USER MANUAL 17xx FLUKE
Auxiliary Input Adapter
for 17xx
Power and Energy Loggers
Instructions
Introduction
The Auxiliary Input Adapter (the Adaptor or Product) is an accessory for the 17xx Power and Energy Loggers. The adapter allows connection of up to two sensors for analog measurements. These measurements include temperature, pressure, and level with an output voltage of 0 V dc to 10 V dc or a current loop of 4 mA to 20 mA (with additional shunt). Alternately, up to two dc voltage inputs in the range of 10 V dc to 1000 V dc can be used for other applications such as solar energy systems.

natural_image
Technical line drawing of a mechanical component with a central connector and four circular ports (no text or symbols)How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
• Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
• Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31 402-675-200
• Japan: +81-3-6714-3114
• Singapore: +65-6799-5566
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com
To view, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.
Safety Information
A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.
Warning
To prevent possible electrical shock, fire, or personal injury:
- Read all safety information before you use the Product.
- Carefully read all instructions.
- Do not alter the Product and use only as specified, or the protection supplied by the Product can be compromised.
-
Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc.
-
Comply with local and national safety codes. Use personal protective equipment (approved rubber gloves, face protection, and flame-resistant clothes) to prevent shock and arc blast injury where hazardous live conductors are exposed.
- Do not use the Product around explosive gas, vapor, or in damp or wet environments.
- Limit operation to the specified measurement category, voltage, or amperage ratings.
- Do not use test leads if they are damaged. Examine the test leads for damaged insulation, exposed metal, or if the wear indicator shows. Check test lead continuity.
- Remove the input signals before you clean the Product.
- Use only cables with correct voltage ratings.
- Do not exceed the Measurement Category (CAT) rating of the lowest rated individual component of a Product, probe, or accessory.
- Do not use the Product if it is altered or damaged.
- Disable the Product if it is damaged.
- Disconnect test leads before opening the housing.
Symbols
| Symbol Description | |
![]() | WARNING. RISK OF DANGER. |
![]() | WARNING. HAZARDOUS VOLTAGE. Risk of electric shock. |
![]() | Consult user documentation. |
![]() | Double Insulated |
| [BATC] | Earth |
| [CWCE] | Measurement Category II is applicable to test and measuring circuits connected directly to utilization points (socket outlets and similar points) of the low-voltage MAINS installation. |
| [SW3C] | Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building's low-voltage MAINS installation. |
| [WZX1] | Measurement Category IV is applicable to test and measuring circuits connected at the source of the building's low-voltage MAINS installation. |
| [H726] | Conforms to European Union directives. |
![]() | Certified by CSA Group to North American safety standards. |
![]() | Conforms to relevant Australian Safety and EMC standards. |
![]() | This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste. |
Direct Input
0 V DC to 10 V DC Configuration
Caution
Connect to the Direct Input only if the Voltage Divider Input is disconnected.
To connect the Adapter:
- Make sure that all 4 mm safety test leads are disconnected.
- Loosen the 4 captive screws ② and remove the lid ① to open the housing.

text_image
FUKE 12XX Audio/USB/PLC 1 2 5 Ø 3-6 mm [0,12-0,24in] 4 3 40mm [1,57in] 10mm [0,39in]-
Cut and strip the cable.
-
Open the grommet strain relief ③ and remove the dummy plug ④.
Note The dummy plug or a cable must be installed to achieve the ingress protection of IP50.
- Insert the sensor cable through the strain relief ③.
The cable diameter must be in the range of 3 mm to 6 mm to ensure proper strain relief.
- Tighten the strain relief.
- To connect the wires to the terminals ⑤, press the opening lever.
- Correctly orient the lid ① and close the housing.
- Tighten the captive screws ②.
- Connect the 4-pin connector (not shown) to the Logger.
- Configure the AUX channel on the Logger to show the correct readings and unit of measurement for the attached sensor. For more information, see the documentation for the Logger and the Sensor.
Voltage Divider Input
10 V DC to 1000 V DC Configuration
Caution
Connect the Voltage Divider Input only if the Direct Input is disconnected.
To connect the Adapter:
- Use correctly rated 4 mm test leads on the safety socket inputs.
Note
Do not connect a signal at the direct input and voltage divider input at the same time. The measured value will be incorrect.
The best accuracy is achieved when the inputs have galvanic isolation from each other.
See the Additional Errors table for more information.
- Connect the 4-pin connector to the Logger.
-
Configure the AUX channel on the Logger to show the correct readings, select an available custom configuration (Custom 1 to Custom 5), and use these settings:
-
Sensor Type: Other
- Unit: V
•Gain: 1000 V/V - Offset: 0 V
How to Clean
Clean the Product with a soft cloth, mild soap, and water.
Caution
To avoid damage, do not clean with abrasives or solvents.
Specifications
General
Calibration Cycle .... 2 years
Dimensions.... 119 mm x 86 mm x 38 mm
(4.69 in x 3.39 in x 1.5 in)
Weight 280 g (0.6 lb)
Output Cable Length ..... 1.9 m (6.2 ft)
Electrical
Direct input
Voltage ....0 V dc to ±10 V dc
Input Impedance..... 1066 kΩ (see block diagram for details)
Cable Diameter .... 3 mm to 6 mm
Connection .... Spring-loaded terminal for solid or braided wires
minimum: 0.2 mm² (AWG 24)
maximum: 1.5 mm ^4 (AWG 16)
Voltage Divider Input
Voltage .... 0 V dc to ±1000 V dc
Input Impedance..... 421 MΩ (see block diagram for details)
Connector.... 4 mm safety socket
Accuracy
| Input Range | Intrinsic Accuracy AUX Adapter + Instrument (% of Reading + % of Range) | |
| Direct Input ±10 V see instrument specification | ||
| Voltage Divider Input ±1000 | V ±(0.7 % + 0.002 %) | |
| NoteReference Conditions for attachment: Individual use of AUX1 or AUX2, or galvanic-isolated sources (for example, dc current clamps) at AUX1 and AUX2.Environmental Reference Conditions: 23 °C ±5 °C, instrument operating for at least 30 minutes, no external electrical/magnetic field, RH <65 %. | ||
Additional Errors:
Influence by galvanic connection of sources
| Typical additional errors for measurements on galvanic-connected sources | ||||
| 2x Divider Inputs | Divider Input/Direct Input | 2x Direct Inputs | ||
| Type of Influence | AUX1 or AUX21000 VCAT III Input | AUX1 & AUX21000 VCAT III Input | AUX1 or AUX2max 30 V to ground Input | AUX1 or AUX2max 30 V to ground Input |
| Common Mode [1] | 1.5 % of V_CM | 3 % of V_CM | 30 ppm of V_CM | 0.15 % of V_CM |
| Voltage difference [2] | 0.7 % of V_diff | 1.5 % of V_diff | 15 ppm of V_diff | 0.15 % of V_diff |
| [1] Common Mode Voltage VCM = Voltage difference between LO potentials of AUX1 and AUX2[2] Voltage difference Vdiff = difference of voltages V_AUX1 - V_AUX2 with connected LO terminals | ||||
Temperature Coefficient
Direct Input ......See specification for Logger
Voltage Divider Input.....150 ppm/°C (for temperatures <18 °C and >28 °C)
Safety
Direct Input
Maximum Voltage between Direct Input Terminals and
Earth Ground.... 30 V
Voltage Divider Input
Maximum Voltage between 4 mm Safety Socket and
Earth Ground..... 1000 V
IEC61010-1 ......Pollution Degree 2
IEC61010-2-030.....CAT III 1000 V
Environmental
IP Rating..... IEC 60529:IP50
Temperature
Operating....-10 °C to +50 °C (+14 °F to +122 °F)
Storage....-20 °C to +60 °C (-4 °F to +140 °F)
Operating Humidity..... <10 °C (<50 °F) non condensing
10 °C to 30 °C (50 °F to 86 °F) ≤95 %
30 °C to 40 °C (86 °F to 104 °F) ≤75 %
40 °C to 50 °C (104 °F to 122 °F) ≤45 %
Altitude
Operating ....≤2000 m (>2000 m to ≤4000 m, derate to
CAT IV 300 V / CAT III 600 V / CAT II
1000 V)
Storage.... 12 000 m
Block Diagram

text_image
Voltage Divider Input/1000 V + Safety plugs - AUX1 Direct Input / 10 V + Terminal Blocks - AUX1 17xx Auxiliary Input Rin= 2.92M Voltage Divider Input/1000 V + Safety plugs - AUX2 Direct Input/10 V + Terminal Blocks -LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke's behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.