VAX020 - Measurement Megger - Free user manual and instructions
Find the device manual for free VAX020 Megger in PDF.
| Product Type | High Voltage Amplifier |
| Brand | Megger |
| Model | VAX020 |
| Application | Insulation diagnostics in electric power systems, moisture assessment for oil-immersed transformers |
| Output Voltage | 2 kV peak (1414 V rms) |
| Output Current | 50 mA peak (above 50 Hz), derating linearly to 30 mA below 10 Hz |
| Frequency Range | DC to 1 kHz |
| Capacitive Load | 0-20 μF capability; max 80 nF at 50 Hz at 2 kV, 67 nF at 60 Hz |
| Mains Voltage | 100-240 V AC, 50/60 Hz |
| Power Consumption | 120 VA max |
| Dimensions (Instrument) | 335 x 300 x 99 mm |
| Weight | 4.4 kg (9.7 lbs) without accessories |
| Operating Temperature | -20°C to +55°C (-4°F to +131°F) |
| Storage Temperature | -40°C to +55°C (-40°F to +131°F) |
| Humidity | < 90% RH, non-condensing |
| Protection Class | IP21 |
| Safety Features | Interlock input, protective ground terminal, overcurrent protection, safety instructions per IEEE 510-1983 |
| Interface | Control via multi-pin connector to IDAX; interlock via 3-pole XLR |
| Compatible Systems | IDAX300, IDAX300S, IDAX206-FR, IDAX206, IDA200 |
| Maintenance | Routine inspection of cables and connectors; clean with damp cloth; disconnect from mains before cleaning |
| Reparability | Do not attempt to service yourself; warranty void if opened; contact Megger for service |
Frequently Asked Questions - VAX020 Megger
User questions about VAX020 Megger
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Measurement in PDF format for free! Find your manual VAX020 - Megger and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. VAX020 by Megger.
USER MANUAL VAX020 Megger
High Voltage Amplifier
User's Manual

Megger
VAX020
High Voltage Amplifier
User's Manual
NOTICE OF COPYRIGHT & PROPRIETARY RIGHTS
© 2013, Megger Sweden AB. All rights reserved.
The contents of this manual are the property of Megger Sweden AB. No part of this work may be reproduced or transmitted in any form or by any means, except as permitted in written license agreement with Megger Sweden AB. Megger Sweden AB has made every reasonable attempt to ensure the completeness and accuracy of this document. However, the information contained in this manual is subject to change without notice, and does not represent a commitment on the part of Megger Sweden AB. Any attached hardware schematics and technical descriptions, or software listings that disclose source code, are for informational purposes only. Reproduction in whole or in part to create working hardware or software for other than Megger Sweden AB products is strictly prohibited, except as permitted by written license agreement with Megger Sweden AB.
TRADEMARK NOTICES
Megger® and Programma® are trademarks registered in the U.S. and other countries. All other brand and product names mentioned in this document are trademarks or registered trademarks of their respective companies.
Megger Sweden AB is certified according to ISO 9001 and 14001.
Postal address:
Megger Sweden AB
Box 724
SE-182 17 DANDERYD
SWEDEN
Visiting address:
Megger Sweden AB
Rinkebyvägen 19
SE-182 36 DANDERYD
SWEDEN
T +46 8 510 195 00
F +46 8 510 195 95
seinfo@megger.com
www.megger.com

Contents
1 Introduction
6
1 General....6
1.1 Description 6
1.2 Receiving instructions 6
2 Safety
8
2.1 Symbols on the instrument 8
2.2 Safety instructions....8
3 Control, indicators and connectors\_10
3.1 Front panel 10
3.2 Back panel....10
4 Run VAX020
12
4.1 Connections ...... 12
4.2 Perform a measurement....13
5 Specifications
14
1 Introduction
1 General
1.1 Description
VAX020 is a 2 kV voltage amplifier designed to work as an accessory to an IDA/IDAX test system, increasing the output voltage from 200 V to 2000 V peak (1414 V rms). The main application is insulation diagnostics in electric power systems with moisture assessment for oil-immersed transformers as a key area.
In particular, VAX020 is recommended in situations where measurements need to be performed during high interference, e.g. measuring low capacitance components in substations with severe AC and/or DC noise/offset.
1.2 Receiving instructions
Check the equipment received against the packing list to ensure that all materials are present. Notify Megger of any shortage. Telephone +46 8 510 195 00 and ask for the Customer Service Department.
Examine the instrument for damage received in transit. If damage is discovered, file a claim with the carrier at once and notify Megger, giving a detailed description of the damage.
This instrument has been thoroughly tested and inspected to meet rigid specifications before being shipped. It is ready for use when set up as indicated in this manual.
2
Safety
2.1 Symbols on the instrument

Caution, refer to accompanying documents.

Caution, risk of electric shock.

Protective conductor terminal.

WEEE, Waste Electrical and Electronic Equipment. Please utilize your local WEEE collection facilities in the disposition of this product and otherwise observe all applicable requirements.
2.2 Safety instructions

Important
Read and comply with the following instructions.
Always comply with local safety regulations.

Warning
1] The test set and the specimen to which it is connected are a possible source of high-voltage electrical energy and all persons making or assisting in tests must use all practical safety precautions to prevent contact with energized parts of the test equipment and related circuits.
2] Persons actually engaged in the test must stand clear of all parts of the complete high-voltage circuit, including all connections, unless the test set is de-energized and all parts of the test circuit are grounded. Persons not directly involved with the work must be kept away from test activities by suitable barriers, barricades or warnings.
3] High-voltage discharges and other sources of strong electric or magnetic fields may interfere with the proper functioning of heart pacemakers. Persons with heart pacemakers should obtain expert advice on the possible risks before operating this equipment or being close to the equipment during operation.
4] Treat all terminals of high-voltage power equipment as a potential electric shock hazard. There is always the potential of voltages being induced at these terminals because of proximity to energized high-voltage lines or equipment.
5] The instrument is equipped with a power cord with integral safety ground pin. The equipment must be connected to a grounded mains outlet.
The instrument case must also be grounded by the separate protective ground wire with connection to the protective earth terminal on the rear panel. This is to eliminate differ-
ence in earth potential between the instrument and the device to be tested. Check the continuity of the protective ground wire before each use.
6] Do not attempt to service the instrument yourself. Opening or removing covers may expose you to dangerous voltage. If you attempt to service the instrument yourself the warranty is no longer valid.
7] Do not use any accessories that are not intended for use together with the instrument.
8] Do not use the instrument for any purpose other than indicated by the manufacturer.
9] Never connect the test set to energized equipment.
10] Operation is prohibited in rain or snow.
11] Do not use the test set in an explosive atmosphere.
12] Disconnect the instrument from the mains before cleaning. Use a damp cloth for cleaning. Do not use liquid cleaners or aerosol cleaners.


1] Always use a safety ground stick to ground the high-voltage conductor. A safety ground jumper must then be installed between all terminals of apparatus under test and ground.
2] Always disconnect test leads from power equipment before attempting to disconnect them at the test set. The ground connection must be the first made and the last removed. Any interruption of the grounding connection can create an electric shock hazard.
3] Before making connection to the power source, determine that the instrument rating matches the voltage of the power source and has a suitable two-pole, three-terminal grounding connector.
4] The power input plug must be inserted only into a mating receptacle with a ground contact. Do not bypass the grounding connection. Any interruption of the grounding connection can create an electric shock hazard. Determine that the receptacle is properly wired before inserting the plug.
5] Routine maintenance is all that is required for these test sets. The cables and connector panel should be inspected frequently to be
sure all connections are tight and all ground connections intact.
6] A qualified operator should be in attendance at all times while the test equipment is in operation.
7] Refer to IEEE 510-1983, "IEEE Recommended Practices for Safety in High-Voltage and High-Power Testing," for information.
8] Make sure that air intake and fan exhausts are not blocked
9] Avoid dusty environments and locations because the accuracy can be affected by long-term exposure to dust
10] The VAX020 is designed for indoor and outdoor use. Although a rugged instrument, it should be well protected when moved or shipped.
11] The Protective Earth/Ground (PE) of the VAX020 is connected to the earth/ground pin of the inlet, making it grounded when the mains cord is connected.
12] Connect VAX020 to ground using the ground wing nut on the front panel to ensure protective grounding independent of the mains cord being connected or not.
3
Control, indicators and connectors
3.1 Front panel
| 1. | ON / OFF switch |
| 2. | Mains inlet |
| 3. | Ground: wing nut/screw (PE) |
| 4. | Control: multi-pin connector to IDAX |
| 5. | Interlock: 3-pole XLR |
| 6. | Output: red LED is lit when high-voltage output is active |
| 7. | Output: HV connector for 0-2000 V |
3.2 Back panel
■Two holes/grids for fans
Air inlet

4 Run VAX020
4.1 Connections
1] Connect VAX020 to protective ground on the IDAX using the Ground cable (green in the figure below)
2] Connect VAX020 (CONTROL) to IDAX (EXTERNAL) with the 12 pole multi-cable (blue in the figure below)
3] Connect an Interlock cable to the Interlock input (violet in the figure below)
4] Connect the VAX020 to the test object using the Generator cable (18 meter) with HV connector and clamp
5] Connect the Measure cable from the IDAX to the test object

4.2 Perform a measurement
1] Turn on the IDAX and the VAX020
2] Follow the standard IDAX procedure for measurement setup
3] When prompted to select voltage: choose VAX
4] Then add required voltage, e.g. 1400 V (RMS) (same as 2 kV peak)
5] Perform the measurement in the same way as a standard IDAX measurement. See IDAX User Manual for more details.

5
Specifications
Specifications VAX020
Environmental
Application field The instrument is intended for use in medium- and high-voltage substations and industrial environments.
Ambient temperature
Operating -20°C to +55°C (-4°F to +131°F)
Storage -40°C to +55°C (-40°F to +131°F)
Humidity < 90% RH, non-condensing
CE-marking
EMC 2004/108/EC
LVD 2006/95/EC
Category CAT 1
IP class IP21
General
Mains voltage 100-240 V AC, 50 / 60 Hz
Power consumption 120 VA (max)
Dimensions Instrument 335 x 300 x 99 mm (17.7" x 6.3" x 16.1")
Transport case 520 x 430 x 220 mm (20.5" x 17" x 8.7")
Weight 4.4 kg (9.7 lbs) without accessories
Interface
Control Connects VAX020 with IDAX
Interlock An external contact must be closed to allow the VAX020 to generate any voltage
Output HV connector that connects VAX020 to the test object
VAX020 can be used to get together with the following IDAX systems: IDAX300, IDAX300S, IDAX206-FR, IDAX206 and IDA200
Output
Voltage 2 kV (peak)
Current 50 mA (peak) above 50 Hz derating linearly to 30 mA below 10 Hz
Frequency range DC-1 kHz
Capacitive load 0-20 μF capability
Max capacitive load 80 nF at 50 Hz at 2 kV 67 nF at 60 Hz
Your "One Stop" Source for all your electrical test equipment needs
■Battery Test Equipment
■Cable Fault Locating Equipment
■Circuit Breaker Test Equipment
■Data Communications Test Equipment
■Fiber Optic Test Equipment
■Ground Resistance Test Equipment
■Insulation Power Factor (C&DF) Test Equipment
■Insulation Resistance Test Equipment
■Line Testing Equipment
■Low Resistance Ohmmeters
■Motor & Phase Rotation Test Equipment
■Multimeters
■Oil Test Equipment
■Portable Appliance & Tool Testers
■Power Quality Instruments
■Recloser Test Equipment
■Relay Test Equipment
■T1 Network Test Equipment
■Tachometers & Speed Measuring Instruments
■TDR Test Equipment
■Transformer Test Equipment
■Transmission Impairment Test Equipment
■Watthour Meter Test Equipment
■STATES® Terminal Blocks & Test Switches
■Professional Hands-On Technical and
■Safety Training Programs
Megger is a world leading manufacturer and supplier of test and measurement instruments used within the electric power, building wiring and telecommunication industries.
With research, engineering and manufacturing facilities in the USA, UK and Sweden, combined with sales and technical support in most countries, Megger is uniquely placed to meet the needs of its customers worldwide.
Megger is certified according to ISO 9001 and 14001. Megger is a registered trademark.
Megger Group Limited UNITED KINGDOM
Dover, Kent CT17 9EN ENGLAND
AUSTRALIA
CANADA
CHINA
FRANCE
GERMANY
INDIA
INDONESIA
JAPAN
KINGDOM OF BAHRAIN
KOREA
MALAYSIA
PAKISTAN
PHILIPPINES
RUSSIA
SINGAPORE
SOUTH AFRICA
SPAIN
SWEDEN
SWITZERLAND
TAIWAN
THAILAND
UNITED ARAB EMIRATES
USA
VIETNAM
CE
Megger
WWW.MEGGER.COM
Postal address:
Megger Sweden AB
Box 724
SE-182 17 DANDERYD
SWEDEN
Visiting address:
Megger Sweden AB
Rinkebyvägen 19
SE-182 36 DANDERYD
SWEDEN
T +46 8 510 195 00
F +46 8 510 195 95
seinfo@megger.com
www.megger.com