Rigel Medical

Citrex H3 - Measuring device Rigel Medical - Free user manual and instructions

Find the device manual for free Citrex H3 Rigel Medical in PDF.

📄 46 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice Rigel Medical Citrex H3 - page 5
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Measuring device for flow, pressure, volume, and ventilation parameters
Brand Rigel Medical (manufactured by IMT Analytics)
Model Citrex H3
Dimensions (W x D x H) 11.4 x 7 x 6 cm
Weight 0.38 kg
Power Supply Input 100-240 VAC, 50-60 Hz via USB; 5 VDC
Battery Operating Time 4 hours
Display 1.7\" color display
Interfaces Ethernet, USB Mini-B, microSD card slot, O2 sensor interface
Flow Measurement Range ±300 sL/min (air, N2, O2/air mixtures); ±80 sL/min (N2O/O2)
Flow Accuracy ±2% of reading or ±0.1 sL/min (air); ±4% or ±0.3 sL/min (N2O/O2)
Pressure Measurement Range (channel) -50 to 150 mbar
Pressure Accuracy (channel) ±0.75% of reading or ±0.1 mbar
Barometric Pressure Range 500-1150 mbar
Barometric Accuracy ±1% of reading or ±5 mbar
Volume Measurement Range 0-10 sL
Volume Accuracy ±2% of reading or ±0.02 sL
Oxygen Measurement Range (optional) 0-100%
Oxygen Accuracy ±1% O2 (absolute)
Gas Temperature Measurement 0-50°C, accuracy ±1.75% or ±0.5°C
Supported Gas Types Air, O2, Air/O2 mixtures, N2O/O2 mixtures
Gas Standards ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPS-A, BTPD
Ventilation Parameters Measured Rate, tidal volume, minute volume, peak flow, pressures (PEEP, mean, peak), I:E ratio, inspiratory time
Calibration Interval Annually
Operating Temperature 15-40°C (59-104°F)
Storage Temperature -10 to 60°C (14-140°F)
Humidity (Operating) 10-90% RH
Humidity (Storage) 5-95% RH
Safety Standards CE, CSA, UL 61010-1, IEC 61010-1, EMC IEC 61326-1
Intended Use Testing and calibration of ventilators and anesthetic equipment in laboratory environment
Service Life 10 years

Frequently Asked Questions - Citrex H3 Rigel Medical

What is the Citrex H3 used for?
The Citrex H3 is a measuring device designed for testing and calibrating medical devices that generate gas flows or pressures, such as ventilators and anesthetic equipment. It measures flow, pressure, volume, temperature, and optionally oxygen concentration in a laboratory environment.
How do I turn on the Citrex H3?
Press the On/Off button briefly. You will hear an audible signal. To turn off, press and hold the button for about 1 second. If unresponsive, press and hold for 6 seconds to force shutdown.
How to perform zero calibration?
Remove all tubes from the device. Wait for warm-up (approx. 10 minutes). Press and hold the Menu button (x) for about 3 seconds until the display shows "Zero Offset – Calibrating, please wait". Ensure no flow or pressure during calibration.
How to change the battery?
The battery is user-replaceable. Remove the two screws on the back of the device, open the compartment, and replace the battery ensuring terminals are aligned correctly. Use only the recommended spare part (301.625.000).
How to install the oxygen sensor?
Remove the protective cap from the sensor port. Screw the oxygen sensor clockwise into the port, ensuring no leaks. Connect the sensor cable to the device's O2 interface. Perform oxygen calibration after installation.
What maintenance is required?
Use the included dust filter during operation. Check the bacterial filter every 4 weeks for soiling (pressure drop ≤2 mbar at 60 L/min). Annually, perform factory calibration and servicing through IMT Analytics or an authorized partner.
Can I use the Citrex H3 on patients?
No. The device is intended for laboratory use only, not for patient monitoring or direct connection to patients. It must not be used during patient treatment with a ventilator.
How to connect to a PC for configuration?
Use the Ethernet interface. Set the device to 'Default' mode (IP 192.168.1.1). Configure your PC with a static IP in the 192.168.1.2-255 range, subnet 255.255.255.0. Open Internet Explorer and enter 192.168.1.1 to access the web server.
What gas standards are supported?
The device supports multiple gas standards including ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPS-A, BTPD. These can be cycled through in the Settings menu.
How to enable the oxygen option?
Connect the device to a PC via Ethernet. Access the web server using Internet Explorer. Enter the enable code (obtained separately) in the 'Software Options' menu. The code is device-specific and entered once.

User questions about Citrex H3 Rigel Medical

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Measuring device in PDF format for free! Find your manual Citrex H3 - Rigel Medical and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. Citrex H3 by Rigel Medical.

USER MANUAL Citrex H3 Rigel Medical

the art of measuring

IMT.Analytics CITREX H3 ATP Air O2 man Ppeak 25 mbar PEEP 5 mbar

User Manual

CITREX H3

IMT.Analytics

IMT Analytics AG

Gewerbestrasse 8

9470 Buchs (SG)

Switzerland

www.imtanalytics.com

Table of Contents

1 Introduction 5
2 Intended use 6
3 Safety instructions 7

3.1 Representation of hazards, cautions and notes 7
3.2 Personnel 7
3.3 Responsibility and guarantee 7
3.4 Service life 7

4 Symbol explanation 8

5 Start-up 9

5.1 Power supply 10
5.2 Mechanical connectors 11

5.2.1 Flow channel 11
5.2.2 Oxygen sensor (option) 12
5.2.3 Installing the oxygen sensor 12

5.3 Electrical interfaces 14
5.4 Change CITREX battery 15

6 Operation 16

6.1 Switching the device on/off 16
6.2 Screen lock 16
6.3 Dim screen 16
6.4 User controls 17
6.5 Settings 18

6.5.1 Info display 18
6.5.2 Battery indicator 18
6.5.3 Ethernet interface 18
6.5.4 Trigger 19
6.5.5 Gas standard 19
6.5.6 Gas type 20
6.5.7 Gas humidity 20
6.5.8 Oxygen calibration (optional) 20

6.6 Numerical readings 21

6.7 Filter 21
6.8 Change parameters and units 22

7 Calibration

7.1 Zero point 23
7.2 Oxygen (O _2 ) calibration 23

7.2.1 Calibration with air 24
7.2.2 Calibration with oxygen and air 24

8 Connecting the device

8.1 General measurement setup 25
8.2 Measurement setup for checking ventilators 26

9 Configuration tool 27

9.1 PC minimum requirements 27

9.2 Web server 27

9.2.1 Default 27

9.2.2 Configured 28

9.2.3 DHCP 28

9.2.4 Enable oxygen option 28

10 Servicing and care 29

10.1 Preventive cleaning and servicing operations 29

10.1.1 During operation 29

10.1.2 Every 4 weeks 29

10.1.3 Every 12 months 29

11 Accessories and spare parts 30

11.1 Accessories table 30

12 Disposal 31

13 Directives and approvals 32

14 Specifications 33

14.1 Measurement parameter 33

14.2 Gas type 36

14.3 Power supply 36

14.4 Battery operation 36

15 Appendix 37

15.1 Principle of flow measurement 37

15.2 Trigger 37

15.2.1 Flow trigger 38

15.2.2 Pressure trigger 38

15.2.3 Base flow 38

15.2.4 Delay 38

15.3 Measurement parameters and units 39

15.4 Gas standards for flow and volume readings 40

15.5 Conversion factors 41

15.6 List of tables 42

15.7 List of figures 42

15.8 Index 43

1 Introduction

CITREX H3 was developed in order to measure flow and various pressures and thus calculate various ventilation parameters. CITREX H3 is a compact, mobile and easy-to-operate measuring instrument. The device was developed in order to calibrate gas flows and pressures by simple means. With the oxygen option it is possible to measure oxygen concentrations.

The descriptions and instructions in this manual refer to the product CITREX H3. In this User Manual the unit "sL/min" is based on ambient conditions of 0°C and 1013.25 mbar in accordance with DIN 1343.

This documentation applies to the following versions:

CITREX H3 software: 4.4.000

CITREX H3 hardware: 4.0

In the case of older or newer versions there may be discrepancies in relation to this User Manual.

Subject to technical modifications without notice.

Rigel Medical Citrex H3 - Introduction - 1

To avoid possible injuries, please read all the safety instructions before you use the product.

Rigel Medical Citrex H3 - Introduction - 2

The device is not intended for use outside a building.

2 Intended use

Rigel Medical Citrex H3 - Intended use - 1

This product is intended for testing and calibration purposes on medical devices or systems that generate gas flows or gas pressures. That includes ventilators and anaesthetic equipment. The user of the device has received training on how to use medical equipment and can perform repairs, maintenance and servicing on medical devices. CITREX H3 can be used in hospitals, in clinics, at device manufacturers or at independent service companies that perform repairs or servicing operations on medical devices.

CITREX H3 is intended for use in a laboratory environment. It may only be used outside the nursing sector. It must not be used directly on patients or devices that are connected to patients. The measuring instrument CITREX H3 is intended for over-the-counter sale.

With CITREX H3 you have the solution for measurements in the following areas:

  • Flow
  • Pressure
    • Volume
  • Ambient pressure
  • Temperature

In addition, various ventilation parameters can be measured:

  • Ventilation rate
    • Time
  • Ratio
  • Tidal volume
  • Minute volume
  • Peak flow
  • Pressure
  • Trigger

Rigel Medical Citrex H3 - Intended use - 2

CITREX H3 is a measuring instrument for checking and calibrating ventilators and anaesthetic equipment. It must not be used for patient monitoring. During patient treatment by the ventilator it is not allowed to connect to CITREX H3.

It is not allowed to measure liquids with CITREX H3.

3 Safety instructions

Please read all the safety instructions carefully before you use CITREX H3.

3.1 Representation of hazards, cautions and notes

This User Manual uses the representation below to specifically draw attention to residual risks during intended use and emphasise important technical requirements.

Information and/or instructions and prohibitions to prevent damage of any kind, as well as useful tips and information for handling the device, will be indicated by the following icon:

Rigel Medical Citrex H3 - Representation of hazards, cautions and notes - 1

3.2 Personnel

Rigel Medical Citrex H3 - Personnel - 1

Work on and with CITREX H3 may only be performed by persons who have undergone appropriate technical training and have the necessary experience.

3.3 Responsibility and guarantee

The manufacturer accepts no responsibility or guarantee and will exempt itself from liability claims accordingly if the operator or any third parties:

  • Fail to use the device in accordance with its intended use.
  • Disregard the specifications.
  • Tamper with the device in any way (conversions, modifications or the like).
  • Operate the device with accessories that are not listed in the associated sets of product documentation.

Rigel Medical Citrex H3 - Responsibility and guarantee - 1

Although the device meets high quality and safety standards and it has been constructed and tested according to the current state of the art, it is not possible to rule out the risk of injuries with serious consequences if the device is used in non-compliance with the intended use (improperly) or is misused.

Therefore please read through this User Manual carefully and keep this documentation in a readily accessible place close to your device.

3.4 Service life

The maximum service life of the device has been specified as 10 (ten) years, provided it is handled properly in accordance with this User Manual.

4 Symbol explanation

The symbols listed below may appear on the packaging material, on the device rating plate and in the User Manual of the CITREX H3 measuring instrument.

Rigel Medical Citrex H3 - Symbol explanation - 1USB interface
SN BBXXXX Serial number
Rigel Medical Citrex H3 - Symbol explanation - 2Ethernet interface
Rigel Medical Citrex H3 - Symbol explanation - 3On/Off button
Rigel Medical Citrex H3 - Symbol explanation - 4SD card
Rigel Medical Citrex H3 - Symbol explanation - 5Fragile contents
Rigel Medical Citrex H3 - Symbol explanation - 6Keep dry
Rigel Medical Citrex H3 - Symbol explanation - 7Read the User Manual
Rigel Medical Citrex H3 - Symbol explanation - 8The device must not be disposed of in household waste
Rigel Medical Citrex H3 - Symbol explanation - 9The device is CE approved
Rigel Medical Citrex H3 - Symbol explanation - 10Caution: observe the safety instructions in the User Manual
Rigel Medical Citrex H3 - Symbol explanation - 11Reusable packaging
Rigel Medical Citrex H3 - Symbol explanation - 12Manufacturer's specification and date of manufacture
Rigel Medical Citrex H3 - Symbol explanation - 13Keep away from heat
Rigel Medical Citrex H3 - Symbol explanation - 14Temperature range for storage and transport
Rigel Medical Citrex H3 - Symbol explanation - 15CSA monogram with C/US indicator
---Direct current

Table 1: Symbol explanation

5 Start-up

Rigel Medical Citrex H3 - Start-up - 1CITREX H3
Rigel Medical Citrex H3 - Start-up - 2Power supply plug with country-specific adapters
Rigel Medical Citrex H3 - Start-up - 3USB cable
Rigel Medical Citrex H3 - Start-up - 4MicroSD card
Rigel Medical Citrex H3 - Start-up - 5Dust filter RT019
Rigel Medical Citrex H3 - Start-up - 6Network cable

Table 2: Scope of delivery

5.1 Power supply

CITREX H3 can be operated from the mains or from the integrated battery.

Power can be supplied via the USB port (Mini-B) on the top of CITREX H3. Use the power supply unit included to charge the battery or operate the device via the USB port. You will find more information about power supply in the section "Electrical interfaces".

During the charging process a green battery symbol is lit on the front.

Please connect the power supply unit included to a voltage of 100 VAC to 240 VAC with a frequency of 50 Hz to 60 Hz.

USB port ©MB100001 Dedicated in the European ©A 2014 GB 1975 CEA® USB port ©MB100001 Dedicated in the European ©A 2014 GB 1975 CEA® USB port

Figure 1: Power supply

Rigel Medical Citrex H3 - Power supply - 2

Before switching on, make sure the operating voltage of the power supply unit agrees with the local mains voltage. You will find this information on the rating plate on the back of the power supply unit. When operating CITREX H3 via the USB port only use the original power supply unit included!

Rigel Medical Citrex H3 - Power supply - 3

The device indicates visually and audibly when the battery has to be charged. Please do not store the battery in a depleted state.

Caution: depletion can damage the battery beyond repair!

5.2 Mechanical connectors

5.2.1 Flow channel

The flow channel can be used bidirectionally. The positive flow direction is from left to right, viewed from the front of the device. The measurements of volume, flow, gas temperature, oxygen (option) and channel pressure are taken in the flow channel. The values, and the ventilation parameters calculated from them, can be displayed on the screen. You will find the relevant setting options in the section "Operation".

Flow (air) Measuring range ±300sL/min
Accuracy ± 2% of reading or ± 0.1 sL/min
Volume Measuring range 0–10sL
Accuracy ± 2% of reading or ± 0.02sL
Temperature Measuring range 0–50°C
Accuracy ± 1.75% of reading or 0.5°C
Oxygen Measuring range 0–100%
Accuracy ± 1% O2
Pressure in flow channelMeasuring range -50 – 150 mbar
Accuracy ± 0.75% of reading or ± 0.1 mbar

0 GB100001 Dialysis Sensor CE Flow channel

Figure 2: Flow channel

5.2.2 Oxygen sensor (option)

CITREX H3 can be fitted with an oxygen sensor. The option is enabled using the configuration tool. You will find further information on the configuration tool in the section "Configuration tool".

After enabling, oxygen concentration can be measured in the flow channel. To do so, an oxygen sensor is screwed into the appropriate port. The oxygen sensor has to be connected to the measuring instrument using the cable included. The following steps explain how to install and replace the oxygen sensor.

Oxygen sensor holder BB100001 CNC 250-1.25-0.000 ISO 163487-000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000

Figure 3: Oxygen sensor holder

Measuring range 0–100%

Accuracy ±1% O _2 (absolute)

5.2.3 Installing the oxygen sensor

  1. Remove the protective cap from the sensor port of the device.

iMT.Analytics CITREX H3 100001895B PRETOMING OF PETERIO 100001895B

Figure 4: Protective cap

  1. Screw the oxygen sensor clockwise into the appropriate port. Make sure the sensor seals off the port and there is no leak.

IMT.Analytics CITREX H3

Figure 5: Screwing in the oxygen sensor

Rigel Medical Citrex H3 - Installing the oxygen sensor - 3

  1. Connect the cable included to the oxygen sensor by pushing the cable into the hole at the top of the sensor until the cable locks into place. Connect the other end of the cable to CITREX H3 by inserting it into the hole provided, which is labelled "O₂".

IMT.Analytics CITREX H3

Figure 6: Oxygen sensor cable

  1. Perform an oxygen calibration. The calibration procedure is described in the section "Calibration". Calibration ensures that the measured values of the new sensor are correct.

5.3 Electrical interfaces

Figure 7 shows the available electrical interfaces of CITREX H3.
4 IMT.Analytics 3 2 1 CITREX H3

Figure 7: Electrical interfaces

1 MicroSD card slotThe firmware of CITREX H3 is stored on the microSD card. It also contains customised configurations.
2 O2interface The optional oxygen sensor is connected to CITREX H3 via the O_2 interface. You will find further information on this in the section "Oxygen sensor".
3 USB port The USB port is used to operate the device from the mains power supply and to charge the device battery. It is a "USB Mini-B port".
4 Ethernet The Ethernet interface is used to configure the device and it is used as a data interface.

Table 3: Description of electrical interfaces

5.4 Change CITREX battery

The battery of CITREX H3 can be changed by the user. To do so, undo and remove the two screws on the back of the device. Then the battery can be removed and replaced. Check whether the new battery has been inserted properly. For this purpose the terminals must be opposite one another.

Rigel Medical Citrex H3 - Change CITREX battery - 1

natural_image 3D mechanical assembly diagram showing a motor with blue directional arrows indicating motion or force (no text or symbols)

Figure 8: Change battery

6 Operation

This section describes how to use the device and what possible uses there are.

6.1 Switching the device on/off

The device is switched on and off at the On/Off button. Figure 8, section "User controls", shows where this button is located on the device. To switch CITREX H3 on you must press the On/Off button briefly. You will hear an audible signal. To switch the device off you must press the On/Off button for about 1 second. If the device can no longer be controlled, you have the option of pressing the On/Off button for about 6 seconds. The device is then forced to shut down.

6.2 Screen lock

Press the context button on the side of the device for 2 seconds. The screen shows a message indicating that the screen is locked. To unlock the screen, press and hold down the context button or one of the four buttons on the front for 2 seconds.

6.3 Dim screen

If the device is not operated by the user, the display of CITREX H3 shuts down after about one minute and the four buttons start to flash. As soon as a button is pressed, the screen comes on again.

The setting for how long it takes until the display is dimmed can be customised using the configuration tool. You will find further information on this in the section "Configuration tool".

6.4 User controls

IMT.Analytics CITREX H3 10 9 8 7 6 5 4 3 2 1 11/16/198-05:20 12 11/16/2000 (10.32.19.000) BIB10001 Citation in EMB/2000 IPT1 IPT1 IPT1 IPT1 IPT1 IPT1 IPT1

Figure 9: User controls

1 Change, Edit
2 Context button; long press: key lock on/off
3 On/Off button
4 Menu button; zero point adjustment
5 Charge indicator
6 Flow direction indicator
7 Screen
8 Display readings and scroll forward
9 Malfunction indicator
10 Display readings and scroll back

Table 4: User controls

6.5 Settings

With the x button you access the "Settings" menu. By pressing the button more than once you can view the various settings of the measuring instrument.

6.5.1 Info display

This display provides information about the owner, the company, the next recommended calibration, the software version and the hardware revision. Settings concerning the owner can be edited with the configuration tool.

CITREX H3 Owner: John Bio Company: IMT Analytics Next Calib.: May 2018 Software: 4.1.000 Hardware: 4

Figure 10: Info display

6.5.2 Battery indicator

The battery indicator informs you about the level of the battery charge.

Battery 100%

Figure 11: Battery indicator

6.5.3 Ethernet interface

Here it is possible to make various settings for the network connection. With the ○ button you can choose between the options "DHCP Client", "Default" and "Configured". The setting does not have to be confirmed and it is enabled as soon as it is visible on the screen. More information about the settings is available in the section "Web server".

Ethernet Default IP: 192.168.1.1 Subnet: 255.255.255.0

Figure 12: Ethernet interface

6.5.4 Trigger

With the trigger settings the start and end points of a ventilation parameter are defined. Two preset triggers are available. With the ○ button you can select the trigger "Adult" or "Pediatric". The trigger settings do not have to be saved and they are enabled as soon as they are displayed on the screen. It is possible to differentiate between flow trigger, pressure trigger and external trigger.

The settings can be changed with the configuration tool. You will find further information on this in the section "Configuration tool".

Trigger Adult Start: rising Flow: 3.0 l/min 60ms End: falling Flow: -3.0 l/min 60ms

Figure 13: Trigger

The preset trigger settings are defined as follows.

Adult Pediatric
Start 3L/minRising edge1 L/minRising edge
Stop -3L/minFalling edge-1 L/minFalling edge
Delay 60ms60ms
Base flow 0L/min 0L/min

Table 5: Trigger settings

6.5.5 Gas standard

The CITREX H3 measuring instrument can convert gas flow and volume readings to various gas standards and display them. Care must be taken to ensure that on the measuring instrument the same gas standard is set as the one on the device being tested. With the O button you can switch between the various gas standards. As soon as a gas standard is displayed, it is enabled. There is a list of available gas standards in the Appendix in the section "Gas standards for flow and volume readings".

Standard ATP Ambient temp. and pressure Current gas temperature Current ambient pressure Current gas humidity

Figure 14: Gas standard

6.5.6 Gas type

Under this menu item the gas type to be measured can be set. With the O button you can switch between gas types. The gas type indicated is enabled and does not have to be saved. In the section "Gas type" there is an overview of available gas types. Gas types with adjustable oxygen concentrations, e.g. "Air O _2 manual", can be changed with the configuration tool.

Gas Type Air O2 auto

Figure 15: Gas type

6.5.7 Gas humidity

The gas humidity of the gas being measured can be set. This has an impact on gas flow measurement. With the ○ button the gas humidity can be changed in steps of 10. The value is enabled as soon as it is displayed on the screen.

Humidity 50.0%

Figure 16: Gas humidity

6.5.8 Oxygen calibration (optional)

Oxygen calibration is only available if the oxygen option has been enabled. The process of oxygen calibration is described in the section "Calibration". With the ○ button you can choose between one-point calibration and two-point calibration. Press the △ button to start calibration.

O2 Calibration O2 and Air Press O to choose mode Press △ to start

Figure 17: O_2 calibration

6.6 Numerical readings

With the □ button on the front of CITREX H3 you can display the various numerical readings. If you press more than once, the view on the screen changes. The different views can be configured by web server. The web server and how to make the settings are explained in the section "Web server".

Rigel Medical Citrex H3 - Numerical readings - 1

flowchart
graph TD
    A["ATP 0.0 Flow l/min"] --> B["1"]
    A --> C["2"]
    A --> D["3"]
    A --> E["4"]
    A --> F["5"]
    A --> G["6"]

Figure 18: Numerical readings

1 Gas Standard The measured volume readings or gas flow readings can be displayed with various gas standards. There is a list of standards in the Appendix in the section "Gas standards for flow and volume readings".
2 Trigger signalThe icon appears as soon as a trigger condition is fulfilled. This means that the time of appearance of the indicator is identified as the start of inspiration. The indicator appears for 0.5 seconds. If this signal is not displayed, the trigger settings should be adjusted for the current ventilation mode.
3 Gas type The gas type currently set is displayed as text. It can be customised on the device under Settings.
4 Reading This shows the current reading in the selected unit of measurement.
5 Measurement parameterIndicates the measurement parameter currently selected. Measurement parameters can be changed in configuration; see section "Configuration tool".
6 Measuring UnitIndicates the unit of measurement currently selected. Units of measurement can be changed in configuration; see section "Configuration tool".

Table 6: Numerical values

6.7 Filter

The screen of CITREX H3 is refreshed every 0.5 s. The recording of readings takes place every 5 ms. Since CITREX H3 can record and display readings very quickly, it is advisable to filter the readings. This is performed using a mean value. The extent to which a reading is filtered can be set using the configuration tool.

The following filters are available:

  • No filter (indication of the last value measured without any threshold)
  • Low (mean above 240 ms)
    • Medium (mean above 480 ms)
    • High (mean above 960 ms)

6.8 Change parameters and units

The "high" filter is set by default.

If the context button (≡) is pressed twice in succession, "Edit Mode" is enabled. This is indicated by a red icon on the screen. The parameter or unit in the red frame can be changed with the □ icon or the ○ icon. The Δ icon allows you to jump to the next element. If the context button or the × button is pressed once, you exit "Edit Mode"

ATP 123.2 Flow Edit Umin

7 Calibration

The various calibration options with CITREX H3 are described in this section. To avoid incorrect measurements you must adhere to the procedures described here.

7.1 Zero point

This adjustment is necessary if indication of the channel pressure sensor ( P_channel ) or a flow through open connection ports shows a value greater or less than zero. This can occur if there are considerable temperature fluctuations, or after the warming-up time. Zero calibration resets all values to zero. To perform zero calibration you must remove all connected tubes from the device. Press the × icon and keep it pressed for approx. 3 seconds. The screen shows a message "Zero Offset – Calibrating, please wait".

Zero Offset

Calibrating, please wait...

Figure 19: Zero calibration

7

Rigel Medical Citrex H3 - Zero Offset - 1

When you have switched on the device, individual displays may deviate slightly from the zero point until operating temperature has been reached. Zero calibration should never be performed with the device cold. Warming-up time is approx. 10 minutes.

Rigel Medical Citrex H3 - Zero Offset - 2

During zero calibration there must be no pressure being applied to a connection port, and care must be taken to ensure that there is no flow through the flow channel.

7.2 Oxygen (O _2 ) calibration

There are two different methods of calibrating the oxygen cell. The variant in which the oxygen cell is calibrated with air only takes about two minutes. The second variant consists of calibrating the oxygen cell with air and 100% oxygen. This so-called two-point calibration adjusts the oxygen sensor more accurately and takes about four minutes. Calibration can be selected by pressing the × button more than once.

7.2.1 Calibration with air

Calibration is only available if the oxygen option has been enabled. Make sure air is flowing through the flow channel at a rate of at least 30 L/min. To start calibration, press the × button until you see the menu item

"O₂ Calibration". With the ○ button you can switch between air and air and oxygen (O₂). Select the ○ button until "Air" appears on the screen in green letters. To start calibration, press the △ button. It takes about 120 seconds before calibration is completed.

O2 Calibration O2 and Air Press O to choose mode Press △ to start

O2 Calibration Air Apply 30 l/min (0) Press △ to continue

O2 Calibration Air calibrating Flow 0.0 l/min O2 21.2 % Time rem.: 98 s

Figure 20: Screen displays "Calibration Air"

7.2.2 Calibration with oxygen and air

For calibration of the oxygen cell with oxygen and air a gas flow of 30 L/min is used in each case. Press the × button until you see the menu item "O₂ Calibration". With the ○ button you can switch between air and air and oxygen (O₂). Press the ○ button until the screen shows "O₂ and Air" in green letters. To start calibration, press the △ button. Calibration for air and oxygen takes 120 seconds for each one.

O2 Calibration O2 and Air Press O to choose mode Press △ to start

O2 Calibration 100% O2 Apply 30 l/min (0) Press △ to continue

O2 Calibration 100% O2 calibr. Flow 0.0 l/min O2 21.2 % Time rem.: 98 s

Figure 21: Screen displays "Calibration Oxygen and Air"

8 Connecting the device

The measurement setup for CITREX H3 has an impact on flow measurement. To obtain results that are as accurate as possible, comply with the instructions in this section. It is important to ensure that the tubing in the measurement setup does not have any radii, kinks or dents. You are also recommended to always use the dust filter included.

Rigel Medical Citrex H3 - Connecting the device - 1

The measured gases must be free of oil, grease and dust.

8.1 General measurement setup

The general measurement setup applies to gas flow measurement. The RT019 filter included and a short piece of tubing must be used. This ensures laminar flow to the flow sensor unit. The filter also prevents dust, oil and grease from contaminating the CITREX H3 measuring instrument and thus prevents discrepancies in measurement results. The measurement setups shown below are dependent on the direction of gas flow being measured.

Technical diagram showing two medical or electrical measurement devices with digital displays and ventilation airflow indicators.

Figure 22: General measurement setup

The measurement setups listed below are unsuitable and produce inaccurate measurement results. Kinks, tees and angle pieces should be avoided in the flow channel. They cause turbulence in the gas being measured and hence inaccurate or incorrect measurement results.

Bad setup: Kinks, tees, angle pieces at the device inlet

Diagram showing three different types of industrial control valves with red X marks indicating failure or absence.

Figure 23: Bad setup

8.2 Measurement setup for checking ventilators
CITREX H3 is ideal for checking ventilators. The best measurement results are achieved with the measurement setup shown below. Make sure the test lung is connected to the grey aluminium connection port of CITREX H3.
Rigel Medical Citrex H3 - General measurement setup - 3

flowchart
graph LR
    A["Breathing circuit"] --> B["Filter"]
    B --> C["Tubing"]
    C --> D["CITREX H3"]
    D --> E["Test lung"]
    D --> F["IMT.Analytics EasyLung"]

Figure 24: Measurement setup for checking ventilators

9 Configuration tool

The configuration tool can only be used with Microsoft Internet Explorer.

9.1 PC minimum requirements

Microsoft® Silverlight 5 or higher

Windows x86 or x64 (64-bit mode only supports IE) 1.6 GHz or higher with 512 MB RAM

Macintosh (Intel based) Intel Core Duo 1.83 GHz or higher with 512 MB RAM

Microsoft® Windows® 10, 8.1, 8, Windows Server 2012, 7, 7 SP1, Windows Server 2008 SP2, Windows Server 2008 R2 SP1, Vista

Macintosh OS 10.6 (Intel based), MacOS 10.7 - 10.11 (Intel based)

Ethernet network connection

Screen resolution 1024 × 768 (1280 × 1024 recommended)

9.2 Web server

The Ethernet port on CITREX H3 enables access to the device via a network. Settings can be made on the device via the web browser using the so-called configuration tool. The oxygen option can also be enabled with the web server. The prerequisite for using the web server is an installed Internet Explorer with Microsoft Silverlight 5.

There are three different setting options to establish a connection between CITREX H3 and a computer. Tap the × button until the "Ethernet" menu item appears. You will find a description of the settings in the following sections.

9.2.1 Default

These are default settings that cannot be changed. These settings are recommended in order to establish a direct connection to the computer via an Ethernet cable. The configuration on CITREX H3 is as follows:

IP Address: 192.168.1.1

Subnet Mask: 255.255.255.0

To establish a connection, the network settings on the computer must be changed. For this purpose open the network settings of the computer, which are located in the Control Panel. Then open the "Internet Protocol Version 4 (TCP/IPv4)" settings. Enter an IP address between 192.168.1.2 and 192.168.1.255 and subnet mask 255.255.255.0 in the form on the screen.

Now open Internet Explorer and enter IP address 192.168.1.1 in the address field. The connection to CITREX H3 is established.

9.2.2 Configured

This setting option is suitable for connecting CITREX H3 to a network that does not have a DHCP server. Define an IP address and a subnet mask using the configuration tool on CITREX H3. Then the device can be connected up to the network and be accessed using the defined IP address via Internet Explorer.

9.2.3 DHCP

To connect CITREX H3 to a DHCP server, first connect CITREX H3 to the network. In the "Ethernet" menu select the setting "DHCP". With the IP address shown on the display it is possible to establish a connection to CITREX H3 via Internet Explorer.

9.2.4 Enable oxygen option

With the PC, establish a connection to CITREX H3, as described above. The enable code obtained can be entered in the "Software Options" menu. The code is device-dependent and only has to be entered once.

CITREX device configuration monitoring SUMMARY SOFTWARE OPTIONS software options Release additional software options with an activation code. The activation code can be obtained from your sales representative based on the device serial number. Oxygen measurement activated

Figure 25: Enable oxygen option

10 Servicing and care

Careful servicing in compliance with the instructions is essential for ensuring that CITREX H3 operates safely and efficiently. Only components recommended by the manufacturer may be used.

Rigel Medical Citrex H3 - Servicing and care - 1

It is absolutely essential to comply with the guidelines and servicing instructions issued by the various manufacturers.

Rigel Medical Citrex H3 - Servicing and care - 2

The servicing operations listed below may only be performed by persons who are familiar with CITREX H3. All further repair work may only be performed by authorised trained professionals. Please also observe the information issued by the various manufacturers.

10.1 Preventive cleaning and servicing operations

To ensure that your device operates with precision and reliability for as long as possible, it is essential to perform the following servicing routines regularly.

10.1.1 During operation

Use of the filter included in order to protect the device against contamination. Make sure the device is only used inside a building.

10.1.2 Every 4 weeks

Check the bacterial filter for soiling. For this purpose the pressure drop above the filter must be measured. The pressure drop must not exceed a value of 2 mbar at a flow of 60 L/min. Otherwise the filter must be replaced.

10.1.3 Every 12 months

Factory calibration and servicing to ensure reliable measurement; it may only be performed by IMT Analytics or an authorised partner.

To have CITREX H3 calibrated at the manufacturer's, IMT Analytics, visit the website www.imtanalytics.com/easycal

The EasyCal service makes it possible for users to have CITREX H3 calibrated and adjusted quickly and easily. The annual servicing procedure is also performed.

11 Accessories and spare parts

On the website www.imtanalytics.com you will find the original spare parts and other products from IMT Analytics.

Ordering address:

IMT Analytics AG

Gewerbestrasse 8

CH-9470 Buchs, Switzerland

Tel: +41 (0) 81 750 67 10

Email: sales @imtanalytics.com

Orders can also be placed in our Webstore.

11.1 Accessories table

Options

305.056.000 Oxygen sensor with activation code
305.055.000 Warranty extension (plus 2 years) CITREX H3

Service

000.000.017 Calibration & Servicing CITREX H3
000.000.022 ISO17025 Calibration & Servicing CITREX H3
000.000.018 Receiving inspection of CITREX H3
305.054.000 Triple Calibration & Servicing Package for CITREX H3

Accessories & Consumables

300.548.000 Adapter set
301.997.000 Car adapter for CITREX
302.077.000 Laminar inlet pipe
304.161.000 Black protective pouch for CITREX
304.161.001 Red protective pouch for CITREX
304.161.002 Blue protective pouch for CITREX
301.851.000 MicroSD memory card
301.655.000 Blind plug for oxygen connector (rubber)
302.178.000 Blind plug for oxygen connector (solid)
301.624.000 Oxygen sensor with mono jack
302.531.000 Bacterial filter RT019
304.714.000 CITREX stand

Spare parts

301.936.000 Carrying case for CITREX H4
301.625.000 Battery for CITREX
301.563.000 Network cable
301.673.000 USB cable for CITREX
301.653.000 Oxygen sensor cable
304.578.000 Power supply plug for CITREX
302.780.000 Flow channel protective cap

Table 7: Accessories

12 Disposal

Disposal of the device is the operator's responsibility. The device can ...

  • be delivered, carriage free and duty paid, to the manufacturer for disposal.
  • be handed over to a licensed private or public collection company.
  • be professionally broken down into its constituent parts by the operator and be recycled or disposed of in accordance with regulations.

In the case of self-disposal the disposal regulations are country-specific and are contained in relevant laws and ordinances. These codes of conduct must be obtained from the authorities responsible.

In this context, wastes must be recycled or destroyed ...

• without endangering human health.
- without using processes or methods that harm the environment, especially water, air, soil, animals and plants.
- without causing noise or odour nuisances.
- without having a detrimental effect on the surroundings or landscape.

13 Directives and approvals

• CE
• CAN/CSA-C22.2 No. 61010-1-12
• UL Std. No. 61010-1 (3rd Edition)
• IEC 61010-1 2010
• IEC 61326-1 2012
• ETSI EN 301 489-17 V3.1.0
• FCC Part 15, Subpart B, Digital Devices, Emission Class B

CE Declaration of Conformity

2014/35/EU (LVD)

DIRECTIVE 2014/35/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 26 February 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of electrical equipment designed for use within certain voltage limits tested according to EN61010-1:2010

2014/30/EU (EMC)

DIRECTIVE 2014/30/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 26 February 2014 on the harmonisation of the laws of the Member States relating to electromagnetic compatibility tested according to EN61326-1:2013

14 Specifications

14.1 Measurement parameter

Flow and pressure measurement Measuring range Accuracy

Air and N_2
Flow measurement ±300 sL/min***±2%* or ±0.1 sL/min**
Temperature compensated Yes
Ambient pressure compensated Yes
Channel pressure compensatedYes-50 -600 mbar

O_2/air mixtures

Flow measurement ±300 sL/min *** ±2%* or ±0.1 sL/min **
Temperature compensated Yes
Ambient pressure compensated Yes
Channel pressure compensatedYes-50 - 600 mbar

N_2O/O_2 mixtures

Flow measurement ±80 sL/min***±4%* or ±0.3 sL/min**
Temperature compensated Yes 25-30°C
Ambient pressure compensated Yes
Channel pressure compensatedYes-50 -600 mbar

Pressure

In flow channel -50-150 mbar ± 0.75%* or ± 0.1 mbar**
Barometer 500-1150 mbar ±1%* or ± 5 mbar**
Additional readings Measuring range Accuracy
Oxygen concentration (pressure compensated ≤ 150 mbar)0–100% ± 1% O2**
Gas temperature****0–50°C± 1.75%* or ± 0.5°C**
Gas typeAir, O_2 , Air/ O_2 , N_2O/O_2
Gas Standard ATP, ATPD, ATPS,AP21, STP,STPH, BTPS, BTPS-A, BTPD

Table 8: Measurement parameters

Units of measurement

Flow L/min, L/s, cfm
Pressure bar, mbar, cmH _2O , mmHg,inH _2 O

Table 9: Additional readings
It is the larger tolerance that applies: * Tolerance in relation to the reading ** Absolute tolerance

*** In this User Manual the unit sL/min is based on ambient conditions of 0°C and 1013.25 mbar (DIN1343)
*** CITREX H3 measures the gas temperature inside the measurement channel. While CITREX H3 is warming up, the temperature of the measurement channel, and hence also the temperature of the gas inside the measurement channel, rises at the same time. The measurement channel volume is relatively small, even for relatively high volumetric flows (e.g. PIF @ 60 L/min). If the temperature of the gas on entering CITREX H3 is compared with gas temperature in the measurement channel, it becomes evident that the temperature in the measurement channel is higher. Therefore the temperature of the gas entering the CITREX H3 measurement channel should not be expected to equal the temperature displayed on the screen because the temperature displayed is measured inside the CITREX H3 measurement channel.

Ventilation parameters MeasuringrangeAccuracy
Rate Breaths/min1-1000breaths/min±1 breath or ±2.5%**
Time Ti0.05-60s±0.02s
RatioI:E1:300-300:1±2.5%*
Tidal volume V_i ±10sL±2%* or ±0.20mL (>6sL/min)**
Minute volume V_i 0-300sL/min±2.5%*
Peak flowInsp./exp.±300sL/min±2%* or ±0.1sL/min**
Pressure P_peak , P_mean , PEEP0-150mbar±0.75%* or ±0.1mbar**
TriggerAdult, Pediatric, Flow, Pres-sure

Table 10: Ventilation parameters

General information

Screen 1.7" colour display
Interfaces Ethernet
AC input 100–240 VAC (50/60Hz)
Battery operation 4 hours
Dimensions (W × D × H) 11.4 × 7 × 6 cm
Weight 0.38 kg
Calibration interval Once a year
Memory card Yes

Operating data

Ambient temperature 15-40°C (59-104°F)
Air humidity 10-90% RH
Ambient pressure 783-1150mbar
Storage and transport conditions-10-60°C (14-140°F) at 5-95% RH
Degree of soilingDegree of soiling 2, to IEC 61010-1

Table 11: General information and operating data

It is the larger tolerance that applies: * Tolerance in relation to the reading ** Absolute tolerance

*** In this User Manual the unit sL/min is based on ambient conditions of 0°C and 1013.25 mbar (DIN1343).

14.2 Gas type

The gas type measured must agree with the setting on CITREX H3. Please select the correct gas type in the settings.

The following gas types are available for selection:

  • Air 100%
  • Air/O 2 -Man. Air/oxygen mixture according to manual input; the default is 100% O 2
  • Air/O _2 -Auto. (optional) Air/oxygen mixture according to sensor measurement of internal oxygen cell
  • N2 O / O2 -Man. Nitrous oxide / oxygen mixture according to manual input; the default is 100 % O2

Standard conditions are understood to mean defined conditions for pressure, temperature and, in some cases, humidity, which constitute the basis for converting the flow actually measured. Therefore it is essential to check which standard condition the value displayed will relate to.

The standard currently set is indicated on the numerical display.

Rigel Medical Citrex H3 - Gas type - 1

A gas type that has not been selected properly and a gas standard that has not been selected properly can lead to measuring errors of up to 20%.

14.3 Power supply

Input voltage of the power supply unit 100 – 240 VAC, 50 – 60 Hz Supply voltage 5 VDC Power input 2.5–6 W

14.4 Battery operation

Operating time in battery operation 4 hours*

Charging the battery A complete charging process takes 5 to 8 h, depending on which connection port is used for charging. The service life of the battery is extended if the battery is charged completely only after a prompt by the device.

15 Appendix

15.1 Principle of flow measurement

Flow in the flow channel is determined by differential pressure measurement. To build up differential pressure a linear flow element is used to provide flow resistance.

Δρ = c₁ × η × Q + c₂ × ρ × Q²

η: dynamic viscosity of the gas [Pa s]

ρ: gas density [kg/m³]

c_1, c_2 : device-specific constants (channel geometry)

Dynamic viscosity

  • The viscosity of the medium is its resistance to flow and shear.
    • Viscosity is extremely dependent on temperature.
  • The viscosity of a medium is slightly dependent on the pressure and moisture content of the medium.

Density

• Density is the unit for the mass per unit volume of the medium.
• Viscosity is extremely dependent on temperature.
- The viscosity of a medium is slightly dependent on the pressure and moisture content of the medium.

15.2 Trigger

Triggers are used to define start and end points of cyclical signals. With regard to pressure and flow curves the trigger makes it possible to determine inhalation and exhalation. The resulting information constitutes the basis for ventilation parameter calculation. If the trigger is not set properly or if it is not possible to detect a trigger, the ventilation parameters will be calculated incorrectly or not at all.

Rigel Medical Citrex H3 - Trigger - 1

line | Phase | Druck | Fluss | | --------- | ----- | ----- | | Start | 0 | 0 | | Stop | 0 | 0 | | Start | 0 | 0 |

Figure 26: Trigger

15.2.1 Flow trigger

On CITREX H3 a flow trigger can be set. When the set flow is reached, the trigger is activated. In this context it is essential to specify whether at the start and end of a cycle the trigger is to be activated by a rising edge or a falling edge. Flow measurement in the flow channel serves as the trigger source. CITREX H3 can be operated bidirectionally.

15.2.2 Pressure trigger

In the case of a pressure trigger it is the pressure measured in the flow channel that serves to activate measurement. Whereby the direction of flow is of no consequence.

15.2.3 Base flow

Base flow is a constant flow that must not be included in volume calculation. If, for example, a system has a defined leak, resulting in a constant discharge of 3 L/min, those 3 L/min are not included in the inspiratory volume. The 3 L/min can be entered as a trigger setting so they will not be taken into account.

15.2.4 Delay

With a delay it is possible to filter out signal errors or noise and prevent false triggering. As a result, a trigger is only activated if the set trigger level continues to apply after the delay time. If the trigger level is no longer reached after the delay time, no trigger is activated. The delay time can be set.

Rigel Medical Citrex H3 - Delay - 1

line | Vti | Vflss | | ------ | ----- | | Vti | 0.5 |

Figure 27: Delay

15.3 Measurement parameters and units

Pressure readings Measurement parameter Designation Units of measurement
Ambient pressurePressure in flow channel high P_atm . P_chane mbar, bar, in H_2O , cm H_2O , mmHg
Flow readings Measurement parameter Designation Units of measurement
Flow Flow L/min, cfm, L/s
Meteorological readings Measurement parameter Designation Units of measurement
TemperatureOxygen content (option)VolumeTemp. O_2 Volume°C, K, °F%mL, L, cf
Gas concentrations Measurement parameter Designation Units of measurement
Gas concentration Gas concentration %
Ventilation parameters Measurement parameter Designation Units of measurement
Positive end-expiratory pressureMean pressureMaximum pressurePEEPPmean P_peak mbar, bar, in H_2O ,cm H_2O , mmHg.
Inspiratory minute volumeInspiratory/expiratory peak flow V_i PF _inep/PF_exp L/min, cfm, L/s,
Inspiratory volumeVentilation rateInspiratory/expiratory ratioInspiratory timeVtiRateI:ETimL, L, cfBreaths/mins

Table 12: Measurement parameters and units

15.4 Gas standards for flow and volume readings

CITREX H3 converts the flow and volume readings measured in the device to match the conditions of the standard selected. The following gas standards are supported by CITREX H3.

Gas StandardAbbreviationPressure TemperatureRelative humidity
Ambient Temperature and Pressure ATP Currentambient pressureCurrent gas temperatureCurrent gas humidity
Ambient Temperature and Pressure Dry ATPDCurrent ambient pressureCurrent gas temperature0%
Ambient Temperature and Pressure SaturatedATPS Current ambient pressureCurrent gas temperature100%
Ambient Pressure at 21°C AP21 Current ambient pressurepressure 21.0°C(70°F)Current gas humidity
Standard Conditions USASTP1013.25 mbar (760 mmHg)21.0°C (70°F)0%
Standard Conditions USA HumidSTPH1013.25 mbar (760 mmHg)21.0°C (70°F)Current gas humidity
Body Temperature and Pressure, SaturatedBTPSCurrent ambient pressure + channel pressure37.0°C(99°F)100%
Body Temperature and (Ambient) Pressure Saturated according to ISO 80601-2-12:2011BTPS-ACurrent ambient pressure37.0°C (99°F)100%
Body Temperature and Pressure DryBTPDCurrent ambient pressure + channel pressure37.0°C(99°F)0%

Table 13: Gas standards for flow and volume readings

15.5 Conversion factors

Value Equivalent
1 mbar 0.001 bar
100 Pa
1 hPa
0.1 kPa
0.75006 torr (760 torr = 1 atm.)
0.75006 mmHg (at 0°C)
0.02953 inHg (at 0°C)
1.01974 cmH _2O (at 4°C)
0.40147 inH _2O (at 4°C)
0.01450 psi, psia
1 bar 1000 mbar
0.1 Pa
1000 hPa
100 kPa
750.06 torr (760 torr = 1 atm.)
750.06 mmHg (at 0°C)
29.53 inHg (at 0°C)
1019.74 cmH _2O (at 4°C)
401.47 inH _2O (at 4°C)
14.50 psi, psia

Table 14: Conversion factors

15.6 List of tables

Table 1: Symbol explanation 8

Table 2: Scope of delivery 9

Table 3: Description of electrical interfaces 14

Table 4: User controls 17

Table 5: Trigger settings 19

Table 6: Numerical values 21

Table 7: Accessories 30

Table 8: Measurement parameters 33

Table 9: Additional readings 34

Table 10: Ventilation parameters 34

Table 11: General information and operating data 35

Table 12: Measurement parameters and units 39

Table 13: Gas standards for flow and volume readings 40

Table 14: Conversion factors 41

15.7 List of figures

Figure 1: Power supply 10

Figure 2: Flow channel 11

Figure 3: Oxygen sensor holder 12

Figure 4: Protective cap 12

Figure 5: Screwing in the oxygen sensor 13

Figure 6: Oxygen sensor cable 13

Figure 7: Electrical interfaces 14

Figure 8: Change battery 15

Figure 9: User controls 17

Figure 10: Info display 18

Figure 11: Battery indicator 18

Figure 12: Ethernet interface 18

Figure 13: Trigger 19

Figure 14: Gas standard 19

Figure 15: Gas type 20

Figure 16: Gas humidity 20

Figure 17: O_2 calibration 20

Figure 18: Numerical readings 21

Figure 19: Zero calibration 23

Figure 20: Screen displays "Calibration Air" 24

Figure 21: Screen displays "Calibration Oxygen and Air" 24

Figure 22: General measurement setup 25

Figure 23: Bad setup 25

Figure 24: Measurement setup for checking ventilators 26

Figure 25: Enable oxygen option 28

Figure 26: Trigger 37

Figure 27: Delay 38

15.8 Index

A

Accessories 30

Approvals 32

B

Battery depletion 10

Battery operation 36

C

Calibration 23

Calibration with air 24

Calibration with oxygen and air 24

Care 29

Change battery 15

Change units 22

Charging the battery 36

Cleaning 29

Configured 28

Connecting the device 25

Conversion factors 41

D

Default 27

DHCP 28

Dim screen 16

Directives 32

Disposal 31

Dynamic viscosity 37

E

Electrical interfaces 14

Ethernet 14

F

Filter 21

Flow and volume readings 40

Flow channel 11

Flow measurement 37

G

Gas concentrations 39

Gas Standard 21

Gas standards 40

|

Intended use 6

M

Measurement parameters 33, 39

Measurements 6

Measurement setup 25

Mechanical connectors 11

Meteorological readings 39

MicroSD 14

N

Notes 7

Numerical readings 21

0

O _2 23

O_2 interface 14

Operating time 36

Operation 16

Options 30

Oxygen 23

Oxygen option

Enable oxygen option 28

Oxygen sensor 12

Oxygen sensor, installing 12

P

Parameter 22

PC minimum requirements 27

Personnel 7

Power supply 10, 36

Pressure readings 39

S

Safety instructions 7

Screen lock 16

Service life 7

Servicing 29

Servicing operations 29

Settings 18

Spare parts 30

Specifications 33

Start-up 9

Switching the device on/off 16

Symbol explanation 8

T

Trigger 37

Trigger signal 21

U

Units 39

USB port 14

User controls 17

V

Ventilation parameters 6

W

Web server 27

Z

Zero point 23

IMT Analytics

IMT Analytics

Table of contents Click a title to access it
Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Rigel Medical

Model : Citrex H3

Category : Measuring device