THERMO FH2 - Thermometer THOMSON - Free user manual and instructions
Find the device manual for free THERMO FH2 THOMSON in PDF.
| Product type | Non-contact infrared forehead thermometer |
| Brand | Thomson |
| Model | THERMO FH2 |
| Dimensions | 34 x 160 x 50 mm |
| Weight (without batteries) | Approx. 68 g |
| Power supply | 2 AAA alkaline batteries (not included) |
| Body measurement range | 34.0 °C to 43.0 °C (93.2 °F to 109.4 °F) |
| Object measurement range | 0 °C to 93.2 °C (32 °F to 199.7 °F) |
| Accuracy (body 35-42 °C) | ±0.2 °C |
| Resolution | 0.1 °C / °F |
| Measurement time | Approx. 1 second |
| Memory | Last 9 measurements |
| Display | Backlit LCD screen with 3 colors (green, orange, red) |
| Sound signal | Can be disabled |
| Automatic shut-off | After 30 seconds |
| Recommended measuring distance | 3 to 5 cm from the forehead |
| Units of measurement | °C or °F (switchable) |
| Operating ambient temperature | 10 °C to 40 °C, humidity 15%-93% |
| Storage temperature | -25 °C to 55 °C, humidity 0%-93% |
| Care and cleaning | Clean the sensor with a cotton swab dipped in 70% isopropyl alcohol |
| Safety | Do not disassemble, do not immerse, keep out of reach of children |
| Spare parts and repairability | No detachable parts, sealed device |
| General information | Intended use: measurement of human body temperature on the forehead. Does not replace a medical diagnosis. |
Frequently Asked Questions - THERMO FH2 THOMSON
User questions about THERMO FH2 THOMSON
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Thermometer in PDF format for free! Find your manual THERMO FH2 - THOMSON and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. THERMO FH2 by THOMSON.
USER MANUAL THERMO FH2 THOMSON
Shenzen Pango Electronic Co., Ltd
N°25 1st industry zone, Fenghuang road, Xikeng village, Henggang town,
Longgang district, Shenzhen, Guangdong China
Authorised representative :
Imported and distributed by :
STAB XTOM SARL, 46 avenue des Frères Lumière, 78190 Trappes - France.
This product has been manufactured and sold under the responsibility of Bigben Interactive, THOMSON, and the THOMSON logo are trademarks of Technicolor (S.A.) or its affiliates and are used under license by Bigben Interactive.
Any other product, service, company, trade or product name and logo referenced herein are neither endorsed nor sponsored by Technicolor (S.A.) or its affiliates.
Date: 2019-01-09
Rev : A/1

0197

Appendix 1 Guidance and Manufacturer Declaration Tables
| Guidance and manufacturer's declaration-electromagnetic emissions | ||
| The Model THERMO F42 Infrared Thermometer the electromagnetic environment specified below. The customer or the user of the Model THERMO FH2 Infrared Thermometer should assure that it is used in such an environment. | ||
| Emissions | Compliance | Electromagnetic environment-guidance |
| RF emissions CISPR 11 | Group 1 | The Model THERMO FH2 Infrared Thermometer uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. |
| RF emissions CISPR 11 | Class B | The Model THERMO FH2 Infrared Thermometer is used in home and it's powered by DC 3V |
| Harmonic emissions IEC 61000-3-2 | N. A. | |
| Voltage fluctuations/ flicker emissions IEC 61000-3-3 | N. A. | |
| Guidance and manufacturer's declaration – electromagnetic immunity | |||
| The Model THERMO For Use Infrared Thermometer the electromagnetic environment specified below. The customer or the user of the Model THERMO FH2 Infrared Thermometer should assure that it is used in such an environment. | |||
| Immunity test | IEC 60601 test level | Compliance level | Electromagnetic environment-guidance |
| Electrostatic discharge (ESD)IEC 61000-4-2 | ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 KV air | ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 KV air | Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %. |
| Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 | 30 A/m, 50/60Hz | 30 A/m, 50/60Hz | Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. |
| NOTE U is the a.c. mains voltage prior to application of the test leve | |||
| Guidance and manufacturer's declaration – electromagnetic immunity | |||
| The Model THERMO FHK2 Infrared Thermometerthe electromagnetic environment specified below. The customer or theuser of the Model THERMO FH2 Infrared Thermometer should assurethat it is used in such an environment. | |||
| Immunity test | IEC 60601test level | Compliance level | Electromagnetic environment - guidance |
| Conducted RF IEC61000-4-6 | 3 Vrms150 kHz to 80 MHz 6 Vrms150 kHz to 80 MHz outside ISM bandsa | N/A | Portable and mobile RF communications equipment should be used no closer to any part of the Model THERMO FH2 Infrared Thermometer, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.Recommended separation distance d = [ 3.5V_1 ] d = [ 3.5E_1 ] 80MHz to 800MHz d = [ 7E_1 ] 800MHz to 2.7GHzwhere P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres(m).Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, ^a should be less than the compliance level in each frequency range ^b Interference may occur in the vicinity of equipment marked with the following symbol: |
| Radiated RF IEC61000-4-3 | 10 V/m80 MHz to 2.7 GHz | 10 V/m | |
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2,7 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 has been incorporated into the formulae used in calculating the recommended separation distance for transmitters in these frequency ranges.
c Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Model THERMO FH2 Infrared Thermometer is used exceeds the applicable RF compliance level above, the Model THERMO FH2 Infrared Thermometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Model THERMO FH2 Infrared Thermometer.
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
| Recommended separation distances between portable and mobile RF communications equipment and the Model THERMO FH2 Infrared Thermometer | |||
| The Mode in THERMO or FH2 Infrared Thermometer an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Model THERMO FH2 Infrar Thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Model THERMO FH Infrared Thermometer as recommended below, according to the maximum output power of the communications equipment. | |||
| Rated maximum output of transmitter W | Separation distance according to frequency of transmitter m | ||
| 150 kHz to 80 MHz d = [ 3.5V_1 ] | 80 MHz to 800 MHz d = [ 3.5E_1 ] | 800 MHz to 2.7 GHz d = [ 7E_1 ] | |
| 0.01 | 0.12 | 0.12 | 0.23 |
| 0.1 | 0.38 | 0.38 | 0.73 |
| 1 | 1.2 | 1.2 | 2.3 |
| 10 | 3.8 | 3.8 | 7.3 |
| 100 | 12 | 12 | 23 |
| For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, when P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.NOTE 2 These guidelines may not apply in all situations.Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. | |||