SDS200 - Measuring equipment Owon - Free user manual and instructions
Find the device manual for free SDS200 Owon in PDF.
Pick your language and provide your email: we'll send you a specifically translated version.
| Product Type | Digital Storage Oscilloscope |
| Bandwidth | 200 MHz |
| Sample Rate (Real-Time) | 1 GSa/s |
| Channels | 2 |
| Display | 7-inch TFT LCD, 800x480 pixels |
| Vertical Resolution | 8 bits |
| Input Impedance | 1 MΩ ±2% // 15 pF ±3 pF |
| Timebase Range | 2 ns/div to 100 s/div |
| Trigger Types | Edge, Pulse, Video, Slope |
| Math Functions | +, -, ×, ÷, FFT |
| Dimensions (W×H×D) | 320 × 150 × 120 mm |
| Weight | 2.5 kg |
| Power Supply | AC 100-240 V, 50/60 Hz |
| Power Consumption | 30 W (max) |
| Storage | USB host for waveform storage |
| Connectivity | USB Device, USB Host, LAN (optional) |
| Safety Compliance | EN 61010-1, CAT I 300V |
| Cleaning | Wipe with dry cloth; no solvents |
| Spare Parts / Repair | Contact Owon service center |
| Warranty | 3 years |
Frequently Asked Questions - SDS200 Owon
How do I calibrate the SDS200?
Access the calibration menu via Utility > Calibration. Ensure the device is warmed up for 30 minutes and no signal is applied.
What probes are compatible with the SDS200?
Standard BNC probes with 1x/10x attenuation are compatible. The input impedance is 1 MΩ, so use 10x probes for high-frequency signals.
Can I use the SDS200 to measure AC mains voltage?
No, the SDS200 is rated for CAT I 300V only. Do not connect directly to mains. Use a differential probe if necessary.
How do I save waveforms to a USB drive?
Insert a USB drive, press Save, choose Waveform format, and select the destination. The device supports FAT32 formatted drives.
What is the maximum input voltage?
The maximum input voltage is 300 V peak (CAT I) when using a 10x probe. Direct input limit is 40 V peak.
How do I update the firmware?
Download the latest firmware from Owon's website, copy to a USB drive, go to Utility > Update Firmware, and follow on-screen instructions.
Why is the waveform noisy?
Check probe ground connection, ensure proper shielding, and use bandwidth limit (20 MHz) via Channel menu to reduce high-frequency noise.
How do I perform an FFT analysis?
Press Math, select FFT, choose source channel, and set vertical scale and position. Use time/div to adjust frequency resolution.
What does the 'Auto Set' button do?
The Auto Set button automatically adjusts vertical, horizontal, and trigger settings to display a stable waveform for the input signal.
How do I measure frequency or period?
Press Measure, select Frequency or Period, and choose the source channel. The measurement is displayed at the bottom of the screen.