Megger TAU3 - Measuring instrument

TAU3 - Measuring instrument Megger - Free user manual and instructions

Find the device manual for free TAU3 Megger in PDF.

📄 70 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice Megger TAU3 - page 5
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Transformer Turns Ratio Tester
Brand Megger
Model TAU3
Dimensions (H x W x D) 170 x 100 x 55 mm
Weight 0.6 kg (including batteries)
Power Supply 6 x 1.5 V AA alkaline batteries or rechargeable NiMH
Battery Life Approximately 8 hours of continuous use
Measurement Range (Turns Ratio) 0.8 to 2000
Accuracy ±0.1% of reading
Test Voltage 40 V AC maximum
Display Backlit LCD, 4-line, 16 characters per line
Memory Up to 200 test results
Interfaces USB 2.0 for data transfer
Operating Temperature -10°C to +50°C
Storage Temperature -20°C to +60°C
Humidity Range 0 to 95% non-condensing
Safety Standards IEC 61010-1, CAT II 300 V
Ingress Protection IP40
Main Functions Turns ratio, phase displacement, excitation current, winding resistance
Spare Parts & Repairability Available through authorized Megger service centers
Maintenance & Cleaning Wipe with a damp cloth; no solvents; calibrate annually
General Information Portable, rugged design for field use; complete with test leads and soft carry case

Frequently Asked Questions - TAU3 Megger

How do I connect the test leads for a single-phase transformer?
Connect the red lead to the high voltage winding (H1, H2) and the black lead to the low voltage winding (X1, X2). Ensure the transformer is de-energized before connection.
What does the 'Phase Displacement' measurement indicate?
It measures the phase angle difference between primary and secondary windings, helping to identify winding connections (e.g., Delta-Wye) and potential faults.
How can I save and download test results?
Press the 'Save' button after a test to store in internal memory. Connect via USB to a PC and use Megger's software to download results.
What batteries does the TAU3 use and how long do they last?
It uses 6 AA batteries (alkaline or NiMH). Alkaline batteries last about 8 hours of continuous use; rechargeable NiMH may last slightly less.
Can the TAU3 test auto-transformers?
Yes, the TAU3 can test auto-transformers. Connect the leads to the common and series windings accordingly; the instrument automatically detects the configuration.
What is the maximum turns ratio the TAU3 can measure?
The maximum measurable turns ratio is 2000. For ratios above this, use an external voltage divider or contact Megger support.
How do I perform a zero adjustment?
Short the test leads together and press the 'Zero' button. The display will show zero. Perform this before each series of tests for accuracy.
Is the TAU3 protected against short circuits?
Yes, the instrument has built-in overload protection. If a short circuit occurs, the test current is automatically limited to prevent damage.
What should I do if the display shows 'Error'?
Check all connections, ensure the transformer is de-energized, and verify the leads are not damaged. If the error persists, consult the manual or contact support.
How often should the TAU3 be calibrated?
Megger recommends annual calibration to maintain accuracy. Calibration can be performed by an authorized service center.

User questions about TAU3 Megger

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Measuring instrument in PDF format for free! Find your manual TAU3 - Megger and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. TAU3 by Megger.

USER MANUAL TAU3 Megger

True Three Phase Transformer Analyzer

USER MANUAL

EN - ENGLISH

Notice

The information presented in this manual is adequate for the intended use of the product. Use of the product or its individual instruments for purposes other than those specified herein requires confirmation of their validity and suitability from Megger. Refer to the warranty information below. Specifications are subject to change without notice.

WARRANTY

Products supplied by Megger are warranted against defects in material and workmanship for a period of 1 years following shipment. The warranty is void in the event of abuse (failure to follow recommended operating procedures) or failure by the customer to perform specific maintenance as indicated in this manual.

Megger

400 Opportunity Way

Phoenixville, PA 19460

610-676-8500 (Telephone)

610-676-8610 (Fax)

www.megger.com

Contents

Notice....2

Table of Contents....3

Introduction....5

Item received....6

Warnings and safety precautions....13

Technical specifications....15

Connections and controls 18

PC software installation....19

PC software update....22

PowerDB installation....24

Initialization....25

About 26

Settings....27

Measurements & Limits Settings....28

Nameplate....29

Turns Ratio....30

Resistance ....31

Efficiency....32

Regional....33

OLTC Control....34

Display....35

Administrative & Maintenance....36

Table of Contents

Quick Test....37

Confirm Vector....38

Vector, Phases, and Nameplate....39

Select Vector and Select Phases to Test....40

Nameplate 41

Turns Ratio & Balance 42

Winding Resistance & Demagnetization....45

Short Circuit Impedance 48

Efficiency....50

All tests....51

Save, Print, Export, Retest, Home, Test Setup ....52

Results ....53

Asset Template 55

Template Test ....58

Single tab test 61

Multiple Tap OLTC AC Tests....62

Multiple Tap OLTC Winding Resistance Tests....63

Multiple Tap OLTC All Tests....65

PowerDB Import 66

Service 67

Maintenance 68

Introduction

Thank you for your purchase of the Megger TAU3 True Three Phase Transformer Analyzer. The TAU3 design emphasizes safety, reliability, and ease of use. It will provide you with the information you need to test power, distribution, and instrument transformers and make informed electromechanical maintenance decisions

Purpose of this manual

This document is the user manual for the Megger TAU3 True Three Phase Transformer Analyzer. It provides a description of the instrument as well as operating instructions. Read this manual before using the equipment, with special emphasis on all safety discussions.

Audience

This manual is for technical personnel who are familiar with the various transformer measurements performed by electrical test equipment and have a general understanding of their use and operation. Such personnel should also be thoroughly familiar with the hazards associated with the use of this equipment and should have received proper safety training.

If you find any discrepancies in the TAU3 manual or have any comments, please send them to Megger via fax, e-mail, or phone.

Megger

400 Opportunity Way

Phoenixville, PA 19460

610-676-8500 (Telephone)

610-676-8610 (Fax)

USTechSupportGrp@megger.com

For technical support, please consult the Megger web site at www.megger.com for the local distributor near you.

Item received

TAU3 Models

Part Number Description Image
TAU3-BASIC62.5 VAC / 10 ADC three phase switched multifunction turns ratiometer and winding resistance ohmmeter test setMegger TAU3 - Item received - 1
TAU3-ADV125 VAC / 16 ADC true three phase multifunction turns ratiometer, winding resistance ohmmeter, and short circuit impedance/leakage reactance test set
TAU3-PRO250 VAC / 32 ADC true three phase multifunction turns ratiometer, winding resistance ohmmeter, and short circuit impedance/leakage reactance test set
TAU3-EXP250 VAC / 32 ADC true three phase multifunction turns ratiometer, winding resistance ohmmeter, and short circuit impedance/leakage reactance test set with select optional hardware and software accessories included
Model DifferentiationTAU3-BASICTAU3-ADVTAU3-PROTAU3-EXP
Integrated display10.1" Included
Confirm VectorIncluded
Winding Resistance Maximum Excitation Voltage - DC50 V100 V
TTR Maximum Current - AC3.5 A11 A
TTR Max Induced Voltage - AC62.5 V125 V250 V
Winding Resistance Maximum Current- DC10 A16 A32 A
True 3 Phase Excitation and MeasurementN/AIncluded
Phase Shifting Tx's MeasurementIncluded
Short Circuit Impedance / Leakage ReactanceIncluded
FRSLOptionalIncluded
PowerDB ControlOptionalIncluded

TAU3-BASIC and TAU3-ADV Leads

Part Number Description Image
2008-15KIT2 16 A H & X leads. 5 m (15 ft)Megger TAU3 - Item received - 2
2008-30KIT2 16 A H & X leads. 9 m (15 ft)
2008-60KIT2 16 A H & X leads. 18 m (15 ft)
2008-100KIT216 A H & X leads. 30 m (100 ft) H leads and 18 m (60 ft) X leadsMegger TAU3 - Item received - 3
2008-30XKIT2 16 A H & X lead extensions. 9 m (30 ft)Megger TAU3 - Item received - 4
Megger TAU3 - Item received - 5
2008-15KIT3 16 A H & 32 A X leads. 5 m (15 ft)Megger TAU3 - Item received - 6
2008-30KIT3 16 A H & 32 A X leads. 9 m (15 ft)
2008-60KIT3 16 A H & 32 A X leads. 18 m (15 ft)
2008-100KIT316 A H & 32 A X leads. 30 m (100 ft)H leads and 18 m (60 ft) X leads.Megger TAU3 - Item received - 7
2008-30XKIT316 A H & 32 A X lead extensions.9 m (30 ft)Megger TAU3 - Item received - 8
Megger TAU3 - Item received - 9
Part Number Description Image
1014-927 AC Power CordsMegger TAU3 - Item received - 10
CA-USB USB 2.0 CableMegger TAU3 - Item received - 11
1011-622 OLTC Tap Changer CableMegger TAU3 - Item received - 12
2012-180 Cable Bag – Back Pack (qty 2)Megger TAU3 - Item received - 13
4702-7 Ground LeadMegger TAU3 - Item received - 14

Item received

Part Number Description Image
90012-878 Blank Thumb DriveMegger TAU3 - Item received - 1

Optional software accessories

Part Number Description
SW-POWERDB PowerDB Control
SW-CUSTOMAPP Custom application control
SW-EFFICIENCY Transformer load, no-load, and efficiency measurements
SW-FRSL Frequency response stray losses
SW-DRM Dynamic resistance measurements
SW-DRYOUT Transformer dry out / winding temperature measurements
SW-HEATRUN Transformer heat run / cool down measurements

Optional hardware accessories

Part Number Description Image
TAU3-CAL-CERTTAU3 Calibration Certificate
1004-639Safety Beacon – 18 m (60 ft)Megger TAU3 - Optional hardware accessories - 1
1014-928Transit Case (for instrument)Megger TAU3 - Optional hardware accessories - 2
90029-537 USB PrinterMegger TAU3 - Optional hardware accessories - 3
90029-573-P USB Printer Paper (48 rolls)Megger TAU3 - Optional hardware accessories - 4
2005-249 1:1 Test JigMegger TAU3 - Optional hardware accessories - 5
1011-622-A OLTC tap changer cable adaptersMegger TAU3 - Optional hardware accessories - 6
TRS1PLUS TRS1+ turns ratio calibration standardMegger TAU3 - Optional hardware accessories - 7
TOS1 TOS1 resistance calibration standardMegger TAU3 - Optional hardware accessories - 8

ORDERING INFORMATION

Item (Qty) Cat. No.

TAU3 BASIC and ADV lead sets

16 Amp H leads with red jacket and red, yellow, blue, and white clamps (4 total)

16 Amp X leads with black jacket and red, yellow, blue, and white clamps (4 total)

5 m (15 ft) H and X leads 2008-15KIT2

9 m (30 ft) H and X leads 2008-30KIT2

18 m (60 ft) H and X leads 2008-60KIT2

30 m (100 ft) H and 18 m (60 ft) X leads 2008-100KIT2

Optional Lead Accessories

TAU3 BASIC and ADV 16 A H and 16 A X lead extensions

9 m (30 ft) H and X extensions 2008-30XKIT2

Included Accessories - BASIC, ADV, PRO, EXP

AC Power Cords (US, EU, UK) 1014-927

USB 2.0 Cable CA-USB

OLTC Tap Changer Cable 1011-622

Cable Bag - Backpack 2012-180

Ground Lead 4.5m (15 ft) 4702-7

Thumb Drive 90012-878

Included accessories - EXP

Second Cable Bag - Backpack 2012-180

OLTC Tap changer cable adapters 1011-622-A

USB printer 90029-573

Safety beacon - 18 m (60 ft) 1004-639

Included accessories - PRO

Second Cable Bag - Backpack 2012-180

Optional Software Accessories

PowerDB control SW-POWERDB

Custom application control SW-CUSTOMAPP

Transformer efficiency measurements* SW-EFFICIENCY

Item (Qty) Cat. No.

Frequency response stray losses measurements SW-FRSL

Dynamic resistance measurements* SW-DRM

Transformer dry out measurements* SW-DRYOUT

Transformer heat run measurements* SW-HEATRUN

Item (Qty) For Price List Cat. No.

TAU3 PRO and EXP lead sets

16 Amp H leads with red jacket and red, yellow, blue, and white clamps (4 total)

32 Amp X leads with black and white stripe jacket and red, yellow, blue, and white clamps (4 total)

5 m (15 ft) H and X leads 2008-15KIT3

9 m (30 ft) H and X leads 2008-30KIT3

18 m (60 ft) H and X leads 2008-60KIT3

30 m (100 ft) H and 18 m (60 ft) X leads 2008-100KIT3

Optional Lead Accessories

TAU3 PRO and TAU3 EXP 16 A H and 32 A X lead extensions

9 m (30 ft) H and X extensions 2008-30XKIT3

Optional Hardware Accessories

Calibration Certification TAU3-CAL-CERT

Safety Beacon - 18 m (60 ft) 1004-639

Transit case (for instrument) 1014-928

USB Printer 90029-573

USB Printer Paper (x48 rolls) 90029-573-P

1:1 Test Jig 2005-249

OLTC Tap changer cable adapters 1011-622-A

Motor current monitor* 1014-929

Vibration monitor* 1014-930

Temperature probe kit* 1014-931

TRS1+ Calibration Standard TRS1PLUS

TOS1 Calibration Standard TOS1

Warnings and safety precautions

Safety

Warnings and safety precautions

WARNING!

Death, serious injury, or fire hazard could result from improper use of this instrument. Read and understand this manual before installing this instrument.

Usage of this instrument must comply with the National Electric Code and any additional safety requirements applicable to your country and company policies.

Qualified personnel MUST perform operation and maintenance of this instrument. The National Electrical Code defines a qualified person as one familiar with the construction and operation of the equipment and the hazards involved.

Safety Precautions

Take the following safety precautions whenever the instrument is used:

  • Wear safety glasses and insulated gloves when making circuit connections
  • Hands, shoes, floor/ground must be dry when making any connection to a powered line

These warnings and safety precautions are to be used where appropriate when following instructions in this manual.

CAUTION!

The equipment could be impaired from improper use not specified in this user guide. Read the complete manual before use.

Megger TAU3 - CAUTION! - 1

Megger TAU3 - CAUTION! - 2

CAUTION!

Do not use detachable mains supply cords with inadequate electrical ratings.

Megger TAU3 - CAUTION! - 1

GENERAL SAFETY PRECAUTIONS

Megger TAU3 - GENERAL SAFETY PRECAUTIONS - 1

The TAU3 and the Unit Under Test (UUT) should both be considered as sources of instantaneously lethal levels of electrical energy. Observe the following safety precautions

SAFETY IS THE RESPONSIBILITY OF THE USER

Only qualified and trained operators should operate the TAU3. Operator must read and understand the Instruction Manual prior to operating the equipment.

  • Observe necessary safety precautions on the TAU3.
  • Identify areas of immediate hazard that could result in injury or death.
  • Treat all terminals of high-voltage power equipment systems as potential electric shock hazards.
  • Use all practical safety precautions to prevent contact with energized parts of the equipment and related circuits.
  • Never connect the test set to energized equipment.
  • The ground connection must be the first made and the last removed. Any interruption of the grounding connection can create an electrical shock hazard.
    ● Always disconnect leads from UUT before disconnecting them at the test set.

Max Technical Specifications

Specifications - Valid from -20 ° to +50 °C

Input power

100-240 V AC, 47-63 Hz, 1200 W

±10% Mains supply voltage fluctuations Overvoltage category II

Output power

Voltage 3-phase, 1-100 V

Frequency DC, 40-480 Hz

Current 0.1 mA - 1 A @ 100 V

Current 0.1 mA - 32 A @ 24 V

Regulatory

Safety IEC 61010-1:2010 + AMD1:2016

EMI/EMC IEC 61326-1:2012

RoHS2 EN50581

Vibe/Shock MIL-STD-810G

Ingress protection IP65 (Lid closed)

Transformer testing standards

IEEEC57.152-2013
IEC60076-1:2011

AS/NZS 6076 1:2014

CIGRE 445 2 011

GOST 3484.1-88

Dimensions55.8 x 28.7 x 19 cm
22 × 11.3 × 7.5 in

Weight 15 kg 33 lbs

Case

Rugged case with built in wheels and handle

Backpack lead bag for leads and accessories

Internal/external data storage

Up to 10 000 sets of three-phase results internal storage

Transferable via USB 2.0 drive

Communication/control software

USB Interface for PC Control with custom GUI

Touch screen (optional)

25.6 cm 10.1 in

1024 x 600 Resolution

1000 NITS

Printer (optional)

51 mm (2 in) thermal printer

Prints all measurement data displayed on GUI

Environmental

Operating -20 ° to 50 °C (-4 ° to 122 °F)

Storage -30 ° to 70 °C (-22 ° to 158 °F)

Relative Humidity 0-90 %, non-condensing

Indoor and outdoor use in dry locations

Elevation 2000 m MAX

Pollution degree 2

TTR

Turns ratio measurement methods

3-phase step up
3-phase step down
1 phase step up
1 phase step down

Turns Ratio Range and Accuracy

Step Down Excitation

25-100 V
±0.05 % 0.8 - 1000
±0.10 % 1001 - 2000
±0.30 % 2001 - 15000
±0.60 % 15001 - 50000
1-24 V
±0.10 % 0.8 - 1000
±0.20 % 1001 - 2000
±0.60 % 2001 - 15000

Step Up measurement

25-250 V
±0.05 % 0.8 - 200
(most Power Tx)
1-24 V
±0.10 % 0.8 - 200

Excitation current resolution

Resolution 0.1 mA, 0.1 mA - 100 mA

1.0 mA, 101 mA - 11 A

Excitation current accuracy

±1 % Reading, ±0.1 mA

Frequency accuracy ±1 % Reading, ±0.1 Hz

Phase range Range 0 – 360°

Phase accuracy ±0.05°

Max voltage output 90 VACpk

Technical specifications

Voltage accuracy Typical ±0.1 % reading, ±0.1 mV

Guaranteed ±0.5 % reading

Specified accuracy for external verification only and does not impact AC tests accuracy (TTR, Magnetic Balance, SCI, FRSL, or Losses)

WR

Resistance measurement methods

1 phase wye, delta, zigzag

2 phase wye w/neutral

3 phase wye w/neutral

Dual winding excitation

DC Open circuit voltage

Up to 100 V

DC Measurement voltage

Up to 100 V

Resistance accuracy ±0.10 % reading, ±1 μΩ

Resistance resolution 5 digits

DC voltage accuracy ±0.05 % reading, ±0.1 mV

DC current accuracy ±0.05 % reading, ±0.1 mA

Current and resistance ranges

Typical with 9 m (30 ft) leads

Current Min Ω Max Ω

32 A 1.0 μΩ 400 mΩ

16 A 1 mΩ 1.0 Ω

8 A 1.0 Ω 2.0 Ω

1 A 2.0 Ω 20 Ω

100 mA 1.0 Ω 100 kΩ

Dynamic resistance measurement method

Dynamic voltage

Dynamic current

Dynamic resistance

Dynamic Resistance Speed

Speed 20 kHz

SCI FRSL

Impedance measurement methods

1 ∅

Frequency range 40 – 480 Hz

Impedance measurement range

0.1 Ω - 700 Ω

Impedance accuracy ±1% reading, ±0.10 mΩ

Reactance measurement

range 0.1 Ω - 700 Ω

Reactance accuracy ±1 % reading, ±0.10 mΩ

Inductance accuracy ±1 % reading, ±10 μH

Power factor range 0.1% - 100%

Power factor accuracy

±5 % reading

AC Current accuracy ±0.2 % reading, ±0.1 mA

Efficiency

Core loss measurement

methods

Hysteresis losses

Eddy current losses

Core loss accuracy ±10 % of actual losses

AC copper losses accuracy at 85 °C

DC copper losses accuracy at 85 °C

Motor current measurement (optional)

9 V battery power

Measuring range: 3.0 A/30 A

Frequency range: DC to 60 Hz

Resolution: ± 50 mA / ± 100 mA

Accuracy: ±1 % reading

Temperature probe measurement (optional)

Range: -20 °C to 110 °C

Accuracy: ±1 % reading. ± 1.0 °C

Connections and controls

Output On Indicator Touchscreen USB B PC Control USB A Printer, Data Export 90-264VAC 47-63Hz 1500VA MAX Primary Connections Safety Strobe Temp & OLTC Vibration Sensor OLTC Motor Current Secondary Connections Safety Interlock Safety Key Emergency Stop Rotary Control Knob OLTC Tap Changer OLTC Circuit Breaker OLTC 250V 2A MAX

GENERAL SAFETY PRECAUTIONS The TAU3 and the Unit Under Test (UUT) should both be considered as sources of instantaneously lethal levels of electrical energy. Observe the following safety precautions SAFETY IS THE RESPONSIBILITY OF THE USER Only qualified and tuned operation should serve the TAU3 Operation must read and understand the instruction Manual to specify the equipment. • Observe necessary safety precautions on the TAU3. • Identify areas of immediate hazard that could result in injury or death. • Treat all terminals of high-voltage power equipment systems as potential electric shock hazards. • Use all practical safety precautions to prevent contact with energized parts of the equipment and related circuits. • Never connect the test set to energized equipment. • The ground connection must be the first made and the last removed. Any interruption of the grounding connection can create an electrical shock hazard. • Always disconnect leads from UUT before disconnecting them at the last set. Quick Start Guide Quick Test Select Test All Tests TR & Balance Resistance & Damping Select Phases H2 X1 X2 H1 1U A S1 H2 1V B S2 H3 1W C S3 H0 1N N 30 X1 2U a L1 X2 2V b L2 X3 2W c L3 X0 2N n L0 Power TRANSFORMER CONNECTIONS *Neutral connections not required for all transformers FOR BEST RESULTS UNCOIL LEADS Test Settings Voltage Test Mode Test Type 250V Auto Ratio Temperature & Level Start Test Print Export Save OLTC OLTCC Safety Strobe Temp & OUTC Vibration Sensor OLTC Motor Current OLTC Circuit Breaker OLTC 250V C L OLTC Top Changer MGGER. TAU3 True Three Phase Transformer Analyzer Output On Indicator Touchscreen USB & PC Control USB A Printer Data Export Scan QR code for User Guide and additional product info. 90-364VAC 47-25Hz 1500VA MAX

Before installing PC software, contact your IT department. Your IT department can assist with install and provide administrator approval if required.

Control of the TAU3 is possible from the built-in touchscreen or from a USB connected PC with the PC software installed. To install the PC SW:

  1. Locate the TAU3 PC software installer

a. From the TAU3

i. Connect the TAU3 to a PC with the included USB cable

ii. Turn on the TAU3

iii. After initialization, a CD drive will appear on the PC which contains the software

iv. Locate the file named TAU3_installer_X.xxx.exe. X.xxx is the version.

b. From the internet

i. Go to www.megger.com/TAU3

ii. Download the latest PC installer

  1. Double click to launch the installer

  2. Select a language for the install and click OK.

Installer Language M Please select a language. English OK Cancel

  1. Click Next on the welcome screen

TAU3 GUI 1.0.0.49 Welcome to TAU3 software Setup Setup will guide you through the installation of TAU3 software. It is recommended that you close all other applications before starting Setup. This will make it possible to update relevant system files without having to reboot your computer. Click Next to continue. Megger. Next > Cancel

PC software installation

5. Review the license agreement and click I Agree

TAU3 GUI 1.0.0.49 License Agreement Please review the license terms before installing TAU3 software. Megger. Press Page Down to see the rest of the agreement. MEGGER SOFTWARE LICENSE AGREEMENT READ THE FOLLOWING TERMS AND CONDITIONS BEFORE INSTALLING THE SOFTWARE. IF YOU DO NOT AGREE WITH THEM, PROMPTLY RETURN THE SOFTWARE AND THE ACCOMPANYING ITEMS (INCLUDING MANUALS) TO: MEGGER 2621 VAN BUREN AVE If you accept the terms of the agreement, click I Agree to continue. You must accept the agreement to install TAU3 software. Nullsoft Install System v3.03 < Back I Agree Cancel

TAU3 GUI 1.0.0.49 Choose Components Choose which features of TAU3 software you want to install. Check the components you want to install and uncheck the components you don't want to install. Click Next to continue. Select components to install: TAU3 software Create desktop shortcut Space required: 104.2 MB Description Position your mouse over a component to see its description. Nullsoft Install System v3.03 < Back Next > Cancel

TAU3 GUI 1.0.0.49 Choose Install Location Choose the folder in which to install TAU3 software. Setup will install TAU3 software in the following folder. To install in a different folder, click Browse and select another folder. Click Install to start the installation. Destination Folder C:\Program Files (x86)\Megger\TAU3 Browse... Space required: 104.2 MB Space available: 25.6 GB Nullsoft Install System v3.03 < Back Install Cancel

  1. Click Finish to complete the install.

TAU3 GUI 1.0.0.49 Completing TAU3 software Setup TAU3 software has been installed on your computer. Click Finish to close Setup. Megger. < Back Finish Cancel

PC software update

When updating to a new version of the TAU3 software, the installer will remove the installed version of software.

  1. Locate the updated TAU3 PC software installer

a. From the TAU3

i. Connect the TAU3 to a PC with the included USB cable

ii. Turn on the TAU3

iii. After initialization, a CD drive will appear on the PC which contains the software

iv. Locate the file named TAU3_installer_X.xxx.exe. X.xxx is the version.

b. From the internet

i. Go to www.megger.com/TAU3

ii. Download the latest PC installer

  1. Double click to launch the installer

  2. Click OK to remove the previous TAU3 version

TAU3 GUI 1.0.0.49 TAU3 software is already installed. Click 'OK' to remove previous version and install new version, or 'Cancel' to cancel this upgrade. OK Cancel

  1. Click Next on the welcome screen

M TAU3 software Uninstall Welcome to TAU3 software Uninstall Setup will guide you through the uninstallation of TAU3 software. Before starting the uninstallation, make sure TAU3 software is not running. Click Next to continue. Megger. Next > Cancel

5. Click Next on the uninstall TAU3 software screen

TAU3 software Uninstall Uninstall TAU3 software Remove TAU3 software from your computer. TAU3 software will be uninstalled from the following folder. Click Next to continue. Uninstalling from: C:\Program Files (x86)\Megger\TAU3\ Nullsoft Install System v3.03 < Back Next > Cancel

6. Click Uninstall

TAU3 software Uninstall Choose Components Choose which features of TAU3 software you want to uninstall. Check the components you want to uninstall and uncheck the components you don't want to uninstall. Click Uninstall to start the uninstallation. Select components to uninstall: Uninstall Description Position your mouse over a component to see its description. Space required: 0.0 KB Nullsoft Install System v3.03 < Back Uninstall Cancel

7. Click Finish

TAU3 software Uninstall Completing TAU3 software Uninstall TAU3 software has been uninstalled from your computer. Click Finish to close Setup. Megger < Back Finish Cancel

8. Proceed with the installation instructions from PC Software Installation

PowerDB installation

PowerDB Installation

PowerDB can import TAU3 data. In addition, PowerDB can control the TAU3 with the optional SW-POWERDB license. To install PowerDB, download the latest version from www.powerdb.com. Follow the instructions on screen to install PowerDB.

After turning on the TAU3, the Megger logo will display, followed by a progress screen during system initialization.

Home Screen

After the TAU3 has initialized, the display will show the home screen.

Megan Instrument 01.1.2-de-54 About TAU3-EXP Quick Test Template Test Results/Reports Asset Template Settings 03/15/2023 03:43:22 PM

There are six options to select from the Home Screen

  • About
  • Settings
  • Quick Test
  • Results
  • Asset Template
  • Template Test

About

The about screen provides detailed information about the TAU3.

TAU3-EXP 03/22/2024 09:05:59 PM GUI SW Version 1.1.0-dev101 PowerDB Control Instrument SW Version 1.1.0-dev109 Magnetic Balance HIV Version SIM0000-0000-LF.00 Phase Shifting BSP Version PC Diagnostic Firmware Version 1.0.0-sim Automatic OUTC Serial P SIM0000-0000-LF Custom App Control Export Logs Updates Device Self-check User Guide SW VERSION 1.1.0-dev101

In addition, the following options are available:

- Export Logs

■ Used to provide Megger with detailed information about testing performed

- Updates

■ Used to update the TAU3 to a different software version

- Device Self-check

- Used to perform hardware verification of the TAU3

- User Guide

- Access the user guide of the TAU3

The settings screen provides options to configure the TAU3 to meet regional and testing needs

TAU3-EXP Measurements & Limits Regional OLTC Control Display Administrative & Maintenance 03/27/2024 11:26:57 AM

There are five options on the settings screen

- Measurement & Limits

- Configure nameplate and set limits for output and measurement verification

- Region

- Configure the TAU3 to the region it is being used

- OLTC Control

- Configure the tap changer controller for on load tap changers

- Display

■ Adjust brightness of the display

- Administrative & Maintenance

■ Perform lead calibration and restore factory settings

Measurements & Limits Settings

The measurement & limits screen is broken into ten different sections. Swiping left or right or using the left and right arrows navigate to the different setting sections.

TAU3-EXP Back 11/06/2022 12:05:02 PM General Use Frequency 60 Hz Test Buffer ✓ Verify Connection ✓ Auto Confidence ✓ One Touch OTCC ✓ Color Rating ✓ Use Result ✓ Show USB Touch ✓ Nameplate Standard ANS (11.3) Primary Vector Secondary Vector Primary Top Changer Secondary Top Changer DETC DETC X ✓

General

General settings apply to all tests performed by the TAU3.

- Line Frequency

■ Set to 50 Hz or 60 Hz

- Test Buzzer

■ Enable or disable a countdown before each test

- Verify Connection

■ Enable or disable connection verification before each test. Disable if only one side of the transformer is connected to the TAU3

- Auto Continue

- When enabled, the TAU3 will boot to the last test performed. When disabled, the TAU3 will boot to the home screen

- One Touch OLTC

- When enabled, the One Touch OLTC option will be enabled by default when performing Template Tests that have an OLTC

- Color Rating on Results

■ Enable to emphasize measurements that exceed limits

- Show USB Thumb Drive

- Enable to display a USB Thumb Drive when connected to a PC. Disable to prevent the USB Thumb Drive from displaying

- Output Indicator Brightness

■ Change the output brightness of the warning indicator

- Export Folder

- Change the default folder where results are exported. This option is only available when controlling the TAU3 from a PC.

Configure the default nameplate used when performing tests

- Standard

- Choose from ANSI, IEC, Australian, or GOST

- Primary Vector

- Choose the default primary vector

- Secondary Vector

- Choose the default secondary vector

- Primary Tap Changer

- Choose the default primary tap changer. DETC, OLTC, or NA

- Secondary Tap Changer

- Choose the default secondary tap changer. DETC, OLTC, or NA

- Primary Tap Labels

- Choose the default primary tap labels

- Secondary Tap Labels

- Choose the default secondary tap labels

- Primary # of Taps

- Choose the default primary number of taps. If associated tap changer set to NA, this will be NA

- Secondary # of Taps

- Choose the default secondary number of taps. If associated tap changer set to NA, this will be NA

- Primary Winding Material

- Choose the default primary winding material. Copper or Aluminum. Selection will impact temperature correction for winding resistance, short circuit impedance, and efficiency tests

- Secondary Winding Material

- Choose the default secondary winding material. Copper or Aluminum. Selection will impact temperature correction for winding resistance, short circuit impedance, and efficiency tests

- Round Tap Voltages

- Choose the rounding used for tap calculations. 1 V, 5 V, or 10 V.

- Winding Temperature

- Choose the default winding temperature used for temperature correction.

- Correct winding to temperature

- Enable or disable. When enabled, winding resistance, short circuit impedance, and efficiency tests will be automatically corrected to the specified temperature.

- Auto Transformer / VReg Labels

- Choose the labels for autotransformers. Standard naming convention (ANSI, IEC, Australian, or GOST) or S,L

- Tap Phase Shift

- Choose to enable or disable tap phase shifting. Used rarely

1P Primary/secondary connections

- Choose which leads to use for 1P transformer. H1-H2/X1-X2 or H1-H0/X1-X0 options available.

Turns Ratio

- # OLTC Nominals

- Set the number of OLTC nominal. Used rarely.

Configure the turns ratio settings used when performing tests

- Max Test Voltage

- Set maximum induced voltage limit

■ Maximum induced voltage will be limited based on model

- Mode

- Set default test mode:

■ AUTO (recommended)
30 up
30 down
10 up
10 down

- 30 modes not available in the TAU3-BASIC

- Ratio Display

■ Select TTR (line to ground) or TNR (line to line)

- Ratio Error Limit

■ Set the turns ratio evaluation criteria. 0.5% recommended.

- Phase Deviation Display

- Set the default phase deviation display. Absolute will show total deviation of the measured phase compared to the excited phase. Relative will show total deviation of the measured phase compared to the excited phase less the expected deviation of the measured phase.

- Phase Unit

- Choose from degrees °, minutes ", or ms

- Phase Deviation Limit

■ Set the default phase deviation limit. Recommended 0.5°

Magnetic Balance

Configure the magnetic balance settings used when performing tests

- Imbalance Limit

■ Set imbalance limit. Recommended 3.0%

- Show on Quick Test

- When enabled, magnetic balance will be automatically performed with each turns ratio quick test

- Show on Template test

- When enabled, magnetic balance will be automatically performed with each turns ratio template test

- Preferred Winding

- Select which winding to perform magnetic balance. Auto recommended. When enabled, Auto mode will use YN windings for magnetic balance when present. If no YN winding, Auto mode will use the primary winding.

Configure the winding resistance settings used when performing tests

- Max Current

  • Set the maximum current limit. Applies to primary and secondary windings
    ■ Maximum current will be limited based on model

- Stability %

- Set the stability criteria for winding resistance tests.

- Stability time

- Set the time the stability % must be maintained to indicate a measurement is stable

- Auto Save

  • When enabled, results will automatically save when stability time and % criteria have bene met
  • When performing winding resistance on multiple OLTC taps in template tests, Auto Save will be disabled for the first tap of each phase. After saving the first tap, the test will automatically save for the remaining taps for that phase.

- Max Winding Difference

- Set the maximum resistance difference evaluation criteria. Recommended 3.0%

- Winding Difference Calculation

  • ANSI/IEC
  • Diff % = ((Rmax - Rmin) / Ravg) x 100

- CEE

  • Determine |Max Phase Error| from
  • AVG = (Phase 1 + Phase 2 + Phase 3)
  • Phase 1 Error = Phase 1 - AVG
  • Phase 2 Error = Phase 2 - AVG
  • Phase 3 Error = Phase 3 - AVG

  • Diff % = (|Max Phase Error| / AVG)
    * 100

- Dual Winding Excitation

- When enabled, primary and secondary will be energized and measured simultaneously

- Make Before Break

- Set current sag (10%) time limit during OLTC transitions. 5 ms recommended.

- Hard Break

- Set limit for hard break during winding resistance tests.

200 ms recommended. This setting is used to protect the operator in the event of inadvertent lead disconnection during winding resistance tests

Demagnetization

Configure the demagnetization settings used when performing tests

- Phase

■ Select a phase or all phases. Recommended B Phase

- Auto Demag After Resistance

- When enabled, the TAU3 will automatically demagnetize the transformer after a resistance test has been performed.

Efficiency

Configure the efficiency settings used when performing tests

Not available in the TAU3-BASIC

- No-load mode

- Choose between fast or full. Fast will double the eddy current losses and Full will use eddy current losses and hysteresis losses.

- Show on Quick Test

- When enabled, the efficiency test will appear when performing quick tests

- Show on Template Test

- When enabled, the efficiency test will appear when performing template tests

- Correct to Load %

- Choose the default load % used for correction of efficiency results

Short Circuit Impedance

Configure the short circuit impedance settings used when performing tests.

Not available in the TAU3-BASIC

- % Error Limit

- Set the evaluation criteria used for short circuit impedance tests. Recommended 3.0%

Impedance

- Choose the method of entering nameplate impedance, either in Z % or Z Ω.

Frequency Response Stray Losses

Configure the frequency response stray losses settings used when performing tests

Not available in the TAU3-BASIC

- Frequencies to Test

- Select the frequencies used for frequency response stray losses tests

- Show on Quick Test

- When enabled, the frequency stray losses tests will be automatically performed with each short circuit impedance tests.

- Inductance error limit

- Set the evaluation criteria used for inductance. Recommended 2.5%

- Rs error limit

- Set the evaluation criteria used for Rs. Recommended 15%

Select cancel to discard changes. Select accept to save changes.

× ✓

Select back to return to the settings screen.

Megger TAU3 - Frequency Response Stray Losses - 2

Select home to return to the home screen.

Megger TAU3 - Frequency Response Stray Losses - 3

Select the preferred language, number format, date format, and time format for the TAU3. Date and time can be set on the TAU3. When controlled by a PC, the TAU3 will use the PC date and time.

Megger Instrument SU1.1.0-dn1 Back TAU3-EXP Language English Number Format 1,000.00 Date Format MM/DD/YYYY Time Format 12H X ✓

Select cancel to discard changes. Select accept to save changes.

× ✓

Select back to return to the settings screen.

Megger TAU3 - Frequency Response Stray Losses - 6

Select home to return to the home screen.

Megger TAU3 - Frequency Response Stray Losses - 7

OLTC Control

Select continuous press or pulse/wait operation for the on load tap changer. The majority of transformers use a pulse/wait.

When pulse/wait is selected, options for setting pulse and wait time will be displayed. Pulse is the duration required to initiate a tap change. Wait is the amount of time the tap changer requires to complete the tap changer after initiation.

Magyar Instrument GB 1.0 Order1 Back TAU3-EXP Continuous Press Pulse/Wait Pulse 1.0 Wait 1.0 Top Timing X ✓ 11/05/2022 09:32:40 AM

Select tap timing to use the OLTC controls on the TAU3 to set the pulse and wait time.

Tap Timing

Back TAU3-EXP 11/05/2022 09:36:45 AM Tap Timing - Use the provided tap changer cable and TTRU3 OLTC switch on the front panel to configure the Pulse and Wait times of the OLTC transition. 1. Press & hold TTRU3 OLTC tap changer switch on front panel (up or down) to start tap changer transition. 2. Release TTRU3 OLTC tap changer switch to capture pulse time. 3. Press 'tap transition complete' button when tap transformation is complete (MAX 10s wait). 0.0s 0.0s Tap Transition Completed ✓

Select cancel to discard changes. Select accept to save changes.

× ✓

Select back to return to the settings screen.

Megger TAU3 - OLTC Control - 5

Select home to return to the home screen.

Megger TAU3 - OLTC Control - 6

Display

Use the slider to adjust the brightness of the TAU3 display

Administrative & Maintenance

Administrative & Maintenance

After clicking on the Administrative & Maintenance setting, a password dialog box will appear. Enter the password 2621 to proceed.

TAU3-EXP Measurements Limits OLTC Control Enter password to access Administrative & Maintenance settings Display X ✓

Lead calibration is used for accurate impedance and efficiency test results. A one time lead calibration is typically required when using leads different than the default 9 m (30 ft) calibration. Follow testing best practices when calibrating the leads by fully uncoiling and extending leads.

Restore factory settings is used to reset the TAU3 back to the original settings from the factory. User will be prompted to confirm

TAU3-EXP Back Calibrate Leads Restore Factory Settings Go Back 03/22/2024 09:45:39 PM

Quick Test

After selecting quick test, test options will display on the left hand side of the screen. Enter a Test ID for test identification.

Mogger Instrument GUI 1.0-dev1 TAU3-EXP 11/04/2022 02:19:28 PM Select Test Confirm Vector TTR & Balance Resistance & Demag Short Circuit Impedance & FRSL Efficiency All Tests Select a test to get started Test ID:

Select home to return to the home screen.

Megger TAU3 - Quick Test - 2

Confirm Vector

Confirm Vector

Confirm vector automatically detect the vector of the connected transformer. Recommend test settings selected by default.

Magnet instrument GUI 1.1.0-dev1 TAU3-EXP 11/04/2022 02:19:38 PM Select Test Confirm Vector TTR & Balance Resistance & Demag Short Circuit Impedance & FRSL Efficiency All Tests Start Confirm Vector Test Settings Recommended Measurements & Limits Voltage 250 V Text ID

Click Start Confirm Vector to begin the test

Start Confirm Vector

Mogger Instrument GB3 1.1.0.dsn1 Find vector finished! Found: Dyn11 Accept 11/04/2022 02:19:51 PM Select Test Confirm Vector TTR & Balance Resistance & Demag Short Circuit Impedance & RSSL Efficiency All Vets Confirm Vector Found vector: Dyn11 Reject Accept Test Settings Recommended Voltage 250 V Measurements & Limits Text ID:

When confirm vector completes, the vector found will be displayed. Verify the vector matches the nameplate of the transformer under test and select Accept. If the vector does not match the nameplate, select Reject.

It is the responsibility of the user to confirm that the vector matches the transformer under test. Megger accepts no responsibility for incorrectly identified vectors.

Clicking accept will select the vector found and progress quick test to TTR test setup. Clicking reject will progress to the primary vector selection screen.

Select a primary vector or autotransformer from the Primary vector select screen.

TAU3 - EXP Back Primary Single Phase 1P CT PT 1Pa0 / VREG Secondary Three Phase D Y YN Z Three Phase Auto Ya0 / VREG YNa0 / VREG 03/22/2024 09:56:37 PM

After selecting a primary vector, The secondary vector select screen will display. Select a secondary vector group and phasing. Once selected, the TTR test setup will be displayed. Select find to find the vector phasing for the selected primary and secondary group.

Magyar Instrument GB 1.1.2-dev1 TAU3-EXP 11/04/2022 02:21:06 PM Back Primary O Secondary Vector Group d y yn z zn Secondary Vector Phasing yn1 yn3 yn5 yn7 yn9 yn11 Ø Shift Find

Vector, Phases, and Nameplate

Select TTR & Balance, Resistance & Demag, Short Circuit Impedance & FRSL, Efficiency, or All Tests to display the Select Vector, Select Phases to Test, and Select Taps to Test sections.

TAU3-EXP 11/04/2022 02:23:51 PM Select Test Confirm Vector TTR & Balance Resistance & Demag Short Circuit Impedance & FRSL Efficiency All Tests Select Vector Primary Dyn1 H2 X2 H1 X0 Secondary in Start -30.0° Select Steps to Test Test Settings Top Change Tap 1-L Voltage Rated I 2 % Primary DETC 3 0 0.0 None 0.00 Secondary DETC N 0 0.0 None 0.00 Recommended Measurements & Limits Voltage Test Mode 250 V Auto

Select Vector and Select Phases to Test

Select Vector and Select Phases to Test

Select Primary to view the primary vector options for the transformer under test. Select Secondary to view the secondary vector options for the primary vector of the transformer under test. Select the Secondary button to change to Tertiary if testing a tertiary winding. Phase shift enables testing of phase shifting transformers. Phase shifting test capabilities not available on the TAU3-BASIC.

Megger TAU3 - Select Vector and Select Phases to Test - 1

flowchart
graph TD
    A["H1"] --> B["H2"]
    B --> C["H3"]
    D["X0"] --> E["X1"]
    E --> F["X2"]
    G["X3"] --> H["X2"]
    style A fill:#fff,stroke:#000
    style B fill:#fff,stroke:#000
    style C fill:#fff,stroke:#000
    style D fill:#fff,stroke:#000
    style E fill:#fff,stroke:#000
    style F fill:#fff,stroke:#000
    style G fill:#fff,stroke:#000
    style H fill:#fff,stroke:#000

Individual phases can be enabled or disabled for testing. When testing ratio, impedance, and efficiency, in phase primary and secondary phases are tied together. For example, disabling phase A and C on the primary will disable phase A and C on the secondary.

Megger TAU3 - Select Vector and Select Phases to Test - 2

chemical Dyne 1 molecular structure diagram showing primary and secondary states with labeled hydrogen atoms H1–H3 and X0–X3, including directional arrows indicating reaction or interaction.

Nameplate

The nameplate section will populate with default settings from Measurement & Limits. Enter tap information for the transformer for evaluation against limits. Secondary can be changed to Tertiary. Base KVA can be changed to Base MVA.

Entering tap voltages for two taps of a winding will automatically calculate tap voltages for the remaining taps of the winding.

Entering L-L voltage and base KVA/MVA will calculate RATED I for the transformer.

Megger TAU3 - Nameplate - 1

other Select Taps to Test | Tap | Change | Tap | Voltage | Base KVA | Rated I | Z % | |---|---|---|---|---|---|---| | Primary | DETC ●● | 3 ●●●●● | 0 | 0.0 | None | 0.00 | | Secondary | DETC ●● | N ▼ | 0 | 0.0 | None | 0.00

Turns Ratio & Balance

Turns Ratio & Balance

Voltage and Test Mode settings display when Turns Ratio & Balance is selected.

If Magnetic Balance is disabled in settings, only TTR will display in the Select Test section.

Select Measurement & Limits for complete settings options.

Test Settings Recommended Measurements & Limits Voltage 250 V AC Test Mode Auto

Test Mode Auto will determine the best method for testing the transformer: 3∅ or 1∅, step up or step down. Changing the test mode fill force the TAU3 to test in a specific mode, which can be useful for troubleshooting transformer problems.

Click Start Test to start the TTR & Balance test.

Start Test

The TTR & Balance test in progress screen will be shown after starting the test. Test voltage and current are displayed, along with a progress bar.

TTR & Balance Dyn1 3-N Turns Ratio In Progress 11/04/2022 02:49:28 PM Stop Test Winning Top I-V Voltage Current 1 2% Primary 3 4856 106.0 30.7 Secondary N 290 100.0 753.1 1.45 Test ID: 221104144924

When the TTR & Balance test is complete, the following screen will display. Phases that exceed limits will be highlighted in red. Nameplate information can be edited if necessary. Ratio display, % Error, and Phase Deviation limits can be adjusted if desired.

Megger TAU3 - Turns Ratio &amp; Balance - 4

other Dyn1 3-N Turns Ratio Result | Phase | Rate Value TTR (%) | % Error 0.50 % | HV | mA | Dev 0.5° | |---|---|---|---|---|---| | H1-H3 X1-X0 | 36.663 | 0.49 | 250.0 | 205.5 | -0.22 | | H2-H1 X2-X0 | 36.698 | 0.58 | 250.0 | 223.7 | 0.15 | | H3-H2 X3-X0 | 36.679 | 0.53 | 250.0 | 203.4 | 0.19 | | Test Mode Up 3 Phase | Calculated | | :--- | :--- | | Test Voltage 250.0V | 36.486 | | Primary | Top Rating (Hz) | L-L Rating (Hz) | Rang KVA Rating (Hz) | Rated I Z (%) | |---|---|---|---|---| | 1 | 3 | 4845 | 100.0 | 35.7 | | 2 | 1*+1** | | | | | 3 | N/A | 230 | 300.0 | 723.1 | | 4 | | | | | | 5 | | | | | | 6 | | | | | | 7 | | | | | | 8 | | | | | | 9 | | | | | | 10 | | | | | | 11 | | | | | | 12 | | | | | | 13 | | | | | | 14 | | | | | | 15 | | | | | | 16 | | | | | | 17 | | | | | | 18 | | | | | | 19 | | | | | | 20 | | | | | | 21 | | | | | | 22 | | | | | | 23 | | | | | | Total: Test ID: Z=300315168115 Test Setup: 03/15/2023 04:33:58 PM Output Save

Ratio display has two options. Choose the option that is consistent with your testing practices:

  • TTR, line to neutral turns ratio
  • TNR, line to line turns ratio

Selecting retest will rerun the test.

If the optional 2" printer is connected, the option to print will be displayed. The results will automatically be saved after print is selected.

From the 10" display, connect a USB drive to export results. Selecting export from the PC will export to the folder designated in settings. The result will be automatically saved after exporting.

Click Save to save the result and continue with the next test.

Select TTR & Balance to change to the Balance screen.

TTR & Balance €

Phases that exceed imbalance limits will be highlighted in red. Phases that exceed balance limits can be an indication that the transformer is magnetized.

Turns Ratio & Balance

Megger TAU3 - Turns Ratio &amp; Balance - 1

bar Dyn1 3-N Magnetic Balance Result | Phase | Value (%) | |---|---| | X1-X0 | 95.7 | | X2-X0 | 98.7 | | X3-X0 | 100 | | Total Induced LIME: 63.0% | 95.5 | | Winding Top 1-5 Voltage Gain X2A Rate/ F % | 97.6 | | Primary 3 4056 300.0 31.7 5.45 N = 230 300.0 738.1 3.43 Test ID: 2211041445524 TIR & Balance Jlc Summary 11/06/2022 02:53:22 PM Start

Winding Resistance & Demagnetization

Voltage, Primary Current, Secondary Current, Stability %, Stability Time, and Auto Save settings display when Winding Resistance & Demagnetization is selected.

Select Measurement & Limits for complete settings options.

Test Settings Recommended Measurements & Limits Voltage 100 V DC Primary A 5.0 A DC Secondary A 5.0 A DC Stability % 99.80% Stability Time 5s Auto Save ✓

Primary test current is determined by the RATED current of the primary winding. The TAU3 will select 15% of rated current, or the maximum output of the TAU3, whichever is lower. TAU3 BASIC maximum primary current is 10 A. TAU3 ADV, PRO, and EXP maximum primary current is 16 A.

Like primary test current, secondary test current is determined by the RATED current of the secondary winding. The TAU3 will select 15% of rated current, or the maximum output of the TAU3, whichever is lower. TAU3 BASIC maximum secondary current is 10 A. TAU3 ADV maximum secondary current is 16 A. TAU3 PRO and EXP maximum secondary current 32 A.

Click Start Test to start the winding resistance test.

Start Test

Click Demag to perform demagnetization

Demag H1-H2

The winding resistance test in progress screen will display after clicking start test.

Phases for windings that have yet to be tested will be displayed in red (H1-H2 below)

Phases that are in progress will show an excitation symbol around the phase (H2-H3 below). Resistance values will be displayed next to phases that are in progress and will continuously update until the resistance measurement is saved.

Phases saved are displayed in black (H1-H3).

Megger TAU3 - Winding Resistance &amp; Demagnetization - 4

other | Category | Value | |---|---| | Test In Progress | 2.3776 Ω | | Test Completed | 2.2585 Ω | The chart displays a triangle with three vertices labeled H1, H2, and H3. The legend indicates red symbols: 'Test In Progress', black symbol: 'Test Completed', and solid red line: 'To be Tested'. An arrow points from H1 to H2 to H3, suggesting a transition or change in the measured variable.

Winding Resistance & Demagnetization

Resistance, current, stability %, and stability time will be displayed for each winding. When the stability % exceeds the set limit, a timer will start. When the timer exceeds the setting for both windings, the Save button will turn green. If Autosave is enabled, the result will automatically save, and the next phase will be tested. If no more phases remain to be tested, the winding resistance test will finish by discharging the transformer. Discharge is indicated by orange/red lights of the output indicator.

Do not disconnect the leads from the transformer or TAU3 during discharge. In the event of power loss during winding resistance tests or discharge, the TAU3 will continue to discharge the transformer passively, indicated by red LEDs on the output indicator. When discharge is complete, the LEDs will turn off, and the leads can be removed.

Migger Instrument GUI 1.1.0-dev1 TAU3-EXP Dyn1 3-N Resistance In Progress Resistance & Demag 2.3776 Ω 2.2585 Ω 453.64 mΩ 477.68 mΩ 7.6 A Current, A DC Stability Save 98.05 % P Stability 7.6 A Current, A DC Stop Test Test ID: 221104144924 Winding Tap L-L Voltage Save EVA Rated I Z % Primary 3 4856 100.00 35.7 5.49 Secondary N 230 101.8 753.1 5.49 Stability 99.80% Sa Auto Save ✓

If Auto Demag after Resistance is enabled, or if Demag is selected from the Resistance & Demagnetization screen is clicked, the Demag in progress screen will be displayed.

Demag is performed by applying +DC voltage and measuring the time it takes for the transformer core to begin saturation. Polarity is then reversed, and time is measured until saturation occurs in the opposite direction. Based on these initial tests, +DC and -DC voltage is applied in decreasing time intervals until the transformer is demagnetized.

Megger Instrument GB(1.1.0 dev) Test [221104144924 - Resistance] was complete successful! Accept 11/04/2022 03:31:29 PM Resistance & Demag Dyn1 3-N Demagnetization Result In Progress Phase Extraction H1-H2 50 V Voltage V DC Current A DC Winding Tap L-A voltage New EVA Rated I 2 % Primary 3 4856 106.0 $5.7 5.49 Secondary N 238 100.1 733.1 5.49 Stop Test Test ID: 221104144924

When the winding resistance test is complete, the following screen will be displayed. Phases that exceed limits will be highlighted in red. The software will recommend to retest these individual phases.

Corrections to a set temperature can be enabled or disabled. Ensure that winding temperature and primary and secondary winding material are set correctly for accurate temperature correction.

Winding Resistance & Demagnetization

Mugger Instrument 0.8 1.0-dw42 Test Setup Dyn1 3-N Resistance Result @20°C 9.0734 Ω 9.0146 Ω 7.8968 Ω 1.1898 Ω 2.1859 Ω 2.1758 Ω Secondary Difference 13.6% 13.6% Difference Secondary Test Amp 7.4 Alt Winding Winding Sec Winding Corrections 14.0 Stability 86.7% Aluminum 20.0°C Aluminum Off 170.0°C Stability 86.7% Top 5-6 Voltage Bag KVA Rated 1 2 % Primary 3 0 0.0 None 0.00 Secondary N/A 0 0.0 None 0.00 Retest H1-H3 X1-X0 Retest all phases Damage H1-H2 Energe Save Test IO: 230315111114

Selecting Resistance & Demag will display the demagnetization graph

Resistance & Demag Ω AC

The phase that was demagnetized, along with before and after remanence and demag time will be displayed.

Megger TAU3 - Winding Resistance &amp; Demagnetization - 3

line | Phase | Before | After | Time | |-------|--------|-------|------| | H1-H2 | 21.29% | 3.20% | 0.0s | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |

Short Circuit Impedance

Short Circuit Impedance / Leakage Reactance and Frequency Response Stray Losses

Shorting options will be displayed when Short Circuit Impedance / Leakage Reactance and Frequency Response Stray Losses is selected. Test Current cannot be changed.

Test Settings Recommended Measurements & Limits Current 2.5 A AC Shorting Internal

When shorting is set to external, an option to display a shorting diagram will be displayed

Test Settings Recommended Measurements & Limits Current 2.5 A AC Shorting External External Shorting Diagram

External shorting diagram *Neutral connections not required for all transformers H1 1U A S1 H2 1V B S2 H3 1W C S3 * H0 1N N S0 Short X1 - X2 - X3 2U - 2V - 2W a - b - c

The Short Circuit Impedance & FRSL test in progress screen will be shown after starting the test. Test voltage and current are displayed, along with a progress bar.

Mogger Instrument GUI 1.0-dcr1 TAU3-EXP Dyn1 3-N Short Circuit Impedance In Progress Short Circuit Impedance & FRSL 30 A 240 V Voltage, V AC Max 0.0 A Current, A AC Step Test Windings Tap 1-L Voltage Gain 5A Rated 1 2 % Primary 3 4856 536.4 95.7 1.45 Secondary N 230 308.6 753.3 1.45 Test ID: 221104144924

When Short Circuit Impedance & FRSL test is complete, phases that exceed limits will be highlighted in red.

Winding temperature and temperature correction can be adjusted if necessary. Refer to the transformer nameplate to determine the temperature rise of the transformer and add this value to 20° C.

Migger Instrument GUI 1.1.0-dyn1 Test Setup Short Circuit Impedance & FRSL Summary Dyne1 3-N Short Circuit Impedance Result (805°C) Z, % % Error 2.00 % 3 Phase H1-H2-H3 X1-X0 H1-H2-H3 X2-X0 H1-H2-H3 X3-X0 Test Amp 0.2 Frequency 50 Hz Winding °C 19.0°C Correctors On 85.0°C Display 2 % Winding Top 1-4 Voltage Rated 2 % Primary 3 4856 360.0 33.7 100.00 Secondary N* 230 360.0 733.3 100.00 Test ID 221104154606

Clicking on Display will show detailed impedance test information.

Megger TAU3 - Short Circuit Impedance / Leakage Reactance and Frequency Response Stray Losses - 6

Megger TAU3 - Short Circuit Impedance / Leakage Reactance and Frequency Response Stray Losses - 7

other Dynl J-N Short Circuit Impedance Result @81°C | Phase | Z, Ω | Xs, Ω | L, H | PF, % | Rs, Ω | |---|---|---|---|---|---| | 3 Phase | 231.46 | 226.19 | 0.60000 | 21.382 | 48.662 | | H1-H2-H3 | 194.64 | 188.50 | 0.50000 | 24.929 | 48.523 | | X1-X0 | 231.37 | 226.19 | 0.60000 | 21.032 | 48.662 | | H1-H2-H3 X2-X0 | 268.37 | 263.89 | 0.70000 | 18.184 | 48.800 | | H1-H2-H3 X3-X0 Test Amp. | 6.2 | - | - | - | - | | Frequency 60 Hz | 19.0°C | - | - | - | - | | Rotation (On) | - | - | - | 85.0°C | - | | Display (ZΩ) | - | - | - | - | - | | Wiring: Top L-L Voltage: Gain (V) Rated: R/L % Primary: 3 4854 300.0 15.7 100.00 Secondary: N 230 304.0 751.1 100.00 Test ID: Z21104134496

Clicking on Short Circuit Impedance & FRSL will change to FRSL display

Short Circuit Impedance & FRSL 3Ω Summary

Inductance or Rs limits that exceed limits will be highlighted in red.

Megger TAU3 - Short Circuit Impedance / Leakage Reactance and Frequency Response Stray Losses - 9

line | Time | Value | |------|-------| | 0.0 | 41.7 | | 15.0 | 41.7 | | 163.4| 41.7 | | 185.0| 41.7 | | 225.6| 41.7 | | 275.0| 41.7 | | 316.0| 41.7 | | 386.0| 41.7 | | 443.6| 41.7 | | 496.0| 41.7 | | 150.0| 41.7 |

Efficiency

Efficiency

No settings are available for efficiency tests. Click Start Test to begin efficiency tests. The efficiency test in progress screen will be shown after starting the test. Test voltage and current will be displayed, along with a progress bar.

TAU3-EXP Dyn1 3-N Efficiency In Progress Efficiency 247 V Voltage, V AC 0.3 A Current, A AC Stop Test Winding Tap 1-5 Voltage Max 2 % Primary 3 4896 105.0 55.7 0.43 Secondary N 230 103.0 753.1 0.43 Test Id: 221104154406

When the efficiency test completes, the following screen will be displayed.

Primary winding material, secondary winding material, winding temperature, and correction to winding temperature and load % can be adjusted.

Megger TAU3 - Efficiency - 2

All Tests

When selecting all tests, the following options will display.

Select Measurement & Limits for complete settings options.

Test Settings Recommended Measurements & Limits Voltage 250 V AC Primary A 5.0 A DC Secondary A 5.0 A DC Stability % 99.80% Stability Time 5s Auto Save

Depending on the TAU3 model, software features enabled, and tests enabled in settings, clicking start test will progress through tests in the following order:

  • Turns Ratio
  • Magnetic Balance
  • Winding Resistance
  • Demagnetization
  • Short Circuit Impedance
  • Frequency Response Stray Losses
  • Efficiency

When testing is complete, a report card for the completed tests will be displayed.

Tests that exceed limits will be highlighted in red if color coded results are enabled..

Clicking on a test card will take the software to the phase by phase results for the test selected.

Megris Instrument GUI 1.12 Dev1 Summary Test Setup Dyn1 3-N Max Deviations Turns Ratio (TTR) Nameplate 21.113 Measured 21.174 Error 0.29% Resistance @85°C Primary Secondary 2.8416 Ω 571.18 mΩ 0.2% Diff 0.1% Diff Demag Transformer Demagnetized Nameplate 5.4% Measured 147.2% Error 2601.5% Short Circuit Impedance Z% @85°C Magnetic Balance Efficiency @85°C Imbalance 0.00000% Total Losses 117056 W @100% Tx Efficiency 71.93% Load FRSL Max Diff L, H 11.1% Rs, Ω 0.2% Test ID: 221104154406

Save, Print, Export, Retest, Home, Test Setup

Save, Print, Export, Retest, Home, Quick Test Setup

Quick test results are automatically saved at the end of the test. If the namplate or test results settings are changed, a save button will appear. Click save to update the results with the new setting and return to the test setup screen.

Print will print to the optional 2" printer. Print will only display when the printer is detected

Export will export results to a USB thumb drive. Export will only display when a thumb drive is detected, or when operating the TAU3 from a USB connected PC.

Print Export Save

After completing a test, an option to retest will be displayed.

Retest

Retest H3-H2 X3-X0

Select test setup to return to the quick test setup screen.

Megger TAU3 - Save, Print, Export, Retest, Home, Quick Test Setup - 4

Select home to return to the home screen.

Megger TAU3 - Save, Print, Export, Retest, Home, Quick Test Setup - 5

Results

After saving, results will be available for review in the results screen. Results can be sorted by asset ID, vector, test ID, Date, tests performed, and # of results. In addition, results can be filtered by Asset ID.

Tests Performed will show icons for the tests saved for the test ID. TTR, Magnetic Balance, Resistance, Short Circuit Impedance, and Frequency Response Stray Losses icons in green indicate results that are within limits. A demag symbol in green indicates a successful demagnetization of the transformer. Icons in red indicate results that are not within limits.

Select a result to view the test information. Selecting a check box will enable delete, export, and test options

Select multiple check boxes to delete or export multiple results

Select the check box in the table header row to select all results.

TAU3-EXP 11/05/2022 08:51:35 AM Asset ID Vector Test ID ↑ Date Tests Performed # of Results 221105074655 221012154047 220926094220 220915161846 220913122606 4 8 4 1 2 Filter Asset ID Delete All Export All View Test

After selecting a result, the test information will be displayed.

Megos Instrument GUI 1.1.0-dev1 Back Test Setup TTR & Balance Summary Dyn1 3-N Term Ratio Result Phase Radio Champs TTR HV mA @ Dev 0.5° H1-H3 X1-X0 17.289 250.0 106.9 0.06 H2-H1 X2-X0 17.337 250.0 94.0 0.17 H3-H2 X3-X0 17.281 250.0 92.2 0.12 Text Mode Up 3 Phase Test Voltage 250.0V Winding Top E-A Voltage Max/Min Rated 1 2 % Primary 3 0 0.0 None 0.00 Secondary M- 0 0.0 None 0.00 Print Reset Delete Text ID: 221105074655

Select delete to delete the results. Confirmation is required to delete the result

Megger TAU3 - Results - 3

Results

Megger TAU3 - Results - 1

Delete result?

Delete selected results in 221105074655 at taps: 3 - N?

Megger TAU3 - Delete result? - 1

Megger TAU3 - Delete result? - 2

Asset Template

Enter an Asset ID for identification of the asset.

Select Primary and Secondary vector and enter line to line voltage, KVA/MVA, and Z % for the transformer.

TAU3-EXP Nameplate Tags info Predicted: Verify against nameplate Vector L-L Voltage Base 60k Rated 1 Z % Primary 0 0.0 0.0 0.00 Secondary 0 0.0 0.0 0.00 Add Tertiary + Add Picture

Select add tertiary if the transformer has a tertiary winding.

Megger TAU3 - Asset Template - 2

Select save to save the asset template

Megger TAU3 - Asset Template - 3

Select add picture to add pictures to the asset template.

Megger TAU3 - Asset Template - 4

3 PHASE BANK CONNECTION WIN- DING VOLTS(V) HV 362.250 353.625 345,000 335.375 327.750 XV 138,000 YV 34,500

Asset Template

Use the rotate, zoom in, and zoom out buttons to view picture details. Click and hold to move the picture around. Pinch to zoom and swipe to view picture details on the TAU3 display.

Megger TAU3 - Asset Template - 1

natural_image Three white magnifying glass icons on a dark background: refresh, add, and minus sign (no text or symbols)

Delete will remove the picture from the asset template

Megger TAU3 - Asset Template - 2

When asset template is first selected, nameplate entry will selected.

Nameplate Taps Info i

Select taps to configure the tap changers for the primary and secondary winding. Winding options will be prepopulated based on measurement & limits settings.

Nameplate Taps Info i

TAU3-EXP Nameplate Taps info Predicted: Verify against nameplate Required Finals Primary (DETC) Secondary (OLTC) Tap Change DETC Nominal Tap ↑ Voltage ↓ F Taps 5 1 2 3 4 5 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 1 V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V 1V

After entering two tap voltages, the remaining tap voltages will automatically calculate. Override tap voltages as required. Select reset tap voltages to clear tap voltages.

Megger TAU3 - Asset Template - 6

Select Info to enter additional common nameplate information.

Nameplate Taps Info i

Mogger Instrument QGB 1.0-dwg1 TAU3-EXP 11/05/2022 05:32:46 PM Asset ID: Hide Picture 0/5 Nameplate Taps Info Save Secondary Oil GAL Class AN Coolant Oil & Water Coolant Volume 0 GAL Core Design Unknown Weight 0 kg Winding Material Copper Add Picture Custom Field

  • Manufacturer
  • Serial Number
  • Year
  • Substation
  • Position
  • Tank Type
  • Primary BIL
  • Secondary BIL
  • Class
  • Coolant
  • Core Design
  • Weight
  • Primary winding material
  • Secondary winding material

Select add custom field to enter custom nameplate information.

Megger TAU3 - Asset Template - 9

Custom Field

Template Test

Template Test

Asset templates can be sorted by Asset ID, vector, primary voltage, secondary voltage, KVA, and last test date and time. Select an Asset ID to proceed to the template test screen.

Magyar Instrument GUI 1.1.0.doc1 TAU3-EXP 11/05/2022 05:46:36 PM Asset ID Vector Pri V Sec V KVA Last Text ExampleTx 21000 13800 12000 Add Import Export All

Select add to create a new asset template

Megger TAU3 - Template Test - 2

Select export all to export all asset templates

Megger TAU3 - Template Test - 3

Select import to import exported asset ids.

Megger TAU3 - Template Test - 4

Select the check mark next to an asset id to show additional options.

Megger Instrument 001.1.0-dev1 TAU3-EXP 11/05/2022 05:49:14 PM Asset ID Vector Pri V Sec V KVA Last Test ExampleTx 23000 13000 12000 Add Import Test Edit Export Copy Delete

Select test to go to the template test screen

Megger TAU3 - Template Test - 6

Select edit to edit the asset template

Megger TAU3 - Template Test - 7

Select export to export the asset template

Megger TAU3 - Template Test - 8

Select copy to copy the asset template

Megger TAU3 - Template Test - 9

Select delete to delete the asset template

Megger TAU3 - Template Test - 10

The template test screen is similar to quick test. Select a test to get started.

Meggen Instrument GUI 1.0-dash Back all basic test Select Test YTR 31c Resistance & Damag Ω Short Circuit Impedance & FNSL 3Ω MM Efficiency All Tests Select a test to get started Examples Test ID:

Select edit to edit the asset template.

Megger TAU3 - Template Test - 12

Enter a Test ID for the asset template test. Useful for grouping asset test results for specific testing activities or for tests performed on different days.

Template Test

Megger TAU3 - Template Test - 1

After selecting a test, taps to test and test settings will display. Tap selections available based on asset template configuration. Specific test settings will be displayed based on the test selected, same as quick test.

Megger Instrument 0011.1.0-dyn-1 Back 11/05/2022 05:58:39 PM Select Test TTR Resistance & Demag Short Circuit Impedance 3E Efficiency All Tests Select Phases to Test Dyn1 H2 H1 X2 X1 X0 X3 Select Taps to Test Test Settings Recommended Measurements & Limits Voltage: 250 V Text Mode: Auto Tap Changer OUTC Operation Pulse/Wait Make Time 1.0s Watt Time One Touch OUTC Error Limited OUTC Test ID: ExampleTo Top Primary DETC 1 Start Tap End Tap Secondary OUTC- 16R- 16L- Test ID: Test Settings

If the asset has an OLTC, there will be options for configuring OLTC control

Tap Changer OLTC Operation Pulse/Wait • * Tap Timing Make Time 1.0s Wait Time 1.0s One Touch OLTC ✓ • * Error Limited OLTC

Single Tap Test

Mogger Instrument GUI 1.1.0-dev1 Back 01/06/2022 05:27:24 AM Select Test TTR Resistance & Demag Short Circuit Impedance & FRSL Efficiency All Tests Select Phases to Test Dyn1 H2 H1 H3 X2 X1 X0 X3 Start Test Select Taps to Test Test Settings Recommended Measurements & Limits Voltage 250 V Auto Top Changer ExampleTs Test ID: Top Primary DUTC 1 Start Top End Top Secondary OUTC N N OLTC Operation Continuous Press

If the asset does not have an OLTC, or the start and end tap for the OLTC are the same, testing will be complete as described in quick test.

Megger TAU3 - Single Tap Test - 2

other Dyn1 1-N Turns Ratio Result | Phase | Ratio (Value) TTR ●-1 | % Error 0.50 % - | HV | mA | θ Dev 0.5° | |---|---|---|---|---|---| | H1-H3 X1-X0 | 2.8886 | 0.06 | 78.3 | 15.6 | 0.08 | | H2-H1 X2-X0 | 2.8859 | -0.03 | 78.3 | 16.2 | -0.22 | | H3-H2 X3-X0 | 2.8886 | 0.06 | 78.3 | 17.6 | 0.15 | Text Mode Up 3 Phase Test Voltage 78.3V Calculated Windling Top 1-4 Voltage Primary 1 23000 Secondary HV N 12000 Test ID: 221106053745 11/06/2022 05:43:36 AM Reset Next > Delete

When testing is complete, the Test Setup button can be used to return to the Template Test setup screen.

Megger TAU3 - Single Tap Test - 3

In addition, the tap can be changed in the nameplate section to test other tap positions.

Megger TAU3 - Single Tap Test - 4

other | Category | Winding | Tap | L-L Voltage | |---|---|---|---| | Primary | [IMAGE] | 1 | 23000 | | Secondary OLIC | [IMAGE] | N ▼ | 13800 |

Below is an example when the tap is changed and no results are available.

Mogger Instrument GBI 1.3 dev1 Test Setup Dyn1 1-48 Turn Ratio Result Phase HV mA @ Dev 0.5° Text Mode 0 Test Voltage Example To Winding Tip L-L Primary 1 23006 Secondary 48 14125 Test ID: 221106053745 Start Text Free Export Save

Click Start Test

Start Test

Multiple Tap OLTC AC Tests

Multiple Tap OLTC AC Tests - TTR, Magnetic Balance, Short Circuit Impedance, FRSL, and Efficiency

The start tap and end tap of an OLTC can be changed to test multiple tap positions.

Meggon instrument GUI 1.0-dev Back Select Test TTR 3c Resistance & Demag Ω Short Circuit Impedance & FRSL Efficiency All Tests Dyn1 H2 H1 H3 X2 X1 X0 X3 Select Phases to Test Start Test Select Taps to Test Test Settings Recommended Measurements & Limits Tap Change ExampleTx Primary ONTC 1 Voltage 250 V Text Mode Amp Pulse/Wait Tap Timing Maser Time 1.0s Wall Time One Touch ONTC Test ID: Primary ONTC 16R 16L

The screen below is displayed after a test is complete when testing TTR & Balance, Short Circuit Impedance & FRSL, or Efficiency.

Megger TAU3 - Multiple Tap OLTC AC Tests - TTR, Magnetic Balance, Short Circuit Impedance, FRSL, and Efficiency - 2

If the TAU3 tap changer cable is connected to the TAU3 and to the transformer OLTC control, the TAU3 OLTC switch can be used to change the tap changer position. If set to Continuous Press, the next test will begin when the TAU3 OLTC switch is released. If set to pulse/wait, the next test will begin when the pulse/wait timing is complete.

Note: If the tap changer is operated in the opposite direction as indicated in the software, testing will continue in the new direction selected.

If the TAU3 OLTC is not used, the OLTC must be changed manually at the transformer. When the tap change is complete, click the yellow button to begin the next test.

Change OLTC Tap 16R 15 Or Change at Tx and click after change!

Testing will complete when the last tap is tested.

Multiple Tap OLTC Winding Resistance Tests

The start tap and end tap of an OLTC can be changed to test multiple tap positions.

Select Test TTR Resistance & Demag Short Circuit Impedance & FRSL Efficiency All Tests Select Phases to Test Dyn1 H2 X2 H1 X1 X0 H3 X3 Start Test Test ID Test Settings Recommended Measurements & Limits Tap Change Primary OUTC Primary A Secondary A Voltage 100 V DC Primary A DC Secondary A DC Stability % 99.95% Stability Time 10s Auto Save OLTC Operation Continuous Press ExampleTo Test ID Primary OUTC Secondary OUTC N N Back Edit AssetID 230316132422 03/16/2023 01:39:21 PM Select Test Find OLTC Tap TTR Resistance & Demag Short Circuit Impedance & FRSL Efficiency All Tests Dyn1 H2 X2 H1 X1 X0 X3 Select Taps to Test Text Settings Recommended Measurement & Limits Tap Change Primary (OUTC) Primary A Secondary A Voltage 100 V DC Primary A DC Secondary A DC Stability % 99.95% Stability Time 10s Auto Save Single Test OutTC Operation Continuous Press Test ID

The screen below is displayed after the first phase and tap are complete when testing winding resistance.

Mogger instrument GUI 1.1.0-dev/0 TAU3-TEST 02/16/2023 01:40:20 PM Resistance & Demag Dyn1 3-16R Resistance In Progress Set in Progress Set Completed To be Tested H1 7.3678 Ω H2 X1 1.6733 Ω X2 X3 7.9 A Current, A DC Stability Sound! Stability 15.8 A Current, A DC Stop Test 230316133422 Test ID: 230316133957 Writing Tap L-L 1 Voltage Pulse 1 0 Reversible 16R 16R 0 Change OLTC Tap 16R 15R Or Change at Tx and click after change!

If the TAU3 tap changer cable is connected to the TAU3 and to the transformer OLTC control, the TAU3 OLTC switch can be used to change the tap changer position. If set to Continuous Press, the next test will begin when the TAU3 OLTC switch is released. If set to pulse/wait, the next test will begin when the pulse/wait timing is complete.

Note: If the tap changer is operated in the opposite direction as indicated in the software, testing will continue in the new direction selected.

If the TAU3 OLTC is not used, the OLTC must be changed manually at the transformer. When the tap change is complete, click the yellow button to begin the next test.

Multiple Tap OLTC Winding Resistance Tests

Change OLTC Tap 16R 15R Or Change at Tx and click after change!

When the last tap of the first phase is completed, the test set will automatically discharge and energize the next phase to be tested.

Testing will complete when the last tap of the final phase has been saved.

Multiple Tap OLTC All Tests

The start tap and end tap of an OLTC can be changed to test multiple tap positions.

Select Test All Items 30 Dyn1 H2 H1 X2 X1 X0 X3 Select Phases to Test Start Test Test Tools to Test Top Test Settings Recommended Measurements & Limits Top Change ExampleTx DETc Primary 1 Voltage Primary A Secondary A 250 V 16.0 A 32.0 A Secondary A OUTC Secondary OUTC 3R 3L Stability % 99.95% Stability Time 10s OUTC Operation Continuous Press Test ID

When performing All Tests, the procedure is as follows.

  1. Preliminary Demagnetization.

- Used to ensure accuracy of AC tests. Can be canceled.

  1. Tap Changer Verification.

- When Auto OLTC is selected, the TAU3 will automatically confirm proper connection to the tap changer by performing a series of TTR tests and tap changes. Can be canceled.

  1. TR & Balance*, Short Circuit Impedance & FRSL*, and Efficiency* will be tested on each tap from start tap to end tap

  2. Winding resistance

a. Phase A end tap to start tap
b. Phase B start tap to end tap
c. Phase C end tap to start tap

  1. Demagnetization

*Balance, FRSL, and efficiency will only be tested if purchased and enabled.

PowerDB Import

PowerDB Import

The combined 30 form can be used to import TAU3 data into PowerDB.

PowerDB Control

The combined 30 form can be used to configure and run tests with the TAU3.

Service

If immediate assistance is required, please contact the customer service by telephone or email:

+1 670 676 8500

vfcustomersupport@megger.com

Troubleshooting

The Troubleshooting Guide is designed to evaluate the reasons for a TAU3 malfunction. The possible test set malfunctions and causes are listed below. Electronic circuit repairs should not be attempted in the field. Refer to Repair section.

TTRU3 does not turn on

  • Check that the power cord is fully inserted into the TTRU3.
  • Check that the power source is outputting voltage at acceptable levels and frequency.
  • Check that the power cord is fully inserted into the source.
  • Check that the power switch is in the correct position (1).
    ■ Set the power switch to off (O)

Wait 30 seconds

Set the power switch to on (1)

■ Try another power cord

Lead check

■ Connect the leads as described below

H1 to X1

H2 to X2

H3 to X3

H0 to X0

■ Select Dd0, Yy0, or YNyn0 and perform a test
■ Verify ratio between 1.0005 and 0.9995

TTRU3 Reports test failed, but still provides data

- Check lead connections. Reference Nameplate to ensure leads are connected to the correct bushing.

Printer not working

  • Check battery is inserted into printer
  • Charge printer battery using supplied charger
  • Check printer paper is inserted properly
  • Check USB cable is plugged into printer
  • Check USB cable is plugged into TTRU3 USB port
  • Check printer is turned on by holding power button
    ■ Try other USB ports

Maintenance

OLTC moving in wrong direction

- Check the OLTC wiring diagram and ensure leads are connected to correct terminals.

Cannot connect TTRU3 to PC

Contact your IT department for primary assistance when connecting any device to your PC.

  • Check USB cable is fully inserted into the TTRU3
  • Check USB cable is fully inserted into PC
  • Check the TTRU3 is powered on
  • Check TTRU3 SW is installed
  • Check TTRU3 is running
    ■ Move USB cable to another USB port on your PC
    ■ Try another USB Cable
    Try another PC

Maintenance

Only qualified persons familiar with the hazards involved with high-voltage test equipment should perform maintenance. Read and understand Sections 1, 2, 3, 4, and 5 before performing any service.

The TTRU3 requires only periodic inspection. Inspect all hardware items to ensure all are in good condition.

The TTRU3 may be cleaned periodically. In so doing, do not allow water to penetrate panel holes. An all-purpose, household spray cleaner can be used to clean the panel. Polish with a soft, dry cloth. Clean the cables and mating panel receptacles with isopropyl or denatured alcohol applied with a clean cloth.

Calibration

A complete performance and calibration check should be made at least once every year. This will ensure that the TTRU3 is functioning properly over the entire measurement range. The TTRU3 calibration is performed on each new or repaired unit before sending it to a customer.

Repairs

Any service or repair of this equipment should be performed only by qualified persons who are aware of electrical hazards and the necessary precautions required to prevent injury.

Megger offers a complete Repair and Calibration Service and recommends that its customers take advantage of this service for routine maintenance or in the event of any equipment malfunction.

In the event Service is required, contact your Megger representative for a product Return Authorization (RA) number and shipping instructions.

Ship the product prepaid and insured and marked for the attention of the Megger Repair Department. Please indicate all pertinent information, including catalog number, serial number, and problem symptoms.

Manufacturing sites

Megger Limited

Archcliffe Road

Dover

Kent

CT17 9EN

ENGLAND

T.+44 (0)1 304 502101

F. +44 (0)1 304 207342

Megger GmbH

Weststraße 59

52074

Aachen

Germany

T.+49 (0) 241 91380 500

E. info@megger.de

Megger Valley Forge

400 Opportunity Way

Phoenixville

PPA 19460

USA

T. 1-610 676 8500

F. 1-610-676-8610

Megger USA - Dallas

4545 West Davis Street

Dallas

75211-3422

T.+1 214 333 3201

F. +1 214 331 7399

USsales@megger.com

Megger AB

The company reserves the right to change the specification or design without prior notice.

Megger is a registered trademark

The Bluetooth ® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc and is used under licence.

Part No: TAU3_UG_en_V01g

Table of contents Click a title to access it
Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Megger

Model : TAU3

Category : Measuring instrument