Product Type Multifunction Installation Tester
Brand Sonel
Model MPI-536
Dimensions (W x H x D) 260 mm x 170 mm x 100 mm
Weight 1.5 kg
Power Supply Rechargeable NiMH battery pack (4.8 V / 4.5 Ah) or mains adapter
Display LCD with backlight
Measurement Functions Insulation resistance, continuity, RCD (AC/A/SI) testing, loop impedance, phase sequence, voltage, frequency
Insulation Resistance Range 0.01 MΩ to 10 GΩ (test voltages 50 V, 100 V, 250 V, 500 V, 1000 V)
Continuity Testing 0.01 Ω to 10 kΩ with 200 mA test current
RCD Testing Type AC, A, SI, F, B; rated currents 10/30/100/300/500/1000 mA; trip time and trip current
Loop Impedance 0.01 Ω to 10 kΩ (L-N, L-L, L-PE) with 2-wire and 3-wire methods
Voltage Measurement 0 V to 600 V AC/DC, TRMS
Safety Rating CAT IV 300 V, CAT III 600 V
Ingress Protection IP54 (dust and splash proof)
Memory Up to 1000 measurement records
Communication USB 2.0 (data transfer to PC)
Standards Compliance IEC 61557, IEC 61010
Operating Temperature -10 °C to +40 °C
Storage Temperature -20 °C to +60 °C
Accessories Included Test leads, alligator clips, mains supply adapter, USB cable, carrying case, user manual
Maintenance Clean with a soft damp cloth; calibration recommended annually
Repairability Spare parts available from Sonel service centers; fixable by qualified personnel

Frequently Asked Questions - MPI-536 Sonel

How do I change the battery in the Sonel MPI-536?
Turn off the device and disconnect all test leads. Remove the battery compartment cover on the back. Replace the NiMH battery pack with a compatible one (4.8 V / 4.5 Ah). Reattach the cover securely.
What does the 'ERR' message mean on the display?
The 'ERR' message indicates a measurement error. Possible causes: test leads not connected properly, voltage present on the tested circuit, or the battery is low. Check connections and ensure the circuit is de-energized before retesting.
Can I test RCDs with trip current of 30 mA?
Yes, the Sonel MPI-536 supports RCD testing for rated currents from 10 mA to 1000 mA, including 30 mA. Select the appropriate RCD type (AC, A, SI, etc.) and test mode (trip time or trip current).
How do I calibrate the MPI-536?
Calibration should be performed by an authorized service center at least once a year. The device has no user-adjustable calibration. Contact Sonel support for service.
What is the maximum insulation resistance it can measure?
The insulation resistance measurement range goes up to 10 GΩ (gigaohms) at test voltages up to 1000 V. For lower voltages, the maximum range may be lower.
How do I transfer data to a PC?
Connect the device to your PC via the included USB cable. Use the Sonel Reader software (downloadable from the Sonel website) to download and analyze measurement records. Ensure the device is turned on.
The device does not turn on. What should I check?
First, check if the battery is charged. Connect the mains adapter and try turning on. If still not working, the battery may need replacement or there is an internal fault. Contact Sonel support.
What is the difference between 2-wire and 3-wire loop impedance testing?
The 2-wire method measures L-PE loop impedance using two test leads. The 3-wire method uses an additional reference lead to improve accuracy, especially in noisy environments. Both are supported by the MPI-536.
Is the MPI-536 safe to use on live circuits?
Yes, it is designed for use on live installations up to CAT IV 300 V / CAT III 600 V. Always follow safety rules: wear appropriate PPE, use insulated test leads, and ensure the device is rated for the measured voltage.
How do I clean the device?
Use a soft, slightly damp cloth with mild detergent if needed. Do not use abrasive cleaners or solvents. Avoid getting moisture inside the device. Dry thoroughly before use.

User questions about MPI-536 Sonel

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Measuring equipment in PDF format for free! Find your manual MPI-536 - Sonel and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. MPI-536 by Sonel.

USER MANUAL MPI-536 Sonel

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Sonel

Model : MPI-536

Category : Measuring equipment