Uni-T UT325 - Environment thermometer

UT325 - Environment thermometer Uni-T - Free user manual and instructions

Find the device manual for free UT325 Uni-T in PDF.

📄 3 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice Uni-T UT325 - page 1
Pick your language and provide your email: we'll send you a specifically translated version.
Type of product Environment thermometer
Brand Uni-T
Model UT325
Display LCD with backlight
Measurement range (K-type) -50°C to 1300°C (-58°F to 2372°F)
Accuracy ±1.5% of reading or ±1°C (whichever greater)
Resolution 0.1°C / 0.1°F
Number of channels 2 (dual-channel)
Probe type K-type thermocouple (included)
Functions Data hold, max/min recording, auto power off, °C/°F switch
Power supply 9V battery (6F22 or equivalent)
Battery life Approximately 200 hours
Dimensions 180 x 72 x 32 mm
Weight 200 g (including battery)
Operating temperature 0°C to 50°C (32°F to 122°F)
Storage temperature -20°C to 60°C (-4°F to 140°F)
Cleaning Wipe with a soft dry cloth; do not use solvents
Safety precautions Do not measure live circuits; use only K-type probes within rated temperature
Accessories included 2x K-type thermocouple probes, 9V battery, user manual
Repairability User replaceable battery; probes can be replaced

Frequently Asked Questions - UT325 Uni-T

What type of battery does the UT325 use?
The UT325 requires a standard 9V battery (6F22 or equivalent). Replace the battery when the low battery indicator appears.
Can I use other thermocouple types besides K-type?
No, the UT325 is designed specifically for K-type thermocouples. Using other types may result in inaccurate readings or damage.
What is the maximum temperature the UT325 can measure?
With a K-type thermocouple, the UT325 can measure up to 1300°C (2372°F). Ensure the probe is rated for the temperature.
How do I switch between Celsius and Fahrenheit?
Press the '°C/°F' button on the meter to toggle between Celsius and Fahrenheit units.
What does 'Data Hold' do?
The 'Data Hold' function freezes the current reading on the display, allowing you to record it conveniently. Press the 'HOLD' button to activate.
How accurate is the UT325?
Accuracy is ±1.5% of reading or ±1°C (whichever greater) for temperatures between -50°C and 1300°C. For best accuracy, allow the meter to stabilize.
Why does the display show 'OL'?
'OL' indicates an over-range condition, meaning the measured temperature is outside the meter's range or the probe is not connected properly.
Can I measure the temperature of a live electrical circuit?
No, the UT325 is not designed for measuring live circuits. Always ensure the circuit is de-energized before using the thermocouple.
How do I clean the UT325?
Wipe the exterior with a soft, dry cloth. Do not use harsh chemicals or solvents, as they may damage the casing.
What does the 'MAX/MIN' function do?
The 'MAX/MIN' function records and displays the highest and lowest temperatures measured since the function was activated. Press 'MAX/MIN' to cycle through readings.

User questions about UT325 Uni-T

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Environment thermometer in PDF format for free! Find your manual UT325 - Uni-T and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. UT325 by Uni-T.

USER MANUAL UT325 Uni-T

Digital Thermometers

Overview

The Uni-Trend Model UT321, UT322, UT323 and UT325 Thermometers (the thermometer) are microprocessor-based, digital thermometers designed to use external J-, K-, T-, E-, R-, S- and N-type thermocouples (temperature probes) as temperature sensors.

  • UT321: single input(T1), suitable for K-, J-, T- and E- type thermocouples
  • UT322: dual input(T1,T2), suitable for K-, J-, T- and E-type thermocouples
  • U1323: single input[11], suitable for K-, J-, T-, E-, R-, S- and N-type thermocouples. Equipped with over-limit alarm, over-limit signal output and user self-debug
    Equipped with over-limit alarm, over-limit signal output and user con developing features.
  • UT325: dual input(T1,T2), suitable for K-, J-, T-, E-, R-, S- and N-type thermocoupling
    Thermocouples. Equipped with over-limit alarm, over-limit signal output and user self debris features

Safety Information

Use the thermometer only as specified in this manual. Otherwise, the protection provided by the meter may be impaired.

Refer to safety information in Table 1 and the international symbols in Table 2.

Table 1. Safety Information

Warning

A warning identified conditions and actions that pose hazards to the user. To avoid electrical shock or personal injury, follow these guidelines:
- Before using the thermometer inspect the case. Do not use the thermometer if it appears damaged. Look for cracks or missing plastic. Pay particular attention to the insulation around the connectors.
- Disconnect the thermocouple(s) from the thermometer before opening the case. - Replace the batteries as soon as the battery indicator (☐) appears. The possibility of false readings can lead to personal injury.
- Do not use the thermometer if it operates abnormally. Protection may be
impaired. When in doubt, have the thermometer serviced.
- Do not operate the thermometer around explosive gas, vapor, or dust. - Do not apply more than the rated voltage, as marked on the thermometer (300.0
- DO not apply more than the failed voltage, as marked on the thermometer (30V) between the thermocouple(s), or between any thermocouple and earth ground.
- When potential differences are anticipated between the thermocouples, use
electrically insulated thermocouples.
- When servicing the thermometer, use only specified replacement parts.

Caution

A caution Identifies conditions and actions that may damage the meter or the equipment under test.

  • Use the proper thermocouples, function, and the range of your thermometer.
  • When carrying two thermocouples measurement, make sure there are no potential differences between two thermocouples.
  • Do not attempt to recharge the battery.
  • Match the + and - polarities of the battery with the battery case.

Table 2. International Symbols

Refer to the manual for information about this feature
CECompiles with European Union directives
Battery

Get Acquainted with Your Thermometer

Unless otherwise specified, all the descriptions apply to all Models of UT321, UT322, UT323 and UT325.

To become familiar with the thermometer, study the following
● Figure 1 and Table 3 for components;
● Figure 2 and Table 4 for LCD display:
● Figure 3-a and 3-b and Table 5 for button functions;

Components
Technical diagram of a digital thermometer with labeled components and measurement ranges

Figure 1. Components

Table 3. Components

1Display
2Buttons
3Thermocouple T1 input
4Thermocouple T2 input (UT323 and UT325 only)
5USB Port
6SIGN Port - Over Limit Signal Output (UT323 and UT325 only)
7NTC
8Battery Door
99V battery (8F22 )

LCD Display

SETUP MAX MIN AVG USB RCL T1-T2 V HOLD TYPE K J T E R S N T1 T2 DATA OFFSET DEBUG INTERVAL T2 T1 88.88 d:hh:mm:ss 0 1 2 3 4 5

Figure 2. Display LCD Display

Table 4. LCD Display

1Setup is in progress when the icon blinks
2Display readings of maximum, minimum and average
3Data Transferring is in progress
4Logged readings are displayed when the icon blinks
5Low battery indicator. Replace the battery
6The thermocouple type
7The temperature units
8Secondary Display 2
9Under Calibration mode when the icon blinks. The displayed reading is fixed
10Secondary Display 1
11The thermocouple measurement includes an offset. See "Changing Setup Options"
12Readings are being logged when the icon blinks
13Primary Display
14Data hold function is active.
15The symbol of display backlight

Buttons

① HOLD MAXMIN STORE TYPE RECALL ENTER CFK SEND SETUP

Figure 3-a UT321 and UT323 Keypad

HOLD MAX MIN T1 T1-T2 SEND STORE RECALL SETUP ENTER

Figure 3-b UT322 and UT325 Keypad

Table 5. Buttons

1Press to turn the thermometer on or ofT
Press to turn the display backlight on and off.
HOLDPress to freeze or unfreeze the displayed readings
°C OFKPress to switch between Celsius (°C), Fahrenheit(°F), and Kelvin (K)
MAX MINPress to step through the maximum, minimum, and average readings. Press and hold to exit this function.
TYPE(UT320UT325)Press to step through K-, J-, T-, E- (R- S- N-) type thermocouple.
3113-37(UT320UT325)Press to select and show T1, T2 and T1-T2 (differential temperature measurement) in the primary or secondary display 1
SENDPress to enter USB mode and the USB icon blinks. Press again to exit USB mode.
STOREPress to start or stop logging(See "Using Memory - Starting and Stepping Logging".)
RECALLPress to show logged readingsPress again to stop.
SETUPPress to start or exit Setup.Press to scroll the Setup option you want to change(See "Changing Setup Options")
After entering the Setup mode, press to increase the displayed setting.(See "Changing Setup Options")
After entering the Setup mode, press to decrease the displayed setting.(See "Changing Setup Options")
ENTERConfirm button.(See "Changing Setup Options")

Using the Thermometer

1) Plug the thermocouple(s) into the input connector(s).
2) Press the power button to turn on the thermometer.
3) Set the type of thermocouple(s) to be the same with the one plugged into the input connector(s).

If no thermocouple is plugged into the selected input terminal or the thermocouple is "open" or the over-range positive deviation is too big, the display shows".

Changing Setup Options

- Use Setup to change the following settings:

- Logging interval→Thermocouple type→Offset (T1) Offset (T2) (UT322 and UT325 only) Sleep Mode Time Frequency Time (S-T) Low Limit Alarm (Lo) (UT323 and UT325 only)→High Limit Alarm (Hi) (UT323 and UT325 only)→Over Limit Signal Output (SI) ON/OFF→UT323 and UT325 only) Normal Temperature Compensation (NTC) ON/OFF→DEBUG ON/OFF (UT323 and UT325 only) have setting and return to normal measurement mode.

Entering and Exiting Setup

When the thermometer is in Setup mode, the display shows and blinks SETUP

- Press SETUP to start or exit Setup

- Press SETUP to scroll through the Setup options you want to change.

Changing the Logging Interval

The logging interval determines how often the thermocouple stores logged readings in memory. You choose the length of the logging interval. See "Using Memory".

Press SETUP until the display shows INTERVAL

The thermometer stores logged readings at the end of each logging interval. You can select a logging interval by pressing ▲or▼. Holding down ▲or▼ causes the number to change more quickly. The maximum interval is 59:59 and the minimum interval is 00:00. When the logging interval is 00:00, you need to store the readings manually as the auto store feature will be disabled.

Changing the Thermocouple Type

  1. Press SETUP until the display shows TYPE.

  2. Press ▲ or ▼ to select the thermocouple type you want including K-, J-, E- type (UT323 and UT325 have extra R-, S- and N- type, UT321/UT325 can use TYPE button to change directly)).

Changing the Offset (T1)

  1. Press SETUP until the display shows OFFSET and T1

  2. Press ▲ or ▼ to change the offset value. The offset range is -6\~6.

Changing the Offset (T2)

  1. Press SETUP until the display shows OFFSET and T2

  2. Press ▲ or ▼ to change the offset value. The offset range is -6\~6.

Sleep Mode

  1. Press SETUP until the display shows SLP.
  2. Press ▲ or ▼ to change until the desired sleep lime is obtained, it is between 5–60 minutes.
  3. Holding down ▲ or ▼ causes the time to change more quickly.
  4. When the sleep mode time is less than 5 minutes, the display shows OFF which means disabling the sleep mode.

Changing the Line Frequency

  1. Press SETUP until the display shows LINE.

  2. Press ▲ or ▼ to select 60Hz or 50Hz

Setting the Time

  1. Press SETUP until the display shows S-T
  2. Press ENTER to toggle between time format 'h:m' and 'm:s'.
    3 Press ▲ or ▼ until the display shows the correct time. The time is 24-hour format.
  3. Holding down ▲ or ▼ causes the number to change more quickly.
  4. If you don't change this option, it means the time that the meter has run through currently.
  5. The system counts the time when the user turns on the thermometer. It will be automatically cleared once the Thermometer is dis-energized.

Setting the Low Limit Alarm (Lo) ((Only UT323 and UT325)

  1. Press SETUP until the display shows Lo.
  2. Press ▲ or ▼ until the display shows the correct low limit alarm.
  3. Holding down ▲ or ▼ causes the number to change more quickly
  4. Press ENTER to turn the low limit alarm off, the display shows OFF. Press ENTER again to turn the low limit alarm on, the display shows the low limit alarm reading
  5. When the tested temperature is over the limit, the thermometer beeps
  6. The Min.value for low alarm is the lower limit value of measuring range of the applied thermocouple.
  7. The Max. value for low alarm is the setting high alarm limit value if high alarm is enabled (if still higher than high alarm limit value, Max. value=High Alarm Limit Value-1), otherwise this Max. value is the upper limit value of measuring range of the applied thermocouple.

For example:

The Min. value of the low alarm of K type is -200°C.
The Max. value of the low alarm of K type is either High Alarm Limit Value-1 or +1372°C

The table below shows the measuring range of each type of thermocouple

K type: -200.0°C to +1372°C (-328.0°F to +2501°F)
J type: -210.0°C to +1200°C (-346.0°F to +2192°F)
T type: -250.0°C to +400.0°C (-418.0°F to +752.0°F)
E type: -150.0°C to +1000°C (-236.0°F to +1832°F)

Setting the High Limit Alarm (HI) (Only UT323 and UT325)

  1. Press SETUP until the display shows HI.

  2. Press ▲ or ▼ until the display shows the correct high limit alarm.

  3. Holding down ▲ or ▼ causes the number to change more quickly

  4. Press ENTER to turn the high limit alarm off, the display shows OFF. Press ENTER again to turn the high limit alarm on, the display shows the high limit alarm reading.

  5. When the tested temperature is over the limit, the thermometer beeps

  6. The Max.value for high alarm is the upper limit value of measuring range of the applied thermocouple.

  7. The Min. value for high alarm is the setting low alarm limit value if low alarm is enabled (if still less than the low alarm limit value, Min. Value=Low Alarm Limit Value+1), otherwise this Min. value is the lower limit value of measuring range of the applied thermocouple.

For example:

The Max. value of the high alarm of K type is +1372°C

The Min.value of the high alarm of K type is either Low Alarm Limit Value+1 or -200°C.

The table below shows the measuring range of each type of thermocouple

K type: -200.0°C to +1372°C (-328.0°F to +2501°F)
J type: -210.0°C to +1200°C (-346.0°F to +2192°F)
T type: -250.0°C to +400.0°C (-418.0°F to +752.0°F)
E type: -150.0°C to +1000°C (-238.0°F to +1832°F)

Enabling or Disabling Over Limit Signal Output (Only UT323 and UT325)

  1. Press SETUP until the display shows SI.

  2. Press ▲ or ▼ to turn the over-limit signal output on or off. The display shows on or OFF.

  3. With High/Low Limit Alarm set up and Over Limit Signal Output turned on already, the SIGN port of the thermometer will output the corresponding signal if the tested temperature is over or under the high or low limit. When the tested temperature is over the high limit, the SIGN port will output the high electric level signal. When the tested temperature is under the low limit, the SIGN port will output around 10Hz pulse signal.

Enabling or Disabling Normal Temperature Compensation (NTC)

  1. Press SETUP until the display shows NTC

  2. Press ▲ or ▼ to turn normal temperature compensation on or off. The display shows On or Off. It defaults to ON status at power-on.

Enabling or Disabling Debug (Only UT323 and UT325)

  1. Press SETUP until the display shows DEBUG.

  2. Press ▲ or ▼ to turn Debug on or off. The display shows on or off. The thermometer defaults to OFF status.

  3. You can self-debug the thermometer when the Debug feature is turned on. (Refer to Measuring Temperatures - Debug)

Measuring Temperatures

Connecting a Thermocouple

Thermocouples are color-coded based on the North American ANSI Colour Code:

TypeColourTypeColour
JBlackRGreen
KYellowSGreen
TBlueNOrange
EPurple
  1. Plug a thermocouple into the input connector(s).

  2. Set the same thermocouple type in the thermometer.

To change the thermocouple type, see "Changing Setup Options".

Displaying Temperatures

  1. Press °C °FK to select the correct temperature unit.

  2. Hold or attach the thermocouple(s) to the measured location.

The temperature shows in the selected unit on LCD.

Holding the Displayed Temperatures

  1. Press HOLD to freeze the readings on the display. The display shows HOLD

  2. Press HOLD again to turn off the HOLD function.

Turning on and off display backlight

  1. Under temperature measurement mode, press to turn the display backlight on.

  2. Press again to turn off the display backlight. The display backlight will not turn automatically off.

Viewing the MIN, MAX, and AVG Readings

  1. Press MIN MAX to step through the maximum (MAX), minimum (MIN), or the average (AVG) readings.

  2. Press and hold MIN MAX to exit MIN MAX mode.

Using the Offset to Adjust for Probe Errors

Use the offset option in Setup to adjust the thermometer's readings to compensate for the errors of a specific thermocouple.

  1. Plug the thermocouple into the input connector.

  2. Place the thermocouple in a known, stable temperature environment (such

as an ice bath or a dry well calibrator).

  1. Allow the readings to stabilize.

  2. In Setup change the offset until the display reading matches the debug temperature. (See "Changing Setup Options.")

Over Limit Alarm (Only UT323 and UT325)

Setting the low alarm and high alarm limit at the SETUP mode, the thermometer beeps when the measured temperature is higher or lower than the pre-set limit. (See "Changing Setup Options.")

Over Limit Signal Output (Only UT323 and UT325)

Setting the low and high limit signal output at the SETUP mode, the SIGN port output the corresponding signal when the measured temperature is higher or lower than the pre-set limit. (See "Changing Setup Options.")

Debug (Only UT323 and UT325)

After entering the Debug mode, you can self-debug. Follow the procedures as below:

- Debug point:

K- and J-type: -180°C, 0°C and 900°C

T type: -180°C, 0°C and 400°C

E type: -140°C, 0°C and 900°C

R- and S- type: 0°C and 1200°C

N type: -160°C, 0°C and 1200°C

Remark: 0°C is only for debugging K type.

- Debug method:

  1. Connect T1 and T2 to standard voltage source.

  2. Standard voltage source sources the voltage corresponding to

debug point.

  1. Press HOLD to freeze the debug data.

  2. Normal temperature debug:

Press MAX MIN when the thermometer is at Debug mode, secondary

display 1 shows the offset reading

Secondary display 2 shows the normal temperature

Press ▲ or ▼ to adjust the offset reading, the offset range is -6\~+6. Remark:

For Temperature VS Voltage table, refer to BS EN 60584 - 1:1996. Before carrying out the debug, please ensure your standard source has the accuracy up to the demand. Our company will not be liable for massive deviation occurring to the accuracy of the thermometer that has been caused by improper selection of standard source on users' side.

Using Memory

During a logging session, the thermometer stores logged readings in its memory.

The thermometer stores 00-99, total 100 sets of temperature readings. The stored readings are from primary display.

Starting and Stopping Logging

  1. Set the logging interval. (See "Changing Setup Options.")

  2. Press STORE to start logging.

DATA shows and blinks on LCD

  1. Press STORE again to stop logging.

  2. If you selected a manual logging interval, set the logging interval as 00:00. Press ENTER each time you want to store logged readings in memory. The secondary display 1 shows the logged reading. Each presses of ENTER will automatically store the logged readings in the next memory location. Press ▲ or ▼ would change the memory location, the empty location shows "----".

  3. If you selected an auto logging interval, set the logging interval to any time except 00:00. After entering logging mode, the secondary display 2 shows "." Press ENTER to start storing readings in sequence according to the pre-set logging interval, the "-" blinks. Press ENTER again to pause auto logging, the "-" stop blinking. Press ENTER again to resume logging, the "-" blinks again.

Viewing Logged Readings

  1. Press RECALL to view logged readings

RCL shows and blinks on LCD

  1. Press ▲ or ▼ to scroll through the logged readings, the default reading is the last stored reading prior to exiting the logging. The temperature unit will switch automatically along with the recalled data based on what have been stored during the logging.

  2. The secondary display 1 shows its memory location. The secondary display 2 shows the system time when the data was logged. The primary display shows each logged reading.

  3. Press ENTER causes the readings to change more quickly until memory location reaches 00.

Clearing Memory

  1. Press STORE to enter logging mode, DATA shows and blinks on LCD.

  2. Press and hold down STORE for 2 seconds, the display shows CLR.

  3. Press ENTER to delete all the logged readings from memory. It is not possible to delete the logged readings one by one.

The display then shows *...*

Communicating with a PC

You can transfer the data from the thermometer's memory to a PC using the included Software.

Press SEND button and USB blinks on the display. It means the Thermometer has already been communicated normally to the PC. Details refer to the Installation Guide file in the Software.

Under the influence of radiated Radio-Frequency electromagnetic Field

the captioned models have a measurement error, it will be back to normal when the interference is removed.

Maintenance

Replacing the Battery

Replace the batteries as soon as the battery indicator appears. The possibility of false readings can lead to personal injury.

  1. Turn off the thermometer.

  2. Loosen the screw and remove the battery door.

  3. Replace the battery with a new 9V battery (6F22). Reinstall the battery door and tighten the screw.

Cleaning the Case

Clean the surface of the thermometer with soft cloth or sponge dampened with water or mild detergent. Do not get water inside the thermometer, otherwise it may cause damage to internal circuits or the instrument.

Specifications

Environmental

Operating Temperature-10°C to 50°C(14°F to 122°F)
Storage Temperature-40°C to +60°C(-14°F to +140°F)
HumidityNon condensing <10°C (<50°F)95% RH: 10°C to 30°C (50°F to 86°F)75% RH: 30°C to 40°C (86°F to 104°F)45% RH: 40°C to 50°C (104°F to 122°F)

General

CertificationCE
SafetyEN61326: 2006, EN55022: 2006 and EM55024: 1998+A1+A2

Electrical

FunctionUT321UT322UT323UT325Input Protection
TypeK, J, T, EK, J, T, E, R, S, N30V
InputT1T1, T2T1T1, T2
Measurement RangeK type: -200.0°C to +1372°C (-328.0°F to +2501°F)
J type: -210.0°C to +1200°C (-346.0°F to +2192°F)
T type: -250°C to +400.0°C (-418°F to +752.0°F)
E type: -150.0°C to +1000°C (-238.0°F to +1832°F)
Display Resolution0.1°C °F/K.... (<1000) (T type below -200°C and R type, S type is 1°C °F/K)1.0°C °F/K.... (>1000)
Measurement Accuracy±(0.5%+0.8°C(1.6°F))K-, J-, T-, E- type: ±(0.2%+0.6°C(1.2°F))
R, S type: ±(0.2%+2°C(4°F))N type: ±(0.2%+1.5°C(3°F))
Below -10°C; add 0.5°C Below -200°C, add 12°C Type T Below -200°C: for reference only.
Sampling Rate50 times per second, Updates 2-3 times per second
TimeRelative Time
Data Store0 ~ 99 sets
SetupLogging Interval, Thermocouple Type, Offset (T1), Offset (T2) (UT322 and UT325 only), Sleep Mode, Line Frequency, Time, High/Low Limit (UT323 and UT325 only), Over Limit Signal Output (UT323 and UT325 only), Normal Temperature Compensation, Debug (UT323 and UT325 only)
Over Limit AlarmN/AYes
Over Limit SignalN/AYes
PowerOne piece of 9V (NEDA1804 or 6F22 or 006P)

Accuracy is specified for ambient temperatures between 18°C(60°F) and 28°C

(82°F) for a period of 1 year, operating temperatures: -10°C \~ 50°C (14°F \~

122°F). The above specifications do not include thermocouple probe error.

*END *

This operating manual is subject to change without notice.

UNI-T

UNI-TREND TECHNOLOGY (CHINA) CO., LTD.

No6. Gong Ye Bei 1st Road.

Songshan Lake National High-Tech Industrial

Development Zone, Dongguan City,

Guangdong Province, Ch

Tel: (86-769) 8572 3888

http://www.uni-trend.com

说明书菲林做货要求:

Manual assistant
Powered by Anthropic
Waiting for your message
Product information

Brand : Uni-T

Model : UT325

Category : Environment thermometer