USER MANUAL 17xx FLUKE

Auxiliary Input Adapter

for 17xx

Power and Energy Loggers

Instructions

Introduction

The Auxiliary Input Adapter (the Adaptor or Product) is an accessory for the 17xx Power and Energy Loggers. The adapter allows connection of up to two sensors for analog measurements. These measurements include temperature, pressure, and level with an output voltage of 0 V dc to 10 V dc or a current loop of 4 mA to 20 mA (with additional shunt). Alternately, up to two dc voltage inputs in the range of 10 V dc to 1000 V dc can be used for other applications such as solar energy systems.

FLUKE 17xx - Introduction - 1

natural_image Technical line drawing of a mechanical component with a central connector and four circular ports (no text or symbols)

How to Contact Fluke

To contact Fluke, call one of the following telephone numbers:

• Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
• Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31 402-675-200
• Japan: +81-3-6714-3114
• Singapore: +65-6799-5566

Anywhere in the world: +1-425-446-5500

Or, visit Fluke's website at www.fluke.com.

To register your product, visit http://register.fluke.com

To view, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.

Safety Information

A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.

Warning

To prevent possible electrical shock, fire, or personal injury:

  • Read all safety information before you use the Product.
  • Carefully read all instructions.
  • Do not alter the Product and use only as specified, or the protection supplied by the Product can be compromised.
  • Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc.

  • Comply with local and national safety codes. Use personal protective equipment (approved rubber gloves, face protection, and flame-resistant clothes) to prevent shock and arc blast injury where hazardous live conductors are exposed.

  • Do not use the Product around explosive gas, vapor, or in damp or wet environments.
  • Limit operation to the specified measurement category, voltage, or amperage ratings.
  • Do not use test leads if they are damaged. Examine the test leads for damaged insulation, exposed metal, or if the wear indicator shows. Check test lead continuity.
  • Remove the input signals before you clean the Product.
  • Use only cables with correct voltage ratings.
  • Do not exceed the Measurement Category (CAT) rating of the lowest rated individual component of a Product, probe, or accessory.
  • Do not use the Product if it is altered or damaged.
  • Disable the Product if it is damaged.
  • Disconnect test leads before opening the housing.

Symbols

Symbol Description
FLUKE 17xx - Warning - 1WARNING. RISK OF DANGER.
FLUKE 17xx - Warning - 2WARNING. HAZARDOUS VOLTAGE. Risk of electric shock.
FLUKE 17xx - Warning - 3Consult user documentation.
FLUKE 17xx - Warning - 4Double Insulated
[BATC]Earth
[CWCE]Measurement Category II is applicable to test and measuring circuits connected directly to utilization points (socket outlets and similar points) of the low-voltage MAINS installation.
[SW3C]Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building's low-voltage MAINS installation.
[WZX1]Measurement Category IV is applicable to test and measuring circuits connected at the source of the building's low-voltage MAINS installation.
[H726]Conforms to European Union directives.
FLUKE 17xx - Warning - 5Certified by CSA Group to North American safety standards.
FLUKE 17xx - Warning - 6Conforms to relevant Australian Safety and EMC standards.
FLUKE 17xx - Warning - 7This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste.

Direct Input

0 V DC to 10 V DC Configuration

Caution

Connect to the Direct Input only if the Voltage Divider Input is disconnected.

To connect the Adapter:

  1. Make sure that all 4 mm safety test leads are disconnected.
  2. Loosen the 4 captive screws ② and remove the lid ① to open the housing.

FLUKE 17xx - Caution - 1

text_image FUKE 12XX Audio/USB/PLC 1 2 5 Ø 3-6 mm [0,12-0,24in] 4 3 40mm [1,57in] 10mm [0,39in]
  1. Cut and strip the cable.

  2. Open the grommet strain relief ③ and remove the dummy plug ④.

Note The dummy plug or a cable must be installed to achieve the ingress protection of IP50.

  1. Insert the sensor cable through the strain relief ③.

The cable diameter must be in the range of 3 mm to 6 mm to ensure proper strain relief.

  1. Tighten the strain relief.
  2. To connect the wires to the terminals ⑤, press the opening lever.
  3. Correctly orient the lid ① and close the housing.
  4. Tighten the captive screws ②.
  5. Connect the 4-pin connector (not shown) to the Logger.
  6. Configure the AUX channel on the Logger to show the correct readings and unit of measurement for the attached sensor. For more information, see the documentation for the Logger and the Sensor.

Voltage Divider Input

10 V DC to 1000 V DC Configuration

Caution

Connect the Voltage Divider Input only if the Direct Input is disconnected.

To connect the Adapter:

  1. Use correctly rated 4 mm test leads on the safety socket inputs.

Note

Do not connect a signal at the direct input and voltage divider input at the same time. The measured value will be incorrect.

The best accuracy is achieved when the inputs have galvanic isolation from each other.

See the Additional Errors table for more information.

  1. Connect the 4-pin connector to the Logger.
  2. Configure the AUX channel on the Logger to show the correct readings, select an available custom configuration (Custom 1 to Custom 5), and use these settings:

  3. Sensor Type: Other

  4. Unit: V
    •Gain: 1000 V/V
  5. Offset: 0 V

How to Clean

Clean the Product with a soft cloth, mild soap, and water.

Caution

To avoid damage, do not clean with abrasives or solvents.

Specifications

General

Calibration Cycle .... 2 years

Dimensions.... 119 mm x 86 mm x 38 mm

(4.69 in x 3.39 in x 1.5 in)

Weight 280 g (0.6 lb)

Output Cable Length ..... 1.9 m (6.2 ft)

Electrical

Direct input

Voltage ....0 V dc to ±10 V dc

Input Impedance..... 1066 kΩ (see block diagram for details)

Cable Diameter .... 3 mm to 6 mm

Connection .... Spring-loaded terminal for solid or braided wires

minimum: 0.2 mm² (AWG 24)

maximum: 1.5 mm ^4 (AWG 16)

Voltage Divider Input

Voltage .... 0 V dc to ±1000 V dc

Input Impedance..... 421 MΩ (see block diagram for details)

Connector.... 4 mm safety socket

Accuracy

Input RangeIntrinsic Accuracy AUX Adapter + Instrument (% of Reading + % of Range)
Direct Input ±10 V see instrument specification
Voltage Divider Input ±1000V ±(0.7 % + 0.002 %)
NoteReference Conditions for attachment: Individual use of AUX1 or AUX2, or galvanic-isolated sources (for example, dc current clamps) at AUX1 and AUX2.Environmental Reference Conditions: 23 °C ±5 °C, instrument operating for at least 30 minutes, no external electrical/magnetic field, RH <65 %.

Additional Errors:

Influence by galvanic connection of sources

Typical additional errors for measurements on galvanic-connected sources
2x Divider InputsDivider Input/Direct Input2x Direct Inputs
Type of InfluenceAUX1 or AUX21000 VCAT III InputAUX1 & AUX21000 VCAT III InputAUX1 or AUX2max 30 V to ground InputAUX1 or AUX2max 30 V to ground Input
Common Mode [1]1.5 % of V_CM 3 % of V_CM 30 ppm of V_CM 0.15 % of V_CM
Voltage difference [2]0.7 % of V_diff 1.5 % of V_diff 15 ppm of V_diff 0.15 % of V_diff
[1] Common Mode Voltage VCM = Voltage difference between LO potentials of AUX1 and AUX2[2] Voltage difference Vdiff = difference of voltages V_AUX1 - V_AUX2 with connected LO terminals

Temperature Coefficient

Direct Input ......See specification for Logger

Voltage Divider Input.....150 ppm/°C (for temperatures <18 °C and >28 °C)

Safety

Direct Input

Maximum Voltage between Direct Input Terminals and

Earth Ground.... 30 V

Voltage Divider Input

Maximum Voltage between 4 mm Safety Socket and

Earth Ground..... 1000 V

IEC61010-1 ......Pollution Degree 2

IEC61010-2-030.....CAT III 1000 V

Environmental

IP Rating..... IEC 60529:IP50

Temperature

Operating....-10 °C to +50 °C (+14 °F to +122 °F)

Storage....-20 °C to +60 °C (-4 °F to +140 °F)

Operating Humidity..... <10 °C (<50 °F) non condensing

10 °C to 30 °C (50 °F to 86 °F) ≤95 %

30 °C to 40 °C (86 °F to 104 °F) ≤75 %

40 °C to 50 °C (104 °F to 122 °F) ≤45 %

Altitude

Operating ....≤2000 m (>2000 m to ≤4000 m, derate to

CAT IV 300 V / CAT III 600 V / CAT II

1000 V)

Storage.... 12 000 m

Block Diagram
FLUKE 17xx - Altitude - 1

text_image Voltage Divider Input/1000 V + Safety plugs - AUX1 Direct Input / 10 V + Terminal Blocks - AUX1 17xx Auxiliary Input Rin= 2.92M Voltage Divider Input/1000 V + Safety plugs - AUX2 Direct Input/10 V + Terminal Blocks -

LIMITED WARRANTY AND LIMITATION OF LIABILITY

This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke's behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.

THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.