MTD - Sphygmomanometer MEDISANA - Free user manual and instructions
Find the device manual for free MTD MEDISANA in PDF.
| Product type | Automatic blood pressure monitor for the arm |
| Brand | Medisana |
| Model | MTD |
| Dimensions | Approximately 15 x 10 x 8 cm |
| Weight | Approximately 300 g |
| Power supply | 4 AA batteries (1.5 V) or mains adapter (not included) |
| Measurement method | Oscillometric |
| Measurement range | Pressure: 0-300 mmHg, Pulse: 40-200 beats/min |
| Accuracy | Pressure: ±3 mmHg, Pulse: ±5% |
| Memory | 2 x 120 measurements with date and time |
| Display | Large LCD screen with risk indicator |
| Main functions | Automatic measurement, arrhythmia detection, calculation of the average of the last 3 measurements, low battery indicator, automatic shut-off |
| Care and cleaning | Clean with a soft, dry cloth, do not immerse in water |
| Safety | Compliant with European directive 93/42/EEC for medical devices, clinically validated |
| Spare parts | Cuff (size chart available), batteries, mains adapter (optional) |
| Repairability | Device not user-repairable, contact Medisana customer service |
| General information | 2-year warranty, domestic use |
Frequently Asked Questions - MTD MEDISANA
User questions about MTD MEDISANA
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Sphygmomanometer in PDF format for free! Find your manual MTD - MEDISANA and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. MTD by MEDISANA.
USER MANUAL MTD MEDISANA
Please read carefully!
1 Safety Information 14
2 Useful Information 16
3 Getting started 18
4 Operating 21
5 Memory 22
6 Miscellaneous 23
7 Warranty 26
FR Mode d'emploi
Please fold out this page and leave it folded out for quick reference.
1 Socket for mains adaptor
2 Touch screen button MEM (green)
3 Touch screen button START (red)
4 ON/OFF switch
5 Display
6 Blood pressure indicator
7 Connector for the hose
8 Air hose
9 Upper arm cuff
10 Batteries low
11 Deflate to measure symbol
12 Inflate to measure symbol
13 Pulse symbol
14 Radio clock symbol
15 Display of systolic pressure
16 Display of diastolic pressure
17 Display of pulse rate
18 Memory location number
19 Time
20 Date
FR
(RCC = Radio Controlled Clock)
MEDISANA Service-Center
Feuerbach KG
Corneliusstraße 75
40215 Düsseldorf
Tel.: 0211 - 38 10 07
Read the instruction manual carefully before using this device, especially the safety instructions, and keep the instruction manual for future use. Should you give this device to another person, it is vital that you also pass on these instructions for use.
Explanation of symbols

This instruction manual belongs to this device. It contains important information about starting up and operation. Read the instruction manual thoroughly. Non-observance of these instructions can result in serious injury or damage to the device.

WARNING These warning notes must be observed to prevent any injury to the user.

CAUTION These notes must be observed to prevent any damage to the device.

NOTE These notes give you useful additional information on the installation or operation.

Classification: Ila, type B applied part

Lot number

Manufacturer

Date of manufacture

Safety Information

Do not take any therapeutic measures based on your own measurements! Never change the quantity of medication prescribed by your doctor!
1.1 Tips for your well being
- Irregularity of pulse or arrhythmia can lead to difficulties in recording a correct reading when measurements are taken using oscillometric blood pressure devices.
- If you should suffer from conditions such as arterial occlusive disease, please consult your doctor before using the unit.
- The unit may not be used to check the heart rate of patients with a pacemaker.
- Pregnant patients should take their own personal state of health into consideration and take all necessary precautions before using the unit. If in doubt, consult your doctor.
- Should you experience discomfort or other complaints during measurement, for example pain in the upper arm, please take the following action: Press the START button ③ to deflate the cuff immediately. Loosen the cuff and remove it from your arm. Please contact either your dealer or MEDISANA directly.
• Patients with a weak pulse wave should raise their arm and open and close their hand around 10 times before taking a reading. This exercise optimizes the pulse wave and the measuring process.
1.2 Important considerations
- The unit is only to be used for the specific purpose described in this instruction manual.
- Any misuse will void the warranty.
- The unit is intended for domestic use only.
- On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff immediately.
- Do not use the unit near to equipment emitting a strong electric or magnetic field, such as radios. This may impair the correct functioning of the unit.
- This device is not designed to be used by persons (including children) with limited physical, sensory or mental abilities, or by persons with insufficient experience and/or knowledge, unless under observation by a person responsible for their safety, or unless they have been instructed in the use of the device.
• Children must be supervised to ensure that they do not play with the device. - If a fault occurs, do not try to repair the unit yourself. Attempts to do so will void the warranty. Refer all servicing to authorized service personnel.
- Protect the unit from moisture. Should moisture enter the unit, remove the batteries and stop using it immediately. In this case, please contact either your dealer or MEDISANA directly. You can find details of how to contact us on the address page.
Thank you very much
Thank you for your confidence in us and congratulations on your purchase! With the MTD upper arm blood pressure monitor you have purchased a MEDISANA quality product. To ensure the best results and long-term satisfaction with your MEDISANA MTD blood pressure monitor, we recommend that you read the following operating and maintenance instructions carefully.
2.1 Items supplied and packaging
Please check first of all that the device is complete and is not damaged in any way. In case of doubt, do not use the appliance and contact your dealer or your service centre. The following parts are included:
• 1 MEDISANA Blood pressure monitor MTD
• 1 Cuff with air hose
• 4 Batteries (AA type, LR 6) 1,5V
- 1 Storage bag
• 1 Holder with adhesive pad
• 1 Instruction manual
The packaging can be reused or recycled. Please dispose properly of any packaging material no longer required. If you notice any transport damage during unpacking, please contact your dealer without delay.

WARNING
Please ensure that the polythene packing is kept away from the reach of children! Risk of suffocation!
2.2 What is blood pressure?
Blood pressure is the pressure created in the blood vessels by each heart beat. When the heart contracts (= systole) and pumps blood into the arteries, it brings about a rise in pressure. The highest value is known as the systolic pressure and is the first value to be measured. When the cardiac muscle relaxes to take in fresh blood, the pressure in the arteries also falls. Once the arteries are relaxed, a second reading is taken – the diastolic pressure.
2.3 How is blood pressure measured?
The MTD is a blood pressure unit which is used to measure blood pressure at the upper arm. The measurement is carried out by a microprocessor, which, via a pressure sensor, measures the vibrations resulting from the inflation and deflation of the cuff over the artery.
2.4 Why should you take your blood pressure at home?
MEDISANA has had many years of experience in the field of blood pressure measurement. The high precision of MEDISANA devices has been confirmed by clinical tests, carried out according to international standards. One strong argument in favour of taking blood pressure at home is the fact that the readings are taken in familiar surroundings under relaxed conditions. The most important reading is the so-called „baseline value“, the one taken before breakfast, directly after getting up in the morning. As far as it is possible, you should always try to take your blood pressure at the same time and under the same conditions each day.
This ensures comparable results and helps in detecting the first signs of high blood pressure. If high blood pressure remains undetected, it increases the risk of further cardiovascular diseases.

THEREFORE WE RECOMMEND:
Take a blood pressure reading daily and on a regular basis, even if you are experiencing no apparent complaints.
2.5
Blood pressure classification
The following table shows the classification of high and low blood pressure, without taking age into consideration. This blood pressure assessment scale corresponds to the guidelines of the World Health Organisation (WHO).
Low blood pressure Normal blood pressure
(green display area 6)
systolic below <100 systolic 100 - 139
diastolic below <60 diastolic 60 - 89
Types of high blood pressure
mild hypertension medium hypertension serious hypertension
(yellow (orange (red
| display area 6) display area 6) display area 6) | |
| systolic 140 – 159 systolic 160 – 179 systolic | >180 |
| diastolic 90 – 99 diastolic 100 – 109 diastolic | >110 |

WARNING
Blood pressure that is too low represents just as great a health risk as blood pressure that is too high! Fits of dizziness may lead to dangerous situations arising (e.g. on stairs or in traffic)!
2.6
Fluctuations in blood pressure
There are many factors that can influence blood pressure. Readings can be affected seriously by stress, fear, heavy physical exertion or the time of day at which the measurements are taken. Your personal blood pressure values are subject to considerable fluctuation during the course of the day and the year. These fluctuations are particularly pronounced in the case of patients suffering from hypertension. Blood pressure is usually at its highest during physical exertion and at its lowest at night when you are asleep.
2.7
Influencing and evaluating readings
- Measure your blood pressure several times, then record and compare the results. Do not draw any conclusions from a single reading.
- Your blood pressure readings should always be evaluated by a doctor who is also familiar with your personal medical history. When using the unit regularly and recording the values for your doctor, you should visit the doctor from time to time to keep him updated.
- When taking readings, remember that the daily values are influenced by several factors. Smoking, consumption of alcohol, drugs and physical exertion influence the measured values in various ways.
• Measure your blood pressure before meals.
- Before taking readings, allow yourself at least five minutes rest.
- If the systolic and diastolic readings seem unusual (too high or too low) on several occasions, despite correct use of the unit, please inform your doctor. This also applies to the rare occasions when an irregular or very weak pulse prevents you from taking readings.
3 Getting Started
3.1 Inserting / removing batteries
Insertion: Open the battery compartment on the underside of the unit by exerting gentle pressure, and insert the four batteries included with the unit (alkaline batteries, AA type, LR 6). Remember to insert the batteries as indicated in the diagram in the battery compartment. Replace the lid, pressing it until it snaps into place.
Removal: When the battery symbol 📄 appears in the display, please replace the batteries with new ones. After batteries be inserted, the display will display the defaulted date & time. Set the date and time as described in 3.3 / 3.4. Please note that the date and time will have to be reset each time new batteries are inserted. The previous measuring results are saved.

WARNING BATTERY SAFETY INFORMATION
- Do not disassemble batteries!
- Remove discharged batteries from the device immediately!
- Increased risk of leakage, avoid contact with skin, eyes and mucous membranes! If battery acid comes in contact with any of these parts, rinse the affected area with copious amounts of fresh water and seek medical attention immediately!
- If a battery has been swallowed seek medical attention immediately!
- Replace all of the batteries simultaneously!
- Only replace with batteries of the same type, never use different types of batteries together or used batteries with new ones!
- Insert the batteries correctly, observing the polarity!
- Remove the batteries from the device if it is not going to be used for an extended period!
- Keep batteries out of children's reach!
- Do not attempt to recharge these batteries! There is a danger of explosion!
- Do not short circuit! There is a danger of explosion!
- Do not throw into a fire! There is a danger of explosion!
- Do not throw used batteries into the household refuse; put them in a hazardous waste container or take them to a battery collection point, at the shop where they were purchased!
3.2 Using a mains adaptor
Alternatively, the unit may also be operated with a special power adapter (MEDISANA item no. 51125), which is plugged into the connector ① provided on the right-hand side of the unit. In this case the batteries remain in the unit. The batteries are disconnected mechanically when the connector is inserted on the right-hand side of the blood pressure monitor. It is necessary, therefore, to plug the mains adaptor into the mains socket first of all and then to connect it to the blood pressure monitor. If the blood pressure monitor is no longer being used, first remove the connector from the blood pressure monitor and then pull the mains adaptor out of the mains socket. Doing this means that the date and time will not have to be reset each time.
3.3 Setting the reception for the time signal (RCC)
(RCC = Radio Controlled Clock)
a. Initial setting:
As soon as you have inserted the batteries, the device automatically switched to the radio clock mode. While searching for the time signal, the radio clock signal flashes. As soon as the clock receives an adequate signal for setting, the time and date are set automatically. When the setting process is finished, the device switches to stand-by mode, in which time and date are displayed permanently. If no signal is received, the device goes directly to stand-by mode without changing the preset time and date.
By pressing the Start button ③ the initial setting reception of the time signal can be switched off. The device goes directly to stand-by mode without changing the preset time and date.
b. Quick radio clock setting
When the device is in stand-by mode and you press the START button ③ and MEM button ② simultaneously for approx. 1 second, you will go to the quick setting mode for the radio clock. The radio clock symbol flashes. When the time signal is received, the time and date are set automatically. If no signal is received, the device switches to the manual time setting mode.
The quick setting can be cancelled by pressing the START button ③. The device then switched to the manual time setting mode.
c. regular signal reception
The device automatically receives the radio signal to update the time setting twice a day, at 2.05 a.m. and 2.05 p.m.
By pressing the START button ③ the regular reception of the time signal can be switched off.
3.4 Setting the date and time
The time and date are normally set automatically by radio along with the time signal. When no time signal is received, or it is too weak, the clock can also be set manually.
- The device is in stand-by mode. Press the START button ③ and MEM button ② simultaneously to go to the quick setting mode for the radio clock. Now press the START button ③ to go to the manual time setting mode.
- The digit for entering the month flashes first. By pressing the START button ③ repeatedly, the digits for the day, hour and minute will flash in succession. While each input position is flashing, press the MEM button ② to enter the corresponding digit. By holding the MEM button ② the setting process can be speeded up.
- Exit the time setting mode by pressing the START button ③ while the minute position is flashing. It will switch to stand-by mode.
After changing the batteries the clock will need to be reset, either automatically by radio signal or manually by time setting mode.
3.5 Adjusting the voice function
The voice function can be adjusted with the device switched on (ON/OFF switch ④ in I position), i.e. with just the date and time displayed. Press and hold down the MEM button ② until the selected language appears on the display (e.g. L1).
| L0 = Language function switched off | L4 = Dutch |
| L1 = German | L5 = French |
| L2 = English | L6 = Italian |
| L3 = Spanish | L7 = Portuguese |
The device speaks the language setting, and changes the display and the announcement to the next language if the MEM button ② is held down. The selected language can be changed by releasing the MEM button ② as soon as the language that you require is displayed or spoken. The selected volume, which is indicated by flashing bars, can be modified by pressing the MEM button ② again and holding it down until the required volume is indicated – the higher the bar, the louder the volume. Release the MEM button ② to confirm your selection. The bar stops flashing and the display switches back to the time display.
3.6 Holder
Using the holder included with the unit you will store the arm cuff and the air hose together with the device. First attach the adhesive pad to the rear side of the unit, then fix the holder on it firmly.
3.7 ON/OFF switch I/O 20
Your device has a touch screen as an equipment. Be careful not to touch the display by mistake. Keep the ON/OFF switch 4 in 0 position when the device is not in use to avoid unintentional adjustments.
4.1 Fitting the cuff

-
Push the end piece of the air tube into the hole on the unit ⑦ prior to use. Ensure that the hose is not damaged and is securely attached. The blood pressure monitor is intended for use as a table-top unit.
-
Slide the open end of the cuff through the metal bracket so that the Velcro fastener is on the outside.
-
Open the cuff ⑨ so as to create a cylindrical shape.
-
Slide the cuff over your left upper arm. Position the air hose ⑧ in the middle of your arm in line with your middle finger (a).
-
Place your arm on the table with the palm uppermost. The lower edge of the cuff should be 1 - 2 cm above the crease of the elbow (b). Pull the cuff tight (c) and close the Velcro fastener.
-
Avoid creating a space between arm and cuff. It may falsify the measured result.
-
Measure the pulse on your bare arm. If wearing tight clothing, ensure that the circulation in the arm is not impaired when rolling up the sleeve.
-
Only position the cuff on the right arm if it cannot be used on the left arm. Always carry out measurements on the same arm.
4.2 Correct position for taking readings
The Correct Measuring Position when sitting
- Carry out the measurement whilst seated.
- Relax your arm and place it comfortably, on a table for instance.
- Ensure that the cuff is at heart height.
- Keep still during the measurement: do not move and do not talk as this may alter the measurements.

Correct position for taking measurements lying down
- In special cases the measurement can also be taken lying down.
- Lie on your back and relax for at least 5 minutes and also relax your arm.

- Ensure that the cuff is at heart height. Differing heights affect the accuracy of the measurement.
- Keep still during the measurement: do not move and do not talk as this may alter the measurements.
4.3 Taking blood pressure readings
- Switch the unit on by using first the ON/OFF switch ④ (I position) and then pressing the START button ③.
- If the START button 3 is pressed, a beep is heard and all display characters are shown (display test). This test can be used to check that the display is indicating properly and in full. The display shows the values from the previous measurement. Information will be given in the language set if the language function is switched on. If no measurement has been taken, zeroes will appear next to the date and time.
- The unit is ready for measurement, the number 0 is displayed, the start of measurement will be announced if the language function is switched on. It automatically inflates the cuff slowly in order to measure your blood pressure. The pressure on the display changes continuously during the time it continues to inflate.
- The unit inflates the cuff until sufficient pressure has built up for a measurement. Then the unit slowly releases air from the cuff and carries out the measurement.
- When the measurement is finished, the cuff is deflated completely. The systolic and diastolic blood pressure and the pulse value with the pulse symbol ♥ appear on the display ⑤. The blood pressure indicator ⑥ flashes next to the relevant coloured bar depending on the WHO blood pressure classification (see page 17). The measured results will be announced in the language set if the language function is switched on.
- The measured values are stored automatically. Up to 60 measured values with date and time may be stored in the memory.
4.4 Switching off the unit
The unit has an automatic switch-off function. The unit switches off automatically after a minute if no buttons are pressed. The unit may also be switched off using the START button ③. When the device is not in use switch the ON/OFF switch ④ in 0 position.
The time and the date are constantly shown in the display.
5 Memory
5.1 Displaying stored results
The memory of your blood pressure monitor will hold up to 60 measurements (blood pressure and pulse with time and date). Press the MEM button 2 to call up the saved measured values. The saved data will appear on the display. Initially the number of memory locations occupied will appear then the values of the most recently saved measurement including the flashing blood pressure indicator 6 will be displayed.
If the voice function has been activated, the device speaks the memory location number, together with the saved blood pressure and pulse values and the blood pressure assessment as per the scale in the World Health Organisation guidelines (see pg. 17).
Repeatedly pressing the MEM button ② displays the respective values measured previously. The unit will switch off if the oldest memory has been reached.
In memory recall mode, the unit will switch off automatically after approx. 120 seconds if no button is pressed. Memory recall mode can be exited at any time by pressing the START button ③, which will also switch the unit off.
If all 60 memory locations in the memory are occupied, a new measured value is saved whilst the oldest value is deleted.
5.2
Deleting memory
If you are in memory recall mode, press and hold the MEM button ② for approximately 8 seconds. All measured data stored will be deleted automatically after three beeps. Only noughts will be shown on the display instead of the measured values.
6 Miscellaneous
6.1
Troubleshooting
The battery change symbol 📄 appears in the display
- The batteries are too low or empty. Replace all four batteries with new 1.5 V alkali AA batteries.
Erratic measurements being displayed
- Cuff incorrectly fitted or not tight enough. Fit cuff correctly and tighten.
- You have changed position or the cuff has moved during the measurement. Keep still and do not move your arm or the cuff.
- You have an irregular heartbeat. Even if your heartbeat is only slightly irregular, repeat the measurement. The unit is not suitable for people with an extremely irregular heartbeat.
- You have spoken during the measurement, you are nervous or you are excited. Do not speak and breathe deeply two or three times in order to relax.
- You are not in the corrected seated or lying position. Please assume the correct position when taking the measurement.
- The air hose is not connected or leaking. Check the connection between the upper arm cuff and the unit and check the air hose for leaks. Repeat the measurement.
"EE" appears in the display
- The unit has not been inflated enough. Repeat the measurement
- A problem occurred during inflation or during the measurement. Repeat the measurement.
No display or display locked
- No batteries have been inserted, batteries wrongly inserted or batteries are completely empty. Insert batteries, check battery position or replace with new batteries.
- A malfunction or fault has occurred. Briefly remove the batteries and then replace them.
Please contact the service centre if the error message appears again when the measurement is repeated.
6.2 Care and maintenance
- Remove the batteries before cleaning the unit.
- Never use strong detergents or hard brushes.
- Clean the unit using a soft cloth lightly moistened with a mild soapy solution. Never use abrasive cleaning agents or alcohol.
- Do not allow any water to get into the unit. Do not use the unit again until it is completely dry.
- Remove the batteries from the unit if you do not wish to use it for an extended period of time, otherwise there is a risk of battery leakage.
- Do not expose the unit to direct sunlight, and protect it from dust and moisture.
- Only inflate the cuff when it is in position around the upper arm.
- The unit is best kept in the storage pouch included in your purchase and should preferably be stored in a clean dry place.
- Servicing and calibration (applies to commercial use):
The unit has been calibrated by the manufacturer for a period of two years' use. In the case of commercial use, the unit must be serviced and recalibrated at least once every two years. This calibration will be charged for and can be carried out by an appropriate authority or an authorised service centre - in accordance with the regulations for users of medical equipment.
6.3 Disposal

This product must not be disposed of together with domestic waste.
All users are obliged to hand in all electrical or electronic devices, regardless of whether or not they contain toxic substances, at a municipal or commercial collection point so that they can be disposed of in an environmentally acceptable manner.
Please remove the batteries before disposing of the device / unit. Do not dispose of old batteries with your household waste, but at a battery collection station at a recycling site or in a shop.
Consult your municipal authority or your dealer for information about disposal.
6.4 Directives / norms
This blood pressure monitor meets the requirements of the EU standard for non-invasive blood pressure monitors. It is certified in accordance with EC Guidelines and carries the CE symbol (conformity symbol) "CE 0297". The blood pressure monitor corresponds to European standards EN 1060-1 and EN 1060-3+A1. The specifications of EU Guideline "93/42/EEC of the Council Directive dated 14 June 1993 concerning medical devices" are met and likewise those of EU Guideline 89/336/EEC.
Intended use:
This unit is suitable for non-invasive blood pressure measurement in adults (i.e., suitable for external use).
Electromagnetic compatibility:
The device complies with the EN 60601-1-2 standard for electromagnetic compatibility. Inquire at MEDISANA for details on this measurement data.
6.5
Technical specifications
Name and model : MEDISANA blood pressure monitor MTD
Display system : Digital display
Memory slots : 60
Measuring technique : Oscillometric
Power supply : 6 V= , 4 x 1,5 V Mignon AA LR 6, Alkaline
Blood pressure measuring
range : 0 - 300 mmHg
Pulse measuring range : 30 - 180 beats / Min.
Maximum error tolerance
for static pressure : ± 3 mmHg
Maximum error tolerance
for pulse rate : ± 5 % of the reading
Pressure generation : Automatic, using pump
Deflation : Automatic
Automatic switch-off : After approx. 1 minute
Operating conditions: + 5 °C to + 40 °C,
< 85 % humidity
Storage conditions: -20°C to +55°C
< 95 % humidity
Dimensions : approx. 142 mm x 98 mm x 52 mm
Cuff size : 22 - 30 cm Adult cuff for average
upper arm circumference
Weight : approx. 310 g without batteries
Article number : 51145
EAN number : 40 15588 51145 5
Accessories : Mains adapter
Art.-No. 51125
EAN-No. 40 15588 51125 7
Cuff, large 30 - 42 cm for
adults with upper arm of large circumference
Art.-No. 51126
In accordance with our policy of continual product improvement, we reserve the right to make technical and optical changes without notice.
Warranty and repair terms
Please contact your dealer or the service centre in case of a claim under the warranty. If you have to return the unit, please enclose a copy of your receipt and state what the defect is.
The following warranty terms apply:
- The warranty period for MEDISANA products is three years from date of purchase. In case of a warranty claim, the date of purchase has to be proven by means of the sales receipt or invoice.
- Defects in material or workmanship will be removed free of charge within the warranty period.
-
Repairs under warranty do not extend the warranty period either for the unit or for the replacement parts.
-
The following is excluded under the warranty:
a. All damage which has arisen due to improper treatment, e.g. non-observance of the user instructions.
b. All damage which is due to repairs or tampering by the customer or unauthorised third parties.
c. Damage which has arisen during transport from the manufacturer to the consumer or during transport to the service centre.
d. Accessories which are subject to normal wear and tear (cuff, batteries etc.).
- Liability for direct or indirect consequential losses caused by the unit are excluded even if the damage to the unit is accepted as a warranty claim.

MEDISANA AG
Itterpark 7-9
40724 Hilden
Germany
Tel.: +49 (0) 2103 / 2007-60
Fax: +49 (0) 2103 / 2007-626
The service centre address is shown on the last page.

REMARQUE IMPORTANTE! TOUJOURS CONSERVER!
(RCC = Radio Controlled Clock)
NOTE IMPORTANT! CONSERVARE IN MANIERA SCRUPOLOSA!
(RCC = Radio Controlled Clock)
(RCC = Radio Controlled Clock)
a. Primer ajuste:
Colocar / retirar as pilhas
(RCC = Radio Controlled Clock)
3.4 Regular a data e as horas
(RCC = Radio Controlled Clock)
OBSERVERA! SPARA!
MEDISANA Service-Center
Feuerbach KG
Corneliusstraße 75
40215 Düsseldorf
Tel.: 0211 - 38 10 07
City Business Centre 41 St.
Olav's Court Surrey Quays
London SE16 2XB
Tel.: +44/207-237 88 99
Fax: +44/207-252 22 99
eMail: info@medisana.co.uk
Internet: www.medisana.co.uk
BE / FR / IT / LU / NL
MEDISANA Benelux N.V.
Euregiopark 30
6467 JE Kerkrade
Nederland
MEDISANA HEALTHCARE, S.L.