IMPACT Pro R - Hearing aid SIEMENS - Free user manual and instructions
Find the device manual for free IMPACT Pro R SIEMENS in PDF.
| Product Type | Behind-the-ear hearing aid |
| Model | IMPACT Pro R |
| Brand | Siemens |
| Number of Hearing Programs | Up to 5 |
| Program Selection | Push button with beep tones |
| Battery Type | Size 312 or 13 (typical) |
| Battery Life | Approximately 5-7 days (typical) |
| Low Battery Indicator | Intermittent beeps or reduced amplification |
| Wireless Technology | e2e wireless 2.0 (binaural), optional miniTek/Tek connectivity |
| Ear Pieces | Click Dome and Click Mould (replaceable) |
| Receiver Unit | Replaceable, with wax guard (HF-4) |
| Wax Protection | HF-4 Wax Guard and Click Mould wax filter |
| Microphone | With protective cover |
| Power-On Delay | Optional (prevents feedback on insertion) |
| Cleaning | Daily cleaning recommended; Click Domes washable; replace every 3 months |
| Safety Precautions | Choking hazard from small parts; avoid extreme heat, humidity, strong magnetic fields, X-rays |
| Certifications | CE mark, compliant with EU directives 93/42 EEC and 99/5/EC (wireless) |
| Warranty | Refer to local NHS Audiologist |
Frequently Asked Questions - IMPACT Pro R SIEMENS
User questions about IMPACT Pro R SIEMENS
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Hearing aid in PDF format for free! Find your manual IMPACT Pro R - SIEMENS and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. IMPACT Pro R by SIEMENS.
USER MANUAL IMPACT Pro R SIEMENS
natural_image
Medical device with attached white cable and connector, no visible text or symbolsIMPACTProR
Instructions and guidance
SIEMENS
Your IMPACT Pro R hearing instrument

Features
1 Receiver Unit
2 Click Dome
3 Battery compartment
4 Push Button
5 Microphone Cover
6 Click Mould
7 Grid
8 HF-4 Wax Guard
Your hearing instruments have one of the following ear pieces:


Changing the batteries
- Open the battery compartment

natural_image
Simple diagram showing a cloud and a blue circle with a plus sign, no text or symbols present.
natural_image
Close-up of a curved, metallic object with a blue arrow indicating upward motion (no text or symbols)- Remove coloured sticker from new battery, place battery in drawer with + symbol on battery + side up. (You should only remove the protective sticker if you actually use the battery)

natural_image
Illustration of a hand holding a blue circular object with a plus sign, next to a curved mechanical component (no text or symbols)- Close battery door

natural_image
Close-up of a white ergonomic toilet with a blue arrow indicating direction (no text or symbols)Low Battery indicator
Your hearing instrument may be programmed to inform you when the battery should be replaced for best system performance. You may hear an intermittent series of beeps and/or experience low amplification power or volume. Change the battery when either of these occur.
Inserting and removing your hearing instrument
NOTICE
Do not use lubricant when you insert the Click Dome/Mould, because it may clog the sound outlet
For hearing instruments with a Click Mould, handle by holding the Click Mould. Do not pull the tube
Inserting your hearing instrument
▶ Turn your hearing instrument off
Hold the receiver with the Click Dome/Mould between your thumb and index finger and insert it carefully into the ear canal. You may find it helpful to pull your ear slightly upward or to the back
Place your hearing instrument behind your ear
▶ Turn your hearing instrument on

natural_image
Illustration of a hand holding a blue medical tube inserted into an ear area, with no visible text or symbols.Removing your hearing instrument
▶ Turn your hearing instrument off
Hold the Receiver Unit/Click Mould between your thumb and index finger and carefully pull it out. You may find it helpful to lightly push your ear from the back

natural_image
Illustration of a human ear with a blue and pink color strip inside, showing anatomical positioning (no text or symbols)
CAUTION
Attach the Click Dome properly. If the Click Dome remains in the ear, have it removed by your NHS Audiologist.
Switching the instrument on/off
On:
Close the battery compartment fully to switch on.

natural_image
Simple illustration of a toilet with a blue arrow pointing downward (no text or symbols)Off:
Press open the battery compartment to the first stop to switch off.

Selecting the hearing program
The IMPACT Pro R can be programmed with up to 5 hearing programs which are selected with a program button. Each time the button is pressed the instrument changes to the next program, which is confirmed with beep tones. The number of beep tones is the same as the selected program.
Your NHS Audiologist will complete your program details below:

natural_image
Blue rounded rectangular button or pad icon on a gray background (no text or symbols)Hearing program
- Everyday listening
- 5.
IMPACT Pro R will always be on program 1 when initially switched on.
Optional
On/Off using the program button
Your IMPACT Pro R can also be programmed by your NHS Audiologist to be turned on and off using the program button.
Optional
Power-on delay
Your hearing instrument may be equipped with a "power-on delay" feature.
This feature delays sound from entering your hearing instrument when it is first turned on, so that you may insert the hearing instrument into your ear without experiencing unpleasant feedback (whistling).
Click Domes and Click Moulds
Cleaning
NOTICE
Click Domes and Click Moulds are easily damaged when handled wrongly.
▶ Never use sharp objects for cleaning
▶ Never wash Receiver unit under a tap
▶ Never use pressure or rotating action while cleaning
Do not spray cerumen solvent directly onto the Grid
Do not pull the receiver out of the Click Mould
The Click Mould is fitted with a wax filter which will require changing regularly
Cerumen (ear wax) on the Click Domes and Click Moulds can affect sound quality.
Clean Click Domes and Click Moulds immediately after taking the hearing instrument off. This way cerumen can not dry and harden.

natural_image
Illustration of two hands performing a medical procedure, one pointing at a blue arrow (no text or symbols present)Wash Click Dome under a tap and a small brush or cloth should be used to remove general wax as part of a daily routine
Clean and dry Click Dome with a suitable cloth
Click Domes are consumable items and should be replaced every 3 months
Changing the Click Dome
Please refer to the Click Dome packaging for instructions on how to remove and connect Click Domes
Changing the Click Mould wax filter
▶ Please refer to the instructions provided with your Click Mould
Click Domes and Click Moulds

CAUTION
Risk of injury caused by incorrect use of Click Domes.
▶ Always wear the Receiver Unit with the Click Dome
Make sure that the Click Dome is completely attached to the Receiver Unit
If the Click Dome remains in your ear, have it removed by a medical professional
Replace the Click Dome if it breaks
Insert the Receiver Unit with the Click Dome carefully and not too deeply into the ear
Use the Receiver Unit and Click Dome only in the manner described in this user guide
Your NHS Audiology Department offers an assortment of receiver lengths and Click Dome sizes to best fit your ear.
Handling of the Receiver Unit
i The Receiver Unit can be changed as needed by your NHS Audiologist. If you need to replace the Receiver Unit by yourself follow these directions carefully.
Remove Receiver Unit
Hold the instrument in one hand and turn the Receiver Unit with the other hand carefully to the left (45°) and remove it

Attach Receiver Unit
- Join the instrument and a new Receiver Unit together with the Receiver Unit turned to the left (45°). The mark on the instrument has to coincide with the mark on the Receiver Unit - Turn the Receiver Unit to the right until it locks into place

e2e wireless™ for IMPACT Pro e2e systems
Optional
The following section applies to binaural fittings only.
Your IMPACT Pro R hearing instrument includes e2e (ear-to-ear) wireless technology. The wireless functionality:
Automatically adjusts both hearing instruments simultaneously to adapt to different listening situations
Provides synchronisation of the hearing instrument settings. If you change the volume or program on one instrument, the other is automatically changed
Your NHS Audiologist can activate the wireless functionality of your hearing system.
NOTICE
The performance of the e2e wireless system could be affected by electro-magnetic interference, such as computer monitors or halogen lamp systems controlled by a switching power supply. Move away from the source of interference if you experience difficulty.

In some countries restrictions for the usage of wireless equipment exist.
▶ Refer to local authorities for further information
miniTek and Tek wireless enhancement
Optional
Your IMPACT Pro R hearing instrument can be used with the miniTek and Tek wireless enhancement system. Both devices provide a wireless connection from your IMPACT Pro R hearing instruments to your television, mobile phone, MP3 player and stereo. miniTek and Tek are also easy-to-use remote controls for your IMPACT Pro R hearing instruments for convenient and discreet program and volume adjustment. The miniTek and Tek packages comprises the miniTek or Tek Connect itself which controls the hearing instruments, a Tek transmitter and all cables necessary to link the miniTek/Tek into a variety of devices such as TVs, MP3 players, etc.

NOTICE
miniTek and Tek are additional products to your hearing instrument; Tek is available to purchase at £300. miniTek is available to purchase at £400. If you think it would assist you in your day-to-day life talk to your NHS Audiologist about how you can obtain a miniTek or Tek. Further information is available on our website www.siemens.co.uk/hearing or email siemenshearing.shi.uk@siemens.com
Siemens Hearing Instruments will also be offering accessories and care products. Please visit the website or call 01293 423700.
General notes on safety

WARNING
Choking hazard posed by small parts.
Keep batteries, hearing instruments and accessories out of children's reach
If swallowed consult a physician or a hospital immediately

If infants, small children or disabled persons need to wear hearing instruments ensure adequate supervision
▶ Check the condition and operation of the hearing instrument regularly.
Consult your NHS Audiologist if the housing is deformed
Keep the battery compartment locked. Verify the proper function of the locking mechanism
NOTICE
Your hearing instruments are sensitive to extreme heat, high humidity, strong magnetic fields (> 0.1T), X-rays and mechanical stress.
Do not expose your hearing instruments to extreme temperature or high humidity
Do not leave them in direct sunlight
Do not wear them in the shower, or when applying makeup, perfume, aftershave, hairspray or suntan lotion
Do not wear your hearing instruments when you are exposed to short-waves, a strong magnetic field, a high frequency field or X-ray
Do not place your hearing instruments in a microwave oven
Leaking batteries damage hearing instruments.
▶ Turn the hearing instruments off when not in use to preserve battery life
Remove batteries when the instruments are not in use for a prolonged period of time
NOTICE
Batteries contain harmful substances that pollute the environment.
Do not throw away batteries into household refuse
▶ Dispose batteries according to national regulations or return to your NHS Audiologist

WARNING
Hazard of explosion!
Do not use your hearing instruments in areas where there is a danger of explosions (e.g. mining)
IN CASE OF INSTRUMENT FAILURE
Is the instrument switched on?
Is the battery inserted correctly?
Is the battery run down?
Is the battery compartment closed?
Is the sound opening in the receiver clogged with wax?
With the CE marking, Siemens confirms that this hearing instrument fulfils all relevant European guidelines, in particular directive 93/42 EEC of the European Council applying to medical products.
For the products with e2e wireless 2.0 Siemens additionally confirms compliance with the European Directive 99/5/EC (R&TTE) concerning radio and telecommunications terminal equipment.
Our quality management system, certified by an independent authority, assures consistently high product quality and reliability to the customer.
Siemens
Represented in the UK by:

0123

Siemens Hearing Instruments
Alexandra House
Newton Road
Manor Royal, Crawley
West Sussex RH10 9TT
United Kingdom
Email: siemenshearing.shi.uk@siemens.com
www.siemens.co.uk/hearing