can I set the auto-off time?
Enter meter configuration by holding DOWN and pressing ON/OFF. Use UP/DOWN to navigate to 'oFF' screen, then set auto-off time to 300 s, 600 s, 900 s, or '-' to disable. Press START to confirm.
What type of batteries does the MRU-11 use?
The meter uses 4 x AA alkaline batteries or 4 x AA NiMH rechargeable batteries. Rechargeable batteries must be charged in an external charger.
Can I measure earth resistance if the interference voltage exceeds 24 V?
No. Measurement is blocked if U_N >24 V. For voltages between 24 V and 40 V, the display shows 'U_N xxV' and blocks measurement. Above 40 V, a continuous beep sounds. Do not connect to voltages above 100 V.
How do I perform a soil resistivity measurement?
Connect four equally spaced auxiliary electrodes in a line: E, ES, S, H. Set the distance L between electrodes in configuration (meters or feet). Press START to measure; the meter automatically calculates resistivity ρ = 2π L R_E.
What should I do if the battery symbol is empty?
Replace all batteries or rechargeable NiMH cells with fresh ones. If 'bAtt' message appears, the batteries are fully discharged and measurements are blocked; replace immediately.
How can I clean and maintain the MRU-11?
Wipe the casing with a soft, damp cloth using mild detergent. Clean auxiliary electrodes with water and dry; lubricate before prolonged storage. Test leads can be cleaned with water and detergent, then dried. No internal maintenance required.
What is the difference between 2-pole and 3-pole measurement?
2-pole method (R_E2P) uses two leads and requires a known reference earth; result is sum of measured and known resistances. 3-pole method (R_E3P) uses two auxiliary electrodes for a true earth resistance value, eliminating reference dependency.
How do I interpret the additional results after a measurement?
After pressing START, the main result (R_E for 3/4-pole or R_E2P for 2-pole) is shown. Press UP to cycle through: R_H (current electrode resistance), R_S (voltage electrode resistance), ER (additional uncertainty in %), and U_N (interference voltage).

User questions about MRU-11 Sonel

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Measurement in PDF format for free! Find your manual MRU-11 - Sonel and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. MRU-11 by Sonel.

USER MANUAL MRU-11 Sonel

natural_image Line drawing of a handheld electronic device with control buttons and display screen (no text or symbols)

USER MANUAL

Symbols of the measurement sockets SET ES S H Additional measurement results ERUN <00.00 kΩ% RHS >00.00 Vft Warnings RE2P3P4P <00.00 s NOISE! LIMIT! VHz Symbol indicating battery charge level Active measure- ment mode Main display field for: • measurement result • setting values

USER MANUAL

EARTH RESISTANCE METER MRU-11

CE

SONEL S.A.

Wokulskiego 11

58-100 Świdnica

Poland

Version 1.07 11.07.2023

The MRU-11 meter is a modern, easy and safe measuring device. Please acquaint yourself with this manual in order to avoid measuring errors and prevent possible problems in operation of the meter.

CONTENTS

1 Safety 4
2 Turning the meter ON and activating screen backlight 5
3 Meter configuration 6
4 Measurements....9

4.1 Measurement of interference voltages DC + AC 9
4.2 Measurement of earth resistances with 3-pole method (R E 3P)....10
4.3 Measurement of earth resistances with 4-lead method (R
E 4P)....14
4.4 Measurement of earth resistances with 2-pole method (R E2P).... 18
4.5 Soil resistivity measurement (ρ) 20

5 Power supply....23

5.1 Monitoring the power supply voltage.... 23
5.2 Replacing (rechargeable) batteries 24
5.3 General principles regarding using Ni-MH rechargeable batteries 25

6 Cleaning and maintenance 25

7 Storage....25
8 Dismantling and disposal 26
9 Technical data....26

9.1 Basic data 26
9.2 Other technical data....27
9.3 Additional data 28

9.3.1 Influence of serial interference voltage on the resistance measurements for function R_E3P , R_E4P , .....28
9.3.2 Influence of the auxiliary electrodes on earth resistance measurements for function R_E3P , R_E4P , .....28
9.3.3 Additional uncertainties according to IEC 61557-5 (BP)....28

10 Manufacturer 29

1 Safety

The following international symbols are used in the Analyzer and in this manual:

Sonel MRU-11 - Safety - 1Warning;See explanation in the manualSonel MRU-11 - Safety - 2Dual insulation(Protection class II)
Sonel MRU-11 - Safety - 3Do not dispose of with other household wasteSonel MRU-11 - Safety - 4Declaration of Conformity with EU directives (Conformité Européenne)

MRU-11 meter is designed for measuring parameters important for safety of electrical installations. Therefore, in order to provide conditions for correct operation and accuracy of obtained results, the following recommendations must be observed:

  • Before you proceed to operate the meter, acquaint yourself thoroughly with the present manual and observe the safety regulations and specifications provided by the producer.
  • MRU-11 meter is designed to measure earth resistance values. Any application that differs from those specified in the present manual may result in a damage to the device and constitute a source of danger for the user.
  • The meter must be operated solely by appropriately qualified personnel members holding required certificates for carrying measurements in electric installations. Unauthorized use of the meter may result in its damage and may be a source of serious hazard to the user.
  • Using this manual does not exclude the need to comply with occupational health and safety regulations and with other relevant fire regulations required during the performance of a particular type of work. Before starting the work with the device in special environments, e.g. potentially fire-risk/explosive environment, it is necessary to consult it with the person responsible for health and safety.

- It is unacceptable to operate the device when:

⇒ a damaged meter which is completely or partially out of order,

⇒ a meter with damaged insulation,

⇒ a meter stored for an excessive period of time in disadvantageous conditions (e.g. excessive humidity). If the meter has been transferred from a cool to a warm environment with a high level of relative humidity, do not start measurements until the meter is warmed up to the ambient temperature (approximately 30 minutes).

- Before measurement make sure that test leads are connected to appropriate measuring terminals.

- Do not power the meter from sources other than those listed in this manual.

- The inputs of the meter are protected electronically against overload e.g. due to having been connected to a live circuit, for all combinations of inputs - up to 276 V for 30 seconds.

- Factory calibration does not include the resistance of the test leads. The result displayed by the meter is a sum of the resistance of the measured object and the resistance of leads.

- The device meets the requirements of standards EN 61010-1 and EN 61557-1, -5.

Sonel MRU-11 - Safety - 5

Due to continuous product development, the manufacturer reserves the right to introduce changes to the functionality, appearance, accessories and technical data of the meter. Due to continuous development of the meter's software, the actual appearance of the display, in case of some of the functions, may slightly differ from the display presented in this operating manual.

2 Turning the meter ON and activating screen backlight


Sonel MRU-11 - Turning the meter ON and activating screen backlight - 1

To turn ON the meter, press ON/OFF button.

100

The screen of the device will activate its all segments (self-test), and then it will briefly show the software version.


Sonel MRU-11 - Turning the meter ON and activating screen backlight - 3

When the meter is ON, a short press on ON/OFF button will turn ON and then OFF the screen backlight;


Sonel MRU-11 - Turning the meter ON and activating screen backlight - 4

To turn OFF the device, press and hold ON/OFF button for approx. 2 seconds.

off

A screen indicating shutting off the device is shown briefly.

3 Meter configuration


Sonel MRU-11 - Meter configuration - 1

Turn on the meter by pressing and holding DOWN button and pressing ON/OFF button.

SET Fn 50 Hz

After switching the meter ON, a screen is shown with network frequency settings Fn.


Sonel MRU-11 - Meter configuration - 3

When the Fn screen is displayed, use UP and DOWN buttons to set the network frequency of 50 Hz or 60 Hz (default: 50 Hz).


Sonel MRU-11 - Meter configuration - 4

Short press START button to accept the selected value.

SET BEEP ON

You will enter the screen with settings of audio messages - bEEP.


Sonel MRU-11 - Meter configuration - 6

Use UP and DOWN buttons to turn the audio messages ON (or OFF ( ) off


Sonel MRU-11 - Meter configuration - 7

Short press START button to accept the selected option.

SET OFF 300 s

The screen of Auto-OFF time settings will be shown: oFF.


Sonel MRU-11 - Meter configuration - 9

Use UP and DOWN buttons to set the auto-off time at 300 s, 600 s, 900 s or "-" (Auto-OFF inactive). Auto-OFF function is used to turn-off inactive meter after a preselected time, which is signalled by a beep.


Sonel MRU-11 - Meter configuration - 10

Short press START button to accept the selected option.

SET Un 50v

You will enter the screen with measuring voltage selection Un.


Sonel MRU-11 - Meter configuration - 12

Use UP and DOWN buttons to set the measuring voltage at 25 V or 50 V. The measuring voltage set relates to all measurement functions in the meter.

9 START SET ρ m

Short press START button to accept the selected option.

You will enter the screen with settings of units soil resistivity.

Sonel MRU-11 - Meter configuration - 14

natural_image Two circular buttons with white arrows, one numbered 10 (no text or symbols on buttons)

Use UP and DOWN buttons to set the unit:

⇒ meters m or

⇒ feet ft.

Sonel MRU-11 - Meter configuration - 15

Short press START button to accept the selected option.

SET 4 m ε

You will enter the screen with settings of distance between auxiliary electrodes.

Sonel MRU-11 - Meter configuration - 17

Use UP and DOWN buttons to set the distance.

Sonel MRU-11 - Meter configuration - 18

Short press START button to accept the selected option and return to the network frequency selection screen Fn.

Sonel MRU-11 - Meter configuration - 19

Long press (> 2 s) START button to the introduced changes and proceed to the measurement screen RE3P.

Sonel MRU-11 - Meter configuration - 20

Long press (> 2 s) ON/OFF button to turn off the meter without accepting the changes introduced at the current setting position.

Sonel MRU-11 - Meter configuration - 21

Short press (> 2 s) ON/OFF button to enter the measurement screen R_E 3P without accepting the changes introduced at the current setting position.

4 Measurements

Sonel MRU-11 - Measurements - 1

Earth resistance measurements significantly different from other measurements performed to assess the protection against electric shock. They require a thorough knowledge of the structure of earthing system, the phenomena occurring during the measurements and the skills of coping with severe outdoor conditions. Earthing system tests/measurements require adequate knowledge and curing equipment, which will be able to maximally facilitate these examinations.

4.1 Measurement of interference voltages DC + AC

E S U_N V R_E2P

In measuring functions, before START button is pressed, the meter monitors the measuring voltage at the terminals (between E socket and S / H sockets) and the interference voltage value is displayed on the screen.

Additional information displayed by the meter

U_N > 100V!, > 100V and a continuous beep ,

NOISE! and

Sonel MRU-11 - Additional information displayed by the meter - 1

Voltage at test terminals is higher than 100 V, the measurement is blocked.

UN xxV!, >40V and a continuous

beep

NOISE! and

Sonel MRU-11 - Additional information displayed by the meter - 2

Where xx is the value of interference voltage. Voltage at test terminals is higher than 40 V but lower than 100 V, the measurement is blocked.

UNxxV!, >24V, NOISE! and

Sonel MRU-11 - Additional information displayed by the meter - 3

Where xx is the value of interference voltage. Voltage at test terminals is higher than 24 V but lower than 40 V, the measurement is blocked.

NOISE!

The interference voltage is lower than 24 V, but with a high value - the measurement result may be affected by additional uncertainty.

4.2 Measurement of earth resistances with 3-pole method (R_E3P)

Sonel MRU-11 - Measurement of earth resistances with 3-pole method (R_E3P) - 1

For earth resistance measurements, the most commonly used method is the 3-pole method, often called the potential drop method, or technical method. During the measurement, the voltage drop at the earthing is measured with current flowing through it, then the Ohm's law is used to calculate the resistance.

Tested earthing system Voltage probe Current probe

The scheme of measuring the earthing resistance with the 'technical method' is shown above. The figure shows the measurement of earthing resistance R E. To perform the measurement, use two additional auxiliary electrodes:

  • H electrode (current electrode) to allow excitation of current flow in the circuit: tested earth electrode R_E meter H current electrode earth tested earth electrode;
  • S electrode (voltage electrode) for measuring the voltage drop across the resistance of the measured earthing as a result of current flow.

Sonel MRU-11 - Measurement of earth resistances with 3-pole method (R_E3P) - 3

Disconnect the tested earth electrode from the installation of the facility.

Technical diagram of a house with grounding and electrical components, including a close-up view of a wall-mounted switch and bolt assembly.


Sonel MRU-11 - Measurement of earth resistances with 3-pole method (R_E3P) - 5

Turn on the meter using ON/OFF button.

E S H U_N V R_E 3P

The meter enters the measurement function screen Re3P.

The meter is in the mode of measuring the interference voltage between the test terminals. The measuring voltage is compatible with the voltage selected when setting up the device.


Diagram showing electrical grounding setup with a house, power source, and connected components labeled E, R_E, S, H.

Test leads should be connected to the measurement terminals in the device, as shown above.

- The current electrode H (driven into earth) should be connected to H socket of the meter.

- The voltage electrode S (driven into earth) should be connected to S socket of the meter.

- The earth electrode being tested should be connected to E socket of the meter with the lead.

- The earth electrode being tested and the current electrode and the voltage electrode should be located in one line.


Sonel MRU-11 - Measurement of earth resistances with 3-pole method (R_E3P) - 8

Press START.

The progress of the measurement is indicated by horizontal lines on the screen.


E S H R_H 6.89 Ω R_E 3P 400 Ω

After completing the measurement, results are displayed for all the measurements carried-out: at the bottom of the screen, the main result is shown for R_E , whereas the upper part of the screen shows additional results for R_H . The result is displayed for 20 seconds. The result can be recalled by pressing UP button.


R_S 6.88 Ω ER 5 % U_N |v

Use UP button to view the component results in the following order: R_H R_S ER U_N , where:

R_H – resistance of electrode H R_S – resistance of electrode S ER – additional uncertainty resulting from the electrode UN - interference voltage


X X S

Repeat the measurements (steps 456) moving the voltage electrode a few meters - placing it farther and closer to the measured earth electrode. If R_E measurement results differ from one another by more than 3%, the distance of the current electrode from the earth electrode being tested should be considerably increased and the measurements should be repeated.

Sonel MRU-11 - Measurement of earth resistances with 3-pole method (R_E3P) - 12

NOTE!

Measurement of resistance-to-earth may be carried out if voltage of interferences does not exceed 24 V. Voltage of interferences is measured up to the level of 100 V, but above 40 V it is signalled as dangerous. The meter must not be connected to voltages exceeding 100 V.

Sonel MRU-11 - NOTE! - 1

  • Particular attention should be paid to quality of connection between the object being tested and the test lead – the contact area must be free from paint, rust, etc.
  • If resistance of electrodes is too high, R_E earth electrode measurement will include an additional uncertainty. Particularly high uncertainty of measurement occurs when a small value of resistance-to-earth is measured with electrodes that have a weak contact with earth (such a situation occurs frequently when the earth electrode is well made and the upper soil layer is dry and slightly conductive). In such a case, the ratio of resistance of the electrodes to resistance of the tested earth electrode is very high and consequently, uncertainty of measurement that depends on this ratio is also very high. Then, you can make a calculations according to the formulas given in sec. 9.3 to estimate the influence of measurement conditions. You can improve the contact between the electrode and soil, for example, by dampening with water the place where the electrode is driven into earth, driving the probe into earth in a different place or using a 80 cm-long electrode. Check also the test leads for possible insulation damage and for corroded or loosened connection between the banana plug and the test lead. In majority of cases the achieved measurement accuracy is satisfactory. However, you should always be aware of the uncertainty included in the measurement.

Additional information displayed by the meter

RE>9999 ΩMeasuring range is exceeded.
UN>100 V, >100 Vand a continuous beep ⬆,NOISE! and [IMAGE]Voltage at test terminals is higher than 100 V, measurement is blocked.
UN xxV, >40 Vand a continuous beep ⬆,NOISE! and [IMAGE]Where xx is the value of interference voltage. Voltage at test terminals is higher than 40 V, measurement is blocked.
UN xxV, >24 V,NOISE! and [IMAGE]Where xx is the value of interference voltage. Voltage at test terminals is higher than 24 V but lower than 40 V, the measurement is blocked.
NOISE!The interfering signal (noise signal) is too high - the measurement result may be affected by additional uncertainty.
LIMIT!and ER together with the value expressed in %Measurement due to the resistance of the auxiliary electrodes >30%. (Measured values are used in calculation of uncertainty.)
LIMIT!and RH or Rs with the value of ΩResistance of auxiliary electrodes H and S, or one of them exceeds 19.9 kΩ, correct measurement is not possible.
Flashing edgesSonel MRU-11 - Additional information displayed by the meter - 1Flashing edges of symbols: E or H or S, two or all three at the same time: disconnected one, two or three leads to the terminals, or the resistance of the auxiliary auxiliary electrode/s is outside the measuring range.

4.3 Measurement of earth resistances with 4-lead method ( R_E4P )

Sonel MRU-11 - Measurement of earth resistances with 4-lead method ( R_E4P ) - 1

The four-lead method is recommended in the case of measurements of earth resistance of very low values. It permits to eliminate the influence of the test leads resistance over the result of the measurement. In order to evaluate the resistivity of the ground it is recommended to use the dedicated measurement function.


Disconnect the tested earth electrode from the installation of the facility.
Technical diagram showing a house with grounding and a close-up of electrical components including resistors and bolts.

Sonel MRU-11 - Measurement of earth resistances with 4-lead method ( R_E4P ) - 3

E ES S H U_N V R_E 4P

Turn on the meter using ON/OFF button.

Press DOWN button until the screen of Re4P measurement appears.

The meter is in the mode of measuring the interference voltage between the test terminals. The measuring voltage is compatible with the voltage selected when setting up the device.


Diagram of an electrical grounding system with labeled components including a house, power source, and terminal blocks S and H.

Test leads should be connected to the measurement terminals in the device, as shown above.

- Connect the current electrode driven into ground to the H socket of the meter.

- Connect the voltage electrode driven into ground to the S socket of the meter.

- Connect the tested earth electrode to the E socket of the meter.

- Connect the ES socket to the earth electrode In question below the E cable.

- The tested earth electrode as well as the current electrode and voltage electrode should be aligned.


Sonel MRU-11 - Measurement of earth resistances with 4-lead method ( R_E4P ) - 6

Press START.

The progress of the measurement is indicated by horizontal lines on the screen.


E ES S H R_H 6.09 Ω R_E 4P 1.13 Ω

After completing the measurement, results are displayed for all the measurements carried-out: at the bottom of the screen, the main result is shown for R_E , whereas the upper part of the screen shows additional results for R_H . The result is displayed for 20 seconds. The result can be recalled by pressing UP button.

⑥ R_S 688 Ω ER 5 % U_N |v

⑦ x x S

Use UP button to view the component results in the following order:

R_H R_S ER U_N , where:

RH – resistance of electrode H

Rs - resistance of electrode S

ER – additional uncertainty resulting from the electrodes

UN - interference voltage

Repeat the measurements (steps) 456 moving the voltage electrode a few meters - placing it farther and closer to the measured earth electrode.

If RE measurement results differ from one another by more than 3%, the distance of the current electrode from the earth electrode being tested should be considerably increased and the measurements should be repeated.

Sonel MRU-11 - Measurement of earth resistances with 4-lead method ( R_E4P ) - 10

NOTE!

Measurement of resistance-to-earth may be carried out if voltage of interferences does not exceed 24 V. Voltage of interferences is measured up to the level of 100 V, but above 40 V it is signalled as dangerous. The meter must not be connected to voltages exceeding 100 V.

Sonel MRU-11 - NOTE! - 1

- Particular attention should be paid to quality of connection between the object being tested and the test lead – the contact area must be free from paint, rust, etc.

- If resistance of electrodes is too high, R_E earth electrode measurement will include an additional uncertainty. Particularly high uncertainty of measurement occurs when a small value of resistance-to-earth is measured with electrodes that have a weak contact with earth (such a situation occurs frequently when the earth electrode is well made and the upper soil layer is dry and slightly conductive). In such a case, the ratio of resistance of the electrodes to resistance of the tested earth electrode is very high and consequently, uncertainty of measurement that depends on this ratio is also very high. Then, you can make a calculations according to the formulas given in sec. 9.3 to estimate the influence of measurement conditions. You can improve the contact between the electrode and soil, for example, by dampening with water the place where the electrode is driven into earth, driving the probe into earth in a different place or using a 80 cm-long electrode. Check also the test leads for possible insulation damage and for corroded or loosened connection between the banana plug and the test lead. In majority of cases the achieved measurement accuracy is satisfactory. However, you should always be aware of the uncertainty included in the measurement.

Additional information displayed by the meter

RE>9999ΩMeasurement range exceeded.
UN>100V, >100V and a continuous sonic signal←,NOISE! and Sonel MRU-11 - NOTE! - 2The voltage on the measurement points exceeds 100 V, the measurement is blocked.
UN xxV, >40V and a continuous sonic signal←,NOISE! and Sonel MRU-11 - NOTE! - 3Where xx is the value of interfering voltage. The voltage on the measurement points exceeds 40 V, the measurement is blocked.
UN xxV, >24V, NOISE! and Sonel MRU-11 - NOTE! - 4Where xx is the value of interfering voltage. The voltage on the measurement points exceeds 24 V, but it's below 40 V, the measurement is blocked.
NOISE!The value of the interfering signal is below 24 V, but it has too high value, so the result may be distorted by additional uncertainty.
Sonel MRU-11 - NOTE! - 5 and ER along with % valueThe uncertainty caused by resistance of the auxiliary electrodes >30%. (Uncertainties calculated on the basis of the measured values.)
LIMIT! and RH or Rs along with Ω valueThe resistance of H and S electrodes, or one of them ex-ceeds 19.9 kΩ, the proper measurement is not possible.
Flashing symbols: Sonel MRU-11 - NOTE! - 6Flashing symbols E or H or S, or both of them, or all three at the same time, one or two or three test leads are dis-connected from the measurement sockets.

4.4 Measurement of earth resistances with 2-pole method (R_E2P)

Sonel MRU-11 - Measurement of earth resistances with 2-pole method (R_E2P) - 1

2-pole method may be also used for measuring earth resistance. When the earthing system is known and earthing is available with a known resistance value, the measurement result will be the sum of the earth resistances: measured and known.

Sonel MRU-11 - Measurement of earth resistances with 2-pole method (R_E2P) - 2

Sonel MRU-11 - Measurement of earth resistances with 2-pole method (R_E2P) - 3

Turn on the meter. After turning the meter ON, the screen with 3-pole method R3P is shown.

Sonel MRU-11 - Measurement of earth resistances with 2-pole method (R_E2P) - 4

Sonel MRU-11 - Measurement of earth resistances with 2-pole method (R_E2P) - 5

Press DOWN button until the screen of 2-pole method Re2P appears.

E S U_N V R_E2P ... ... ... ...

The meter is in the mode of measuring the interference voltage between the test terminals. The measuring voltage is compatible with the voltage selected when setting up the device.

Sonel MRU-11 - Measurement of earth resistances with 2-pole method (R_E2P) - 7

Diagram showing a simple electrical setup with a house, connected to an orange device labeled with terminals E, S, ES and a red wire labeled 'S'.

Test leads should be connected to the measurement terminals in the device, as shown above.

Sonel MRU-11 - Measurement of earth resistances with 2-pole method (R_E2P) - 9

Sonel MRU-11 - Measurement of earth resistances with 2-pole method (R_E2P) - 10

In order to start the measurement, press START button.

Sonel MRU-11 - Measurement of earth resistances with 2-pole method (R_E2P) - 11

E S UN ΩV RE2P 2.08 Ω

After the measurement, its result is shown: at the bottom of the screen, the main result is shown for RE2P, whereas the upper part of the screen shows the measured interference voltage UN. The result is displayed for 20 seconds. The result can be recalled by pressing UP button.

Additional information displayed by the meter

R > 9999 ΩMeasuring range is exceeded.
UN>100 V, >100 V and a continuous beep ↔, NOISE! andVoltage at test terminals is higher than 100 V, measurement is blocked.
UNxxV, >40 V and a continuous beep ↔, NOISE! andWhere xx is the value of interference voltage. Voltage at test terminals is higher than 40 V, measurement is blocked.
UNxxV, >24 V,NOISE! andWhere xx is the value of interference voltage. Voltage at test terminals is higher than 24 V but lower than 40 V, the measurement is blocked.
NOISE!The interfering signal (noise signal) is below 24 V, but with a high value - the measurement result may be affected by additional uncertainty.

4.5 Soil resistivity measurement (ρ)

For the purpose of earth resistivity measurements – which are used as a preliminary measure for the project of earthing systems or in geology - there is a separate function: earth resistivity measurements . The function is metrologically identical as the four-lead earth resistance measurement, but it includes an additional procedure of storing of the distance between the auxiliary electrodes. The result of the measurement is the resistance value which is calculated automatically in accordance with the following formula:

$$ \rho = 2 \pi L R _ {E} $$

which is used in the Wenner's measurement method. The method in question assumes equal distances between electrodes.

① Connect test leads according to the drawing.

E ES S H ES S H L L L

Connect the four aligned and equally spaced auxiliary electrodes, which are driven into the ground, to the meter, and do so in accordance with the diagram above.

  • Connect the current electrode driven into ground to the H socket of the meter.
  • Connect the voltage electrode driven into ground to the S socket of the meter.
  • Connect the voltage electrode driven into ground to the ES socket of the meter.
  • Connect the current electrode driven into ground to the E socket of the meter.

Sonel MRU-11 - Soil resistivity measurement (ρ) - 2

Sonel MRU-11 - Soil resistivity measurement (ρ) - 3

Turn on the meter using ON/OFF button.

Set measurement unit and distance L between electrodes according to sec. 3 steps ⑨. . ⑬

Press DOWN button until the screen of p measurement appears.


E ES S H U_N V ρ

The meter is ready for measurement.


Sonel MRU-11 - Soil resistivity measurement (ρ) - 5

Press START to commence measurement.


E ES S H U_N Ωv 270 Ωm

After finishing the measurement, read out the result. The results of all the measurements that have been carried out will be displayed on screen.


R_S 6.00 Ω ER 5 % U_N |V

Use UP button to view the component results in the following order: R_H R_S ER U_N , where:

R_H – resistance of current electrode R_S - resistance of voltage electrode ER – additional uncertainty caused by the resistance of the electrodes U_N – interfering (noise) voltage

Sonel MRU-11 - Soil resistivity measurement (ρ) - 8

NOTE!

Earth resistivity measurement may be performed if the interference voltage does not exceed 24 V. The interference voltage is measured up to 100 V, but over 40 V is it signaled as hazardous. Do not connect the meter to a voltage exceeding 100 V.

Sonel MRU-11 - NOTE! - 1

  • Calculations are based upon the assumption that the distances between the specific auxiliary electrodes are equal (the Wenner's method). If this is not the case the earthing resistance measurement must be carried out by means of the four-pole method and calculations must be performed individually.
  • Pay particular attention to the quality of the connection of the tested object with the

test leads – the contact area must be cleaned of paint, rust, etc.

- If resistance of electrodes is too high, R_E earth electrode measurement will include an additional uncertainty. Particularly high uncertainty of measurement occurs when a small value of resistance-to-earth is measured with electrodes that have a weak contact with earth (such a situation occurs frequently when the earth electrode is well made and the upper soil layer is dry and slightly conductive). In such a case, the ratio of resistance of the electrodes to resistance of the tested earth electrode is very high and consequently, uncertainty of measurement that depends on this ratio is also very high. Then, you can make a calculations according to the formulas given in sec. 9.3 to estimate the influence of measurement conditions. You can improve the contact between the electrode and soil, for example, by dampening with water the place where the electrode is driven into earth, driving the probe into earth in a different place or using a 80 cm-long electrode. Check also the test leads for possible insulation damage and for corroded or loosened connection between the banana plug and the test lead. In majority of cases the achieved measurement accuracy is satisfactory. However, you should always be aware of the uncertainty included in the measurement.

Additional information displayed by the meter

>xxxk m or >xxxk ft Measurement range exceeded, where xxx is maximum value that can be measured for the selected settings.
U_N >100V, >100V and a continuous sonic signal ,NOISE! and [IMAGE]The voltage on the measurement points exceeds 100 V, the measurement is blocked.
U_NxxV, >40V and a continuous sonic signal ,NOISE! and [IMAGE]Where xx is the value of interfering voltage. The voltage on the measurement points exceeds 40 V, the measurement is blocked.
U_NxxV, >24V, NOISE! and Where xx is the value of interfering voltage. The voltage on the measurement points exceeds 24 V, but it's below 40 V, the measurement is blocked.
NOISE!The value of the interfering signal is below 24 V, but it has too high value, so the result may be distorted by additional uncertainty.
LIMIT!and ER along with % valueThe uncertainty caused by resistance of the auxiliary electrodes >30%. (Uncertainties calculated on the basis of the measured values.)
LIMIT!and RH or Rs along with Ω valueThe resistance of H and S electrodes, or one of them ex-ceeds 19.9 kΩ, the proper measurement is not possible.
Flashing symbols:Sonel MRU-11 - NOTE! - 2Flashing symbols E or ES or H or S, or two of them, or three, or all of them at the same time – one or two or three or four test leads are disconnected from the measurement sockets.

5 Power supply

Sonel MRU-11 - Power supply - 1

Before the measurements, make sure that the status of the batteries or rechargeable batteries meter is sufficient for performing tasks related to the operation of the device.

5.1 Monitoring the power supply voltage

E S H U_N V R_E 3P

The charge level of the batteries or rechargeable batteries is indicated by the symbol in the right upper corner of the display on a current basis:

Sonel MRU-11 - Monitoring the power supply voltage - 2

When all segments of the battery symbol are lit, it means that the batteries/rechargeable batteries are fully charged.

Sonel MRU-11 - Monitoring the power supply voltage - 3

When all segments of the battery symbol are dark, it means that the batteries/rechargeable batteries are discharged and must be replaced.

bA+B

Message bAtt indicates that batteries/rechargeable batteries are totally discharged, all measurements are blocked. The meter switches off automatically after 5 sec.

5.2 Replacing (rechargeable) batteries

MRU-11 is powered by four AA alkaline LR6 batteries or rechargeable batteries of NiMH type. The (rechargeable) batteries are placed in the compartment at the bottom of the enclosure. The device is not equipped with an internal battery charger. Rechargeable batteries must be recharged in an external charger.

Sonel MRU-11 - Replacing (rechargeable) batteries - 1

NOTE!

Do not power the meter from sources other than those listed in this manual. Before replacing the (rechargeable) batteries, disconnect the test leads from the meter.

① E S E ES S H

Disconnect the unit from the object!

Technical diagram of a car body with exploded view and component labels, including numbered annotations ② and ③.

Turn off the device using ON/OFF button.

Remove the screws that secure the battery cover at the bottom of the compartment(4 pcs),

Remove all batteries (rechargeable batteries). Observe the correct polarity when inserting new batteries/rechargeable batteries.

Place and tighten the battery compartment cover.

Sonel MRU-11 - NOTE! - 3

NOTE!

Reverse polarity will not damage the meter or the batteries, but the meter will not work. Have the meter serviced in case of battery leakage inside the compartment.

5.3 General principles regarding using Ni-MH rechargeable batteries

  • Store the he rechargeable batteries (the meter) in a dry, cool and well ventilated place and protect them from direct sunlight. The temperature of the environment in the case of prolonged storage should not exceed 30^ C. If the rechargeable batteries are stored for a long time in a high temperature, then the occurring chemical processes may reduce their lifetime.
  • Rechargeable batteries NiMH usually lasts for 500-1000 charging cycles. The rechargeable batteries reach their maximum capacity after being formatted (2-3 charge and discharge cycles). The most important factor which influences the lifetime of rechargeable batteries is the level of their discharge. The deeper the discharge level of the batteries, the shorter their lifetime.
  • The memory effect is limited in the case of NiMH batteries. These batteries may be charged at any point with no serious consequences. However, it is recommended to discharge them completely every few cycles.
  • During storage of Ni-MH rechargeable batteries they are discharged at the rate of approximately 20% per month. Keeping rechargeable batteries at high temperatures may accelerate this process even 100%. In order to prevent excessive discharge of rechargeable batteries, after which it would be necessary to format them, it is recommended to charge them from time to time (even if they are not used).
  • Modern fast chargers detect both too low and too high temperature of the battery pack and react to the situation adequately. Too low temperature should prevent starting the process of charging, which might irreparably damage rechargeable batteries. An increase of the temperature of the rechargeable batteries is a signal to stop charging and is a typical phenomenon. However charging at a high ambient temperature apart from reducing batteries' lifetime causes an accelerated increase of their temperature and the result is that the batteries are not charged to their full capacity.
  • Please note that when the batteries are charged with a fast-charger they are charged only to approx. 80% of their capacity - better results can be achieved by continuing charging: the charger enters trickle-charging mode and during the next few hours batteries are charged to their full capacity.
  • Do not charge or use the batteries in extreme temperatures. Extreme temperatures reduce the lifetime of batteries and rechargeable batteries. Avoid placing devices powered by rechargeable batteries in very hot environments. The nominal working temperature must be absolutely observed.

6 Cleaning and maintenance

Sonel MRU-11 - Cleaning and maintenance - 1

NOTE!

Use only the maintenance methods specified by the manufacturer in this manual.

The casing of the meter may be cleaned with a soft, damp cloth using all-purpose detergents. Do not use any solvents or cleaning agents which might scratch the casing (powders, pastes, etc.).

Clean the auxiliary electrode with water and dry it. Before the probe is stored for a prolonged period of time it is recommended to grease it with any machine lubricant.

The reels and test leads should be cleaned with water and detergents, and then dried.

The electronic system of the meter does not require maintenance.

7 Storage

In the case of storage of the device, the following recommendations must be observed:

  • Disconnect all the test leads from the meter.
  • Clean the meter and all its accessories thoroughly.
    • Wind the long test leads onto the reels.
  • If meter is to be stored for a prolonged period of time, the batteries must be removed from the device.
  • In order to prevent a total discharge of the battery pack in the case of a prolonged storage, charge it from time to time.

8 Dismantling and disposal

Worn-out electric and electronic equipment should be gathered selectively, i.e. it must not be placed with waste of another kind.

Worn-out electronic equipment should be sent to a collection point in accordance with the law of waste electrical and electronic equipment.

Before the equipment is sent to a collection point, do not dismantle any elements.

Observe local regulations concerning disposal of packages, waste batteries and accumulators.

9 Technical data

  • The specified accuracy relates to the meter terminals.
  • The abbreviation „m.v.” in the accuracy definition means the measured value.

9.1 Basic data

Measurement of interference voltage U_N (RMS)

RangeResolutionAccuracy
0...100 V1 V±(10% m.v. + 1 digit)

• Measurement for f_N 45...65 Hz.
- The frequency of measurement - at least 2 measurements / sec.

Measurement of earth resistances - 2-pole method (R_E2P)

RangeResolutionAccuracy
0.01 Ω ... 19.99 Ω0.01 Ω±(3% m.v. + 3 digits)
20.0 Ω...199.9 Ω0.1 Ω
200 Ω...1999 Ω1 Ω±5%
2000 Ω...9999 Ω1 Ω±8%
  • Measuring current at the short circuit of >20mA .
  • Measuring frequency of 125 Hz or 150 Hz.
  • Selected test voltage: 25 V or 50 V.
  • Maximum interference voltage for the R_E measurement is 24 V.

Measurement of earth resistances - 3-pole method (RE3P), 4-wire method (RE4P)

Measurement method: 3-pole, conforming to EN 61557-5.

Measurement range according to EN 61557-5: 0.53 Ω...9999 Ω for Un = 50 V.

RangeResolutionAccuracy
0.00 Ω...19.99 Ω0.01 Ω±(3% m.v. + 3 digits)
20.0 Ω...199.9 Ω0.1 Ω
200 Ω...1999 Ω1 Ω±5%
2000 Ω...9999 Ω1 Ω±8%
  • Measuring current at the short circuit of >20mA .
  • Measuring frequency of 125 Hz or 150 Hz.
  • Selected test voltage: 25 V or 50 V.
  • Maximum interference voltage for the R_E measurement is 24 V.

Ground resistivity measurements

The measurement method: Wenner's, = 2 LR_E

RangeResolutionAccuracy
0,00..9,99 Ωm0,01 ΩmDepends on the accuracy of the R 4P measurement but not less than ±1 digit.
10,0..99,9 Ωm0,1 Ωm
100..999 Ωm1 Ωm
1,00..9,99 kΩm0,01 kΩm
10,0..99,9 kΩm0,1 kΩm
100..999 kΩm1 kΩm

• distance between auxiliary electrodes (L): 1...50 m or 1...150 ft

Measurement of resistance of auxiliary electrodes R_H and R_S

RangeResolutionAccuracy
0...999 Ω1 Ω±(5% + 8 digits)
1.00...9.99 kΩ0.01 kΩ
10.0...19.9 kΩ0.1 kΩ

9.2 Other technical data

a) type of insulation acc. to EN 61010-1 and IEC 61557.....double
b) measurement category (for 2000 m a.s.l.) acc. to EN 61010-1....III 300 V
c) degree of housing protection acc. to EN 60529 IP67
d) maximum interference of AC + DC voltages, for the measurement .....24 V
e) maximum measured voltage of interferences ....100 V
f) measuring current frequency 125 Hz for 50 Hz networks 150 Hz for 60 Hz networks
g) measuring voltage for R_E2P , R_E3P , R_E4P ....25 V or 50 V
h) measuring current (short circuit) for R E2P, RE3P, RE4P....>20 mA
i) measurement range according to EN 61557-5: 0.53 Ω...9999 Ω for Un = 50 V
j) maximum resistance of auxiliary electrodes 20 kΩ
k) meter power supply ....4 x AA alkaline battery ....or 4 x AA NiMH rechargeable battery
I) number of measurements for R E 3P ....>3000 .... (R E =10 Ω, R H =R S =100 Ω, 25 V 50 Hz, 2 measurements/minute)
m) time of performing the resistance measurement with 2-pole method ....<4 s
n) time of performing the resistance measurement with 3-pole method ....<8 s
o) time of performing the resistance measurement with 4-lead method ....<8 s
p) dimensions 221 x 102 x 62 mm (without test leads)
q) weight of the meter with batteries 660 g
r) operating temperature....-10°C..+50°C
s) reference temperature....23 ± 2°C
t) storage temperature....-20..+60°C
u) relative humidity 20.90%
v) nominal relative humidity 40.60%
w) altitude (above sea level) .... ≤2000 m*
x) quality standard ......design and manufacturing are ISO 9001 compliant
y) measurement method ...... technical, conforming to EN 61557-5
z) the device meets the EMC requirements according to: ...... EN 61326-1 and EN 61326-2-2

NOTE

\* Information about the use of meter at altitude from 2000 m to 5000 m

As for voltage inputs E, ES, S, H the instrument is to be considered downgraded to measurement category CAT III 150 V to ground (max 150 V between inputs) or CAT IV 100 V to ground (max 100 V between inputs). Markings and symbols indicated on the instrument are to be considered valid when using it at altitude lower than 2000 m.

9.3 Additional data

Data on additional uncertainties are useful mainly when the meter is used in non-standard conditions and for metrological laboratories for the purpose of calibration.

9.3.1 Influence of serial interference voltage on the resistance measurements for function RE3P, RE4P, ρ

R_E U_N Additional uncertainty [Ω]
0.00 ... 10.00 Ω25 V±(0.001 R_E +0.01) U_Z +0.007 U_Z^2
50 V±(0.001 R_E +0.01) U_Z +0.004 U_Z^2
10.01..2000 Ω25 V, 50 V±(0.001 R_E +0.01) U_Z +0.001 U_Z^2
2001..9999 Ω25 V, 50 V±(0.003 R_E + 0.4) U_Z

9.3.2 Influence of the auxiliary electrodes on earth resistance measurements for function R_E3P , R_E4P ,

R_H, Rs Additional uncertainty [%]
R_H ≤ 5 k and Rs ≤ 5 k ± (_SR_S + 100000 · 150 + _H · 004.0R_E + ·s^-8 · R_H^2)_1
R_H > 5 k or Rs > 5 k or R_H and Rs > 5 k ± 5(7_H · 004.0R_E + ·s^-8 · R_H^2)_1

R_E[], R_S[] and R_H[] are values displayed by the device.

9.3.3 Additional uncertainties according to IEC 61557-5 ( R_E 3P)

Significant parameterDesignationAdditional uncertainty
Position E_1 0%
Supply voltage E_2 0% (BAT is not lit)
Temperature E_3 ±0.2 digit/°C for R<1 kΩ ±0.07%/°C ±0.2 digit/°C for R≥1 kΩ
Serial interference voltage E_4 According to the formulas shown in par. 9.2.1 ( U_N =3V 50/60Hz)
Resistance of earth contact probes E_5 According to the formula in par. 9.2.2

10 Manufacturer

The manufacturer of the device and provider of guarantee and post-guarantee service:

SONEL S.A.

Wokulskiego 11

58-100 Świdnica

Poland

tel. +48 74 884 10 53 (Customer Service)

e-mail: customerservice@sonel.com

web page: www.sonel.com

Sonel MRU-11 - Manufacturer - 1

NOTE!

Service repairs must be performed only by the manufacturer.

NOTES

NOTES

NOTES

SONEL S.A.

Wokulskiego 11

58-100 Świdnica

Poland

Customer Service

tel.+48 74 884 10 53

e-mail: customerservice@sonel.com

www.sonel.com

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Sonel

Model : MRU-11

Category : Measurement