FLUKE

721 - Uncategorized FLUKE - Free user manual and instructions

Find the device manual for free 721 FLUKE in PDF.

📄 28 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice FLUKE 721 - page 9
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Pressure Calibrator
Brand Fluke
Model 721
Pressure Range 0 to 30 psi (2 bar)
Pressure Media Non-corrosive gases
Accuracy ±0.05% FS
Pressure Connections 1/8 inch NPT female
Loop Power Supply 24V DC, 25 mA max
Power Source 4 AA alkaline batteries
Battery Life Approximately 300 hours typical
Dimensions (H x W x D) 200 x 100 x 60 mm
Weight 1.0 kg (2.2 lb)
Display Dual LCD with backlight
Key Functions Pressure measurement, pressure generation, 24V loop supply, mA measurement
Min/Max Hold Yes
Auto Shutoff After 30 minutes idle
Operating Temperature -10°C to 55°C (14°F to 131°F)
Storage Temperature -20°C to 60°C (-4°F to 140°F)
Humidity 0% to 90% non-condensing
Safety Rating CAT II 30V
Overpressure Protection Up to 45 psi
Cleaning Wipe with damp cloth and mild detergent
Spare Parts Battery cover, O-rings, test leads (if applicable)
Repairability Field-replaceable batteries; service by qualified personnel for other repairs
Warranty 1 year (standard)

Frequently Asked Questions - 721 FLUKE

What is the Fluke 721 used for?
The Fluke 721 is a pressure calibrator used to measure and generate pressure, as well as provide 24V loop power for testing transmitters.
What is the maximum pressure the Fluke 721 can measure?
The Fluke 721 can measure pressure up to 30 psi (2 bar). It also generates pressure up to its full range.
How do I power the Fluke 721?
The device uses 4 AA alkaline batteries. Insert them in the battery compartment at the rear. Battery life is approximately 300 hours.
Does the Fluke 721 measure current?
Yes, it can measure mA for transducer outputs. It also supplies 24V loop power to powered transmitters.
How do I zero the pressure reading?
Ensure no pressure is applied, then press the ZERO button. The device will tare the current reading to zero.
What type of pressure connections does the Fluke 721 have?
It uses 1/8 inch NPT female connections. Adapters for other fittings are available separately.
Is the Fluke 721 safe to use with flammable gases?
No, the Fluke 721 is designed for non-corrosive gases only. Do not use with flammable or corrosive media.
How do I clean the Fluke 721?
Clean the exterior with a damp cloth and mild detergent. Do not use solvents or immerse the device.
What does the error message 'OL' mean?
It means overload or overpressure. The applied pressure exceeds the safe limit. Reduce pressure immediately.
Can I replace the batteries myself?
Yes, batteries are field-replaceable. Open the battery door, replace the 4 AA cells, and close the door securely.

User questions about 721 FLUKE

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Uncategorized in PDF format for free! Find your manual 721 - FLUKE and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. 721 by FLUKE.

USER MANUAL 721 FLUKE

Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without interruption.

Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country.

Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.

To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures caused by use outside the product's specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).

THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.

Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.

Fluke Corporation

P.O. Box 9090

Everett, WA 98206-9090

U.S.A.

Fluke Europe B.V.

P.O. Box 1186

5602 BD Eindhoven

The Netherlands

Table of Contents

Title

Page

Introduction....1

Contact Fluke 1

Safety Information 2

Symbols.... 3

Specifications 4

Environmental....4

Electrical and Temperature Measurement (1 year) 4

Physical 4

Required Equipment.... 6

Performance Verification Tests 6

Pressure Verification 7

DC Voltage Measure Verification 9

DC Current Measure Verification.... 10

RTD Measure Verification 10

720RTD Probe Verification.... 11

mA Loop Performance Check 11

Calibration Adjustment 12

Initiate Communication.... 12

Adjust Voltage Input 13

Adjust mA Input 13

Adjust RTD 14

Adjust Pressure 15

Maintenance.... 16

Replace the Batteries 16

Clean the Product.... 18

User-Replaceable Parts 18

Accessories 19

List of Tables

TableTitlePage
1.Symbols......3
2.Required Equipment for Verification and Calibration......6
3.Pressure Verification Tests......7
4.DC Voltage Measure Verification Tests......9
5.DC Current Measure Verification Tests......10
6.RTD Measure Verification Tests......10
7.720RTD Probe Verification Tests......11
8.User-Replaceable Parts and Accessories......18
9.Accessories......19

List of Figures

Figure

Title

  1. Battery Replacement 17
  2. User-Replaceable Parts 20

Introduction

The 721 Pressure Calibrator (the Product) is a simple to use and versatile pressure calibrator. The two internal pressure sensors are configured with different pressure measurement ranges. One range for lower pressure (P1) and one range for higher pressure (P2). The Product features inputs for mA, switch contacts, and an RTD probe. An external pressure module option gives a wider range of pressure calibration options that include absolute and differential pressure measurement.

Contact Fluke

To contact Fluke, call one of the following telephone numbers:

• Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
• Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31 402-675-200

• Japan: +81-3-6714-3114

• Singapore: +65-6799-5566

• Anywhere in the world: +1-425-446-5500

Or, visit Fluke's website at www.fluke.com.

To register your product, visit http://register.fluke.com.

To download manuals, or to view, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.

Safety Information

A Warning identifies conditions and procedures that are dangerous to the user.

A Caution identifies conditions and procedures that can cause damage to the

Product or the equipment under test.

Warning

To prevent possible electrical shock, fire, or personal injury:

  • Only assemble and operate high-pressure systems if you know the correct safety procedures. High-pressure liquids and gases are hazardous and the energy from them can be released without warning.
  • Read all safety information before you use the Product.
  • Carefully read all instructions.
  • Do not use the Product around explosive gas or vapor.
  • Use the correct terminals, function, and range for measurements.
  • Do not apply more than the rated voltage, between the terminals or between each terminal and earth ground.
  • Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc.
  • Remove all probes, test leads, and accessories before the battery door is opened.
  • Do not exceed the Measurement Category (CAT) rating of the lowest rated individual component of a Product, probe, or accessory.
  • Disable the Product if it is damaged.
  • Do not use the Product if it is damaged.
  • Remove the input signals before you clean the Product.
  • Use only specified replacement parts.
  • Have an approved technician repair the Product.
  • Use the Product only as specified, or the protection supplied by the Product can be compromised.
  • Pressure sensors can be damaged and/or personnel injury can occur due to improper application of pressure. Vacuum should not be applied to any gauge pressure sensor. The Product display shows “OL” when an inappropriate pressure is applied. If “OL” is shown on any pressure display, the pressure should be reduced or vented immediately to prevent Product damage or possible personnel injury. “OL” is shown when the pressure exceeds 110 % of the nominal range of the sensor or when a vacuum in excess of 2 PSI is applied on gauge range sensors.

  • Push the ZERO button to zero the pressure sensor when vented to atmospheric pressure.

  • Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures that exceed the battery manufacturer's specifications. If the batteries are not removed, battery leakage can damage the Product.
  • Replace the batteries when the low battery indicator shows to prevent incorrect measurements.
  • Be sure that the battery polarity is correct to prevent battery leakage.
  • Repair the Product before use if the battery leaks.
  • The battery door must be closed and locked before you operate the Product.
  • Use only specified replacement parts.

Symbols

Symbols used on the Product or in this manual are shown in Table 1.

Table 1. Symbols

SymbolDescriptionSymbolDescription
FLUKE 721 - Symbols - 1Risk of Danger. Important information. See Manual.FLUKE 721 - Symbols - 2Conforms to relevant North American Safety Standards.
[OWHS]Hazardous voltage. Risk of electric shock.[W287]Double Insulated
[S42Y]Conforms to European Union directives.FLUKE 721 - Symbols - 3Conforms to relevant Australian EMC standards.
FLUKE 721 - Symbols - 4Inspected and licensed by TÜV Product Services.[35X6]Battery
FLUKE 721 - Symbols - 5This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste.Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste. Go to Fluke's website for recycling information.FLUKE 721 - Symbols - 6Conforms to relevant South Korean EMC Standards.

Specifications

Conditions for specifications are 15^ C to 35^ C unless otherwise noted.

Environmental

Operating Temperature ....-10 °C to +50 °C (14 °F to +122 °F)

Storage

With Batteries .... Per battery manufacturer's specification, not to exceed storage specification without batteries.

Without Batteries ....-20 °C to +60 °C (-4 °F to +140 °F)

Altitude....2000 m

Power Requirements 6 V dc

Batteries 4 AA batteries (alkaline)

Battery Life ....>35 hours, typical usage

Electrical and Temperature Measurement (1 year)

FunctionRangeResolutionAccuracy
mA Measure 0 to 24 mA 0.001 mA±0.015 % of rdg ±0.002 mA
V dc Measure 0 to 30 V dc 0.001 V±0.015 % of rdg ±0.002 V
*Temperature measure (RTD/Ohms)-40 °C to 150 °C (-40 °F to 302 °F)0.01 °C, 0.01 °F±0.015 % of rdg ±0.02 Ω; ±0.1 °C (±0.2 °F) ±0.25 °C (±0.45 °F) combined uncertainty when using 720 RTD probe accessory
Loop Power Supply 24 V N/A 24 mA @ 24 V
*Temperature measurement requires optional 720RTD Pt-100 RTD probe available as an accessory.

Physical

Dimensions (with holster) ....(H x W x D) (20 x 11 x 5.8) cm, (7.9 x 4.3 x 2.3) in

Weight (with holster) 0.539 kg (1 lb 3 oz)

EMI/RFI Conformance ...... EN50082-1 and EN55022: Class B

Ingress Protection IP40

Connectors/Ports

Pressure ...... Two, 1/8 in NPT, 700P Series Pressure Module

RTD.... RTD probe

Temperature Effect (all functions)....No effect on accuracy on all functions from 15 °C to 35 °C

Add ±0.002 % F.S./°C for temperatures outside of 15 °C to 35 °C

1-year Specifications LowPressure Sensor High Pressure Sensor
ModelCalibrator DescriptionRange Sensor 1Resolution Sensor 1Accuracy Sensor 1Range Sensor 2Resolution Sensor 2Accuracy Sensor 2
721-160116 PSIG, 100 PSIG-14 psi + 16 psi -0.97 bar to 1.1 bar0.001 psi, 0.0001 bar0.025 % of full scale-12 psi to +100 psi -0.83 bar to 6.9 bar0.001 psi 0.0001 bar0.025 % of full scale
721-160316 PSIG, 300 PSIG-12 psi to +300 psi -0.83 bar to 20 bar0.01 psi 0.001 bar
721-160516 PSIG, 500 PSIG-12 psi to +500 psi -0.83 bar to 34.5 bar0.01 psi 0.001 bar
721-161016 PSIG, 1000 PSIG0 psi to +1000 psi 0.00 bar to 69 bar0.01 psi 0.001 bar
721-161516 PSIG, 1500 PSIG0 psi to +1500 psi 0.00 bar to 103.4 bar0.1 psi 0.01 bar
721-163016 PSIG, 3000 PSIG0 psi to +3000 psi 0.00 bar to 200 bar0.1 psi 0.01 bar
721-165016 PSIG, 5000 PSIG0 psi to +5000 psi 0.00 bar to 345 bar0.1 psi 0.01 bar0.035 % of full scale
721-360136 PSIG, 100 PSIG-14 psi + 36 psi -0.97 bar to 2.48 bar-12 psi to +100 psi -0.83 bar to 6.9 bar0.001 psi 0.0001 bar0.025 % of full scale
721-360336 PSIG, 300 PSIG-12 psi to +300 psi -0.83 bar to 20 bar0.01 psi 0.001 bar
721-360536 PSIG, 500 PSIG-12 psi to +500 psi -0.83 bar to 34.5 bar0.01 psi 0.001 bar
721-361036 PSIG, 1000 PSIG0 psi to +1000 psi 0.00 bar to 69 bar0.01 psi 0.001 bar
721-361536 PSIG, 1500 PSIG0 psi to +1500 psi 0.00 bar to 103.4 bar0.1 psi 0.01 bar
721-363036 PSIG, 3000 PSIG0 psi to +3000 psi 0.00 bar to 200 bar0.1 psi 0.01 bar
721-365036 PSIG, 5000 PSIG0 psi to +5000 psi 0.00 bar to 345 bar0.1 psi 0.01 bar0.035 % of full scale

Electromagnetic Environment ...... IEC 61326-1: Portable

Electromagnetic Compatibility....Applies to use in Korea only. Class A Equipment (Industrial Broadcasting & Communication Equipment) [1]

[1] This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes.

Required Equipment

The equipment listed in Table 2 is necessary to do the performance verification tests and calibration adjustment.

Table 2. Required Equipment for Verification and Calibration

Equipment Minimum Specifications Recommended Model
Low-thermal EMF banana to banana leads-Pomona 2948-36-2 (red) and 2948-36-0 (black)
RTD adapter - Fluke 720URTDA
Calibrator/Resistance Standard0 V to 30 VAccuracy: 0.006 % rdg ±0.5 mV0 mA to 24 mAAccuracy: 0.006 % rdg ±0.25 μA-40 °C to 150 °C RTD0.033 °CFluke Calibration 7526A
Pressure Controller/Calibrator-14 PSI to 1000 PSI0.005 %Ruska 7250xi
Piston Gauge3000 PSI to 5000 PSI0.005 %Fluke Calibration PG7302 or PG7202
Metrology Well-30 °C to 140 °C0.06 °CFluke Calibration 9170

Performance Verification Tests

Warning

To prevent possible electrical shock, fire, or personal injury, do not perform the performance test or verification test procedures unless the Product is fully assembled.

The performance tests verify the complete operation of the Product and measure the accuracy of each function against Product specifications. If the Product fails a part of the test, calibration adjustment or repair is necessary. See “Calibration Adjustment Procedure”.

The performance verification tests check the accuracy of each Product function against its specifications. If the Product fails any of these tests, calibration adjustment or repair is necessary.

Pressure Verification

To do the pressure verification test:

  1. Carefully attach the pressure fitting of the Pressure Controller/Calibrator or Piston Gauge (called pressure source for the remainder of this procedure) to the first pressure port of the Product.

Note

Use plenty of TEFLON tape when attaching pressure fitting.

The display should read 0.00 PSI with the deadweight tester opened up to ambient air. If not, push ZERO until both displays read 0.00 PSI.

  1. Set up the deadweight tester for the sequence of PSI inputs from Table 3 to be injected into the pressure port of the Product.
  2. Ensure the pressure has stabilized at each input before you verify the display reading.
  3. Carefully vent all pressure and disconnect the first pressure port.
  4. Connect and test the second pressure port.
  5. Set up the deadweight tester for the sequence of PSI inputs from Table 3 to be injected into the pressure port of the Product.
  6. Ensure the pressure has stabilized at each input before you verify the display reading.
  7. Set up the deadweight tester for the sequence of PSI inputs from Table 3 to be injected into the pressure port of the Product.
  8. Ensure the pressure has stabilized at each input before you verify the display reading.

Table 3. Pressure Verification Tests

16 PSI 36 PSI
Input PressureLower LimitUpper Limit
16.00015.99616.004
13.00012.99613.004
10.0009.99610.004
6.0005.9966.004
3.0002.9963.004
0.000-0.0040.004
-5.000-5.004-4.996
-10.000-10.004-9.996
-13.750-13.754-13.746
Input PressureLower LimitUpper Limit
36.00035.991
29.00028.991
22.00021.991
14.00013.991
7.0006.991
0.000-0.0090.009
-5.000-5.009
-10.000-10.009-9.991
-13.750-13.759-13.741

Table 3. Pressure Verification Tests (cont.)

100 PSI 300 PSI
Input PressureLower LimitUpper Limit
100.0099.975100.025
80.0079.97580.025
60.0059.97560.025
40.0039.97540.025
20.0019.97520.025
0.00-0.0250.025
-6.00-6.025-5.975
-12.00-12.025-11.975
Input PressureLower LimitUpper Limit
300.00299.92300.08
240.00239.92240.08
180.00179.92180.08
120.00119.92120.08
60.0059.9260.08
0.00-0.08
-6.00-6.08-5.92
-12.00-12.08-11.92
500 PSI 1000 PSI
Input PressureLower LimitUpper Limit
500.00499.87500.13
400.00399.87400.13
300.00299.87300.13
200.00199.87200.13
100.0099.87100.13
0.00-0.130.13
-6.00-6.13-5.87
-12.00-12.13-11.87
Input PressureLower LimitUpper Limit
1000.0999.751000.25
800.0799.75800.25
600.0599.75600.25
400.0399.75400.25
200.0199.75200.25
0.0-0.250.25
1500 PSI
Input PressureLower LimitUpper Limit
1500.01499.61500.4
1200.01199.61200.4
900.0899.6900.4
600.0599.6600.4
300.0299.6300.4
0.0-0.40.4
3000 PSI
Input PressureLower LimitUpper Limit
3000.02999.23000.8
2400.02399.22400.8
1800.01799.21800.8
1200.01199.21200.8
600.0599.2600.8
0.0-0.80.8

Table 3. Pressure Verification Tests (cont.)

5000 PSI
Input PressureLower Limit Upper Limit
5000.04998.25001.8
4000.03998.24001.8
3000.02998.23001.8
2000.01998.22001.8
1000.0998.21001.8
0.0-1.81.8
  1. Carefully vent all pressure and disconnect the second pressure port.

DC Voltage Measure Verification

To do the dc voltage measure verification test:

  1. Push F2 repeatedly until VOLTS is shown on the third row of the Product display.
  2. Connect the Product red banana jack (V mA) to the Calibrator HI VOLTS output.
  3. Connect the Product black banana jack (COM) to the Calibrator LO VOLTS output.
  4. Set the Calibrator for the voltage setting in Table 4, and verify the display reading on the Product.

Table 4. DC Voltage Measure Verification Tests

Applied Voltage from 7526A Lower limit Upper Limit
0.000 V -0.002 V 0.002 V
15.000 V 14.996 V 15.004 V
30.000 V 29.994 V 30.007 V
  1. Set the Calibrator to Standby.

DC Current Measure Verification

To do the dc current measure verification test:

  1. Push F2 repeatedly until mA is shown in the third row of the Product display.
  2. Connect the Product red banana jack (V mA) to the Calibrator HI mA output.
  3. Connect the Product black banana jack (COM) to the Calibrator LO mA output.
  4. Set the Calibrator for the current setting in Table 5, and verify the display reading on the Product.

Table 5. DC Current Measure Verification Tests

Applied Current from the CalibratorLower limit Upper Limit
4.000 mA 3.997 mA 4.003 mA
12.000 mA 11.996 mA 12.004 mA
24.000 mA 23.994 mA 24.006 mA
  1. Set the Calibrator to Standby.

RTD Measure Verification

  1. Push F2 repeatedly until RTD is shown in the third row of the Product display.
  2. Connect the RTD Adapter to the Product.
  3. Connect the RTD Adapter Current HI to the Calibrator HI RTD output.
  4. Connect the RTD Adapter Sense HI to the Calibrator HI RTD output.
  5. Connect the RTD Adapter Current LO to the Calibrator LO RTD output.
  6. Connect the RTD Adapter Sense LO to the Calibrator LO RTD output.
  7. Set the Calibrator to the 400 Ω range, output the resistance setting in Table 6, and verify that the display reading on the Product is within the limits.
  8. Set Calibrator to standby.

Table 6. RTD Measure Verification Tests

Applied Current from the CalibratorLower limit Upper Limit
84.270 Ω-40.10 °C-39.90 °C
100.000 Ω-0.10 °C0.10 °C
157.325 Ω149.90 °C150.10 °C

720RTD Probe Verification

This is an optional test for a Product that is paired with a 720RTD. To do this test:

  1. Push F2 repeatedly until RTD is shown in the third row of the Product display.
  2. Plug the 720RTD probe into the Product.
  3. Insert the probe into the Metrology Well.
  4. Set the Metrology Well to the temperatures in Table 7, after the Well stabilizes, verify that the display readings are within tolerance.

Table 7. 720RTD Probe Verification Tests

Applied Temperature Lower Limit Upper Limit
-30.00 °C-30.25 °C-29.75 °C
0.00 °C-0.25 °C0.25 °C
70.00 °C69.75 °C70.25 °C
130.00 °C129.75 °C130.25 °C
  1. Set the Metrology Well to Standby.

mA Loop Performance Check

The Loop Voltage function does not have traceable specifications, this is a functional check.

  1. Push F2 repeatedly until mA Loop is shown in the third row of the Product display.
  2. Connect the Product red banana jack (V mA) the Calibrator HI input.
  3. Connect the Product black banana jack (COM) to the Calibrator LO input
  4. Set the Calibrator to the 100 V range, the voltage measured should be between 22 V and 30 V.

Calibration Adjustment

The Product has an electronic calibration process. There are no mechanical adjustments and the calibration is done with the case closed. A serial communications port is used for the calibration process to send commands and receive readings. The normal RS232 interface is used. Calibration can be done with a terminal program or an automated-calibration program can be written with programs like MetCal. In this manual, only the serial terminal mode is described.

Table 2 lists the required equipment.

Initiate Communication

Terminal communications can be set up with terminal communication software on a PC such as Hyperterminal or Ucon. Connect the RS232 cable to the 5-pin Lemo on the side of the Product. Connect the other end of the cable to the terminal/PC serial port.

The terminal settings need to be set as follows:

  • Bits per second: 9600
  • Data bits: 8
  • Parity: None
  • Stop bits: 1
  • Flow control: none
  • Local echo: on

First, enable the full-command set by sending HC_GASUI_OFF. Check that it turned off by sending HC_GASUI?, The Product should reply OFF. All calibration commands are done on the functions selected on the lower display. Use the following command sequence to turn off all but the lower display:

DISPLAY UPPER, OFF DISPLAY MIDDLE, OFF DISPLAY LOWER, ON

As long as the Product has been at a stable temperature, within the range of 20 °C to 26 °C for an hour or longer, the Product only needs 5 minutes to warm up. If temperature conditions were previously below 10 °C or higher than 40 °C , then the unit must be allowed to stabilize for a minimum of 3 hours prior to calibration.

Adjust Voltage Input

To adjust the voltage input:

  1. Use the test lead set to attach the voltage output of the Calibrator to the input of the Product.
  2. Send this command to put the Product into Voltage mode: FUNC LOWER, DCV
  3. Use the Calibrator to input 0 V.
  4. When the reading is stable, send this command: CAL_POINT[0] [0]
  5. Use the Calibrator to input 30 V.
  6. When the reading is stable, send this command: CAL_POINT[0][1]
  7. The voltage range is now calibrated.

Adjust mA Input

To adjust the mA input:

  1. Use the test lead set to attach the voltage output of the Calibrator to the input of the Product.
  2. Send this command to put the Product in Current mode: FUNC LOWER, DCI
  3. Use the Calibrator to input 0 mA.
  4. When the reading is stable, send this command: CAL_POINT[0][0]
  5. Use the Calibrator to input 24 mA.
  6. When the reading is stable, send this command: CAL_POINT[0][1]
  7. The voltage range is now calibrated. Use the Calibrator to verify calibration.

Adjust RTD

Before the adjustment, make these connections:

  1. Plug the RTD Adapter into the Product.
  2. Connect the high side of the current output (ohms measure) jacks of the RTD Adapter to the high side of the current input of the Resistance Standard.
  3. Connect the low side of the current output (ohms measure) of the RTD Adapter to the low side of the current input of the Resistance Standard.
  4. Connect the high side of the sense jacks of the RTD Adapter to the high side of the sense output of the Resistance Standard.
  5. Connect the low side of the sense jacks of the RTD Adapter sense input to the low side of the sense output of the Resistance Standard.

To make the adjustment:

  1. After you have made your connections, send this command to put the unit in RTD mode: FUNC LOWER,RTD
  2. Use the Resistance Standard to input 90 Ω.
  3. When the reading is stable send this command: CAL_POINT[0][0]
  4. Use the Resistance Standard to input 140 Ω.
  5. When the reading is stable, send this command: CAL_POINT[0][1]
  6. The RTD range is now calibrated. Use the Resistance Standard to verify calibration.

Adjust Pressure

The Product uses a 1/8 inch NPT female connection in the pressure input port. Various adapters may be necessary to connect to the pressure standard. Always make sure the hose, tubing, fittings, etc., have a rated working pressure at or above the pressure of the Product. It is important that there are no leaks. Use Teflon tape where appropriate.

To adjust the pressure:

  1. Send this command to put the Product into pressure mode:
    FUNC LOWER, {port} where port is either P1 or P2.
  2. Send the OFFSET_ADJ? command.
  3. Note the value returned.
  4. Use the Pressure Standard to input a value equal or close to the noted value.
  5. When the pressure is stable, send the OFFSET_ADJ n command where n is the entered pressure.
  6. Send the GAIN_ADJ? command.
  7. Note the value returned.
  8. Use the Pressure Standard to input a value equal or close to the noted value.
  9. When the pressure is stable, send the command GAIN_ADJ n where n is the applied pressure.

The pressure is now adjusted. Restore normal operation by turning the other displays back on. Send DISPLAY UPPER, ON then DISPLAY MIDDLE, ON. Check the unit display. There should be three bars across the display, the upper three show reading, the bottom showing the F key functions. Then, send HC_GASUI_ON. Verify that it is on by sending HC_GASUI?. The unit should reply ON.

Maintenance

Replace the Batteries

If the batteries discharge too far, the Product automatically shuts down to prevent battery leakage.

Note

Use only AA size alkaline batteries.

Warning

To prevent possible electrical shock, fire, or personal injury:

  • Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can damage the Product.
  • Replace the batteries when the low battery indicator shows to prevent incorrect measurements.
  • Be sure that the battery polarity is correct to prevent battery leakage.
  • Repair the Product before use if the battery leaks.
  • The battery door must be closed and locked before you operate the Product.

To change the batteries, see Figure 1:

  1. Turn off the Product.
  2. Turn the Product so that the display is down.
  3. With a flat-head screwdriver, remove the battery door screw.
  4. Replace the four AA batteries with new batteries. Make sure that the polarity on the batteries is correct.
  5. Replace the battery door.
  6. Tighten the battery door screw.

FLUKE 721 - Warning - 1

natural_image Illustration of a yellow battery pack with three batteries and one battery, showing internal structure and battery stack (no text or symbols)

Figure 1. Battery Replacement

hmq061.eps

Clean the Product

Caution

To avoid damaging the plastic lens and case, do not use solvents or abrasive cleansers.

Clean the Product with a soft cloth dampened with water or water and mild soap.

User-Replaceable Parts

⚠ Warning

To prevent possible electrical shock, fire, or personal injury, use only specified replacement parts.

User-replaceable parts are listed in Table 8 and shown in Figure 2. For more information about these items, contact a Fluke representative. See the “Contact Fluke” section of this manual.

Table 8. User-Replaceable Parts and Accessories

Item NumberDescriptionQtyPart Number
1Case Top Assembly For 72114409796
2Fluke-721 Lens, 72114364743
3Screw, 5-20k.312, Ph, Threaded4494641
4Fluke-7xx, Display Module14404450
5Fluke-721 Keypad14364568
6Conn, Panel Mount, Circ,1942714
7 Insulator, Input Receptacle, Red 1 664257
8 Insulator, Input Receptacle, Black 1 664265
9Battery, AA, Alkaline4376756
10Fluke-721 Beltline Gasket, Black14364546
11Case Bottom Assembly For 72114409801
12Fluke-721-2003,Holster, Yellow143
13Fluke-721 Battery Door, Gray 814364522
14 Screw, 4x5/16, Pan Head 5 642931
15Fluke-721 Foot, Black24364579
Not ShownDecal, Manifold Range,100 Psi,14110705
Not ShownDecal, Manifold Range,300 Psi14110710
Not ShownDecal, Manifold Range,500 Psi14110722
Not ShownDecal, Manifold Range,1000 Psi14110731
Not ShownDecal, Manifold Range,3000 Psi14110746

4364505

Table 9. User-Replaceable Parts and Accessories (cont.)

Item NumberDescriptionQtyNumber
Not ShownDecal, Manifold Range,5000 Psi14110754
Not ShownDecal, Manifold Range 16psi 1.1bar14374274
Not ShownDecal, Manifold Range 36psi 2.5bar14374288
Not ShownDecal, Manifold Range 1500psi 100bar14374301
Not ShownFluke-721,Decal, Rear, Serial Number Plate14364755
Not ShownTie, Cable, 4"1172080
See www.fluke.com for more information about the test leads and alligator clips available for your region.

Part

Accessories

Table 9 is a list of available accessories for the Product.

Table 10. Accessories

Part NumberDescription
4382695Fluke-720URTDA,Universal RTD Adapter, For 719Pro And 721
4380630 721Carrying Case
See www.fluke.com for more information about the test leads and alligator clips available for your region.

Exploded view diagram of a car body panel with numbered components for identification

Figure 2. User-Replaceable Parts
hmq101.eps

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : FLUKE

Model : 721

Category : Uncategorized