PCE Instruments

PH 26F - Measurement PCE Instruments - Free user manual and instructions

Find the device manual for free PH 26F PCE Instruments in PDF.

📄 15 pages English EN Download 💬 AI Question
Notice PCE Instruments PH 26F - page 5
Pick your language and provide your email: we'll send you a specifically translated version.

User questions about PH 26F PCE Instruments

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Measurement in PDF format for free! Find your manual PH 26F - PCE Instruments and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. PH 26F by PCE Instruments.

USER MANUAL PH 26F PCE Instruments

text_image QR code image containing encoded data, no visible human-readable text

User manuals in various languages (français, italiano, español, português, nederlands, türk, polski, русский, 中文) can be found by using our product search on: www.pce-instruments.com

Last change: 19 August 2021

v1.0

Contents

1 Safety notes....1
2 Specifications.... 2
3 Delivery scope.... 3
3.1 PCE-PH 26F 3
3.2 PCE-PH 28L 3
4 System description....3
4.1 Display....3
4.2 Keys.... 4
5 Measurement preparation....4
5.1 Insert batteries .... 4
5.2 Preparation of the buffer solution.... 4
5.3 Electrode control 4
5.4 Switch meter on / off.... 5
5.5 Basic settings.... 5
6 pH calibration 6
6.1 1-point calibration....7
6.2 2-point calibration....8
6.3 3-point calibration....9
7 Temperature calibration....10
8 pH measurement....11
9 Care and maintenance of the electrodes....11
10 Troubleshooting.... 11
11 Warranty....12
12 Disposal 12

1 Safety notes

Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel.

Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.

  • The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
  • The instrument may only be used if the environmental conditions (temperature, relative humidity, ...) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
  • Do not expose the device to shocks or strong vibrations.
  • The case should only be opened by qualified PCE Instruments personnel.
  • Never use the instrument when your hands are wet.
  • You must not make any technical changes to the device.
  • The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
  • The device must only be used with accessories from PCE Instruments or equivalent.
  • Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
  • Do not use the instrument in explosive atmospheres.
  • The measurement range as stated in the specifications must not be exceeded under any circumstances.
  • Non-observance of the safety notes can cause damage to the device and injuries to the user.

We do not assume liability for printing errors or any other mistakes in this manual.

We expressly point to our general guarantee terms which can be found in our general terms of business.

If you have any questions please contact PCE Instruments. The contact details can be found at the end of this manual.

2 Specifications

pH measurement
Measurement range-1.00 ... 15.00 pH
Accuracy±0.01 pH
Resolution0.01 pH
Calibration points1, 2 or 3
Buffer solutionsUSA (pH 4.01 / 7.00 / 10.01) and NIST (pH 4.01 / 6.86 / 9.18)
Automatic recognition of the buffer solutionyes
Temperature measurement
Measurement range0 ... 60 °C / 32 ... 140°F
Accuracy±1 °C
Resolution1 °C
Calibration points1
General specifications
Temperature compensation0 ... 60 °C / 32 ... 140°F
Temperature of medium0 ... 60 °C / 32 ... 140°F
Environmental conditions0 ... 60 °C / 32 ... 140°F; <80 % RH
Batteries2 x 1.5 V AAA
Weight100 g / <1 lb
Total dimensions
PCE-PH 26F185 mm x ∅ 40 mm / 7.3 x 1.6 in
PCE-PH 28L240 mm x ∅ 40 mm / 9.44 x 1.6 in
Electrode dimensions
PCE-PH 26F40 mm x ∅ 26 mm / 1.6 in x ∅ 1.02 in
PCE-PH 28L75 mm x ∅ 14 mm / 2.95 in x ∅ 0.55 in

3 Delivery scope

3.1 PCE-PH 26F

1 x pH meter PCE-PH 26F

1 x buffer 4.01 pH powder for 250 ml solution

1 x buffer 7.00 pH powder for 250 ml solution

1 x buffer 10.01 pH powder for 250 ml solution

2 x batteries 1.5 V AAA

1 x storage box

1 x user manual

3.2 PCE-PH 28L

1 x pH meter PCE-PH 28L

1 x buffer 4.01 pH powder for 250 ml solution

1 x buffer 7.00 pH powder for 250 ml solution

1 x buffer 10.01 pH powder for 250 ml solution

3 x container for calibration solution

2 x batteries 1.5 V AAA

1 x carrying case

1 x user manual

4 System description

4.1 Display

PCE Instruments PH 26F - Display - 1

text_image MEASCALSETUP HOLD pH 8.8.8.8 % ATC 888.8 °C
IconDescription
MEASMeasuring mode
CALCalibration mode
SETUPSetup mode
HOLDMeasured value is stable
ATCAutomatic temperature compensation enabled
Battery full (the icon disappears when the battery is flat)

4.2 Keys

KeyFunction
MEAS / Hold- Switches the meter ON/OFF- Freezes the reading. Press the key again to continue measurement.- Ends the calibration or adjustment and returns to measuring mode
CAL- Press the key to start the calibration.- Press and hold the key to open the setup menu.- In setting mode, press the key to select the default option.
EnterConfirms the entries

5 Measurement preparation

5.1 Insert batteries

PCE Instruments PH 26F - Insert batteries - 1

text_image 1 2 3
  1. Unscrew the union nut of the electrode and pull the electrode off the measuring device, observing correct axial alignment.

  2. Insert two AAA batteries into the battery compartment. Observe correct polarity.

  3. Align the guide groove with the electrode and carefully push the electrode into the meter. Screw the union nut of the electrode tight.

5.2 Preparation of the buffer solution

Prepare 3 sealable containers with at least 250ml content. Open the bags with the buffer powder and put the contents of one bag into each container. Fill each container with 250ml of distilled water and stir the liquid until the powder has completely dissolved.

5.3 Electrode control

Remove the protective cap from the electrode and check for damage.

Damaged electrodes must be replaced.

Rinse off any salt crystals that may have settled from the storage solution under running water.

The salt crystals develop naturally and are not a deficiency.

If the electrode has become dry, place it in a 3M KCL solution for at least 15 minutes. (storage solution for pH electrodes).

5.4 Switch meter on / off

To switch on the meter, press and hold the "MEAS" key until the display switches on.

Press and hold the "MEAS" key for 5 seconds to switch it off.

The auto-off function switches the unit off after 8 minutes without a key being pressed. The function can be enabled or disabled in the basic settings.

PCE Instruments PH 26F - Switch meter on / off - 1

text_image MEAG pH 7.00 ATC 25.0 °C Waterproof Call Enter Meas hold Power On

PCE Instruments PH 26F - Switch meter on / off - 2

text_image Waterproof Call Enter Moss Hold 5 sec. Power Off

5.5 Basic settings

Press and hold the "CAL" key for 3 seconds to call up the basic settings.

Press the "CAL" key to call up the possible options of the adjustable items.

Press the "ENTER" key to confirm the selection and jump to the next item.

PCE Instruments PH 26F - Basic settings - 1

text_image 1 USA BUF Option Cal Waterwater Cal Control Minor Hold

PCE Instruments PH 26F - Basic settings - 2

text_image 2 Cal N 15t BUF Waterproof Cal Turb/ Wise wolt

PCE Instruments PH 26F - Basic settings - 3

text_image 3 CAL Wakereproof Enter CAL Enter Mean Install

After confirming all selected items, the meter returns to measuring mode. By pressing the "MEAS" key, you can return directly to measuring mode.

MENU ITEMDESCRIPTIONOPTIONSFACTORY SETTING
BUFSelection of the calibration pointsUSAUSA (pH4.01/7.00/10.01)USA
NISTNIST (pH4.01/6.86/9.18)
CALNumber of calibration points11 point2 points
22 points
33 points
UNITSelection of the temperature unit°CCelsius°C
°FFahrenheit
°CTemperature calibrationCAL------
HOLDEnables the hold function when the measured value is stableYESOnOff
NOOff
OFFThe auto-off function switches the unit off after 8 minutes without any key being pressedYESOnOn
NOOff
-5tReset to factory settingsYESYesNo
NONo

After resetting the meter to factory settings, calibration is mandatory.

6 pH calibration

The devices allow a calibration of up to 3 points. To achieve high measurement accuracy, we recommend at least a 2-point calibration. The meter automatically recognises and calibrates to the following standard buffer values.

USA standard buffer pH 4.01, 7.00, 10.01

NIST standard buffer pH 4.01, 6.86, 9.18

The 1-point calibration can only be performed with pH 7.00 or pH 6.86, otherwise the calibration will not be accepted.

The meter must be calibrated before first use.

Regular calibration is recommended to ensure measurement accuracy. DO NOT reuse the pH buffer solution after calibration as impurities in the solution will affect the calibration and eventually the accuracy of the measurement.

6.1 1-point calibration

PCE Instruments PH 26F - 1-point calibration - 1

text_image Cal Waterproof pH 7.00 CAL 1 Cal Enter Meas Hold 7.00 Enter Cal 7.00 ATC 25.0℃ Waterproof Cal Enter Meas Hold Cal End Waterproof Cal Enter Meas Hold

PCE Instruments PH 26F - 1-point calibration - 2

6.1.1 The 1-point calibration must be selected in the basic settings.
6.1.2 Rinse the electrode with distilled water. Press the "CAL" key.

Depending on the setting, the display shows "pH 7.00/CAL1" or "pH 6.86/CAL1".

6.1.3 Immerse the electrode in the pH 7.00 (or pH 6.86) buffer solution with the end of the electrode completely immersed in the calibration solution. Gently stir the solution with the electrode to obtain a homogeneous solution.

6.1.4 Press the "ENTER" key. The meter starts the calibration.

6.1.5 After the measured value has stabilised, the display shows "End". The calibration is completed.

6.2 2-point calibration

PCE Instruments PH 26F - 2-point calibration - 1

text_image CAL pH CAL2 Waterproof Cal Enter Meas Hold 4.01 CAL pH 4.01 ATC 25.0℃ Waterproof Cal Enter Meas Hold Enter CAL 99 % SLOP Waterproof Cal Enter Meas Hold

6.2.1 The 2-point calibration must be selected in the basic settings.
6.2.2. Carry out the calibration of the first calibration point as described under 6.1.2. ... 6.1.4.
The display shows "CAL 2".

6.2.3 Rinse the electrode with distilled water. Immerse the electrode in the pH 4.01 or pH 10.01 (pH 9.18) buffer solution with the end of the electrode completely immersed in the calibration solution. Carefully stir the solution with the electrode to obtain a homogeneous solution.
6.2.4 Press the "ENTER" key.

The meter automatically detects the selected calibration liquid and starts the calibration.

6.2.5 After the measured value has stabilised, the display shows the slope of the electrode in % and then displays "End".

The 2-point calibration is completed.

6.3 3-point calibration

PCE Instruments PH 26F - 3-point calibration - 1

text_image 4.01 CAL2 Waterproof Cal Enter Meas HOLD 4.01 Enter CAL PH 4.01 ATC 25.0℃ Waterproof Cal Enter Meas HOLD 10.01 Waterproof Cal Enter Meas HOLD 10.01

6.3.1 The 3-point calibration must be selected in the basic settings.
6.3.2. Carry out the calibration of the first calibration point as described under 6.1.2. ... 6.1.4.
The display shows "pH 4.01" and "CAL 2".

6.3.3 Rinse the electrode with distilled water Immerse the electrode in the pH 4.01 buffer solution with the end of the electrode completely immersed in the calibration solution. Stir the solution carefully with the electrode to obtain a homogeneous solution.

6.4.4 Press the "ENTER" key. The meter starts the calibration.

6.3.5 After the measured value has stabilised, the display shows the slope of the electrode in %, followed by "pH 10.01" (pH 9.18) and "Cal 3".

6.3.6 Rinse the electrode with distilled water Immerse the electrode in the pH 10.01 (pH 9.18) buffer solution with the end of the electrode completely immersed in the calibration solution. Gently stir the solution with the electrode to create a homogeneous solution.

PCE Instruments PH 26F - 3-point calibration - 2

text_image PCE BRD CAL pH 10.0 °C ATIC 25.0 °C Waterproof Enter Cal Enter Mean Field

6.3.7. press the "ENTER" key.

The meter starts the calibration.

6.3.8 After the measured value has stabilised, the display shows the slope of the electrode in % and then displays "End".

The 3-point calibration is completed.

Press the "MEAS" key to return to measuring mode even during calibration.

7 Temperature calibration

When deviating temperature values are displayed, you can adjust them in the corresponding menu item.

PCE Instruments PH 26F - Temperature calibration - 1

text_image Cal USR BUF Waterproof Cal Enter Move Hold SETUP CAL Waterproof Cal Enter Move Hold Enter Cal 26.0°C Waterproof Cal Enter Move Hold SETUP NO HOLD Waterproof Cal Enter Move Hold Enter

7.1. Press and hold the "CAL" key for 3 seconds to call up the basic settings.

7.2. Press the "Enter" key repeatedly until the display shows "°C" and "CAL" or "°F" and "CAL".

7.3 Press the "CAL" key.

7.4 Press the "CAL" key again until the temperature display matches the actual temperature.

7.5 Press the "ENTER" key to confirm your entry.

7.6 Press the "MEAS" key to return to measuring mode.

8 pH measurement

Check the electrode as described in chapter 5.3 and switch on the meter, see chapter 5.4.

Rinse the electrode with distilled water.

Immerse the electrode in the medium with the end of the electrode completely immersed in the medium. Move the electrode slightly back and forth to make good contact with the measuring medium.

Read the result after stabilisation.

With the automatic "Hold" function switched on, the measurement result remains on the display after stabilisation. (see basic settings)

9 Care and maintenance of the electrodes

For best results, always keep the glass-sensitive membrane moist.

Make sure that the electrode is thoroughly rinsed with distilled or deionised water after use.

If you do not use the meter for a long period of time, store the electrode in the 3M KCL solution.

(Storage solution PCE-SSOxxx)

If the meter cannot be calibrated or shows fluctuating readings for calibration standards, you must replace the electrode.

Remove and attach the electrode in the same way as described in chapter 5.1.

10 Troubleshooting

Display indicationCause of errorTroubleshooting
---Electrode too dryPlace the electrode in a 3M KCL solution for at least 15 minutes. (Storage solution)
OverrangeCheck whether the electrode is dirty or the glass body is defective.
ErrWrong buffer solutionUse the correct buffer solution.
Electrode defectiveReplace the electrode.

11 Warranty

You can read our warranty terms in our General Business Terms which you can find here: https://www.pce-instruments.com/english/terms.

12 Disposal

For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.

In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law.

For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.

If you have any questions, please contact PCE Instruments.

PCE Instruments PH 26F - Disposal - 1

text_image Pb www.pce-instruments.com

PCE Instruments PH 26F - Disposal - 2

PCE Instruments contact information

Germany France Spain

PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L.

Im Langel 4 23, rue de Strasbourg Calle Mayor, 53

D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete)

Unit 11 Southpoint Business Park

Ensign Way, Southampton

Hampshire

United Kingdom, SO31 4RF

Tel: +44 (0) 2380 98703 0

Fax: +44 (0) 2380 98703 9

info@pce-instruments.co.uk

www.pce-instruments.com/english

Italy

PCE Italia s.r.l.

Via Pesciatina 878 / B-Interno 6

55010 Loc. Gragnano Pehlivan Sok. No.6/C

Capannori (Lucca)

Italia

www.pce-instruments.com/turkish

The Netherlands

PCE Brookhuis B.V.

Institutenweg 15

7521 PH Enschede

Nederland

Telefoon: +31 (0)53 737 01 92

info@pcebenelux.nl

www.pce-instruments.com/dutch Fax: +1 (561) 320-9176

info@pce-americas.com

United States of America

PCE Americas Inc.

1201 Jupiter Park Drive, Suite 8

Jupiter / Palm Beach

33458 FL

USA

Tel: +1 (561) 320-9162

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : PCE Instruments

Model : PH 26F

Category : Measurement