CT-561 - Thermometer CITIZEN - Free user manual and instructions
Find the device manual for free CT-561 CITIZEN in PDF.
| Product Type | Digital Thermometer |
| Brand | Citizen |
| Model | CT-561 |
| Measurement Type | Contact (oral, rectal, underarm) |
| Temperature Range | 32.0°C – 42.9°C (89.6°F – 109.2°F) |
| Accuracy | ±0.1°C (between 35.5°C – 42.0°C) |
| Display | LCD, 3.5 digits |
| Power Supply | 1× CR2032 3V lithium battery |
| Auto Power Off | After approximately 1 minute of inactivity |
| Memory | Recalls last measurement |
| Measurement Time | Approximately 60 seconds (oral) |
| Dimensions | 15.0 × 2.0 × 1.0 cm |
| Weight | 20 g (including battery) |
| Material | ABS plastic body, stainless steel sensor tip |
| Water Resistance | Splash proof (do not submerge) |
| Cleaning | Wipe with 70% isopropyl alcohol before and after use |
| Storage Conditions | Dry place, 10°C – 40°C, avoid direct sunlight |
| Warranty | 2 years from date of purchase |
| Certifications | CE, FDA listed |
| Intended Use | For household use, not for continuous monitoring |
Frequently Asked Questions - CT-561 CITIZEN
User questions about CT-561 CITIZEN
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
Download the instructions for your Thermometer in PDF format for free! Find your manual CT-561 - CITIZEN and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. CT-561 by CITIZEN.
USER MANUAL CT-561 CITIZEN
OPERATION
-
Press the ON/OFF switch to activate the thermometer and 188.8 oC will show on the LCD display window to insure all parts of the display are functioning properly. By holding down the ON/OFF switch, the display will show the last temperature until the switch is released.
-
When release the ON/OFF switch, 36.5°C(97.7°F) will show on the display and then the L with flashing°C(°F) indicates the thermometer being ready to take a temperature. The temperature reading should take about ninety seconds.
-
When the peak temperature has been reached, ° C( ° F) will stop flashing and beeper will sound for four seconds.
-
To turn the unit off, press the ON/OFF switch. If the thermometer is inadvertently left on after use, it will automatically shut-off in 10 minutes.
Remark: 1. If the ambient temperature is higher than 32.0°C(89.6°F) , the display will not show L°C ( L°F ) after showing 36.5°C(97.7°F) .
- If the temperature goes lower than 32.0°C(89.6°F), the display will only show L°C (L°F), but if the temperature goes higher than 44.0°C(111.2°F), it will show H°C (H°F) only.
3.Temperatures should only be taken under the supervision of an adult and it is not suggested to take a rectal temperature with our thermometer.
ORAL USE
Put the probe tip under the rear part of the tongue and close the mouth tightly to insure accurate readings. Normal oral temperature is 37.0°C(98.6°F) .
AXILLARY USE
The probe tip should be placed under the armpit and hold the temperature parallel to the body and put the arm on the chest so the probe tip is well covered by the arm. NOTE: axillary temperature is usually 0.5°C(1.0°F) lower than oral readings.
Remark: When taking your temperature, please use the thermometer in the same position to measure in the same time period.
THERMOMETER CLEANING
You can clean thermometer by wiping it with soap and water or 70% Isopropyl Alcohol whenever you take your temperature.
CHANGING THE BATTERY
When the battery is weak and shows the symbol, change the battery to insure precise temperature readings. To replace the battery, use a small sharpened item to push the small hole to unload the battery cover. Pull out the battery compartment with the hook on the cover. Remove the battery and insert a new battery with positive (+) side facing up and slide back the cover.
WARNING
1.Do not let the thermometer fall or twist the unit.
- Keep it off all chemical solvents, direct sunshine or high temperature.
3.Do not operate the product near transmitter which have high voltage or high frequency wave.
4.Do not recharge the battery or dispose it in fire.
5.Cleaning the thermometer prior to each use to evade cross infection in oral use.
SPECIFICATIONS

DISPLAY: liquid crystal display 3 1/2 digits
DISPLAY RESOLUTION: 0.1°C OR 0.1°F
MEASUREMENT RANGE: 32.0°C(89.6°F) to 44.0°C(111.2°F)
ACCURACY:±0.1°C from 35.5°C to 42.0°C within the ambient
temperature 18°C-28°C; otherwise±0.2°C (for CT561C)
±0.2°F from 95.9°F to 107.6°F within the ambient
temperature 64.4°F-82.4°F; otherwise±0.4°F (for CT561F)
BATTERY: one 1.5v button size battery (LR41 or equivalent)
BATTERY LIFE: about 18 months while used 10 minutes per day
RECOMMENDED MINIMUM MEASURING TIME: Oral -- 90 seconds
Armpit -- 3 minutes
AUTOMATIC SHUT-OFF: after approximately ten minutes
OPERATING TEMPERATURE: 10 -40°C (50.0-104.0°F) HUMIDITY: 15 - 95% RH
STORAGE TEMPERATURE: -20 -60°C (-4.0-140.0°F) HUMIDITY: 15 - 95% RH
DIMENSIONS: 130.8 mm X 18.2 mm X 11.6 mm
WEIGHT: 11.2 grams (including battery)
This clinical digital thermometer comply with EN 60601-1,EN60601-1-2,EN 12470-3; and Technical document refer to our EU representative.
● CITIZEN is a registered trademark of Citizen Holdings CO., Japan.
● Design and Specification are subject to change without notice.
3F-2. No.26, Lane 513, Jui-kuang Road, Taipei, Taiwan, R.O.C
Tel:886-2-87975035 Fax:886-2-26589280
EU Representative:
SIR'S PELLE
VIA RAVENNA 3
21017 SAMARATE (VA) ITALY
TEL:39-0331-235899 FAX:39-0331-238476
CE0197