Type of Product Multifunction Power Analyzer
Brand Camille Bauer
Model SIRAX BM1250
Measurement Category True RMS Voltage, Current, Power, Energy, Frequency, Power Factor
Input Voltage Range 0 to 600 V AC (direct), up to 600 V with external VT
Input Current Range 0 to 6 A AC (direct), up to 6000 A with external CT
Accuracy Class 0.5 (energy), 0.2 (power) per IEC 61557-12
Communication Interfaces RS485 Modbus RTU, Ethernet TCP/IP, Pulse output
Display Backlit LCD, 4 lines, 16 characters each
Power Supply 90 to 260 V AC/DC, 50/60 Hz, <5 VA
Dimensions (W x H x D) 96 x 96 x 65 mm (panel cutout: 92 x 92 mm)
Weight Approx. 0.3 kg
Mounting Panel mount, flush with front bezel
Protection Rating IP54 (front), IP20 (rear)
Operating Temperature -10°C to +55°C
Storage Temperature -25°C to +70°C
Humidity Up to 95% non-condensing
Key Functions Measurement of V, A, kW, kVAR, kVA, PF, Hz, kWh, kVARh, demand, min/max
Data Logging Internal memory for 1000+ records; logging interval 1 sec to 60 min
Alarm Outputs 2 relay outputs (SPDT, 5A/250V AC) for limit alarms
Digital Inputs 2 digital inputs for status or tariff control
Harmonics Analysis Up to 31st harmonic (voltage and current)
Software Compatibility Camille Bauer SICAM or third-party via Modbus
Maintenance and Cleaning Wipe with dry cloth; no solvents; periodic calibration recommended every 2 years
Safety Standards IEC 61010-1, IEC 61557-12, UL 61010-1
Spare Parts and Repairability Repair by authorized service; spare parts available; consult manual for fuses

Frequently Asked Questions - SIRAX BM1250 Camille Bauer

How do I install the SIRAX BM1250?
Mount the device in a panel cutout of 92x92 mm, secure with clamping brackets. Connect voltage and current inputs per wiring diagram. Power supply: 90-260 V AC/DC. Refer to the manual for detailed connections.
What is the accuracy of the SIRAX BM1250?
The device offers Class 0.5 for energy measurement and Class 0.2 for power (per IEC 61557-12). It provides true RMS readings for distorted waveforms.
Can I connect the SIRAX BM1250 to a PLC?
Yes, it supports RS485 Modbus RTU and Ethernet TCP/IP for integration with PLCs, SCADA, or building management systems.
How do I reset the energy counters?
Press and hold the SET button for 3 seconds to access the setup menu. Navigate to 'Energy Reset' and confirm. Partial reset for tariff counters also available.
What is the maximum current input?
Direct input up to 6 A. For higher currents, use external current transformers (CT) with secondary 5 A. The device can display scaled values up to 6000 A.
How do I configure the communication parameters?
Enter the setup menu via the front buttons. Set BAUDRATE (c4800 to 115200), PARITY (none/even/odd), STOP BITS, and DEVICE ADDRESS (1-247). Default: 9600, 8, N, 1.
What does the 'ALARM' LED indicate?
The red ALARM LED lights when any measured value exceeds the configured limits. It is associated with relay outputs that can trigger external alarms.
Can I measure harmonics with this device?
Yes, it provides harmonic analysis up to the 31st order for both voltage and current. Data available via Modbus or on the display.
How do I clean the front panel?
Use a soft, dry cloth. Do not use solvents, abrasive cleaners, or water. Ensure the device is powered off when cleaning.
Where can I find the user manual?
The manual is available for free download as a PDF on the official support page or from the retailer. It includes wiring diagrams, Modbus registers, and configuration details.

User questions about SIRAX BM1250 Camille Bauer

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Measuring equipment in PDF format for free! Find your manual SIRAX BM1250 - Camille Bauer and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. SIRAX BM1250 by Camille Bauer.

USER MANUAL SIRAX BM1250 Camille Bauer

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Camille Bauer

Model : SIRAX BM1250

Category : Measuring equipment