HS60 - Ultrasonic cleaner SAMSUNG - Free user manual and instructions
Find the device manual for free HS60 SAMSUNG in PDF.
| Product Type | Ultrasonic Cleaner |
| Brand | Samsung |
| Model | HS60 |
| Dimensions (L x W x H) | Approx. 200 x 150 x 120 mm |
| Tank Capacity | 600 ml |
| Power Supply | AC 220-240V, 50/60 Hz |
| Ultrasonic Frequency | 42 kHz |
| Timer Setting | Up to 30 minutes, adjustable |
| Main Functions | Ultrasonic cleaning, degassing, timer, auto shut-off |
| Operation Method | Touch button or dial control |
| Material | Stainless steel tank, ABS plastic body |
| Weight | Approx. 1.2 kg |
| Maintenance | Clean tank after each use; wipe dry; descale monthly |
| Cleaning Agent Compatibility | Water or mild cleaning solution; avoid flammable solvents |
| Safety Features | Auto shut-off, overheat protection, water level sensor |
| Spare Parts Available | Replacement tank, transducer, control board |
| Repairability | Modular design; user-serviceable with manual |
| General Information | Ideal for cleaning jewelry, eyeglasses, coins, small parts |
Frequently Asked Questions - HS60 SAMSUNG
User questions about HS60 SAMSUNG
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Ultrasonic cleaner in PDF format for free! Find your manual HS60 - SAMSUNG and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. HS60 by SAMSUNG.
USER MANUAL HS60 SAMSUNG
Version 1.00
HS50/HS60
User Manual
English
Table of Contents – Volume 2
Chapter 6 Starting Diagnosis
System Power 6-3
Turning the Power On 6-3
Shutting Down the System....6-3
Probes and Applications 6-4
Selecting Probes and Applications 6-5
Creating UserPreset 6-5
Modifying UserPreset....6-5
Deleting UserPreset....6-5
Patient Information 6-6
Patient Information for Each Application 6-9
Searching Patient Information....6-16
Chapter 7 Diagnosis Modes
Information....7-3
Types of Diagnosis Modes....7-3
Basic Use 7-5
Basic Mode 7-8
2D Mode 7-8
M Mode 7-23
Color Doppler Mode 7-26
Power Doppler Mode....7-31
PW Spectral Doppler Mode 7-33
CW Spectral Doppler Mode 7-38
TDI Mode 7-40
TDW Mode 7-41
ElastoScan Mode 7-42
Combined Mode 7-51
2D/C/PW Mode 7-51 2D/PD/PW Mode 7-51 2D/C/CW Mode 7-51 2D/C/M Mode 7-51 2D/TDI/TDW Mode 7-52 Dual Live Mode 7-52
Multi-Image Mode 7-53
Dual Mode 7-53 Quad Mode 7-54
3D/4D Mode 7-55
3D Stand By 7-59 3D View – MPR 7-61 5D Follicle 7-69 VOCAL 7-71 3D XI 7-78 XI VOCAL 7-88 XI STIC 7-95 4D 7-98 3D Utility Menu 7-99
Chapter 8 Measurements and Calculations
Measurement Accuracy 8-3
Causes of Measurement Errors 8-3 Optimization of Measurement Accuracy 8-5 Measurement Accuracy Table 8-7
Basic Measurements 8-9
Distance Measurement 8-13 Circumference and Area measurement 8-23 Volume Measurement 8-26
Calculations by Application 8-29
Things to Note 8-29
Common Measurement Methods 8-33
OB Calculations 8-37
Fetal Heart Calculations 8-46
Gynecology Calculations 8-49
Cardiac Calculations 8-52
Abdomen Calculations 8-66
Pediatric Hips Calculations 8-68
Urology Calculations 8-70
Small Parts Calculations 8-71
Musculoskeletal Calculations 8-75
Vascular Calculations 8-76
Reports 8-85
Viewing Reports 8-85
Report Preview 8-86
Editing Report (Worksheet) 8-89
Adding Comments 8-92
Print Reports....8-94
Saving Reports....8-94
Store SR 8-95
Images 8-95
OB Graph 8-97
Description 8-102
Closing Reports....8-104
Chapter 9 Image Managements
Cine/Loop 9-3
Annotating Images....9-5
Annotation 9-5
BodyMarker 9-10
Saving, Playing and Transferring Images 9-12
Saving Images....9-12
Playing Images 9-13
Transferring Images....9-14
Printing Images 9-15
Printing Images 9-15
SonoView Mode....9-16
Exam List....9-17
Review – Exam Mode 9-19
Review – Compare Mode....9-24
DICOM Q/R....9-31
**Reference Manual
Samsung Medison is providing an additional HS50/HS60 Reference Manual (English version).
Starting Diagnosis
System Power 6-3
Turning the Power On 6-3
Shutting Down the System 6-3
■ Probes and Applications 6-4
Selecting Probes and Applications....6-5
Creating UserPreset 6-5
Modifying UserPreset 6-5
Deleting UserPreset 6-5
■ Patient Information ...... 6-6
Patient Information for Each Application....6-9
Searching Patient Information......6-16
:: System Power
Boot up the system for use.

CAUTION: Make sure to connect the probes and peripheral devices that will be used before powering on the system. If you attempt to connect them during system use, it may lead to patient injury or irreparable damage to the console.
Turning the Power On
Press the On/Off button when the power is off. Booting begins, and the product logo appears on the screen. When booting is completed, the 2D mode screen appears in End Exam status.

CAUTION: Before starting the diagnosis, you must register the patient information.

NOTE:
The product should turn on about 10 seconds after the power switch at the back of the product is turned on. During system booting, do not press any key on the keyboard. It may cause product malfunction. If you turn on the power after turning it off suddenly, the system may turn on and off momentarily. This is one of the characteristics of the Intel PC main board, not a system error.
Shutting Down the System
Press the On/Off button while using the system to initiate shutdown.

CAUTION:
When the system is not turned off, holding down the On/Off button for five seconds or longer forces the product to power off. If the ultrasound system is unplugged while working or turned off by forcibly shutting down, the files may get corrupted and it may result in system down or data loss. To ensure that the product is safely cut off from electrical power, set the power switch at the rear of the product to Off position after using the product.
Probes and Applications
Before scanning, select a probe and an application.

CAUTION: Refer to 'Chapter 5. Probes' for more information on probes and applications supported by the system.
Probe Selection Screen
Press the Probe button on the Touch Screen to load the Probe Selection screen. In this screen, you can select/change probe or application you want to use and change the probe presets.
The Probe Selection screen is divided into four lists.
① Probe list: Displays a list of probes currently connected to the system. ② Application list: Displays a list of applications the selected probe supports. ③ Preset list: Displays a list of presets the selected application supports. ④ UserPreset list: Displays a list of presets you can configure according to your preferences.

[Figure 6.1 Probe Selection]
Selecting Probes and Applications
The selected probe, preset, or userpreset appear in the screen area of the monitor screen.
- Select the items in the following order: Probe → Application → Preset/UserPreset.
- When you select Preset/UserPreset, the system automatically applies your selections and switches to Diagnosis mode. Press the Probe button to cancel.
Creating UserPreset
- Tap + on the touch screen to create a UserPreset.
- The Preset setup window will open.
- Enter the name you want to use and tap Save. Tap Close to cancel.
Modifying UserPreset
- Enter the Preset Setup by tapping Preset Setup on the touch screen.
- Select a Preset to rename and press Rename.
- The Preset setup window will open.
- In the Preset setup window, enter a new Preset name and press Save. To cancel it, tap Close.
Deleting UserPreset
- Enter the Preset Setup by tapping Preset Setup on the touch screen.
- After selecting a preset to delete, tap Delete.
- The Preset setup window will open.
- Tap Delete to delete the UserPreset. Tap Close to cancel it.
Patient Information
Press the Patient button on the control panel to display the Patient Information screen.
This screen allows you to enter, search, or modify patient information. Patient information includes basic details such as patient ID, name, date of birth, and gender, as well as additional application-specific information.

NOTE: The ID is required.
Entering Basic Patient Information
Enter or modify general patient information in the General Information section.
Use the trackball to move the pointer to the desired field. Tap a button on the touch screen to enter information using the touch screen.
After entering information, click OK on the screen or tap OK on the touch screen to save. Click Cancel or tap Cancel on the touch screen to cancel.

[Figure 6.2 General Patient Information]

[Figure 6.3 General Patient Information – Touch Screen]
Patient ID
Enter a patient ID.
Last Name
Enter the patient's last name.
First Name
Enter the patient's first name.
Middle Name
Enter the patient's middle name.

NOTE: The name that you have entered will appear in the title area and reports.
Date of Birth (Age)
Enter the patient's date of birth in the specified format.
Gender
Select the patient's gender.
Accession Number
When viewing the worklist for a patient via the DICOM server, this information is automatically displayed in the appropriate fields.
Diag. Physician
Enter the name of the physician who diagnosed the patient.
Ref. Physician
Enter the name of the referring physician.
Operator
Enter the name of the operator who scanned the patient. Up to 20 persons may be entered.
Search
Search through the information stored in the system.
Worklist
Perform a search by connecting to the DICOM Modality Worklist server in the hospital network.
New
Enter new patient information.
Auto ID Creation
Create temporary patient information using a temporary ID.
Keyboard
Use the keyboard displayed on the touch screen to enter text.
Start Ez Exam+
This function helps you to link an exam with Ez Exam+ when the exam starts.
Patient Information for Each Application
In Study Information, enter additional patient information or change the existing patient information for each application.
- Use Application on the monitor to select an application.
- Enter additional information required for diagnosis.
OB
Select OB from Application. Enter the additional obstetrical information.
LMP
Enter the last menstrual period for a patient.
DOC
Indicates the patient's date of conception.
EDD
Indicates the patient's expected date of delivery.
GA (LMP)
Indicates the gestational age of a patient.
Ovulation Date
Enter expected ovulation date in accordance with defined format
Day of Cycle
Enter length of menstrual cycle in days (Date).
Fetuses No.
Enter the number of fetuses. Up to four may be entered.
Gravida
Enter the number of pregnancies.
Para
Enter the number of deliveries.
Aborta
Enter the number of miscarriages.
Ectopic
Enter the number of ectopic pregnancies.
Description
Enter a description of the diagnosis.

NOTE:
Refer to 'Chapter 3. Utilities' Setup > System > Patient > User Defined List. Up to 20 persons may be entered per application.

[Figure 6.4 Study Information - OB]

GYN
Select GYN from Application. Enter additional information for Gyn. This is the same information as for OB.

[Figure 6.5 Study Information - GYN]

Urology
Select Urology from Application. Enter urological information.
PSA
Enter the Prostate Specific Antigen (PSA) value.
PPSA Coefficient 1
Enter predicted PSA coefficient 1.
PPSA Coefficient 2
Enter predicted PSA coefficient 2.

[Figure 6.6 Study Information – Urology]

Vascular
Select Vascular from Application. Enter vascular-related information.
Left Systole
Enter left systole blood pressure.
Left Diastole
Enter left diastole blood pressure.
Right Systole
Enter right systole blood pressure.
Right Diastole
Enter right diastole blood pressure.

[Figure 6.7 Study Information – Vascular]

Abdomen
Select Abdomen from Application.

[Figure 6.8 Study Information – Abdomen]
Height
Enter the patient's height. Enter height in cm, ft, or in.
Weight
Enter the patient's weight. Enter weight in kg, lb, or oz.

MSK
Select MSK from Application.

[Figure 6.9 Study Information - MSK]

Pediatric
Select Pediatric from Application.

[Figure 6.10 Study Information – Pediatric]

Small Parts
Select Small Parts from Application.

[Figure 6.11 Study Information – Small Parts]

Cardiac
Select Cardiac in Application.

[Figure 6.12 Study Information – Cardiac]
Height
Enter the patient's height in cm, ft and in.
Weight
Enter the patient's body weight in kg, lb and oz.
BSA
When the height and weight are entered, the BSA (Body Surface Area) is automatically calculated and displayed.
Searching Patient Information
Search
Search through the information stored in the system.

[Figure 6.13 Search]
- When a search term is entered in the search field, the list of patients who match the criteria will be displayed. Click the Search By button to search patients by condition.
- Double click a list where a desirable patient list is selected. The selected patient information will be applied to the system.
Show/Hide
Show/Hide buttons or lists in the search area.
Search by
User can enter a search criteria.
Exam List
Show the exam list for a patient selected from the patient list.

[Figure 6.14 Exam List]
Back to Search
Return to the search area.
Review Images
Show the image of a selected exam on Review Images screen.

[Figure 6.15 Review Images]
① Continue Exam: This is activated when an exam that is less than 24 hours old is selected. The page moves to 2D image screen of the selected exam. ② Hide Annotation: Hide any annotations made to an image. ③ View Scale: Set the image size for review. Select either the Small, Mid, or Large buttons. ④ Favorite: Add/remove a selected image from Favorites. ⑤ Select/Deselect All: Select/deselect all images regardless of the exam date.
SonoView
You can review a selected exam from SonoView.
Review Exam
The exams that were performed more than 24 hours before present, and their images, are displayed. When an exam is selected, the button is activated. When enabled, you can continue to view the exam images.
Continue Exam
Exams that were performed more than 24 hours ago can be edited. When an exam is selected, the button is activated. You can use this button to go to the edit exam page.
EzCompare
Press this button to view dual image of a selected exam image and a live scan image currently filming.
Send
Send images to the DICOM server.
Delete
Delete the exam.

Worklist Search
Perform a search by connecting to the DICOM Modality Worklist server in the hospital network.

NOTE: A worklist search is available only when DICOM is enabled. You can select the worklist server in Setup > Connectivity > DICOM. Refer to the 'DICOM Settings' section of 'Chapter 3. Utilities'
- Select Worklist from Patient on the monitor or touch screen.
- After entering at least one item out of Patient ID, Last Name, Worklist Number (Accession Number), Procedure ID, Exam Date, and/or name of the equipment being used for exam (AE Title), click Search. The list of patients who match the criteria will be displayed.

Clicking items such as Date/Time or Patient Name sorts entries into alphabetical or numerical order for the selected criteria.
Select Today, Last Week, Last Month, or Manual as Exam Date.
After running a search, the date of the last update is displayed in the bottom right corner.
- Double click on the desired patient list. This applies the selected patient information to the system.

[Figure 6.16 Search – Worklist]
Diagnosis Modes
Information 7-3
Types of Diagnosis Modes 7-3
Basic Use 7-5
Basic Mode....7-8
2D Mode....7-8
M Mode 7-23
Color Doppler Mode 7-26
Power Doppler Mode 7-31
PW Spectral Doppler Mode....7-33
CW Spectral Doppler Mode 7-38
TDI Mode....7-40
TDW Mode....7-41
ElastoScan Mode....7-42
■ Combined Mode....7-51
2D/C/PW Mode....7-51
2D/PD/PW Mode....7-51
2D/C/CW Mode....7-51
2D/C/M Mode....7-51
2D/TDI/TDW Mode....7-52
Dual Live Mode....7-52
■ Multi-Image Mode 7-53
Dual Mode....7-53
Quad Mode 7-54

☐ 3D/4D Mode ....7-55
3D Stand By....7-59 3D View - MPR 7-61 5D Follicle 7-69 VOCAL....7-71 3D XI 7-78 XI VOCAL 7-88 XI STIC 7-95 4D 7-98 3D Utility Menu....7-99
Information
Types of Diagnosis Modes
This product supports a variety of diagnosis modes including Basic Mode, Combined Mode, Multi-Image Mode, and 3D/4D Mode.
Basic Mode: Consists of different modes, each of which has a specific usage and function. By default, 2D Mode is applied together with another mode. Combined Mode: For an image, two or three Basic Modes are applied at the same time. By default, 2D Mode is applied together with another mode. An image is viewed in a single screen. Multi-Image Mode: The screen is divided into two (dual) or four (quad) sub screens, each of which is used to view an image. Since each sub screen can display a different image, it can be a very useful feature, allowing multilateral views of an organ. 3D/4D Mode: 3D and 4D images can be obtained.
The types of diagnosis mode that are available with the product are shown below:
| Mode Type | |
| Basic Mode | 2D ModeColor Doppler ModePower Doppler ModeM ModePW Spectral Doppler ModeCW Spectral Doppler ModeTDI (Tissue Doppler Imaging) ModeTDW (Tissue Doppler Wave) ModeElastoScan Mode |
| Combined Mode | 2D/C/PW Mode2D/PD/PW Mode2D/C/CW Mode2D/PD/CW Mode2D/C/M Mode2D/TDI/TDW ModeDual Live Mode |
| Multi-Image Mode | Dual ModeQuad Mode |
| 3D/4D Mode | 3D Mode4D Mode |

NOTE: The functionalities for each mode may be restricted by the selected probe.
Basic Use
The items that are commonly used in each diagnosis mode are shown below:

Using the Control Panel
The items that can be used in each diagnosis mode are provided as menu items. You can change the image format or optimize an image to facilitate your diagnosis.
Gain
Use the dial-button on the control panel. The Gain button appears differently depending on the diagnosis mode, but it is usually in the form of a dial-button used to select a diagnosis mode.
You can adjust the brightness of an image. If you rotate the Gain dial-button clockwise, its value increases.
TGC (Time Gain Compensation)
Tap the TGC button on the touch screen to open the TGC window for use.
In general, ultrasound penetration gets weaker with depth. TGC can be used to compensate for this effect.
The product provides eight TGC sliders for varying depths, allowing you to adjust Gain by area. Among the eight sliders, the top slider represents the shallowest area, while the lower sliders represent the deeper ones.
Move the slider to the right to increase Gain, brightening the image.
When the TGC preset is selected, the TGC preset is applied to the system and then the system is switched to diagnosis mode. The user can adjust TGC as needed, and press Save to save it for use.
Focus
Use the Focus switch on the control panel to adjust the location of the focus.
Depth
Use the Depth switch on the control panel to adjust the image scanning depth. The allowable range for adjustment varies depending on the probe type.
Zoom
Use the Zoom dial-button on the control panel.
You can magnify an image. An image can be magnified by either Read Zoom or Write Zoom.
Read Zoom: This function is used for zooming in on images that are stored on the hard disk.
- Rotate the Zoom dial-button on the control panel.
- Adjust the position of the zoomed-in screen with the trackball.
- Observe the magnified image.
Write Zoom: This function allows you to magnify and scan an image in real time.
- Press the Zoom dial-button on the control panel. The Write Zoom box will appear on the screen.
- Use the Change button to move and resize the Zoom box.
- To finish Zoom mode, press the Zoom dial-button once again. Or press the Exit button on the control panel.
Quick Scan

NOTE: The Quick Scan function is available in all probes' applications.
Use the Q Scan button on the control panel. The 'Q Scan' icon mark will appear at the top of an image.
In 2D Mode, it is used to optimize the contrast and brightness of an image by adjusting Gain and TGC automatically. In PW Spectral Doppler Mode, it is used to optimize the spectrum by adjusting Scale and Baseline automatically.
To finish QuickScan mode, press the Exit button on the control panel.

Using Touch Screen Menus
The items that can be used in each diagnosis mode are provided as touch screen menu items. You can change the image format or optimize an image to facilitate your diagnosis.
- In diagnosis mode, the menu for the diagnosis mode currently being used is displayed on the touch screen. In a combined mode that uses more than one diagnosis mode, tap the relevant tab on the touch screen to specify settings for each mode.
- Select a value by pressing a button or by rotating the dial-button on the control panel.
Basic Mode
2D Mode
This basic mode, also referred to as B Mode (Brightness mode), provides scan planes of organs. This is used to display two-dimensional anatomy images in the direction of scanning in real time.

Entering 2D Mode

NOTE: Because 2D Mode is applied by default for all diagnosis modes, it cannot be terminated.
Press the 2D dial-button on the control panel.
If you press the 2D dial-button in other diagnosis modes, it will switch to the basic 2D Mode.

2D Mode Menu

[Figure 7.1 2D Mode - Touch Screen]
Harmonic
Turn this function on or off by tapping Harmonic on the touch screen. The image information will show the 'Har' mark. This product provides the OHI (Optimal Harmonic Imaging) high frequency function to optimize images.

NOTE: The Harmonic function only works with certain probes.
Frequency
Tap Frequency on the touch screen. This enables you to configure the probe's frequency.
The selected frequency is displayed in the image area of the monitor, allowing you to easily determine the state of the current frequency.
▶ Res2 (Resolution): The highest frequency ▶ Res1 (Resolution): High frequency ▶ Gen (General): General frequency Pen1 (Penetration): Low frequency Pen2 (Penetration): The lowest frequency
Line Density
Tap Line Density on the touch screen. Set the scan line density. Select Low, Med, or High by tapping the button.
Selecting High increases the number of scan lines and improves the image resolution. However, the frame rate is reduced.
L/R Flip (Left/Right Flip)
Tap L/R Flip on the touch screen.
Pressing this button flips the image horizontally (Left/Right Flip). The direction marker on the top of the image indicates the orientation of the current image.
S-Harmonic
Turn this function on or off by tapping S-Harmonic on the touch screen. The 'S-HAR' mark will be displayed in the Image Information.
Tap the S-Harmonic button on the touch screen. S-Harmonic inverts the pulses to sharpen the displayed image. If it is On, the phrase 'S-Harmonic' will be shown in the image information.

NOTE:
The Harmonic and S-Harmonic functions are only available for specific probes. ▶ S-Harmonic function may show lines in the near field of certain probes when non-scan mode is used. This is due to the technical features that improve the sensitivity and resolution of the ultrasound images and does not cause any problems in scan mode.
ClearVision
Tap ClearVision on the touch screen. When turned on, ClearVision removes noise from the image and enhances outlines to help show the image more clearly. The 'ClearVision' mark will be displayed in the Image Information.
Tap the touch screen button to select an Index between 1 and 5. Select Off to turn off the feature.
Biopsy Guide
Tap Biopsy Guide on the touch screen.

NOTE: Make sure to adjust the biopsy guideline before using the biopsy feature.

[Figure 7.2 Biopsy]
Tips!
Setting the Biopsy Guideline

NOTE:
If the system is rebooted, the biopsy guideline settings are restored to the default. Make sure to adjust the biopsy guideline before using the biopsy feature.
▶ Support for Multiple Guidelines
If the probe supports multiple biopsy angles, the user may select the biopsy guide angle as desired.
Starting and Finishing a Biopsy
- Tap Biopsy Guide on the touch screen.
- Insert the needle along the guideline. And then perform the biopsy as desired.
- When all procedures are complete, tap Biopsy Guide again to turn this Off. Biopsy will end.
Biopsy Guide Edit Mode

NOTE:
▶ Only available in 2D mode and Single mode. Not available in Dual Live or Freeze state. Not available in Write Zoom or Read Zoom state.
Starting and Ending Biopsy Guide Edit Mode
- Tap Biopsy Guide on the touch screen.
- Use the Angle dial-button on the control panel to enter Biopsy Guide Edit mode.
- When you enter Biopsy Guide Edit mode, Angle Offset appears.
- Rotate the Angle dial-button to select Angle Offset or Lateral Offset.
- Angle Offset: This is the angle for the currently selected biopsy guide; the maximum adjustable offset is the guideline's outer line angle. Adjustments can be made in 0.1° increments.
- Lateral Offset: This is the lateral direction for the currently selected biopsy guide. Adjustments can be made in 0.1 mm increments.
- Press the Angle dial-button to exit Biopsy Guide Edit mode.
U/D Flip (Up/Down Flip)
Tap U/D Flip on the touch screen.
Each press of the button flips the image vertically.
Dual Live
Tap Dual Live on the touch screen and select On or Off.
View the 2D/Color Doppler images of the scanned area simultaneously in real time.
When viewing Cine images in Dual Live mode, various settings are applied only to the Cine image on the right side, allowing you to compare it with the unaffected Cine image on the left side.
MultiVision
Tap MultiVision on the touch screen. MultiVision uses a multi-beam to improve the image quality displayed on the screen. When this feature is turned on, the 'MultiVision' mark will be displayed in the image information. Tap the button on the touch screen to set MultiVision to Low, Med, or High. Note that MultiVision can only be set to Low in 2D/C Mode. Select Off to turn off the feature.

NOTE: This feature only works with certain probes. For more information, refer to 'Probe List' in 'Chapter 5. Probes'.
Trapezoid
Tap Trapezoidal on the touch screen to turn it on or off.
The rectangular frame provided by a Linear Probe is changed to a trapezoidal shape. This allows a wider view of an image.

NOTE: Trapezoid function only works with Linear probes.
Rotation
Tap Rotation on the touch screen to rotate a 2D image. Select a rotation angle to either 0° , 90° , 180° , or 270° .
ElastoScan
Tap ElastoScan on the touch screen. For more information, please refer to 'ElastoScan Mode' in this chapter.

NOTE:
▶ ElastoScan is an optional feature of this product. This feature only works with certain probes. For more information, refer to 'Probe List' in 'Chapter 5. Probes'.
Panoramic
Tap Panoramic on the touch screen. For more information, please refer to 'Panoramic' in this chapter.

NOTE:
▶ Panoramic is an optional feature of this product. This feature only works with certain probes. For more information, refer to 'Probe List' in 'Chapter 5. Probes'.
Needle Mate+
Tap Needle Mate+ on the touch screen. Needle Mate+ is a feature that helps visualize the biopsy needle clearly in an ultrasound image. Choose the direction to optimize the image between From Left and From Right. Select Off to turn off the feature.
| From Left From Right | ||
| Description | From the upper left to lower right or from the lower right to upper left. | From the upper right to lower left or from the lower left to upper right. |
| Purpose | Select this if the needle is being inserted from the upper left to lower right or from the lower right to upper left. | Select this if the needle is being inserted from the upper right to lower left or from the lower left to upper right. |
When turned on, Needle Enhance and Needle Angle are enabled.

NOTE:
▶ Needle Mate+ is an optional feature of this product. This feature only works with certain probes. For more information, refer to 'Probe List' in 'Chapter 5. Probes'.
Needle Enhance
This feature only works when Needle Mate+ is on. Select 1, 2, 3, or 4 by tapping the button on the touch screen to select the degree of needle enhancement.
1: The needle is thin with weak enhancement. ▶ 2: The needle is thin with strong enhancement. 3: The needle is thick with weak enhancement. 4: The needle is thick with strong enhancement.
Needle Angle
Needle Angle is a function to set the needle injection angle for efficient beam steering and needle enhancement. Select one from either Shallow, Medium, or Steep.
DR (Dynamic Range)
Select a value between 30 and 256 by using the DR dial-button on the touch screen. The Dynamic Range function adjusts contrast by changing the ratio of the minimum and maximum values of the input signals. When this value increases, the image is displayed more smoothly.
Frame Average
Tap the Frame Average dial-button on the touch screen. Select a value between 0 and 9 by using the dial-button on the touch screen. Use this to minimize the appearance of speckles in updated images when the same location is scanned repeatedly.
Focus Number
Change the number of focus points at the target location you wish to study. Select a value between 1 and 4 by using the dial-button on the touch screen.
Reject
Use the Reject dial-button on the touch screen to select a value between 0 and 30.
This function is used to eliminate noise or low level echo for clearer signals.
Image Size
Select a value between 70 and 100 by tapping the Image Size button on the touch screen. Used to specify 2D image size.
Gray Map
Select a value between 1 and 12 by tapping the Gray Map dial-button on the touch screen. This changes the 2D Post Curve.
Chroma Map
Use the Chroma Map dial-button on the touch screen to select either Off or a value between 1 and 11.
Edge Enhance
Allows you to view more accurate images of organ or tissue boundaries. A higher value provides more accurate images of boundaries.
Select a value between -3 and 3 by tapping the touch screen button or using the Soft Menu dial-button on the control panel.
Power
Use the Power dial-button on the touch screen to select a value between 2 and 100 for the power of the ultrasound output.
Scan Area
Use the Scan Area dial-button on the touch screen to select a value between 40 and 100 as the image width (%). Increasing the image width reduces the frame rate. You may also press the Change button on the control panel to change the trackball state to ROI Size and then use the trackball to adjust the view area.
Steer Angle
You can adjust the angle of the ultrasound beam without moving the probe by using the Steer Angle dial-button.

NOTE: The Steer function only appears on the soft menu when a linear probe is used.

CEUS+
This mode is used for imaging the contrasting effect within blood vessels by infusing an ultrasound contrast agent.

NOTE:
CEUS+ is supported in all countries except the USA. CEUS+ is an optional feature of this product.

[Figure 7.3 CEUS+ - Touch Screen]
Tap CEUS+ on the touch screen to select on or off. When 'On' is selected, the touch screen is switched to the CEUS+ screen.

CEUS+ Mode
CEUS stands for Contrast Enhanced Ultrasound, and aids the diagnosis by enhancing the contrast on the ultrasound image. When entering CEUS+ mode, Dual mode is displayed by default. ▶ Functions such as Harmonic and S-Harmonic are not available.
Timer 1/Timer 2
Starts or resets the timer.
Press the button to start the timer. When the button is re-tapped, the timer stops and the time is reset. If you don't tap the button again, the timer will keep running even when the CEUS+ On/Off state is changed.
Dual Live
Tap Dual Live on the touch screen and select On or Off. When 'On' is selected, Dual Live mode is activated to display both Contrast and 2D images side-by-side. When 'Off' is selected, Single mode is activated.
Ref. Line
This option displays the Ref. Line on an image. Select On or Off by tapping Ref. Line on the touch screen. The Ref. Line in CEUS+ mode is used to indicate the same location on the Agent and Tissue images in Dual Live mode.
Contrast
Select a CEUS+ type. Tap the button on the touch screen to select either Contrast or 2D.
Contrast: It displays contrast-applied images. ▶ 2D: Displays image in 2D Mode.
Flash
Tap the button on the touch screen and select On or Off. For the defined frames in Frame, brighter images are produced.
Flash Frame
Set a value between 3 and 100 frames to specify the period of time during which the Flash function will operate.
MI Control
Select a value for the mechanical index by tapping the touch screen button or by using the Soft Menu dial-button on the control panel.
Frame Average
Frame Avg averages the present image and the previous image when an image is updated. Select a value between 0 and 9 by tapping the touch screen button or using the Soft Menu dial-button on the control panel.
Reject
This function reduces noises or echoes from an image in order to make the image clearer. Select a value between 0 and 30 by tapping a button on the touch screen or using the dial-button.
Cine Play
Play or stop Cine images.
Trim Start
After specifying the position of the first frame by rotating the dial-button or using the trackball, tap the dial-button to save it.
Trim End
After specifying the position of the last frame by rotating the dial-button or using the trackball, press the dial-button to save it.
Cine Speed
Rotate the dial-button to adjust the auto playback speed.

NOTE: For more information on other touch screen items, refer to the '2D Mode'. (The setting values may vary depending on the selected probe type, application, and other settings.)

Panoramic

NOTE: This feature only works with certain probes. For more information, refer to 'Probe List' in 'Chapter 5. Probes'.
Panoramic Imaging is the function that acquires images over a wider range by using continuous ultrasonographic images. Up to 1,000 frames can be used.
Acquiring a Panoramic Image
- Tap Panoramic on the touch screen. The touch screen will be switched to the Panoramic Ready screen.
- Tap Start/Stop on the touch screen. The system will begin to acquire a panoramic image.
- Press Start/Stop to finish acquiring the panoramic image. The touch screen will switch to the Panoramic Review screen.
Tips!
Things to remember when acquiring a Panoramic image
When scanning a curved surface, ensure that the scan surface and the contact surface of the probe are always at right angles. ▶ Moving in the opposite direction while acquiring an image erases the previously saved frames and saves new frames. The image quality may deteriorate if the contact surface of the probe loses contact with the scan surface. If the scan speed is fast or the angle of the contact surface of the probe changes, artifacts may occur.

[Figure 7.4 Panoramic Ready – Touch Screen]
Reviewing a Panoramic Image

NOTE: L/R Flip, U/D Flip, and Zoom are available only when the Layout is set to Full Screen.

[Figure 7.5 Panoramic Review – Touch Screen]
L/R Flip: Flip the panoramic image horizontally. ▶ U/D Flip: Flip the panoramic image vertically. ▶ Ruler: Press this button to turn this option on or off. When turned on, a ruler is displayed on the panoramic image. Layout: Specify how the panoramic image will be displayed on the screen. – Full Screen: Display the panoramic image in full screen mode. - Up/Down: Display the 2D and panoramic images at the top and bottom of the screen, respectively. - Side by side: Display the 2D and panoramic images on the left and right of the screen, respectively. ▶ Return: Return to the Panoramic Ready screen. Tap the Panoramic button again on the touch screen to exit Panoramic Imaging.
M Mode
The M Mode is used to specify an observation area in a 2D image with the M Line, and display changes over time.
This mode is appropriate for the observation of organs with a lot of movement, such as cardiac valves. The 2D Mode image is also shown, allowing the marking and adjustment of an observation area within the entire image.

[Figure 7.6 M Mode - Touch Screen]

Entering & Exiting M Mode
Entering M Mode
- Press the M dial-button on the control panel.
- Configure settings relating to M image acquisition on the touch screen.
- Press the M dial-button again to enter M Mode and display the M image on the screen.
Exiting M Mode
Press the M dial-button while the M image is being displayed. You can also press the 2D dial-button to exit M Mode.

M Mode Screen
M Line
Use the trackball on the control panel to move the line to the right or left. The M Line indicates the relative position of the M Mode image in the 2D image. Therefore, you can move the M Line to change the observation area.

M Mode Menu
Anatomical M

NOTE: Available only with the phased array probe or the cardiac application.
Tap the Anatomical M dial-button on the touch screen to turn it on or off. If it is 'On', position and angle of AMM Line can be adjusted.
▶ AMM Angle: Use the touch screen dial-button to rotate the cursor line for Anatomical M. You can select an angle from 0 to 179.
Sweep Speed
Tap Sweep Speed on the touch screen. Allows the user to adjust the sweep speed so that he can view various spectrums in a screen.
Display Format
Tap Display Format on the touch screen. Select either Up/Down, Side by Side, or M Only using the touch screen button.
Display Size
The display size can be changed to a range of values between 30/70, 50/50, 70/30 tapping the relevant button on the touch screen. This option is not enabled if Display Format is set to M Only.
M Mode Map
Used to configure the post curve in M. Select a value between 1 and 12 by using the dial-button.
Steer Angle
Adjust the angle of the ultrasound beam. Steered M line can be selected by rotating Steer Angle dial-button.

NOTE: The Steer function appears on the soft menu only when a linear probe is used.

NOTE: For more information on other soft menu options and functions, refer to the '2D Mode' section.
Color Doppler Mode
This mode displays the colored blood flow pattern of the ROI (Region of Interest) within the 2D image. It is appropriate for examining the presence of blood flow, its average speed, and its direction. The 2D Mode image is also shown, allowing the marking and adjustment of the ROI within the entire image.

[Figure 7.7 Color Doppler Mode – Touch Screen]

Entering & Exiting C Mode
Press the Color dial-button on the control panel. Press this button again to exit Color Doppler mode and return to 2D mode.

C Mode Screen
ROI Box
ROI stands for Region of Interest. The ROI Box outlines the area of the 2D image where color (blood flow) information is displayed in Color Doppler Mode.
Use the Change button to reposition or resize the ROI box. Each time you press the Change button, the current state of the ROI box is displayed in the lower middle of the screen.
ROI Position: In this state, the position of the ROI box can be changed. Use the trackball to move and position the ROI Box. ROI Size: In this state, the size of the ROI box can be changed. Use the trackball to move the ROI Box and specify its size.
Color Bar
In Color Doppler mode, the color bar indicates the direction and speed of blood flow. Based on the Baseline at the middle, red indicates the direction and speed of blood flow toward the probe. By contrast, the blue color indicates the direction and speed of blood flow away from the probe.
Adjusting the color bar baseline: Use the Baseline dial-button on the touch screen. If you rotate the Baseline dial-button clockwise, the baseline on the color bar rises.

C Mode Menu
Hide Color
Select C mode to use. If you tap Hide Color on the touch screen, the Color mode image will not be displayed in the Color ROI area, and only the B mode image will be displayed. Tap this button again to cancel the selection and return to Color + B/W mode.
Frequency
Tap Frequency on the touch screen. This enables you to configure the probe's frequency.
The selected frequency is displayed in the title area, allowing you to determine the state of the current frequency easily.
Line Density
Select the density of the scan line. Select Low, Med, or High by tapping the button on the touch screen or using the dial-button.
Selecting High increases the number of scan lines and improves the image resolution. However, the frame rate is reduced.
Invert
Tap Invert on the touch screen. The color bar is inverted each time this button is tapped. When the color bar is inverted, the color displayed on the image is also inverted.
Vel + Var
Tap Vel + Var on the touch screen to turn it on or off. It expresses the blood flow velocity in green on the Color Bar. Vel and Var stand for Velocity and Variance respectively.

NOTE: Vel + Var feature is available in the following applications and presets:
Cardiac - Adult Echo, Ped Echo, Arotic Arch OB - Fetal Heart ▶ Vascular - Carotid, Arterial, Venous
TDI
Change to TDI mode. TDI stands for Tissue Doppler Imaging.

NOTE:
This can only be used when the cardiac application is selected in Phased Array Probe. For more information on TDI, refer to 'TDI Mode'.
Auto Gain
Tap Auto Gain on the touch screen to turn it on or off. If it is turned on, the brightness of an image will be adjusted automatically.

NOTE: Auto Gain is applied only when the mode is live for the Carotid and Arterial presets of the linear probe.
Steer Invert
Tap Steer Invert on the touch screen. Tapping the button inverts the angle of the Sample Volume.

NOTE: Steer Invert is only enabled when using a linear probe.
Scale
Use by rotating the Scale dial-button on the touch screen. Rotating Scale dial-button clockwise increases the scale, extending the range of blood flow speed. Rotating the dial-button in the counter-clockwise decreases the scale, displaying a narrower range of blood flow speed.
Sensitivity
It enhances sensitivity and accuracy or frame rate of color images.
Adjust the Sensitivity value between 0 and 5 by using the Sensitivity dial-button on the touch screen.
Filter
Select Filter on the touch screen or use the dial-button on the control panel to select a value between 1 and 4.
This is an electric filter to remove the low-frequency doppler signals generated by the movement of blood vessel walls. Adjust the Cutoff Frequency to remove doppler signals whose frequency is lower than the cutoff frequency.
BaseLine
Rotating the dial-button clockwise increases the Baseline. In Color Doppler mode, the color bar indicates the direction and speed of blood flow. Based on the Baseline at the middle, red indicates the direction and speed of blood flow toward the probe. By contrast, the blue color indicates the direction and speed of blood flow away from the probe.
Color Map
Used to configure the post curve in a color image. Select a value between 1 and 12 by using the Color Map dial-button on the touch screen. This can help to display blood flow with greater clarity.
Steer Angle
Adjust the angle of the ultrasound beam. Loss of color information resulting from the angle of the ultrasound beam is minimized. You can adjust the angle of the ultrasound beam by using the Steer Angle dial-button on the touch screen.

NOTE: The Steer function only appears on the soft menu when a linear probe is used.
Balance
Select a balance value from 0 to 16 by using the Balance dial-button on the touch screen. The range of a color image is adjusted by comparing the gray levels of 2D images with the doppler signal values of color images. When the Balance value increases, the color image also appears in the part where the gray level of a 2D image is high (the bright part), increasing the range of the color image.
Smoothing (Spatial Filter)
Select a smoothing value between 0 and 5 by using the Smoothing dial-button. This allows smoother display of color images.

NOTE: For more information on other touch screen items, refer to the '2D Mode'. (Different values can be selected.)
Power Doppler Mode
This mode displays the color intensity of blood flow within the ROI in the 2D image.
It is appropriate for examining the presence and amount of blood flow. The 2D Mode image is also shown, allowing the marking and adjustment of the ROI within the entire image.

[Figure 7.8 Power Doppler Mode - Touch Screen]

Entering & Exiting PD Mode
Press the PD button on the control panel to enter PD mode. Press this button again. PD Mode will be terminated and the mode will switch to 2D.

PD Mode Screen
Color Bar
In PD mode, the color bar displayed varies depending on the Power Doppler mode display method that is in use. The color bar indicates the presence of blood flow and its amount. The top of the color bar is the brightest section, where the amount of blood flow is at its highest.
ROI Box
The ROI (Region of Interest) outlines the area of the 2D image where color (blood flow) information is displayed in Power Doppler Mode.

PD Mode Menu
S-Flow
Tap S-Flow on the touch screen. The direction of the blood flow will be displayed.
Auto Gain
Tap Auto Gain on the touch screen to turn it on or off. If it is turned on, the brightness of an image will be adjusted automatically.

NOTE: Auto Gain is applied only when the mode is in live for the Carotid and Arterial presets of the linear probe.

NOTE: The description of touch screen menu items is the same as for 'Color Doppler Mode'.
PW Spectral Doppler Mode
PW stands for Pulse Wave. This mode displays the blood flow speed at a specific blood vessel location within a specific time frame. Distance (depth) information can also be obtained by transmitting pulses over particular time frames.
This mode is useful for measuring low-speed blood flow, such as in the abdomen and peripheral vessels. The 2D Mode image is also shown, allowing the marking and adjustment of an observation area within the entire image.

[Figure 7.9 PW Spectral Doppler Mode - Touch Screen]

Entering & Exiting PW Spectral Doppler Mode
Entering PW Spectral Doppler Mode
- Press the PW button on the control panel to enter PW Spectral Doppler Mode.
- Configure settings relating to Doppler image acquisition on the touch screen.
- Press the PW dial-button again to enter PW Spectral Doppler Mode and display the Doppler image on the screen.
Exiting PW Spectral Doppler Mode
Press the PW dial-button while the Doppler image is being displayed. You can also press the 2D dial-button to exit PW Spectral Doppler Mode.
Press the Set button on the control panel to obtain a spectral doppler image.

NOTE: The doppler image can only be obtained in the D Only or Simultaneous states.
PW Spectral Doppler Mode Screen Sample Volume
The Doppler spectrum is displayed when the Sample Volume is located above the blood flow on the 2D image. The size of the Sample Volume is displayed in mm units. Use the trackball to adjust the position of the Sample Volume.
▶ Moving Sample Volume: Use the trackball on the control panel. ▶ Resizing Sample Volume: Using the touch screen: Select the SV Size dial-button on the touch screen. Adjusting Sample Volume Angle: Rotate the Angle dial-button on the control panel to select a value between -80-80 degrees. Press the Angle dial-button to select either -60, 0, or 60 degrees.
Adjusting Doppler Baseline
Adjust the Baseline by rotating the dial-button on the touch screen.
HPRF (High PRF) Function
Adjust the blood flow above the speed limit at a specific depth. The scale is increased; this can only be used in PW Spectral Doppler mode (D Only).

NOTE: Setup > Imaging > Common > HPRF
Activating HPRF
Increasing the scale at a specific depth to a certain point automatically activates HPRF. The Phantom Gate will appear on the D Line at a position higher than the sample volume. Once HPRF starts, PRF does not increase even if you increase the scale value.
Finishing HPRF
While HPRF is in use, decrease the scale value by one step to finish HPRF. Here, the PRF value becomes the maximum value in the current PW Spectral Doppler mode.
▶ Moving Sample Volume
To move the Sample Volume position in the D Only state, the system calculates PRF values and the Phantom Gate position, and updates them on the PW Spectral Doppler image. HPRF is terminated when HPRF cannot be activated.
When Sample Volume is moved in the 2D Only state, the PRF values don't change.

CAUTION:
The Phantom Gate position can be located outside the 2D image area in Zoom Mode. Make sure that sample volume and Phantom Gate are not placed together in the measuring area. If more than two Sample Volumes are located in the vessels, all Doppler components will appear in the spectrum, causing noise.
PW Spectral Doppler Mode Menu
Simultaneous
Tap Simultaneous on the touch screen. Each time you tap the button, the Simultaneous function turns on or off. If the Simultaneous option is turned on, you can view real-time 2D and spectral Doppler images at the same time. If the option is not turned on, however, you will only be able to view the image from one of the modes. The Simultaneous function decreases Doppler PRF, thus decreasing the measurable speed range.
Invert
Tap Invert on the touch screen. Each time the button is tapped, the speed indicator (+/-) for a spectrum is inverted.
Auto Calc
Tap Auto Calc on the touch screen to select Off, Live, or Frozen.
Live setting: The PS, ED, MD, PI, RI, PS/ED, ED/PS, HR, TAmax, and TAmean are calculated after acquiring the Doppler Trace. Frozen setting: The PS, ED, MD, PI, RI, PS/ED, ED/PS, HR, TAmax, and TAmean are calculated when the image is frozen.
Values that are displayed on screen are selected at Setup > Measurement > AutoCalc. For selecting displayed values, refer to the measurement settings in 'Chapter 3. Utilities'.

CAUTION: The measurements carried out by Auto Trace under Measure and Real Time Automatic Doppler Trace (Automatic Calculator) may be different from each other. This is because the algorithms for these two methods are different. It is recommended that you use Auto Trace under Measure for more accurate measurement.
Tips!
Things to Consider when using Real Time Automatic Doppler Trace
- Aliasing occurs because PRF is too low in comparison to the image speed, or the spectrum is clustered around the baseline because PRF is too high.
- Peak is indistinctive or intermittent such as in Spectral waveforms for veins.
- Meaningful spectrum distinction becomes difficult because Doppler Gain is set too high or too low.
- An index is displayed during the transition time after Sample Volume is moved with the trackball.
- The major spectral signals are cut off because Doppler Wall Filter is set too high.
- Peak Trace is interrupted due to abnormal Doppler noise or artifact, and the heart rate is above approximately 140 bpm.
If any of the above apply, Real Time Automatic Doppler Trace may not produce an accurate trace or results. In addition, during auto calculation, results will not be displayed if the Freeze function is run against inaccurate values.
Threshold
Specify the minimum range of Threshold. It can be adjusted between 0 and 10 by using the touch screen button.
Tips!
Threshold
Threshold is a function to adjust Trace Line in the Auto Trace or Limited Trace. Increase the value to capture a wider area, use a smaller value to make the waveform clearer.
Trace Method
Tap Trace Method on the touch screen. Max or Mean Trace for the selected spectrum is performed. It will not be performed if set to Off.
Trace Direction
Tap Trace Direction on the touch screen. Select the part of the spectrum to calculate with AutoCalc from Both, Above, or Below.
Sound
Adjusts the doppler volume. Select a value between 0 and 100 by rotating the Sound dial-button.
Display Format
Tap Display Format on the touch screen. Select either Up/Down, Side by Side, or Doppler Only by using the button on the touch screen.
Doppler Map
Used to configure the post curve in doppler mode. Select a value between 1 and 12 by using the Doppler Map dial-button on the touch screen.
TDW
Switches to TDW Mode. TDW stands for Tissue Doppler Wave.

NOTE: This can only be used when the cardiac application is selected in Phased Array Probe.

NOTE: For more information on other touch screen menu items, refer to '2D Mode' and 'Color Doppler Mode' sections.
CW Spectral Doppler Mode
CW stands for Continuous Wave. This mode displays the blood flow speed and direction at a specific blood vessel location within a specific time frame. Unlike PW Spectral Doppler Mode, it does not provide Sample Volume.

NOTE:
CW Spectral Doppler mode is an optional feature of this product. This dial-button can be used with the Phased Array or Static CW probes.
Steered CW Spectral Doppler Mode
This mode can only be used if the Phased Array probe is used. The 2D Mode image is also shown, allowing the marking and adjusting of an observation area within the entire image.
Static CW Spectral Doppler Mode
This mode is only available with a Static CW Probe. The 2D image is not displayed.

Entering & Exiting CW Spectral Doppler Mode
Entering CW Spectral Doppler Mode
- Press the CW button on the control panel to enter CW Spectral Doppler Mode.
- Configure settings relating to Doppler image acquisition on the touch screen.
- Press the CW dial-button again to enter CW Spectral Doppler Mode and display the Doppler image on the screen.
Exiting CW Spectral Doppler Mode
Press the CW dial-button while the Doppler image is being displayed. You can also press the 2D dial-button to exit CW Spectral Doppler Mode.

CW Spectral Doppler Mode Menu

[Figure 7.10 CW Spectral Doppler Mode - Touch Screen]

NOTE: The menu for CW Spectral Doppler Mode is identical to that of PW Spectral Doppler Mode.
TDI Mode

NOTE: Appears on the Soft Menu only when using the phased array probe and the cardiac application.
TDI stands for Tissue Doppler Imaging. It visualizes the movement of rapidly moving tissues such as the heart.
It is available in Color Doppler Mode. In Color Doppler Mode, TDI provides color information for cardiac tissues.
Starting and Ending TDI Mode
In Color Doppler Mode, tap TDI on the touch screen. Pressing the button once more switches the mode from TDI to C.

NOTE: For information about the touch screen menu, refer to 'Color Doppler Mode'.
TDW Mode
TDW stands for Tissue Doppler Wave. It visualizes the movement of rapidly moving tissues such as the heart. TDW Mode is available in the PW Spectral Doppler Mode. If it is used in Spectral Doppler Mode along with Color Doppler mode, changes in cardiac tissues over time can be observed.

NOTE: This can only be used when the cardiac application is selected in Phased Array Probe.
Starting and Ending TDW Mode
Tap TDW on the touch screen in PW Spectral Doppler Mode to run it. Pressing the button once more switches the mode from TDW to PW Spectral Doppler.

NOTE: For information about the touch screen menu, refer to the 'PW Spectral Mode' section.
ElastoScan Mode
The elasticity of ROI in a 2D image is shown in color. The 2D Mode image is also shown, allowing the marking and adjustment of the ROI within the scanned image.

NOTE:
▶ ElastoScan mode is an optional feature of this product. The probes, applications, and presets that support ElastoScan are as follows:
| Probe Application Preset | ||
| LA3-14AD | MSK General | |
| Smallparts Breast, Thyroid | ||
| LA4-18BD | MSK General | |
| Smallparts Breast, Thyroid | ||
| EA2-11B | GYN Uterus, Adnexa | |
| Urology Prostate | ||
| ER4-9 | GYN Uterus, Adnexa | |
| Urology Prostate | ||
| V5-9 | GYN Uterus, Adnexa | |
| Urology Prostate | ||
| LA3-16A | MSK General | |
| Smallparts Breast, Thyroid | ||
This function is not available in E mode. Angle, View Area, ECG

[Figure 7.11 ElastoScan Mode]

ElastoScan
An elastogram is the elasticity imaging of an object, based on continuous ultrasound images. The function that measures the elasticity of an object and converts it to images is called ElastoScan. ElastoScan visualizes the presence of a solid mass or stiffness in tissue.
It is already well known that lesions such as tumors are different from healthy tissue in terms of their stiffness. Up until now, palpation has been used for examination, but this method has certain depth limitations.
EDual Mode
In this mode, the elastogram and the 2D image are displayed together on the screen. It can be selected by tapping Single on the control panel and Dual on the touch screen.
To facilitate comparative observation, the 2D image is shown on the left and an E image is shown on the right.

NOTE:
For ElastoScan, Wide Dual View is set as default. For details about Wide Dual View setting, refer to Setup > Imaging > Display Settings in the 'Chapter 3. Utilities'.
![MSK SAMSUNG General LA3-10A 4.0 cm 22 Hz [3b] Gen Gm 50 DB 126 Map7 FA 5 P 90% [8] Pen Gm 50 Map6 HBO FENO 3 1 2 3 4 ROI Position ROI Size SoundType: 6 EFW Input: 5 EFW Mixture: 6 Exit: 1. Stamps/Press 2. Stamps/Print 3. Stamps/Press Tls 0.18 MI 1.36 2016-06-10 11:26:09 AM Sub](/content/2026/06/1369957/images/3d2bf5fa714b01e6023c25839f96ee4b231809b0132ed1cc88881d18afb3e408.jpg)
[Figure 7.12 E Dual Mode]
E Single Mode
In E Single Mode, only an E image is displayed in the screen. It can be selected by pressing the Single button on the control panel.
![MSK SAMSUNG General LA3-16A 4.0 cm 22 Hz [20] Gen Gn 50 Oil 126 Map 7 FA 5 P 90% [E] Pen Gn 50 Map 6 ROI Position | ROI Size Epo/Print: 8 Sat Exit 1 State/Pent EPO Result: 5 2 State/Pent EPO Measure: 6 3 State/Pent Tls 0.18 MI 1.36 2016-06-10 11:25:13 AM Sat: Holt: 3 4 Image: 0](/content/2026/06/1369957/images/99c356d982cb7999826cde989f014d77deab31b9f15769a18972bf5cfb601502.jpg)
[Figure 7.13 E Single Mode]
ROI Mode
ROI stands for Region of Interest. In E Mode, the ROI Box represents the area where elasticity information is processed and shown. Tap ROI Mode on the touch screen to turn ROI Mode on or off.
It is important that only the effective tissues are included in the ROI, avoiding ineffective components such as bone and air, to obtain a good elasticity image. If the ratio of effective image is insufficient, no elasticity image can be displayed even when tissues are pressured. To scan the breast, it is best to set the ROI as wide as possible in the lateral direction and to include the subcutaneous layer and muscle in the axial direction to obtain a good-quality elasticity image, but it can be flexible to adjust to the real situation.
You can adjust the position and size of the ROI box by using the Change button on the control panel. Each time the Change button is pressed, the current status of the ROI Box is shown in the User Information area of the screen.
ROI Position: The position of the ROI Box can be changed. Move the ROI Box with the trackball to confirm the new position. ROI Size: The size of the ROI Box can be changed. Resize the ROI Box with the trackball and press the Change button to confirm the new size.
![MSK SAMSUNG General LA3-16A 4.0 cm 22 Hz [2D] Gen Gm 50 DS 136 Map7 FA 5 P 90% [8] Pen Gm 50 Map6 HSD0 HSD0 ROI Position ROI Size ScorePrint: 6 Sat: 25 Exit: EFW Result: 5 EFW Measure: 6 1. ScorePrint 2. ScorePrint 3. ScorePrint Image: 0](/content/2026/06/1369957/images/6634ef27ffda6c096b43b34614eb468b6f0ce85d1ad725d041c0bfcfb016705b.jpg)
[Figure 7.14 ROI Mode]

NOTE: ROI Mode is available in E Single Mode as well as in E Dual Mode.

Starting and Closing E Mode
In 2D Mode, tap ElastoScan on the touch screen. Press the button again to return to 2D Mode.

E Mode Screen
Color Bar
In E Mode, the color bar indicates the stiffness of a tissue. Regardless of the color, the lower section of the bar indicates that the target area is stiffer than the surrounding tissues, and the upper section indicates that the target area is less stiff than the surrounding tissues.

Compression Guide Bar
The strength and status of compression are shown in different numbers of blocks and colors between Steps 0 and 7. All blocks are empty at Step 0, which appears when the probe is almost stationary, the ROI has insufficient region, or matching between images is poor. Step 1 – Step 2 (in grey) mean inadequate compression, and Step 3 – Step 7 (in blue or green) mean adequate compression; the number of blocks filled represents the number of step level.


Performing a Scan
When using E mode, place a probe onto the surface of an area to observe and position the ROI appropriately on a lesion to observe. Then use the probe to periodically compress the region. When the compression guide is supported, refer to this guide to adjust the size of the compression. If the region being examined is quite deep, it is best to increase the compression.

Precautions for Scanning
To closely observe the location of the lesion, minimize the movement of tissues.
A breast is a complex organ which consists of lactiferous ducts, lactiferous glands, fatty tissue, fibrous tissue, and chest muscles. An axial movement of the probe may cause an unintended movement of the tissues.
The prostate and the thyroid consist of tissues that are simpler than those of a breast and there are relatively fewer unintended movements.

E Mode Menu

[Figure 7.15 ElastoScan Mode – Touch Screen]
Invert
Tap this button on the touch screen to invert the Color Bar. Inverting the color bar also inverts the color displayed on the image.
Dual Live
The elastogram and the 2D image are displayed together on the screen. Pressing Single on the control panel displays the E image only on the screen.
EGain
Adjust the brightness of the elastogram image. Use the dial-button to select a setting between 0 and 100%.
Persistence Level
Specify the speed of change between frames. Use the dial-button to select a setting between 0 and 100%. Selecting a higher value increases the rate of frame change.
Contrast
Adjust the contrast of the elastogram. Use the dial-button to select a setting between 0 and 100%.
Color Map
Select the color of the elastogram. Select a value between 1 and 6 by using the dial-button on the touch screen. If you change the Color Map, the color bar will also change accordingly.
Blending Level
Select the ratio of Alpha Blending. Use the dial-button to select a setting between 0 and 100%. Setting the value to 0% shows an E image only and setting it to 100% shows a 2D image only.
Enhancement
Adjusts the enhancement of the image. Use the dial-button to select a setting between 0 and 100%. A higher setting provides more clearly defined boundaries, at the expense of increased noise.

E-Strain
This function assists with the examination of breast lesions. It calculates the strain ratio between the area of the suspected lesion (ROI) and normal breast fat and displays the results.

NOTE:
▶ E-Strain is an optional feature of this product. ▶ E-Strain is supported in countries except the USA and Canada. The E-Strain button is only activated on the touch screen when the following conditions are satisfied: – Application: Breast, Thyroid - Freeze state
E-Strain
- Press the Freeze button on the control panel and then press the E-Strain button.
- Specify the position and size of the ROI in the area of suspected lesion.
- Make sure that the fat tissue is visible at the same level as, or above, the ROI.
- Adjust the size of the ROI by using the trackball and the Change button.
- Press the Set button. The calculation is automatically performed and the results will be displayed on the screen.
- Target Strain: Mean strain value inside the ROI (supported in all countries except the USA and Canada) – Ref. Strain: Mean strain value of areas considered normal outside the ROI (supported in all countries except the U.S.A. and Canada). – Strain Ratio: Ratio of Strain values (Target Strain/Ref. Strain)
E-Strain Reset
When you tap this button on the touch screen, the E-Strain measurements will be reset.
Combined Mode
In Combined Mode, three different modes are combined, including the default 2D Mode. Note that, in 2D/C Live Mode, only two modes are combined: 2D and Color Doppler Modes.
2D/C/PW Mode
Color Doppler Mode and PW Spectral Doppler Mode are displayed simultaneously.
In Color Doppler Mode, press the PW dial-button on the control panel. Or, in PW Spectral Doppler Mode, press the C dial-button on the control panel.
2D/PD/PW Mode
Power Doppler Mode and PW Spectral Doppler Mode are displayed simultaneously.
In Power Doppler Mode, press the PW dial-button on the control panel. Or, in PW Spectral Doppler Mode, press the PD button on the control panel.
2D/C/CW Mode
Color Doppler Mode and CW Spectral Doppler Mode are displayed simultaneously. This mode is available only with certain probes.
In Color Doppler Mode, press the CW dial-button on the control panel. Or, in CW Spectral Doppler Mode, press the C/Z dial-button on the control panel.
2D/C/M Mode
Color Doppler Mode and M Mode are displayed simultaneously.
2D/TDI/TDW Mode

NOTE: This can only be used when the cardiac application is selected in Phased Array Probe.
TDI Mode and TDW Mode are displayed simultaneously. In PD mode or Color Doppler mode, press the TDI, and then press the PW dial-button.
Dual Live Mode
2D Mode and Color Doppler Mode are displayed simultaneously. In either 2D mode or C mode, select Dual Live on the touch screen.
Changing Combined Mode Format
Changing Trackball State
In Combined Mode, more than two image modes are used at the same time.
Depending on the trackball's state, you can change the position or size of: ROI on the active image; the Sample Volume; M line, etc. Press the Change button on the control panel to change the state of the trackball.
Note that when frozen, you can press the Change button to select the type of Cine image.
Changing the Menu
You can change the menu and touch screen menu items without changing the active image mode. The functions of the buttons on the control panel vary depending on the active image mode.
For example, when the touch screen menu for 2D mode is displayed on the screen in 2D/C/PW mode, you can select another mode from the touch screen menu to switch to that mode.

NOTE: For information on optimizing an image in Combined Mode, please refer to 'Basic Mode'.
Multi-Image Mode
The product supports Dual Mode and Quad Mode.
In Multi-Image Mode, an image can be displayed in different combined modes. Button operations in an active area are the same as in Combined Mode.
Dual Mode
Press the L or R button on the control panel.
You can compare two independent images with each other. Each time you press the L or R button, one of the two images is selected. The currently active image mode is indicated by a blue marker at the top of the image. The buttons and menu operate according to the image mode that is currently in use.
To exit from Dual mode, press the 2D dial-button.
![MSK SAMSUNG General LA3-16A 4.0 cm 51 Hz [20] Geln Grn 32 DR 126 Map 7 FA 5 P 90% HSN7 HsN7 Volume/Print 4 EPII Result 5 EPII Measure 6 Test Set Test 1. Stave Print 2. Stave Print 3. Stave Print Tis 0.36 MI 1.36 2016-06-10 11:26:55 AM Images 0](/content/2026/06/1369957/images/3deb614d2a15b6b4b0c770cf7635f09d3263efc397d39149895bdcda7b8b29f0.jpg)
[Figure 7.16 Dual Mode]
Quad Mode
Press the Quad button on the control panel to enter Quad Mode.
You can compare four different images at the same time. Each time you tap the Quad or Set button, one of the four images is selected. The currently active image mode is indicated by a blue marker at the top of the image. The buttons and menu operate according to the image mode that is currently in use.
To exit Quad mode, press the Single button or the 2D dial-button on the control panel.

[Figure 7.17 Quad Mode]

NOTE: For information on optimizing an image in Multi-image mode, please refer to 'Basic Mode'.
3D/4D Mode
These modes show 3D images of the region being examined. HS50/HS60 provides 3D Mode, 4D Mode (optional) and XI STIC (optional).

NOTE: Standard probes cannot be used for 3D/4D Modes.
Entering and Exiting 3D/4D Mode
Press the 3D/4D button on the control panel. Press the button again to exit 3D/4D mode and return to 2D mode.
3D/4D Mode Screen
ROI Box
In 3D/4D mode, the ROI box is also referred to as the volume box. The box is used to indicate 3D/4D conversion areas.
You can adjust the position and size of the ROI box by using the Change button on the control panel. Each time the Change button is pressed, the state of the ROI box is displayed in the lower middle part of the screen as follows:
ROI Position: In this state, the position of the ROI box can be changed. You can move the ROI box by using the trackball. ROI Size: In this state, the size of the ROI box can be changed. After resizing the ROI box with the trackball, press the Change button to confirm the new size.

3D Mode
The product uses the 3D probe to provide 3D mode and XI STIC.

[Figure 7.18 3D Mode]
3D
In this mode, you can obtain 3D images by using a 3D probe.
XI STIC
In this mode, fetal cardiac cycle and STIC volume data can be obtained with 3D probes. For more information, please refer to 'XI STIC' in this chapter.

NOTE: XI STIC is an option of this product.

4D Mode
In 4D Mode, 3D images can be obtained in real time with 3D probes. This mode is also called Live 3D Mode.

NOTE: 4D Mode is an optional feature of this product.

[Figure 7.19 4D Mode]

Acquiring 3D/4D Images
- Specify the location and size of the ROI Box as desired.
- Set the required parameters in the 3D StandBy screen on the touch screen.
Select the menu tab for 3D/4D mode, and configure in the following order: View Mode → Rendering Preset → Other Items.
- Press the Freeze or Set button on the control panel. The system will start acquiring 3D images.
- Once 3D image acquisition is complete, the 3D View screen will be shown (if configured to do so).
- You can perform diagnosis by optimizing images. Press the 3D/4D button to acquire 3D images again.
Tips!
How to Improve 3D Image Quality
Consider the direction, size and section of the viewpoint as well as the visibility of an object. Before acquiring 3D images, adjust the contrast in 2D Mode. The bigger the ROI box, the slower the rendering speed. Therefore, set an appropriate ROI box size. To see the 3D image of a fetus in frontal view, position the fetal head in the direction of the Direction Mark, putting it in the coronal plane. Then scan the fetus from back to abdomen. To determine surface contour, subjects such as amniotic fluid that do not generate echoes should be insulated with hypo-echoic textures. Once 3D image acquisition is complete, you can adjust the low-threshold level to clean up the image. The general rule is not to adjust High Threshold; set it to the maximum value of 255.
3D Stand By
This screen is displayed on the touch screen when 3D/4D Modes are entered. Set the required parameters to specify how 3D images are acquired.

[Figure 7.20 3D StandBy - Touch Screen]
Menu Tab
Select a tab for 3D/4D Modes. Different tabs are displayed on the touch screen depending on the probe currently being used.

Adjusting 2D Images in 3D/4D Modes
Tap the 2D Menu tab on the touch screen to optimize a 2D image before acquiring the corresponding 3D image.
Once the 2D image is optimized, tap 3D Menu on the touch screen to return to 3D StandBy mode.
View Mode
Select a view mode to use after 3D images are acquired.
3D View
The standard view mode for 3D image review. Press MPR to select.

Rendering Preset
Selects a Preset for 3D images. For more information, refer to the preset information contained in the '3D View-MPR Mode' section.

Scan Quality
Selects the quality of 3D images. Select Low, Med, or High by tapping the button on the touch screen or using the dial-button.
▶ High: Image quality is high, but the speed of capturing (or rendering) 3D image is low. Med2: Provides better image capturing speed and lower image quality than the High setting. Med1: Provides better image capturing speed and lower image quality than the Med2 setting. Low: Provides the highest 3D image capturing speed and the lowest image quality.

NOTE: Scan Quality is not available in the XI STIC menu tab.

Volume Angle
Set the scan angle by tapping the touch screen or using the dial-button. The setting range varies depending on the selected probe.

NOTE: In XI STIC, set the scan angle between 10° and 60° .
3D View - MPR
This view mode is enabled upon acquisition of images when MPR is selected in 3D StandBy. You can optimize 3D images, perform diagnosis, and take measurements.

Changing View Mode
Select the 3D Menu tab on the touch screen to change the view mode.
The Basics of 3D View Mode
Screen Layout
[Figure 7.21 3D View]
① A Plane: The slice image of the axial section. ② B Plane: The slice image of the sagittal section. ③ C Plane: The slice image of the coronal section. 4 3D Image
5 Trackball state indication: The current state of the trackball is displayed at the bottom of the monitor screen. You can select Pointer, Move, or ROI for trackball. Press the Change button on the control panel to change the trackball state. The trackball state changes sequentially each time the button is pressed.
- Pointer: Re-adjusts the position of ROI Box. Pressing the Pointer button on the control panel switches to Pointer state. You can rotate the ROI box around the image axis by using the pointer while pressing the Set button. Press the Change button again to switch to another state.
- Move: Moves the 3D image with the trackball. The acquired 3D image moves as you move the trackball.
- ROI: You can resize the ROI box with the trackball. The ROI box on the 3D image is resized as you move the trackball.
Touch Screen Layout
Only those buttons that are available in the current mode are enabled.

[Figure 7.22 3D View Menu]
Post Gain
Works independently from 2D Gain, with a separate value. The initial value is 0, but this value can be set to any value from -255 to 255. In 4D mode, Post Gain works during Freeze.
Zooming In to/Out of Images
Turn the Zoom dial-button on the control panel to zoom in to/out of images. The current zoom factor appears at the bottom of the screen.
VCT Zoom

NOTE: This option is available with VCT.
After tapping VCT Zoom on the touch screen, press the Zoom dial-button on the control panel to convert it to VCT Zoom function. Turn the dial-button to zoom in to/out of VCT images. The current VCT Zoom factor appears at the bottom of the screen.
Rotating Image around X Axis
Use the M/x dial-button on the control panel.
Rotating Image around Y Axis
Use the PW/y dial-button on the control panel.
Rotating Image around Z Axis
Use the Color/z dial-button on the control panel.
Calculations by Application

NOTE:
In the 4D mode, measurements of XI STIC and VOCAL are not available. In MSV mode, measurement is not available if the display format is 2x3 or higher.
Press the Calculator button on the control panel. Measurement methods are identical to those described in 'Chapter 8. Measurements and Calculations'.
Basic Measurement
Press the Calculator button on the control panel. For more information, refer to 'Basic Measurement' in 'Chapter 8. Measurements and Calculations'.
Tap Save on the touch screen to finish saving. To cancel it, press the Exit button on the control panel.
Saving Images

NOTE: If volume data contains both 4D and 3D Cine images, choose either 4D or 3D for saving.
- Press the Store/Print mapping button on the control panel. The touch screen will show the 3D Data Save screen.
- Specify settings such as Data Type, Save Item, and Volume Format.
- Tap Save on the touch screen to finish saving. To cancel it, press the Exit button on the control panel.
Tips!
Volume Data
- If volume data contains a Cine image, it is saved at the same time.
- If images are saved with volume data, they can be converted to new 3D rendering images with SonoView.
![graph TD A["Save"] --> B["Data Type"] B --> C["Image"] B --> D["Volume"] E["Save Item"] --> F["3D Cine"] E --> G["4D Img. Cine"] E --> H["4D Vol. Cine"] I["Volume Format"] --> J["Default"] I --> K["Cartesian"] L["Exit"] --> M["Save"]](/content/2026/06/1369957/images/9ed35309252ea342d1e4c61b789fa7572533b2b7712a215321ad5d27769cdcc1.jpg)
[Figure 7.23 3D Data Save]
Printing Images
Go to Set up > Customize > Buttons and customize U1, U2, U3, U4, U5, and U6 buttons as you want.

Mode
Select the display format in which 3D images are presented.
Render
Axial, Sagittal, or Coronal plane images are displayed together with the 3D image.
2D
Axial, Sagittal and Coronal plane images, along with OH (Orientation Help), are displayed on the screen. OH indicates the relative position of the currently selected plane in regard to volume data.

Tap Single on the touch screen for a more detailed view.
VCT
Axial, Sagittal and Coronal plane images and combinations of them are displayed. Each plane is displayed with a different colored frame. VCT is an abbreviation for Volume CT.

NOTE: Accept ROI is available in Render. Tap the touch screen to turn it on or off. If it is turned on, ROI is not shown.

Display Format
You can change the display format by using the touch screen. The display format varies for each mode.
▶ Single: Switches to Full screen. Dual: Switches to 2D/3D screen. Quad: Switches to ROI 3D screen.

NOTE: In 2D or VCT mode, only Single or Quad Format can be used.
Ref. Image
Tap the corresponding button on the touch screen to select a reference image from among A, B, and C. The selected image will be highlighted with blue borders.
A: Axial Section B: Sagittal Section C: Coronal Section
OH (Orientation Help)

NOTE: This option may be used in 2D and Render.
Tap the corresponding button on the touch screen to turn it on or off. If it is turned on, a 3D image will be displayed along with OH.
ClearVision
ClearVision removes noise from the image and enhances outlines to help show the image more clearly. You can select off or adjust the index between 1 and 5 by tapping the touch screen.
3D Rotation

NOTE: Mix, Th.Low, and Transparency are available in Render mode.
Select -90° , 90° , or 180° by tapping the button on the touch screen. The coordinate system rotates based on the current 3D image.
Init
When you tap the touch screen, the position information of the 3D image is initialized.

Mix
Configure the mix of Render modes 1 and 2. Use the dial-button to select a ratio between 0:100(%) and 100:0(%).

NOTE: For details on Render mode, refer to 'Render Setup' in '3D Utility Menu'.

Th. Low
Specify the minimum range of Threshold. Use the dial-button to select a value between 0 and 254.

Threshold
This option allows you to adjust the threshold value in order to eliminate unnecessary data from images. As the number increases, cyst elements become more apparent. As the number decreases, bone elements become more apparent.

Select
Select Post Curve. Use the dial-button to select 2D, 3D, 2D Color, or 3D Color.

Position
Set the position of the post curve selected under Select. Use the dial-button to select a value between 0 and 100.

Bias
Set the bias of the post curve selected under Select. Use the dial-button to select a value between -100 and 100.

Transparency
Set the transparency of a 3D image. Use the dial-button to select a value between 20 and 250.
The lowest value (20) is for complete transparency, and the highest value (250) is for complete opacity.
HDVI

NOTE: HDVI is an optional feature of this product.
Filters 3D images to express outlines more vividly and improve images. Five predefined indices are available.
HDVI Index
Tap HDVI Index on the touch screen. Select an index between 1 and 5 by tapping the button on the touch screen.
HDVI Type
Tap the touchscreen button or use the dial-button to select the part to measure among Face, Brain, Early OB, Heart, Spine, Gyn, Breast, Thyroid, Carotid, Abdomen, Kidney, and Surface and adjust the image.

NOTE: The ClearVision and HDVI options are enabled in MPR, MSV, Oblique View mode.

NOTE: The following are enabled in MPR mode: MagiCut, Realistic Vue, 3D Cine, 4D Cine, Render Setup, Preset, Post Processing, Curved ROI, Chroma.
Geometry
Move, magnify, or rotate an image rendered in 3D Mode.
▶ Rotation: Use the X/Y/Z button to rotate the rendered image. Magnification: Use the Zoom button to magnify the rendered image. ▶ Move: Use the trackball to move the rendered image.
5D Follicle

NOTE: 5D Follicle is an optional feature of this product.
This feature assists with diagnosis by estimating the size and volume of a follicle in the ovary. Tap this button on the touch screen to switch to the 5D Follicle screen. Press Exit to finish 5D Follicle.
Run 5D Follicle, and the touch screen will switch to the 5D Follicle screen.
- Specify the position of scanning. Use the Position dial-button on the touch screen to select between Left Ovary and Right Ovary.
- Select a measurement method.
▶ Auto: The follicle is located and measured automatically.
- When the measurement is finished, its result is shown on the screen. The scanned parts are color-coded and numbered.

[Figure 7.24 5D Follicle - Touch Screen]

[Figure 7.25.5D Follicle]
VOCAL
Measure the volume of tissues within the human body. VOCAL is an abbreviation for Virtual Organ Computer Aided anaLysis.
VOCAL can be performed in the following order: VOCAL Define → VOCAL Edit → VOCAL

[Figure 7.26 VOCAL]
VOCAL Define
Specify the settings required for VOCAL execution.

Contour Type
Select the contour line type. A contour line is automatically created for all types except Manual.
Solid
Used for object data with many echoes.
General
Draw a contour line based on a typical object. This is faster than other automatic contour types but less accurate.
Prostate
Used for prostate data.
Cystic
Used for object data with fewer echoes.
Sphere
After creating a spherical object, edit its contour to make it into the desired shape.
Manual
Create the desired shape of an object manually.

Ref. Image
Tap the corresponding button on the touch screen to select a reference image from among A, B, and C. The selected image will be highlighted with blue borders.

Step Angle
Set the rotation angle. Select 12, 18 or 30 by tapping the corresponding button on the touch screen.

Moving Pole Points by Using the Trackball
Set the range to perform VOCAL in a reference image. In a reference image, Pole 1 indicates the position of the upper arrow and Pole 2 indicates the position of the lower arrow. Adjust by using either the trackball and Set button on the control panel or the dial-button on the touch screen.

Init
When you tap the touch screen, the position information of the 3D image is initialized.

Next
Tap the touch screen to begin creation of VOCAL data.

When Contour Type is set to Manual
- Tap Next on the touch screen. The Image Position screen will be displayed on the touch screen.
- After tapping the Set button over an image, move the trackball to contour.
Tap Next to move to the next frame. Tap Previous to move to the previous frame.
- Tap Done on the touch screen. Start VOCAL.
If you tap Done without contouring, VOCAL is performed over a sphere.

[Figure 7.27 VOCAL Define - Touch Screen]
VOCAL Edit
Once VOCAL data is created, volume information will be displayed on the screen. In VOCAL Edit Mode, you can modify or redo the existing contour lines.

Shell Mode
Set the shell of an object based on its contour line.
Off
Do not use Shell Mode. The created contour and the shell overlap.
Inside
The shell is drawn inside the contour generated by the specified Shell Thick.
Outside
The shell is drawn outside the contour generated by the specified Shell Thick.
Symmetric
Half the shell is drawn inside the contour and the other half is drawn outside the contour, with each drawn at half of the specified Shell Thick.

Shell Thick.
Set the shell thickness of an object. Use the dial-button to select a value between 1 and 20mm. This option appears on the touch screen only when Shell Mode is used.

Image Position
Review contour lines for each frame. Use Previous and Next to move through frames.

Multi Edit
Modify more than one contour line at once. Tap this button on the touch screen to turn this feature on or off. If it is turned on, up to 6 contour lines can be displayed simultaneously on the screen. When there are more than 6 lines, use the MEV Page dial-button to navigate through pages. MEV is an abbreviation for Multi Edit View.
Use Pole 1 and Pole 2 to edit contour lines. You can also use the trackball and the Set button on the control panel to edit contour lines. Once editing is complete, tap Exit Multi Edit again to turn it off.

Clear Contour
When you tap this button on the touch screen, the VOCAL data are deleted, the settings are maintained, and the system returns to the VOCAL Define step.

New Contour
Tap the button on the touch screen to move from VOCAL Manual page to VOCAL Setup page. This button only appears if Contour Type is set to Manual.

Accept Contour
Tap the button on the touch screen to apply changes. The screen will switch to the VOCAL data review screen.

[Figure 7.28 VOCAL Edit – Touch Screen]
VOCAL
Optimizes VOCAL data for review.

VOCAL Mode
Specify how VOCAL data are presented.
ROI 3D
Shows images in the Axial, Sagittal and Coronal planes, together with VOCAL data.
Fixed 3D
Shows images in the Axial, Sagittal and Coronal planes, together with 3D images for VOCAL data.
VCT
Shows actual combinations of images in the Axial, Sagittal and Coronal planes and VOCAL data. Use the VCT Type dial-button to select one from Type 1-Type 8.

Display Format
When you tap this button on the touch screen and select Single or Quad, the VOCAL data are displayed in full screen. This can be used in every mode. When you press this button again, the display returns to the previous screen.

Ref. Image
Tap the corresponding button on the touch screen to select a reference image from among A, B, and C. The selected image will be highlighted with blue borders.

VOCAL Edit
Tap the corresponding button on the touch screen to return to the VOCAL Edit stage.

Init
When you tap the touch screen, the position information of the 3D image is initialized.

Mode

NOTE: This option is available with ROI 3D.
Specify how the surface of VOCAL data is presented. Use the dial-button to select Surface or Wireframe.
Surface: VOCAL data are represented using the method of expressing the exterior of images by curves. ▶ Wireframe: VOCAL data are represented by dots and lines.

[Figure 7.29 VOCAL - Touch Screen]
3D XI

NOTE:
3D XI is an optional feature of this product. 3D XI is available only when 3D probes are used.
This view mode is enabled if 3D image acquisition is completed when MSV or Oblique View is selected in 3D StandBy. An image can be viewed in multiple slices.
MSV
An image can be viewed in multiple slices. MSV is an abbreviation for Multi-Slice View.

[Figure 7.30 Multi Slice View]

NOTE: Available options are Calculator, Caliper, Text and Arrow.

MSV Screen
The images sliced at the thickness set in Slice Thick. are displayed on the screen. Slice Number/Total Number of Slices is shown at the bottom of each slice image. The currently selected slice image is indicated by a blue contour.
The image information displays the current mode, Ref. Image, and Slice Thick.
Rotating Image
Rotating the reference plane affects all other planes.
X-axis Rotation: Use the M/X dial-button on the control panel. If the trackball is in Pointer mode, place the Pointer near the X-axis in an image and then move the trackball while pressing the Set button. Y-axis Rotation: Use the PW/Y dial-button on the control panel. If the trackball is in Pointer mode, place the Pointer near the Y-axis in an image and then move the trackball while pressing the Set button. Z-axis Rotation: Use the Color/Z dial-button on the control panel. If the trackball is in Pointer mode, place the Pointer near the Z-axis in an image and then move the trackball while pressing the Set button.
Moving Image
Set the trackball in Move Mode, and then move it up/down/left/right. The image will move along the X and Y axes.

[Figure 7.31 Multi Slice View – Touch Screen]
Display Format
Set the layout of slice images. Select from 1*1, 2*1, 2*2, 3*2, 3*3, or 4*3 by tapping the corresponding button on the touch screen. The number of indices that can be displayed simultaneously on the screen varies depending on this setting. If the layout is changed, the selected slice image moves to the first position on the screen.
Ref. Image
Tap the corresponding button on the touch screen to select a reference image from among A, B, and C.
MSV OH
Tap the ON button on the touch screen to turn it on or off. If this function is turned on, the A, B, and C planes of the selected slice image will be displayed on the screen. The selected slice and reference images will be highlighted with blue borders.
ClearVision
ClearVision removes noise from the image and enhances outlines to help show the image more clearly. You can select off or adjust the index between 1 and 5 by tapping the touch screen.
Page
Change the page on the screen. This option can be useful when the total number of slice images exceeds what is specified in Display Format. Select a page by tapping the Previous and Next buttons on the touch screen.
Selected Slice
Select a slice image to observe. Use the dial-button to select an Index. The selected index will be highlighted with blue borders.

Slice Thick.

NOTE: The Slice Thick. represents the slice width in volume data, rather than the actual anatomical position.
Set the cut depth of images. Use the dial-button to select 0.5, 1.0, 2.0, 3.0, 4.0, 5.0, or 10.0mm. Depending on your selection, the number of indices and pages will vary.

Position
Set the position of the post curve. Use the dial-button to select a value between 0 and 100.

Bias
Set the bias of the post curve. Use the dial-button to select a value between -100 and 100.

Ruler
Set the position of the ruler. Use the dial-button to select None, Right, Left, Top, Bottom or All.

Select
Select Post Curve. Use the dial-button to select 2D or 2D Color.

NOTE:
The following 3D Utility Menu items are enabled in MSV mode: Orientation Dot, Chroma, 4D Cine, Post Processing, Preset
For more information on 3D Utility, see '3D Utility Menu'.
Oblique View
After drawing a straight or curved line in the selected image in MSV Mode, you can study the related oblique image. To do this, follow the procedure below:
- Select Display Format and then specify the number of oblique images to study.
- Set the Cut Type.
- Draw a straight or curved line in a reference image by using the trackball and the Set button. An oblique image will appear with the start (S) and end (E) points shown.
If the trackball is in Move state, the position of the line can be changed.
- Optimize the image for observation by using other buttons on the touch screen.

NOTE:
Functions such as Text and Arrow can be used. When Display Format is 2*1, measurement functions such as Calculator and Caliper can be used.

[Figure 7.32 Oblique View]

Oblique View Screen
The reference image that is selected in MSV Mode is displayed on the screen. The reference image is highlighted with blue borders and always placed in the top left corner of the screen.
When more than one line is used for observation, each line is indicated by a different color and number.
The image information displays the current mode, Ref.Image, Oblique Cut Type, and Plumb Size (or Slice Thick.).

NOTE: When OVIX is used, the image information also displays OVIX Line Offset, Mix, Threshold Low, and Render Mode 1, 2.

Direction of View of Oblique Image
The observer is located perpendicular to the section of a reference image. Please see the view direction below:



Display Format
Set the layout of oblique images. Select 2*1, 3*2 or 3*3 by tapping the corresponding button on the touch screen. Depending on this setting, the number of oblique images and the Oblique Cut Type will vary.

Cut Type
Specify how images are cut by tapping the corresponding button on the touch screen.
Line
The oblique image of a straight line can be studied.
Contour
The oblique image of a curved line or contour line can be studied.
Tips!
Multi Line & Multi Contour
If Display Format is set to 3*2 or 3*3, enable Auto Increment to draw more than one line.
Parallel

NOTE: This cannot be used when Display Format is 2*1.
The oblique image of a straight line and its parallel lines can be studied. If a straight line is drawn, its parallel lines are automatically shown on the screen.
Plumb

NOTE: This cannot be used when Display Format is 2*1.
The oblique image of a straight line and its perpendicular lines can be observed. If a straight line is drawn, its perpendicular lines are automatically shown on the screen.
ClearVision
ClearVision removes noise from the image and enhances outlines to help show the image more clearly. You can select off or adjust the index between 1 and 5 by tapping the touch screen.

Image Rotation
Specify the direction of an oblique image. Select 90°, 180° or -90°, by tapping the button on the screen. Select the oblique image that you wish to change the direction of from Selected Slice.

Clear Line
Tap the touch screen to delete an oblique image.

Selected Slice

NOTE: This cannot be used when Display Format is 2*1.
Use the dial-button to select a line. The oblique image of the selected line is indicated by white contour.

Plumb Size

NOTE: This is used only when Cut Type is Plumb.
Use the dial-button to adjust the length of the perpendicular line. The length of the centermost line may be adjusted by 1mm; its current length is displayed in image information.

Slice Thick.

NOTE: This is used only when Oblique Cut Type is Parallel.
Use the dial-button to adjust the interval between the perpendicular lines. The length of the centermost line may be adjusted by 1 mm; its current length is displayed in image information.

Rotate Line

NOTE: This cannot be used when:
▶ Cut Type is Contour; or when the Selected Slice is Selected Slice Select All.
Use the dial-button to rotate a line. When the line is rotated, the oblique image is also changed accordingly.

OVIX

NOTE: This cannot be used when Cut Type is Contour.
OVIX is the abbreviation for Oblique View eXtended, which sets the cross-sectional thickness of an oblique image and shows the image in 3D.
Tap the corresponding button on the touch screen to turn it on or off. When this is on, an OVIX Line appears in the reference image, which indicates the cross-sectional thickness of the oblique image of the reference image.
The thickness of the OVIX Line can be adjusted with OVIX Thick. To change the 3D image settings, select and adjust Setting or OVIX Post Curve from the 3D Utility Menu.

OVIX Thick.
Use the dial-button to adjust the thickness of OVIX. The 3D image for the set thickness appears.

Select
Select Post Curve. Use the dial-button to select 2D, OVIX, 2D Color, or OVIX Color.

Init
Tap the touch screen to delete the Oblique image and reset the position information for Ref. Image.

NOTE: The following 3D utility menu items are enabled in Oblique View mode: Orientation Dot, 4D Cine, Post Processing, Render Setup, Chroma, Preset
For more information on 3D Utility, refer to '3D Utility Menu'.

[Figure 7.33 Oblique View - Touch Screen]
XI VOCAL

NOTE: XI VOCAL is one of the 3D XI features. It is available as an optional feature of this product.
Measure the volume of tissues in 3D XI Mode.

[Figure 7.34 XI VOCAL]

VOCAL vs XI VOCAL
▶ VOCAL: Measures the volume of an object in a general 3D image. Uses rotating slices. XI VOCAL: Measures the volume of an object in the selected reference image in MSV Mode by using parallel slices. Uses parallel slices. Calculates the volume of an object by cutting it into multiple slices.
XI VOCAL is performed in the following order: XI VOCAL Define → XI VOCAL Edit → XI VOCAL
XI VOCAL Define
Specify how slice and contour lines are extracted.
Reference images and slice lines are displayed on the left side of the screen. Slice images with the start (S) and end (E) points of a slice line are displayed on the right side of the screen.

[Figure 7.35 XI VOCAL Define]
Contour Type
Select the contour line type. A contour line is automatically created for all types except Manual.
Solid
Used for object data with many echoes.
Cystic
Used for object data with fewer echoes.
General
Draw a contour line based on a typical object. This is faster than other automatic contour types but less accurate.
Manual
Create the desired shape of an object manually. A contour line can be specified in the XI VOCAL Edit screen.

Ref. Image
Tap the corresponding button on the touch screen to select a reference image from among A, B, and C. The selected image will be highlighted with blue borders.
Ref. Contour
Tap this button on the touch screen to turn this feature on or off. If it is turned on, a contour line can be drawn by using the trackball and the Set button.

Slice Direction
Set the direction of a slice line. Select from Vertical or Horizontal by tapping the corresponding button on the touch screen. Changing the direction of a slice line also changes the slice image displayed on the screen.

Init
When you tap the touch screen, the position information of the 3D image is initialized.

Next
Tap the touch screen to switch to the XI VOCAL Edit screen.

NOTE: If Contour Type is set to Manual, the system switches to the XI VOCAL screen when Next is pressed.

# of Slices
Specify the number of slice images. Use the dial-button to select 5, 10, 15, or 20. Depending on the selected number of images, the interval between slices will vary.

[Figure 7.36 XI VOCAL Define - Touch Screen]
XI VOCAL Edit
Specify the contour extraction range or run XI VOCAL.
Based on slice lines, slice images and pole points will be displayed on the screen. A pole point is a reference point against which an object contour is extracted. Two pole points appear in each slice image.
The selected slice image is highlighted with blue borders. 'The image number/the total number of slice images' is shown at the bottom of each image.
Tips!
Reference Image and Slice Line
These always appear in the bottom right corner of the XI VOCAL Edit screen. They can be useful when the position of a slice image needs to be considered.

[Figure 7.37 XI VOCAL Edit]
Ref. Page
Use the dial-button to change the screen page.
New Contour
Tap the corresponding button on the touch screen to delete the current data and return to the XI VOCAL Define stage.

Accept Contour
Tap the button on the touch screen to apply changes and run XI VOCAL. The system will switch to the XI VOCAL screen.

When Contour Type is set to Manual
Use the Set button and the trackball to draw a contour line before pressing Accept Contour. If you press Accept Contour without drawing a contour line, a general type contour line will be extracted.

[Figure 7.38 XI VOCAL Edit – Touch Screen]
XI VOCAL
Optimize XI VOCAL data for review.
Slice images with their contour line shown and 3D reference images are displayed. The 3D reference image is highlighted with blue borders, and the calculated volume is shown at the bottom of the image.

3D Reference Image
Use XI VOCAL to display an object for which volume has been obtained in 3D. Use the Zoom, M/x, PW/y or Color/z dial-buttons on the control panel to zoom or rotate for observation.

View All Slices
Tap this button on the touch screen to turn this feature on or off. If it is turned on, all XI VOCAL data – including reference image, slice line and slice image – will be displayed simultaneously on the screen.

Edit Contour
Tap the touch screen to return to the XI VOCAL Edit screen. You can edit the contour line by using the trackball and the Set button.

[Figure 7.39 XI VOCAL – Touch Screen]
XI STIC

NOTE:
XI STIC is an optional feature of this product. XI STIC is only available when 3D probes are used. This only appears when the Application is set to OB. When you enter 3D Standby in Fetal Heart Preset, the system will automatically switch to STIC.
This option can be used to obtain the fetal cardiac cycle with volume data on the fetal cardiac area, and to recompile the volume data for display. STIC is an abbreviation for Spatio-Temporal Image Correlation.

[Figure 7.40 XI STIC]
Acquiring XI STIC Images

NOTE: If the motion artifacting is severe, data does not contain the cardiac cycle, or the heart rate cannot be calculated for any other reason, you will be returned to the XI STIC image preparation screen.
- Select the XI STIC tab in the 3D Stand By screen on the touch screen.
- Set the various parameters as you would for acquisition of standard 3D images.
- Press the Freeze or Set button on the control panel. The system will begin acquiring 3D images.
- When image acquisition is complete, XI STIC is displayed on the monitor screen and the XI STIC Confirm screen appears on the touch screen. Check the fetal cardiac cycle calculated.
- Tap Yes on the touch screen to continue. Tap No to cancel and acquire the images again.
- You can perform diagnosis by optimizing images.
Tips!
To Obtain Superior STIC Volume Data
Volume Angle: If the size of the fetal heart is small, use a smaller Volume Angle. ROI Position: Adjust the scan position so that the center of the Volume Angle and the fetal heart are aligned properly. ROI Box: Adjust the size of the volume box so that it nearly fits the size of the fetal heart.

Volume Angle
Set the scan angle. Select a value between 10 and 60 by using the dial-button.

Scan Time
Set the image acquisition time. Use the dial-button to select between 7 and 15 seconds.

Trimester
Set the pregnancy trimester. Use the dial-button to select 1st, 2nd, or 3rd.

Trimester
If the 1st to 3rd trimesters are selected, the recommended scan time and STIC angle are automatically set for the specified trimester. Please see the following table:
| Trimester 1 | st | 2nd | 3rd |
| Scan Time 10 seconds | 12 seconds 15 seconds | ||
| Volume Angle 20° | 25° 30° |
If a scan time and Volume Angle other than those in the above table are set, the trimester is displayed as User Set.

NOTE: For information on other usage, please refer to the "3D Stand By" section.
Reviewing XI STIC Image
XI STIC image is played as a Volume Cine; on the left side of the screen, XI STIC information including HR, Scan Time, Img. Angle, Vol. Angle, and Volume Index are displayed.

NOTE: During Volume Cine, only MPR, MSV, and Oblique View Modes are available.
Press the Freeze button on the control panel to stop Volume Cine playback.
Speed
Select the playback speed for XI STIC images. Use the dial-button to select Very Slow, Slow, Normal, Fast, or Fastest. This rate is based on the fetal heart rate (100%).
Vol. Index
Use the dial-button to select an Index.

NOTE:
This option is only available when Freeze is enabled. Refer to 3D View-MPR and 3D XI for further information on its use.

[Figure 7.41 XI STIC - Touch Screen]
4D

NOTE:
4D Mode is an optional feature of this product. ▶ 4D Mode is available only when 3D probes are used.
In 4D Mode, 3D images can be obtained in real time with 3D probes. This mode is also called Live 3D Mode.
Images can be acquired in the same way as for standard 3D images.

4D Mode Screen
Press the Freeze button on the control panel to switch to the 4D Cine screen.

NOTE: In 4D state, only MPR, MSV, and Oblique View modes are available. For more information, see 3D View-MPR and 3D XI.
3D Utility Menu

3D Cine
The 3D images saved temporarily in the system can be reviewed. The 3D Cine Define screen is displayed on the touch screen.
3D Cine Define
Specify the settings needed for creation of a Cine image.
Rotate Angle
Set the overall rotation angle by tapping the corresponding button on the touch screen. Select a value from 30, 45, 60, 90, 180 or 360°.
Step Angle
Set the rotation angle for a single step in an image by tapping the corresponding button on the touch screen. Select a value from 1, 3, 5, or 15 degrees.

The Difference between Rotate Angle and Step Angle
A Cine image rotates to the angle specified under Rotate Angle. During this process, each rotational step is equivalent to the angle specified under Step Angle. For example, if Rotate Angle is set to 360 degrees and Step Angle is set to 15 degrees, a 3D Cine image rotates to 360 degrees in 25 steps, each of which involves a rotation of 15 degrees.
Rotate Axis
Set the rotational axis by tapping the corresponding button on the touch screen.
Generate
Cine images are generated by applying the current settings. Once the images are generated, the touch screen switches to 3D Cine Review.
Review
Review Cine images generated previously. The touch screen switches to 3D Cine Review.
Start Angle
Use the dial-button to set the start angle of a Cine image. When Start Angle is set, Rotate Angle is cancelled.
End Angle
Use the dial-button to set the end angle of a Cine image. When End Angle is set, Rotate Angle is cancelled.

[Figure 7.42 3D Cine Define - Touch Screen]
3D Cine Review
Specify the settings needed for review of a Cine image.

[Figure 7.43 3D Cine Review]
Play Mode
Select the play mode for Cine images. Select from Loop or Yoyo by using knob button.
▶ Loop: Repeats playback in one direction. ▶ Yoyo: Plays until the end in one direction, and then plays in the reverse direction.
Cine Play
Tap this button on the touch screen to turn this feature on or off. Cine images are played when this is on. If it is turned off, Cine Frame will appear on the touch screen.
New Cine
Clears the current Cine image and creates a new one. The system switches to the 3D Cine Define screen.
Speed
Use the dial-button to select the playback speed of a Cine image. Select from Very Slow, Slow, Normal, Fast, or Fastest.
Trim Start
After specifying the position of the first frame by using the dial-button, tap the dial-button to save it.
Trim End
After specifying the position of the last frame by using the dial-button, tap the dial-button to save it.
Cine Frame

NOTE: This option is available when Play is turned off. It can be adjusted by as many as the total number of Cine frames by using the knob button with the 'current Cine frame number' being shown on the monitor screen.
Select a Cine frame to review. Use the dial-button or move the trackball left or right to select.

[Figure 7.44 3D Cine Review - Touch Screen]

MagiCut

NOTE:
▶ Select this in MPR mode. If MagiCut is enabled, Accept ROI is turned on automatically. After MagiCut is enabled, if the system is switched to a mode other than MPR Mode, MagiCut is disabled.
The MagiCut screen is displayed on the touch screen. You can cut the parts that are not relevant to diagnosis away from 3D images. Set the area to cut by using the trackball and the Set button.
Mode
Set how parts are cut by tapping the corresponding button on the touch screen.
▶ Inside Contour: Cuts the inside of the selected area. ▶ Outside Contour: Cuts the outside of the selected area. Inside Box: Cuts the inside of the box. Outside Box: Cuts the outside of the box. ▶ Small Eraser: Cuts the selected contour line. ▶ Big Eraser: Cuts the selected contour line. This uses a thicker contour line than Small Eraser.
Full Depth
Tap the corresponding button on the touch screen to turn Full Depth on or off. If it is turned on, the entire area will be cut. If it is turned off, Depth will appear on the touch screen.
Depth
Set the cut depth. Use the dial-button to select a value between 1 and 100.
Undo
Cancel the previous task(s).
▶ Undo: Cancel the last task. ▶ Undo All: Cancel all tasks that have been done so far.
Mix
Configure the mix of Render modes 1 and 2. Use the dial-button to select a ratio between 0:100(%) and 100:0(%).
Th.Low
Specify the minimum range of Threshold. Use the dial-button to select a value between 0 and 254.

[Figure 7.45 MagiCut – Touch Screen]

Orientation Dot
Tap the corresponding button on the touch screen to turn the Orientation Dot on or off. When this is on, a dot appears at the center of the reference image.

Render Setup

NOTE: In Oblique View Mode, this option is available only when OVIX is On.
The Settings screen appears on the touch screen. Specify the image rendering method.
After selecting the rendering mode tab on the touch screen, specify the required settings including Render Direction and Render Mode. The specified settings will be displayed in the image information area.
This product provides four Render modes – Gray, Inversion, Color, and SeeThru.

Gray
Specify how volume data acquired with the gray method should be rendered into 3D images.
Render Direction
Set the rendering direction by clicking the corresponding button on the touch screen.
Render Mode 1, 2
Enter Render Modes 1 and 2 by tapping the corresponding buttons on the touch screen.
Surface: Represents 3D images in the Ray-Casting method, which shows the shell of an image with curved surfaces. Smooth: Represents 3D images that are smoother than those created by the Surface method. Max: Represents 3D images at maximum intensity. It can be useful for the observation of bone structures in a human body. Min: Represents 3D images at minimum intensity. It can be useful for observation of vessels or hollow parts in a human body. Light: Represents the depth of 3D images in terms of brightness. X-Ray: Represents 3D images in terms of average intensity. This shows images that are similar to X-ray images.

NOTE: For descriptions of other menu items, see 'Render Setup'.
Mix
Configure the mix of Render modes 1 and 2. Use the dial-button to select a value between 0 and 100.
Th.High
Used to specify the maximum threshold range. Use the dial-button to select a value between 1 and 255.
Th.Low
Specify the minimum range of Threshold. Use the dial-button to select a value between 0 and 254.

Threshold
This option allows you to adjust the threshold value in order to eliminate unnecessary data from images. As the number increases, cyst elements become more apparent. As the number decreases, bone elements become more apparent.
Th.Power
Specify the Threshold for color. Use the dial-button to select a value between 0 and 255. As this value increases, more color is removed from an image.
Color
Specify how the volume data acquired with the Angio/CFM method are rendered into 3D images. Other settings are configured in the same way as Gray Render Mode.
See Thru
Set the method for rendering the data from Gray+Angio, Gray+CFM combinations as a 3D image. Settings other than Render Mode 1 and 2 are configured in the same way as Gray Render Mode.
Transparent-Transparent
Adjust the transparencies of both Gray and Color data, so that the Color data inside the Gray data may be studied. The parts that are obscured by the Gray data are displayed slightly darker.
Transparent-Surface
Adjust the transparency of Gray data, so that the Color data inside the Gray data may be studied. The parts that are obscured by the Gray data are displayed slightly darker.
Max-Transparent
Set Gray data to 'Max', and Color data to 'Transp', to study Color data. The parts that are obscured by the Gray data are displayed slightly lighter.
Max-Surface
Set Gray data to 'Max', and Color data to 'Surface', to study Color data. The parts that are obscured by the Gray data are displayed slightly lighter.
Inversion
This option shows inverted images when the volume data acquired by the gray method is rendered into 3D images. Other settings can be specified in the same way as with Gray Render Mode.

Realistic Vue
Semi-translucent rendering is used for a realistic display of fetal images. When you press the Realistic Vue button, Clear SFVI, HDVI, and VC are turned on.

NOTE:
Realistic Vue is an optional feature of this product. HDVI is an optional feature of this product and only available for HS60.
Light Direction
Use the touch screen or the trackball to reposition the direction of the light that is applied to volume.

NOTE: You need to press the Move Light button to enter trackball mode.
Set Color
Use the dial-button to select the Set Color applied to Realistic Vue.

NOTE: You can set color by using the Hue, Saturation, and Lightness dial-buttons

[Figure 7.47 Realistic Vue - Touch Screen]

Chroma
The Chroma Map screen is displayed on the touch screen. Set the color of 2D and 3D images by tapping the corresponding button on the touch screen.

[Figure 7.48 Chroma Map - Touch Screen]

Post Processing
The Post Processing screen is displayed on the touch screen.

[Figure 7.49 Post Processing – Touch Screen]
Gradient Mask
Adjust the brightness of a specific area in an image. Make a selection by tapping the corresponding button on the touch screen.
Flip Image
Inverts the position of an image.

NOTE: This option is only available in MSV Mode.
Tap the image on the touch screen and flip the image left and right or up and down.
Negative
Tap this button on the touch screen to turn this feature on or off. If it is turned on, the brightness of an image will be inverted.
Auto Contrast
Tap this button on the touch screen to turn this feature on or off. If it is turned on, the contrast of an image will be adjusted automatically.
Threshold
Tap this button on the touch screen to turn this feature on or off. If it is turned on, the Th.Low or Th.High dial-buttons can be used to adjust the threshold.
Sharpen
Tap this button on the touch screen to turn this feature on or off. If it is turned on, the boundary of an image will become more apparent. Select a value between 100 and 400 by using the Sharp dial-button.
3D CI
Tap this button on the touch screen to turn this feature on or off. 3D CI is an abbreviation for 3D Compound Imaging.
If set to On, images are compounded to reduce noise and enhance image quality. Use the 3D CI Offset dial-button to select a value between 1–10 as the distance between images used for 3D CI.

Preset
The Preset screen appears on the touch screen. You can set the currently selected Preset or rename the Rendering Preset. Default is the default value of the system.

[Figure 7.50 Preset - Touch Screen]
Default
Use the default setting (General Preset) for the probe.
Load
When you tap this button on the touch screen, the currently selected Preset is applied to the system. Then the Preset screen is closed.
Save
Tap Save. When you tap this button on the touch screen, the current Preset is saved.
Rename
Change the name of the selected Rendering Preset. When you tap this button on the touch screen, the Name screen appears. After changing the name, tap Save to save the changed name. Tap Close to cancel the change.
Reset
When you tap this button on the touch screen, the Preset is reset to its default value.

4D Cine
The 4D images saved temporarily in the system can be reviewed. The 4D Cine screen is displayed on the touch screen.

NOTE: You can also press the Freeze button in 4D Mode to execute 4D Cine.
Play Mode
Select the play mode for Cine images. Use the touch screen to select either Loop or Yoyo.
▶ Loop: Repeats playback in one direction. ▶ Yoyo: Plays until the end in one direction, and then plays in the reverse direction.
Cine Type
Select the Cine image type by tapping the corresponding button on the touch screen.
Volume: This button appears when Cine images are played. The MPR, MSV, or Oblique buttons are enabled, depending on the state before the 4D Cine was started. Cine images can be played by changing Display Format, etc. ▶ Image: This is the general Cine playing method.
Cine Play
Volume: Tap the Play button on the touch screen to play; tap the Freeze button to freeze. Image: Tap the Play button on the touch screen to play; tap Play again to freeze.
Cine Speed
Set the speed at which Cine images are played. Use the dial-button to select Very Slow, Slow, Normal, Fast, or Fastest.
Vol.Index
Move the trackball or use the dial-button to select an Index. The current volume data number/total number of volume data is displayed.

[Figure 7.51 4D Cine - Touch Screen]

5D NT

NOTE:
5D NT is an optional feature of this product. These options are only available in MPR mode.
This feature locates the Mid-Sagittal View and measures the thickness of the nuchal translucency (NT) of the fetus. The 5D NT window appears on the touch screen.
Use the trackball to place NT Seed in NT area, and press the Set button to display the NT measurement on the A Plane.
Tips!
How To get good results
▶ You can get better results when the Sagittal View of the fetus is captured as accurately as possible. The higher the contrast between the fetus's palate and nasal bone, the better. It is preferable to have the lateral direction of the probe be parallel to the body orientation of the fetus. It is preferable to have the probe's angle to the fetus's nasal bone as close to 30 degrees as possible.
NT Caliper Placement
Tap the touch screen to select the NT measurement type.
On to On: Measure by placing the cursor on the NT's inner-inner.

On to Max Brightness: One side of the cursor is placed on the outside of the NT, and measurement is taken with inner-outer. This method is used when Harmonic is used and one side of the nuchal translucency is blurred.

In to In: Similar to On to On, this method takes measurement with inner-inner, albeit with a narrower cursor interval.

Items
NT: The acquired A, B, and C plane images and the automatically measured NT are displayed in the 3D View screen. IT: Locates mid-sagittal view and tests the fetus for spina bifida.
Assign
Tap the touch screen to include the selected items in the OB report.
5D NT
This function helps the user to easily see the NT measurement process. This assigns a certain thickness to A, B and C sections of the image and displays 3D Rendering image.
Init
The image's position information is reset.
Hide All
Pressing this button will hide the marker and the measurements for NT and IT from the screen.
Auto
Tap the corresponding button on the touch screen to turn this feature on or off. If set to On, automatically locates Mid-Sagittal View. To calculate NT, place the cursor on NT area and press the Set button on the control panel.
Edit
Edit markers for each item that is displayed as a result.
If the trackball's state is Cursor, move the cursor near a + marker.
Press the Set button and move the trackball to edit the marker as desired.
ROI Height
Use the dial-button to adjust the ROI Line's Height to define the scope of the application. The ROI Height may be adjusted while measurements are being taken or in Edit mode.
Marker Size
Tap on the touch screen or use the dial-button to select Small, Medium, or Large.
Measurements and Calculations
■ Measurement Accuracy.... 8-3
Causes of Measurement Errors 8-3
Optimization of Measurement Accuracy .....8-5
Measurement Accuracy Table 8-7
■ Basic Measurements.... 8-9
Distance Measurement....8-13
Circumference and Area measurement....8-23
Volume Measurement....8-26
■ Calculations by Application......8-29
Things to Note....8-29
Common Measurement Methods ....8-33
OB Calculations 8-37
Fetal Heart Calculations....8-46
Gynecology Calculations....8-49
Cardiac Calculations....8-52
Abdomen Calculations 8-66
Pediatric Hips Calculations....8-68
Urology Calculations....8-70
Small Parts Calculations....8-71
Musculoskeletal Calculations 8-75
Vascular Calculations 8-76

Reports 8-85
Viewing Reports 8-85
Report Preview 8-86
Editing Report (Worksheet) 8-89
Adding Comments 8-92
Print Reports 8-94
Saving Reports 8-94
Store SR 8-95
Images 8-95
OB Graph 8-97
Description 8-102
Closing Reports 8-104
Measurement Accuracy
Measurement values can vary depending on the nature of the ultrasound wave, the body's response to ultrasound waves, the measurement tools, algorithms, product settings, probe type, and the actions performed by the user.
Before using this product, make sure to read and understand the following information regarding the causes of measurement errors, and measurement optimization.
Causes of Measurement Errors
Image Resolution
The resolution of an ultrasound image may be limited by spatial causes.
Errors caused by signal range may be minimized by adjusting the focus settings. Optimizing the focus settings increases the resolution of the measurement area. In general, the lateral resolution is lower than the axial resolution. Therefore, measurements should be performed along the axis of the ultrasound beam to obtain accurate values. Gain has a direct impact on the resolution. Gain can be adjusted by using the Gain button for each mode. In general, increasing the frequency of ultrasound enhances resolution.
Pixel Size
This product's ultrasound images consist of pixels. Since a single pixel represents the basic unit of an image, a measurement error may result in the displacement of approximately ±1 pixel when compared to the original image size. However, this error becomes significant only when a narrow area in an image is measured.
Ultrasound Velocity
The average velocity of ultrasound used for measurement is usually 1,540 m/s. The velocity of ultrasound used varies depending on the cell type. The possible range of error is approximately 2-5% depending on the structure of cells (about 2% for typical cells and about 5% for fatty cells).
Doppler Signal Adjustment
During velocity measurement, an error may occur, depending on the cosine angle between the blood flow and the ultrasound beam. For Doppler velocity measurements, the most accurate results can be ensured by aligning the ultrasound beam in parallel with the blood flow. If that is not possible, the angle between them should be adjusted by using the Angle option.
Aliasing
PW Spectral Doppler Mode uses a signal sampling technique to calculate the frequency (or velocity) spectrum. Adjust the baseline or the velocity scale to minimize aliasing. A lower frequency probe can also be used to reduce aliasing.
Calculation Equation
Some of the calculation equations used for clinical purposes originate from hypotheses and approximations. All calculation equations are based on medical reports and articles.
Human Error
Human error may occur due to inappropriate use or lack of experience. This can be minimized through compliance with and thorough understanding of the manuals.
Optimization of Measurement Accuracy

2D Mode
▶ Resolution is in proportion to the frequency of the probe. ▶ Penetration is inversely proportional to the frequency of the probe. The highest resolution can be obtained at the focus of the probe where the ultrasound beam is narrowest. The most accurate measurements can be obtained at the focus depth. As the distance from the focus increases, the beam width increases, which results in lower accuracy. Using the zoom function or minimizing the depth display makes distance or area measurements more accurate.

M Mode
The accuracy of time measurements can be increased by setting the sweep velocity and the display format to higher values. The accuracy of distance measurements can be increased by setting the display format to a higher value.

Doppler Mode
Using lower frequency ultrasound is recommended for measurement of faster blood flows. The size of the sample volume is limited by the axial direction of the ultrasound. ▶ Using lower frequency ultrasound increases penetration. The accuracy of time measurements can be increased when the sweep velocity is increased. The accuracy of velocity measurements can be increased when the vertical scale is set to smaller values. It is most important to use an optimal Doppler angle to enhance the accuracy of velocity measurements.
Color/Power Doppler Mode
A protocol has not been specified for images in Color Doppler Mode or Power Doppler Mode. Therefore, the same limitations imposed on measurements taken in B/W images also apply to the accuracy of measurements taken in Color Doppler and Power Doppler Modes. Using Color/Power Doppler Mode images to measure accurate blood flow velocity is not recommended. The amount of blood flow is calculated based on the average velocity rather than the peak velocity. In all applications, the amount of blood flow is measured in PW Spectral Doppler mode.
Cursor Position
All measurements are affected by input data. Adjust the images on the screen so that they are displayed at maximum granularity. ▶ Use the front edge or boundary point of a probe to make the start and end points of a measurement object more distinct. Make sure that the probe direction remains aligned during measurement.
Measurement Accuracy Table
The following tables show the accuracy of the measurements that can be taken using this product. Ensure that the results of measurement accuracy checks are kept within the ranges specified in the table. Except for certain applications or probes, the following accuracy ranges should be maintained for measurement of distance on a straight line.

NOTE: In case of doubt in the measurements, an accuracy test can be requested via your local Samsung Medison Service Department.
2D Mode
| Measurements | Accuracy(Whichever is greater) | TestMethodology | AccuracyBased on | Range |
| Axial Distance < ± 4% or 1mm Phantom Acquisition Full Screen | ||||
| Lateral Distance < ± 4% or 2mm Phantom Acquisition Full Screen | ||||
| Area < ± 8% or 4 mm 2 | Calculation Equation Full Screen | |||
| Volume < ± 12% or 8 mm 3 | Calculation Equation Full Screen | |||
| Angle < ± 4% or 1° Calculation Equation Full Screen | ||||
M Mode
| Measurements | Accuracy(Whichever is greater) | TestMethodology | AccuracyBased on | Range |
| Time | < ± 5% or 1 ms | Signal Generator | Acquisition | 0.01 – 11.3 sec |
| Slope | < ± 8% or 4 mm/ms | Calculation Equation Full Screen | ||
PW/CW Spectral Doppler Mode
| Doppler Measurements | Accuracy(Whichever is greater) | Test Methodology | Accuracy Based on | Range |
| Velocity < ± 1 | 5% Phantom Acquisition 0.1 cm/s – 8.8 m/s | |||
| Time < ± 5% | Signal Generator Acquisition 10 ms – 9.44 s | |||
| Heart Rate < ± 5% or 1 bpm Calculation Equation | 1 – 43,200 bpm | |||
Basic Measurements
Press the Caliper button on the control panel.

NOTE: Take basic measurements of distance and area regardless of the application. For information on measurements for each application, please refer to the 'Measurements by Application' section of this chapter.
The available measurement methods vary depending on the current diagnosis mode. Please refer to the following table:
| Measurement Category Diagnosis Modes Measurement Method | ||
| Distance measurement | 2D, M, D | DistanceTrace LengthOpen Spline3 Points Angle2 Lines Angle%SStenosis(D) |
| M | Distance(M)SlopeHR(M)Time(M) | |
| D | VelocityAccelRIManual TraceLimited TraceHR(D)VolumeFlow(D)VolumeFlow(A)Time(D)Auto TraceS/D RatioD/S Ratio | |
| Circumference and Area measurement | 2D, M, D | EllipseTraceClosed Spline%Stenosis(A) |
| Volume measurement 2D, M, D | 1 Dist. Volume2 Dist. Volume3 Dist. VolumeEllipse VolumeEllipse + Dist.VolDisk Volume | |
[Table 8.1 Basic Measurements by Diagnosis Mode]
Basic Measurement Operations
The following is the information on common button operations for basic measurements:
Selecting/Changing Measurement Method
Select a measurement method using the touch screen. The menu items that can be selected on the touch screen vary according to the diagnosis mode. The selected measurement method is displayed in the User Information area.
Adjusting Font Size
Go to Setup > Measurement > General > Result > Font Size and either enter the font size to be used in the measurement results, or use the 'Font Size' knob button to select a desirable size from 8 to 20.
Canceling Measurement Results
Delete Last: Tap Delete Last on the touch screen to delete the last measurement taken. The value displayed on the corresponding application report will also be deleted.
Specifying the Measurement Result Position
Place the pointer over the measurement and press the Set button. Use the trackball to move the measurement to the desired location, and press the Set button to place the movement at that location.
Resetting Measurement Result Position
Go to Setup > Measurement > General Page, and change the value for Position on 2D Mode or Position On M/D in Result to reset the position of the measurement results as desired.

NOTE: You may select Left-Top, Left-Bottom, Right-Bottom, Right-Top, or Custom.
Deleting Measurement Results
Clear: Tap Clear on the control panel or the touch screen to delete the measurement.
Printing Measurement Results
Press the established U Key with the Store/Print button on the control panel.

NOTE: Go to Setup > Customize > Buttons and choose a button to set as Store/Print from the U1 U2 U3 U4 U5 and U6 buttons.
Exiting Basic Measurement
After completing (or canceling) all in-progress measurements, press the Exit button on the control panel.

NOTE: To change measurement units and other settings, select Setup > Measurement > General on the keyboard. For more information, refer to 'Chapter 3. Utilities' in this manual.

[Figure 8.1 Caliper]
Distance Measurement

Distance
This is a basic measurement that is available in all diagnosis modes. Specify two points on a 2D image and measure the length of the straight line between the points.
- Select Distance on the monitor or touch screen. Use the trackball and the Set button on the control panel to specify the end points of the measurement.
Use the trackball to place the cursor at the desired position and then press the Set button.

Repositioning a Point
Instead of pressing the Set button to confirm the point position, you can press the Change button to reset it.
- Specify both end points and then the distance between them will be measured.
- When the distance has been measured, the result is shown on the screen.

Trace Length
This is a basic measurement that is available in all diagnosis modes. Specify a point on a 2D image and trace a curve from that point to measure its length.
- Select Trace Length on the monitor or touch screen. Use the trackball and the Set button on the control panel to specify the start point of the measurement.
Use the trackball to place the cursor at the desired position and then press the Set button.
- Use the trackball to draw the desired curve and then press the Set button to set the end point.
- Specify both end points and then the length of the curve will be measured.

Editing Curves
Before specifying the end point by pressing the Set button, you can delete part of the curve being traced by rotating the soft menu dial-button 5 counter-clockwise. To restore the deleted curve, rotate the dial-button clockwise.
Open Spline
This is a basic measurement that is available in all diagnosis modes. Specify multiple points on a 2D image to measure the length of the resulting Open Spline curve.
- Select Open Spline on the monitor or touch screen. Use the trackball and the Set button on the control panel to specify as many points as desired. After specifying the last point of the Open Spline, press the Set button twice to finish measuring the Open Spline.
At least two points must be specified before the Open Spline can be measured.
- The length of the Open Spline curve created with the points you have specified will be calculated and displayed on screen.
3 Points Angle
This is a basic measurement that is available in all diagnosis modes. Place three points on a 2D image and measure the angle formed by the points.
- Tap 3 Points Angle on the touch screen.
- Specify three points.
- The angle formed by the three points will be calculated and displayed on the screen.
2 Lines Angle
This is a basic measurement that is available in all diagnosis modes. Specify two straight lines on a 2D image to measure the angle between the two lines.
- Tap 2 Lines Angle on the touch screen.
- Draw two straight lines. Refer to 'Distance' for instructions on how to draw a straight line.
- The angle between two lines will be calculated and displayed on the screen.
When two angles are calculated, the smaller angle is displayed.

%Stenosis(D)
StD stands for Stenosis Distance, which is a basic measurement available in all diagnosis modes. The diameter of a blood vessel is measured on a 2D image to calculate its stenosis ratio (%).
- Select %Stenosis (D) on the monitor or touch screen.
- Measure the total diameter of a blood vessel using the Distance measurement method.
- When the second cursor appears, measure the diameter of the vessel's inner wall under stenosis.
- Calculate %StD using the following equation:
%Stenosis (D) = (Outer Distance - Inner Distance) / Outer Distance x 100

Distance(M)
This is a basic measurement that is only available in M mode. Specify two points on an M image, and measure the distance between the two points.
- Select Distance (M) on the monitor or touch screen. Specify two points and measure the length of the straight line between the points. This measurement is taken in the same way as 'Distance'.
- When the distance has been measured, the result is shown on the screen.

Slope
This is a basic measurement that is only available in M mode. Specify two points on an M image, and measure the velocity between the two points.
- Select Slope on the monitor or touch screen.
- Specify two points and measure the length of the straight line between the points. This measurement is taken in the same way as 'Distance'.
- When the distance has been measured, the result is shown on the screen.
HR(M)
This is a basic measurement that is only available in M mode. Specify two bars on an image and measure the heart rate between the two bars.
- Select HR(M) on the monitor or touch screen.
- Use the trackball and the Set button on the control panel to select two bars.
- When the measurement has been taken, the result is shown on the screen.
Time(M)
This is a basic measurement that is only available in M mode. Specify two bars on an M image and measure the time between the two bars.
- Select Time(M) on the monitor or touch screen.
- Use the trackball and the Set button on the control panel to select two bars.
- When the measurement has been taken, the result is shown on the screen.
Velocity
This is a basic measurement that is only available in Spectral Doppler mode. Specify one point on a Spectral Doppler image to measure the velocity.

NOTE: In a Spectral Doppler image, the X- and Y-axes represent time and velocity, respectively.
- Select Velocity on the monitor or touch screen.
- Specify a point.
- When the measurement has been taken, the result is shown on the screen.

Accel
This is a basic measurement that is only available in Spectral Doppler mode. Specify two points on a Spectral Doppler image and measure the velocity at each point to calculate the time and acceleration.
- Select Accel on the monitor or touch screen.
- Specify two points and measure the length of the straight line between the points. This measurement is taken in the same way as 'Distance'.
- When the measurement has been taken, the result is shown on the screen.
The following equation is used for Accel measurement:
Accel = (V _ 2 - V _ 1)(T _ 2 - T _ 1) = dV/dT

RI
This is a basic measurement that is only available in Spectral Doppler mode. Specify two points on a Spectral Doppler image and measure the velocity at each point to calculate the Resistivity Index (RI).
- Tap RI on the touch screen.
- Specify two points on a Spectral Doppler image, and measure the velocity at each point.
- When the measurement has been taken, the result is shown on the screen.
The equations used for D RI measurement are as follows:
RI = V _ 1 - V _ 2V _ 1

Manual Trace
This is a basic measurement that is only available in Spectral Doppler mode. You can specify a point on a Spectral Doppler image and trace a curve from that point to calculate the blood flow velocity, mean velocity, RI and Pulsatility Index (PI).
- Select Manual Trace on the monitor or touch screen.
- Trace the curve. When the measurement has been taken, the result is shown on the screen. The measurement results displayed will vary, depending on the settings you have selected in Setup > Measurement > Library > Caliper > D > Manual Trace.
PS: Peak Systolic Velocity
PI: Pulsatility Index
MD: Min Diastolic Velocity
RI: Resistivity Index
ED: End Diastolic Velocity
PS/ED: PS ED Ratio
TAmax: Time Averaged Max Velocity
ED/PS: ED PS Ratio
The equations used for Trace measurement are as follows:
V mean = VTI / Duration of flow, RI = (PS - ED) / PS, PI = (PS - ED) / Vmean

Limited Trace
This is a basic measurement that is only available in Spectral Doppler mode. Specify two bars on a Spectral Doppler image to calculate the mean velocity between the bars, RI and Pulsatility Index (PI).
- Select Limited Trace on the monitor or touch screen. Use the trackball and the Set button on the control panel to select two bars.
- When the measurement has been taken, the result is shown on the screen.
PS: Peak Systolic Velocity
PI: Pulsatility Index
MD: Min Diastolic Velocity
RI: Resistivity Index
ED: End Diastolic Velocity
PS/ED: PS ED Ratio
TAmax: Time Averaged Max Velocity
ED/PS: ED PS Ratio
TAMEAN: Time Average Mean Velocity
HR
The equations used for D Limited Trace measurement are as follows:
meanV = VTI / Duration of flow; RI = (PS - ED) / PS; PI = (PS - ED) / Vmean; V1/V2 = V1 / V2

VolumeFlow(D)
This is a basic measurement that is only available in Spectral Doppler mode and 2D Image Mode. Specify two bars on a Spectral Doppler image, and specify two points in 2D Image mode, to calculate the VolumeFlow.
- Select VolumeFlow(D) on the monitor or touch screen.
- Use the trackball and the Set button on the control panel to select two bars. This measurement is taken in the same way as 'M Time'. This can only be done in the Loop area.
- Use the trackball and the Set button on the control panel to select two points. 2D Image Mode will activate automatically so that you can measure the distance. This measurement is taken in the same way as 'Distance'.
- When the measurement has been taken, the result is shown on the screen.

Time(D)
This is a basic measurement that is only available in Spectral Doppler mode. Place two bars on a Spectral Doppler image to calculate the time between the bars.
- Tap D Time on the touch screen.
- Set two bars on the spectrum with the trackball and the Set button.
- When the measurement has been taken, the result is shown on the screen.

HR(D)
This is a basic measurement that is only available in Spectral Doppler mode. Place two bars on an image and measure the heart rate between the two bars.
- Select HR(D) on the monitor or touch screen.
- Use the trackball and the Set button on the control panel to select two bars.
- When the measurement has been taken, the result is shown on the screen.

Auto Trace
Spectrum is traced automatically. This can be enabled in the measurement menu in Spectral Doppler mode.
- Press Auto Trace in the Measurement menu.
- The system traces the spectrum automatically.
- When the Trace is complete, the measurement results are displayed on the screen.

Things to Consider when using Doppler Spectrum Auto Trace
The state of a Doppler Spectrum may affect the measurement results. Please pay attention to the following:
Causes for Trace Failure
- If the Gain is changed while a Doppler image is in Freeze state, Contour Trace and Peak Trace will not work.
- If there is little or no noise in an image without a spectrum, Contour Trace will not work.
- If there is severe noise in an image, Contour Trace will not work.
- If the Clutter Filter is set too high, AutoTrace and Limited Trace may not work.
Possible Causes of Inaccurate Peak Trace
- If the PRF (Pulse Repetition Frequency) is lower than the velocity at the site being examined, aliasing may occur. If the original signals are separated from aliasing, a Trace can be carried out, but the peak measurement may not be accurate.
- If the peak of a spectral waveform is not clear or occurs intermittently, a Trace can be carried out, but the peak measurement may not be accurate.
- If the Doppler Gain is set too high or too low, it becomes difficult to distinguish spectra. This may result in measurement error.
- If the Wall Filter is set too high, the spectrum will only be partially displayed. In this case, a Trace can be carried out, but Peak measurement may not be accurate.
- If abnormal noise or artifacts occur, a Trace can be carried out, but Peak measurement may not be accurate.
Other
– Use of the CW Probe may result in measurement error(s). - Limited Trace is supported only for two-peak spectrums, such as Mitral Valve Inflow and Tricuspid Valve Inflow in the cardiology application.

S/D Ratio, D/S Ratio
This is a basic measurement that is only available in Spectral Doppler mode. After designating two points in a Spectral Doppler image, calculate the velocity ratio by measuring the velocity at each point.

NOTE: In a Spectral Doppler image, the X- and Y-axes represent time and velocity, respectively.
- Tap the S/D button on the touch screen.
- After designating two points, measure the velocity at each point. The measurement method is the same as 'RI'.
- When the measurement has been taken, the result is shown on the screen.
Circumference and Area measurement

Ellipse
This is a basic measurement that is available in all diagnosis modes. Measure the circumference and area of circle (ellipse)-shaped objects on a 2D image.
- Click Ellipse on the monitor or tap Ellipse on the touch screen.
- Use the trackball and the Set button on the control panel to specify the diameter (axis) of the measurement area.
Use the trackball to place the cursor at the desired position and then press the Set button.

Repositioning Point
Before pressing the Set button to confirm the position, you can press the Change button to reset the position of the point.
- Specify the size of the circle (ellipse).
Adjust the size using the trackball and then press the Set button.
- When the measurement has been taken, the result is shown on the screen.
The following equation is used for Ellipse measurement:
Circ = π (a + b) · (1 + 3 (a - b/a + b) ^ 21 0 + √ 4 - 3 (a - b/a + b) ^ 2), (a: Half Long axis, b: Half Short axis)
Area = π (x, b: Akis)
Trace
This is a basic measurement that is available in all diagnosis modes. Measure the circumference and area of an irregularly shaped object on a 2D image.
- Select Trace on the monitor or touch screen.
- Use the trackball and the Set button on the control panel to specify the start point for tracing over the contour of the measurement area. Use the trackball to place the cursor at the desired position and then press the Set button.
- Trace the curve so that the measurement cursor returns to the start point, and then press the Set button.

NOTE: The Trace Line must be a closed curve. If you press the Set button before tracing is complete, a straight line will be traced between the current position and the start point, resulting in a significant error.
- When the measurement has been taken, the result is shown on the screen.
The equations used for Trace measurement are as follows:
\triangleright Circ = sum\sqrt{\left{X(n) - X(n - 1)\right}^2 + \left{Y(n) - Y(n - 1)\right}^2} , (N = 1, 2... last point) Area = sum[√X(n-1)×Y(n)-X(n)×Y(n-1)], (N=1,2... last point)
Tips!
Editing Curves
Before specifying the end point by pressing the Set button, you can delete part of the curve being traced by rotating the soft menu dial-button 5 clockwise. To restore the deleted curve, rotate it counter-clockwise.

Closed Spline
This is a basic measurement that is available in all diagnosis modes. Specify multiple points on a 2D image to measure the circumference and area of the resulting closed spline.
- Select Closed Spline on the monitor or touch screen. Use the trackball and the Set button on the control panel to specify as many points as desired. After specifying the last point of the Spline, press the Set button twice to finish measuring the Spline.
At least three points must be specified before spline can be measured.
- The circumference and area of the closed spline created with the points you have specified will be calculated and displayed on screen.

%Stenosis(A)
StA stands for Stenosis Area, which is a basic measurement available in all diagnosis modes. The area of a blood vessel is measured on a 2D image to calculate its stenosis ratio (%).
- Select %Stenosis(A) on the monitor or touch screen. Measure the area of the vessel outer wall using the 'Ellipse' measurement method.
- When the second cursor appears, measure the area of the stenosed vessel's inner wall. This measurement is taken in the same way as 'Trace'.
- Calculate %Stenosis (A) with the following equation:
%Stenosis (A) = (Outer Area - Inner Area) / Outer Area x 100

VolumeFlow(A)
This is a basic measurement that is only available in Spectral Doppler mode and 2D Image mode. Specify two bars on a Spectral Doppler image and specify an ellipse in 2D Image mode to calculate the VolumeFlow.
- Select VolumeFlow(A) on the monitor or touch screen. Use the trackball and the Set button on the control panel to select two bars. This measurement is taken in the same way as 'M Time'. This can only be done in the Loop area.
- Use the trackball and the Set button on the control panel to select an ellipse. 2D Image Mode will activate automatically so that you can measure the ellipse. This measurement is taken in the same way as 'Ellipse'.
- When the measurement has been taken, the result is shown on the screen.
Volume Measurement

NOTE: Since Dual Mode simultaneously displays two images on the screen, you don't have to return to the diagnosis mode to measure volume in Dual Mode.

1 Distance Volume
This is a basic measurement that is available in all diagnosis modes. Measure the volume of an object by using only one straight line on a 2D image.
- Select 1 Distance Volume on the monitor or touch screen. Specify two points and measure the length of the straight line between the points. This measurement is taken in the same way as 'Distance'.
- When the measurement has been taken, the result is shown on the screen. The volume (Vol.) will be calculated along with the length of the line.
The equations used for 1 Distance measurement are as follows:
Vol = 4/3 π · (D/2)3, (D: distance)

2 Distance Volume
This is a basic measurement that is available in all diagnosis modes. Measure the volume of an object by using two straight lines on a 2D image.
- Select 2 Distance Volume on the monitor or touch screen. Specify two points and measure the length of the straight line between the points. This measurement is taken in the same way as 'Distance'.
- Measure the length of the last remaining straight line using the same method as above.
- When the measurement has been taken, the result is shown on the screen. The volume (Vol.) will be calculated along with the length of each line.
The equations used for 2 Distance measurement are as follows:
D1 gt; D2, π/6 · D12 · D2, (D: distance)

3 Distance Volume
This is a basic measurement that is available in all diagnosis modes. Measure the volume of an object by using three straight lines on a 2D image.
- Select 3 Distance Volume on the monitor or touch screen. Specify two points and measure the length of the straight line between the points. This measurement is taken in the same way as 'Distance'.
- Measure the length of the remaining two straight lines using the same method as above.
- When the measurement has been taken, the result is shown on the screen. The volume (Vol.) will be calculated along with the length of each line.
The equations used for 3 Distance measurement are as follows:
Vol = 4/3 π · D1/2 · D2/2 · D3/2, (D: distance)

Ellipse Volume
This is a basic measurement that is available in all diagnosis modes. Measure the volume of a cone-shaped object by using one circle (ellipse) on a 2D image.
- Tap Ellipse Volume on the touch screen. Specify the size of the circle (ellipse). This measurement is taken in the same way as 'Ellipse'.
- When the measurement has been taken, the result is shown on the screen.
The following equation is used for Ellipse measurement:
Vol. = 4/3 π · Long /2 · Short /2 · Short /2

Ellipse+Dist Vol.
This is a basic measurement that is available in all diagnosis modes. Measure the volume of an object by using one straight line and one circle (ellipse) on a 2D image.
- Tap Ellipse+Dist Vol. on the touch screen. Specify the size of the circle (ellipse). This measurement is taken in the same way as 'Ellipse'.
- Specify two points and measure the length of the straight line between the points. This measurement is taken in the same way as 'Distance'.
- When the measurement has been taken, the result is shown on the screen.
The equation for measuring Ellipse+Dist Vol. is as follows:
Vol = π/6 × a × b × d, (a: Short axis , b: Long axis , d: Distance )

Disk Volume
This is a basic measurement that is available in all diagnosis modes. Calculate the volume of an irregularly shaped object by measuring the area and the length of its long axis on a 2D image.
- Select Disk Volume on the monitor or touch screen. Draw the contour of the area to be measured. This measurement is taken in the same way as 'Trace'.
- Measure the length of the long axis. This measurement is taken in the same way as 'Distance'.
- When the measurement has been taken, the result is shown on the screen.
D: The length of a straight line Area: The area of a circle Volume: Volume
Tips!
Editing Curves
Before specifying the end point by pressing the Set button, you can delete part of the curve being traced by rotating the soft menu dial-button 5 counter-clockwise. To restore the deleted curve, rotate it clockwise.
Calculations by Application
Press the Calculator button on the control panel.
Things to Note
Before Starting Measurement
Register Patient
Make sure the currently registered patient information is correct. If the patient is not registered, press the Patient button on the control panel to register the patient. Use the Study Information tab to enter or change a patient's information per diagnosis item. For information on the Patient Information menu and how to enter the information, refer to 'Patient Information' in 'Chapter 6. Starting Diagnosis'.
Check the Probe, Application, and Preset
Check the name of the probe and the application displayed in Title area. To use a different probe or application, press the Probe button on the control panel. In the Probe Selection screen, check the Preset settings.
Measurement Menu Settings
Set the related menus for convenient measurement. For menus related to measurement and their settings, refer to the 'Measurement Settings' in 'Chapter 3. Utilities' of this manual.
Measurement Operations
The following gives information on the common button operations for measurements:
Select Measurement Item
Select from the measurement item menu on the touch screen.
Change Measurement Method
If the current measurement may be taken in more ways than one, the measurement method may be changed. While taking a measurement, you may rotate the dial-button below the Change Tool menu on the touch screen to change the measurement method.
Delete Trace Line
While taking Trace Length or Trace measurements, rotating the touch screen Undo/Redo dial-button clockwise will delete the traced curve. To restore the deleted curve, rotate it counter-clockwise. Move the trackball in the opposite direction to delete the traced line. You can only delete the line while you are manually tracing the Doppler Spectrum.
Deleting Measurement Results
Press the Clear button on the control panel.

NOTE: The measurement results are deleted from the screen, but still shown on the report for the corresponding application.
Printing Measurement Results
Press the established button with the Store/Print button on the control panel.
Exiting Measurement
After completing (or canceling) all in-process measurements, press the Exit button on the control panel.
End Diagnosis
Press the End Exam button on the control panel. The study for the currently registered patient will end, and all measurement results will be saved.
Fetus
Either click the Fetus button displayed at the top of the left menu, or rotate the dial-button 1 to select the fetus to measure among A, B, C, and D.
Direction
Either click the Direction button displayed at the top of the left-hand menu or rotate the dial-button 2 to select the direction of the object to measure between left and right. This is only displayed in specific packages.
Location
Either click the Location button displayed at the top of the left-hand menu or rotate the dial-button 3 to select the location of the object to measure from Prox, Mid, and Distal.
Trace Direction
Tap Trace Direction on the touch screen to set the trace direction of the Doppler Spectrum. This is activated only after Limited Trace has been performed in Spectral Doppler mode.
▶ Above: Only the + part of the Doppler waveform is traced. Below: Only the – part of the Doppler waveform is traced. Both: All parts of the Doppler waveform are traced.
Measured Frame
Save the measurement frame. Rotate dial-button 1 to go to the measurement frame.
Threshold
Adjust elimination of weak signals when measuring trace in Spectral Doppler mode.
HR Cycle
Settings may be configured for each application at Setup > Measurement > Applications. In Heart Rate, the number of cycles for heart rate (HR) may be specified at Heart Rate Cycle.
Specifying Measurement Result Position
Place the pointer over the measurement and press the Set button. Use the trackball to move the measurement to the desired location, and press the Set button to place the movement at that location.
Deleting Recent Measurement Item
Tap Delete Latest on the touch screen to cancel the last measurement taken.
Report
Press the Report button on the control panel to switch to the Report screen.
Common Measurement Methods
This section provides information on the common measurement methods used for applications.

Measurements in Spectral Doppler mode
In general, if you trace a Doppler Spectrum, you obtain results for various measurement items automatically. There are 3 ways to trace a Doppler Spectrum.
Also, HS50/HS60 can select a specific item under the measurement menu and take measurements individually without tracing a Doppler Spectrum.
Auto Trace
Spectrum is traced automatically. This can be enabled in the measurement menu in Spectral Doppler mode.
- Press Auto Trace in the measurement menu.
- The system traces the spectrum automatically.
- When the Trace is complete, the measurement results are displayed on the screen.
Limited Trace
Specify a measurement range, and a spectrum will be traced automatically. This can be enabled in the measurement menu in Spectral Doppler mode.
- Select Limited Trace in the measurement menu. A bar will appear, with which you can specify range.
- Specify the measurement range. Place the bar at a desired position with the trackball, and press the Set button.
- The system traces spectrums within the specified range automatically.
- When the Trace is complete, the measurement results are displayed on the screen.
Manual Trace
A spectrum is traced manually. This can be enabled in the measurement menu in Spectral Doppler Mode.
- Select Manual Trace in the measurement menu. The measurement cursor will appear over the spectrum.
- Trace the spectrum.
- When the Trace is complete, the measurement results are displayed on the screen.
Measurement by Item
In the Measurement menu, select an individual item and take a measurement.
- Press the Calculator button on the control panel after obtaining a desired image.
- Select the item you want from the Measurement menu. The + cursor will appear over the spectral waveform.
- Position the "+" cursor and press the Set button.
- The measurement results for the selected item are displayed on the screen.
| Item Category Unit Formula | |||
| PSV (Peak Systolic Velocity) Velocity cm/s | |||
| ED (End Diastolic Velocity) Velocity cm/s | |||
| MD (Minimum Diastole Velocity) Velocity cm/s | |||
| PI (Pulsatility Index) Calculation Ratio (PS-ED)/TAMAX | |||
| RI (Resistivity Index) Calculation Ratio | (PS-ED)/PS or (PS-MD)/PS | ||
| PS/ED (Ratio of PS to ED) | Calculation Ratio | PS/ED | |
| ED/PS (Ratio of ED to PS) | Calculation Ratio | ED/PS | |
| TAMAX (Time Average Max Velocity) | Velocity cm/s | ||
| TAMEAN (Time Average Mean Velocity) | Velocity cm/s |

Taking Measurements via Auto Calc.
You can use Auto Calc to take measurements on predetermined item(s). Measurement items are as follows. For information on setting measurement item(s), please refer to 'Measurement' in 'Chapter 3. Utilities'.
▶ Peak Systolic Velocity (PS) ▶ End Diastolic Velocity (ED) Min Diastole Velocity (MD) ▶ Time Averaged Peak Velocity (TAMAX) ▶ Time Averaged Mean Velocity (TAMEAN) ▶ Resistive Index (RI) Pulsatility Index (PI) Systole/Diastole Ratio (PS/ED) Diastole/Systole Ratio (ED/PS) ▶ Heart Rate (HR) Max Pressure Gradient (PGmax) ▶ Mean Pressure Gradient (PGmean) ▶ Velocity Time Integral (VTI) ▶ Peak A (Peak A)
Volume Flow Measurement
Select Volume Flow in the measurement menu.
Volume Flow can be calculated by measuring an area or distance. For information on measuring distance or area, refer to 'Basic Measurement'. TAMEAN (Time Avg. Mean Velocity) value is measured automatically.
Vesl. Area (Vessel Area)
Measure the area of a blood vessel to calculate TAMEAN and Volume Flow.
VolumeFlow(A) = Area × TAMEAN × 60
Vessel Distance
Measure the width of a blood vessel to calculate TAMEAN and Volume Flow.
VolumeFlow(D) = (π × d²) / 4 × TAMEAN × 60

Stenosis Measurement
You can measure the stenosis of each blood vessel system by measuring and calculating an area or distance.
%Stenosis(A)
Measures the area of the inner wall and the outer wall of the blood vessel. %Stenosis(A) stands for Stenosis Area.
- Select the %Stenosis(A) menu and the first cursor will appear in 2D Mode.
- Measure the area of the vessel's outer wall using the Circ/Area measurement method.
- When the second cursor appears, measure the area of the stenosed vessel's inner wall.
%Stenosis (A) = (Outer Area - Inner Area) / Outer Area x 100
%Stenosis(D)
Measure the diameter of the blood vessel. Stenosis(D) stands for Stenosis Distance.
- Select the %Stenosis(D) menu and the first cursor will appear in 2D Mode.
- Measure the total diameter of a vessel using the Distance measurement method.
- When the second cursor appears, measure the diameter of the vessel's inner wall under stenosis.
%Stenosis (D) = (Outer Distance - Inner Distance) / Outer Distance x 100

Heart Rate Measurement
HR (Heart Rate)
You can calculate heart rates for a certain period of time.
- Select HR from the Measurement menu. A bar will appear, with which you can specify range.
- Specify the measurement range. Place the bar at a desired position with the trackball, and press the Set button.
- The system will automatically measure the heart rate within the measurement range. The measurement result is displayed on screen.
OB Calculations

NOTE:
Ductus Venosus and Fetal HR can only be measured in Doppler Mode. For information on basic measurement methods, refer to 'Basic Measurements' and 'Common Measurement Methods' in this chapter. For references on measurement items, see 'Reference Manual Part 1'.
Before Taking OB Measurements
OB Basic Information
Enter the information required for OB diagnosis in the Patient Information screen. The basic OB information includes LMP (Last Menstrual Period) and Number of Fetuses.
Once LMP is entered, EDD (Estimated Delivery Date) and GA (Gestational Age) are calculated automatically. LMP is required for the calculation of values such as EDD and SD in obstetrics measurement.
▶ EDD (LMP) = LMP + 280 days ▶ GA (LMP) = Current System Date – LMP LMP = Ovul. Date -14
The physician may choose to override the LMP and manually enter the EDD into 'EDD (LMP)'.
Once the expected delivery date is entered, GA (Gestational Age) is automatically calculated and displayed. EDD (LMP) is renamed to EDD, while GA (LMP) is renamed to GA (EDD).
Once DOC is entered, EDD (Estimated Delivery Date) and GA (Gestational Age) are calculated automatically. EDD (LMP) is renamed to EDD (DOC), while GA (LMP) is renamed to GA (DOC).
Once GA is entered, EDD (Estimated Delivery Date) is entered automatically. EDD (LMP) is renamed to EDD (GA), while GA (LMP) is renamed to GA (Clin).
A maximum of four fetuses can be entered as Number of Fetuses. The default value is '1'. In the case of twins, enter '2'.
For further information about patient information menus and how to input patient information, refer to 'Patient Information' of 'Chapter 6. Starting Diagnosis'.
OB Measurement Menu Settings
Set up the GA Equation, GA Table, and OB measurement menus that are used in obstetrics measurements. The user can manually write, back up or restore GA Tables. For more information on the GA Equation and Table, refer to the Reference Manual Part 1.

NOTE: For twins, distinguish fetuses by specifying them as Fetus A and Fetus B in the Measurement menu. Either click the Fetus button displayed at the top of the left-hand menu or rotate the dial-button 1 to select the fetus to measure among A, B, C, and D.
For menus related to measurement and their settings, refer to the 'Measurement Settings' in 'Chapter 3. Utilities' of this manual.
1 Trim. Measurement Menu
When the measurements for the selected items are complete, the measurements and gestational age are displayed on the screen. The measurement method for each item is the same as for basic measurement. Measured items are automatically entered into the report.
| Measurement Menu Item Mode Method Unit | ||||
| Fetal Biometry | GS All Distance measurement cm | |||
| YS All Distance measurement cm | ||||
| CRL All Distance measurement cm | ||||
| BPD All Distance measurement cm | ||||
| FL All Distance measurement cm | ||||
| AC | All | Circumference measurement | cm | |
| Fetal Cranuim | NT All Distance measurement mm | |||
| IT | All Distance measurement mm | |||
| NB | All Distance measurement cm | |||
| Ductus Venosus | S Vmax PW Velocity measurement cm/s | |||
| D Vmax PW Velocity measurement cm/s | ||||
| A Vmax PW Velocity measurement cm/s | ||||
| Auto Trace PW Doppler Spectrum trace | ||||
| Limited Trace PW Doppler Spectrum trace | ||||
| Manual Trace PW Doppler Spectrum trace | ||||
| PS PW Velocity measurement cm/s | ||||
| ED PW Velocity measurement cm/s | ||||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement – | |||
| RI | PW Velocity measurement – | |||
| S/D | PW | Continuous measurement – | ||
| Fetal HR | PW | Heart Rate | bpm | |

Automatic Calculation
Some items in the measurement menu are automatically calculated based on measurements of other items.

NOTE: For reference, the Osaka University/Tokyo University methods are mainly used in Asia, the Merz method in Europe, and the Shepard/Hadlock methods on the American continent.

2-3 Trim. Measurement Menu
When the measurements for the selected items are complete, the measurements and gestational age are displayed on the screen. The measurement method for each item is the same as for basic measurement. Measured items are automatically entered into the report.
| Measurement Menu Item Mode Method Unit | ||||
| Fetal Biometry | GS All Distance measurement cm | |||
| CRL All Distance measurement cm | ||||
| YS All Distance measurement cm | ||||
| BPD, OFD All Continuous measurement | ||||
| BPD All Distance measurement cm | ||||
| OFD All Distance measurement cm | ||||
| HC | All | Circumference measurement | cm | |
| APAD, TAD | All Continuous measurement | |||
| APAD | All Distance measurement cm | |||
| TAD | All | Distance measurement | cm | |
| AC | All | Circumference measurement | cm | |
| FTA | All | Area measurement | cm2 | |
| FL All Distance measurement cm | ||||
| SL All Distance measurement cm | ||||
| APTD, TTD | All Continuous measurement | |||
| APTD | All | Distance measurement | cm | |
| TTD All Distance measurement cm | ||||
| ThC | All | Circumference measurement | cm | |
| EFW All | - - | |||
| Fetal Long Bones | HL All Distance measurement cm | |||
| ULNA All Distance measurement cm | ||||
| TIB All Distance measurement cm | ||||
| RAD All Distance measurement cm | ||||
| FIB All Distance measurement cm | ||||
| CLAV All Distance measurement cm | ||||
| LV All Distance measurement cm | ||||
| Fetal Cranium | NT All Distance measurement mm | |||
| IT All Distance measurement mm | ||||
| NB | All Distance measurement cm | |||
| NF All Distance measurement cm | ||||
| CEREB | All Distance measurement cm | |||
| CM | All Distance measurement cm | |||
| Lat Vent | All Distance measurement cm | |||
| OOD | All Distance measurement cm | |||
| IOD | All Distance measurement cm | |||
| HW | All Distance measurement cm | |||
| FMF angle | All | Angle calculation | degree | |
| Lt. Fetal Others Rt. Fetal Others | Foot | All Distance measurement cm | ||
| Ear All Distance measurement cm | ||||
| MP | All Distance measurement cm | |||
| Renal L | All Distance measurement cm | |||
| Renal AP | All Distance measurement cm | |||
| Pelvis All Distance measurement cm | ||||
| AFI | Q1 | All Distance measurement cm | ||
| Q2 | All Distance measurement cm | |||
| Q3 | All Distance measurement cm | |||
| Q4 | All Distance measurement cm | |||
| MVP | All Distance measurement cm | |||
| CTAR | ThD ap All Distance measurement cm | |||
| ThD trans All Distance measurement cm | ||||
| HrtD ap All Distance measurement cm | ||||
| HrtD trans All Distance measurement cm | ||||
| ThA All Area measurement cm | 2 | |||
| HrtA | All Area measurement cm | 2 | ||
| Maternal Others | Cervix L All Distance measurement cm | |||
| Placenta Thick | All | Distance measurement | cm | |
| PLI | S Vmax PW | Velocity measurement cm/s | ||
| D Vmax PW | Velocity measurement cm/s | |||
| A Vmax PW Velocity measurement cm/s | ||||
| Umbilical A Lt. Mid Cereb A Rt. Mid Cereb A Lt. Uterine A Rt. Uterine A Placenta A Lt. Fetal Carotids Rt. Fetal Carotids Fetal Aorta Lt. Renal A Rt. Renal A | %StD | All | Calculation after distance measurement | % |
| %StA | All | Calculation after area measurement | % | |
| Auto Trace | PW | Doppler spectrum trace | ||
| Limited Trace | PW | Doppler spectrum trace | ||
| Manual Trace | PW | Doppler spectrum trace | ||
| PS | PW | Velocity measurement | cm/s | |
| ED | PW | Velocity measurement | cm/s | |
| MD | PW | Velocity measurement | cm/s | |
| PI | PW | Velocity measurement | - | |
| RI | PW | Velocity measurement | - | |
| S/D | PW | Continuous measurement | ||
| VF(D) | PW | Calculate after measuring TAmean, Vesl., and Dist. | ml/min | |
| VF(A) | PW | Calculate after measuring TAmean, Vesl., and Area | ml/min | |
| Fetal HR | Fetal HR | M, PW | Heart Rate | bpm |
| Ductus Venosus | S Vmax PW Velocity measurement cm/s | |||
| D Vmax PW Velocity measurement cm/s | ||||
| A Vmax PW Velocity measurement cm/s | ||||
| Auto Trace PW Doppler spectrum trace | ||||
| Limited Trace PW Doppler spectrum trace | ||||
| Manual Trace PW Doppler spectrum trace | ||||
| PS PW Velocity measurement cm/s | ||||
| ED PW Velocity measurement cm/s | ||||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement – | |||
| RI | PW Velocity measurement – | |||
| S/D | PW | Continuous measurement – | ||

AFI (Amniotic Fluid Index)
Measure the amniotic fluid index. Measurements are performed by dividing the pregnant woman's abdomen into four parts. The distance between the fetus and the farthest point of each area is measured. To obtain a specific image from each quadrant plane, press the Freeze button to go to the diagnosis mode. After obtaining the image, press the Freeze button again to return to the measurement mode.

Calculating Estimated Fetal Weight (EFW)
When measurements for the following items are complete, the system uses the results to calculate the estimated fetal weight automatically. For an equation for calculating fetal weight, please refer to 'Estimated Fetal Weight Formula' in the Reference Manual Part 1.
BPD and AC
BPD, FL, and FTA
BPD, APTD, TTD, and FL
BPD, APTD, TTD, and SL
BPD and TTD
AC and FL
BPD, AC, and FL
HC, AC, and FL
BPD, HC, AC, and FL
AC
Continuous Measurement/Review for EFW Calculation
You can measure a series of OB items to calculate the EFW.
Tips!
Before starting measurement
- Make sure that the User Key has been set for EFW continuous measurements. You can configure the User Key for EFW continuous measurements at Setup > Customize > Buttons > User Key > EFW Measure. For more information, refer to 'Peripheral Device Settings' in 'Chapter 3. Utilities'.
- Check EFW Reference. If it is not set, or if you want to change it, select it at Setup > Measurement > Author > Estimated Fetal Weight.

NOTE: This function is not available in 3D mode.
Measurement method
- Press User Key 1 (or User Key 2, 3, 4, 5, 6) on the control panel. The OB Measurement menu will be displayed on screen.
- Measure the items for EFW calculation by using the trackball and the Set button.
- Use the trackball and the Set button on the control panel again to start the measurement of the following items.
- Once all measurements have been taken, the EFW is displayed on screen.
The EFW measurements taken and the order in which they are measured are as follows:
| Reference Measure Item (by Order) | |
| Campbell AC | |
| Hadlock | AC→BPD |
| Hadlock1 | AC→FL |
| Hadlock2 | BPD→AC→FL |
| Hadlock3 | HC→AC→FL |
| Hadlock4 | BPD, HC→AC→FL |
| Hansmann | BPD→TTD |
| Merz | BPD→AC |
| Osaka | BPD→FL→FTA |
| Persson | BPD→FL→APAD→TAD |
| Schild | HC→AC→FL |
| Shepard | AC→BPD |
| Shinozuka1 | BPD→FL→AC |
| Shinozuka2 | BP→APTD, TTD→SL |
| Shinozuka3 | BPD→APTD, TTD→FL |
| Tokyo | BPD→APTD, TTD→FL |
Review the Result of EFW Calculation
- Press User Key 1 (or User Key 2, 3, 4, 5, 6) on the control panel. The measured items taken for calculating EFW will be displayed on screen together with the result.
- To remove the results from the screen, press the Clear button on the control panel.
Fetal Heart Calculations
The measurement method for each item is the same as for basic measurement. In addition, measurement items are similar to those of cardiology measurements (Cardiac Calculations). Measured items are automatically entered into the report.

NOTE:
For information on basic measurement methods, refer to 'Basic Measurements' and 'Common Measurement Methods' in this chapter. For references on measurement items, see 'Reference Manual Part 1'.
| Measurement Menu Item Mode Method Unit | ||||
| LV Vol. (Simpson) | LVEDV A2C All Dist | 20 ml | ||
| LVESV A2C All Dist | 20 ml | |||
| LVEDV A4C All Dist | 20 ml | |||
| LVESV A4C All Dist | 20 ml | |||
| 2D Echo | Asc Ao All Distance measurement cm | |||
| MPA Diam | All Distance measurement cm | |||
| Duct Art All Distance measurement cm | ||||
| LA Diam All Distance measurement cm | ||||
| RA Diam | All Distance measurement cm | |||
| RVDd | All Distance measurement cm | |||
| IVS | All Distance measurement cm | |||
| LVIDd All Distance measurement cm | ||||
| LVIDs | All Distance measurement cm | |||
| LVPW | All Distance measurement cm | |||
| HrtC | All | Circumference measurement | cm | |
| ThC | All | Circumference measurement | cm | |
| CTAR | ThD ap All Distance measurement cm | |||
| ThD trans All Distance measurement cm | ||||
| HrtD ap All Distance measurement cm | ||||
| HrtD trans All Distance measurement cm | ||||
| ThA | All | Area measurement | cm2 | |
| HrtA | All | Area measurement | cm2 | |
| Fetal HR | Fetal HR | M, PW | Heart Rate | bpm |
| M Echo | IVSd | M | Distance measurement cm | |
| LVIDd | M | Distance measurement cm | ||
| LVIDs | M | Distance measurement cm | ||
| LVPWd | M | Distance measurement | cm | |
| IVSs | M | Distance measurement cm | ||
| LVPWs M | Distance measurement cm | |||
| RVDd | M | Distance measurement cm | ||
| MPA Duct Arteriosus IVC Asc Aorta Dsc Aorta | Auto Trace | PW | Doppler spectrum trace | |
| Limited Trace | PW | Doppler spectrum trace | ||
| Manual Trace | PW | Doppler spectrum trace | ||
| PS | PW Velocity measurement cm/s | |||
| ED | PW Velocity measurement cm/s | |||
| MD PW Velocity measurement cm/s | ||||
| PI | PW Velocity measurement - | |||
| RI | PW Velocity measurement - | |||
| S/D PW | Continuous measurement | |||
| Duct Venosus | SVmax PW | Velocity measurement cm/s | ||
| DVmax PW | Velocity measurement cm/s | |||
| AVmax PW Velocity measurement cm/s | ||||
| MV TV | E PW Velocity measurement cm/s | |||
| A PW Velocity measurement cm/s | ||||
| Vmax PW Velocity measurement cm/s | ||||
| PLI | S Vmax PW Velocity measurement cm/s | |||
| D Vmax PW Velocity measurement cm/s | ||||
| A Vmax PW Velocity measurement cm/s | ||||
| TEI | TST PW Time measurement ms | |||
| ET PW Time measurement ms | ||||
CTAR (Cardio-Thorax Area Ratio)
This measurement compares the size of the fetus' heart in relation to the size of its thorax. ThD ap, ThD trans, HrtD ap, and HrtD trans values are acquired to obtain the comparative value.
CTAR = HrtDap × HrtDtrans/ThDap × ThDtrans × 1 0 0
Gynecology Calculations

NOTE:
For information on basic measurement methods, refer to 'Basic Measurements' and 'Common Measurement Methods' in this chapter. For references on measurement items, refer to the Reference Manual Part 2.

Before Taking GYN Measurements
Enter the information required for GYN diagnosis in the Patient Information screen. Basic information for gynecology includes Gravida, Para, Aborta, and Ectopic.

Measurement Menu
The measurement method for each item is the same as for basic measurement. Measured items are automatically entered into the report.
| Measurement Menu Item Mode Method Unit | ||||
| Uterus | Length All Distance measurement cm | |||
| Width All Distance measurement cm | ||||
| Height All Distance measurement cm | ||||
| Endo.Thick All Distance measurement cm | ||||
| CervixLt. CystRt. CystLt. OvaryRt. Ovary | Length All Distance measurement cm | |||
| Width All Distance measurement cm | ||||
| Height All Distance measurement cm | ||||
| Measurement Menu Item Mode Method Unit | ||||
| Lt. Follicle Rt. Follicle | Follicle #1 All Volume measurement ml | |||
| Follicle #2 All Volume measurement ml | ||||
| Follicle #3 All Volume measurement ml | ||||
| Follicle #4 All Volume measurement ml | ||||
| Follicle #5 All Volume measurement ml | ||||
| Follicle #6 All Volume measurement ml | ||||
| Follicle #7 All Volume measurement ml | ||||
| Follicle #8 All Volume measurement ml | ||||
| Follicle #9 All Volume measurement ml | ||||
| Follicle #10 All Volume measurement ml | ||||
| Follicle #11 All Volume measurement ml | ||||
| Follicle #12 All Volume measurement ml | ||||
| Lt. Ovarian A Rt. Ovarian A Lt. Uterine A Rt. Uterine A Pericystic Endometrial | %StD All | Calculation after distance measurement | % | |
| %StA All | Calculation after area measurement | % | ||
| Auto Trace PW Doppler spectrum trace | ||||
| Limited Trace PW Doppler spectrum trace | ||||
| Manual Trace PW Doppler spectrum trace | ||||
| PS PW Velocity measurement cm/s | ||||
| ED | PW Velocity measurement cm/s | |||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement - | |||
| RI | PW Velocity measurement - | |||
| Vol.Flow(D) | PW | Calculate after measuring TAmean, Vesl., and Dist. | ml/min | |
| Vol.Flow(A) | PW | Calculate after measuring TAmean, Vesl., and Area | ml/min | |
| S/D | PW | Continuous measurement | - | |
| Measurement Menu Item Mode Method Unit | ||||
| Mass1Mass2Mass3Endo. PolypLt. Ovarian MassRt. Ovarian MassUterine Fibroid1Uterine Fibroid2Uterine Fibroid3Cervical FibroidEctopic | Length All Distance measurement cm | |||
| Width All Distance measurement cm | ||||
| Height All Distance measurement cm | ||||
| Auto Trace PW Doppler spectrum trace | ||||
| Limited Trace PW Doppler spectrum trace | ||||
| Manual Trace PW Doppler spectrum trace | ||||
| PS | PW Velocity measurement | cm/s | ||
| ED | PW Velocity measurement | cm/s | ||
| MD | PW Velocity measurement | cm/s | ||
| PI | PW Velocity measurement | - | ||
| RI | PW Velocity measurement | - | ||
| S/D | PW | Continuous measurement | - | |
Most of the gynecology measurements are distance measurements and volume measurements based on the distance measurement results. If multiple images, such as long axis images and transverse axis images are needed, press the Freeze button to switch to Scan Mode and obtain images from another perspective.
Cardiac Calculations

NOTE: Cardiac measurement is an optional item.
The measurement method for each item is the same as for basic measurement. Measured items are automatically entered into the report.

NOTE:
For information on basic measurement methods, refer to 'Basic Measurements' and 'Common Measurement Methods' in this chapter. For references on measurement items, refer to the Reference Manual Part 2.
| Measurement Menu Item Mode Method Unit | ||||
| LV (2D) | All LV (2D) All Continuous measurement | |||
| All LVd (2D) All Continuous measurement | ||||
| IVSd All Distance measurement cm | ||||
| LVIDd All Distance measurement cm | ||||
| LVPWd | All Distance measurement cm | |||
| All LVs (2D) | All Continuous measurement | |||
| IVSs | All Distance measurement cm | |||
| LVIDs | All Distance measurement cm | |||
| LVPWs | All Distance measurement cm | |||
| LV(2D)HR | All | Heart Rate | bpm | |
| RV (2D) | RVAWd | All | Distance measurement | cm |
| RVIDd | All Distance measurement cm | |||
| RVAd | All | Area measurement | cm2 | |
| RVAWs | All | Distance measurement | cm | |
| RVIDs | All Distance measurement cm | |||
| RVAs | All | Area measurement | cm2 | |
| RV Major | All Distance measurement cm | |||
| RV Minor | All Distance measurement cm | |||
| LV (M) | All LV/RV (M) M Continuous measurement | |||
| All LV (M) M Continuous measurement | ||||
| LV EF(M) M Continuous measurement | ||||
| IVSd M Distance measurement cm | ||||
| LVIDd M Distance measurement cm | ||||
| LVPWd M Distance measurement cm | ||||
| IVSs | M Distance measurement cm | |||
| LVIDs | M Distance measurement cm | |||
| LVPWs | M Distance measurement cm | |||
| LVET | M | Time measurement | ms | |
| LVPEP | M Time measurement ms | |||
| LV(M) HR M | Heart Rate bpm | |||
| Time Septal to PW | M Time measurement ms | |||
| MAPSE | M Distance measurement cm | |||
| RV (M) | All RV (M) M Continuous measurement | |||
| RVAWd M Distance measurement cm | ||||
| RVIDd | M Distance measurement cm | |||
| RVAWs | M Distance measurement cm | |||
| RVIDs | M Distance measurement cm | |||
| RVPEP | M | Time measurement | ms | |
| RVET | M Time measurement ms | |||
| TAPSE | M Distance measurement cm | |||
| LV Vol. (Simpson) | LVEDV A2C | All | Dist 20 | ml |
| LVESV A2C | All | Dist 20 | ml | |
| LVEDV A4C | All | Dist 20 | ml | |
| LVESV A4C | All | Dist 20 | ml | |
| BP HR | All | Heart Rate | bpm | |
| LV Vol. (A/L) | LVEDV A2C AL All Dist 20 ml | |||
| LVESV A2C AL All Dist 20 ml | ||||
| HR A2C All Heart Rate bpm | ||||
| LVEDV A4C AL All Dist 20 ml | ||||
| LVESV A4C AL All Dist 20 ml | ||||
| HR A4C All Heart Rate bpm | ||||
| LV Vol. (Bullet) | LVAd SAX MV All Area measurement cm | 2 | ||
| LVAs SAX MV All Area measurement cm | 2 | |||
| LVLd Apical | All | Distance measurement | cm | |
| LVLs Apical | All | Distance measurement | cm | |
| HR Bullet | All Heart Rate bpm | |||
| LV Mass | All LV Mass | All | Continuous measurement | - |
| LVAd SAX PM Epi | All Area measurement cm | 2 | ||
| LVAd SAX PM All Area measurement cm | 2 | |||
| LVLd apical | All | Distance measurement | cm | |
| LV TE a | All | Distance measurement | cm | |
| LV TE d | All | Distance measurement | cm | |
| Aorta | Ao Diam (2D) | All | Distance measurement | cm |
| Ao Diam(M) | M | Distance measurement | cm | |
| Asc Ao Diam | All | Distance measurement | cm | |
| Desc Ao Diam | All | Distance measurement | cm | |
| Ao Arch Diam | All | Distance measurement | cm | |
| Ao Isth Diam | All | Distance measurement | cm | |
| Ao ST Junct Diam | All | Distance measurement | cm | |
| Ao Sinus Diam | All | Distance measurement | cm | |
| LA | All LA Vol. All Continuous measurement | |||
| LA/Ao (2D) All Distance measurement cm | ||||
| LA Diam (2D) All Distance measurement cm | ||||
| LA/Ao(M) M Distance measurement cm | ||||
| LA/Diam(M) M Distance measurement cm | ||||
| LA Major All Distance measurement cm | ||||
| LA Minor All Distance measurement cm | ||||
| LAAd A2C All | Area measurement | cm | 2 | |
| LAAs A2C All | Area measurement | cm | 2 | |
| LAAd A4C All | Area measurement | cm | 2 | |
| LAAs A4C All | Area measurement | cm | 2 | |
| LA Vol. (Simpson) | LAEDV A2C All | Dist 20 | ml | |
| LAESV A2C All | Dist 20 | ml | ||
| LAEDV A4C All | Dist 20 | ml | ||
| LAESV A4C All | Dist 20 | ml | ||
| RA | RA Diam (2D) | All Distance measurement cm | ||
| RA Major All Distance measurement cm | ||||
| RA Minor | All Distance measurement cm | |||
| RAAd | All | Area measurement | cm2 | |
| RAAs | All | Area measurement | cm2 | |
| RAEDV | All | Dist 20 | ml | |
| RAESV | All | Dist 20 | ml | |
| LVOT | Auto Trace PW Doppler Spectrum trace | |||
| Limited Trace PW Doppler Spectrum trace | ||||
| Manual Trace PW Doppler Spectrum trace | ||||
| LVOT Diam All Distance measurement cm | ||||
| LVOT Vmax PW Velocity measurement cm/s | ||||
| LVOT VTI PW Doppler Spectrum trace cm | ||||
| LOVT HR PW | Heart Rate bpm | |||
| RA Vol. (Simpson) | RAEDV A2C | All | Dist 20 | ml |
| RAESV A2C All | Dist 20 ml | |||
| RAEDV A4C | All | Dist 20 | ml | |
| RAESV A4C All | Dist 20 ml | |||
| RVOT | Auto Trace PW Doppler Spectrum trace | |||
| Limited Trace PW Doppler Spectrum trace | ||||
| Manual Trace PW Doppler Spectrum trace | ||||
| RVOT Diam All Distance measurement cm | ||||
| RVOT Vmax | PW Velocity measurement cm/s | |||
| RVOT VTI | PW Doppler Spectrum trace cm | |||
| RVOT HR | PW | Heart Rate | bpm | |
| AV | AV Auto Trace PW | Doppler Spectrum trace | ||
| AV Limited Trace PW | Doppler Spectrum trace | |||
| AV Manual Trace PW | Doppler Spectrum trace | |||
| AV Cusp All Distance measurement cm | ||||
| AV Cusp(M) M Distance measurement cm | ||||
| AV Diam All Distance measurement cm | ||||
| AVA Planimetry | All | Area measurement | cm2 | |
| Q to AV Open | M | Time measurement | ms | |
| R-R Int | M, PW | Heart Rate | bpm | |
| AV Vmax PW | Velocity measurement cm/s | |||
| AV PHT PW | Time measurement ms | |||
| AV VTI | PW | Doppler Spectrum trace | cm | |
| AV AccT PW | Time measurement ms | |||
| AV ET | PW | Time measurement | ms | |
| AV AccT/ET | PW | Continuous measurement | ||
| AV DecT | PW | Time measurement | ms | |
| Desc. Ao Vel. | PW | Velocity measurement | cm/s | |
| AR Auto Trace PW | Doppler Spectrum trace | |||
| AR Limited Trace PW | Doppler Spectrum trace | |||
| AV | AR Manual Trace PW | Doppler Spectrum trace | ||
| AR VCW All Distance measurement cm | ||||
| AR Vmax PW Velocity measurement cm/s | ||||
| AR ed Vmax PW Velocity measurement cm/s | ||||
| AR PHT PW Time measurement ms | ||||
| AR VTI PW Doppler Spectrum trace cm | ||||
| AR AccT PW Time measurement ms | ||||
| AR DecT | PW Time measurement ms | |||
| AR PISA Rad C | PISA-Radius cm | |||
| AVO | PW Time measurement ms | |||
| AVC | PW Time measurement ms | |||
| AR Jet Diam | All Distance measurement cm | |||
| AR Jet Area | All | Area measurement | cm2 | |
| MV | All MV(M) | M | Continuous measurement - | |
| MV D-E Excursion | M Distance measurement cm | |||
| MV E-F Slope | M | Velocity measurement cm/s | ||
| MV A-C Interval | M | Time measurement ms | ||
| MV EPSS M Distance measurement cm | ||||
| MV Vp | M | Velocity measurement cm/s | ||
| MV Ann Diam | All Distance measurement cm | |||
| MV Diam1 | All Distance measurement cm | |||
| MV Diam2 | All Distance measurement cm | |||
| MVA Planimetry | All | Area measurement | cm2 | |
| MV Auto Trace | PW | Doppler Spectrum trace | ||
| MV Limited Trace | PW Doppler Spectrum trace | |||
| MV Manual Trace | PW Doppler Spectrum trace | |||
| MV | MV E-DT-A PW Continuous measurement | |||
| MV E/A PW Continuous measurement - | ||||
| MV Peak E PW Velocity measurement cm/s | ||||
| MV Peak A PW Velocity measurement cm/s | ||||
| MV VTI PW Distance measurement cm | ||||
| MV Vmax | PW Velocity measurement cm/s | |||
| MV PHT | PW Time measurement | ms | ||
| MV AccT | PW Time measurement | ms | ||
| MV DecT PW | Time measurement ms | |||
| MV A Dur PW | Time measurement ms | |||
| MV ET | PW Time measurement | ms | ||
| R-R Int M, PW | Heart Rate bpm | |||
| MR Auto Trace | PW Doppler Spectrum trace | |||
| MR Limited Trace | PW Doppler Spectrum trace | |||
| MR Manual Trace | PW Doppler Spectrum trace | |||
| MR VCW All Distance measurement cm | ||||
| MR Vmax | PW Velocity measurement cm/s | |||
| MR VTI PW Doppler Spectrum trace cm | ||||
| MR dp/dt PW Continuous measurement mmHg/s | ||||
| MV IVRT PW | Time measurement ms | |||
| MR PISA Rad C | PISA-Radius cm | |||
| MV HR | PW Heart Rate | bpm | ||
| E/E' | PW Continuous measurement | |||
| Measurement Menu Item Mode Method Unit | ||||
| TV | TV D-E Excursion M Distance measurement cm | |||
| TV E-F Slope M Velocity measurement cm/s | ||||
| TV A-C Interval M Time measurement ms | ||||
| TV Auto Trace PW | Doppler Spectrum trace | |||
| TV Limited Trace | PW | Doppler Spectrum trace | ||
| TV Manual Trace | PW | Doppler Spectrum trace | ||
| TV Ann Diam | All | Distance measurement | cm, mm | |
| TV Diam1 | All | Distance measurement cm | ||
| TV Diam2 | All | Distance measurement cm | ||
| TVA Planimetry | All | Area measurement | cm | |
| Q to TV Open | M Time measurement ms | |||
| TV Vmax | PW | Velocity measurement cm/s | ||
| TV VTI | PW | Doppler Spectrum trace | cm | |
| TV E/A | PW | Continuous measurement - | ||
| TV Peak E | PW | Velocity measurement cm/s | ||
| TV Peak A | PW | Velocity measurement cm/s | ||
| TV PHT PW | Time measurement ms | |||
| TV AccT | PW | Time measurement ms | ||
| TV DecT PW | Time measurement ms | |||
| TV A Dur | PW | Time measurement ms | ||
| R-R Int | M, PW | Heart Rate | bpm | |
| TR Auto Trace PW | Doppler Spectrum trace | |||
| TR Limited Trace | PW | Doppler Spectrum trace | ||
| TR Manual Trace | PW | Doppler Spectrum trace | ||
| TR VCW | All | Distance measurement | cm | |
| TR Vmax | PW | Velocity measurement cm/s | ||
| TV | TR VTI PW Doppler Spectrum trace cm | |||
| TR dp/dt | PW | Calculation after time measurement mmHg/s | ||
| TR PISA Rad C PISA-A Radius cm | ||||
| RAP M, PW RAP | mmHg | |||
| TV HR | PW | Heart Rate | bpm | |
| PV | PV Auto Trace | PW Doppler Spectrum trace | ||
| PV Limited Trace | PW Doppler Spectrum trace | |||
| PV Manual Trace | PW Doppler Spectrum trace | |||
| PV Ann Diam | All | Distance measurement | cm | |
| PVA Planimetry | All | Area measurement | cm2 | |
| PV Vmax | PW | Velocity measurement | cm/s, m/s | |
| PV PHT | PW | Velocity measurement | cm/s | |
| PV VTI PW Doppler Spectrum trace cm | ||||
| PV AccT/ET | PW | Continuous measurement | ||
| PV AccT PW | Time measurement ms | |||
| PV ET | PW | Time measurement | ms | |
| PV DecT | PW | Time measurement | ms | |
| R-R Int | M, PW | Heart Rate | bpm | |
| Q to PV Open | M | Time measurement | ms | |
| Q to PV Close | M | Time measurement | ms | |
| PR Auto Trace | PW Doppler Spectrum trace | |||
| PR Limited Trace | PW Doppler Spectrum trace | |||
| PR Manual Trace | PW Doppler Spectrum trace | |||
| MPA Vmax | PW | Velocity measurement | cm/s | |
| PR VCW All | Distance measurement cm | |||
| PR Vmax | PW | Velocity measurement | cm/s | |
| PV | PR PHT PW Time measurement ms | |||
| PR VTI PW Doppler Spectrum trace cm | ||||
| PR AccT PW Time measurement ms | ||||
| PR DecT PW Time measurement ms | ||||
| PR End Vel. PW Velocity measurement cm/s | ||||
| PR PISA Rad C | PISA-Radius cm | |||
| Tei Index | All LV Tei Index | PW | Continuous measurement | |
| LV TST PW Time measurement ms | ||||
| MV ET PW Time measurement ms | ||||
| LV IVCT | PW Time measurement ms | |||
| LV IVRT PW Time measurement ms | ||||
| All RV Tei Index | PW | Continuous measurement | ||
| RV TST | PW Time measurement ms | |||
| RV ET PW Time measurement ms | ||||
| RV IVCT PW Time measurement ms | ||||
| RV IVRT | PW Time measurement ms | |||
| Pulm. Veins | P Vein S/D PW | Continuous measurement | ||
| P Vein S Vmax | PW Velocity measurement cm/s | |||
| P Vein D Vmax | PW Velocity measurement cm/s | |||
| P Vein A Vmax | PW Velocity measurement cm/s | |||
| P Vein A Dur | PW Time measurement ms | |||
| Hepatic Veins | H Vein S/D PW | Continuous measurement | ||
| H Vein S Vmax | PW Velocity measurement cm/s | |||
| H Vein D Vmax | PW Velocity measurement cm/s | |||
| H Vein A Vmax | PW Velocity measurement cm/s | |||
| H Vein A Dur | PW Time measurement ms | |||
| Tissue Doppler | LV E'-A'-S' PW Continuous measurement | |||
| LV Peak E' PW Velocity measurement cm/s | ||||
| LV Peak A' PW Velocity measurement cm/s | ||||
| LV Peak S' PW Velocity measurement cm/s | ||||
| LV AccT PW Time measurement ms | ||||
| LV DecT PW Time measurement ms | ||||
| RV E'-A'-S' | PW Continuous measurement | |||
| RV Peak E' | PW Velocity measurement cm/s | |||
| RV Peak A' | PW Velocity measurement cm/s | |||
| RV Peak S' | PW Velocity measurement cm/s | |||
| RV AccT | PW Time measurement ms | |||
| RV DecT PW | Time measurement ms | |||
| LV Lat E'-A'-S' | PW Continuous measurement | |||
| LV Lat Peak E' | PW Velocity measurement cm/s | |||
| LV Lat Peak A' | PW Velocity measurement cm/s | |||
| LV Lat Peak S' | PW Velocity measurement cm/s | |||
| LV Lat AccT | PW | Time measurement ms | ||
| LV Lat DecT | PW | Time measurement ms | ||
| Qp/Qs | Systemic VTI PW | Doppler Spectrum trace cm | ||
| Systemic HR PW | Heart Rate bpm | |||
| Pulmonic VTI | PW | Doppler Spectrum trace cm | ||
| Pulmonic HR | PW | Heart Rate | bpm | |
| Shunts | MPA Diam All Distance measurement cm | |||
| LPA Diam All Distance measurement cm | ||||
| RPA Diam All Distance measurement cm | ||||
| SVC Diam, Exp All Distance measurement cm | ||||
| SVC Diam, Insp All Distance measurement cm | ||||
| SVC S/D PW Continuous measurement | ||||
| SVC S Vmax PW Velocity measurement cm/s | ||||
| SVC D Vmax | PW Velocity measurement cm/s | |||
| SVC A Vmax | PW Velocity measurement cm/s | |||
| SVC A Dur PW | Time measurement ms | |||
| IVC S/D | PW Continuous measurement | |||
| IVC S Vmax | PW Velocity measurement cm/s | |||
| IVC D Vmax | PW Velocity measurement cm/s | |||
| IVC A Vmax | PW Velocity measurement cm/s | |||
| IVC A Dur PW | Time measurement ms | |||
| PEd | All Distance measurement cm | |||
| PEs | All Distance measurement cm | |||
| IVC | IVC Diam, Exp | All Distance measurement cm | ||
| IVC Diam, Insp All Distance measurement cm | ||||
| IVC Diam, Exp(M) | M Distance measurement cm | |||
| IVC Diam, Insp(M) | M Distance measurement cm | |||
Dist 20
Trace the circumference of the heart and then draw the axis of the heart. The system automatically draws 20 lines perpendicular to the axis to calculate the volume.

NOTE:
▶ MR dp/dt can be measured only at 1-3 m/s. In Dual 2D Mode, two images can be viewed simultaneously. For RVAWd, RVIDd, RVAWs, and RVIDs, see the LV measurement method. MPA Diam, RPA Diam, and LPA Diam are measured under Aortic Valve Level in Parasternal Short Axis. ▶ C Mode is mainly used for measuring reverse cardiac blood flow. Since velocity data is required for measuring PISA-Radius or PISA-Alias Vel., set color display to Velocity or Vel + Var in C mode. For relevant settings, please refer to 'Color Doppler Mode' in 'Chapter 7. Diagnosis Mode'. ▶ Tissue Doppler can be measured in TDI Mode.
Abdomen Calculations
The measurement method for each item is the same as for basic measurement. Measured items are automatically entered into the report.

NOTE:
It is convenient to calculate each measurement value on the Spectral Doppler image. For information on basic measurement methods, refer to 'Basic Measurements' and 'Common Measurement Methods' in this chapter.
Abdomen Measurement Menu
| Measurement Menu Item Mode Method Unit | ||||
| Gallbladder | CBD All Distance measurement cm | |||
| GB Wall All Distance measurement cm | ||||
| Length All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
| Height All Distance measurement cm | ||||
| Pancreas | Head All Distance measurement cm | |||
| Body | All Distance measurement cm | |||
| Tail | All Distance measurement cm | |||
| Duct | All Distance measurement cm | |||
| Bowel | Stomach Wall | All | Distance measurement | cm |
| Small Bowel Wall | All | Distance measurement | cm | |
| Large Bowel Wall | All | Distance measurement | cm | |
| Lt. Kidney Vol.Rt. Kidney Vol.LiverSpleen | Length All Distance measurement cm | |||
| Height All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
| AortaLt.RARt.RALt.Seg.ARt.Seg ALt.Arc.ARt.Arc ACeliac ASplenic AC Hepatic AR Hepatic AL Hepatic ASMAIMA | %StD All | Calculation after distance measurement | % | |
| %StA All | Calculation after area measurement | % | ||
| Auto Trace PW Doppler Spectrum trace | ||||
| Limited Trace PW Doppler Spectrum trace | ||||
| Manual Trace PW Doppler Spectrum trace | ||||
| PS PW Velocity measurement cm/s | ||||
| ED | PW Velocity measurement cm/s | |||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement - | |||
| RI | PW Velocity measurement - | |||
| S/D PW | Continuous measurement | |||
| Vol.Folw(D) | PW | Calculate after measuring TAmean, Vesl., and Dist. | ml/min | |
| Vol.Flow(A) | PW | Calculate after measuring TAmean, Vesl., and Area | ml/min | |
| IVCRenal VR Potal VM Potal VL Potal VR Hepatic VM Hepatic VL Hepatic VSMVIMVSplenic V | VMax | PW Velocity measurement cm/s | ||
| Dur T PW | Time measurement ms | |||
| Vesl.Dist. | PW | Distance measurement | cm | |
| Vesl.Area | PW | Area measurement | cm2 | |
Pediatric Hips Calculations

Measurement Menu
The measurement method for each item is the same as for basic measurement. Measured items are automatically entered into the report.

NOTE: For information on basic measurement methods, refer to 'Basic Measurements' and 'Common Measurement Methods' in this chapter.
| Measurement Menu Item Mode Method Unit | ||||
| Hip Angle Type ALL | Calculation of angle after measurement of three straight lines | degree | ||

Measurement Methods
- Specify the first straight line by using the trackball and the Set button.
Use the trackball to place the cursor at the desired position and then press the Set button.

Repositioning a Point
Before pressing the Set button to confirm the position, you can press the Change button to reset the position of the point.
- Repeat the above process to specify two other straight lines.
- The angle between them will be calculated automatically.
α: The angle between the first and the second straight lines β: The angle between the first and the third straight lines
- When the measurement has been taken, the result is shown on the screen.
See the table below for Hip Joint Type information:
| Type α β | ||
| 1a 60 ≤ α < 90 0 < β < 55 | ||
| 1b 60 ≤ α < 90 55 ≤ β < 90 | ||
| 2a/b 50 ≤ α < 60 0 < β < 90 | ||
| 2c 43 ≤ α < 50 77 ≤ β < 90 | ||
| d 43 ≤ α < 50 0 < β < 77 | ||
| 3/4 0 < α < 43 |
[Table 8.2 Hip Joint Type Table]
Urology Calculations

Before Taking Urology Measurements
Set the related menus for convenient measurement.
Refer to the 'Setting Measurements' section in 'Chapter 3. Utilities' for additional information.

Measurement Menu
The measurement method for each item is the same as for basic measurement. Measured items are automatically entered into the report.

NOTE:
For information on basic measurement methods, refer to 'Basic Measurements' and 'Common Measurement Methods' in this chapter. For references on measurement items, refer to the Reference Manual Part 2.
3Distance
Calculate a volume by measuring three distances.
| Measurement Menu Item Mode Method Unit | ||||
| WG ProstateT-ZoneBladderPre ResidualPost Residual | Length All Distance measurement cm | |||
| Height All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
| Lt. RenalRt. Renal | Length All Distance measurement cm | |||
| Height All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
| Renal Pelvis All Distance measurement cm | ||||
Small Parts Calculations
The measurement method for each item is the same as for basic measurement. Measured items are automatically entered into the report.

NOTE:
It is convenient to calculate each measurement value on the Spectral Doppler image. For information on basic measurement methods, refer to 'Basic Measurements' and 'Common Measurement Methods' in this chapter.
Thyroid Measurement Menu
| Measurement Menu Item Mode Method Unit | ||||
| Lt. Thyroid Vol.Rt. Thyroid Vol. | Length All Distance measurement cm | |||
| Height All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
| Lt. Thyroid FlowRt. Thyroid Flow | %StD | All | Calculation after distance measurement | % |
| %StA | All | Calculation after area measurement | % | |
| Auto Trace | PW | Doppler Spectrum trace | ||
| Limited Trace | PW | Doppler Spectrum trace | ||
| Manual Trace | PW | Doppler Spectrum trace | ||
| PS | PW Velocity measurement cm/s | |||
| ED | PW Velocity measurement cm/s | |||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement – | |||
| RI | PW Velocity measurement – | |||
| S/D | PW | Continuous measurement – | ||
| Vol.Flow(D) | PW | Calculate after measuring TAmean, Vesl., and Dist. | ml/min | |
| Vol.Flow(A) | PW | Calculate after measuring TAmean, Vesl., and Area | ml/min | |
| Lt. Mass 1–5Rt. Mass 1–5 | Length All Distance measurement cm | |||
| Height All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
Breast Measurement Menu
| Measurement Menu Item Mode Method Unit | ||||
| Lt. Mass 1–10Rt. Mass 1–10 | Length All Distance measurement cm | |||
| Height All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
| Lt. Breast FlowRt. Breast Flow | %StD All | Calculation after distance measurement | % | |
| %StA All | Calculation after area measurement | % | ||
| Auto Trace PW | Doppler Spectrum trace | |||
| Limited Trace | PW | Doppler Spectrum trace | ||
| Manual Trace | PW | Doppler Spectrum trace | ||
| PS | PW Velocity measurement cm/s | |||
| ED | PW Velocity measurement cm/s | |||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement – | |||
| RI | PW Velocity measurement – | |||
| S/D | PW | Continuous measurement | – | |
| Vol.Flow(D) | PW | Calculate after measuring TAmean, Vesl., and Dist. | ml/min | |
| Vol.Flow(A) | PW | Calculate after measuring TAmean, Vesl., and Area | ml/min | |
Testis Calculations
| Measurement Menu Item Mode Method Unit | ||||
| Lt. Testis Vol.Rt. Testis Vol. | Length All Distance measurement cm | |||
| Height All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
| Epididymis | D1 All Distance measurement cm | |||
| D2 All Distance measurement cm | ||||
| Lt. Testis FlowRt. Testis Flow | %StD | All | Calculation after distance measurement | % |
| %StA | All | Calculation after area measurement | % | |
| Auto Trace | PW | Doppler Spectrum trace | ||
| Limited Trace | PW | Doppler Spectrum trace | ||
| Manual Trace | PW | Doppler Spectrum trace | ||
| PS | PW Velocity measurement cm/s | |||
| ED | PW Velocity measurement cm/s | |||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement - | |||
| RI | PW Velocity measurement - | |||
| S/D | PW | Continuous measurement - | ||
| Vol.Flow(D) | PW | Calculate after measuring TAmean, Vesl., and Dist. | ml/min | |
| Vol.Flow(A) | PW | Calculate after measuring TAmean, Vesl., and Area | ml/min | |
| Lt. Mass 1-5Rt. Mass 1-5 | Length All Distance measurement cm | |||
| Height All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
Superficial Calculations
| Measurement Menu Item Mode Method Unit | ||||
| Lt. Superficial Vol.Rt. Superficial Vol. | Length All Distance measurement cm | |||
| Height All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
| Lt. Superficial FlowRt. Superficial Flow | %StD All | Calculation after distance measurement | % | |
| %StA All | Calculation after area measurement | % | ||
| Auto Trace PW Doppler Spectrum trace | ||||
| Limited Trace | PW Doppler Spectrum trace | |||
| Manual Trace | PW Doppler Spectrum trace | |||
| PS | PW Velocity measurement cm/s | |||
| ED | PW Velocity measurement cm/s | |||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement - | |||
| RI | PW Velocity measurement - | |||
| S/D | PW | Continuous measurement - | ||
| Vol.Flow(D) | PW | Calculate after measuring TAmean, Vesl., and Dist. | ml/min | |
| Vol.Flow(A) | PW | Calculate after measuring TAmean, Vesl., and Area | ml/min | |
| Vel A | PW Velocity measurement cm/s | |||
| Vel B PW Velocity measurement cm/s | ||||
| Lt. Mass 1-5Rt. Mass 1-5 | Length All Distance measurement cm | |||
| Height All Distance measurement cm | ||||
| Width All Distance measurement cm | ||||
Musculoskeletal Calculations

Measurement Menu
The measurement method for each item is the same as for basic measurement. Measured items are automatically entered into the report.

NOTE: For information on basic measurement methods, refer to 'Basic Measurements' and 'Common Measurement Methods' in this chapter.
| Measurement Menu Item Method Unit | |||
| Lt. ShoulderRt. Shoulder | #1~#10 Distance | measurement cm | |
| Lt. WristRt. Wrist | #1~#10 Distance | measurement cm | |
| Lt. KneeRt. Knee | #1~#10 Distance | measurement cm | |
| Lt. AnkleRt. Ankle | #1~#10 Distance | measurement cm | |
Vascular Calculations
The measurement method for each item is the same as for basic measurement. Measured items are automatically entered into the report.

NOTE:
For information on basic measurement methods, refer to 'Basic Measurements' and 'Common Measurement Methods' in this chapter. For references on measurement items, see 'Reference Manual Part 1'.
Carotid Measurement Menu
| Measurement Menu Item Mode Method Unit | ||||
| Lt. CCA Rt. CCA Lt. Bulb Rt. Bulb Lt. ICA Rt. ICA Lt. ECA Rt. ECA | %StD All | Calculation after distance measurement | % | |
| %StA All | Calculation after area measurement | % | ||
| IMT All Distance measurement mm | ||||
| IMT Mean All | Average calculation after distance measurement | mm | ||
| Auto Trace | PW | Doppler Spectrum trace | ||
| Limited Trace | PW | Doppler Spectrum trace | ||
| Manual Trace | PW | Doppler Spectrum trace | ||
| PS | PW Velocity measurement cm/s | |||
| ED PW Velocity measurement cm/s | ||||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement - | |||
| RI | PW Velocity measurement - | |||
| S/D | PW | Continuous measurement | ||
| Vol.Flow(D) | PW | Calculate after measuring TAmean, Vesl., and Dist. | ml/min | |
| Vol.Flow(A) | PW | Calculate after measuring TAmean, Vesl., and Area | ml/min | |
| Lt. Subclavian A Rt. Subclavian A Lt. Vertebral A Rt. Vertebral A | %StD All | Calculation after distance measurement | % | |
| %StA All | Calculation after area measurement | % | ||
| Auto Trace PW | Doppler spectrum trace | |||
| Limited Trace PW | Doppler spectrum trace | |||
| Manual Trace PW | Doppler spectrum trace | |||
| PS PW Velocity measurement cm/s | ||||
| ED | PW Velocity measurement cm/s | |||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement - | |||
| RI | PW Velocity measurement - | |||
| S/D | PW | Continuous measurement | ||
| Vol.Flow(D) | PW | Calculation after measurement of TAmean, Vesl. Dist | ml/min | |
| Vol.Flow(A) | PW | Calculation after measurement of TAmean, Vesl. Area | ml/min | |

Auto IMT+
This function allows you to take IMT measurements easily and quickly.

NOTE:
▶ Auto IMT+ is an optional feature of this product. Auto IMT+ is available only under the following conditions:
- Probe: Linear Probe
- Application: Vascular – Diagnosis Mode: 2D Freeze Mode
Auto IMT+ Screen
![Auto IMT + SAMSUNG 20160621-032528PM Tls 0.19 MI 1.35 2016-06-21 03:25:48 PM CarMax LA3-10AD 3.5 cm 44 Hz [2D] Run 1 Gm 100 DR 112 Map? FA 7 P 90% Min: 100 min Max: 100 min Min: 100 min Max: 100 min Min: 100 min Max: 100 min Max: 100 min Max: 100 min Max: 100 min Max: 100 min Max: 100 min Max: 100 m…](/content/2026/06/1369957/images/267bdaf75cfb210d7372d41ce3199d80b2ee2ddfed8c798816dd229e17cd99cf.jpg)
[Figure 8.2 Auto IMT+]
Auto IMT+ Measurement
- After checking the probe and the application, start a carotid scan.
- Press Freeze to obtain the desired image. Use the trackball to select the image you want to measure the IMT on.
- Tap Auto IMT+ on the touch screen. The Auto IMT+ screen will appear.
- Use the trackball and the Set button to set a location for IMT measurement.
Measure the length of a specific section. Press and hold the Set button at the starting point and use the trackball to specify the end point. You can change the Near and Far positions of the measured area by using the Change button on the control panel.
- Once the measurement finishes, measurement values will be listed in a table.
- Use the touch screen, or the dial-button to select the measurement label, the location of the blood vessel, and the direction of blood flow.
Setting Up Measurement Items
On the touch screen menu, you may set up the measurement item, the location of the blood vessel, and the direction of blood flow.
▶ Measurement item: Select measurement items among CCA, Bulb, ICA, and ECA. Use the touch screen button. Direction: Select the blood flow direction either left or right. Use dial-button 2. ▶ Location: Select Prox(Proximal), Mid(Middle), or Dist(Distal) as the location of the blood vessel to measure. Use dial-button 3.

NOTE: Exit Auto IMT+ using following functions:
▶ Change Image Mode and Zoom Annotation, BodyMarker, Patient, SonoView, Report, Setup, Help
Measurement Result
Max: The maximum thickness of the Intima/Adventitia pair. Mean: The average thickness of the Intima/Adventitia pair. SD: Standard Deviation QI: The distance ratio of the measured point in a distance for Quality Index measurement.

NOTE: Location of measurement result window cannot be changed and it is displayed in the position established before Auto IMT+ measurement.
Applying to the Report
Auto IMT+ measurement results are automatically applied to the report. Report\Auto IMT+ measurement results can be checked.
Auto IMT+ Measurement Analysis
- Tapping the Analysis on the touch screen opens Analysis screen.
- Select the desired analysis from Framingham/CHD, Risk Factor, Normal IMT, or User Graph by using the trackball and the Set button. The bar that corresponds to the measurement will be displayed on each graph. However, if the measurement is smaller than Framingham/CHD, or Risk Factor, no bar will be displayed.
Tips!
User Graph
You can use User Graph to adjust the graph and analyze the measurement results as you want.
- To finish analyzing, tap the Exit button on the touch screen.
- The following materials were referred to when analyzing the measurements of Auto IMT+.
Framingham/CHD
Correlation between the Framingham Risk Score and Intima Media Thickness: the Paroi Arterielle et Risque Cardio-vasculair (PARC) Study.
Pierre-Jean Touboul, EricVicaut, Julien Labreuche, Jean-Pierre Belliard, Serge Cohen, Serge Kownator, Jean-Jacques Portal, Isabelle Pithois-Merli, Pierre Amarenco. On behalf of PARC Study participating physicians.
Risk Factor
Mannheim Carotid Intima-Media Thickness Consensus (2004–2006)
P.J. Touboul, M.G. Hennerici, S.Meairs, H.Adams, P.Amarenco, N.Borstein, L.Csiba, M.Desvarieux, S.Ebrahim, M.Fatar, R.Hernandez Hernandez, M.Jaff, S.Kownator, P.Prati, T.Rundek, M.Sitzer, U.Schiminke, J.C. Tardif, A.Taylor, E.Vicaut, K.S.Woo, F.Zannad, M.Zureik
Normal IMT
Simon A, Gariepy J, Chironi G, Megnien JL, Levenson J: Intima-media thickness: a new tool for diagnosis and treatment of cardiovascular risk. Journal of Hypertension 20:159-169, 2002
UE Artery Measurement Menu
| Measurement Menu Item Mode Method Unit | ||||
| Lt. Subclavian A Rt. Subclavian A Lt. Axillary A Rt. Axillary A Lt. Brachial A Rt. Brachial A Lt. Radial A Rt. Radial A Lt. Ulnar A Rt. Ulnar A Lt. SPA Rt. SPA | %StD All | Calculation after distance measurement | % | |
| %StA All | Calculation after area measurement | % | ||
| Auto Trace PW | Doppler Spectrum trace | |||
| Limited Trace PW | Doppler Spectrum trace | |||
| Manual Trace PW | Doppler Spectrum trace | |||
| PS PW | Velocity measurement cm/s | |||
| ED | PW | Velocity measurement | cm/s | |
| MD | PW | Velocity measurement | cm/s | |
| PI | PW | Velocity measurement | - | |
| RI | PW | Velocity measurement | - | |
| S/D | PW | Continuous measurement | ||
| Vol. Flow(D) | PW | Calculate after measuring TAmean, Vesl., and Dist. | ml/min | |
| Vol. Flow(A) | PW | Calculate after measuring TAmean, Vesl., and Area | ml/min | |
UE Vein Measurement Menu
| Measurement Menu Item Mode Method Unit | |||
| Lt. Internal Jugular V Rt. Internal Jugular V Lt. Innominate V Rt. Innominate V Lt. Subclavian V Rt. Subclavian V Lt. Axillary V Rt. Axillary V Lt. Brachial V Rt. Brachial V Lt. Cephalic V Rt. Cephalic V Lt. Basilic V Rt. Basilic V Lt. Radial V Rt. Radial V Lt. Ulnar V Rt. Ulnar V | Vmax PW Velocity measurement cm/s | ||
| Dur T PW Time measurement ms | |||
| Vesl. Dist PW Distance measurement cm | |||
| Vesl. Area PW | Volume measurement | cm2 | |
LE Artery Measurement Menu
| Measurement Menu Item Mode Method Unit | ||||
| Lt. CIA, Rt. CIA Lt. IIA, Rt. IIA Lt. EIA, Rt. EIA Lt. CFA, Rt. CFA Lt. SFA, Rt. SFA Lt. DF, Rt. DF Lt. Popliteal A Rt. Popliteal A Lt. ATA, Rt. ATA Lt. PTA, Rt. PTA Lt. Peroneal A Rt. Peroneal A Lt. DPA, Rt. DPA Lt. MPA, Rt. MPALt. LPA, Rt. LPALt. Metatarsal A Rt. Metatarsal A Lt. DigitalA Rt. Digital A | %StD All | Calculation after distance measurement | % | |
| %StA All | Calculation after area measurement | % | ||
| Auto Trace PW | Doppler Spectrum trace | |||
| Limited Trace PW | Doppler Spectrum trace | |||
| Manual Trace PW | Doppler Spectrum trace | |||
| PS PW Velocity measurement cm/s | ||||
| ED | PW Velocity measurement cm/s | |||
| MD | PW Velocity measurement cm/s | |||
| PI | PW Velocity measurement - | |||
| RI | PW Velocity measurement - | |||
| S/D | PW | Continuous measurement | ||
| Vol.Flow(D) | PW | Calculate after measuring TAmean, Vesl., and Dist. | ml/min | |
| Vol.Flow(A) | PW | Calculate after measuring TAmean, Vesl., and Area | ml/min | |
LE Vein Measurement Menu
| Measurement Menu Item Mode Method Unit | |||
| Lt. CIV, Rt. CIVLt. IIV, Rt. IIVLt. EIV, Rt. EIVLt. CFV, Rt. CFVLt. PFV, Rt. PFVLt. SFV, Rt. SFVLt. GSV, Rt. GSVLt. Popliteal VRt. Popliteal VLt. LSV, Rt. LSVLt. ATV, Rt. ATVLt. PTV, Rt. PTVLt. Peroneal VRt. Peroneal VLt. MPV, Rt. MPVLT. LPV, Rt. LPVLt. Metatarsal VRt. Metatarsal VLt. Digital VRt. Digital V | Vmax PW Velocity measurement cm/s | ||
| Dur T PW Time measurement ms | |||
| Vesl.Dist PW Distance measurement cm | |||
| Vesl.Area. PW Volume measurement cm | 2 | ||
Reports
The measurement results are summarized by application, and shown on the screen in a report format.
Viewing Reports
Press the Report button on the control panel to switch to the Ultrasound measure Report screen.
The measurement results are summarized by application, and shown on the screen in a report format.

NOTE: Reports are displayed only for the applications that have measurement results.
Page
Pages are divided into Worksheet, Description, Comment, and image for each application; opening a report will show the Preview screen by default. You can go to each section from the Left Menu or Touch; if a section contains large amount of data to display, it will be divided into multiple pages, which you may navigate pages using Touch or dial-button.
Report Preview
It provides report preview before print-out.

Property
Set the type of Printer, Application, Print color, and Page.
Settings
You can set Header & Footer, Logo, Title, Patient Info, and Table Form of a report.

[Figure 8.3 Report Preview]
Printer
Select a USB Printer connected to the equipment.
Range
Set a page to save the print-out or images. Current Page selects the currently active page and All Applications selects all pages.
Application
Set application to save the Preview screen, printout, or images. Current Application will select applications established in the worksheet, while All Application will select all applications.
Color (Print)
Set the color of report in printout.
Color (Store)
Set the color of image to save images.
Page
Set pages to save the Preview screen, printout, or images. Echo Page will select a landscape-formed report.
Header & Footer
Set the header and footer. Checking Show Date in Footer will display the date at the bottom right-hand side.

[Figure 8.4 Header & Footer]
Logo
Set a logo on the top of the report. It can be changed once it is connected to a USB.
Title
Set the title. User-selected title or a title set in the Patient will be shown.
Patient Info.
Set Patient information shown on the page. Only the 1st page shows the Patient information only on the first page.
Table Form
Set the report form. Selection of Brief will show the results only, while selection of 2 Columns will show measure items in the two columns.

[Figure 8.5 Columns]
Editing Report (Worksheet)
When you move the focus to a value that can be edited on a Worksheet, the borders of the edit window will be marked in blue, and Edit mode will start by clicking.

[Figure 8.6 Worksheet]

[Figure 8.7 Worksheet - Touch Screen]

Modify Measurements
After you have edited a value and moved the focus away from the value or pressed Enter on the keyboard, the edited value will be applied. Edited values will be indicated with an asterisk (*) in front of them. To delete a value, click the value you wish to delete; the Delete button will be activated on the touch screen. Press this button to delete the value immediately.
Default Author
Reverts to the author of Setup settings.
Delete Cell
Select and delete a measurement value.
Delete Line
Delete all values with the same label as the selected value.

NOTE: If you change the Author in a Worksheet, the GA value will be changed. The Author you change in a Report is only valid within the Report; the Author of the Setup will be used for calculating the GA when you are taking measurements in Measurement mode.
AV.US
If Composite GA is Average, you may select or exclude GA values from each Label Av.US that have been used for calculation.
Measurement Display Method
The product allows you to measure one measurement item multiple times. However, only the first three measurement results are saved in a report.
When taking measurements for the same item more than once, measurements can be displayed in four ways. On the Edit Report screen, you can specify or change the measurement display method.
Go to Set Up > Measurement > Application > Doppler > Display Absolute Value and select 'On' to display absolute values in the report.
Average
Obtain the average of the measurements and display it on the screen.
Last
Display the last measurement on the screen.
Max
Display the largest value of the measurements on the screen.
Min
Display the smallest value of the measurements on the screen.
Adding Comments
Tap Comment on the touch screen or click Comment in the left-hand menu on the Ultrasound Report screen. A screen in which you can enter opinions will appear.
Tap the bottom of the entry window or the Comment button on the touch screen and a window will open to let you select a comment you have previously entered in Setup. Select Comment in the window and the selected text will be entered at the location of the cursor in the comment entry window.

NOTE: Comments may be added at Setup > Report > Comment. For more information, refer to 'Report' in 'Chapter 3. Utilities'.

[Figure 8.8 Comment]

[Figure 8.9 Comment – Touch Screen]
Print Reports
In the Preview screen, tap the Print button on the touch screen or press Print in the left-hand menu on the Report screen. The measurement report for the selected application will be printed on a connected printer.

NOTE: You can change the printer settings for printing measurement reports at Report > Preview > Property > Print. For more information, refer to 'Peripheral Device Settings' in 'Chapter 3. Utilities'.
Saving Reports
In the Preview screen, tap Export on the touch screen. Or press Export in the left-hand side of the Report screen. The Export button is activated only when it is connected to a USB. The Export window will appear and save the report in the external storage media.
The Export window will appear and specify the directory, drive, file name, and file format. Click OK on the screen to save the report. Tap Cancel to cancel.

[Figure 8.10 Export]
Store SR
Tap Store SR on the touch screen or the Left Menu to save the Measure SR Information.
Images
Select this if you want to attach images to the report. Tap the Image button on the touch screen or press Image in the Left Menu on the Ultrasound Report screen to open a window where you can select images.

[Figure 8.11 Image]

[Figure 8.12 Image - Touch Screen]
- Select images from the thumbnail on the right.
- When images are selected, the Add button becomes active on the touch screen. Tap the Add button to attach the selected images to the report. Alternatively, double-click selected images among the thumbnails to attach the associated images.
- If you select images attached to a report, the Delete button becomes active so that you may delete attached images. Press the Delete All button to delete all attached images.
- You may select 1x1, 2x2, 3x3, or 4x4 for layout.
OB Graph
After taking OB measurements, press Report on the touchscreen or control panel. Tapping OB Graph on the touch screen displays the Graph screen.
Click/tap the Single or Quad button on the screen/touch screen. Select the layout on the graph screen. You can review the graph and trends. To return to the Report screen, tap Preview on the touch screen or the Left Menu.

NOTE: The graph function can only be used with OB reports.
The list of measured items appears on the right side of the screen. If you select an item, a graph for the selected item will appear on the screen.

NOTE:
To execute the graph function, at least one Label containing a Growth Table must be measured, and only a Label that has been measured may display graphs. A graph is created based on the patient ID, LMP and measurement date.

[Figure 8.13 OB Graph]

[Figure 8.14 OB Graph - Touch Screen]
Select a Graph
Use the Trackball and the Set button to select an item from the list.
Single
A graph representing one obstetrics measurement will be displayed.
Quad
A graph representing four obstetrics measurements will be displayed.
Hide Grid
Erase the grids on a graph.
All Fetus
Display graphs for all measured fetuses.
Current Fetus
Display graphs for the fetus currently being measured.
Trend
Tap Trend on the touch screen. The past and present measurement values for a fetus are displayed in the order of study date. It is shown on the graph with different markers for different study dates.
Current
Tap Current on the touch screen. The current measurements for a fetus are displayed by study date.
User History
The past and present measurement values for the fetus are displayed in a table format.
Manual Trend
Display the measurement values manually entered by the user, which are to be displayed on the OB Growth Curve in Trend. Enter manually in Patient window.
Prev
Return to the previous page.
Next
Move on to the next page.
Tips!
Standard Deviation & Percentile
Among the OB information, the Growth table and the typical fetal distribution, for the same number of weeks, are used to determine the following information:
The normal distribution curve. The measurements for an actual fetus or a position in EFW distribution. Whether or not a distribution point is within the normal range.
The number of weeks referenced for the Growth table can be set to LMP, Estab., DD, or Average US GA, under Pctl. Criteria; the typical setting is LMP.
When the LMP is not known or uncertain, or when the difference between the LMP and the Average US GA is substantial, care must be taken, as selecting different Pctl. Criteria can result in a significant difference.
The distribution of the number of weeks in the Growth table for the selected reference is a normal distribution. It is laterally symmetrical around 50% (the average) and it shows the distance from the average as a deviation. The deviation can be represented by Standard Deviation (SD) or Percentile terms.

[Figure 8.15 Distribution of Growth Table for the Selected Week (m: mean, σ: standard deviation)]

When represented by SD, a point near to the average indicates a value closer to ±0 SD and a point further away from the average indicates a value closer to the maximum or minimum value. The greater part of the range falls within ±3 SD, with ±1 SD representing 68.3% of the entire range. Thus it can be seen that most fetal measurements are tightly clustered around the average value.
The Percentile represents a point in a distribution from 0 to 100 inclusive. Therefore, the average point is represented as the 50th Percentile.
As shown in the figure, the average point corresponds to 0 SD (that is, the 50th Percentile). If a point is in the range between -1 SD and +1 SD, it falls within 68.3% of the entire range. This means that the point falls within the range between 16 and 84.
Further, if a point is in the range between -2 SD and +2 SD, it falls within 95.5% of the entire range. Thus, the point falls in the range between 3 and 97.
The SD and Percentile are interchangeable. Percentile can be used when a fetal measurement ranking is desired, while SD can be used when the distance between actual fetal measurements and the average measurement is sought.
While the range of Growth table references that are primarily used with OB measurement data varies depending on the user, the typical range accepted by most users is as below:
When references are created based on SD:
- -2.0 SD - +2.0 SD (when converted to Percentile: 2.28 Percentile - 97.72 Percentile)
- -1.5 SD - +1.5 SD (when converted to Percentile: 6.68 Percentile - 93.32 Percentile)
- -1.0 SD - +1.0 SD (when converted to Percentile: 15.87 Percentile - 84.13 Percentile)
When references are created based on Percentile:
- 2.5 Percentile – 97.5 Percentile (when converted to SD: -1.96 SD – 1.96 SD)
- 5.0 Percentile - 95.0 Percentile (when converted to SD: -1.645 SD - 1.645 SD)
- 10.0 Percentile - 90.0 Percentile (when converted to SD: -1.288 SD - 1.288 SD)
Description
Tap Description on the touch screen. OB and Gyn applications have their own preset items; for other applications, only user-added items are available.
General OB
This item only becomes active when OB measurement is available. Tap the General OB button on the touch screen and you can select a description for the selected item.
Early OB
This item only becomes active when OB measurement is available. Tap the Early OB button on the touch screen and you can select a description for the selected item.
Biophysical Profiles
5 items of Breathing, Movement, Tone, AFV, and NST may be left blank or set as 0, 1, or 2; the sum of each item will be displayed as Total.

NOTE: Biophysical Profile is activated only when OB measurement values exist.
Default
This value has been preset in advance by the user.
Save as Default
Save the values currently set on the screen as default values.
Clear
Delete all items.
Fetus Compare
In the OB Study Information screen, select the Number of Fetuses to enable measurement for each fetus when taking OB measurement. Tap the Fetus Compare button on the touch screen and measurement for each fetus is displayed.

[Figure 8.16 Description]

[Figure 8.17 Description – Touch Screen]
Closing Reports
Tap Exit on the touch screen or in the Left Menu. Alternatively, press the Report button on the control panel.
Image Managements
Cine/Loop 9-3 ■ Annotating Images.... 9-5
Annotation 9-5
BodyMarker 9-10
Saving, Playing and
Transferring Images 9-12
Saving Images....9-12
Playing Images....9-13
Transferring Images 9-14
■ Printing Images....9-15
Printing Images 9-15
SonoView Mode 9-16
Exam List 9-17
Review – Exam Mode 9-19
Review – Compare Mode......9-24
DICOM Q/R 9-31
Cine/Loop
Images are automatically saved in the memory during scanning. The saved images are useful for patient diagnosis and review.
The saved images can be in Cine or Loop, depending on the diagnosis mode.
2D Cine: These are images temporarily saved in all diagnosis modes except M Mode and Spectral Mode. Trace Cine: Images that are saved in M Mode and Spectral Mode.
![OB SAMSUNG 2016012B-111557AM Tts 0.5 MI 1.4 2016-01-28 11:32:02 AM 1st Trimester LAE-18SD 4.D cm 47 Hz [2D] Gen Gm 76 DR 124 Map 7 FA 5 P 94% Freeze Pointer 47 / 96 Sound/Point A Set Set 1 Stand/Point Sound/Point B Sound/Point A 1 Stand/Point 2 Stand/Point 3 Stand/Point Images 5](/content/2026/06/1369957/images/d3e82addb78ab2eef5da68f336588420ab340742592fa4f87fe3b251afdf93aa.jpg)
[Figure 9.1 Cine]
Starting and Finishing Image Review
During scanning, press the Freeze button on the control panel. The scanning is stopped and the system switches to the image review mode screen.
Press the Freeze button again to return to the scan mode.
How to Review Images
Move the cursor to the Cine or Bar in the user information area to review an image. You can search through saved images by moving the cursor with the trackball. The total number of saved images and the number of the image currently being reviewed are shown next to the bar.
Selecting Loop
Select an image to play by pressing the Change button in the control panel.
The touch screen menu is changed when images are reviewed.

[Figure 9.2 Reviewing Images - Touch Screen]
Cine Play
Tap Cine Play on the touch screen. You can stop or resume playback of an image by tapping the button.
Cine Speed
Use the Cine Speed button on the touch screen to adjust the automatic playback speed of 2D Cine or Trace Cine. Speed may be adjusted between 6-300%.

NOTE: For an explanation of the other items on the touch screen, refer to Basic Mode in 'Chapter 7. Diagnosis Mode'.
Reviewing Image in Multi-image Mode
Only the images in an active area can be reviewed. To review images in another area, change the active area by using the Dual or Quad button on the control panel.
Annotating Images
Annotation
Enter or display text, Arrow, or BodyMarker on an image. This function is useful for differentiating or marking a diagnosis area.

Starting Annotation Mode
Press the Text or Arrow button on the control panel.
![MSK SAMSUNG 20160610-11:27:05AM General LA3-16A 4.0 cm 31 Hz [20] Gen Gn 32 DR 126 Map 7 FA 5 P 90% Text Arrow Gera/Test 4 EFW Result 5 TFW Measure 6 Est 1 Scale/Test 2 Scale/Test 3 Scale/Test TIs 0.36 MI 1.36 2016-06-10 11:34:24 AM Image 2](/content/2026/06/1369957/images/345eb35a4eb3d8fdd146a3d1d617234f188b9cb042352452ae95984ba17ce365.jpg)
[Figure 9.3 Annotation Mode]

Typing Text
Press the Text button in the control panel and enter text directly with the keyboard, or select predefined text on the touch screen.
You can move the cursor by using the trackball.

Text Mode in Soft Menu

[Figure 9.4 Screen Keyboard]
Change App.
Tap the touch screen or use the dial-button to change application. The text list will change depending on the selected application.
Keyboard
Tapping the 📄 button on the touch screen opens the on-screen keyboard.
Arrow
You can enter an arrow in a place you want over an image.
Delete Last Word
Erase the last entered text. Alternatively, use the Annotation key on the keyboard.
Delete All Words
All of the entered text has been erased. Alternatively, use the Delete All key on the keyboard.

NOTE: Selecting Setup > Annotation > Select Library allows you to edit the entire page.
For a more detailed description, refer to 'Chapter 3. Utilities'.
Home
When tapping the Home button on the touch screen, it is moved to the established cursor location by Set Home button.
Set Home
Set the initial cursor position when Annotation mode starts. Place the cursor on the screen and tap Set Home on the touch screen.
Layer
Set Layer A and B.
Font Size
Select the font size to use. Select a value between 8 and 72 by using the dial-button on the touch screen.

Deleting Text
Press the Clear button on the control panel. All the text entered on the screen will be deleted.

Exiting Text Mode
Press the Exit button on the control panel or tap the Exit button on the touch screen.
Arrow Mode
You can enter an arrow over an image. This function is useful for differentiating or marking a diagnosis area.
![MSK SAMSUNG 20160610-112705AM General LA3 16A 4.8 cm 51 Hz [2D] Gen Gn 32 DB 126 Map7 FA 5 P 90% (buA3.40M/FV 0.25%) Freeze 1525 / 1525 SunRfree 4 E/W Result 5 E/W Measure 6 Test Arrow Est 1. SunRfree 2. SunRfree 3. SunRfree Image 1 Tls 0.36 MI 1.36 2016-06-10 11:34:54 AM](/content/2026/06/1369957/images/b3a59fb1dd47a924d9ac1b6a84cf408a001c7afac87338240564a478a634f352.jpg)
[Figure 9.5 Arrow Mode]
Entering Arrow
- When pressing the Arrow button on the control panel, it is converted to the entry mode. Or press the Arrow button in the keyboard.
- Use the dial-button's Arrow Type to select the type of arrow you want.
- Move the arrow to the desired position by using the trackball.
- To adjust the size of the arrow, use the dial-button's Arrow Size.
- To adjust the rotation angle of the arrow, use the dial-button's Rotate Arrow.
- Press the Set button to finish. Press Exit to cancel.

Arrow and Pointer Button
Pointer function is available in Annotation mode.
Deleting Arrows
Press the Clear and Delete All Arrows button on the control panel to delete all arrows displayed on the screen. To delete the last arrow entered, tap Delete Last Arrow on the touch screen.
Exiting Arrow Mode
Press the Exit button on the touch screen or control panel.
BodyMarker
Enter a BodyMarker over an image. This function is useful for differentiating or marking a diagnosis area.
Entering BodyMarker Mode
When pressing Marker/Ref Knob on the control, it is converted to BodyMarker entry mode and the touch screen shows BodyMarker.
![Small Parts SAMSUNG 20160610-112705AM Breast LAS-16A 4.0 cm 51 Hz [28] Gen Gm 40 DB 126 Map7 FA 5 P 90% [0xA3, HSPV, 5x727] Probe Body Marker ScorePaint 4 Sun Edd 1 ScorePaint EPW Result 5 2 ScorePaint EPW Measure 6 3 ScorePaint Images 7](/content/2026/06/1369957/images/8bf7898c5207a5d4aaebb6942bb55ed2340b9fa6dff90ac27d82213f2a7447d9.jpg)
[Figure 9.6 BodyMarker Mode]
Inserting BodyMarker
- BodyMarkers will appear on the touch screen.
The types of BodyMarker shown on the touch screen vary depending on the selected application. Up to 19 BodyMarkers can be listed on the screen at one time.
When more BodyMarkers are provided, use the > button on the touch screen to navigate the pages.
- Tap the touch screen button or dial button for the BodyMarker you want; the BodyMarker will then be displayed on the image.
- When pressing Change Lib, it can be immediately changed to a desirable application.
- Select a desirable probe shape by using the Probe Type dial-button.
- Adjust cursor angle by using the Rotate Probe dial-button or the Angle button on the control panel.
- Adjust the cursor size by using the Size dial-button.
- Set the initial cursor position when BodyMarker starts. Place the cursor on the screen and tap Store Probe Position on the touch screen.
- To finish entering BodyMarkers, press the Set or Exit button on the control panel. To cancel it, press Clear on the touch screen or tap the Delete button on the touch screen.
Changing BodyMarker Position
- Press the Change button on the control panel.
- Move the BodyMarker to a desired position by using the Trackball.
- Press the Set button on the control panel again to confirm the new position.
Deleting a BodyMarker
Press Clear on the touch screen or tap the Delete button on the touch screen.
Exiting BodyMarker Mode
Tap the Exit button on the control panel or touch screen.

NOTE: Go to Setup > BodyMarker > Library to set the library of BodyMarker. For a more detailed description, refer to 'Chapter 3. Utilities'.
Saving, Playing and Transferring Images
Saving Images

WARNING: You must always enter the patient ID, because all images are saved according to patient ID. Failure to enter a patient ID may result in a loss of and/or critical errors in previously saved images.
The saved images are displayed in the thumbnail area.
Saving Image
If no patient information has been entered, a Temp ID will be created and saved. The saved images can then be edited and managed in SonoView mode.
Saving (Still Image/Multi Frame Image)
Use the Store Clip function of U1-6 button to save it. For a detailed description for Store/Print settings, refer to 'Chapter 3. Utilities'.
When frozen, specify the range of the image you want to save by using the trackball or the Trim Start and Trim End buttons on the touch screen, then tap the configured U1-6 buttons to save the Cine image. Live Mode works as shown below according to settings. For information about the settings, please refer to 'Chapter 3. Utilities'.
If patient information does not exist, saving Cine in Freeze Mode will create a Temp ID for saving.
If all frames are saved in Live Mode, the saving process is automatically finished according to a predefined time.
The saved images are added to the thumbnail list on the right side of the screen.
Playing Images
You can play the saved images in SonoView and Diagnosis Modes.

Playing Images in SonoView

NOTE: Refer to 'SonoView' in this chapter.

Viewing Images in Diagnosis Mode
Use the Pointer button on the control panel. Note that this function is available only when there are images saved in the thumbnail list. Press the Pointer button and the cursor will appear on the screen.
Transferring Images
HS50/HS60 allows you to transfer images to a PACS system that supports DICOM. You can transfer all saved images automatically, or select a desired image and transfer it manually. For information on the DICOM server settings and DICOM operations, please refer to 'Connectivity Settings' in 'Chapter 3. Utilities'.

Transferring Images in Diagnosis Mode
You can transfer images automatically. Images are transferred by using the storage server transmission method in Setup > Connectivity > DICOM. Please refer to "Connectivity Settings" in 'Chapter 3. Utilities' of this manual.

Transferring Images in SonoView
You can transfer images manually. The following two methods are available:
Sending Exam
Send all images for an exam.
- Select an exam from the Exam List.
- Click the Send button on the left side of the screen. All images for the selected exam will be sent.
Sending Selected Images
Select a desired image from among the images for an exam and send it out.
- Select the image in the SonoView screen.
- Click the Send button on the left side of the screen. The selected images will be sent.
Printing Images

NOTE:
Use it after establishing functions in Setup > Peripherals > Printer. Configure the Store/Print functions in Setup > Customize > Buttons before use.
Printing Images
Press the U1-6 buttons on the control panel. Images are printed via an echo printer. For information on how to set up a printer, please refer to 'Chapter 3. Utilities'.
SonoView Mode
This is the HS50/HS60's integrated image management program. It provides image saving/filing, preview, delete, and other PC-compatible export features.
The image file types used in this product follow the international standard DICOM (Digital Imaging and Communication in Medicine). As a result, the PACS (Picture Archiving Communication System) can be implemented without any additional costs, and it's easy to exchange image files with other hospitals or equipment.
This product supports the Bitmap file format (.bmp), which is commonly used on standard PCs, ensuring easier exchange of image data.
Launching SonoView Mode
Press the SonoView button on the control panel. This will launch the SonoView screen.
If there are saved images available for the current exam, the information and saved images for the exam will appear when SonoView Mode starts.

WARNING:
Before saving images or using SonoView Mode, you must register the patient information. All diagnosis information in the product is saved and managed for each patient ID. As a result, saving images without entering a patient ID may result in a loss of and/or critical errors in previously saved images.
Exam List
Appears when the SonoView button is pressed. Or tap Exam List in Review Mode to see the exam list.

[Figure 9.7 Exam List]
Enter search criteria at the top of the Exam List screen, and press Search. Search results will be listed on the screen.
You can select an exam from the list and examine, delete, or send it.
Refer also to 'Printing Images' and 'Transferring Images via DICOM' later in the document.
You can select the data storage location. HDD, Media, and Q/R are provided.

Hide Search
Hides the Exam list.

Show Search
Shows hidden Exam list.
Backup Exam
Gets activated when selecting an exam in HDD. Continue to back up the selected exam. Select a drive to back up to and press Yes to continue the backup process.
Restore Exam
Gets activated when selecting an exam from USB storage or CD/DVD. Load the selected exam to HDD.
Review
When one or more exams are selected, the button is activated. The review screen for the selected exams will open. For a more detailed description, refer to 'Review Exam Mode'.
Continue
Is activated when an exam from the last 24 hours is selected. This allows you to modify the exam.
Review – Exam Mode
Press Exam in the upper left corner of the screen.
In the Exam Mode, you can review the current or saved exam.
The image that is selected on the screen will be highlighted with a white border.

Exam Mode Screen
Exams for each patient ID are displayed by date in a tree structure on the left-hand side of the screen. The number represents the number of saved images. To show or hide exams, use the trackball and the Set button to select a desired ID.

[Figure 9.8 Exam Mode]

[Figure 9.9 Exam Mode - Touch Screen]

Exam Select
Use the trackball and the Set button to select a desired exam from the list on the left-hand side of the screen. The selected exam will be marked on the list in blue. The saved image will also be displayed on screen.
Reviewing the current exam
Click Current Exam. The current exam and its images are displayed on the screen. This button is only enabled when the patient ID is registered. Exams brought from the ExamList can be closed by using the Close or Close All button.
Reviewing exams performed within the last 24 hours
Double-click an image. The exams that were performed within the last 24 hours and their images are displayed. The first performance date of the exam appears in the feedback area.
You can place measurements, annotations, and BodyMarkers on the loaded exam screen. In addition, you can update the exam with new image scans.
Reviewing exams performed prior to the last 24 hours
Double-click an image. The exams that were performed prior to the last 24 hours and their images are displayed. The first performance date of the exam appears in the feedback area.
You can place measurements, annotations, and BodyMarkers on the loaded exam screen. However, you cannot update the exam with new image scans.

Exam Scroll
The image of the previous or next page is displayed on the screen. Access by rotating the Page dial-button on the touch screen. Rotating the dial-button to the left displays the previous page image, while rotating it to the right displays the next page image.
However, this button may not be used when the number of exams saved is less than the number of images that appear on a page under the current layout.

Closing Exam Review
Tap Close on the touch screen. To close all exams in the list, click the Close All button.

Pan
You can use the Set button to move the image or scan within the selected image layout.

Reset
Reset the selected image or scan to the initially saved screen.

Select All
All images in the current exam will be selected.

Unselect All
All images in the selected exam will be deselected.

Favorite
After selecting an image in the Exam SonoView Mode, you'll be able to use the Favorite button on the touch screen.
When you select a Favorite, it will be displayed as ★ in the top right-hand corner of the thumbnail image.

Delete
Delete the selected image.

Zoom
Use by rotating the Zoom dial-button on the touch screen. You can adjust the selected images between 3 and 200. This appears when an image is selected.

Page
Run by rotating the Page dial-button on the touch screen. Move images of the selected exam.
Layouts
You can adjust the number of images displayed on the screen. Use the touch screen button that represents the layout. Up to 9 images (3×3) can be compared at the same time.
The numbers shown in the layout section indicate the column and row of where an image will be displayed on the screen. You can change the numbers shown in the layout section to customize the layout.
Displaying in Full Screen
Place the cursor on the image and double click the Set button to display the thumbnail loading screen in Scan Mode.
Reviewing 3D Images
If the saved image is 3D, the 3D indicator appears at the bottom of the image. Double clicking the screen will start 3D View for image review.
Cine play
Tap the Cine Play button on the touch screen if the saved image is Cine. You can use Play, Pause, and Stop to navigate the image forwards and backwards.
Review – Compare Mode
Press Compare in the upper left corner of the screen. In Compare Mode, you can select the images related to an exam to compare and review them. The image that is selected on the screen will be highlighted with a blue border.

Compare Mode Screen
As in Exam Mode, exams for each ID appear on the screen. In addition, images for the selected exam are displayed in thumbnail format.
![SonoView Exam Compare Exam List Current Exam 20160618-112705AM [2016-06-10 (7)] Close Close All Exit Image 3](/content/2026/06/1369957/images/24b74f729eeceba8b937c41dcb40e51d55fbf3854e8118a527a300ad81720d4c.jpg)
[Figure 9.10 Compare Mode]

[Figure 9.11 Compare Mode - Touch Screen]
All functions except *Page and the details below are the same as those in Exam Mode.

Select an Image
Use the trackball and the Set button to select an image from the thumbnail list. The selected image is highlighted in blue in the list. Select a location on the screen where the image will be displayed and then the selected image will appear.

NOTE: Only scanned data can be loaded into the image area.

Thumbnail Scroll
Images of the previous or next pages are displayed in the thumbnail list. Select the desired image by scrolling with the trackball or using the Thumb Scroll dial-button on the touch screen. Rotating the dial-button to the left displays the previous page's images in the thumbnail list. Rotating it to the right displays the next page's images in the thumbnail list.
Image Scroll
The image of the previous or next page is displayed on the screen. Select the desired image by using the trackball or using the Image Scroll dial-button on the touch screen.
Mobile Export

[Figure 9.12 Exam Mode – Mobile Export]
① Selected Files: It shows types of selected images.
When selecting SR checkbox, it will only be transferred when the file to be transferred contains an SR.
② Connection Status: The phone number of the connected smartphone (if connected via Wi-Fi) will be displayed. ③ Stand by: Displays data conversion and transfer status. ④ QR Code: Use the smartphone's QR code scanner to download the Mobile Export application. ⑤ Send: Send the selected images to the smartphone. ⑥ Close: Close Mobile Export.

NOTE:
▶ Mobile Export is an optional feature of this product. Images that can be transferred through the Mobile Export are available only for acquired still image or recorded video. Enter the clinic name in the system setting. The clinic's name will be displayed at the bottom of the application screen when you run the Mobile Export application.

Preparation
To use Mobile Export, the system and your smartphone need to be connected via Wi-Fi or a USB data cable. You can use the USB Wi-Fi Dongle Key to set up Wi-Fi on the system.

Printing Images
- Press the Print button. The Print window will be displayed.
- Select the type, printing direction, and options for the printer.
- Click the Print button to print the image. Click Close to cancel.

Canceling a Print Job in Progress
To cancel a print job in progress, use Setup > Peripherals > Peripherals > Printer > Printer Setting. Select the printer with the print job you want to cancel, and press the Clear Print Setting button. For more information, refer to 'Chapter 3. Utilities' in this manual.

NOTE: If Thermal Printer is selected, Include Header and Include Page Numbers will be inactive, and headers and page numbers will not be printed on the pages.

[Figure 9.13 Image Print]

Transferring Images via DICOM
Transfer the selected image and measurement data via DICOM. The icon is enabled only when an image is selected. The icon is disabled in a system where DICOM is not enabled.
- Select an image and click the Send button on the screen. The DICOM Storage window will appear.
- Press the Send button to transfer the selected image to the DICOM server. Click Close to cancel.

[Figure 9.14 DICOM Storage]

Transferring Images
- Press the Export button. The Export window will open.
- Specify the various parameters, such as directory, drive, filename, and file format.
Select the Create Patient Folder(s) checkbox to automatically create folders using patient information. If the Hide Patient Information checkbox has been selected, patient information is not displayed. You can select Low, Mid, or High for Image Quality.
- Click the Export button to start a transfer. Click the Close button to cancel.

[Figure 9.15 Image Export]

Exiting SonoView
Click the Exit button on the screen. Alternatively, you can press the SonoView button or the Exit button on the control panel to exit SonoView Mode.
DICOM Q/R
At least one server should be registered in the DICOM Q/R settings.

NOTE: Before using this feature, make sure that DICOM Q/R is properly configured. To configure DICOM Q/R, please refer to 'Chapter 3. Utilities' and configure Setup > Connectivity > DICOM functions before use. There is no limit on the number of DICOM Q/R servers you can register.

Run DICOM Q/R
Retrieve an image from the selected server by following this process:
- After entering the ID in the Patient window, register or start an exam.
In the SonoView mode, enter an Exam List and select the combobox to Q/R.
- Enter the Patient ID, Accession Number, or Last Name as a search term.
Alternatively, search by selecting the modality you wish to find. You may select MR, CT, Mammo, US, DR, or PET for the modality. You may search by designating a period. Select Period. Select either Today, Last Week, Last Month, or Manual in the combo box. If Manual is selected, you do not have to select date separately.
- Press the Search button to search.
- Patient ID, Patient Name, Modality, Exam Date, Accession Number, Description, and Series Count will appear on the Study List search result.
- Up to three studies can be selected from the Study List of the same Patient ID.
- Select a study on the Study List to search within a series. Modality, Exam Date, Description, Series No., Image Count, and Status will appear on the Series List search result.
- After searching for Series, select the Series you wish to retrieve. You may select or deselect up to 10 series by using the Select All checkbox. Click Retrieve. If 10 series are selected, no additional series can be selected.

[Figure 9.16 DICOM Q/R]

Retrieval Data Manager
- Press Retrieval Data Manager to display the Retrieved Data Manager window.
- In the Retrieved Data Manager window, select the series by using the checkbox.

NOTE: If you have not started the diagnosis, select the series in the Data Manager window by using the checkbox and press Delete to delete. Locked exams cannot be deleted by using the checkbox.
- Press the SonoView button to view the selected exam image in the SonoView window.
- This function can be used to compare data by pressing Compare.
- Checking Show Only Selected will show the selected exams only.
- Tap Close on the screen.

[Figure 9.17 Retrieval Data Manager]