| Product Type | Undercounter Auto Defrost Freezer |
| Brand & Model | American BioTech Supply ABT-HC-UCBI-0420A-LH |
| Capacity | 4.2 Cu. Ft. |
| Door Type & Hinge | Solid swing door, left hinged |
| Temperature Range | -15°C to -25°C (adjustable) |
| Temperature Controller | Microprocessor with digital display |
| Alarms | Audible and visual high/low temperature alarms |
| Remote Alarm Contacts | Yes |
| Temperature Probes | 2 probes (air and glycol bottle) |
| Shelves | 3 adjustable shelves with guard rail |
| Security | Keyed door lock |
| Handle | Heavy duty door handle |
| Gasket | Magnetic door gasket |
| Leveling Legs | Yes |
| Probe Access Port | 3/8" diameter |
| Exterior Finish | Powder coat |
| Insulation | High density urethane foam |
| Refrigerant | HFC-free R600a |
| Defrost Type | Auto defrost |
| Electrical | 115V, 60 Hz, 1.5 Amps, 1/8 HP |
| Dimensions (W x D x H) | 23 3/4" x 24" x 33 1/8" |
| Shipping Weight | 108 lbs. |
| Certifications | UL/C-UL listed |
| Warranty | Two years parts and labor |
| Compatibility | Pyxis, Omnicell, AcuDose RX |
| Maintenance | Clean condensor coils periodically; check door gasket seal |
| Safety Features | Keyed lock, temperature alarms, remote alarm contacts |
| Spare Parts | Available from manufacturer |
Frequently Asked Questions - ABT-HC-UCBI-0420A-LH American BioTech Supply
How do I set the operating temperature on the ABT-HC-UCBI-0420A-LH?
The temperature is set via the microprocessor controller. Press the set button, use the arrows to adjust between -15°C and -25°C, and confirm with the set button. The digital display shows the current temperature.
What do the alarms indicate?
The unit has audible and visual alarms for high and low temperature conditions. The alarm triggers if the internal temperature deviates from the set range. Suspect door left open, power failure, or overloading. Press the alarm mute button to silence.
How often should I clean the condenser coils?
Clean the condenser coils every 6 months or more if the unit is in a dusty environment. Use a soft brush or vacuum. Unplug the unit before cleaning.
Can this freezer be built-in under a counter?
Yes, this model is designed for built-in installation. Ensure proper ventilation per the installation instructions. The dimensions are 23 3/4" W x 24" D x 33 1/8" H.
What is the warranty coverage?
American BioTech Supply provides a two-year parts and labor warranty. Defective parts will be replaced free of charge. Labor costs for repair by an authorized service provider are covered.
Is the ABT-HC-UCBI-0420A-LH compatible with Pyxis, Omnicell, or AcuDose RX?
Yes, this freezer is compatible with Pyxis, Omnicell, and AcuDose RX medication dispensing systems. It meets the interface requirements for these systems.
What type of refrigerant does it use?
It uses R600a, a HFC-free refrigerant. This is environmentally friendly and complies with current regulations.
How do I level the freezer?
Turn the leveling legs counterclockwise to raise or clockwise to lower the unit. Place a level on top to adjust until it is level. Then tighten the lock nuts if present.
Can the door hinge be reversed?
No, this model has a left-hinged door and is not reversible. Ensure the left-hinge orientation matches your cabinet layout before installation.
Where can I download the full user manual?
The user manual can be downloaded for free from notice-facile.com. Search by model ABT-HC-UCBI-0420A-LH or visit the American BioTech Supply support page.
User questions about ABT-HC-UCBI-0420A-LH American BioTech Supply
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
No questions yet. Be the first to ask one.
Download the instructions for your Freezer in PDF format for free! Find your manual ABT-HC-UCBI-0420A-LH - American BioTech Supply and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. ABT-HC-UCBI-0420A-LH by American BioTech Supply.
USER MANUAL ABT-HC-UCBI-0420A-LH American BioTech Supply
| Product Name: | ABT-HC-UCBI-0420A-LH |
| Description: | Premier Undercounter Auto Defrost Freezer Built-In-LEFT HINGED 4.2 Cu. Ft. CapacityOne solid swing door, left hingedMicroprocessor temperature controllerDigital temperature displayAudible and visual high and low temperature alarmsRemote alarm contacts2 probes; 1 in air and 1 in glycol bottleAdjustable operating temperature range: -15°C – -25°CThree adjustable shelves with guard rail on back to prevent product from being pushed against the back wallBack wall evaporator coverKeyed door lockHeavy duty door handleMagnetic door gasket for positive sealLeveling legsProbe access port (3/8")Powder coat exteriorHigh density urethane foam cabinet and door insulationHFC-free refrigerant (R600a), foam insulation and packagingTwo year parts and labor warrantyPyxis®, Omnicell® and AcuDose RX® compatibleExterior dimensions: 23 3/4" W x 24" D x 33 1/8"HAuto defrost115V, 60 Hz, 1.5 Amps 1/8 HPUL/C-UL listedShipping weight: 108 lbs. |