Product Type Biological Microscope
Brand Kern
Model OBT 105
Optical System Infinity corrected optical system
Viewing Head Binocular, 30° inclined, 360° rotatable
Eyepieces WF10x/20mm, pair
Objective Lenses 4x, 10x, 40x, 100x (oil immersion) - Plan achromatic
Magnification Range 40x to 1000x
Focusing Coaxial coarse and fine focus, tension adjustable
Stage Double layer mechanical stage, 140x140mm, with low drive controls
Condenser Abbe condenser N.A. 1.25 with iris diaphragm
Illumination LED, adjustable intensity, built-in
Power Supply AC 100-240V, 50/60Hz, via external adapter
Dimensions (W x D x H) Approx. 250 x 350 x 400 mm
Weight Approx. 8 kg
Cleaning Use lens paper with mild cleaning solution; avoid solvents
Safety Features Auto-off after 30 min inactivity, low voltage
Warranty 2 years manufacturer warranty
Accessories Included Dust cover, immersion oil, Allen key
Spare Parts Availability Bulbs, fuses, eyepieces, objectives available from Kern
Repairability Serviceable by authorized technicians; user-replaceable parts: bulb, fuse

Frequently Asked Questions - OBT 105 Kern

How do I change the LED bulb?
Turn off and unplug the microscope. Remove the bulb cover on the back. Gently pull out the old LED module and insert the new one. Replace cover and plug in.
What is the correct way to clean the lenses?
Use lens paper moistened with a drop of lens cleaning solution (or distilled water). Wipe in a circular motion from center out. Never use paper towels or alcohol.
Why can't I see anything through the eyepieces?
Check that the objective lens is clicked into place, the diaphragm is open, and the light intensity is turned up. Ensure the specimen is properly positioned.
How do I adjust the interpupillary distance?
While looking through both eyepieces, move the binocular tubes together or apart until you see a single circular field of view.
What is the maximum magnification?
The maximum total magnification is 1000x using the 100x oil immersion objective and the 10x eyepiece.
How do I use the oil immersion objective?
Focus on the specimen with the 40x objective first. Then rotate the nosepiece halfway between 40x and 100x, place a drop of immersion oil on the slide, and gently rotate the 100x objective into the oil. Use fine focus only.
The image is blurry and won't focus. What should I do?
Check that the slide is not upside down. Clean the lenses. Adjust the condenser height and diaphragm. Use the coarse focus knob first, then fine-tune with the fine focus knob.
How do I replace the fuse?
Unplug the microscope. The fuse holder is near the power socket. Pry it open, remove the old fuse, and insert a new one of the same rating (usually T1A/250V). Push back in.
Can I use batteries?
No, the OBT 105 requires an AC power adapter (100-240V). There is no battery compartment.
What maintenance is needed?
Regularly clean lenses and stage. Keep the microscope covered when not in use. Lubricate mechanical parts only as per service manual. Check for loose screws.

User questions about OBT 105 Kern

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Microscope in PDF format for free! Find your manual OBT 105 - Kern and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. OBT 105 by Kern.

USER MANUAL OBT 105 Kern

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Kern

Model : OBT 105

Category : Microscope