Digalox DPM72-MPN+-DIN - Energy meter TDE Instruments - Free user manual and instructions
Find the device manual for free Digalox DPM72-MPN+-DIN TDE Instruments in PDF.
Pick your language and provide your email: we'll send you a specifically translated version.
| Product Type | Energy Meter (Panel Mount) |
| Brand | TDE Instruments |
| Model | Digalox DPM72-MPN+-DIN |
| Dimensions (WxHxD) | 72 x 72 x 65 mm (panel cutout 68x68 mm) |
| Weight | Approx. 150 g |
| Power Supply | 230 V AC, 50/60 Hz (or optional DC) |
| Display | LCD, multi-segment, backlit |
| Measured Parameters | Voltage (L-N, L-L), Current, Active Power, Reactive Power, Apparent Power, Power Factor, Frequency, Active Energy (kWh), Reactive Energy (kVArh) |
| Accuracy Class | Class 1 (for active energy) per IEC 62053-21 |
| Input Voltage Range | Up to 500 V AC (L-L) via direct connection or PT |
| Input Current Range | Up to 5 A AC (via external CTs, primary up to 9999 A) |
| Communication | Modbus RTU (RS485), pulse output |
| Operating Temperature | -10°C to +55°C |
| Storage Temperature | -25°C to +70°C |
| Protection Rating | IP40 (front), IP20 (rear) |
| Mounting | Panel mount, DIN rail adapter included |
| Electrical Safety | Double insulation, CAT III |
| Standards | IEC 61010-1, IEC 61326, IEC 62053-21 |
| Maintenance | Clean with dry cloth; no user-serviceable parts inside |
| Spare Parts | Contact TDE Instruments for replacement connectors, fuses |
| Reparability | Repair by authorized service center only |
Frequently Asked Questions - Digalox DPM72-MPN+-DIN TDE Instruments
What is the Digalox DPM72-MPN+-DIN used for?
It is a multifunction energy meter for measuring electrical parameters like voltage, current, power, and energy in industrial or commercial installations.
How do I install the meter?
Mount it in a 68x68 mm panel cutout, secure with the provided brackets, and connect wires per the wiring diagram in the manual. Use external current transformers (CTs) as required.
What is the power supply requirement?
The meter requires 230 V AC (50/60 Hz) for the auxiliary supply. Alternatively, a DC option may be available; check the model label.
Can I connect it directly to high voltage?
The voltage inputs accept up to 500 V AC directly (L-L). For higher voltages, use potential transformers (PTs). Always follow local safety regulations.
How do I read the energy consumption?
The display shows cumulative active energy (kWh). Press the button to scroll through other parameters like reactive energy, instantaneous power, etc.
What communication protocols does it support?
It features Modbus RTU over RS485 for remote monitoring and a pulse output for energy pulsing. Baud rate and address are configurable.
How do I reset the energy counters?
To reset partial counters, hold the button for 5 seconds. Full cumulative counters cannot be reset; they are factory-calibrated.
What should I do if the display is blank?
Check the auxiliary power supply (L/N terminals) and ensure the voltage is present. If still blank, measure the backup fuse (if any). Contact support if needed.
How do I clean the meter?
Wipe the front panel with a dry, soft cloth. Do not use solvents or water. Keep ventilation slots unobstructed.
Where can I find spare parts or accessories?
Contact TDE Instruments or an authorized distributor for replacement terminal covers, fuses, CTs, or the optional Modbus adapter.
User questions about Digalox DPM72-MPN+-DIN TDE Instruments
0 question about this device. Answer the ones you know or ask your own.
Ask a new question about this device
No questions yet. Be the first to ask one.
Download the instructions for your Energy meter in PDF format for free! Find your manual Digalox DPM72-MPN+-DIN - TDE Instruments and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. Digalox DPM72-MPN+-DIN by TDE Instruments.